Documente Academic
Documente Profesional
Documente Cultură
Intensive
Care Medicine
© Springer-Verlag1985
Accepted: 25 September1984
Abstract. The effects of two inspiratory flow wave- blood oxygenation might occur. Some of this contro-
forms (WFs), decelerating and constant have been versy may be related to the fact that in these studies,
studied in 14 patients undergoing intermittent posi- changing the inspiratory flow waveforms produced
tive-pressure ventilation (IPPV). With tidal volume changes in tidal volume, frequency and I/ E ratio. The
(VT), inspiratory time, inspiratory-expiratory (I/E) principle objective of the present study therefore, was
ratio and frequency being kept constant, the deceler- to determine whether the profile of the inspiratory
ating waveform produced statistically significant re- flow waveforms could influence gas exchange in pa-
duction of peak pressure, total respiratory resistance, tients on IPPV, when the tidal volume, minute vol-
work of inspiration, ratio of dead space to tidal vol- ume, I/ E ratio and frequency are kept constant. It
ume (VD/VT) and alveolar-arterial gradient for oxy- should be then possible to assess in isolation the effect
gen (A-a)PO2. There was significant increase in total of different inspiratory flow waveforms on gas ex-
static and kinetic compliances and PaO2, with no sig- change.
nificant changes in PaCO2, in cardiac output (CO)
and in other haemodynamic measurements.
constant flow in the range of 0 - 6 0 1/min against a Table 1. Patients' sex, age and diagnosis
previously calibrated rotameter.
Patients Sex Age Diagnosis
3. The ventilator. This is a volume-cycled, flow
regulated ventilator (BOC, Medishield IMV Pneumo- 1 F 68 Respiratory failure secondary to acute
tron). The accurate measurement of tidal volume and mycoplasma pneumonia
minute volume are the key functions on which this 2 M 58 Neurofibromatosis, bilateral broncho-
ventilator relies for its operational performance. To pneumonia
3 F 53 Respiratory failure due to pneumonia
measure these volumes, this ventilator uses two flow 4 F 64 Acute on chronic respiratory failure
transducers, one on the inspiratory limb and the other 5 F 32 Barbiturate overdose
on the expiratory limb. These two transducers are the 6 M 20 Sickle cell lung disease
same as those of the electronic respirometer [8] and 7 M 62 Adult respiratory distress syndrome
are relatively insensitive to the density, viscosity, tem- 8 M 24 Adult respiratory distress syndrome
9 M 59 Acute on chronic respiratory failure
perature and humidity of the gases they measure. The 10 F 67 Acute on chronic respiratory failure
two transducers are connected via a feedback system 11 M 27 Adult respiratory distress syndrome
that compares the inspiratory VT with the expired 12 M 63 Adult respiratory distress syndrome
VT. 13 F 71 Cardio-respiratory arrest
14 F 60 Systemic lupus erythematosus, liver
To test the accuracy of VT delivery, the ventilator
cirrhosis, septicaemia and
was connected to a standard lung model [7] which pneumothorax
consisted of two 20-1 bottles, each with a compliance
of 0.018 l/cm H 2 0 and a combined compliance of
0.037 1/cm H20, and a circuit resistance of 10.47 cm
H20/1 per second. The Fleisch pneumotachograph
constant. In two of these patients on separate occa-
with a differential pressure transducer was inserted
sions and in three further patients the sequence was
between the Y-piece of the ventilator and the two
decelerating, constant and then decelerating.
bottles. Another pressure transducer was connected
to the third port of the Fleisch pneumotachograph.
Constancy of tidal volume (VT) and minute volume
The Wright spirometer and the electronic respirome-
(11). In order to compare any effects of changing the
ter were connected to the expiratory port of the venti-
inspiratory WFs, it was considered essential to main-
lator. Different combinations of compliances and re-
tain the constancy of both VT and V in each patient.
sistances were used. VT was changed from 0.4 1 to 1.2
Constant WF produces peak and mean flows of
1 in steps of 0.2 1. With each VT different ranges of
the same value whilst the decelerating WF results in
peak flow were used (10, 20, 30 . . . . up to 80 1/min).
the peak flow being greater than the mean flow.
