Documente Academic
Documente Profesional
Documente Cultură
Automated 5-part
Hematology Analyzer
Thank you for choosing the Norma Icon5 automated hematology analyzer.
We believe that this instrument will definitely help you in your daily laboratory
routine.
Symbol Meaning
WARNING
Risk of personal injury
BIOHAZARD
Risk of biological infection, contamination
LASER RADIATION
Risk of exposure to laser radiation
Intended Use
Icon5 is a 60 tests/hour laser and impedance based IVD hematology analyzer
for laboratory use, using specific reagents, performing 5-part differential, 26
parameters analysis of anti-coagulated human whole blood samples introduced
in closed vials.
Disclaimer
The manufacturer reserves the right to:
Please note that this Manual may be revised without prior notification. The
Manufacturer reserves the right to change the specifications of the product
and/or the contents of this manual as deemed necessary, without prior notice.
Pictures included in this manual may differ from the actual delivered product.
LASER RADIATION
Class 1 laser product
CE mark
Technical contact
Icon5 is a precise and accurate laboratory instrument. Its built-in safety
measures guarantee safe and reliable operation. Being a complex device, there
is limited access to internal structures provided to the end user.
Icon5 is a Class 1 laser device, the embedded laser source is not exposed to the
user or service personnel. Under no circumstances should the Operator
attempt to open or remove the outer cover of the analyzer, as this may
influence reliable operation and might fool built-in safety measures and can
invalidate warranty.
There are no user serviceable parts inside this analyzer. Adjustments of internal
structures and repairs of these structures should only be performed by
qualified service personnel.
Your local dealer is always ready to help you in case of malfunction. The contact
of your local dealer is available on the insert of this Manual.
Norma will be doing its best to resolve your problems either directly or
indirectly with the help of its local representative.
Warranty
Your Icon5 comes with a one year manufacturer’s warranty against
workmanship defects. Warranty claims must be made through your local
representative.
Your supplier will only be able to repair and warrant operation of the analyzer
as per its specifications if all service actions are performed by qualified service
personnel, and the Warranty Label is intact.
Warranty is void if any of the below can be identified, or inoperability, malfunction can be traced
back to any of the events listed below:
The manufacturer does not recommend and support modifying the operating software of this
hematology analyzer. The analyzer is able to log and track modifications to the operating software.
- In the case, when the End User required or performed installation of any software
product not controlled by the manufacturer and would afterwards experience erroneous
or unreliable operation of the analyzer, or would experience erroneous results, then the
manufacturer cannot accept any complaints about operation, reliability or performance.
Exclusions: The below listed items are not covered by warranty unless they were found damaged
upon arrival to the End User. In such cases, proper complaint must be filed along with a complaint
registered with the forwarder/Distributor company:
When there is only liquid in the aperture, a certain electric resistance can be
measured due to the conductive characteristic of the liquid.
Particles (cells) do not conduct direct current. When a particle (red) passes
through the aperture (yellow) then the conductive cross section of the aperture
filled with conductive liquid will decrease, and a resistance different from that
of the “empty” aperture can be registered over the liquid.
Interpreting scattergrams
Icon5 presents scattergrams on its screen,
in printed or in electronic format. Cells
scatter light when they pass through laser
beam. The level of light scattering in
straight (forward) direction is proportional
to cell size. Cells also include granules,
compartmented nuclei and other
organelles. This leads to side scattering of
light. Side scattering depends on cell
structure (e.g. granularity). The populations
are marked with colors.
Interpreting histograms
Icon5 presents histograms on its screen,
in printed or in electronic format.
Hematology parameters
Parameter Explanation Typical units
name
3 9
WBC White blood cell count 10 /µL & 10 /L
6 12
RBC Red blood cell count 10 /µL & 10 /L
HGB Hemoglobin g/dL, g/L, mmol/L
HCT Hematocrit %, L/L
MCV Mean Corpuscular Volume fL
MCH Mean Corpuscular HGB pg, fmol
MCHC Mean Corpuscular HGB g/dL, g/L, mmol/L
Concentration
RDWsd/cv Red blood cell distribution width fl/%
3 9
PLT Platelet (thrombocyte) count 10 /µL & 10 /L
MPV Mean Platelet Volume fl
PCT Plateletcrit/thrombocrit %
PDWsd/cv Red blood cell distribution width fl/%
P-LCR% Platelet Large Cell Ratio; %
The ratio of PLT’s with volume above
12fl compared to the total PLT count
3 9
P-LCC Platelet Large Cell Count 10 /µL & 10 /L
number of PLT’s with volume above
12fl
3 9
LYM# Lymphocyte count 10 /µL & 10 /L
3 9
MON# Monocyte count 10 /µL & 10 /L
3 9
NEU# Neutrophil count 10 /µL & 10 /L
3 9
EO# Eosinophil count 10 /µL & 10 /L
3 9
BAS# Basophil count 10 /µL & 10 /L
LYM% Lymphocyte percentage (of WBC) %
MON% Monocyte percentage (of WBC) %
NEU% Neutrophil percentage (of WBC) %
EO% Eosinophil percentage (of WBC) %
BAS% Basophil percentage (of WBC) %
5-part differentiation of White Blood Cells
Five-part white blood cell differential counts (so-called diffs), performed during
flow cytometry counting of blood cells, can accurately classify lymphocytes,
monocytes, neutrophils, eosinophils and basophils with an error rate not
exceeding that of conventional diffs. The differential count is made on diluted
and hemolyzed specimens; hemolyzing is performed to break down RBC
membranes, form stable HGB complex and shrank WBC cells. Normal RBC’s
have no nucleus; only WBC cells remain in the solution. The size (volume) and
complexity of various WBCs are detected by the analyzer to classify cells.
