Documente Academic
Documente Profesional
Documente Cultură
Service Manual
(Models 110 - 410)
DINAMAP® PRO 100-400 V2
Patient Monitor
Service Manual
NOTE
The Model Numbers 100 through 400 are generic in nature and reflect the range of product codes
available. Your product may be labeled with a specific product code such as DINAMAP® PRO Model 410.
“V2” refers to the second version of the product’s design.
2009829-001 Revision A
This Page Intentionally Left Blank
The model of the Monitor determines which parameters are in your monitor.
Please refer to applicable sections.
The content of this document including all figures and drawings is proprietary
information of GE Medical Systems Information Technologies, provided
solely for purposes of operation, maintenance or repair, and dissemination
for other purposes or copying thereof is prohibited without prior written
consent by GE Medical Systems Information Technologies, Tampa, Florida.
European Representative
GE Medical Systems
Information Technologies GmbH
Postfach 60 02 65
D-79032 Freiburg, Germany
Tel. +49 761 45 43 - 0
Fax +49 761 45 43 - 233
Table of Contents
Section 1 - Introduction
4.6.6 Temperature System Check (PRO 200 and 400) ........... 4-20
Table of Contents
1.1. Introduction ....................................................................... 1-5
Section 1. INTRODUCTION
1.1. SCOPE OF MANUAL
This service manual provides service, parts, and repair information
and functional descriptions for the DINAMAP® PRO Series Monitors,
Models 100, 200, 300, and 400. The model number identifies the vital
signs monitoring capabilities built into the unit.
WARNING
Section 1 describes this manual and tells you how to use it. Information
is also provided about how to get assistance if the unit fails to function
properly.
Section 2 provides the physical and functional characteristics of the
PRO Series Monitor.
Section 3 explains the principles of operation for the PRO Series
Monitor, including an overall system description and principles of
operation at the PWA (printed wiring board) level.
Section 4 provides information about periodic and corrective
maintenance of the PRO Series Monitor. Procedures include calibration
checks, recalibration, performance tests, and fault isolation.
Section 5 provides component information about the PRO Series
Monitor, including disassembly procedures, parts lists, assembly
diagrams, and electrical schematics.
The first time trademarks and trade names are used in a section of
this document, they are followed by “*” or “**”. A footnote will identify
the owner of the trademark or trade name.
1.6. ASSISTANCE
If the product fails to function properly or if assistance, service, or
spare parts are required, contact Customer Support. Before
contacting Customer Support, it is helpful to attempt to duplicate the
problem and to check all accessories to ensure that they are not the
cause of the problem. If you are unable to resolve the problem after
checking these items, contact Customer Support. When you call,
please be prepared to provide:
1.7. SERVICE
If your product requires repair service, call Customer Support and a
representative will assist you. Prior to returning any product for repair,
you must request a Return Authorization number. Call GE Medical
Systems Information Technologies at: 1-877-274-8456 Monday
through Friday, 8:00 a.m. to 6:00 p.m. EST, excluding holidays.
Remove the battery and all hoses, cables, sensors, and power
cords from the PRO Series Monitor before packing.
Pack only the accessories you are requested to return. Place
them in a separate bag and insert the bag and the product inside
the shipping carton.
Use the original shipping carton and packing materials, if
available.
If the original shipping carton is not available, place the product in
a plastic bag and tie or tape the bag to prevent loose particles or
materials from entering openings such as the hose ports.
Use a sturdy corrugated container to ship the product; tape
securely to seal the container for shipping.
Pack with 4 in. to 6 in. of padding on all sides of the product.
1.7.2. Insurance
Insurance is at the customer's discretion. Claims for damage to the
product must be directed to the shipper.
http://www.gemedical.com
Power Off/On
Silence Alarms
START/STOP BP
AUTO BP
SpO2
Temperature Value
Battery Power
External AC or DC power
AC Alternating Current
ADU Analog to Digital Units: The Main Board has 65,536 steps which
equates to 5V full scale (76 µV/ADU) Secondary Processor ADC
has 256 steps which equates to 5V full scale (19.351 mV/ADU)
AM Amplitude Modulation
ASSY Assembly
BP Blood Pressure
DC Direct Current
mb Millibar
Hz Hertz
OSC Oscillatory
Table of Contents
2.1. Introduction ....................................................................... 2-5
16. ON/OFF switch: Controls on/off state of PRO Monitor; push for
power on and push again for power off.
21. Rotor: Used to highlight and select items in LCD menus; if monitor
is off, pressing rotor will switch monitor on.
22. Pulse BPM display: 3-digit yellow LED shows pulse rate in beats
per minute.
25. LCD (liquid crystal display): Displays all alarms, user interface
messages, and configuration options.
Not Shown: The printer access door is located on the left side of the
monitor. Push the latch to open the door and access the paper supply
and print head. For details, see Section 5, Component Information.
Note: When using remote alarm, the PRO Series Monitor should be
considered the primary alarm source. The secondary alarm is used
for secondary purposes only.
Pin Assignments
Pin Function
1 Ground
2 Inverted TTL Transmit Data
3 Inverted TTL Receive Data
4 Fused +5 Volts
5 No connection
6 No connection
7 Ground
8 Remote Alarm
9 RS232 Request to Send (RTS)
10 RS232 Clear to Send (CTS)
11 RS232 Transmit Data (TxD)
12 No connection
13 RS232 Receive Data (RxD)
14 No connection
15 No connection
NIBP:
Includes: 1 each: Infant, Child Small ADult, Adult, Large Adult, Thigh Cuff
Includes: 2 Neo #1, 3 Neo #2, 5 Neo #3, 5 Neo #4, 5 Neo #5
Includes: 1 Infant, 1 Child, 2 Small Adult, 2 Adult, 2 Large Adult, 1 Thigh, 1 Adult Long
Includes: 2 Neo #1, 3 Neo #2, 5 Neo #3, 5 Neo #4, 5 Neo #5
Includes: 1 each: Infant, Child, Small Adult, Adult, Large ADult, Thigh Cuff
Includes: 1 each: Infant, Child, Small Adult, Adult, Large Adult, Thigh Cuff
TEMPERATURE
SpO2:
NELLCOR®:
Masimo®:
2.6. SPECIFICATIONS
This product conforms to the essential requirements of
the Medical Device Directive. Accessories without the CE
Mark are not guaranteed to meet the Essential
requirements of the Medical Device Directive.
