Documente Academic
Documente Profesional
Documente Cultură
MELODY III D C
MELODY III D C
OPERATOR’S MANUAL
MODM10C-U02
© 2018 Metaltronica
VILLA SISTEMI MEDICALI S.p.a.
All rights are reserved.
Via Delle Azalee, 3
Reproduction or transmission in
20090 Buccinasco (MI) - ITALY
whole or in part, in any form or by
Tel. (+39) 02 48859.1
any means, electronic, mechanical
Fax (+39) 02 4881844
or otherwise, is prohibited without
the prior written consent of the
Manufacturer’s web site copyright owner.
Printed in Italy
www.metaltronica.com
Manufacturer´s address
Metaltronica
X-Ray Equipment & Supplies
Via delle Monachelle, 66
00071 Pomezia (Rome)
Italy
Customer contact
metaltronica@metaltronica.com
DOCUMENT STATUS SHEET
The Manufacturer declares its responsibility concerning safety, reliability and equipment
features included in this manual only if the following items are fully satisfied:
1)-installation;
- room temperature and humidity;
- updates;
-recalibrations;
-repairs and/or modifications
carried out by technical personnel officially authorized from the Manufacturer.
2) Electrical pre-installation performed in the site where the system has to be put into
operation, carried out according to prescription given by IEC rules concerning Medical
Application.
3) Service personnel must ensure that they receive training on the mammographic unit with
the Manufacturer training programs prior to service the Unit.
4) Use of the equipment according to instruction in this manual. In case of difficulties
concerning interpretation, English translation shall prevail.
MODM10C-U02
Page 2 of 226
OPERATOR’S MANUAL
Table of Contents
SECTION 1: GENERAL INFORMATION ............................................................................................................................... 6
INTRODUCTION ..................................................................................................................................................................... 7
INTENDED USE ...................................................................................................................................................................... 7
CONTRAINDICATIONS ........................................................................................................................................................... 7
POTENTIAL ADVERSE EFFECTS ............................................................................................................................................... 7
GENERAL DESCRIPTION ......................................................................................................................................................... 8
SAFETY INFORMATION ........................................................................................................................................................ 13
INTERNATIONAL SYMBOLS .................................................................................................................................................. 15
ACRONYMS ......................................................................................................................................................................... 18
DEVICE CLASSIFICATION ...................................................................................................................................................... 19
MELODY III D C IDENTIFICATION LABELS ............................................................................................................................. 20
ANTI‐X BARRIER/AWS IDENTIFICATION LABELS .................................................................................................................. 21
MELODY III D C DIMENSIONAL DRAWINGS ......................................................................................................................... 23
MELODY III D C DIMENSIONAL DRAWINGS .......................................................................................................................... 24
ANTI‐X PROTECTIVE BARRIERS: DIMENSIONS AND WEIGHT (OPTIONAL) ............................................................................... 25
ACQUISITION WORKSTATION: DIMENSIONS AND WEIGHT (OPTIONAL) ................................................................................ 26
SECTION 2: CONFIGURATIONS ....................................................................................................................................... 28
“FULL DICOM” CONFIGURATION ......................................................................................................................................... 30
“FULL DICOM” CONFIGURATION + DIAGNOSTIC STATION .................................................................................................. 31
“MINI‐PACS” CONFIGURATION ............................................................................................................................................. 32
SECTION 3: MAIN PARTS DESCRIPTION .......................................................................................................................... 33
MELODY III D C MAIN PARTS ................................................................................................................................................ 34
MELODY III D C MAIN PARTS ................................................................................................................................................ 36
ACQUISITION WORK STATION MAIN PARTS ........................................................................................................................ 38
ANTI‐X PROTECTIVE BARRIER WITH OPTIONAL MONITOR AND KEYBOARD (ACQUISITION WORKSTATION) .............................. 39
C‐ARM MOVEMENTS KEYBOARDS....................................................................................................................................... 41
CONTROL PANEL (ACQUISITION SOFTWARE) ...................................................................................................................... 42
KNOB FOR FINE MANUAL COMPRESSION ........................................................................................................................... 44
FOOT‐CONTROLS ................................................................................................................................................................. 45
CONNECTOR PANEL ............................................................................................................................................................. 45
SECTION 4: GENERAL OPERATIONS ................................................................................................................................ 46
EXAMINATION ROOM ......................................................................................................................................................... 47
RELEVANT INSTRUCTIONS BEFORE INITIAL START‐UP ......................................................................................................... 47
SWITCHING ON.................................................................................................................................................................... 47
BLACK OUT .......................................................................................................................................................................... 51
SWITCHING OFF .................................................................................................................................................................. 53
EMERGENCY PUSHBUTTONS ............................................................................................................................................... 54
C‐ARM MOVEMENTS ........................................................................................................................................................... 55
C‐ARM VERTICAL TRASLATION ............................................................................................................................................ 55
THE OPERATOR CAN MOVE VERTICALLY THE C‐ARM BY MEANS OF THE BUTTONS N° 13 (PARAGRAPH “MAIN PARTS”). ................................ 55
C‐ARM ROTATION ............................................................................................................................................................... 55
OPTIONAL MOTORIZED ROTATION ................................................................................................................................ 55
C‐ARM ROTATION FOR STEREO BIOPSY (ONLY FOR MELODY III D C EQUIPPED WITH SBD FFDM) ............................................... 57
COMPRESSION PADDLES INSERTION AND REMOVAL .......................................................................................................... 60
BREAST COMPRESSION SYSTEM .......................................................................................................................................... 62
Motor driven compression ............................................................................................................................................. 62
Manual compression with fine adjustment .................................................................................................................... 64
Compression release ...................................................................................................................................................... 64
LIGHT BEAM INDICATOR ..................................................................................................................................................... 66
AUTOMATIC COLLIMATION DEVICE .................................................................................................................................... 67
MODM10C-U02
Page 3 of 226
OPERATOR’S MANUAL
DEVICE FOR GEOMETRIC MAGNIFICATION (VARIABLE 1.5X/2X) .......................................................................................... 67
PROTECTIVE SCREEN ........................................................................................................................................................... 68
FOCAL SPOT SELECTION ...................................................................................................................................................... 69
RADIOGRAPHIC EXPOSURE ................................................................................................................................................. 69
DOSE CALCULATOR ............................................................................................................................................................. 70
OPERATING AND ERROR MESSAGES ................................................................................................................................... 70
OPERATING ERRORS ............................................................................................................................................................ 70
ERROR MESSAGES ............................................................................................................................................................... 72
SECTION 5: EXPOSURE TECHNIQUES .............................................................................................................................. 74
MANUAL MODE .................................................................................................................................................................. 76
AUTOMATIC EXPOSURE MODE ........................................................................................................................................... 77
AUTOMATIC FILTER DEVICE RH/AG ...................................................................................................................................... 79
SECTION 6: TYPICAL OPERATING CONFIGURATIONS ...................................................................................................... 81
MELODY III D C 24X30CM FORMAT ....................................................................................................................................... 83
CONFIGURATION FOR STANDARD EXAMINATION (LARGE BREAST) .............................................................................................. 83
CONFIGURATION FOR STANDARD EXAMINATION (NORMAL BREAST) .............................................................................................. 84
CONFIGURATION FOR SPOT CONTACT EXAMINATION .................................................................................................................. 84
CONFIGURATION FOR GEOMETRIC MAGNIFICATION ................................................................................................................... 86
CONFIGURATION FOR BIDIMENSIONAL BIOPSY .......................................................................................................................... 86
MELODY III D C 24X30 CM FORMAT ...................................................................................................................................... 88
CONFIGURATION FOR STEREOTACTIC BIOPSY ............................................................................................................................. 88
SECTION 7: ACQUISITION SOFTWARE OPERATING INSTRUCTIONS ................................................................................. 89
1. STARTING THE ACQUISITION SOFTWARE ........................................................................................................................ 91
2. USER INTERFACE DESCRIPTION ....................................................................................................................................... 92
2.1. Status Bar ................................................................................................................................................................ 93
2.2. Operation Bar .......................................................................................................................................................... 95
