Sunteți pe pagina 1din 8

Christopher R. I. Clayton, 1 Suleiman A. Khatrush, 1 Adriano V. D.

Bica, l and A b u Siddique I

The Use of Hall Effect Semiconductors in


Geotechnical Instrumentation

REFERENCE: Clayton, C. R. I., Khatrush, S. A., Bica, A. V. D., and Hall Effect Devices at the University of Surrey
Siddique, A., "The Use of Hall Effect Semiconductors in Geotechnical
Instrumentation," Geotechnical Testing Journal, GTJODJ, Vol. 12, Linear output Hall effect semiconductors were first used at the
No. 1, March 1989, pp. 69-76. University of Surrey in 1983 to control radial strains during Ko tri-
axial compression tests on sand [1]. The sensors were ideally suited
ABSTRACT: For the past five years or so Hall effect semiconductors for these triaxial tests because they were small, very light, and
have been increasingly used in the geotechnical engineering laboratories
could be sealed to work under water at pressures of up to 1000 kPa.
at the University of Surrey. They have been incorporated as sensing ele-
ments in local radial and axial strain measuring devices, for the small- Experience gained during the use of this early sensor system
strain instrumentation of triaxial specimens, and in small diameter showed that the semiconductors were relatively reliable, inexpen-
boundary normal and shear stress cells. Triaxial internal load cells are sive, and easy to implement.
currently being built incorporating Hall effect semiconductors. The first semiconductor used was found to be satisfactory for
This paper describes the Hall effect principle and the methods of con-
figuring magnet/sensor systems to achieve suitable measuring systems. short-term tests, or in an environment with good temperature con-
Some geotechnical instruments built at the University of Surrey are de- trol. But the temperature sensitivity of the device was found to be
tailed, and their characteristics discussed. The calibrations of the in- unacceptable for long-term tests carried out with the level of tem-
struments described in the paper show a performance generally at least perature control provided by standard air-conditioning. Subse-
as good as might be expected from some commercially available instru-
quently, since 1985, temperature compensated semiconductors
ments. Hall effect semiconductors are shown to be of use in a range of
situations where displacement can form the basis of measurement. The manufactured by Micro-Switch have been used (Fig. 1). These
displacements measured can vary from as little as 5 #m to as much as 10 semiconductors are light (0.35 g), small, and in some cases regu-
mm, and the best repeatability so far obtained has been of the order of lated against changes in DC supply voltage. They have a single DC
1/10o of a micrometre. output that varies linearly with magnetic flux density from - - 4 0
m T to + 4 0 roT, and can work with any D C voltage supply from 8
KEYWORDS: instrumentation, strain, stress, calibrations, errors, tri-
axial tests, deformation gages, pressure cells to 16 V.
These latter semiconductors were used not only in devices to
sense radial strain changes on triaxial specimens. They have also
The Hall Effect Principle been used (since 1986) for measuring simultaneously both radial
Consider a metallic or semiconductor plate through which elec- and axial local strains on 102-mm-diameter specimens in a re-
trical current is flowing. If this is placed in a magnetic field where search program to investigate the yielding of a fine Leighton Buz-
the flux lines are perpendicular both to the plate and to the current zard sand [2]. Subsequently, the linear range of the local axial
flow, the charge carriers will be deflected. A voltage will thus be strain gage has been increased (1987). A similar Hall effect sensor
produced across the plate in a direction normal to the current flow. has been successfully used in a 25-ram-diameter boundary total
This voltage is termed the Hall voltage, after E. H. Hall who dis- normal stress cell.
covered the effect in 1879.
Hall effect semiconductors are used widely as switches and to
measure magnetic flux density. Linear Hall effect sensors are typi- Implementation
cally direct current (DC) energized and deliver a DC output that
varies linearly with magnetic flux density over a specified range. The output from a Hall effect semiconductor will change, some-
Various types are commercially available. The cost of a sensor typi- times by up to 2- to 3-V DC, when a magnet is moved close to its
cally ranges from £8 to £10 ($14 to $18). They are currently used as sensing position. To produce a displacement sensor, only a semi-
the sensing element in a range of applications including com- conductor and one or more magnets are required, together with
puters, machine tools, and medical equipment. mechanical parts to hold and encapsulate them. Four basic config-
urations of sensor and magnets have so far been used in geotechni-
eal instrumentation at the University of Surrey (Fig. 2):
nOTE: Discussion is encouraged and should be submitted by Sept. 1,
1989. (1) single magnet, head-on,
~Reader in geotechnical engineering, research student (formerly), re-
(2) double magnet, bi-polar slide-by,
search student, and research student, respectively, Civil Engineering De-
partment, University of Surrey, Guildford, Surrey GU2 SXH, United King- (3) single magnet, bi-polar slide-by, with pole pieces, and
dom. (4) tandem double magnet bi-polar slide-by.
© 1989 by the American Society for Testing and Materials 0149-612318910003-0069502.50
69
Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
70 GEOTECHNICAL TESTING JOURNAL