2.6.2. Environmental
Operating Temperature +5° C to +40° C (+41° F to + 104° F)
Radio Frequency Complies with IEC Publication 601-1-2 (April 1993) Medical
Electrical Equipment, Electromagnetic Compatibility
Requirements and Tests and CISPR 11 (Group 1, Class A)
for radiated and conducted emissions.
2.6.3. Mechanical
Dimensions Height: 9.8 in (25.0 cm)
2.6.4. BP
Cuff Pressure Range 0 to 290 mmHg (adult)
(Normal operating range) 0 to 140 mmHg (neonate)
Systolic 30-245
MAP 15-215
Diastolic 10-195
Neonate
Systolic 40-140
MAP 30-115
Diastolic 20-100
Blood Pressure Accuracy Meets or exceeds ANSI/AAMI standard SP-10 (mean error ≤5
mmHg, standard deviation ≤8 mmHg)
Range
Use only IVAC probes and probe covers. The size, shape, and thermal characteristics of the probe
covers can affect the performance of the instrument. Inaccurate readings or retention problems may
occur unless IVAC probes and probe covers are used.
2.6.6. SpO2
NELLCOR SpO2
Measurement Range
SpO2 1 to 100%
Saturation
Pulse Rate
Default Settings
Sat Seconds 0
Sensor Connect/Disconnect From Patient Monitor detect attachment or disconnection of sensor from
patient within 15 seconds.
Pulse Detection Monitor will detect pulse or enter no signal state within 15
seconds of being attached to patient
Loss of Pulse Monitor will detect loss of pulse from patient and enter no
signal state within 10 seconds
Measurement Range
SpO2 1 to 100%
Saturation
Pulse Rate
*The Masimo SET®® SpO2 parameter with LNOP-Adt sensors has been validated for no motion accuracy in human blood studies on healthy
adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation
equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
**The Masimo SET® SpO2 parameter with LNOP-Adt sensors has been validated for motion accuracy in human blood studies on healthy
adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non
repetitive motion before 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory
co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population.
†The Masimo SET® SpO2 parameter with LNOP-Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies
on neonates while moving the neonate’s foot at 2 to 4 cm against a laboratory co-oximeter and ECG monitor. This validation equals plus or
minus, one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
‡The Masimo SET® SpO2 parameter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator
and Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70
to 100%. This variation equals plus or minus, one standard deviation. Plus or minus one standard deviation encompasses 68% of the
population.
LNOP ±2 digit
LNOP-ADT ±2 digit
LNOP-PDT ±2 digits
LNOP-NEO ±3 digits
LNOP-NEO PT ±3 digits
Resolution
Saturation (% SpO2) 1%
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the
amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterial
pigmentation may cause erroneous readings.
Default Settings
Table of Contents
LIST OF FIGURES
3.1 INTRODUCTION
This section provides overall theory of operation and functional
description of the DINAMAP PRO Monitor (hereinafter referred to as
PRO Monitor). The PRO Monitor comes in four different models, as
follows:
The PRO Monitor is a portable unit that receives input power from an
external AC source, external DC source, or internal rechargeable
battery.
When the ON/OFF button is pressed, the Main Board is brought out
of a sleep mode and turns on the power regulators. The power
regulators provide conditioned power from one of the input power
sources: AC Mains, External DC, or the Lead Acid Battery. The
regulated power is routed to the Printed Wiring Assemblies (PWAs)
via the cable harnesses. Once the Pro Monitor is energized, a self-
test is performed. The self-test automatically tests the main functions
of the PRO Monitor. Failure of the self-test will set the PRO Monitor
into a fail-safe mode with an audio alarm.
A reset signal to the SPO2 PWA is also provided so that the power up
sequencing is corrected. If the SPO2 circuit quits communicating to
the Main Board, the Main Board will attempt to reset the SPO2 PWA.
The Main Board backup voltage, derived from either NiMH battery or
an external power source, appears on the Main Board supervisory
circuit. The Main Board supervisory circuit generates the RESET and
HALT signal for the Primary Processor. It also, allows EXTERNAL DC
voltage to the RAM and Secondary Processor when the system is ON
or battery backup voltage to power the RAM and Secondary
Processor when the system is OFF.
If the Primary Processor fails to reset its watchdog timer, the Primary
Processor sends a low level on its watchdog output to the Pneumatic
Control Gate Array Logic (PCGAL). If the Primary Processor detects
system faults or overpressure, it will notify the Secondary Processor
of system failure condition. The Secondary Processor in turn, sends a
FAILSAFE signal to the PCGAL and Tone Generator. When the
PCGAL receives the FAILSAFE signal, it sends a DEFLATE signal to
the Pneumatic Valves to depressurize the system to atmosphere.
Once the Tone Generator receives the FAILSAFE signal, it sends an
audio frequency (ALARM signal) to the speaker. During the fail-safe
mode the following will also occur:
Patient data received from the finger sensor is filtered, amplified, and
analyzed on the SPO2 PWA. The information is sent to the Main
Board via the optically coupled electrically isolated serial connection.
The Primary Processor receives the data and distributes it to the
appropriate display. The data is also sent to the printer via the
Secondary Processor.
Appropriate valve and pump control signals are sent from the Primary
Processor to the PCGAL. The PCGAL then generates the appropriate
drive signals to the pump and valves.
The LCD assembly is equipped with a variable lighting feature for the
graphic display. The Primary Processor controls the lighting feature,
automatically. When a low or medium ambient lighting condition
occurs, the Primary Processor sends a BACKLIGHT signal to adjust
the graphic display lighting with respect to the room lighting condition.
If the lighting condition is high, the Primary Processor sends a
BACKLIGHT_1 signal to switch off the graphic display lighting.
The printer has a built-in sensor to monitor the printer paper level.
When the printer is out of paper, it sends a PAPER OUT signal to the
Secondary Processor.
