Sunteți pe pagina 1din 3

DIC MICROSCOPY

Principles and Applications of Differential


Interference Contrast Light Microscopy
A. Lasslett, Microscopy Division, Olympus UK Ltd, Southall, Middlesex, UK

BIOGRAPHY INTRODUCTION 2. This light enters a Wollaston prism situ-


After graduating from Most life scientists recognise phase-contrast ated in the front focal plane of the condenser.
Leicester University with images of unstained specimens and many will The actual components are described below.
a degree in Biological Sci- be adept at adjusting the microscope to The prism interacts with the polarized light to
ences Alan Lasslett went achieve good images. Differential interference produce two separate wavefronts polarised
on to work in pharma- contrast (DIC), on the other hand, is much less perpendicularly to each other. These are
ceutical research. Subse- frequently encountered, yet is in many ways a termed the ordinary (O) and extraordinary (E)
quently he has under- superior technique. Possibly the main reason rays. Furthermore, these two wavefronts are
taken several commercial roles, and has held why DIC is not often used is familiarity. Day-to- separated by a very small difference (less than
the position of marketing manager with day microscopy remains, on the whole, a sub- the resolution of the system). This separation
Olympus UK for a number of years. He is jective matter so knowing what a specimen is termed shear and is an important character-
actively involved in running microscopy should look like is an advantage. Switching to istic of the system (Figure 3).
courses in research institutes and hospital a new technique may lead to difficulties in 3. The two wavefronts pass through the
laboratories. interpretation, or worse, the acceptance of specimen, and are retarded to varying extents
poor images. The aim of this article is to intro- in doing so.
ABSTRACT duce DIC to those life scientists who are not 4. The light now enters a second Wollaston
Differential interference contrast (DIC) light well acquainted with the technique and to prism set-up which recombines the wave-
microscopy is a technique which produces give tips on how to achieve the best images. fronts. If there has been a phase shift between
impressive 3D-like images of unstained the two rays as they pass through areas of dif-
specimens. The shadowing effects of the ORIGINS AND ADVANTAGES OF DIFF- ferent refractive index then elliptically
technique are remarkable, yet often misun- ERENTIAL INTERFERENCE CONTRAST polarised light is the result.
derstood. The additional components Differential interference contrast optics have 5. Finally the light enters a second polarizing
required to enable DIC observation on a been with us since the 1950s. Georges filter, termed an analyzer. The initial polarizer
research-level microscope are precision- Nomarski, a French physicist, modified the and this analyzer form crossed polars. The ana-
made and expensive, so DIC is often over- Wollaston prism (used to detect optical gradi- lyzer will permit the passage of some of the
looked in favour of the more common ents in specimens) in order to create differ- elliptically polarized light to form the final
phase contrast technique. This article pro- ences in intensity. The DIC microscope as such image. All the remaining light will be blocked
vides an overview of DIC microscopy, includ- was devised by Francis Smith in 1955. by the analyzer.
ing guidelines for setting up the system The major advantage of DIC over phase con- An additional component, a first-order red
correctly. trast is that the full aperture of the microscope plate, may also be part of this arrangement. It
is used. In phase contrast the condenser’s permits adjustment of contrast and colour.
KEYWORDS annular stop restricts the aperture, and there- The above description merely touches on
light microscopy, polarized light, differen- fore the resolution of the system. the physics of the technique; for a fuller
tial interference contrast, phase contrast, Compared with phase-contrast images, dif- description refer to the excellent Florida State
Nomarski, Wollaston prism ferential interference contrast: University Molecular Expressions website.
· Produces high resolution images
ACKNOWLEDGEMENTS · Shows good contrast
The author thanks Esther Ahrent of Olym- · Can be used with thick specimens
pus Life and Material Science Europa GmbH · Lacks the distracting halo of phase contrast
for providing the images used in this article. · Can be further processed (video enhanced).

A U T H O R D E TA I L S HOW DOES DIC WORK?


Alan Lasslett, DIC microscopy is essentially a development of
Microscopy Division, polarising microscopy which employs sophisti-
Olympus UK Ltd, cated optical principles to produce 3D-like
Dean Way, images of unstained cells and tissues. How-
Great Western Industrial Park, ever, it is essential to realise that these images
Southall, Middlesex, UB2 4SB, are not all they seem; the apparent peaks and
United Kingdom. troughs seen in the image are not actual rep-
Tel: +44 (0)207 250 4697 resentations of the morphology of the cell or
Email: a.lasslett@olympus.uk.com tissue. Instead, they are the product of the
optical gradient through the specimen and
the wavefront path distance. So while DIC
images are impressive, care must be taken in
their interpretation.
This is briefly how Nomarski DIC images are
produced (see Figure 3 for a schematic):
Microscopy and Analysis 20(5):S9-S11 (UK), 2006 1. Light passes through a standard polarizer Figure 1:
before entering the condenser, producing A typical research-level upright microscope for brightfield, darkfield,
plane-polarized light. phase contrast, DIC and fluorescence applications.