Hall EFfect eensor

(a)

FIG. l--Photograph of a Hall effect transducer.


I:
-~l metier

Hall Effect s e n s o r s

G
t/®

FIG. 3--Electrical connections of sendor/magnet systems: (a)for sys-


tems with one sensor and (b)for the tandem double magnet bi-polar slide-
by system.

G
r-,

v (=)
® 4J
D
12=

//~ (d) r e l a t i v e di=placement oF 9ou9e ende (mm)

FIG. 2--Basic configurations of sensor~magnet systems: (a) single mag-


net, head-on, (b) double magnet, bi-polar slide-by, (c) single magnet, bi- 4
polar slide-by, with pole pieces, and (d) tandem double magnet bi-polar
slide-by. > (b)

Electrical connections to the power supply and the read-out in-


strument are shown in Fig. 3. The typical output voltage/displace-
ment characteristics of the head-on and slide-by configurations are
shown in Fig. 4. For the head-on configuration, used in our early o
triaxial lateral strain sensors and in the first prototype for a bound-
ary total stress cell, the change in output is of the order of several r e l a t i v e dieplaoement of 9auge ends (mm)
volts. But the resulting voltage/displacement relationship is FIG. 4--Output voYage~displacement relationships: (a) head-on and
strongly nonlinear. However, this type of geometry may still be of slide-by(b).

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
CLAYTON ET AL. ON HALL EFFECT SEMICONDUCTORS 71

use if software-based conversion to engineering units is used, or if [7-9] and strain gages [10]. Two such devices have been developed
the required range of displacement is small relative to the total at Surrey, using Hall effect semiconductors, and these are de-
range. It may even be considered an advantage where differing sen- scribed in detail below.
sitivities are required at different displacement levels.
The various slide-by configurations, (2) to (4), use two magnetic
A Radial Strain Measuring Device
poles, either by using two magnets with parallel north/south axes
or by extending the poles sideways towards the sensor by means of The Hall effect principle has been used for the measurement of
soft ferro-magnetic "pole pieces." The Hall effect semiconductor is radial strains on the central section of 102-mm-diameter triaxial
displaced in such a way that it passes over first one pole and then specimens. The mechanical parts of the system are essentially simi-
the opposite pole. This system has the advantage that it can, with lar to those originally designed by Bishop and Henkel [11], but
careful design, produce an output that has a portion with good lin- with the original mercury indicator column replaced by a slide-by
earity with respect to displacement. Figure 5 shows the influence of configuration, as in Fig. 2 b. The design of the device can be seen in
varying both the separation S between the magnets and the gap G Fig. 6. Two 3- by 3-ram magnets, 6 mm long, separated by a 3-
between the magnets and the semiconductor for a double magnet mm-thick polytetrafluoroethylene (PTFE) spacer are mounted
bi-polar slide-by configuration. Up to a separation S approximat- within a magnet holding block, which can be adjusted with respect
ing to the minimum face width of the pole of the magnet, an in- to the Hall effect sensor by use of the screws at either end of the
crease in the gap results in an increase in the linear range of the assembly. The Hall effect sensor is encapsulated in epoxy resin,
device. Beyond this spacing the output becomes strongly nonlin- within a brass container, and is attached to the other arm of the
ear. The sensitivity of the output can be increased by decreasing caliper. The gap between the magnets and the face of the semicon-
the gap G. For highly sensitive displacement measurement the ductor is maintained constant, at approximately 1 mm, partly by
spacing is reduced to zero, and the gap is reduced to as little as is the PTFE block separating the two magnets and partly by a 0.5
practical. mm thick PTFE strip covering the surface of the semiconductor.
The two parts of the device are maintained in contact because of
the use of uneven spring forces at the caliper hinge.
Local Strain Measuring Instruments

