Documente Academic
Documente Profesional
Documente Cultură
Measurement
Standards & Units
Chapter 1
INTRODUCTION
Measurement is the process of determining the amount, degree or
capacity by comparison with the accepted standards of the system
units being used.
Absolute instruments
The quantity to be measured in terms of the deflection and
physical constant of the instrument.
e.g tangent galvanometer.
No calibration needed
Secondary instruments
The quantity to be measured will be indicated in term of the
deflection of the pointer which indicates the value on the
calibrated dial from the observation of the output from
instrument.
Without calibration of such an instrument, the deflection is
meaningless.
(i)Indicating (ii)Integrating (iii) Recording
Different types of SECONDARY INSTRUMENTS :
Indicating
- visualize the process/operation
The quantity measured is indicated on a graduated scale
e.g. ammeter, voltmeter and wattmeter.
Recording
- observe and save the measurement reading
The variation in the quantity to be measured during a specified period is
recorded continuously on a graph.
e.g. recording voltmeters used in supply station.
Integrating
- measure and register the total quality of electricity (in ampere-hour) or
the total amount of electrical energy (in-watt-hours) supplied to circuit
over a period of time.
e.g. empere-hour meters, energy meters.
Effect used in the secondary Instruments
Null
Require two input – measured and balance input.
Must have feedback operation that compare the measured with
standard value.
More accurate and sensitive compared to deflection type
instrument.
Signal Type
Analog
Produce the signal that vary in continuous way.
Infinite range of value in any given range.
Digital
Produce the signal that vary in discrete steps.
Finite different values in a given range.
Advantages of Electronic Measurement
Static Characteristic
2) Primary Standard
Maintained at the National Standard Lab (different for every
country)
Function: the calibration and verification of secondary Standard
Each lab has its own secondary Standard which are periodically
checked and certified by the National Standard Lab.
For example, in Malaysia, this function is carried out by SIRIM.
3) Secondary Standard
Secondary standards are basic reference standards
used by measurement and calibration laboratories in
industries.
Each industry has its own secondary standard.
Each laboratory periodically sends its secondary
standard to the National standards laboratory for
calibration and comparison against the primary
standard.
After comparison and calibration, the National
Standards Laboratory returns the secondary standards
to particular industrial laboratory with a certification of
measuring accuracy in terms of a primary standard.
4) Working Standard
Used to check and calibrate lab instrument for
accuracy and performance.
For example, manufacturers of electronic components
such as capacitors, resistors and many more use a
standard called a working standard for checking the
component values being manufactured.
STANDARD OF MEASUREMENT
SI UNIT
SI meas International System Unit and
mandated by ISO. These SI units are:
ERROR IN MEASUREMENT
Measurement Errors
Xn – measured value
% error = x 100
The precision of a measurement is a quantitative or numerical
indication of the closeness with which a repeated set of
measurement of the same variable agree with the average set of
measurements.
Example 1
Example
Given a 600 V voltmeter with accuracy ±2% full scale.
Calculate limiting error when the instrument is used to
measure a voltage of 250V?
Solution
The magnitude of limiting error, 0.02 x 600 = 12V
Therefore, the limiting error for 250V = 12/250 x 100 = 4.8%
Cont’d…
Example
Given for certain measurement, a limiting error for voltmeter at
70V is 2.143% and a limiting error for ammeter at 80mA is
2.813%. Determine the limiting error of the power.
Solution
The limiting error for the power
= 2.143% + 2.813%
= 4.956%
Exercise
i.If the voltmeter reads 200V on 500V range, what is the absolute
error?
ii.What is the percentage error of the reading in (i).
Rules regarding significant
figures in calculation
1)For adding and subtraction, all figures in columns to the right of the
last column in which all figures are significant should be dropped
Example
V1 = 6.31 V
+ V2 = 8.736 V
Example
From the value given below, calculate the value for R1, R2
and power for R1?
I = 0.0148 A ===> 3 s.f
V1 = 6.31 V ===> 3 s.f
V2 = 8.736 V ===> 4 s.f
eg: an observer who tend to hold his head too far to the left
while reading the position of the needle on the scale.
3. Random Error
Transducers
Device that converts a change in physical quantity into a
change of electrical signal magnitude.
Power Supply
Provide energy to drive the transducers.
Signal Conditioning Circuits
Electronic circuits that manipulate, convert the output from
transducers into more usable electrical signal.
Amplifiers
Amplify low voltage signal from transducers or signal
conditional circuit.
Cont’d…
Recorders
Used to display the measurement for easy reading and
interpretation.
Data Processors
Can be a microprocessor or microcontroller.
Process Controllers
Used to monitor and adjust any quantity of the specified level
or value.
Command Generator
Provide control voltage that represents the difference of the
parameter in a given process.
INSTRUMENT APPLICATION GUIDE