Sample vials with cap is lowered into the sample rotor that turns the vial upside
down to allow small sample volume aspiration. The vial is pushed onto a fixed
metal sampling needle and is pierced in the upside down position. The fixed
needle is equipped with a washing head to clean the needle’s external and
internal surfaces.
One of the 1.2µl blood samples is mixed with a lyse reagent for a dilution of
approximately 1:200. The temperature of the mixture is controlled. This
mixture is then driven through the WBC measurement head. The lysed WBC
solution prior to entering the measuring flow cell passes through an absorption
measurement window and is measured for HGB content at 540nm wavelength.
The lysed WBC solution is then focused using a combination of a 100 µm orifice
and a double hydrodynamic focusing system and is led through a perpendicular
red (660 nm wavelength) laser beam. The detector tube and amplifier collects
scattered data about each individual passing cell during the 15 seconds of
measurement.
The second 1.2µl processed blood sample is moved into the mix chamber using
diluent, creating a dilution of approximately 1:1000. This mix dilution is then
forced through a 70 µm aperture with the help of a sheathing fluid, creating an
overall dilution of 1:20000 The electrical signal of each individual cell is
measured (through stainless steel and platinum electrodes and is collected for
8 seconds) for RBC and PLT counts.
As a final step, the system empties the measuring elements and is ready to take
the consecutive sample.
Icon5’s closed vial mode supports the following closed sample vials:
Please consult the Instructions for use of relevant primary sample vial types.
WARNING
Vials with rubber stoppers (caps) are designed for a limited number
of penetrations with a needle. Using a closed vial more than 5 times
poses a risk of damaging the rubber cap and causing liquid damage
in the analyzer.
Samples must be thoroughly mixed and homogenized adhering to general
blood sample collection guidelines prior to analyzing them on Icon5.
Package contents
Icon5 comes in a double cardboard packaging. The outer box provides
protection against shipping damage to the analyzer’s internal box.
The internal box contains the analyzer and its accessories inside protective
foams. The analyzer comes with a protective film covering the scratch resistant
polycarbonate cover elements.
Upon arrival, carefully check the contents of the package and look for visible
damages, even on the outer packaging box. File any obvious damage through
the shipping company to be eligible for compensation and support from your
local dealer and from the manufacturer.
Packing list:
Accessories
external power supply
(SYS1443-6512-T3)
10
6. USB connectors
7. RJ45 (network) connector
8. Power connector
9. Reagent connector
10. ID label of analyzer with S/N and electrical data
Installing the analyzer
Note
Icon5 is a precise and accurate hematology analyzer. Incorrect
handling or accidental falling of the analyzer may damage parts
inside and could influence performance.
1. Carefully remove Icon5 from the shipping carton. Look for signs of
damage, such as cracks on the outer covers or loose/missing screws. If
you find such signs, please file a complaint to the shipping company to
be eligible for compensation and support from your local dealer and
from the manufacturer. Make sure you can find all the accessories
listed in the packing list.
2. Prior to powering on Icon5, allow the analyzer to reach room
temperature to avoid dew condensation. Sudden temperature
changes may cause dew to condensate on colder internal structures
and can lead to damage of electronic components.
3. Place the analyzer on a desktop bench and find a nearby grounded
power outlet. Please avoid power extension cords, use direct
connection to the power outlet. Always use the power supply
packaged along with the analyzer.
4. Connect the power supply to the socket on the rear plate of the
analyzer.
5. Connect the power cord to the power outlet.
6. Remove the protective foil from the analyzer’s front.
Attention
If you experience any error, like smoke, immediately disconnect the
power cord from the power outlet.
Use a fire extinguisher if necessary.
Keyboard (optional)
Icon5 supports connecting external USB keyboards that can
facilitate data entry. The keyboard can be connected any
time.
Mouse (optional)
Icon5 supports connecting an USB mouse. The mouse can be connected any
time. If a mouse is connected, the touch screen remains operating, yet a small
arrow cursor will be displayed to track the movement of the mouse.
Network (“Ethernet”)
Connect the cable (not included) from the computer network
to Icon5’s RJ45 socket. Network settings must be revised.
Reagent pack
The Norma Icon5 reagent pack provides
sufficient reagent to run a specified number of
tests with Icon5. The pack contains all necessary
reagents. Waste must be collected in a separate
container.
Each tube is equipped with a label indicating which bottle to connect it to. The
caps match the standard Icon5 reagent bottles.
Attention
For best results and performance, place the reagents to the same
level (on the same desk) as the analyzer.
Upon connecting the power supply, Icon5 is in low power consumption mode
and ready for operation.