Therefore, in order to maintain frequency and (r
Patients constant it was necessary to increase the peak flow
Fourteen patients (Table 1), 7 males and 7 females when the decelerating WF was in use. Expired VT was
with an age range of 2 0 - 7 1 years (mean
52.00_+ 17.95) were studied. They were in respiratory
failure, intubated or tracheotomized and required 1oo7
IPPV as part of their therapy. 'llmin |
Artificial ventilation
Flow~ N,. "/ 50
Ventilator. All patients were ventilated with the IMV
P n e u m o t r o n ventilator (see equipment testing) which
allowed variation in inspiratory flow waveforms (Fig.
1) [7]. Standard connections and tubings were used in- CmH203012010
-- LO
cluding a heat-moisture exchanger (Bennett cascade II
humidifier).
O
Ventilatory pattern employed. The choice of the ven- Seconds
tilatory pattern used in each patient was determined Fig. 1. The upper trace shows signals obtained from the Fleisch
by the clinical state of the patient. pneumotachograph for both constant and decelerating WFs when
the ventilator was connected to the lung model. The lower trace
Two inspiratory flow WFs were compared, i.e. shows flow and airway pressure signals in one patient for both
constant and decelerating. In 11 patients the sequence WFs. Flow signals were obtained from the ventilator recording out-
of WF studied was constant, decelerating and back to put
70 N. A1-Saady and E. D. Bennett: Decelerating waveform in patients on IPPV
po,/
procedure).
Airway pressure. Airway pressure was recorded con- method of Ingelstedet et al. [18]. Total static com-
tinuously with an Elcomatic transducer (Elcomatic SE pliance (Cst) (chest wall + lung) was calculated [11]
750), calibrated in the usual way. This was connected by dividing the introduced volume (VT) by the pres-
between the Y-piece of the breathing tubes and the sure (P2-P0) established at the end of end-inspiratory
endotracheal/or tracheostomy tube. The signal was pause (EIP).
amplified and recorded on a multichannel recorder A second index of compliance was calculated from
(UV Oscilloscope, SE Labs.). the pressure (P3-P0). This pressure was found by pro-
jecting the slope of the slow decline in pressure back-
End-inspiratory pause (EIP). An end-inspiratory wards (Fig. 2) to meet a vertical line from PI. When
pause (EIP) of 0.55 s was inserted for 15 s during the the introduced volume is divided by this pressure it
time of measurement of breathing mechanics. To test provides a measure of kinetic compliance (Ck). This
the effect of using an EIP for short period on the term is used because dynamic compliance is calculated
breathing mechanics, a pilot study on two patients from the pressure difference between points of zero
was carried out, in which an EIP of 0.5 and 1.0 s were flow during a tidal volume [11]. Pressure stress relax-
used one at a time and with each WF. Measurement ation was presented in two ways:
of airway pressure, VT and I / E flow was made every 1. as the fall in pressure per unit initial pressure ((P3-
minute for 5 rain, then every 5 rain for 30 rain. P2/(P3-P0) x 10) m m H 2 0 / c m H 2 0 ;
2. as the rate of fall in pressure, obtained by dividing
(P3-P2) by EIP (t3-t2) (cm H 2 0 / s ) .
Measurements
All the patients were studied in the supine position. In Arterial blood gas analysis
all 14 patients the mechanics of breathing were mea- Five ml of arterial blood were collected in heparinised
sured in the following way: each WF was applied for plastic syringes from an indwelling arterial catheter,
20 min and at the beginning and the end of this peri- the volume in the dead space having previously been
od, records of airway pressure, flow and a measure- discarded. The time from collection to analysis was
ment of VT was made. never more than 10 min, and analysis was carried out
Figure 2 illustrates the type of airway pressure using an Instrumentation Laboratory machine (Mod-
tracing produced. During the inflation of VT, airway el 415). This was calibrated in the standard fashion
pressure changes in the following way: immediately prior to each measurement. Each sample
As flow ceases (t2) there is a fall in pressure (P1- was measured twice, the mean value being calculated.