Conventional measurements of both axial and volumetric defor- An Axial Strain Measuring Device
mations of triaxial specimens from outside the triaxial cell intro-
duce significant errors, particularly in the computation of small Unlike the radial strain device, the axial measuring device
strains [2,3]. Some of these errors (for example, caused by bed- passed through several stages of development before a successful
ding) are of a random nature and are difficult to evaluate, even
with careful calibration. And the extensive calibrations needed to
eprin3-]oaded hin3e / p o l l a h o d hln~e-pin
evaluate the magnitude of systematic strain errors (for example,
because of internal load cell compression) are generally inconve-
nient. Other errors, which might be thought to be systematic (for ~ (a)

example, frictional and volume losses in the lower chamber of a


Bishop/Wesley stress path cell) in fact turn out to vary with use of
the equipment.
These problems have created the need for alternative techniques
of measurement, from points directly located on the specimen. A
number of different instruments have been devised in the United
Kingdom, including within them linear variable differential trans-
formers (LVDT) [3,4], proximity transducers [5,6], electrolevels

¢11
5 inoreoein9
AT TA
a'}
o
.o
0
> a]uminium rin~
.p
Q_
.i.)
braoa oontainer Ho]] Ef?eot 6ensor
0

Iinear range
For" e o l i d l i n e
I I

coble
dieplacement of the ~ensor
r e l o t i v e t o the magneto eliding b|o~k PTFE~pa~er

FIO. 5--Effectofva~ingmagnetgapandspacingonou~utfrom adou- FIG. 6--Hall effect radial strain measuring device: (a) plan and (b) Sec-
b~ magnetslide-byconfigurat~n. tion A-A.

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
72 GEOTECHNICTESTI
AL NGJOURNAL
design was achieved, because of the need to achieve good mechani- The pendulum arm below is bent, so that it may pass beneath
cal performance. The factors taken into account were the lateral strain caliper described in the previous section. The
PTFE spacer maintains the separation of the two magnets, and
(1) reduction, and if possible avoidance, of both friction and
also the gap between the face of the semiconductor and the magnet
slack in any bearings incorporated in the design;
faces. "Wings" projecting from the sides of the magnet assembly
(2) reduction of size, in order to ensure that this gage and the
(but not shown in Fig. 7) ensure that the magnets track over the
radial strain measuring gage described above could both be fitted
center of the semiconductor in the lower assembly.
on a standard 102-ram-diameter specimen within a 102-mm-diam-
2. A container to hold the encapsulated Hall effect semiconduc-
eter triaxial cell, and in order to minimize the self-weight of the
t o r - T h e lower part of the gage consists of a brass container
device (and thus the forces applied to the side of the specimen);
mounted on the specimen by means of a pinned fixing pad, and
(3) provision of a good contact between the device and the speci-
containing a linear output Hall effect semiconductor encapsulated
men, while minimising the influence of any pad rotation on the
in epoxy resin. In this version of the device the surface of the semi-
measured strains; and
conductor was covered by thin self-adhesive PTFE tape, but in
(4) flexibility and incompressibility of the electrical cabling exit-
later versions a thin glass cover slip (as used in microscopy) has
ing from the device.
been used. The cabling consists of three single-strand small diame-
The best initial solution to these problems [12] was the gage shown ter insulated copper leads within a "heatshrink" plastic cover.
in cross section in Fig. 7, which will subsequently be denoted axial
Typical output of this device is shown in Fig. 4b. The linear
gage I. It consists of two parts:
range is approximately 2.5 mm. The sensitivity of the device might
1. A spring-mounted pendulum which holds two bar magnets, be thought to be affected by mis-alignment of the two parts of the
each 3-mm-square in cross section, separated by 3 mm is sus- assembly, and so experimental work has been carried out to assess
pended from an upper pad fixed to the specimen by pins, and the precision with which the gage must be mounted on the speci-
bonded to the membrane by adhesive. The spring, which should men. When deliberate mis-alignment of the two halves of the gage
ideally be positioned as close to the side of the specimen as possi- was 5% it was found that the change in sensitivity was only 1.25%
ble, allows relative motion between the fixing pad and the pendu- of the sensitivity when properly aligned. In practice it is relatively
lum (which would occur, for example, if the specimen were to bar- easy to achieve an alignment much better than this, as a S° mis-
rel at large strains) without the need to introduce a bearing. This is alignment is extremely noticeable. Thus potential friction between
an important feature of the device, as it guarantees that there will the wings and the semiconductor casing can he avoided since these
be no slack in the system, while also ensuring that friction remains need not be a close fit.
very low. During the early stages of the design of this device, exper- The need for pins on the mounting pads, driven through the
iments showed the very large friction that could be induced in bali- membrane, was based initially on a requirement to define the gage
joint bearings when relative rotation of the pads of the device oc- length. Our first tests were carried out on saturated sand speci-
curs during mounting, on an axis perpendicular to the bearing mens, but subsequent tests on unsaturated clay specimens, when
axis. The spring has the function of gently pressing the PTFE pins were not used, have shown that the pins have a further impor-
spacer against the face of the encapsulated semiconductor, in the tant function in holding the pads against the side of the specimen
lower part of the device. until a cell pressure is applied. If the upper pad is allowed to move
away from the side of the specimen during setting up, a convolute
may develop in the membrane above the pad, and the pad will not
then reseat itself until the cell pressure is raised by a considerable
membrane ~ adhesive, amount. During this period, pad rotation occurs and can be mis-
interpreted as an extensive strain.
?ixin 9 pin