Power button
It is located on the right side of the
beveled front panel. When the external
power supply is connected both to the
analyzer and to the mains outlet, the
analyzer can be turned on by pushing
the power button.
Power on
The single push of the START button will power on the electronics and Icon5
will display the welcome screen and the main menu automatically.
When Icon5 is turning on, the screen remains dark, only backlight illumination
can be seen at the edges of the screen. This is normal.
Power off
Icon5 is a sensitive device with liquids inside. A simple power off might leave
the system in a state that can represent potential risk to moving components
and liquids inside.
Icon5 can be turned off by tapping the local menu button and initiating
Shutdown in the menu. To power off the analyzer, select “Shutdown”. The
automatic process will start, and the analyzer is safely powered off within 2-3
minutes. Do not interrupt this sequence.
Handling reagents
Icon5 operates with special reagents. The reagents have been carefully
designed to meet the requirements of the analyzer. Icon5’s quality assurance
system requires the use of these special reagents. Icon5 recognizes and accepts
genuine reagents only.
Reagent characteristics
When using bulk reagents (from external bottles) the shelf life is up to 4 years
for the reagents, but open stability is 180 days.
An EDTA quantity below or above the recommended ratio will cause false or no
results due to improper chemical treatment of blood cells.
Samples can be stored refrigerated (between 2-8°C) for not more than 24hours.
Samples must not be mixed while cold. Doing so may physically damage the
cells.
Mixing
Immediately prior to inserting the vial into the analyzer, gently invert the vial
then rotate it back. Repeat the cycle five-six (5-6) times; not less and not more.
Insert the vial into the analyzer’s sample opening and press START.
Menu system
1- Status bar
Displays active menu and
instrument status
2- Page count area
(when left-right swipe is available)
3- Main information
4- Quick links
Displays further functions available
related to displayed data
Menu tree
Daily routine Sampling Run FastBlank, closed vial samples
(including QC samples and calibrator)
Results Database functions
Gestures
Icon5’s touch screen user interface allows operating the analyzer with light taps
and swipe actions listed below.
Swipe (quickly moving your finger along the screen without lifting
your finger)
All actions (touch and release) are acknowledged by a small circular symbol
appearing at the point where the user tapped the screen, and an audible “tick”
is head if sound was enabled.
Using the above basic actions, Icon5 can interpret various combined actions, so-
called gestures.
Tap
Touching (tapping) an active area (icon, data input
field) will activate the corresponding function, or
open a sub-menu. For data input fields, an on-screen
keyboard is displayed at the top or bottom quarter
of the screen depending on the position of the data
input field.
Swipe
Icon5 is able to display data that spans over several
screens (pages). Such pages are indicated with small
circles in the center near the top half of the screen.
Alphanumeric input
Input fields are marked with an underscore.
Tapping into any field allows editing data.
The on-screen keyboard will be displayed
either in the top or in the bottom quarter of
the screen to make sure that the input filed
remains visible. The cursor is indicated by a light underscore below the
character.
The on-screen keyboard allows changing symbols entered by changing the keys.
The keyboard can be changed by swiping left or right to reveal capital letters,
small letters and symbols.
Actual keyboard layout depends on the language of the user interface. Always
the standard keyboard layout is used to interpret keystrokes.
Barcode input
If an external barcode scanner is connected, then the read ID will be entered as
Sample ID for the consecutive measurement.
To input barcodes:
When a barcode has been recognized and interpreted, the bar-code will be
entered into the Sample ID input field.
Numeric input
Some data fields may require modification on a larger scale. For this purpose,
Icon5 offers another intuitive setting method.
Symbols ---, -- and – buttons will have decreasing effect on first, second and the
third digit, respectively.
Symbols +, ++ and +++ buttons will have increasing effect on first, second and
the third digit, respectively.
Description of icons displayed
[Options] Settings
However, you can customize these settings to make Icon5 fit into your daily
routine even better. Here is a brief list of options that you are recommended to
adjust or verify to make the use of the analyzer simple and convenient.
Initial setup
Below is a recommended sequence of settings so that you can start using the
analyzer easily:
You can skip each stage; you can leave the setup wizard at any time. You can
also (re)run the wizard at any time later on.
Settings – Customize
Appearance
[Page1]
Icon5 will confirm the new wallpaper file, or will display an error message if the
image is not of the right size, format, or it cannot be found in the root folder of
the USB memory device.
Languages
[Page2]
Units
[Page3]
E-mail settings
[Page2]
- Sender address
- Account e-mail address
- Account password
Enabling “Print Technical Data” will make Icon5 print result related technical
information on the bottom of the page.
You can force Icon5 use ESC/POS language for unrecognized printers.
WARNING
Forcing an incorrect language to a printer may result in scrambled
printouts, and empty pages, even one single symbol per page!
It is possible to exclude (hide) histograms from printouts to save paper.
Customization
[Page4]
Profiles
[Page5]
This page also allows setting up Norma Access parameters. For details, please
contact your service personnel.
Setting up reagents
Icon5 requires reagents to run. Connected reagents
can be managed from the Reagents menu. Reagents
menu has two separate branches: one for packed
and one for individual (“Bulk”) reagents.