2.3. Menu Panel ............................................................................................................................................................. 96
3. OPENING STUDY ............................................................................................................................................................ 102
3.1 Opening study from Worklist ................................................................................................................................. 103
3.1.1 Remote Worklist ................................................................................................................................................................ 103
3.1.2 Local Worklist .................................................................................................................................................................... 108
3.2 Opening local study ................................................................................................................................................ 111
4. RADIOGRAPHIC EXPOSURE ............................................................................................................................................ 112
4.1 Setting the exposure parameters ........................................................................................................................... 115
4.2 Setting the ACR code .............................................................................................................................................. 116
5. IMAGE TOOLS ................................................................................................................................................................ 123
5.1 Display & Images .................................................................................................................................................... 124
5.2 Fit to Window, Pan/Scroll, 1:1, Effective size, Lens ................................................................................................ 126
5.3 Set anchor point ..................................................................................................................................................... 127
5.4 Window/Level, Contrast, Histo, Restore original ................................................................................................... 128
5.5 Process, Info, Change ACR settings ........................................................................................................................ 130
5.6 Downsize to 18x24, Delete, Recover 24x30, Measure, Plot Line, Alt Images ......................................................... 133
6. CLOSING STUDY ............................................................................................................................................................. 138
7. LOADING STUDY FROM PC ............................................................................................................................................. 139
8. DICOM FUNCTIONS ....................................................................................................................................................... 141
9. DAILY CHECK .................................................................................................................................................................. 154
SECTION 8: MAINTENANCE ......................................................................................................................................... 164
GENERAL CONCEPTS ......................................................................................................................................................... 165
GRID MOVEMENT TEST ..................................................................................................................................................... 166
CLEANING AND DISINFECTION .......................................................................................................................................... 167
DETECTOR FANS FILTERS CHECK AND REPLACEMENT ....................................................................................................... 169
MOVING THE UNIT TO ANOTHER PLACE ........................................................................................................................... 170
MODM10C-U02
Page 4 of 226
OPERATOR’S MANUAL
SECTION 9: ACCESSORIES & OPTIONAL PARTS ............................................................................................................. 171
LIST OF PARTS .................................................................................................................................................................... 172
LIST OF OPTIONS ............................................................................................................................................................... 172
CONFIGURATIONS AVAILABLE ........................................................................................................................................... 174
COMPRESSION PADDLES ................................................................................................................................................... 177
DEVICE FOR GEOMETRIC MAGNIFICATION WITH VARIABLE (1,5X/2X) .............................................................................. 178
FOOT‐CONTROLS ............................................................................................................................................................... 178
PHANTOM FOR PERIODICAL SYSTEM CALIBRATION .......................................................................................................... 179
PHANTOM FOR DAILY CHECK WITH COMPRESSION .......................................................................................................... 180
STEREOTACTIC BIOPSY DEVICE .......................................................................................................................................... 180
ANTI‐X PROTECTIVE BARRIER ............................................................................................................................................ 181
ACQUISITION WORKSTATION ............................................................................................................................................ 181
SECTION 10: SPECIAL HINTS – IMAGE RECOVERY STRATEGY – TROUBLESHOOTING ...................................................... 182
PATIENT POSITIONING ...................................................................................................................................................... 183
IMAGE RECOVERY STRATEGY – TROUBLESHOOTING ........................................................................................................ 192
Incorrectly sized images acquired with 18x24 compression paddle ............................................................................. 192
Incorrectly sized images acquired with 24x30 compression paddle ............................................................................. 193
Images with metallic clips, surgical marker and cluster of micro calcification ............................................................ 194
Images with surgical anatomical parts ........................................................................................................................ 195
Images with breast specimen ....................................................................................................................................... 196
Wrong ACR setting ....................................................................................................................................................... 197
Spot compression device (if available) ......................................................................................................................... 198
SECTION 11: TECHNICAL SPECIFICATIONS .................................................................................................................... 201
TECHNICAL SPECIFICATIONS .............................................................................................................................................. 202
ELECTROMAGNETIC INFORMATIONS ACCORDING TO IEC 60601‐1‐2 ............................................................................... 214
WEEE INFORMATIONS ACCORDING TO DIRECTIVES 2012/19/EU AND 2002/96/EC ........................................................... 218
(WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT) ..................................................................................................................... 218
SECTION 12: WARRANTY CONDITIONS ........................................................................................................................ 219
WARRANTY CONDITIONS .................................................................................................................................................. 220
FAILURE REPORT ............................................................................................................................................................... 222
SAFETY FEEDBACK ............................................................................................................................................................. 224
MODM10C-U02
Page 5 of 226
OPERATOR’S MANUAL
SECTION 1:
GENERAL INFORMATION
INTRODUCTION
Proper use of this equipment presupposes that the operating personnel is thoroughly familiar
with the operating instruction, in particular the sections Safety, Function and Safety Check,
and Protective Measures and able to use Personal Computers and their applications.
If legally-binding regulations govern the installation and/or the operation of the above
equipment, it is responsibility of the installer or the operator to observe these regulations.
INTENDED USE
The MELODY III D C is a system indicated for generating mammographic images that can be
used for screening and diagnosis of breast cancer or for needle localization in case of
stereotactic biopsy. The MELODY III D C is intended to be used in the same clinical
applications as traditional film/screen systems.
CONTRAINDICATIONS
No absolute contraindications are given for mammography. Due to the nature of X-ray
procedures the patient is exposed to radiation and adverse health effects exist and are
known.
GENERAL DESCRIPTION
The FFDM DIGITAL MELODY III D C is complete mammography solutions optimized for
digital imaging. They are composed of:
Mammography Unit with X-Ray tube and Indirect Conversion Amorphous Silicon
detector 24x30 cm format or Direct Conversion Amorphous Selenium 24x30 cm format
with not Isocentric (MELODY III D C) or Isocentric C-Arm (MELODY III D C) usable for
Stereo biopsy in combination with SBD FFDM Stereo biopsy device;
Integrated X-Ray control and image Acquisition Software (AS);
Although not as a part of the FFDM MELODY III D C and systems (which are just complete
mammography solutions when composed by components described above), an optional
acquisition station (Anti-x protective barrier equipped with devices for remote control of the
unit) may be provided on request.
Indirect Conversion Amorphous Silicon Detector is recommended for installation in mobile
environment (E.g. Truck, Van, etc) or critical installation in general.
Histograms
Override ACR settings
Mirroring operations on ACR standard views (CC, MLO)
Post-processing filters dedicated to detection of metallic clips, specimens and
surgical breast parts
- DICOM tools:
Worklist studies opening and closing
Sending of locally stored studies to the appropriate RIS server
Studies printing
Studies recording on CD/DVD
QC tools based on EUREF protocol
Storage commitment (SC)
Modality Performed Procedure Step (MPPS)
2 MPixel Color LCD Display System (21.3” viewable size)
Wireless Keyboard with touch pad
Integrated UPS
- UPS 180 / 266 Vac, 650 VA (230 Vac) or 700 VA (115 Vac), 10 minutes
(typically)
- Intelligent power management system with safe automatic shutdown
HIS-RIS-PACS Interface
Conformities:
- FULL DICOM 3.0 MG:
- IHE (PIR-Patient Information Reconciliation, SWF-Scheduled Workflow, MI-
Mammography Image)
A Viewing/Reporting Station is available in option (not part of this Medical
Device)
SAFETY SYMBOLS
SAFETY INFORMATION
13. Frequently verify the wear of the compression paddles to prevent damages as
cracks and tears, and consequent risks for the patient.
14. Do not modify this equipment without authorization of the manufacturer
15. Use only original accessories and spare parts.
16. Do not remove plastic coverings of mammography unit that give a protection
against the electrical, thermal and mechanical hazards.
17. Do not insert in the x-ray beam devices other than compression paddles or
magnification device.
18. Use lead apron for patient protection.
19. During x-ray emission, operator must be behind anti-X protective barrier and in
a position where it is possible to watch patient and unit.
20. Respect the storage conditions.
21. Pay attention to video monitor of mammography unit that: it is the most fragile
part of the system.
22. Use the mammography unit and its software according to the instruction given
in this manual. Do not try to install unauthorized software, access to operating
system configuration or perform other potentially dangerous operations.
23. Do not use other equipment or network/data couplings (to which a signal
input/output part may be connected), other than those forming part of
MELODY III D C.
24. This is a medical equipment and it should not be considered to be a general-
purpose computer: when operating the system, do not attempt to make any
change to system software and do not use it as a personal internal PC. Any
modification or attempt to modify the operating system software will make the
unit unsuitable for its intended use.
25. No computer virus protection or network security is provided. anti-virus
protection and network security are the ongoing responsibility of the user.