Further Developments

E i~ spring The gages described above have been in use at the University of
Surrey for over two years and have performed well. Their manufac-
t"-
v ture involves, however, a large number of machining operations,
which makes them costly to produce in quantity. Figure 8 shows a
¢- Q
.~ E new version of this gage, redesigned with production in mind.
¢- O II v..ticol More than 40 of these devices have either been installed in com-
• @
@
I\ / o ju.t..nt mercial soil testing laboratories, or are currently being manufac-
"o''"
@ tured.
The basic concept of the new version of the gage (subsequently
referred as axial gage II) remains the same, but a number of im-
provements have been made. These include the following:
/'iliad---J-" PTFEa d h e , i v e • replacement of the metal spring strip at the top of the arm
8trip with two lengths of stainless steel spring wire,
--electrical oable • the use of an arm consisting of bent brass strip, to reduce ma-
chining operations,
• substitution of smaller magnets, to reduce size, weight and
F1G. 7--Hall effect axial gage 1. complexity of the holding block,

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
CLAYTON ET AL. ON HALL EFFECT SEMICONDUCTORS 73

Calibration, Accuracy, and Performance of Hall Effect


Strain Devices

Both the radial and axial gages are calibrated using a microme-
ter (minimum resolution 2.54/tin) in a specially built jig. Both the
micrometer and the jig were manufactured of nonmagnetic materi-
als, in this case aluminum alloy, in order avoid any local influence
on magnetic fields in the area of the sensor. Linear regression anal-
yses were carried out on the linear portion of the output, and typi-
cal regression error plots are shown in Figs. 10a and b. Expected
calibration characteristics for the four types of gage are shown in
Table 1. It can be seen that the maximum linear regression error of
the axial gages is of the order of 0.24 to 0.42% of the full-scale
output in the linear range, which is a good achievement for a non-
commercial transducer. The maximum linear regression error of
the radial strain gage is poorer, largely as a result of the mechani-
cal arrangement of the system, which causes the magnets to skew
over the sensor rather than to run in line.
The overall accuracy of these devices is rather more difficult to
assess, since it is controlled not only. by the performance of the de-
FIG. 8--Hall effect axial gage H. vice in calibration under the ideal conditions described above. Fac-
tors, which may be relevant are as follows:
(1) ability to determine the gage length over which the strains
• introduction of an adjustment system in the lower pad, to al- are calculated,
low proper pad alignment, subsequent placement of the semicon- (2) the vertical position of suitable reference points from which
ductor, and final adjustment after fixing to achieve any desired po- to calculate strains free of end effects,
sition in the linear range of the device, and (3) the need for three devices per specimen (as opposed to the
• optional pinning. two currently commonly used), to avoid tilt effects, and