Table of Contents
4.1. Introduction ....................................................................... 4-5
4.2. FUSES
The PRO Monitor power circuits incorporate five fuses. None of the
fuses are externally accessible. Two mains fuses are mounted on the
AC mains converter board, and are not serviceable by the user.
If a mains fuse blows, the Monitor will not change from battery power
to mains power when it is plugged in to an AC power source. To
confirm a fuse problem, remove the battery, plug the unit in, and turn
it on. If the PRO Monitor does not operate, the probable cause is an
open fuse. Mains fuses should be replaced by a qualified service
technician.
The exterior of the PRO Monitor may be wiped clean with a cloth
slightly dampened with mild detergents or standard hospital
bactericides.
CAUTIONS
Moisture or foreign substances introduced into the
pneumatic system will cause damage to the unit. Calibration
equipment should always be kept dry and clean.
Speaker Check
During power up, the audio system generates three short beeps,
followed, after a brief pause, by three more beeps. If the speaker
generates distorted sounds or no sound, it is faulty.
1. Power on, and wait for the PRO Monitor to initialize. Confirm the
external power indicator is lit and the battery LED is off.
3. Fit the battery and disconnect the external power supply. Confirm
the external power indicator is off; the battery LED glows yellow; and
the battery icon appears on the LCD, toggling with the time indicator.
If the battery power is low, the battery LED flashes every four
seconds and the battery icon on the LCD changes.
4. Reconnect the external power supply, and verify that the battery
LED is off.
Press Button
Selecting the Press button displays the target pressure dialog box,
pictured above, which sets the default target inflation pressure for a
BP cycle. The factory default is 160 mmHg for adults and 110 mmHg
for neonates. This is indicated by the “Auto” label at either end of the
adjustable range. Choosing “Retain Values” allows the user to preset
the number of minutes for which displayed values are retained on the
Monitor.
Selecting the Info button displays the most recent calibration dates of
the BP. Choose OK to return to the Service menu.
Selecting the Silence button displays the dialog box, pictured above,
which mutes all the alarms except the fail-safe alarm. The alarms are
disabled until the Monitor is either powered off and on again or the
The More... menu accesses more options that allow the user to
permanently change the default settings.
Main Button
Select this button to exit the Service menu and return to the Main
menu.
Trend Button
Print Button
Set BP Button
SpO2 Button
Alarms Button
Select Reset to return all the alarm limits to the default settings. The
illustration above shows the default settings.
The user can adjust the alarm limits within the ranges specified in the
table below. The high limit must be at least one step higher than the
low limit.
Select Save to save the displayed settings and return to the More…
menu. Select Cancel to abandon any changes and return to the
More… menu.
NIBP Button
The NIBP button accesses the NIBP Service menu. Use this menu to
calibrate and check the NIBP transducers and the overpressure
circuitry (Section 4.5.3).
Info Button
This button displays the date of the most recent NIBP calibration.
Select OK to return the display to the Service menu.
System Button
This button accesses the System menu. The options on this menu set
the display language, communications protocols, test the EEPROM,
and print the error log (Section 4.5.4).
Silence Button
Yes mutes all patient alarms until the PRO Monitor is powered off and
on again or the alarm Silence button is selected again to enable the
alarms. A confirmation menu appears (pictured above) on the display.
Selecting either Yes or No exits the menu. If silence is confirmed, the
alarm Silence button on the front panel illuminates momentarily.
The options on the NIBP menu, pictured above, are used to calibrate
and test BP functions of the PRO Monitor.
Check Button
With no hose attached, the systolic and diastolic displays will indicate
“000.” See Section 4.7 for a more detailed description of this mode
and the calibration procedure.
Cal Button
This button initiates the calibration procedure. See Section 4.7 for a
step-by-step description of this procedure.
This button functions in a similar manner to the Check button, but the
overpressure point is set to the neonate value. See Section 4.7 for a
more detailed description of using this function.
OK Button
The System menu, pictured above, accesses the options that modify
the basic configuration of the PRO Monitor, test the EEPROM, and
print the error log.
Language Button
Comms Button
The Standard check box sets the host comms protocol to standard
format, and the baud rate at 9600 bps.
Selecting the 1846 button sets the comms protocol to 1846 format,
and the baud rate to 600 bps.
The baud Rate can be set to 300, 600, 1200, 2400, 4800, 9600, or
19200.
This button initiates a test of the read and write functions on the
EEPROM. The test is repeated 20 times. The results are displayed on
the LCD and recorded in the error log.
This button generates a printout of the last 20 system errors that have
occurred on the unit. The system and secondary processor software
versions are also listed on the error log.
Test Equipment
Test Conditions
Pulse value
Saturation value
6. Verify the sound has stopped but the error remains displayed.
Pulse Value
Saturation value
Signal Strength bar Graph
Equipment required:
1. Secure the Adult-size cuff to the 12-foot air hose. Attach to the
PRO Monitor. Secure the BP cuff around a rigid, unbreakable
object that measures at least 2” in diameter.
The leak test sequence closes both internal valves and initiates the
pneumatic pump. The Monitor will self-pressurize the pneumatic
setup to approximately 200 mmHg. After 5 seconds the target
Figure 4-1
4. From the Service menu, select the NIBP button, then select the
Popoff button.
7. Switch off the PRO Monitor and disconnect the calibration kit.
Set the probe simulator to 98.6 and verify that the LCD temperature
display reads 98.6°F ±1.0°F. The numbers on the temperature LED
displays should be flashing at this point, indicating the monitor is in a
real-time monitor mode.
±0.2°F. Next. press BROKEN PROBE on the simulator, verify that the
LED display on the Monitor changes to four dashes, indicating a fault
condition.
If the PRO Monitor temperature system does not pass these tests,
the Main PWA needs to be replaced.
Note: If the PRO Monitor displays the language choice menu, select
a language. Select OK and reboot the system.
13. Apply the pressure (200 mmHg) and HOLD the pressure. Press
Accept to continue or Cancel to quit this procedure.
The zero reading and the 200 mmHg reading are the only points used
for calibration. It is important to ensure that the correct pressures are
applied at these two points.