26. When Mammography System Is Installed In Mobile Environment (E.G. Van,
Truck), It’s Necessary Before Cruise Remove Any Accessory From Unit, Move
The C-Arm To The Bottom Position And Lock It By Means Of Its Special
Support.
27. The device incorporates a low risk LED (risk group 1) for light beam centering
device (according to CEI EN 62741).
28. Never Leave Unsolved Problems That May Affect The Safety Of The Product.
29. The pictures present in this manual are only indicative and may be subjected
to changes (for example the background color) which are not significant for the
procedure described
INTERNATIONAL SYMBOLS
The meaning of symbols appearing on the plate and on some components of the machine is
specified in the following table.
SYMBOLS CEI–IECSTANDARD DEFINITION
ISO 7000-0434A
Caution
IEC 60417-5008
“OFF” Power
IEC 60417-5010
“ON” Power
IEC 60417-5840
Type B Applied Part
IEC 60417-5032
Alternating Current
IEC 60417-5019
Protective Earth (ground)
IEC 60417-5017
Earth (ground)
ISO 15223-1
Manufacturer
ISO 15223-1
Date of Manufacture
ISO 15223-1
Serial Number
ISO 7010-M002
Refer to instruction
manual/booklet
NOTE On ME EQUIPMENT “Follow
instructions for use”
IEC 60417-5367
C-ARM POSITION
IEC 60417-5638
Emergency Stop
IEC 60417-5326
Large Focus
IEC 60417-5325
Small Focus
IEC 60417-5384
Indication of radiation field by light
IEC 60417-5349
Radiodiagnostic compression device,
movement
IEC 60417-5350
Radiodiagnostic compression device,
pressure applied
IEC 60417-5351
AUTOMATIC COMPRESSION RELEASE
IEC 60417-5353
Grid movement
IEC 60417-5339
IEC 60417-5172
Class II equipment (according to IEC
60536)
Breast THICKNESS
ERROR MESSAGE
Focal Spot
ACRONYMS
OD Optical density
DEVICE CLASSIFICATION
The identification label is placed on the back of the device and a true copy of it is also placed
inside.
X-RAY TUBE
(alternative labels are
provided in case of optional
tube chosen)
X-RAY tube
H.S. STARTER
(optional)
COLLIMATION
DEVICE
C-ARM
COMPRESSION
SYSTEM
MOTORIZED
ROTATION
OF C-ARM (optional)
POTTER-BUCKY
INVERTER
UNIFIED FRAME
DETECTOR
Z150-dwg050-00
MAGNIFICATION DEVICE
HIGH VOLTAGE
GENERATOR
INTERNAL LABELS
Identification labels are
NOTE important for a safe use of the
mammographic unit.
X-RAY TUBE
(alternative labels are
provided in case of optional
tube chosen)
X-RAY tube
H.S. STARTER
(optional)
COLLIMATION
DEVICE
C-ARM
COMPRESSION
SYSTEM
MOTORIZED
ROTATION
OF C-ARM (optional)
INVERTER POTTER-BUCKY
UNIFIED FRAME
DETECTOR
Z150-dwg051-00
MAGNIFICATION DEVICE
H.IGH VOLTAGE
GENERATOR
Weight: 315 Kg
Weight: 360 Kg
DIMENSIONS IN mm
DIMENSIONS IN mm
Weight: 85 kg
If examination room is not already provided with operator protection barrier, standalone
protective glass is available as accessory.
According to local regulation for radiological protection, the access to the equipment and to
the protected area must be restricted to the authorized personnel only.
SECTION 2:
CONFIGURATIONS
Digital Imaging and Communications in Medicine (DICOM) is a standard for handling, storing,
printing, and transmitting information in medical imaging. It includes a file format definition
and a network communications protocol. The communication protocol is an application
protocol that uses TCP/IP to communicate between systems. DICOM files can be exchanged
between two entities that are capable of receiving image and patient data in DICOM format.
DICOM enables the integration of scanners, servers, workstations, printers, and network
hardware from multiple manufacturers into a picture archiving and communication system
(PACS). DICOM has been widely adopted by hospitals and is making inroads in smaller
applications like doctors' offices.
DICOM consists of many different services, most of which involve transmission of data over a
network, and the file format below is a later and relatively minor addition to the standard.
MELODY III D C is classified FULL DICOM 3.0 MG; it means that this device implements the
totality of the Classes of Service provided for mammography.
Depending on the installation site and preferences, MELODY III D C can be configured as
described in the following pages.
Three different configurations are possible. The differences are based on the presence or not
of the DICOM net and on the supplying or not of the diagnostic station.
This configuration of the MELODY III D C can be used when a DICOM net and a remote
diagnostic station are already available on the installation site.
The mammography system, connected to the PACS, allows to open a study from the Worklist
(remotely generated); closing the study, the acquired images and all exam parameters are
sent to the PACS, synchronized with the Worklist and stored on the appropriate server of the
RIS. Then, the images can be examined and reviewed from remote diagnostic station.
If the DICOM Worklist server is out of service, the mammography system is anyway able to
work. In this case, the images and study data are stored in internal memory of computer.
When the normal functionality of the net is reactivated, the studies are newly synchronized
with the Worklist and stored on the server.
This configuration of the MELODY III D C can be used when a DICOM net already exists on
the installation site but there is not a remote diagnostic station specific for mammographic
review.
With the mammography system, set up as in the “FULL DICOM” configuration, is supplied
also a diagnostic station with a dual headed display system at very high resolution and a
review software dedicated to the mammography.
Mammography system and diagnostic software are both connected to the RIS/PACS servers.
The Software can be configured to use mammographic specific hanging protocol and reading
workflow. It displays also images from other modalities (digitized prior mammogram films,
breast US, breast MRI, breast NM, breast CT) and allows selection, display, manipulation,
markup, print composition and media exchange of breast images.
The software can be integrated with an optional Computer Aided Detection (CAD).
“mini-PACS” CONFIGURATION
This configuration of the MELODY III D C must be used when a DICOM net is not available
on the installation site.
MELODY III D C in mini-PACS configuration represents a complete individual island of digital
imaging. It has the form of a localized, modality-specific network of modalities and storage
directly connected to reading workstations without long-term storage or management.
Often, Full Field Digital Mammography (FFDM) request a similar approach, due to the large
image size, highly specialized reading workflow and display requirements.
Also in this case, the mammography system supports connections with all DICOM modalities.
SECTION 3:
MAIN PARTS DESCRIPTION
5
2
12
3
8
18
4
14
13
7
15 10 16
17
6
11
MAMMOGRAPHY UNIT
1. STAND
2. C-ARM
3. X-RAY TUBE
4. PROTECTIVE SCREEN
5. C-ARM ROTATION HANDLES (on both sides)
6. FULL FIELD FLAT PANEL DETECTOR 24x30 cm FORMAT
7. MAMMO UNIT SWITCH ON
8. LCD COLOR MONITOR (if provided)
9. COMPRESSION FOOT PEDALS
10. CONTROLLER FOR FINE MANUAL COMPRESSION (on both sides)
11. AUXILIARY DISPLAY
12. C-ARM ROTATION PUSH-BUTTONS
13. C-ARM UP/DOWN
14. EMERGENCY PUSH-BUTTONS (on both sides)
15. COMPRESSION PADDLE
16. WIRELESS KEYBOARD
17. CONNECTOR PANEL
18. X-RAY PUSH-BUTTON
1
2
5
3 12
18
4 13
10
14
7
15
16
6
17
11
MAMMOGRAPHY UNIT
1. STAND
2. C-ARM
3. X-RAY TUBE
4. PROTECTIVE SCREEN
5. C-ARM ROTATION HANDLES (on both sides)
6. FULL FIELD FLAT PANEL DETECTOR 24x30 cm FORMAT
7. MAMMO UNIT SWITCH ON
8. LCD COLOR MONITOR (if provided)
9. COMPRESSION FOOT PEDALS
10. CONTROLLER FOR FINE MANUAL COMPRESSION (on both sides)
11. AUXILIARY DISPLAY
12. C-ARM ROTATION PUSH BUTTONS
13. C-ARM UP/DOWN
14. EMERGENCY PUSH-BUTTONS (on both sides)
15. COMPRESSION PADDLE
16. WIRELESS KEYBOARD
17. CONNECTOR PANEL
18. X-RAY PUSH-BUTTON
22
21
20
19
19. X-RAY PUSH-BUTTON
20. WIRELESS KEYBOARD
21. LCD COLOR MONITOR
22. ANTI-X BARRIER
Double configuration
AUXILIARY DISPLAY
1 2 3 6
deg cm N
L R
XCC CV AT ID TAN S RL RM
5 4
1. C-ARM INCLINATION
2. COMPRESSED BREAST THICKNESS (mm)
3. SELECTED COMPRESSION FORCE (less bright)
REALLY EXERTED COMPRESSION FORCE (most bright)
4. LATERALITY L/R SELECTION
5. PROJECTIONS, PREFIXES AND SUFFIXES (ACR protocol)
6. COMPRESSION FORCE ADJUSTMENT
Z150-dwg055-00
1
ON BOTH SIDES
Z150-dwg056-00
10 11
2
13
3
5
6 7 8 9 12
1
2
ON BOTH SIDES
1. ROTATING CONTROLLER (TO SELECT COMPRESSION FORCE OR TO ALLOW FINE
MANUAL COMPRESSION)
2. PUSH-BUTTON TO ACTIVATE THE FUNCTION:
COMPRESSION FORCE TARGET SELECTION (ON)
FINE MANUAL COMPRESSION (OFF)
FOOT-CONTROLS
CONNECTOR PANEL
1. UPS SWITCH ON 2
2. RX PUSHBUTTON CONNECTOR
3. LAN 3
4. USB
5. FOOT-CONTROLS CONNECTORS 4
(one for standard configuration)
5
OPTIONAL
SECTION 4:
GENERAL OPERATIONS
EXAMINATION ROOM
X-Ray units can only be operated inside dedicated room provided with X-ray protection that meets
local standards and regulations.