P3) which is the total flow resistance (R) at the maxi-
mal lung volume. The pressure then slowly declines to
Haemodynamic measurements
reach a stable value. This decline in pressure (P3-P2)
is called pressure stress relaxation [17]. Total flow In six patients, pulmonary artery balloon thermo-di-
resistance of the respiratory system (gas flow + lung lution flotation catheters (Edwards laboratories) were
tissue + chest wall tissue) was obtained by the same inserted for clinical reasons. Cardiac output measure-
N. A1-Saadyand E. D. Bennett: Deceleratingwaveformin patients on IPPV 71
ments were made at the beginning and the end of each the same type of flow pattern. All the parameters
study period. Five ml of ice-cold 5°70 dextrose were were compared in the following set of pairs, accord-
rapidly injected into the right atrium, and values were ing to the two types of gas flow used: Constant-decel-
obtained from an Edwards Laboratories cardiac out- erating, decelerating-constant. Paired differences
put computer (Model 9520). The mean of 3 sets of were also calculated between these start and end set-
measurements were calculated. tings, p values of <0.05 were regarded as significant.
Both pulmonary and systemic arterial pressures
were measured using a Statham P 23 Dc transducer
connected to a Kontron medical 128A monitor. Dy-
Results
namic and mean pressures were measured on each oc-
casion. Central venous pressure was monitored from
a right atrial catheter. Equipment testing
Table 2 shows the correlation equations for VT de-
Expired gas collection
livered by the ventilator measured with the Wright
In five patients, expired gas was collected from the spirometer and the electronic respirometer when two
ventilator outlet into a 100-1 Douglas bag over a 5-rain peak flows (10 and 80 l/rain) were used with the con-
period. The volume was determined with the Wright stant and decelerating WFs respectively. It is evident
spirometer which was fitted between the expiratory from the table that there is a good correlation between
port of the ventilator and Douglas bag. The expired the two methods of VT measurement, although the
CO2 concentration (FECO2) was measured using an intercept values were high for the lower peak flows
infra-red gas analyzer (P K Morgan, Model 901), particularly with decelerating WF. However, during
which was calibrated with air, 5°70 CO2 in N2 and the course of this study the minimum VT used was
10% CO2 (BOC, special gases). 600 ml and the lowest peak flow 30 1/min. Examina-
tion of the equations show the error of measurement
Derived parameters [(A-a)PO2 gradient]. The ideal of such VT to be less than 5°70. In the same table the
alveolar air equation [27] was used to calculate the al- lower equation represent the correlation equation of
veolar oxygen tension (PAO2) using the measured in- the Wright spirometer and the graduated syringe and
spired oxygen tension and assuming RQ to be 0.85 shows an excellent correlation with an intercept value
[25]. of zero.