MICROSCOPY AND A N A LY S I S L I G H T M I C R O S C O P Y S U P P L E M E N T • S E P T E M B E R 2 0 0 6 S9
THE DIC MICROSCOPE a b
A typical research level upright microscope for
brightfield, darkfield, phase contrast, DIC and
fluorescence applications is shown in Figure1.
The configuration for DIC varies between
models and manufacturers, and in particular
between upright and inverted microscopes.
While most utilise the conventional Nomarski
DIC system others use the related de Sénar-
mont set-up. The following description refers
to the Olympus BX upright microscope config- Figure 2:
(a) DIC prisms designed to fit the condensers of Olympus BX and IX (b) DIC sliders which are inserted into the nosepiece of Olympus BX
uration; other makes and models will have
microscopes. microscopes. The lefthand slider has a recess to accept an analyzer.
comparable components.

Condenser DIC slider Locate the field iris diaphragm in the base of
Although a separate polarizer is often The second Wollaston prism arrangement is a the microscope and partly close it. Use the con-
described, in fact for practical reasons con- slider (Figure 2b) fitted above the objectives trols on the condenser to focus and centre the
densers designed for DIC usually have a built- but below the Telan lens (in an infinity-cor- image of the blades of the iris. Open the iris
in polarizer. This can be slid out of the light rected microscope). In this case only one prism fully after this procedure.
path for brightfield illumination. The polarizer is required, and it is provided with a means of
can fully be rotated, but is marked to permit sliding it across the light path. The DIC slider is Initial procedure
correct east-west orientation and a locking orientated northwest – south east, i.e. diago- Firstly ensure that the aperture iris fitted into
screw is provided. nally in the light path. the condenser is fully open. Next, insert the
The main body of the condenser is the rotat- polarizer into the light path but ensure that
ing, phase contrast type, but supplied without Analyzer the condenser is in the brightfield position (i.e.
inserts. The insert used for DIC is a Wollaston The second polarizer in the system, termed the no Wollaston prism in the light path).
prism, a kind of beamsplitter (Figure 2a). Each analyzer, is installed above the DIC slider. In Slide the upper Wollaston prism into posi-
prism consists of two precision made wedges most cases it is fitted directly into the slider, tion (this is the the DIC slider, usually in the
of quartz, cemented together so that their and its north – south orientation is guaranteed nosepiece) along with the analyzer.
axes of birefringence are at right angles to by a lug and corresponding slot. Remove an eyepiece and look down the eye-
each other. The prism itself is mounted in a cir- The optical components are now in the con- piece tube; you can see the back focal plane of
cular cell. figuration shown in Figure 3. the objective. If required a centring telescope
Appropriate Wollaston prisms need to be can be inserted into the eyepiece tube to pro-
inserted, taking care to orientate correctly SETTING UP DIC vide a magnified view.
using the protruding pin as a guide. These Although differences exist between models Turn the knob on the DIC slider until a black
prisms are specific for the objectives to be and manufacturers, the following procedure band on a light background extends diago-
used, so if DIC observation at 103, 403 and (for Olympus BX microscopes) will have para- nally across the back focal plane. This is the ini-
1003 is required then three matching prisms llels in most other models. tial starting point for correct set-up; if the
need to be installed. diagonal band does not appear check for
One advantage of this condenser type is that Köhler illumination errors elsewhere, for example the polarizer
generally there are additional mounting posi- It is a prerequisite for DIC microscopy that cor- may not be correctly orientated. Replace the
tions so phase-contrast annuli can be installed. rect Köhler illumination is first set up. To do eyepiece.
This will permit observation of a specimen first this move all DIC elements out of the light Finally rotate the condenser to bring the
by DIC then by phase contrast – a very instruc- path (the polarizer can be left) then focus on a appropriate Wollaston prism into the light
tive exercise. stained specimen using a 103 objective. path; for example select the DIC20 position if
a 203 objective is being used.
Objectives
Theoretically any objectives can be used, but in DIC IMAGES
practice higher grade objectives (Fluorite and Monochrome or colour?
apochromatic types) are generally specified to A correctly set up a DIC microscope will pro-
benefit from the high-resolution potential. In duce monochrome images with good resolu-
many cases phase-contrast fluorite objectives tion and contrast. However more aesthetically
are chosen, permitting brightfield, DIC, phase pleasing images can be achieved by moving
contrast and fluorescence observation with a the DIC slider through the light path. Colours
single set of objectives. varying from magenta to blue and yellow are
Note that since DIC is a polarized-light tech- formed, reflecting the optical gradients in the
nique then ideally strain-free objectives specimen. Note, however, that while these
should be used (these are to be found on the images are very attractive (and frequently
manufacturer’s list of objectives for quantita- published) they do not tell us anything useful
tive polarized-light microscopy). However, about the specimen. From a scientific point of
most users find standard objectives accept- view the best DIC images are monochrome.
able.
Peaks and troughs
Specimen slides and vessels Another consideration when interpreting
One important restriction of DIC is that plastic these images is understanding how the
vessels cannot be used due to the strain they shadow effects are formed. They result from
exhibit under crossed polars. For upright two characteristics: the difference in refractive
microscopes this is not an issue, but users of index within the specimen or between speci-
inverted microscopes may consider using plas- Figure 3: men and medium, and the distance the wave
tic vessels which have a glass insert of coverslip Schematic showing the major wavefront-splitting optical components path travels within these regions. Thus a ‘hill’
thickness – these are available commercially. and pathways in the differential interference contrast microscope. may indeed be a raised feature, or it may sim-