Reagent: Pack
Tap on the Reagent icon in the main menu then
select “Pack”.
Activation and deactivation of reagents is possible through the local menu. Tap
the Deactivate icon to stop using a pack. Tap the Activate icon to start using a
pack.
- open
- shared
- WPA2 Personal
- WPA Personal
Power on
Icon5 will mostly be waiting in stand-by mode.
The fluidic system will always require initialization, a programmed startup. This
is indicated by a yellow analyzer in
measurement mode.
Introducing a Sample
The opening on top of Icon for Closed vials is designed to receive
blood collection vials of defined dimensions with caps on.
To avoid personal injury, do not put your finger into the opening.
Avoid placing objects different from a blood collection vial to avoid
damage to the analyzer.
Avoid putting vials without caps into the sampling mechanics.
Avoid putting more than one barcode label onto the blood
collection vial to avoid the vial getting stuck in the sampling
mechanics.
Icon’s sampling mechanics is equipped with a cap detector.
Cap detector operation may be impaired by additional labels put on
the sample vial, and Icon5 may refuse to run such vials.
Tapping the analyzer symbol in the center will start the measurement.
Icon5 will lower the sample vial into the sampling mechanics and will take a
sample with its built-in cap piercing needle.
When the sample has been taken, the vial will be returned to the Operator.
Results
Icon5 can display results in two ways:
A small red flag indicates that the system detected a condition that requires
attention. Flags next to result values indicate that the numerical result is
beyond actual minimum or maximum accepted values (normal range). Flags
with yellow text indicate technical flags, messages related to the evaluation
process of the sample. Most of these flags require the sample to be rerun.
Technical flags
Icon5 is able to display technical warning flags. Flags can appear in various
locations, depending on significance and relevance. Flags by the Sample ID refer
to sampling or sample handling mistakes. Flags displayed in WBC and RBC
groups refer to parameter related irregularities. Technical flags rather give
guidance for the user, an indication of not operating the analyzer as
recommended.
Flags have different levels of significance. Each flag has two confidence levels.
As long as results are not negatively influenced by the detected conditions,
results will be displayed along with the flags. As soon as severity of the problem
goes beyond safe result interpretation, and would impact readings, values will
be dashed out.
Flag name Type Flag meaning Letter
The scattergram was found abnormal. This may
Abnormal Diff WBC be the result of an over-mixed sample. See S
Sample handling section earlier in the manual.
HGB The reference value of temperature sensor on
Temperature RBC HGB head is unset. Contact Service. R
reference error
MCHC is over 500 g/L. There was an inadequate
quantity of sample in the sampling vial. Rerun the
High MCHC Sample sample if possible. Please observe minimum m
sample volume guidelines. May also indicate a
block in the sampling system.
Internal measurement related structures are not
working properly. Your analyzer might be
Improper
WBC operating outside the described environmental V
Voltage conditions. Please observe environmental
requirements. Maybe cleaning is due.
Internal measurement related structures are not
working properly. Your analyzer might be
Improper
RBC operating outside the described environmental v
Voltage conditions. Please observe environmental
requirements. Run RBC Cleaning.
Low level measurement structure error in the
Laser Packets
System optical (laser) module. Rerun sample. If the E
Empty problem persist contact Service.
There was an inadequate quantity of sample in
the sampling vial. Rerun the sample if possible.
Low sample
Sample Please observe minimum sample volume B
volume guidelines. May also indicate a block in the
sampling system.
Too many particles have been detected during
RBC/PLT measurement. Some internal bubbles
Overload System x
may cause the error. Do RBC Cleaning. Rerun the
sample. If the problem persists, contact Service.0
Pressure sensor detected high pressure during
Pressure Sensor measurement process and released a valve to
WBC K
High Value avoid system damage. Rerun sample. If the
problem persist contact Service.
Range Results are outside the linearity range of the
WBC
analyzer. L
exceeded
Range Results are outside the linearity range of the
RBC
analyzer. l
exceeded
Flag name Type Flag meaning Letter
Verify reagents. Some bubbles may have formed
inside the analyzer. If the problem persists, run
RBC Noise RBC n
RBC cleaning, prime DILUENT and rerun the
sample.
Low level measurement structure error in the
RBC packets
System impedance (aperture) module. Rerun sample. If e
empty the problem persist contact Service.
RBC/PLT measurement encountered an internal
RBC Slice errors System homogeneity error. Rerun the sample. If the s
problem persists, contact Service.
The internally prepared sample did not meet
proper measurement criteria; sample flow in the
flow cell was not homogenous.
Unintelligible Rerun the sample.
WBC W
histogram If persists, run WBC Debubbling. If it cannot be
resolved with repeated WBC Debubbling, contact
service, as temperature stability or liquid flow
might have been influenced.
Improper internal homogenization of sample.
Rerun the sample. If maintenance related
Unstable HGB WBC H
cleaning actions have been omitted, please do
them now. Run WBC Debubbling
The HGB measurement signaled an error.
Unstable HGB
WBC Probably some bubbles got stuck in the flow cell. h
baseline Run WBC Debubbling.
Internal liquid flow was disturbed, or temporary
Unstable
WBC block of the liquid flow was detected. Rerun the A
Voltage sample.