In addition, the desire to measure larger strains on soft clay speci-


mens, without recourse to external LVDT measuring systems has
lead to the development of a similar device with a linear range of r~

some 7 mm (or 10% strain on a 70 mm gage length). The configu-


ration used is a single magnet bi-polar slide by (as shown in Fig. 9)
(a)
mounted instead of the small magnets on the new flat gage arm.
This configuration will subsequently be referred to as axial gage
IIl.
r ~
0 II ~' " 1 i I
3

distomce (mm)

[ ~ - 10

tn

0
' '
T
- t
o~

clistance (mm)
-2
O9
0
L

0
£

FIG. 9--Single magnet bi-polar slide-by configuration used in axial FIG. lO--Linear regression error plots for (a) axial gage l and (b) axial
gage 11I. gage III.

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
74 GEOTECHNICAL TESTING JOURNAL

TABLE 1--Calibration performance of local strain devices.

Maximum Linear
Regression
Resolution Error
Full-Scale at 1 mV Gage Maximum (% Linear
Device Output V Linear Range, mm Output #m Length, mm Strain, % Range)

Radial gage 1.8 1.5 0.83 100 1.5 --+0.80


Axial gage type I 2.0 2.5 1.25 70 3.6 ±0.24
Axial gage type II 1.6 2.5 1.56 70 3.6 -+0.38
Axial gage type lII 1.9 7.0 3.68 70 10.0 -+0.42

(4) effect of closure of fissures at low strains, and barrelling and Small-Diameter Boundary Earth Pressure Cell
necking at high strains.
As part of an experimental research program on the pressures on
The influences of such factors have been described, in part, by free embedded cantilever walls, small boundary total normal stress
Clayton and Khatrush [13]. and shear stress cells were required. Several such cells have been
The actual performance of axial gage type I was examined ini- described in the literature, using either hydraulic [14] or inductive
tially by conducting special triaxial tests on a dummy specimen of [15], but mostly strain gaged type sensing units [16-21]. However,
rubber, subjected to deviatoric total stress paths. In the first test previous experience at the University of Surrey, described above,
the ends of the specimen were in direct contact with the triaxial suggested that it might be feasible to design these devices using
platens, whereas in the second test two greased latex rubber disks Hall effect semiconductors as the basic sensors, with the advantage
(0.3 mm thick) were placed at either end to the specimen to simu- that this would result in a low-cost robust device. Trials of the
late the effect of lubricated end platens. It can be seen in Fig. 11 boundary total normal stress cell have recently been completed,
that for both tests the axial and radial strains recorded by the local and the prototype for the shear cell is, at the time of writing, in
strain devices are the same, and that they differ from the external production.
axial strain measurement up to a strain level of about 0.4%. A sin- The basic principle chosen for the total normal stress cell was the
gle value of Poisson's ratio (0.45) results from all local strain mea- measurement of the displacements of the center of a diaphragm
surement, at whatever strain level, whereas external axial strain encastr6 with a thick surrounding ring. Since the Waterways Ex-
measurement introduces significant errors. Further comparisons periment Station criterion [22] suggests that the maximum deflec-
made, on specimens of saturated fine Leigthon Buzzard sand, be- tion/diameter ratio for such a cell should be 1/2000 and since the
tween external volume change measurement and volumetric diameter of the diaphragm of the proposed cell was planned to be
strains calculated as the sum of the axial and twice the radial local of the order of 20 mm, it was clear that displacements amounting
strain have also confirmed that at low strain levels volumetric to no more than one tenth of a micrometre must be measured accu-
strains must be measured locally. rately and repeatably.
Initially the simple unipolar head-on configuration was used
(Fig. 2a). However, experience with a first prototype of the cell
suggested that the more efficient bipolar slide-by configuration
+ t e a t 1 ~ avara~e meoaurement o~ two would be required in order to measure the very small displace-
o t a a t 2 J Hall E ~ o c t a x i a l 9augaa ments involved. Small (2- by 2 mm- by 1-mm) rare earth magnets
were used, with the spacing S reduced to zero and the gap G re-
X taat I l external axLal dioplacement duced to the minimum feasible (about 0.1 mm). Although this con-
• toot 2 J tranaduoar
figuration gave a relatively large sensitivity, when compared with
that of the previous gages, the maximum range of the output volt-
0.8 age remained at only about one-tenth of the values previously ob-
8 tained, and temperature sensitivity was a potential problem be-
5
cause of the semiconductor characteristics. Thus the type of Hall
effect semiconductor was once again changed; the type used in the
further development of the cell had better internal temperature
0.4
'3 compensation but had no input voltage stabilization. To overcome
0 the voltage compensation problem, two semiconductors with two
i.
slide-by magnet configurations were used in each cell (Fig. 2d),
with the output of the cell being obtained from the difference in
output between the two semiconductors (Fig. 3 b). In this configu-
0.8 1.6
ration, the output voltage of the completed cell was sufficiently sta-
axial attain CZ) bilized against voltage changes, but two further advantages were
obtained. First, the large zero offset of the individual semiconduc-
FIG. 1l--Radial and axial strains measured locally and externally on a
dummy rubber triaxial specimen. Test 1: top cap in direct contact with tors was removed. Second, by an appropriate arrangement of the
rubber specimen. Test 2: two membrane disks inserted in between speci- magnets the overall sensitivity of the cell was doubled. Finally,
men and top and bottom ends. some further contribution to temperature stability could be