In case of first installation, the Technical Service or the person authorized has to follow the specific
instructions contained in the Service Manual.
SWITCHING ON
the dedicated paragraph) with integrated lamp, lamp is bright when UPS is ON.
2. Switch ON the unit by means of push button n°7 (see paragraph “main parts”), system
The software runs automatically. At the beginning, it shows a thermo graph with temperature
room trend in the time. If room temperature exceeded upper or lower limits, according to the
range fixed with detector on board, an error message appears (refer to Section 7:
Maintenance).
NOTE thermo graph. The two red lines in the thermo graph window
indicate the acceptable temperature range of FFDM unit
The thermo graph window disappears in few second, then the home page of DME (Digital
Mammography Equipment) launcher is shown.
The operator can select the fast launch of DME Acquisition software clicking on icon at the
middle of display.
DME Acquisition
DME Toolkit
Do not click on the icon located on the lower left corner of the display (crossed tools), this is a
service function only for technicians.
If the operator wrongly clicks on the icon DME Toolkit, he can easily return on the main menu
clicking on the following icon, located in the upper right corner of the next window:
During system initialization, after the DME Acquisition software launch, the unit checks the
presence and the functionality of all devices and components to which is connected.
If one of the following warning icons appears, start sequence will proceed after clicking “OK”
on both icons.
If both icons are displayed, switch off the unit and restart it again.
The unit will not be usable for normal use e.g. image acquisition as evident by the lower bar
shown below:
After about 5 minutes, the User Interface is shown on display of acquisition Software and a
green message on the down bar reveals that mammography system is “ready” to use.
BLACK OUT
The unit is provided of Intelligent Power Management (IPM) with Automatic shutdown
capability.
During a black out, the operator is advised by an audible sizzle noise and by a series of four
short beeps emitted by UPS every 30 seconds. Beeps become continuous when the battery
is low and it is about to switch off.
In case of intermittent or permanent black out there is no need of closing the study in
progress and manually shutting down the unit, it will be done automatically on low battery
condition (manufacturer set 50% percentage of battery charge as lower limit for automatic
shutdown) .
If blackout persist, please switch off the UPS for saving the
NOTE
remaining battery charge after automatic shutdown.
If necessary, percentage of available battery charge (%) can be read by clicking on the
specific icon of Windows OS
During black out no C-arm movements, compression and exposure are possible.
Emergency patient release is possible by means of fast unlock of compression paddle holder
as described in section 4: General operation - “Emergency patient release”
SWITCHING OFF
Verify that any study, open locally or from worklist (see Section 7 of this manual), is closed.
Switch OFF takes place in two steps:
1. Close DME Acquisition software clicking on red icon located on the upper right
corner of the display.
1 2
2. Switch OFF the mammography system by clicking again on the upper right
corner of the display.
EMERGENCY PUSHBUTTONS
There are two emergency push-buttons on the MELODY III D C, one on each side of the
Mammography Unit (n° 14 – paragraph “Main Parts”).
In case of emergency, to immediately stop any movement and disable HV generator, press
one of the emergency push buttons.
The following message appears immediately when the emergency pushbutton is pressed:
Once the emergency has been solved, before to re-use the mammography system, it’s
necessary to follow these steps:
C-ARM MOVEMENTS
All movements UP/DOWN and rotation are disabled when compression force is applied.
On the MELODY III D C the C-Arm rotation is isocentric. It allows all breast projections
without moving the patient and without adjusting the height of the C-Arm.
“ ” = UP “ ” = DOWN
C-ARM ROTATION
Manual rotation is available operated by means of push button n°12 (see paragraph “main
parts”). The mammography unit has a safety system with a sliding clutch calibrated on a
limited force to avoid any lesion or damage if C-Arm during rotation runs into a person or an
object.
MELODY III D C version has limited rotation +/-130°.
The rotation is available operated by means of push button n°12 (see paragraph “main parts”)
With a single pressure on “CC” button the C-Arm rotates automatically to “0°” angular position
(factory default).
With a single pressure on “LAT” or “ OBL” button the C-Arm rotates automatically at pre-
selected angular positions.
“CC” = 0°
A continuous pressure of “CW” or “CCW” buttons rotates the C-arm in ±180° (±130° for
isocentric C-arm) range respect to 0° position.
“ ” = CW “ ” = CCW
After a rotation movement on a pre-selected angle (“LAT” or “OBL”) the operator can adjust
the C-Arm inclination pressing the Fine Adjustment angle buttons”.
In this case, the pressure of opposite “OBL” button activates the C-Arm rotation at the
opposite position respect to Fine Adjusted angle.
The rotation angle and the standard code of mammography view (according to ACR
standard) are shown on all displays.
deg cm N
L R
XCC CV AT ID TAN S RL RM
“S” = SCOUT
To insert a compression paddle tighten the internal knobs of locking system in order to retract
the four pins.
Place the compression paddle on locking system and release the knobs to lock it.
To remove compression paddle tighten the internal knobs of locking system and pull it.
TO LEFT TO RIGHT
To move the compression paddle with lateral shifting from middle to right or to left push it
towards desired direction. The paddle stops into place with an audible click.
The compression force can be set by Control Panel by means of buttons n° 8 (see the
paragraph “Control Panel”) or rotating the encoder for fine manual compression (see the
paragraph “controller for fine manual compression”) after having activated the function by
means of push-button n° 2.
The selected value (in N) is shown on all displays:
deg cm N
L R
XCC CV AT ID TAN S RL RM
The range of compression can be set from 70 to 200 N by means of internal setting.
Motorized compression is operated by the foot –controls n°9 (see the paragraph “Main
Parts”).
1 3
Pushing the pedal n°1 (see the paragraph “foot control for motorized compression”), the
compression paddle goes down with a speed that decrease at a certain height from the table.
When the compression paddle comes in contact with the breast, it has a very brief stop and
start to apply an increasing force.
With the function of Tissue Strength Evaluation FTSE, a microprocessor checks the breast
tissue strength.
if the breast is soft, the compression force increases up until value set by the operator.
if the breast is dense, the compression force stops to 100 N. In this case, the operator
can proceed the compression up until the force value set releasing and pushing again
the pedal n°1.
If the operator release and push again the pedal n°1, the compression force increases
up to the maximum value set (200 N).
Pushing the pedal n°2, the operator reduces the compression force.
Pushing the pedal n°3, he operates a fast release of breast compression.
Manual compression is possible acting on the controllers placed on both sides of C-arm.
The continuous rotation of the controller n° 1 in a direction (see the paragraph “Controller for
fine manual compression”) allows to bring down the paddle, touch the breast and press it until
to reach the desired force value. The compression force can increase up to maximum value
set (200 N).
The continuous rotation of the controller n° 1 in the other direction (see the paragraph
“controller for fine manual compression”) allows to rise the paddle reducing the force.