S10 MICROSCOPY AND A N A LY S I S L I G H T M I C R O S C O P Y S U P P L E M E N T • S E P T E M B E R 2 0 0 6


DIC MICROSCOPY

Figure 4: Figure 5:
A thin layer of cultured cells imaged using a DIC system designed to enhance contrast. Anterior region of a nematode worm – a thick specimen imaged using high-resolution DIC.

ply be an area of high refractive index. Simi- light is used as it penetrates deeper into the CONCLUSIONS
larly a ‘crater’ may be a hollow or may be a tissue slice than visible light. Appropriate Differential interference contrast is a tech-
vacuole of very low refractive index. optics with high IR transmission must be nique which deserves wider applications in life
The orientation of features in the specimen utilised, and the image is captured using an science. Most modern research level micro-
with respect to the DIC elements is also an infrared camera. To the untrained eye these scopes, both upright and inverted, can be
important aspect of the image. Rotating the IR-DIC images look blurred but they are highly updated to carry out DIC observation, in many
stage (possible to a limited extent with most valued by neurophysiologists. cases with just the addition of DIC prisms and
standard stages) will often produce signifi- A further advantage of this technique is that sliders. The drawbacks of complexity and cost
cantly different images. the microscope can be equipped with a sensi- of DIC are offset by the impressive high reso-
tive camera to capture the fluorescence image lution, high contrast images, and there is
Contrast versus resolution (often of very short duration) while the IR ample scope for combining DIC with other
The shear distance between the two wave- camera gives structural information. applications such as fluorescence.
fronts emerging from the condenser Wollas-
ton prism is a fixed feature of the system and DIC in reflected-light microscopy FURTHER READING
determines the balance between contrast and Although this article is written primarily with Abramowitz, M. Differential Interference Contrast Microscopy.
resolution. Most systems are optimised for life scientists in mind there are, of course, In: Contrast Methods in Microscopy: Transmitted Light,
specimens of medium thickness, such as tissue applications of DIC in materials science using Olympus America Inc., 1987.
sections. However if the specimen is very thin, reflected light. Here, the principles are the Hartley, W. G. The Light Microscope, Its Use and Development.
such as cultured cells spread on a slide, it is same but the configuration is easier. A non- Senicio Publishing Co., Oxford, 1993.
useful to employ a DIC slider designed to rotating polarizer is fitted into the illumina- Heath, J. P. Dictionary of Microscopy. John Wiley and Sons, 2005.
enhance contrast and permit fine structures tor. A reflected-light DIC slider fits into the Lacey, A. J. Light Microscopy in Biology. Oxford University Press,
to be observed (Figure 4). Likewise, thick spec- nosepiece (as for life sciences) but in this case 1999
imens such as small animals and plants are it has the function of both Wollaston prisms in Molecular Expressions Optical Microscopy Primer website:
best observed using a slider designed to the life-science configuration. Finally a rotat- http://micro.magnet.fsu.edu/primer/index.html
reduce contrast and enhance resolution with ing analyzer is installed just below the Telan Murphy, D. B. Fundamentals of Light Microscoipy and Electronic
the specimen (Figure 5). lens in the observation tube. Imaging. Wiley-Liss, 2001.
DIC is used in metallurgy, materials and Rubbi, C. P. Light Microscopy – Essential Data. John Wiley and
A P P L I C AT I O N S O F D I C semiconductors, producing good images of Sons, 1994.
Like phase contrast, DIC is a very useful tool surface features such as scratches (Figure 6). ©2006 John Wiley & Sons, Ltd
for visualising unstained specimens. This is
clearly an advantage when observing living
specimens, such as small organisms, tissues or Figure 6:
cells. In addition to simply observing such Closed-circuit television chip
specimens DIC can be used effectively in sev- imaged using DIC to show surface
eral other specialised applications listed features which would otherwise
below. be difficult to distinguish.

Fluorescence microscopy
Locating the specimen or even the focal plane
using fluorescence illumination can be a chal-
lenge. DIC, like phase contrast, can be used at
low illumination levels for this task, but more
importantly for indication where in the speci-
men a labelled component resides.

Infrared DIC
One interesting application of DIC is the imag-
ing of cells inside tissues, such as brain slices
used in electrophysiology. Here, infrared (IR)

MICROSCOPY AND A N A LY S I S L I G H T M I C R O S C O P Y S U P P L E M E N T • S E P T E M B E R 2 0 0 6 S11

S-ar putea să vă placă și