If the calibration fails, turn off the Monitor, check the calibration
equipment and repeat the process. Refer to a GE Medical Systems
Information Technologies service representative if calibration is still
unsuccessful. When the calibration is successful, the following will be
displayed on the LCD, “Wait...Storing values,” then after a few
seconds the LCD displays, “Turn Monitor off.” The PRO Monitor is
now accurately calibrated and can be switched off.
Figure 4-2
e. Select NIBP.
e. Select NIBP.
9. The unit will display three zeros in the systolic and diastolic
windows.
11. Set the cuff pressure to 157 mmHg ±2 mmHg using your
connected NIBP test setup.
12. The air pressure within the system may fluctuate initially. Verify
pressure with your external manometer.
13. Adjust VR2 clockwise until the pressure is released and the
Monitor displays Systolic: 000 & Diastolic: 000.
Ground Leakage
Clip, Retaining
Spacer, LED
Battery Support
Actuator Switch
Pole Clamp
Filter, 40 Micron
Rotor Shaft
Spring
Spacer, push-mount
Fascias
Description Part Number
English, US, DP410 2012419-001
RAMCSB[L]
OE[L] OE[L]
LWE[L] LWE[L]
+5VD VRAM
C71 VRAM C73
+5VD L1 +5VP
HZ0805E601R U10 U12
AM29F040B 32 10nF KM681000B 32 10nF
VCC VCC
A[1:19]
22
CS1
C105 C2 C57 C63 C64 22
CE
30
CS2
47uF 100NF 22uF 10nF 10nF 31 29
25V 25V 16V WE WE
24 24
OE OE
+5VD A1 12 13 D0 A1 12 13 D0
A0 DQ0 A0 I/O1
A2 11 14 D1 A2 11 14 D1
UWE[L] A1 DQ1 A1 I/O2
A3 10 15 D2 A3 10 15 D2
R114 +5VP A2 DQ2 A2 I/O3
10K A4 9 17 D3 A4 9 17 D3
A3 DQ3 A3 I/O4
C65 C68 A5 8 18 D4 A5 8 18 D4
A4 DQ4 A4 I/O5
D1 A6 7 19 D5 A6 7 19 D5
A5 DQ5 A5 I/O6
FAILSAFE[L] C66 10nF 10nF C69 A7 6 20 D6 A7 6 20 D6
D[0:15]
A6 DQ6 A6 I/O7
A8 5 21 D7 A8 5 21 D7
BAS16 A7 DQ7 A7 I/O8
10nF C67 C70 10nF A9 27 A9 27
A8 A8
A10 26 A10 26
R77 A9 A9
49K9 10nF 10nF A11 23 A11 23
83 62 60 36 13 7 A10 A10
A12 25 A12 25
WDOG[L] 3 94 A11 A11
VCCD1
VCCQ2
VCCQ1
VCCA1
VCCS1
VCCP1
WDOG CSO A13 4 A13 4
D2 4 95 A12 A12
CHARGER_ON EX_DC TOUT2 /PB6 CS1 A14 28 A14 28
5 96 A13 A13
TIN2/PB5 CS2 A15 29 A15 3
UNFIL_0VP[L] BAS16 6 97 A14 A14
TIN1/PB3 CS3 A16 3 A16 31
LATCHED_0VC +5VP 8 A15 A15
WEH/UD5 A17 2 A17 2
LATCHED_0VP 9 93 A1 A16 A16
WEL/LD5 A1 A18 30
+5VP 11 92 A2 A17
D_FROM_TEMP R190 AS A2 A19 1 VSS
12 91 A3 A18
C96 10K OE/R/W A3 16
15 90 A4
IPL0/IRQ1 A4 VSS
16 U43 89 A5
22pF IPL1/IRQ6 A5 16
MC68LC302
DCLK_X2 R88 17
IPL2/IRQ7 A6
87 A6
R51 18 86 A7
EXTAL A7
180R 10M CS2[L]
XL2 19
XTAL A8
85 A8
4.915MHz 20 84 A9
C97 R155 +5VP CLK0 A9
+5VD 21 82 A10
DISCPU A10
22pF 1K0 22
BUSW A11
81 A11
23 80 A12 +5VD VRAM
R111 C1 C45 GNDSYN A12 C72 C74
24 79 A13 VRAM
10K XFC A13
100nF 25V 2n2 25 78 A14 U11 10nF U13 10nF
VCCSYN A14
HALT[L] 26 77 A15 AM29F040B 32 KM681000B 32
HALT A15
RESET[L] VCC VCC
27 75 A16 22
RESET A16 CS1
28 74 A17 +5VD 22 30
DTACK A17 CE CS2
HOST_RTS[L] 29 73 A18 31 29
RTS1 A18 WE WE
+5VD R110 32 72 A19 24 24
RCLK1 A19 R113 OE OE
33 10K
TCLK1
HOST_TXD
10K
34 37 A1 12 13 D8 A1 12 13 D8
TXD1 PA7 A0 DQ0 A0 I/O1
HOST_RXD 38 47 SPRXD A2 11 14 D9 A2 11 14 D9
CD2 /PA6 PA8 A1 DQ1 A1 I/O2
HOST_CTS[L] 39 46 SPTXD A3 10 15 D10 A3 10 15 D10
RTS2 /PA5 PA9 A2 DQ2 A2 I/O3
40 30 SPCLK A4 9 17 D11 A4 9 17 D11
CTS2 /PA4 PA10 A3 DQ3 A3 I/O4
41 35 A5 8 18 D12 A5 8 18 D12