In any time, the operator can pass from motor driven to manual compression and vice versa.
The force really exerted on the breast at the end of compression is shown both on the display
of control panel and on the auxiliary display.
Compression release
There are two different operating modes for compression release at the end of the exposure:
- AUTO SOFT RELEASE (normal mode)
- MANUAL RELEASE (for 2D breast biopsy)
The operator can choose the release mode on control monitor by means of button n° 7 (see
the paragraph “Control Panel”). The selected mode is shown on the area n° 6 of the display.
When necessary and when it is present, is possible to rise quickly and automatically the
compression paddle, pushing the button n° 3 of foot-controls (in case of motorized
compression).
In case of power failure and breast just under compression, the operator may get free the
patient using the fast mechanical unlock as described in the previous paragraph
“Compression Paddle insertion and removal”.
In case of power failure of mammography system with breast under compression, the
operator may get free the patient using the fast mechanical unlock.
To remove compression paddle tighten the internal knobs of locking system and pull it.
Every time the compression down pedal is used, the luminous field indicator light switches on
for about 20 seconds. The compressor paddle goes down with a little delay, so that if the
breast is already compressed the operator can switch on the light, avoiding additional
pressing. The compression movement is immediate when the light beam is already on.
Magnification ratio 1,5x or 2x is selected by pulling the two lateral spring-loaded plungers at
the same time, and fix table height at requested position. Release the plungers and verify that
locator pins lock the device for geometric magnification.
PROTECTIVE SCREEN
Mammography unit is provided of a protective screen to keep patient’s face and extraneous
parts (i.e. shoulder, earring, hairs, etc.) out of X-Ray beam. The presence of extraneous parts
in the image filed negatively affects the functioning of presentation algorithm, with impact on
breast border recognition, worse contrast and dynamic range.
In order to fix the protective screen, insert the rear side of his frame in the guide ways under
the X-Ray window and push the protective screen reward to lock it.
On MELODY III D C the selection of focal spot is automatic; large focus with Potter-Bucky
and small focus with device for geometric magnification.
No manual selection allowed.
The selected focal spot is shown on control panel.
RADIOGRAPHIC EXPOSURE
Radiographic exposure is allowed only when study is open from Acquisition Software and
breast is compressed. Otherwise error messages like “Open study first” or “Apply
compression force” alarm will appear.
In semi-automatic and fully automatic exposure technique (see Section 5 of this Manual), the
pre-exposure pulse is announced by a short sound signal and by a bright signal (see
paragraph “Control Panel”).
Exposure is announced by a long sound signal and by a bright signal on control panel (see
paragraph “Control Panel”).
After X-ray pushbutton is pressed there’s a few delay, depending on the detector
synchronization, before the tube starts running.
If the X-ray pushbutton is released, the X-ray emission stops immediately.
At the end of exposure, the compression is released (with the function “Auto release”
enabled) and the paddle rises automatically.
DOSE CALCULATOR
The MELODY III D C mammography unit automatically provides calculated values of the
Average Glandular Dose (AGD).
After each exposure, the dose in mGy is stored in the DICOM header and total dose is
displayed in real time after any exposure on the Operation Bar (Menu Panel of Acquisition
Software).
For proper dose calculator operation, it’s necessary to use the appropriate compression
paddle as described in the chapter “Typical Configurations”.
Shifted paddles have to be used with large focus and straight paddles with small focus
because focal spot to skin distance measurement depend on it and dose calculation too.
All the functions of MELODY III D C Mammography System are controlled by microprocessor
with unique safety features. The monitor shows textual error messages (in several languages
selectable in installation phase) related to an acoustic signal.
OPERATING ERRORS
APPLY COMPRESSION FORCE
Before making exposure breast must be compressed and compression force must be higher
than 0
DOOR OPEN
Close the door of diagnostic room.
MOVE C-ARM UP
Only for ISO C_ARM move UP to avoid collision with floor when rotating more than 130
degrees to Caudo Cranial position.
REMOVE COLLIMATOR
Additional magnetic collimation plate must be removed for a proper operation of current
operating configuration
Emergency RED pushbutton/s has been activated; C-arm movements, compression and
exposure are disabled. Emergency pushbutton/s require a manual action for restoring normal
position and normal operation will be available after 30 seconds delay.
MAINS UNSTABLE
Mains unstable message appears when the supply conditions are not satisfactory because
supply voltage is low or supply voltage is unstable. The operative conditions are not
guaranteed with this error message.
ERROR MESSAGES
DMELauncher
Thermo Alert (optional)
In case of this kind of warning message please follow the instruction shown and turn off the
unit. Numbers and message shown are purely illustrative
In case of this kind of warning message please follow the instruction. Numbers and
message shown are purely illustrative
Click “OK” to reset error if the reason for it has been removed.
If not, after few seconds, it will show up again except for messages related to power
problems or black out for which error messages, if persistent, will not show up again after first
reset but only when attempting to star x-ray exposure.
SECTION 5:
EXPOSURE TECHNIQUES
To select the Exposure Technique, press repeatedly the button n° 5 (paragraph “Control
Panel”) on mammography unit or click on the button n°1 (paragraph “Remote Control Panel”)
on Acquisition Software.
MANUAL MODE
In manual mode the operator can select kV and mAs in the appropriate range allowed. The
unit takes care automatically of max tube loading conditions.
Manual mode selected is shown by small round hand-shape icon, as shown below:
In manual mode the exposure parameters (kV and mAs) can be selected on the control panel
of the Acquisition Software, clicking on the arrow icons shown below:
The kV values change through 0.5 kV steps. The selectable standard range is 20 ÷ 35 kV
(20-40 kV optional). The mAs values change in log scale.
A Fast selection of mAs and Kv can be used in manual mode holding down the left button of
keypad or the button on arrow icons shown below:
The Automatic Exposure mode (AEC) is the normal operating mode of the unit, having Full
Auto (kV, mAs) and Auto mAs (manual kV and automatic mAs) modes.
Exposure parameters are defined by two different techniques as follows:
PRE-exposure technique is based on breast density identified by means of a short X-
ray pre-exposure. Breast density is averaged over a mosaic of 96 sensitive areas and
exposure parameters (kV, mAs) are selected accordingly.
FAST exposure technique is based on compressed breast thickness and kV, mAs are
selected accordingly.
Manual, Auto mAs, Full Auto, Pre/Fast are selected by means of pushbuttons below, modes
and techniques currently selected are shown by means of small round hand-shape icon:
Exposure parameters and Dose are automatically selected in order to full fill the EUREF
Protocol limits.
AGD in mGy and W/Rh Anode/filter combination: W/Rh
Pre and Fast techniques are provided to manage specific conditions as follows:
Pre is the standard technique for screening and is also recommended for post surgical
intervention examination for which compression force could be painy or dangerous.
Fast is highly recommended for breast inplant being kV and mAs, thickness related.
The automatic filter Rh/Ag provides the most appropriate ANODE/FILTER (W/Rh-W/Ag) to
guarantee the lowest Dose to the patient.
W/Rh anode/filter combination is the standard for small and medium breasts.
W/Ag anode/filter combination is provided for large or dense breast.
Trip point after which W/Ag combination is selected, corresponds to combination of thickness
and density corresponding to about 6 cm of RMI Tissue Equivalent phantom.
Anode/filter combinations
Anode/Filter combinations are selected automatically and combination used for the last
exposure will be shown at the right side of the icon of Control Panel.
Selected mode will be shown at the right side of the icon of Control Panel:
SECTION 6: TYPICAL
OPERATING
CONFIGURATIONS
The different exposure techniques available with mammography unit MELODY III D C are
shown in the next six pictures.
LARGE FOCUS
PROTECTIVE SCREEN
24X30 cm FORMAT
POTTER-BUCKY
LARGE FOCUS
PROTECTIVE SCREEN
24x30 cm FORMAT
POTTER-BUCKY
PROTECTIVE SCREEN
24X30 cm FORMAT
POTTER-BUCKY
SMALL FOCUS
24x30 cm GEOMETRIC
MAGNIFICATION DEVICE
0,3 mm
LARGE FOCUS
PROTECTIVEE SCREEN
24X30 cm FORMAT
POTTER-BUCKY
STEREOTACTIC
BIOPSY DEVICE
SBD FFDM 24X30
SECTION 7: ACQUISITION
SOFTWARE OPERATING
INSTRUCTIONS
A the end of tuning phase, after about 5 minutes, the User Interface is shown on display of
Acquisition Software and a green message on the down bar reveals that mammography
system is “ready” to use.