Constant WF
)( 29.5 24.2 1.56 5.75 20.9 8.72 48.8 40.0 798 0.38 13.30 2.49
+SD 11.4 11.2 1.4 3.6 10.4 3.7 28.3 15.4 122 0.06 5.9 1.16
Decelerating WF
24.9 22.1 1.35 4.55 19.27 6.00 56.9 46.1 802 0.32 14.04 2.03
+SD 11.3 10.7 1.13 2.83 9.5 2.9 39.7 20.7 121 0.04 "6.0 1.14
N 16 16 16 16 16 16 16 16 16 5 15 16
AX 4.62 2.05 0.22 1.2 1.62 2.7 -8.1 -6.1 - 3.7 0.06 - 0.7 0.45
+SD 3.4 2.38 0.39 1.8 1.7 1.95 12.9 8.3 13.1 0.05 1.9 0.55
% Change 16% 9% 13% 21% 8% 31% 17% 15% 0.5%16% 6% 18%
p <0.0001 0.0037 0.04 0,016 0.0016 <0.0001 0.02 0.01 0.3 0.035 0.14 0.004
Constant WF
25.5 22.4 1.39 4.67 19.72 6.02 55.1 45.1 814 0.32 13.82 2.11
+SD 11.0 10.4 1.18 2.6 9.5 3.1 37.9 20.2 127 0.04 6.4 1.15
Decelerating WF
30.4 25.0 1.52 5.93 21.59 9.08 46.9 38.4 810 0.37 13.22 2.49
_+SD 11.2 10.7 1.35 3.8 9.9 3.9 25.2 14.3 124 0.04 6.2 1.2
N 17 17 17 17 17 17 17 17 17 5 15 17
A)( -- 4.86 -- 2.6 - 0.13 - 1.27 - 1.87 - 3.06 8.2 6.63 3.7 - 0 . 0 5 0.61 -0.38
___SD 2.77 1.98 0.31 1.62 1.5 2.19 14.3 9.18 12.7 0.02 1.35 0.2
%0 Change 19% 12% 9% 27% 9% 51% 15% 15% 0.5070160/0 4O7o 18o70
p <0.0001 <0.0001 0.95 0.005 <0.0001 <0.001 0.031 0.009 0.25 0.005 0.1 < 0.0001
PP = pressure peak; PK = pressure kinetic; SR = pressure stress relaxation; p P L = pressure plateau, R = total resistance, Cst = total static,
Ck = total kinetic compliances, Pawp = mean airway pressure, W = work of breathing
N. A1-Saady and E. D. Bennett: Decelerating waveform in patients on IPPV 73
Constant W F
J~ 73.73 36.36 313.9 246.3 5.02 88.5 84.5 22.33 11.0 10.0
+_SD 25.4 5.5 65.9 69.1 1.7 10.8 13.5 5.05 5.8 6.0
Decelerating W F
84.30 35.3 315.0 236.1 4.98 84.4 83.2 23.33 12.7 9.3
+SD 31.2 4.9 67.4 72.3 1.6 12.3 12.5 5.7 4.6 5.9
N 15 15 9 9 5 8 6 6 6 5
AJ~ - 1.41 1.06 - 1.14 10.21 0.034 4.13 1.33 - 1.0 - 1.67 0.7
_+SD 1.3 2.7 3.18 9.28 0.17 6.17 6.3 1.4 2.8 0.6
070 Change 14% 3% 0.4% 4°70 1°70 5% 2% 4°70 15070 7%
p 0.0009 0.16 0.31 0.011 0.67 0.1 0.63 0.14 0.21 0.052
Constant W F
J~ 88.65 35.27 321.8 236.1 5.04 86.0 84.2 22.67 12.0 9.7
+ SD 38.3 4.9 67.0 69.7 1.5 11.8 12.1 5.5 5.0 5.9
Decelerating W F
J~ 80.03 35.3 322.2 245.0 5.08 85.3 82.4 26.17 11.8 9.3
_+SD 34.7 4.7 66.1 66.8 1.6 10.0 13.5 5.4 5.3 6.5
N 16 16 10 10 5 8 6 6 6 5
AX 1.15 -0.03 -0.38 -8.93 -0.04 0.75 1.79 -3.5 0.17 0.4
_+SD 1.06 1.85 2.37 7.86 0.19 7.3 5.8 6.19 1.17 1.19
07oChange 10070 0.1%o 0.1°7o 4070 l°7o l°7o 2o70 15070 l°7o 40/0
p 0.0006 0.95 0.63 0.0058 0.69 0.78 0.49 0.22 0.74 0.5
these parameters do not occur. In most reported stud- a high initial peak flow which is followed by a slow
ies these criteria have not been fulfilled. decline in flow allowing more time for gas to enter the
Watson [30-33] and Adams et al. [1] had used alveoli [28]. This may be particularly relevant to those
their own experimental ventilators. Watson's ventila- alveoli with prolonged time constants. Furthermore,
tor [34] was a pressure generator, the performance of this early delivery of the VT is accompanied by a high-
which would be modified by the subjects respiratory er initial airway pressure allowing a longer residence
mechanics. Watson's observations on the dead space time for the fresh gas in the alveoli thus favouring gas
were based on using the rebreathing technique to ob- distribution [29].