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
CLAYTON ET AL. O N HALL EFFECT SEMICONDUCTORS 75

achieved, if necessary, by using semiconductors with matched tem- fluid pressure up to 250 kPa, at which stress level the diaphragm
perature sensitivity. deflection/diameter ratio is computed as 1/2260 and the com-
The final design of the device is shown in Fig. 12. The dia- puted deflection at the center of the diaphragm is 10.6/xm. The
phragm and its surrounding ring, and the back plate were ma- fluid calibration showed an output of approximately 280 mV at 250
chined from EN24 steel (similar to American Iron and Steel Insti- kPa, which is relatively high for a device of this type. The levels of
tute [AISI] 4340) and heat-treated to bring the steel to a high level nonlinearity, hysteresis, nonrepeatability, and zero deviation upon
of hardness (approximately 40-43 Rockwell C). Following this, the unloading are all less than 0.25% of the full-scale output (at 250
face of the diaphragm was milled to reduce the diaphragm to its kPa), and their combined effect is well within 1 °70. A typical linear
final thickness. This simple design aimed to produce a high-quality regression error plot is shown in Fig. 13a. The temperature stabil-
cell at low cost, but necessarily included standard design features ity of the cell was tested by mounting it inside a fan assisted oven-
such as refrigerator, and applying a complete cycle of temperature change
(1) radiused corners at the edge of the diaphragm, to reduce from 6 to 36°C and back. The zero output voltage error corre-
overstress at this position, thus limiting hysteresis; sponding to this cycle was about 0.75% of the full-scale output as
(2) a diaphragm of sufficient stiffness--Initial experimental can be seen from Fig. 13b. This corresponds to about 0.025% of
work had shown that a diaphragm deflection/diameter ratio of at the full-scale output per 1° C. The low sensitivity to temperature of
the Young's Modulus of steel (about --0.025%/°C), of the Hall
least 1/5000 would be necessary if the device was to have an ade-
quate cell action factor in some sands; effect semiconductor calibration (less than ___0.02%/°C) and of
(3) a thick outer ring to minimize the sensitivity of the cell to the rare earth magnets total flux density (about --0.06%/°C), to-
external loads other than stresses normal to the diaphragm; gether with their somewhat self-compensating combined influence,
suggest a similarly small influence upon the sensitivity of the cell.
(4) machining of the diaphragm and its ring from a single block
of steel to avoid the nonlinearity and hysteresis characterized by
bolted diaphragms; and Current and Future Developments
(5) use of hardened steel, to minimize nonlinearity, hysteresis,
and zero deviation upon unloading. At this time the first boundary shear cells are being produced,
but have yet to be calibrated. Prototypes simultaneously are being
The sensing element, later mounted inside the cell (Fig. 12), prepared for a new Hall effect internal load cell, for use in triaxial
consists of two Hall effect sensors bonded with epoxy resin to the chambers. It has recently been suggested that a Hall effect dis-
central steel stud on the underside of the diaphragm. Steel strips, placement measuring system may provide the higher levels of accu-
bonded to the inside of the ring with epoxy resin, support the mag-
net assemblies on either side of the semiconductors. During assem-
bly the pairs of magnets are carefully positioned near the center of .$0
the linear range of output voltage with, as far as possible, the same E
initial output voltage in each sensor. The magnets are initially held Z (a)
v
to these strips by their magnetism only, but following precise posi-
tioning they can be permanently bonded with adhesive. The system L
must be mounted so that the semiconductor sensors remain central