TCLK2/PA3 PA12 A4 DQ4 A4 I/O5
SPO2_RESET EEPROM_CS 43 A6 7 19 D13 A6 7 19 D13
RCLK2/PA2 A5 DQ5 A5 I/O6
D_TO_SPO2 44 2 A7 6 20 D14 A7 6 20 D14
TXD2/PA1 PB8 A6 DQ6 A6 I/O7
D_FROM_SPO2 45 1 A8 5 21 D15 A8 5 21 D15
RXD2/PA0 PB9 A7 DQ7 A7 I/O8
48 99 A9 27 A9 27
RXD1 PB10 A8 A8
49 98 A10 26 A10 26
CTS1 PB11 A9 A9
50 A11 23 A11 23
CD1 A10 4 1 30 A10
51 D0 A12 25 A12 25
D0 A11 5 A11
52 D1 A13 4 29 A13 4
D1 A12 A12
53 D2 A14 28 A14 28
D2 A13 A13
54 D3 A15 29 U10 A15 3
SECP_CS D3 A14 U11 A14
56 D4 A16 3 A16 31
SECP_HS D4 A15 A15
57 D5 A17 2 A17 2
D5 A16 A16
58 D6 A18 30 13 21
D6 A17
59 D7 A19 1 14 20 VSS
D7 A18
63 D8 16
D8
64 D9 VSS
D9
65 D10 16
PNEU_RESET D10
66 D11
D11
68 D12
REMOTE_ALARM[L] D12
69 D13
D13
70 D14
+5VD D14
71 D15
D15
GNDD1
GNDD2
GNDQ2
GNDQ1
GNDA2
GNDA1
GNDS1
GNDS2
GNDP2
GNDP
1 5
4 CS3[L]
2 88 76 67 61 55 42 31 14 10
100 76
U2 100
3 BU4S81 1 75
U43 SPRXD
REMOTE[L]
SPTXD
ADC_BUSY[L]
25 51 SPCLK
26 50
LCD_BACKLIGHT
+5VD
U8
ST93C46A 8
VCC
3 4
D Q
2
C
1
S
6
ORG
PL8
VSS
R112
CONNECTOR
NOT INSTALLED
3
4
R191
30K
1% Main Board Assembly
P/N: 2008238-001 C
10K
Sheet 1 of 9
VRAM
U28 64 41
65 40
MAX809R 3
VCC
U5 PDG[0:9] +VLED
R177
2 C95 80
RESET 25
12pF C116
4K7
1 24
GND R141 15pF
VRAM VRAM
1 100K R124 R115 R117 R116 R121 R120 R119 R118 R123 R122
FAILSAFE[L] +5VD XL3 R165 C75 C76 XL1 10K 10K 10K 10K 10K 10K 10K 10K 10K 10K
4.0MHz 1M0 32.768kHz
10nF 10nF Q37 Q28 Q30 Q29 Q34 Q33 Q32 Q31 Q36 Q35
C115 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188 2SB1188
R104
R156 18K U5
C94 15pF R98 R89 R91 R90 R95 R94 R93 R92 R97 R96
1K0 H8/3724
DG0
DG1
DG2
DG3
DG4
DG5
DG6
DG7
DG8
DG9
12pF 76 57 180R 180R 180R 180R 180R 180R 180R 180R 180R 180R
LDR
10 AVCC VCC 8
OSC1 X1
PDG0
PDG1
PDG2
PDG3
PDG4
PDG5
PDG6
PDG7
PDG8
PDG9
11 7
LDR1 C106 OSC2 X2
R52 MPY54C569 47uF 6 Q19 Q10 Q12 Q11 Q16 Q15 Q14 Q13 Q18 Q17
TEST
10M 25V 12 32 PS13
DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA DTC143ZKA
RES P50
33 PS12
P51
SECP_HS 74 34 PS11
PA0 P52
75 35 PS10
PA1 P53
H8_RES[L] 36 PS9 DG[0:9]
DG0
P54
DC_INPUT 77 37 PS8
P00 P55
78
P01 P56
38 PS7 DO NOT STUFF FOR PRO110/210 DO NOT STUFF FOR PRO110/310
DC_BATTERY 79 39 PS6
P02 P57
80
P03 LED22 LED23 LED21 LED28 LED27
PTH 1 41 PS5
P04 P60 LMG53W LMR53W-D LUY53C LMR53W-D LUY53C
PUMPC 2 42 PS4 DC_SUPPLY F BPM ALARM SPO2
P05 P61 LED20 LED24 LED25 LED26
BATTERY_C 3 43 PS3
P06 P62 LUY53C LMR53W-D LMR53W-D LUY53C
ANALOG_V LED5 LED4 LED9
4 44 PS2 BATTERY C ALARM MAP
S5A
P07 P63 DMR14A-1 DMR14A-1 DMR14A-1
SECP_CS 45 PS1
S0
S1
S2
S3
S4
S5
S6
S7
P64 S8 7 S0 7 S8 7
VRAM VRAM VRAM 13 46 PS0 A A A
IRQ0 P65 S9 6 S1 6 S9 6
14 47 PDG0 B a B a B a
IRQ1 P66 S10 4
C
S2 4
C
S10 4
C
DO NOT STUFF FOR PRO110/310 DO NOT STUFF FOR PRO110/210
R126 R125 15 48 PDG1 f b f b f b
IRQ2 P67 S11 2 S3 2 S11 2
10K 10K 16 D D D
IRQ3 S12 1 g S4 1 g S12 1 g
5 E E E
ROTOR[L]
POWER[L]
1
4
17
IRQ4 P70
49 PDG2
SYSTOLIC S13 9
F e c
S5 9
F e c
S13 9
F e c
2 18 50 PDG3
U3 IRQ5 P71 S14 10 S6 10 S14 10
3 G G G LED13 LED14 LED15
BU4S81 19
EVENT P72
51 PDG4
d d d
S15 5 S7 5 S15 5 DUR09A DUR09A DUR09A
EXT_DC_INPUT[L] 40 52 PDG5 DP DP DP DP DP DP
P17 P73 S8 10 S0 10 S8 10
ALARM_FREQ 53 PDG6 A A A
P74 3 3 3 S9 9 S1 9 S9 9
ALARM0 23 54 PDG7 COM COM COM B a B a B a