Internal liquid flow was disturbed, or temporary
Unstable
RBC block of the liquid flow was detected. Rerun the a
Voltage sample.
The flow cell detected an abnormal number of
disturbing signals. This may be the result of air
WBC Noise WBC N
bubbles blocking the flow cell. Run a Prime All
function.
Reports
PDF
Icon5 is able to send reports
ready to be printed via e-mail or
is able to save reports to USB HD.
Reports include all entered
patient data, along with reported
parameters, scattergram,
histograms and additional notes,
comments created using the
analyzer during or after analysis.
Printer (Thermal)
Icon5 is able to send report to connected
thermal printer. When Icon5 detects
Thermal printer is connected proper
symbol will be shown on status bar.
It is possible to hide
histograms from the printout.
For details, see Settings /
Printout.
Note:
- PLC5
- ESC/POS
To power off, tap the local menu button. A quick menu will be displayed
offering various actions including shutdown (power off).
Shutdown will prepare the system for a longer period of inactivity. The process
takes about a minute, after which the system will automatically power off,
leaving Icon ready to be disconnected from the mains if necessary.
Calibration
Icon5 as any laboratory analyzer can be calibrated. For this purpose, Icon5
offers two ways to perform this task. Icon5 arrives to your laboratory factory
calibrated.
Target values
Calibration is performed comparing the reported values of the analyzer to a
known specimen, or standard. Hematology analyzers have their own calibrator
materials called Hematology Calibrators.
These target values represent the standard to which your analyzer’s reported
values are compared.
Automatic calibration
Icon5 offers an automatic calibration method to keep the process as simple as
possible.
Icon5 will mix and run the sample and will calculate the mean values of the
measured parameters. The mean value will be used to calculate the calibration
factors using the formula below
𝑃𝑎𝑟𝑎𝑚𝑒𝑡𝑒𝑟𝑡𝑎𝑟𝑔𝑒𝑡
𝐶𝐴𝐿𝑝𝑎𝑟𝑎𝑚𝑒𝑡𝑒𝑟(𝑛𝑒𝑤) = ∗ 𝐶𝐴𝐿𝑝𝑎𝑟𝑎𝑚𝑒𝑡𝑒𝑟(𝑜𝑙𝑑)
𝑃𝑎𝑟𝑎𝑚𝑒𝑡𝑒𝑟𝑚𝑒𝑎𝑛
Expected values
Quality Control is performed measuring the same specimen over a period of
time. Hematology analyzers have their own control material called Control
Blood.
Control blood is made by using human samples mixed with specially treated
particles to guarantee stable values for parameters over the stability period of
the control material. All control blood products come with a list of the expected
values of parameters.
Browsing records
The database allows browsing records listed in rows. Swiping (scrolling)
gestures can be used to access various views of the database.
↑ ↑ ↑
Human Date Record Details
↓ ↓ ↓
↑ ↑ ↑
QC Date Record Details
↓ ↓ ↓
Results
↑ ↑ ↑
Calibration Date Record Details
↓ ↓ ↓
↑ ↑ ↑
Fast Blank Date Record Details
↓ ↓ ↓
Swiping the screen in the left or right direction will switch pages displayed.
Swiping the screen up or down will scroll the view up or down: you can scroll
between dates, groups of records and individual records.
Database context menu
Tap the local menu button in the lower right corner
of the screen brings up a so-called context menu
screen.
Selection
By tapping “Same Day” you can select
records measured on the day of the
actual result.
By tapping “Same SID” you can select records matching same SampleID of the
actual result.
Tapping “Selection Start” will allow multi-selection of results. You will be taken
back to the database table view. You can select the END of the selection by
ticking the box in front of any sample. All samples between the “first” (where
selection started) and the “last” (the one you ticked) will be selected.
The number of selected records will be shown under the list of records. By
(un)ticking the box by the (e.g.) 16 of 273 will deselect ALL records.
Moving to specific date in database
After selecting date with date
picker tap “Jump” button to move
the corresponding results in
database list. Icon5 will execute
the jump operation only if
database contains records with
the selected date.
Managing data
Tap the corresponding icon to
activate the function:
Sending e-mail
To e-mail, enter the address of
the recipient. E-mail settings are
available under Options menu
Functionality submenu.
Keeping “CC” checkbox selected will send a copy to the default e-mail address
(as defined under e-mail settings)
Maintenance
The external plastic parts can be cleaned with lint free cloths soaked in light
soapy water, or alcohol based cleaning materials, typically alcohol-based
window or computer display cleaning solutions.
Cleaning the stainless steel back plate should be done using alcohol based
cleaning materials, typically alcohol-based window or computer display
cleaning solutions.
Daily maintenance
- Clean outer surface
- Check reagent connector
Emergency procedures
- Cleaning
- Power off
Software updates
You can initiate this process by tapping the proper button in Supervision menu
About submenu. Contact your Service team for details.
Management menu
Maintenance
Maintenance menu allows running processes
related to maintaining reliable operation:
- RBC cleaning
- WBC debubbling
Service
This menu is password protected and it is only
available for and should be used by authorized
service personnel.