iol
with respect to the magnets, and a magnet/sensor spacing of about o~
0.1 mm is used. It is necessary to construct metallic parts, such as
the steel strips and the outer cell ring, from materials with the same
coefficient of thermal expansion, if unacceptably high thermal sen-
sitivity is to be avoided.
~
a app1 iod preaeuro (kN/m2)
The cell shown in Fig. 12 was designed for a maximum normal
stress (that is, when the diaphragm diameter/deflection ratio -.$

equals about 1/5000) of 100 kPa. The cell was calibrated under

diaphragm
ate.1 .trips ~[< 024 mm ~'I/ (b)
/} °utsr tin9
z
v

//'
,I - ° over C
o
L
o
n
@ temperature (degreea C)
two magnets

stud
H a l l E£Fect s e n s o r
FIG. 13--Calibration of Hall effect total normal stress cell: (a) linear
FIG. 12--Hall effect boundary total normal stress cell. regression error and (b) zero output temperature sensitivity.

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize
76 GEOTECHNICAL TESTING JOURNAL

racy that are currently required to improve the measurement of


total stress (the so-called lift-off pressure) in the self-boring [2] Khatrush, S. A., "The Yielding of a Fine Sand in Triaxial Stress
Space," PhD thesis, University of Surrey, Guildford, England, 1987.
pressuremeter test. [3] Daramola, C., "The Influence of Stress-History on the Deformation
of a Sand," PhD thesis, Imperial College, University of London, Lon-
don, England, 1978.
Conclusions [4] Brown, S. F. and Snaith, M. S., "'The Measurement of Recoverable
From the descriptions of the devices so far successfully de- and Irrecoverable Deformations in the Repeated Loading Triaxial
veloped at the University of Surrey it can be seen that Hall effect Test," Geoteehnique, Vol. 24, No. 2, 1974, pp. 255-259.
15] Brown, S. F., Austin, G., and Over),, R. F., "'An Instrumented Tri-
semiconductors can fairly easily be used in the development of axial Cell for Cyclic Loading of Clays," Geotechnical Testing Journal,
" h o m e - b r e w " geotechnical instrumentation. Their small-size, Vol. 3, No. 4, 1980, pp. 145-152.
light-weight, low-cost, temperature and voltage compensation, [6] Hird, C. C. and Yung, P., "Discussion," Geotechnique, Vol. 37, No.
and apparent tolerance to mis-treatment make them ideal for use 3, 1987, pp. 413-414.
in the often hostile environment in which geotechnical engineers [7] Symes, M. J. P. R., "Rotation of Principal Stresses in Sand," PhD
thesis, Imperial College, University of London, London, England,
require their instruments to work. 1983.
To date, these devices have been used as the sensing elements in [8] Symes, M. J. P. R. and Burland, J. B., "Determination of Local Dis-
radial and axial triaxial local strain measuring devices, and in placement on Soil Samples," Geotechnical Testing Journal, Vol. 7,
small-diameter boundary total normal stress cells. Devices to mea- No. 2, 1984, pp. 49-59.
[9] Jardine, R. J., Symes, M. J., and Burland, J. B., "The Measurement
sure boundary shear stress, and the deviator load in triaxial tests of Soil Stiffness in the Triaxial Apparatus," Geotechnique, Vol. 34,
are currently under development, and the possibility of using this No. 3, 1984, pp. 323-340.
type of sensor in the self-boring pressuremeter has been discussed. [10] Ackerly, S. K., Hellings, J. E., and Jardine, R. J., "Discussion,"
The calibrations of the instruments described in the paper show Geotechnique, Vol. 37, No. 3, 1987, pp. 414-415.
