Battery charger

The battery charger is fit for operation at an electric


voltage of 90 V to 264 V. It can be equipped with the feed
cable corresponding to the requirements on site. Green, yellow and orange-coloured LED

• In order to charge the batteries, move the operating


table out of the operating area.

h The batteries cannot be charged as long as the


main switch is switched ON.

h The charging current is max. 2 A. Make sure


that the correct fuse (3.15 A, slow-blow type)
has been inserted into the fuse holder F2. This
is especially important for the more ancient
tables of the OPX mobilis series.

• Switch the main switch on the control panel OFF.

The charging cable of the charging unit is connected to


the operating table by means of a plug-type connector.
You will find the socket for the charging unit on the control
panel at the left hand side of the floor pan.
A 3-coloured LED at the charging unit serves as a mode
The plug and the charging socket are marked by dots. indicator.
The plug will lock inside the charging socket if it is
plugged into the charging socket with the dots in
alignment. Charging diagram

• Plug the plug of the battery charger into the charging Charge voltage

socket so that the marking dots are in alignment and


press the plug in until it locks inside the charging Voltage (V)
socket. Quick Charge Compensation

h
charging LED=yellow charge
LED=orange- LED=green
Electric current coloured Charge
The control panel is equipped with a covering (A) current

sheet, which is designed to protect the control


electronics against electrostatic discharge Capacity
current.

• Connect the battery charger to the mains power


supply.

Operating table OPX mobilis


If the batteries are drained, it will take approx. 8 to
10 hours to recharge them. The batteries are charged as
soon as the LED on is green. The battery charger can now
be disconnected.

• In order to disconnect the charging cable, pull the


front section of the plug back and disconnect the plug
from the socket.

The table-top can now be adjusted again by means of the


electro-hydraulic pump.

h With fully charged, practically new batteries, the


operating table can carry out all its functions
approx. 50 times when charged with a patient’s
weight.