tain PaCO2 which requires disconnecting the patient We found that the decelerating WF produced a
from the ventilator and assumed fixed a-v CO2 ten- higher mean intrathoracic pressure which might be ex-
sion differences. With small changes in V D / V T ratio pected to overdistend those alveoli with a short time
this technique is too insensitive to detect such constant, thus compressing neighbouring alveoli and
changes. their capillaries. This would result in an increased
In Adams et al. [1] study the geometry of the WF physiological dead space. Furthermore, the higher
was different from the one used in our study. Three of mean intrathoracic pressure also has the potential for
the WFs used all had their peak flow in the middle reducing cardiac output. In our study we found no ev-
third of inspiration, i.e. the variation in timing may idence of this occurring. This is in agreement with
be too small to have any significant effect on the in- Fairley [14] who stated that it should be possible to
spiratory WF. Apparently the peak flows were also rase intrathoracic pressure enough to improve gas dis-
very similar. tribution, lowering V D / V T and improving oxygena-
Finally, in their protocol they stated that they al- tion without exceeding the critical closing pressure of
ways started with the sine WF, but they do not indi- the pulmonary capillaries.
cate if after changing the WF they return to the origi- End-inspiratory pause (EIP) is a maintained infla-
nal WF, to verify the reproducibility of the system. tion of VT. This is associated with a slow fall in air-
The lung volume in their animals was also distorted way pressure, although lung volume is not altered
by periodic hyperventilation, so that the compliance [10]. This fall in pressure results in part from stress re-
and lung volume at the beginning of each investiga- laxation. Other contributing factors are the redistri-
tory period presumably varied. This in turn could bution of the inspired gas, absorption of gas into the
override any differences between the effects of differ- blood stream and adiabatic compression of the gas in
ent WFs. Nevertheless Adams et al. [1] had suggested the lung. Don and Robson [11] reported that in anaes-
that the inspiratory flow pattern, with a fixed inspira- thetized patients, a maintained inflation of the respi-
tory time and constant VT might affect the cardio-re- ratory system of 1 I, introduced at a rate of I I/s, is as-
spiratory function by changing the cardiac output or sociated with a decline in airway pressure of approxi-
altering the distribution of ventilatiOn or blood flow mately 0.2 cm H20 for each cm H20 inflation pres-
within the lung. sure. The rate of fall in pressure was found to be ap-
Bergman [ 4 - 6 ] used different ventilators (pres- proximately 1.5 cm H20/s.
sure generators) to produce different WFs resulting in The findings in our study were 0.16 cm H 2 0 / c m
variations in inspiratory time, VT, frequency and ex- H20 and 5.75 cm H 2 0 / s respectively.
piratory pressure, thus making it difficult to relate the Fuleihan et al. [15] studied the effect of EIP on
results to change in the WFs. gas exchange in 10 ventilated adult patients with acute
Similarly, there were some inconsistencies in Jo- respiratory insufficiency. They showed that when
hansson's studies particularly in the constancy of the they used an EIP of 0.6 and 1.2 s compared to a zero
VT, as values indicated in the table differ from those EIP there were no changes in arterial oxygenation,
calculated from the values given for compliance and but V D / V T ratio and PaCO2 dropped significantly.
end-inspiratory pressure [ 2 0 - 2 1 ] . Furthermore, The measurement in their study were made 45 to 60
Johansson and LOfstr6m [19] used low VT min after EIP application.