Supervision
History
History will list all user activity in a log, grouped
into activity types. You can review Information,
User action, Errors.
History details
Swiping left on a specific line of History will show
the details of a specific message. You can sweep
up and down to access further pages, lines.
About
The About screen will display instrument
specific information
- SW version,
- Product ID,
- Manufacturer data,
- Service contact to your local service
personnel.
- You can restart the setting wizard
- You can initiate Remote
Management (refer to your service
personnel)
1 2 3 4 5 6
The status bar contains symbols and shortcut to main menu. Instrument
statuses are including following informations (from left to right):
WARNING: Only qualified service personnel may replace the internal fuse and battery.
When not in use or relocating, always store the analyzer in its original packaging to
avoid physical damage.
Storing the analyzer outside the above specified environmental conditions or outside
its original packaging may impair operation and may also cause erroneous or faulty
operation.
Do not place the analyzer near a direct heat source or into direct sunlight
The desk supporting the instrument should be flat, horizontal and stable enough to
support the weight of the analyzer and accessories
Operating the analyzer outside the above specified environmental conditions may
impair operation and may also cause erroneous or faulty operation.
Reagent consumption
All values in ml (milliliter). Values indicate worst case aspirated volumes.
Parameter Carryover: %
WBC <1%
RBC <1%
HGB <1%
PLT <1%
Display range
Values measured below or beyond the limits listed above will be marked with <
or > symbols and the actual limit itself. E.g. PLT 1200 will be displayed as: “PLT
1000<” accompanied by a linearity range flag.
Note (1): display precision is 1 decimal digit for diff absolute counts
Note (2): If WBC is below 0.1, the diff values will NOT be displayed.
0.1<WBC<1.0 the diff values get an * mark
Error messages
The analyzer can report various errors. Error messages are displayed in a small
pop-up bubble in the lower right corner of the screen. Errors have technical
details which are useful when contacting Technical Support.
Technical Support will ask for diagnostic files to facilitate problem resolution.
Either have the analyzer connected to the Internet, or have a USB storage
device ready for saving and emailing data.
Mechanical Errors
Error message Reason for error Remedy
Shear Valve error. An error Mechanical Error Rerun operation. If
occurred while testing persists, contact Service.
Shear Valve backlash.
An error occurred while Pump not operating, pump Rerun operation. If
attempting to use pump. jammed. persists, contact Service.
Sampling Rotor error. An Mechanical error. Rerun operation. If
error occurred while persists, contact Service.
attempting to use Sampling Check whether closed
Rotor. mode sampling opening is
obstructed.
Sampling Rotor error. An Mechanical error, sensor Rerun operation. If
error occurred while induced. persists, contact Service.
attempting to use Sampling
Rotor (Sampling Rotor
invalid position).
Valve error. An error Valve controller Error Rerun operation. If
occurred while attempting persists, contact Service.
to set valve.
Shear valve recovery The shear valve had Confirm, and run
problems moving, but the initialization.
system resolved the error
7
Operation related errors
Error message Reason for error Remedy
Communication error. An No Internet access Consult network
error occurred while administrator.
sending an e-mail
(connection exception).
Control assay sheet already You scanned a sheet that
exists. has been scanned
previously.
Communication error. An Network error. Consult network
error occurred while administrator.
attempting to download
software upgrade file
(unspecified error).
Measurement error General measurement Contact service with error
error. details.
QC LOT field cannot be You did not enter a LOT ID. Enter a LOT ID.
empty.
QC with defined LOT Verify LOT ID
already exists.
An error occurred while QR code faulty. Try scanning the sheet
decoding QR code (invalid again. If you fail, load data
or broken QC code). from a file. Contact tech
support for details.
Reagent detection error. Bubbles detected. Check that reagent
An error occurred while container is not empty.
attempting to calibrate
reagent sensor at Diluent
inlet.
Reagent detection error. Bubbles detected. Check that reagent
An error occurred while container is not empty.
attempting to calibrate
reagent sensor at Lyse
inlet.
Reagent detection error. Bubbles detected. Check that reagent
An error occurred while container is not empty.
attempting to calibrate
reagent sensor at Cleaner
inlet.
Vial Detector warning Incorrect vial inserted in Insert a correct vial.
(missing condition). sample rotor.
Vial Detector warning No vial detected. Insert a vial, check for
(insert vial). compatibility.
8
Error message Reason for error Remedy
Vial Detector warning A vial is detected in sample Remove sampling vial.
(remove vial). rotor, yet selected action
requires the vial to be
removed
User error A vial got blocked, or was Check for obstruction. Run
incorrectly inserted “Eject” from local menu.
An error occurred while Incorrect email data Verify settings, contact
sending an e-mail specified, or unsupported network administrator.
(username or password email client was set up. Observe and follow
exception). network and email setting
instructions.
Electronic errors
Error message Reason for error Remedy
Camera error. An error The camera could not be Power off and on again. If
occurred while activating detected. the problem persists, use
camera. electronic files to load
data.
Contact service to schedule
a visit.
Touch screen does not Faulty touch screen Connect an external mouse
react to gestures for temporary operation.
Contact Service
Uncertain operation of Faulty touch screen Connect an external mouse
touch screen for temporary operation.