The User Interface is configured in order to show information regarding the functionality of
mammography system and all applicable items.
It is divided in three areas: “Status Bar”, “Operation Bar” and “Menu Panel”.
Operation Bar
Menu Panel
Status Bar
The Status Bar is divided in five sections and includes a calendar with local time.
The first section indicates the “Status” of the device. The possible states are:
Message Meaning
Operational Normal condition. The system is ready and able to
make exposures.
The second section indicates the operating state of the device. The possible states are:
Message Meaning
DICOM Normal operating mode; the connection of the
mammography unit from and towards the outside
(DICOM net or Diagnostic Subsystem, depending on
the configuration) is working correctly.
The third section indicates if the mammography system is sending a study to a DICOM
server. The possible states are:
Message Meaning
First phase of building of the DICOM file.
The fifth section indicates the status of the detector. The possible states are:
Message Meaning
ready The detector is ready to receive the X-rays. This is
the normal condition when there is an open study
(and no elaboration is going on).
not ready The detector is not ready to receive the X-rays. After
the detector receives the X-rays and acquires the
image, it passes through this state before being ready
for a new exposure.
The following status diagram shows the sequence of states which the detector passes
through.
The Operation Bar contains also the button to log off the software.
The Menu Panel includes the most important tools to operate with the Acquisition Software.
It includes four submenus: “Mammo”, “PC”, “Graphic” and “DICOM”.
After switching the unit, during the initial tuning of the system, the menus are not available.
Then, all the menus are available, but two exception exist:
In such circumstance only Load study function is available, for loading and reviewing
studies from the local archive.
2. The “DICOM” menu is not available and all buttons of this menu are grey-colored
The problem can be due, for example, to the PACS of the medical center that is
temporarily not available, or to the connection cable that has been cut. Click on the
menu to verify if the connection with the DICOM net has been restored,. If after a few
seconds the question point disappears, the connection is available and it is possible to
open the menu clicking on it. Otherwise the question point persists and the menu is not
available. In this case it is not necessary to stop working. Anyway, it is possible to
continue acquiring images and store studies locally, on the internal memory of the
Computer, and when the DICOM connection is restored the studies can be sent to the
PACS.
A quick description of each menu follows; the functions are explained better in other sections
of this manual, or other chapters of this section.
Mammo: it serves to set the exposure parameters, to choose the achievable dose
and to check the ACR view.
Clicking on upper side of Mammo menu the Control Panel appears, and related commands
are available.
Clicking on lower side of Mammo menu the operator can choose between “Low dose” and
“High contrast” operating modality.
Both modalities allow to obtain dose values less then established by European Guidelines;
the “High contrast” modality is preferable to view micro-calcifications, the “Low dose” modality
in preferable in screening or diagnostic routine exams in order to reduce X-ray emission to
the patient.
NOTE preferable that all projections are performed with the same
operating modality to obtain the maximum omogeneity of images.
By means of this menu it is possible to open new studies locally (i.e. not from the worklist),
save the studies locally (i.e. in the internal memory), reload the local studies (i.e. resident in
the internal memory), perform three-dimensional biopsy procedures (only on MELODY III D C
with isocentric arm). The functions are available depending on the status of the device.
Moreover, always by means of the PC menu, it is possible to insert the following information:
Operator’s name;
Institution’s name.
To allow this data integration: click on PC label (CTRL right + mouse double click), fill the
corresponding fields and then click on “OK” to save them:
Graphic: this menu serves to manipulate the images shown on the display.
It is used to select the images to display, to position them in a desired layout. The
functions are available only if there is a study open (from worklist or locally) or loaded,
otherwise the menu is partially disabled.
DICOM: this menu includes all the commands of the Acquisition Software that
involve DICOM functions.
It is used to open and close a study from the Worklist, send studies stored on the local PC
to an appropriate server of the RIS, change the exam typology of the study, integrate a
study with other images, print images and make CD of the studies according to the
DICOM protocol.
The menu functions are available depending on the state of the device. In particular, the
menu items are available or not if there is a study presently open, or there is a study
presently loaded, or nothing.
3. OPENING STUDY
To acquire radiological images, it is necessary that a new study is opened either from the
Worklist (that is remotely generated), or from the local data storage system using the
Acquisition software.
The normal operating mode is opening the study from the Worklist, but if the patient’s data
are not available from the Worklist (as for example because the DICOM Worklist server is out
of service), the unit is able to work: the acquired images and the data of the study are stored
in its internal memory.
When the connection with the Worklist server is restored, the data previously memorized in
the internal memory of the local PC are sent to the PACS and synchronized with the Worklist.
Then the memory of the local PC gets free.
To open a study from the Worklist, it is necessary to know the functions of the “Worklist
Query Dialog” window. To open a study from the local database, it is necessary to know the
functions of the “Study Data” window.
3.1.1 Remote Worklist
To open a study from the Remote Worklist, click on the “Worklist” button and a new submenu
appears.
Clicking on “Remote” button, a “Worklist Query Dialog” window will appear, the Software
automatically will ask the Worklist to the configured server and it will be possible charge data.
To open a study from the worklist, first of all it is necessary to query the worklist server to
obtain the data. For this scope, it is necessary to set the query parameters using one or more
fields present in the “Query Parameters” section.
If you do not want set any parameters for this field, it’s
Requested Procedure Description: this field identifies the kind of procedure (for
example 4 views ACR CC…...)
Accession Number: when the acceptance creates and inserts a new study in the
worklist, this information is automatically created and associated to the study. It is
possible to filter the query to the worklist server with this information.
Once set the parameters, click the button “Worklist Query” and all search results will be
presented.
To open a study double click on the corresponding row or select it and press the button
“Done”. If you want to exit without opening any study, do not select any row and click “Done”.
To open more studies in automatic mode, it is sufficient to select all their correspondent rows
and then press the button “Done”.
In this way, the selected studies from remote Worklist will be automatically and sequentially
opened when closing the previous one.
NOTE verify that the windows of the query parameters section are
correctly filled in.
The Acquisition Software loads the data of the patient, and the Operation Bar shows the data
of the currently open study.
3.1.2 Local Worklist
To open a study from the Local Worklist, click on the “Worklist” button and a new submenu
appears.
Clicking on “Local” button, a window will appear and the operator will be able to create a new
study without a Worklist item and perform an unscheduled IHE operation later.
Additionally, it is possible to identify the local study also by means of an Other Patient ID.
In this case, select the related window and write inside it the other’s identification number or
code used to identify the patient univocally:
To open a study locally, click on the “Local Study” button of the “PC” menu.
Identify the study associating to it a unique accession number, then click on OK button.
Before acquiring images, it is possible to abort the study by clicking on the “Abort Study”
button of the “PC” menu. After the first image has been acquired, this button becomes
disabled.
Opening a local study, at the end of the study the data and
radiological images are saved only in the internal memory.
The study is not yet matched with the worklist request and
WARNING the images are not yet sent to the store server of the RIS.
To complete the work, making the information safe and
available, it is necessary to send the study to the RIS by
means of the DICOM menu, as explained in next chapters.
4. RADIOGRAPHIC EXPOSURE
With a study open (from the Worklist or locally, as explained in the previous chapter 3), it is
possible to expose the patient to the X-rays to acquire the diagnostic image. If there is no
study open, the exposure is not possible.
The operator must position the patient and then set the exposure parameters (operating
mode, kV and mAs) from the control panel of the mammography unit or from the control
panel of the Acquisition Software (as explained in the following paragraph 4.1).
After settings, the breast can be exposed by means of the X-rays pushbuttons on Acquisition
Software. The operator must press down them at the same time up to the end of acoustic
signal.
After few seconds for processing, the image is displayed by Acquisition Software and the
related Average Glandular Dose is reported on Operation Bar.
If the quality of image is good, the operator can proceed with the next exposure positioning
newly the patient and setting the parameters, otherwise will be necessary repeat the
exposure trying to obtain a better quality.
The new image is displayed on the main part of screen, meantime the overall dose of the
study is updated in real time on the Operation Bar.
The orientation of the image depends on the laterality: if the laterality set in the ACR code is
“R” (refer to the following paragraph 4.2), the image is shown with the nipple on the left side;
if the laterality is “L”, corresponding to the left breast, the image is shown with the nipple on
the right.