P30 P75 8 8 8 S10 8 S2 8 S10 8
ALARM1 22 55 PDG8 COM COM COM C C C
P31 P76 f b f b f b
S11 5 S3 5 S11 5
DG9
DG9
DG8
ALARM2 21 56 PDG9 D D D
P32 P77 S12 4 g S4 4 g S12 4 g
ALARM3 20 E E E
P33 S13 2 S5 2 S13 2
F e c F e c F e c
P80
58 PMOTORA[H]
LED10 LED11 LED12 S14 3
G
S6 3
G
S14 3
G
MAP
PS14 31 59 PMOTORA[L]
P40 P81 DMR14A-1 DMR14A-1 DMR14A-1 S15 7 d S7 7 d S15 7 d
PS15 30 60 PMOTORB[H] DP DP DP DP DP DP
P41 P82 S0 7 S8 7 S0 7
PSTB1 29 61 PMOTORB[L] A A A
P42 P83 S1 6 S9 6 S1 6 1 1 1
PSTB2 28 62 PLATCH[L] B a B a B a COM COM COM
P43 P84 S2 4 S10 4 S2 4 6 6 6
PSTB3 27 63 PDATAIN C C C COM COM COM
P44 P85 f b f b f b
S3 2 S11 2 S3 2
DG6
DG6
DG5
PSTB4 26 64 PCLOCK D D D
P45 P86 S4 1 g S12 1 g S4 1 g
PSTB5 E E E
PSTB6
25
P46 P87
65 PDATAOUT
DIASTOLIC S5 9
F e c
S13 9
F e c
S5 9
F e c
24
P47 S6 10 S14 10 S6 10
66 G G G
PWM S7 5 d S15 5 d S7 5 d
67 DP DP DP DP DP DP
SCK1
68
SI1 3 3 3
69 COM COM COM
SO1
R178
R181
R182
R185
R186
4K7
4K7
4K7
4K7
4K7
8 8 8
70 COM COM COM
SCK2
DG8
DG7
DG7
71
CS
LED1 LED2 LED3
R179
R180
R183
R184
R187
4K7
4K7
4K7
4K7
4K7
72
SO2 DMY14A-1 DMY14A-1 DMY14A-1
73
UD S0 7 S8 7 S0 7
A A A
S1 6 a S9 6 a S1 6 a
AVSS VSS B B B
5 9 S2 4 S10 4 S2 4
C C C
HEAD_UP f b f b f b
S3 2 S11 2 S3 2
D D D
PAPER_OUT LED16 LED17 LED18 LED19 S4 1 g S12 1 g S4 1 g
E E E
DUR09A DUR09A DUR09A DUR09A
PAPER_OUT_CONTROL
S5 9
F e c
S13 9
F e c
S5 9
F e c BPM
S0 10 S8 10 S0 10 S8 10 S6 10 S14 10 S6 10
A A A A G G G
5VDIGITAL_ENABLE S1 9 S9 9 S1 9 S9 9 S7 5 d S15 5 d S7 5 d
B a B a B a B a DP DP DP DP DP DP
PRINTER_LED_ENABLE S2 8 S10 8 S2 8 S10 8
C C C C
CHARGER_ON f b f b f b f b
S3 5 S11 5 S3 5 S11 5 3 3 3
D D D D COM COM COM
SPRXD
TEMP S4 4
E
g S12 4
E
g S4 4
E
g S12 4
E
g 8
COM
8
COM
8
COM
S5 2 S13 2 S5 2 S13 2
DG3
DG2
DG2
F e c F e c F e c F e c
SPTXD S6 3 S14 3 S6 3 S14 3
G G G G
SPCLK S7 7 d S15 7 d S7 7 d S15 7 d
DP DP DP DP DP DP DP DP
DG4
DG4
DG3
S9 6 a S1 6 a S9 6 a
B B B
S10 4 S2 4 S10 4
C C C
f b f b f b
S11 2 S3 2 S11 2
D D D
S12 1 g S4 1 g S12 1 g
PS[0:15] DO NOT STUFF FOR PRO110/310 E E E
S13 9
F e c
S5 9
F e c
S13 9
F e c SP02
S14 10 S6 10 S14 10
G G G
S15 5 d S7 5 d S15 5 d
S[0:15] DP DP DP DP DP DP
3 3 3
S5A
S10
S11
S12
S13
S14
S15
COM COM COM
S0
S1
S2
S3
S4
S5
S6
S7
S8
S9
8 8 8
COM COM COM
R33 R34 R36 R35 R40 R39 R49 R38 R37 R48 R47 R46 R45 R44 R43 R42 R41
DG1
DG1
DG0
22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R 22R
PS11
PS12
PS13
PS14
PS15
PS0
PS1
PS2
PS3
PS4
PS5
PS6
PS7
PS8
PS9
3
VCC 11
R65 R73 R66 A0
2 10
VOUT A1
1 0R0 3K3 0R0 9
GND C49 A2 +5VD
R85 C102 NOT INSTALLED R10
6
CAP 6 C16 C15 510R
47NF EN
+12VA 1K0 680P0 R12
100nF 25V 100nF 25V 510R
0.5V = 0mmHg R105 V-
5 OPA2340 8
4.0V = 300mmHg 16 38
18K U26 7
TF = Vs * ((P * 0.00233) + 0.1) R67
6 VCC1 VCC2
REF-02 5 SWBUS7 3 PROBE_SW
0R0
SWBUS6 41
TRIM NOT INSTALLED ALARM_SILENCE_SW
SWBUS5 15 8
DCLK
2 6 5.0V START_STOP_SW DCLK_OUT
VIN U17 VOUT SWBUS4 42 12 CS_ADC[L]
STAT_SW CS_ADCL
C11 C100 SWBUS3 43 10 ADC_BUSY[L]
C9 GND POWER_SW ADC_BUSYL
C108 100NF 4 SWBUS2 9
CHA_SW
47uF PT2 100nF 25V 470NF 24 D7
25V SWBUS1 37 DATA7
25V XFPM-050 CHB_SW U19 18 D6
SWBUS0 44 CY37064 DATA6
3 ROTOR_SW 22 D5
VCC 7 DATA5
R74 SERIAL_DIN 23 D4
2 5 OPA2340 36 DATA4
VOUT RESETL V2_CHIP 21 D3
DATA3
GND