that their performance is generally at least as good as might be [11] Bishop, A. W. and Henkel, D. J., The Measurement of Soil Properties
in the Triaxial Test, Edward Arnold, London, England, 1957.
expected from some commercially manufactured instruments, and [12] Clayton, C. R. I. and Khatrush, S. A., "A New Device for Measuring
in some cases better. Local Axial Strains on Triaxial Specimens," Geotechnique, Vol. 36,
Hall effect semiconductors are clearly of use in any situation No. 4, 1986, pp. 593-597.
where displacement can form the basis of measurement; the range [13] Clayton, C. R. I. and Khatrush, S. A., "Discussion," Geotechnique,
Vol. 37, No. 3, 1987, pp. 415-417.
of displacement measurement can vary from as little as S ttm to as [14] Peaker, K. R., "A Hydraulic Earth Pressure Cell," Instruments and
much as 10 ram, and the m i n i m u m repeatability so far obtained Apparatus for Soil and Rock Mechanics, STP 392, American Society
has been of the order of 1/100 of a micrometre. Despite the quality for Testing and Materials, Philadelphia, 1965, pp. 75-81.
of these results, transducers based on Hall effect semiconductors [15] Rowe, P. W., "Anchored Sheet Pile Bulkheads," PhD thesis, St. An-
drews University, St. Andrews, Scotland, 1950.
are relatively simple to assemble.
[16] Redshaw, S. C., "A Sensitive Miniature Pressure Cell," Journal of
Sc&ntific Instruments, Vol. 31, No. 12, 1954, pp. 467-469.
[17] Arthur, J. F. R. and Roscoe, K. H., "An Earth Pressure Cell for the
Acknowledgments
Measurement of Normal and Shear Stresses," Civil Engineering and
The authors gratefully acknowledge the help of Mr. P. F. Public Works Review, Vol. 56, No. 659, 1961, pp. 765-770.
Cheesman in the development of the instruments described in this [18] Dunn, C. S. and Billam, J., "A Study of the Response Characteristics
of a New Earth Pressure Cell," Civil Engineering and Public Works
paper. The work of Dr. S. A. Khatrush was supported by the gov- Review, Vol. 61, No. 715, 1966, pp. 181-186.
ernment of Libya. The work of A. V. D. Bica has been supported [19] Brown, S. F., "Discussion," Proceedings, Symposium on Field In-
by both The British Council and Conselho Nacional de Desenvolvi- strumentation in Geotechnical Engineering, British Geotechnical So-
mento Cientifico e Tecnologico (CNPq) of Brazil. A. Siddique is ciety, Butterworths, London, England, 1974, pp. 543-545, 578-579.
supported by the Commonwealth Scholarship Commission of the [20] Brown, S. F., "State-of-the-Art Report on Field Instrumentation for
Pavement Experiments," Transportation Research Record, No. 640,
United Kingdom. 1977, pp. 13-28.
[21] Hanna, T. H., Field Instrumentation in Geotechnical Engineering,
References Series on Rock and Soil Mechanics, Vol. 10, Transtech, Clausthal-
Zellerfeld, Germany, 1985.
[1] Hababa, M. B., "The Dynamic Penetration Resistance and Com- [22] U.S. Army Engineer Waterways Experiment Station, "Soil Pressure
pressibility of Sand," PhD thesis, University of Surrey, Guildford, Cell Investigation," Technical Memorandum 210-1, Vicksburg, MS,
England, 1984. 1944.

Copyright by ASTM Int'l (all rights reserved); Tue Oct 15 11:56:04 EDT 2019
Downloaded/printed by
(UFRGS) Universidade Federal do Rio Grande do Sul ((UFRGS) Universidade Federal do Rio Grande do Sul) pursuant to License Agreement. No further reproductions authorize

S-ar putea să vă placă și