Repeat the steps “Position the patient – set the exposure parameters – expose – evaluate
the quality”, till the end of the views needed.
NOTE For any study the operator can acquire a maximum of ten images.
To review a previous exposure is enough to select the image clicking on its thumbnail.
The exposure parameters (operating mode, kV and mAs, and anode/filter combination) can
be set on the control panel of the Acquisition Software.
To access the control panel of the Acquisition Software, click on the “Mammo” icon of the
Menu Panel. The control panel will appear to the side.
The operator must choose the preferred operating technique (Full Auto mode, manual mode
or Auto mAs mode), the operating modality (Low dose or High contrast) and choose the
values of kV and mAs. To make this operation refer to the previous section 5 “Exposure
Techniques”.
The ACR code is a standard code created by the American College of Radiology (ACR) to
identify the view, the operating technique and other information of the mammographic image.
Any code is composed by three parts: the prefix, the name of the view, and the suffix.
The complete list of codes is:
A C R C O D E
PREFIX VIEW SUFFIX
CODE MEANING CODE MEANING CODE MEANING
R Right CC CranioCaudal ID Implant Displaced
L Left ML MedioLateral S Spot
M Magnification MLO MedioLateral RM Roll Lateral
Oblique
LM LateroMedial RL Roll Medial
LMO Lateromedial TAN Tangential
Oblique
FB From Below
SIO Superlateral to
Inferomedial
Oblique
XCC Exaggerated
CranioCaudal
CV Cleavage
AT Axillary tail
A complete ACR code that identifies an image contains at least the laterality (“R” or “L”) and
the view, while the other information are not mandatory. For example, to identify a
MedioLateral Oblique projection on the right breast, without other information, the complete
ACR code is:
RMLO
Some codes exclude each other, for example, the prefixes of the laterality “R” and “L” can’t
be present both in the same string. There are also codes that do not exclude any other code;
for example the code for magnification “M” can be present or not in any view. Each code of
the view exclude all the others, for example if a view is “LM”, it can’t be also “CC”, “ML”,
“MLO”, “LMO” etc.
The arrow represents the X-ray beam direction (from the tube
NOTE
to the detector).
According to the DICOM standard, a mammographic image must have an header containing
many information, including the ACR code.
MELODY III D C is able to set some codes automatically, but other codes must be selected
by the operator. For example the insertion of the magnification device automatically sets the
code “M” in the prefix (Magnification); and the inclination of the C-arm automatically sets the
name of the view (but some of them depend also on the laterality, and others are not unique
and can be changed).
In the following list it is evidenced which codes are automatically selected and which not.
R No
M Yes, when the magnification device is
inserted
CC Yes, depending on inclination of C-arm
ML Yes, depending on laterality and inclination
of C-arm
MLO Yes, depending on laterality and inclination
of C-arm
LM Yes, depending on laterality and inclination
of C-arm
LMO Yes, depending on laterality and inclination
of C-arm
VIEW
inserted
RM No
RL No
TAN No
MODM10C-U02 Page 118 of 226
OPERATOR’S MANUAL
The following circles show the sectors of C-arm rotation in which any ACR Code is
selectable:
After positioning the patient, the C-arm inclination and the related default ACR code are
shown both on the auxiliary display and on the control panel of the Acquisition Software.
To set the laterality, click the little patient icon, highlighting the desired breast or push the
button on auxiliary display.
To set an alternative view (different from the default view) click on the ACR code. It changes
cyclically.
To set a further ACR code, select it from the list of the control panel of the Acquisition
Software or from auxiliary display.
Answering “Yes” or “No”, the operator will confirm or not the previous selection.
Clicking on “Yes”, the laterality, chosen before the exposure, will be maintained.
Whereas, clicking on “No”, the SW will rotate the image automatically adapting it to the new
one and correcting, if necessary, the corresponding ACR projection.
The tag relative to the confirmed laterality, will be shown on the corresponding acquired
image as below:
After the confirmation, the operator will modify again the laterality
of an acquired image using the pushbutton “Change ACR
5. IMAGE TOOLS
When one or several images are displayed, it is possible to manipulate them in terms of
visualization as contrast, sharpness, zoom etc. It is also possible to visualize the desired
images choosing among the ones of the currently open study, and set up them in the
preferred position on the screen.
The modifications are not permanent: it is always possible to return to the initial visualization
by means of the “Restore Original” function. The manipulated images are never saved: the
copies of the images saved and stored are the same preexisting before the graphic
manipulation.
To manipulate the images shown on the display, the operator must open a pop-up menu
clicking with the mouse right button on the image
5.1 Display & Images
When a study is open (locally or from Worklist), and the images are obtained by exposing the
patient, they are visualized in the same order as they are acquired. The number of graphic
windows, in which the display is subdivided, increases as the images increase.
When a study is loaded from the local database, the display is soon subdivided in enough
graphic windows to contain all the images of the study. The order in which the images are
set-up in the windows is the same order in which the images have been acquired.
It is also possible to choose the number and position of the graphic windows in which the
screen is divided. And it is possible to position any image in any window, with the maximum
flexibility.
The “Display” submenu of the “Graphic” menu let you choose the number and shape of the
windows in which the screen should be divided.
From the “Display” submenu the operator can select the preferred number and shape of
windows to divide the screen clicking on the corresponding icon. For example, clicking on
“2x2”, the screen is divided as in the following picture.
Any window shows an image, otherwise is empty (black). It is also possible to insert the same
image in more than one window. The operator must drag and drop the same thumbnail to
several graphic windows and the same desired image will be visualized in the chosen
windows.
To understand how the graphic windows are engaged, suppose that they are numbered from
1 to 25 (25 is the maximum number of windows that can be shown at the same time). Any
display visualizes the windows from the number 1 to the number of the displayed images. For
example in the visualization “1x1” only one graphic image is visualized, and it is the number
1; in the visualization “1x2” two windows are shown, so they must be the number 1 and
number 2. And so on.
For example suppose that more than one graphic window is engaged, as for example six
graphic windows contain images.
Now if we choose to divide the screen in less graphic windows than the presently engaged,
for example we choose the “2x2” visualization, the windows remaining are the number 1, 2, 3
and 4, and they remain taken by the same images that were present before.
MODM10C-U02 Page 125 of 226
OPERATOR’S MANUAL
Now if we change again the number of windows visualized, the windows remain taken with
the same concept just explained.
Four further visualization exist: “ACR CC”, “ACR MLO”, “ACR x 4” and “USER”. This
visualizations serve to show the images of the right and left breast having the same ACR
view code. Selecting the “ACR CC” view, the screen visualizes the two images of the
CranioCaudal view for the right and left breast, as shown below.
Selecting the “ACR MLO” view, the two MLO images are visualized.
Selecting the “ACR x 4” view, the two CC and the two MLO images are displayed.
In this window select the checkbox corresponding to the couples of images that you want
see, than click OK. The couples of images selected will be visualized.
To access these functions right click with the mouse on the desired image. All the function
are active only for the window from which the pop-up menu has been activated.
The “Fit to Window” function changes the magnification of the image so that it is adapted to
the graphic window in which is contained.
The “Pan/Scroll” function is active only if the image is not fitted to the window. After selecting
that tool, move the mouse up and down keeping the left button pressed to scroll (i.e. move
the image upward and downward), and right and left to pan (i.e. move the image rightward
and leftward). The operator can use this function even during the execution of another
function by pushing [Alt] and clicking on image.
The “1:1” function visualizes each pixel of the detector on one pixel of the monitor.
The “Effective size” function makes the image displayed as the same size of breast when
exposed.
The “Lens” function opens a square window of 300 x 300 dpi where the image is magnified
with a ratio of 3x. The window can be positioned on any side of the image according to the
mouse position.
To access these functions right click with the mouse on the desired image. All the functions
are active only for the window from which the pop-up menu has been activated.
The “Contrast” function allows to modify the contrast of image in a specific side of breast.
Clicking on “Contrast” button, a submenu appears.
Click on “High” button to set high value of contrast on thorax side; click on “Very High” button
to set very high value of contrast on thorax side; click on “Skin” button to set high value of
contrast at skin level; click on “Flat” button to undo all contrast changes; click on “Restore
original” button to go back to original image before any manipulation.
The “Histo” function allows to display the histogram of grey level distribution of the whole
image and of a selected range of the same.
Click on “Histo” button to open a window with two diagrams.