(mean = 407+63 and 411 + 55 ml) with a calculated In our study when EIP of 0.5 and 1.0 s was tested
VD of 273 ml. Despite these inconsistencies the on two patients there were no changes in total compli-
(A-a)PO2 was lower with the decelerating WF. ance and resistance after the first min of EIP applica-
It seems therefore, that there is now good evidence tion, but there was 5% increase in total compliance
demonstrating that decelerating WF improves gas dis- with no changes in total resistance after 2 rain of EIP
tribution and uptake in the lung without deleterious application. It is highly unlikely therefore, that the
effect on the circulation. use of 15 s EIP to obtain measurement could influ-
One explanation for this improvement is that the ence the gas distribution in our patients. Further-
bulk of the VT is delivered early in inspiration due to more, the expired gas collection for CO2 concentra-
N. A1-Saady and E. D. Bennett: Decelerating waveform in patients on IPPV 75
tion measurement was made after EIP had been dis- 14. Fairley HB (1970) Artificial ventilation. In: Scurr C, Feldman
continued. S (eds) Scientific foundations of anaesthesia. William Heine-
mann, London, p 192
Clinical implications 15. Fuleihan SF, Wilson RS, Pontoppidan H (1976) Effect of me-
chanical ventilation with end-inspiratory pause on blood-gas
As the application of the decelerating WF results in a exchange. Anesth Analg 55(1): 122
lower airway resistance and improved gas distribution 16. Herzog P, Norlander OP (1968) Distribution of alveolar vol-
umes with different types of positive pressure gas flow pat-
it may be of benefit in those patients with high airway terns. Opuscula Medica 13:27
resistance. Similarly the improvement in PaO2 and '17. Hughes R, May A J, Widdicombe JG (1959) Stress relaxation in
(A-a)PO2 gradient also implies probable benefit in rabbit's lungs. J Physiol (Lond) 146:85
patients with severe "~g/(~ abnormalities. This would 18. Ingelstedt S, Jonson B, Nordstrom L, Olsson S-G (1972) A ser-
vo-controlled ventilator measuring expired minute volume, air-
allow reduction in FIO2. In addition the prolonged al-
way flow and pressure. Acta Anaesth Scand (Suppl) 47:7
veolar residence of the gas could be beneficial in 19. Johansson H, L6fstr6m JB (1975) Effects on breathing me-
patients with impaired gas diffusion. Similarly, those chanics and gas exchange of different inspiratory gas flow pat-
patients with reduced compliance due to atelectasis terns during anaesthesia. Acta Anaesth Scand 19:8
may also benefited by the higher mean airway pres- 20. Johansson H (1975a) Effects on breathing mechanics and gas
exchange of different inspiratory gas flow patterns in patients
sure. undergoing respirator treatment. Acta Anaesth Scand 19:19
Our results suggest therefore, that the decelerating 21. Johansson H (1975b) Effects of different inspiratory gas flow
WF should be more generally employed in patients re- patterns on thoracic compliance during respirator treatment.