Contact Service
9
System (software) errors
Error message Reason for error Remedy
Database error. An error Software error, corrupted The system will try to
occurred while attempting file system. recover the error. If the
to read/write database. error persists, contact Tech
Support.
Database error. An error Software error, corrupted The system will try to
occurred while attempting file system. recover the error. If the
to create record in error persists, contact Tech
database. Support.
Communication error. An A key system element is Contact Tech Support.
error occurred while missing.
attempting to read/write
ProductID.
Fluidics error. An unknown A key system element is Contact Tech support for
low-level error occurred missing or is guidance.
while attempting to malfunctioning.
execute fluidics function.
Fluidics error. An unknown A key system element is Contact Tech support for
error occurred while missing or is guidance.
attempting to execute malfunctioning.
fluidics function.
Fluidics error. An unknown A key system element is Contact Tech support for
error occurred while missing or is guidance.
attempting to get/set malfunctioning.
arguments of fluidics
subsystem.
Processing error. An error A key system element is Contact Tech support for
occurred while attempting missing or is guidance.
to process raw malfunctioning.
measurement data.
System error. An error A key system element is Contact Tech support for
occurred while attempting missing or is guidance.
to query active QC LOT. malfunctioning.
Communication error. An A key system element is Contact Tech support for
error occurred while missing or is guidance.
attempting to open serial malfunctioning.
port.
System error. An error A key system element is Contact Tech support for
occurred while attempting missing or is guidance.
to play sound file. malfunctioning.
10
Error message Reason for error Remedy
Communication error. An The printer is not Connect the printer.
error occurred while connected. Contact Tech support for
attempting to print with A key system element is guidance.
thermal printer. missing or is
malfunctioning.
Reagent detection error. A key system element is Contact Tech support for
An error occurred while missing or is guidance.
attempting to detect malfunctioning.
Diluent.
Exporting error. An error No USB storage device Connect USB storage
occurred while attempting attached. device
to export Tab Separated A key system element is Contact Tech support for
File. missing or is guidance.
malfunctioning.
Communication error. An A key system element is Contact Tech support for
error occurred while missing or is guidance.
attempting to send e-mail malfunctioning.
(unspecified error).
An error occurred while A key system element is Contact Tech support for
attempting to set volume. missing or is guidance.
malfunctioning.
Software error A key system element is Contact Tech support for
missing or is guidance.
malfunctioning.
System error. An error A key system element is Contact Tech support for
occurred while attempting missing or is guidance.
to initiate system malfunctioning.
(unspecified error).
Reagent detection error. A key system element is Contact Tech support for
An error occurred while missing or is guidance.
attempting to detect Lyse. malfunctioning.
Reagent detection error. A key system element is Contact Tech support for
An error occurred while missing or is guidance.
attempting to detect malfunctioning.
System Solution.
Reagent system error. An A key system element is Contact Tech support for
error occurred while missing or is guidance.
attempting to activate malfunctioning.
reagent (expired).
System error. An error A key system element is Contact Tech support for
occurred while attempting missing or is guidance.
to modify reagent’s malfunctioning.
tracking level.
11
Error message Reason for error Remedy
System error. An error A key system element is Contact Tech support for
occurred while attempting missing or is guidance.
to query active reagents malfunctioning.
(inactive or expired
reagents).
Reagent detection error. A key system element is Contact Tech support for
An error occurred while missing or is guidance.
attempting to read reagent malfunctioning.
sensor data.
Firmware Error A key system element is Contact Tech support for
missing or is guidance.
malfunctioning.
Remote Management error IP address of Access Server Contact Service for proper
is incorrect, or a Firewall setup
blocks connection
12
Troubleshooting
Measurement problems
Symptom Possible cause Remedy
High BLANK values Contaminated Replace reagents
reagents
Cold reagents Make sure reagents reached room
temperature
13
Electrical problems
Symptom Possible cause Remedy
Instrument cannot be Power is not connected Connect power source,
powered on Power voltage is too high, verify connections
or too low Replace / check power
Fuse is blown inside supply (is LED on?)
analyzer Contact Service
Indicator on top/front of Power voltage is too Replace power supply
the analyzer glows in RED high/low
or BLUE color BEFORE
power ON
Instrument powers down Power system fluctuations Try to start the analyzer
during operation again. Use a UPS.
Contact Service
Software error message Corrupted SW system. Contact Service
Mechanical problems
Symptom Possible cause Remedy
Sample vial is stuck in Power outage during Start analyzer again,
sample rotor operation initiate measurement
sequence
Oversized vial, vial stuck Go to local menu and
initiate “Eject”
14
Symptom Possible cause Remedy
Analyzer reports empty Too low blood sample Run vials with correct
measurements volume sample volume
Rerun sample.
Clogged sampling paths Contact Service
Leakage beneath analyzer Blocked sampling Try running system
valve/needle system cleaning.
If it still fails, contact
Service
Leaking reagent connector Check reagent connector
Hydraulic problems
Symptom Possible cause Remedy
Instrument reports Bad quality sample used Try to initiate a
clogging measurement. Built in
functions will attempt to
unclog needle and
apertures.