1 3K3 U27 7 A7 29
ADDR7
R68 DATA2
25 D2
R86 C103 6 A6 2
ADDR6
6 C50 34 D1
CAP A5 26 DATA1
0R0 ADDR5
1K0 680P0 30 D0
NOT INSTALLED A2 11 DATA0
47NF ADDR2
A1 31 20 LCD_CS
ADDR1 LCD_CS
C46 A3 1 19
ADDR3 PNEU_CSL
27 44 34
WEL
1nF 40 1 33
CS3L 32
13 MUXADDR2
CS2L 35 D[0:7]
PUMPC 4
CLK_IN
MUXADDR1 U19
33
MUXADDR0
GND1 GND2 GND3 GND4
11 23
SWBUS[0:7] 6 17 28 39
12 22
MUXADDR2
MUXADDR1
MUXADDR0
LWE[L]
CS3[L]
CS2[L] A[1:19]
DCLK_X2
U30
HC259
A1
4 12V_ON
+12VA Q0
13 5 DEFLATE
D Q1
PNEU_CS[L] 14 6 DUMP[L]
G Q2
R1 Q3
7 PUMP_ON
10R A2 1 9 D_TO_TEMP[L]
+12VA S0 Q4
C3 C4 A3 2
S1 Q5
10 LCD_BACKLIGHT_1
A4 3
S2 Q6
11
100nF 25V +5VD 12 ADULT[L]
100nF 25V R83 C8 Q7
CLR
21K
R64
3 0.1% 15
8 TLC2272 R108 R76 100nF 25V
10K 1 5 TLC2272 U32 RESET[L]
0.1% U39 LM393
2 887R 2K74 U39 7 3 8
4
R146 R50 0.1% 0.1% 6 1
11K 90K9 R149
2
1% 0.1% 4
UNFIL_0VP[L]
30K
1%
C5
100NF
25V
+5VD
VR1 +5VD
10K D3 R53
CW R87
R161 VR2 1K0 10M
1.5K NEO OVP ADULT OVP R128
10K BAS16
1% CW 10K
U32
R166 R127
5 LM393
FLT_0VP[L]
Q7 330K 10K 7
2N7002 C59 6
+5VD
+5VD 1uF
R59 16V
47K
R84
68K
1%
P/N: 2008238-001 C
Sheet 3 of 9
+5VD
+5VD
C77
R72
R168 10nF 1R0
+5VD 20K C52
R75
U22
3
LM358
8
1
2
4
R169 +5VD
20K
R99
7K5
1% PL1
PDATAOUT 1
PLATCH[L] 2
PCLOCK 3
PSTB1 4
PSTB2 5
PSTB3 6
PRINT HEAD
PTH 7
8 CONNECTOR
PSTB4 9
PSTB5 10
PSTB6 11
PDATAIN 12
+5VD
PL10
PMOTORA[H] 1
PMOTORA[L] 2
PMOTORB[H] 3
PMOTORB[L] 4
PAPER_OUT 5 PRINT MOTOR / SENSOR PCB
6
CONNECTOR
7
PAPER_OUT_CONTROL 8
HEAD_UP 9
C22
100NF
25V
3 +5VD 1
2
+5VD
3
C25 Q1
SI9956DY 7 4
+5VD 8 R61 5
100nF 25V
U42 2 10K 6
+5VD PALCE16V8 20
R173 1
1 VCC
20K CLK
11 U15
OE
DEFLATE_VALVE 5 6
DO NOT POPULATE
+12VP +5VD
UNFIL_0VP[L] 17 15 +5VD C32
FLT_0VP[L] 8
I/O5 I/O3
16 DUMP_VALVE
HC14 J1
I7 I/O4
LATCHED_0VC 12 9 OVERCURRENT R142 100nF 25V
I/O0 I8 +12VP 1
LATCHED_0VP 13 14
100K PROBE1
I/O1 I/O2 2 U15
12V_ON 3
I2 3 SEC_ALARM 1 2 1
DEFLATE Q2
4
I3 R170 WD1
SI9956DY 5 Q5
DUMP[L] 5 DTA143XKA HC14 2 KBS-20DB-4P
I4 6
PUMP_ON
20K
6 4 5
I5 PUMP_SWITCH
PNEU_RESET 7
I6 Q8 6 Switch is closed when
FAILSAFE[L] 2
I1
3 1 19 3 SI9410DY 7 probe is in the holder
18
4 8
GND U42 4
10
8 14 2 DO NOT STUFF FOR PRO110/310 +5VD
9 13
R150 3
+5VD
30K
1%
R134
+5VD +5VD 10K
C26
100NF RESET[L]
25V U15
R106 R172
18K 20K RESET2[L] 3 4 Q20
U23 DTC143ZKA +5VD
C62 LM392 HC14
3 8 U23
R102
1uF 1 5 LM392
R135
2.63V 2
4
1.25A = OVERCURRENT 3K01 7 10K
6
R8 C28 HALT[L]
0R1 100NF
25V
R148 Q21
R143 DTC143ZKA
100K 200K U15
9 8
R171 C27
20K 100NF R131
10K HC14
25V
U15
11 10
HC14
PUMPC
DTA143XKA is a digital transistor with a 4K7 base resistor and a 10K base pull-down resistor Schematic, PRO Monitor Series 1-400
DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor
Main Board Assembly
P/N: 2008238-001 C
Sheet 5 of 9
PL7
VDC_OP
U38
MIC5203-4.7 HOST_TXD 1
3 4 HOST_RXD 2
IN VOUT
HOST_CTS[L] 3 HOST COMMS
1
GND HOST_RTS[L] 4
C33
100NF
25V
2
EN
C61
1uF REMOTE_ALARM[L] 5 CONNECTOR
16V 6
NORMAL
JP1
D10
R6
D17
3K3
BAS16 1 2 3
+12VP
D9
R80
C79
BAT54C VRAM
10nF
3K3
BAS16
D21 C87
BZX84C B1
5V1 3V6 -17VLCD
10uF
20V PL9
+VLED
+5VD RESET2[L]
1
VRAM 2
C35 C83
10nF 3
U18 +5VD
100nF 25V ADM691A 5 1
C34 4
R138
10K 5
3
VCC
BATTON VBATT
2
VOUT
100nF 25V +5VD 6 POWER
RAMCSB[L]
RAMCS[L] 13 12 GND 7
7
CEIN
OSCIN LOWLINE
CEOUT
6