quiring IPPV. Acta Anaesth Scand 19:89
22. Lyager S (1968) Influence of flow pattern on the distribution of
Acknowledgements. We acknowledge the help and cooperation of the respiratory air during intermittent positive-pressure ventila-
all the staff in the Intensive Therapy Unit. In addition the clerical tion. Acta Anaesth Scand 12:191
assistance of Mrs. Betty Simpson, the word processing expertise of 23. Mushin WW, Rendell-Baker L, Thompson PW, Mapleson
Mr. Nawzer Mehta, and the graphical expertise of Mr. Mike Birch. WW (1980) Automatic ventilation of the lung. Third edn,
Blackwell, Oxford
References 24. Nunn JF (1977) Applied respiratory physiology. Butterworths,
London
1. Adams AP, Economides AP, Finlay WEI, Sykes MK (1970) 25. Nunn JF, Bergman NA, Coleman AJ (1965) Factors influenc-
The effects of variations of inspiratory flow waveform on car- ing the arterial oxygen tension during anaesthesia with artificial
diorespiratory function during controlled ventilation in nor- ventilation. Br J Anaesth 37:898
too-, hypo- and hypervolaemic dogs. Br J Anaesth 42:818 26. Otis AB, Mckerrow CB, Bartlett RA, Mead J, Mellroy MB,
2. Baker AB, Babington PCB, Colliss JE, Cowie RW (1977) Ef- Selverstone N J, Radford EP (1956) Mechanical factors in dis-
fects of varying inspiratory flow waveform and time in inter- tribution of pulmonary ventilation. J Appl Physiol 8:427
mittent positive pressure ventilation. I. Introduction and meth- 27. Otis AB (1964) Quantitative relationship in steady-state gas ex-
ods. Br J Anaesth 49:1207 change. In: Fenn O, Rahn H (eds) Handbook of physiology,
3. Baker AB, Colliss JE, Cowie RW (1977) Effects of varying in- section 3. Respiration, vol 1. American Physiological Society,
spiratory flow waveform and time in intermittent positive pres- Washington D.C., p 681
sure ventilation. II. Various physiological variables. Br J An- 28. Robinson JS (1978) Respiratory care and the principle of venti-
aesth 49:1233 lation. In: Hanson GC, Wright PL (eds) The medical manage-
4. Bergman NA (1963) Effect of different pressure breathing pat- ment of the critically ill. Academic Press, London, p 5
terns on alveolar-arterial gradients in dogs. J Appl Physiol 29. Robinson JS (1973) The influence of differing flow patterns on
18:1049 the gas distribution and uptake during artificial ventilation. 7th
5. Bergman NA (1967) Effects of varying inspiratory waveforms International Anaesthesia Postgraduate Course. H. Egermann,
on gas exchange. Anesthesiol 28:390 Vienna, p 277
6. Bergman NA (1969) Effect of varying respiratory waveforms 30. Watson WE (1962a) Some observations on dynamic lung com-
on distribution of inspired gas during artificial ventilation. Am pliance during intermittent positive pressure respiration. Br J
Rev Respir Dis 100:518 Anaesth 34:153
7. Chackrabarti MK, Sykes MK (1976) Evaluation of the lung 31. Watson WE (1962b) Transmural central venous pressure dur-
ventilator performance analyser. Anaesthesia 31:521 ing intermittent positive pressure respiration. Br J Anaesth
8. Cox LA, Almeida AP, Robinson JS, Horsley JK (1974) An 34:278
electronic respirometer. Br J Anaesth 46:302 32. Watson WE (1962c) Observations on physiological dead space
9. Cox LA, Chapman EDW (1974) A comprehensive volume cy- during intermittent positive pressure respiration. Br J Anaesth
cled lung ventilator embodying feedback control. Med Biol 34:502
Eng 12:160 33. Watson WE (1962d) Observations on the dynamic lung compli-
10. Don H (1977) The mechanical properties of the respiratory sys- ance of patients with respiratory weakness receiving intermit-
tem during anaesthesia. Int Anaesthesiol Clin 15(2): 113 tent positive pressure respiration. Br J Anaesth 34:690
11. Don HF, Robson JG (1965) The mechanics of the respiratory 34. Watson WE, Spalding JMK, Smith AC (1962) A respirator
system during anesthesia. Anesthesiol 26(2):168 pump to provide a variable waveform. Br J Anaesth 34:149
12. Enghoff H (1938) Volumen Inefficax Bemerkungen zur Frage Dr. Naab A1-Saady
des schadlichen Raumes. Upsala Lak Foren Forh 44:191 Department of Medicine 1
13. Fahri LE (1966) Ventilation-perfusion relationships and its role St. George's Hospital Medical School
in alveolar gas exchange. In: Collin G Care (ed) Advances in London SW17 ORE
Respiratory Physiology. Edward Arnold, London, p 178 UK