If the problem persists,
contact Service
Network problems
Symptom Possible cause Remedy
Network error message at HL7 or email Verify HL7 / email settings.
the end of every sample communication is enabled Disable email (clear
(forced) but account data) or clear HL7
settings/network does not settings
comply.
Remote Management error Remote Management data Contact Service for correct
message are incorrect settings
WIFI dongle not recognized Driver error Check supported list of
WIFI chipsets
15
Printer problems
Symptom Possible cause Remedy
Scrambled printout Incorrect printer model Use Printers with PLC5 or
ESC/POS language support
Empty pages printed Incorrect printer model Install compatible printer.
Contact Service for
instructions
Printer connected, but no Printer is not supported, Test the printer and USB
printer Icon in top row of USB cable is wrong cable with a PC.
screen
Printer is OFF Restart Icon5
Make sure to use the same
grounded electrical outlet
for Icon5 and peripherals.
Test USB sockets with an
external Keyboard or with
a USB stick (and try to save
diagnostic files to the USB)
16
Certificates, standards
Icon5 complies with the following standards and bears corresponding
certificates as listed below.
Declaration of Conformity
The supplier Norma Instruments Zrt. (Arany János u. 11-13, 3530 MISKOLC,
HUNGARY) hereby declares that Norma Icon5 fulfils the essential obligations
imposed by Article 3 and Annex 1 of the Directive 98/79/EC on „In vitro
diagnostic medical devices” and achieves the performance intended. The
Manufacturer declares that the use of the a.m. device under appropriate
conditions will not compromise the health or safety of the patient, the user or
other involved person.
17
Appendix A – Signs of improper sample handling
Pre-analytical factors
Distinct layers
When blood is left standing for a certain time (10-60
minutes), sample dependent), the sample will sediment;
cells inside will obey gravity and will tend to sink down
forming a yellowish liquid on top with a thin layer of
White Blood Cells and a thick layer of dark red colored
Red Blood Cells below.
Samples must be manually mixed prior to analysis using Sedimented (left) and
the analyzer. homogenized sample (right)
Immediately prior to inserting the vial into the analyzer, gently invert the vial,
hold it and make sure all the liquid has flown down; then rotate it back and
again make sure all the liquid has flown down. Repeat the cycle five-six (5-6)
times.
Doing more than 6 inversions or less than 5-6 inversions will negatively
influence the result.
If you can see big bubbles in the sample, then most probably there are invisibly
small bubbles also present in the sample. These can also reduce the required
sample amount.
Samples with bubbles inside should be let relax more before running it on the
analyzer.
18
Pink plasma (hemolysis)
Plasma above the blood cells must be clear somewhat
yellowish. A discolored plasma indicates sample specific
problems, or handling problems (temperature, time
elapsed since taking the sample). Hemoglobin is
normally located inside the Red Blood Cells. With time,
or storage temperature beyond storage range will cause
the RBCs degrade and release HGB – a red colored
substance. Once HGB appears in the plasma, the
sample is not acceptable for analysis: cell counts will Normal (left) and hemolysed
be incorrect. (right) plasma
19
Post-analytical factors
Such samples should be rested, analyzed for possible hemolysis, and rerun for
verification if possible.
20
Elevated MON, lowered NEU:
- over-mixing
- Cause: shifting of NEU population, appearance of pseudo-populations.
- Can be sample-specific.
Remedy:
Leave the sample rest for 30 minutes and rerun it to obtain much better scatter
and WBC differential results.
21
Errata
This manual is under continuous revision and improvement. The sections listed
below are known to describe functions focusing on convenience of use that
may not yet be available in your analyzer.
References:
- 98/79 EN Directive
- pubmed.org
- eurlex.eu
22
Version Change Sections Date Edited by
Affected
1.0 Original version All 2015-12-16 CsMagyar
1.1b2 Revision A 2016-07-25 CsMagyar
- added optical meas method sketch p6
- review screens…
- hold gesture removed
- ER mode and references removed
- Flags discussed (CSJ)
- Troubleshooting take1 review (MCS)
1.1b5 QuickResult screen 2016-08-19 Atremmel
Flags update
1.1b6 PreLaunch Proofing all 2016-08-26 CsMagyar
1.1b7 Scattergram color explanation added; image replaced to match colors p13 2016-08-31 CsMagyar
1.1b8 Operating temperature max set to 25°C p85 2016-10-19 CsMagyar
1.2 Intermediate revision – proof, screenshot update all 2016-11-10 RND
1.3 First Publication Release all 2016-12-01 CsMagyar
1.31 Applicable standards updated p95 2017-01-04 CsMagyar
1.32 Norma Direct contact removed p5 2017-01-13 CsMagyar
Warranty: list of excluded parts added
1.330 Measurement Flags updated p58-59 16-02-2017 CsMagyar
Sample handling added p28-29
Appendix (signs of improper sample handling)added p95-98
1.331 Power off at the end of the day p26 02-03-2017 CsMagyar
WBC N flag added p59
1.333 Performance Data added 02-JUN-2017 CsMagyar
1.334 Sarstedt Monovette vial removed from list of supported vial types p17 29-AUG-2017 CsMagyar