U14 +5VD
ANA- 8 CONNECTOR
8 15 1 ANA+ 9
OSCSEL RESET 3
11 16 2 VDC_OP 10
WDI RESET R174
DC_INPUT 9
PFI WDO
14 HC132 20K
10 +5VD
PFO
GND FAILSAFE[L]
4 C84 C81
R139 10nF 10nF
10K U14
4 C101 C85 C82 C110 C80
6 Q24 10nF 10nF 47uF 10nF
5 DTC143ZKA 25V
HC132 33nF
R140
U14 10K
VRAM U15 9 U14
8 12
13 12 10 11 SEC_ALARM
13
HC132
R109 HC14 HC132
47K WDOG[L]
EXT_DC_INPUT[L]
Q23
DTC143ZKA
PL4
DC_INPUT 1
DC_BATTERY 2
3
VBC 4 POWER MONITOR
BATT_OFF 5
R144
CHARGER_ON 6 CONNECTOR
PRINTER_LED_ENABLE 7
100K
5VDIGITAL_ENABLE 8
+5VD BATTERY_C
SK3 12V_ON 9
C36 R63
1 24K
100nF 25V 2
3
-17VLCD 4
LWE[L] 5
OE[L] 6
LCD_CS 7
A4 8
+VLED PL6
9
RESET[L]
D0
10
LCD 1
LCD BACKLIGHT
11
2
D1
D2
12
CONNECTOR CONNECTOR
13
D3 14 R70
D4 15
2R7
1.3W
D5 16
Q3
SI9956DY 7 C86
D6 17
8
10nF
D7 18
LCD_BACKLIGHT 2
19
D[0:15]
20 1
R71
2R7
1.3W
Q3
SI9956DY 5
6
LCD_BACKLIGHT_1 4
DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor
C38
+5VIT
R15 100nF 25V
10K
17 1 0VITA
AVDD DVDD 6
5
V_THERM 11 20 D_TO_T_ADC TEMP PGM CONNECTOR
AD0 DIN 4
12 21 D_FROM_T_ADC NOT INSTALLED
+5VIT AD1 DOUT 3
13 22 SCLK_TO_T_ADC
AD2 SCLK 2
14
R157
AD3
DRDY-0
23 DRDY_FROM_T_ADC 1 ISOLATED SIDE ISOLATION BARRIER +5VD NON-ISOLATED SIDE
1K0 15 19
AINCOM CS-0
U31 +5VIT
0VITA
J4
9 Q6 D_TO_TEMP[L]
REF+ ADS1240 +5VIT DTA143XKA
10
REF- R21
6
U36 DSYNC-0
TL431 CLK_TADC 3 18 10K C37
1 X_IN POL
4 24 U25
X_OUT BUFEN R22 R159
8 7 100nF 25V 4N35
PDWN-0 1K0
5 10K 0VITA
RESET-0 5 1
2 3 6 7 2 C AN
C54 DGND1 6
16 8 B
10uF AGND DGND2 4 2
16V E CTH
0VITA
0VITA
0VITA 0VITA R20
+5VIT 10K
+5VD
R158
100
+5VIT
0VITA
R163 R16
U6 10K
4N35
R23 D11 13K7
10K BAS16
0.1% 1
AN C
5 D_FROM_TEMP
+5VIT 6
L5 R107 B
R17 2
CTH E
4
0VITA 0VITA
3
+5VIT R19
PWM_HTR
10K
R18
D12 MON_HTR
BAS16
10K
L4
C40
BLM11B221SB Q4 100NF
25V
D14 SI2306DS
BAS16 3
1 0VITA
R24
10K EN_HTR
R153
0VITA 499K
C93
2U2
25V
C41
100NF R152 0VITA
25V 499K
Schematic, PRO Monitor Series 1-400
0VITA 0VITA
L7 D18 MIC39101-5V
2 1 2 VIN VOUT3
+5VIT
3K01 16 G4 G2 8
7 SG PG 10 SP_DG
SP
ISOLATION BARRIER
SP SP
C56 C127
C55 C122 R29 10uF 100NF
10uF 100NF MIC5270-5.0 16V 25V -5VSPO2
16V 25V 4K99 2 GND
D20 U4 -VOUT4
1 2 5 -VIN
BYG22
DO NOT POPULATE
1N0 1N0
3K15V 3K15V
0VITA
1N0 1N0
3K15V 3K15V
SP
+5VSPO2
FL2 2
RCAL_1 3 2 RCAL_2
MASIMO
J2 D27
1 DIGICAL_1 4 1 DIGICAL_2
DET + A BAV199
2 Z>200/100MHZ 3
DET - C
3
INNER SHIELD
4
INNER SHIELD 1
5
TO J1 ON MASIMO MS-5 N/C NC
6
OUTER SHIELD
7
IR LED CATH SP
8
OUTER SHIELD
9
RED LED CATH
10
N/C NC
ISOLATED SIDE
+5VSPO2
HOST INTERFACE CONNECTOR
U7 R27
+5VD -5VSPO2 +5VSPO2
4N35 10K
NELLCOR
J6
R101 1 5 1 C-LOCK
AN C
GND
249 B
6 2
2 4 3 GND
CTH E
4 RESET
SP 5 GND
SPO2 RESET SPO2_RESET Q26
DTC143ZKA 6 +12V (N/A)
7 ANALOG OUT (N/A)
TO J2 ON NELLCOR MP506
8 TRANSMIT (MP506 OUT)
+5VSPO2
9 RECEIVE (MP506 IN)
+5VD +5VD 10 +5 ANALOG
11 CTS (MP506 IN)
C113 U24
12 GND
R25
HCPL-2601 R82 13 +5V DIGITAL
10K 100nF 25V 8
VCC
3K01 14 GND
2
7 AN
VE Q25 R162
D_FROM_SPO2 6 DTC143ZKA 1K0
DATA FROM SPO2 VO 3
5 CTH
GND
SP SP
R9 J3
MASIMO
499 1 +5V
2 GND
+5VD +5VSPO2
SP 3
4 GND
U9
5 TO J3 ON MASIMO MS-5
HCPL-2601 C114
DATA TO SPO2 D_TO_SPO2 Q27 6 GND
R26
DTA143XKA
Logic
R100 2 VCC
8 100nF 25V 1K0 7 RESET
AN 7 SP 8
VE
249 VO
6 9 RECEIVE (MS-5 IN)
3
CTH 5 10 TRANSMIT (MS-5 OUT)
GND
SP SP