The upper diagram represents the grey level distribution of the whole image. The lower
diagram represents the grey level distribution of selected range in the above diagram.
In order to select the range click on the left or right edge of the above histogram window to
drag the rulers.
Then, click on “Apply” button to save changes or click “Cancel” button to exit.
NOTE printed.
Operator is prompted to accept or not window level manual
changes when closing study.
In the DICOM fields visible on the review station, Window level/width is included.
In the upper left image default values of “Window level/width: 4096/8192”.
In the upper right image manually modified values of “Window level/width: 6574/9781”.
Images for which Window level/width has been manually modified and saved, will look
different in the same study.
To access these functions right click with the mouse on the desired image.
The “Process” function allows to modify some parameters of reconstruction algorithm in order
to enhance all structures and reducing noise to show the images on display of Acquisition
Software with a reading better quality.
Clicking on “Process” button, a submenu appears.
The operator can choose between five optimized algorithm for magnification, biopsy,
examination of surgical anatomical parts, examination of specimen and in case of metallic
clips, surgical makers or numerous clusters of micro calcifications.
Clicking on “Info” button, a window appears showing all DICOM data and parameters
regarding the selected image.
The “Change ACR settings” function allows to modify the preset viewing configuration.
Clicking on “Change ACR settings” button, a window appears showing the ACR settings
about the image from which the Pop-up menu has been reached.
.
View and laterality of image can be changed selecting new identification codes and clicking
on “OK” button to accept the new configuration.
5.6 Downsize to 18x24, Delete, Recover 24x30, Measure, Plot Line, Alt Images
To access these functions right click with the mouse on the desired image.
The “Recover 24x30” function is available only for the mammography unit with 24x30 cm
detector, when an images is incorrectly downsized to 18x24 cm format. It allows to restore
the original 24x30 cm format of acquired image.
Clicking on “Recover 24x30” button, the image is restored at original acquisition format.
The “Delete” function allows to remove the selected image from the study transfer it in a local
storage.
Clicking on “Delete” button a 18x24 cm a window with 16 reason codes appears.
The operator must select the reason of deletion and clicking on “OK” the image will be erased
from the study.
Clicking on desired kind of measure, a little hand symbol appears on the image.
Click on one edge of the detail and drag the hand symbol till reaching the other edge of the
object to measure. Releasing the button, a window with the measurements results appears.
The “Plot line” and “Alt images” functions allow to make preliminary investigations on image
quality.
Clicking on “Plot line” button, a little hand symbol appears on the image.
Click on one edge and drag the hand symbol till draw a line. Releasing the button, a window
with a profile of the gray levels on the drawn line appears.
Clicking on “Alt images” button, the unprocessed image (RAW format before the application
of enhancement algorithm) is displayed.
6. CLOSING STUDY
For any study it is possible to acquire a maximum of ten images. If the maximum number of
images is reached, a pop-up message appears to inform the operator.
When the images acquisition is complete the operator must close and save the study.
If the study has been open locally (by means of the “Study Data” window), it is saved locally.
The study can be saved also on DICOM server in a second time by “Resend” function.
If the study has been open from the Worklist (by means of the “Worklist Query Dialog”
window), it is saved on DICOM server. A copy of the study is also saved locally for a time
configurable in installation phase.
At the end of time of local storage, all studies are erased from
NOTE
the Hard Disk.
At the end of exam, clicking on “Close Study” button of “DICOM” menu a window appears to
allow to send and save study with data and images into store server.
However, if the connection is not available or the net is busy, data are stored locally and sent
to the server when connection is newly available.
A study open locally is saved on the internal memory of local PC clicking on the “Save Study”
button of “PC” menu.
It’s possible to load the studies saved into the internal memory of local PC only for reading.
Clicking on the “Load Study” button of the “PC” menu a window appears with the list of
studies saved locally in the local PC.
Studies that were open from the Worklist (by means of the “Worklist Query Dialog“ window)
are displayed on light green background; the studies opened locally (by means of the “Study
Data” window) are displayed on orange background.
On those Studies, it is possible to do a search setting one or more parameters between the
following:
Patient Name;
Patient ID;
Birthdate;
Study Date
To exit from the study, simply make the next operation (load or open another study, etc.).
8. DICOM FUNCTIONS
By “DICOM” menu, various functions (Resend, Append, Print, Make CD and export) can be
applied on the studies open from the Worklist or locally.
8.1 Resend
The “Resend” function allows to send a local stored study to the DICOM server.
Clicking on “Resend” button, a window appears with the list of studies saved locally in the
Acquisition Software.
Studies open from the Worklist are displayed on light green background; the studies open
locally are displayed on orange background.
Select the study that has to be resend to the PACS checking the corresponding box on the
left and then click on “OK”.
Selecting a study open locally, the “Worklist Query Dialog” window appears; the operator can
associate the study at a patient and send the images to the server.
Selecting a study open from the Worklist, it is sent directly to the server.
8.2 Append
The “Resend” function allows to recall a study earlier open from Worklist in order to integrate
it with other images. In any case, the information associated to the study are not modifiable.
The studies open locally (by means of the “Study Data” window) are displayed on grey
background and they are not selectable.
The studies open from the Worklist are displayed on light green background. The operator
must double click on desired study to open it.
The study images are shown on display and the mammography system is already to take X-
rays exposures.
Once the study has been integrated, close it according to the foreseen modalities. It is always
possible to iterate the “Append” function until the study remains locally stored on the local
PC.
8.3 Print
The “PRINT” function allows to send to a DICOM printer the images of an open study.
The thumbnail of the images, acquired in the current study, are displayed on its left side while
the printer parameters are displayed on the upper side. The parameters adjustable by the
operator are the following:
Printer model
Film type and size
Direction (landscape or portrait)
Resolution (DPI)
The operator can, for example, increase the number of rows and columns in which the area
will be separated:
In some particular case, the system could go out of resource (e.g. if the operator increase
both the film size and the DPI to an high value). In this case, the following pop-up message
appears:
Choosing an image from the customized print layout and clicking on the corresponding icon,
the following window will be displayed:
in which it will be possible to modify the image view features by means of the commands bar
at the bottom. Clicking on “OK” it will be possible to exit and save.
The images can be printed integrating them with many exposure parameters.
Clicking on “Set tags” button on the bottom-left, a window appears with a list of printable
parameters:
In particular, it is possible to choose, between the available tags (left column “Available”),
those ones which have to be included moving them to the right column “Print” using the
pushbuttons “◄” or “►”. Once selected, their order of presentation can be modified using the
pushbuttons “▲” or “▼” and their dimensions clicking on “+” or “-“.
It is also possible to insert a ruler of 10 cm on the image to print:
Finally, clicking on “PRINT” button, all the images in the customized layout are sent to the
printer.
8.4 Make CD
Insert a CD/DVD or a removable drive of adequate capacity (approximately 100 Mb for each
patient’s study).
The system automatically shows one of the following messages after finding the drive
inserted:
8.5 Export
The “Export” function allows to export the DICOM studies on CD/DVD or removable drives.
It drives in native format for PACS Store and is mainly used for patient studies transfer from
Mobile VAN to PACS at the end of working day.
Clicking on “EXPORT” button, a window appears with the list of studies saved locally in the
internal memory of local PC.
Select the study to send to the media checking the corresponding box on the left and then
click on “OK”.
The following message appears:
To select all studies insert tick into virtual box “select all”, to
NOTE
unselect all studies push the tick into “unselect all”
To allow the export of “for processing” images, click on DICOM label (CTRL right + mouse
double click) and open the DICOM server list:
Select the CDBOX Server associated to the EXPORT FUNCTION, enable the flag “send “for
processing” images and click on OK to save:
In this way, the study/ studies chosen will be exported with the “for processing” and “for
presentation” images.
Follow the same procedure detailed in the previous paragraph 8.5.
At the end of EXPORT, the previous flag must be disabled.
9. DAILY CHECK
Daily, prior to start to perform the exams, the operator must verify the right operating of
mammography unit, through a specific tool achievable from “Operation Bar”.
To perform the check procedure is necessary to use an optional phantom (see Section 8:
accessories) to place on Potter-Bucky.
This tool has been developed according with Euref Protocol point 2b.2.2.3.1“Image
Receptor Homogeneity”.
Detector health percentage is related to the number of ROI passing the test , if the value is
lower than 100% call Service for periodic Detector Calibration.
This tool has been developed for daily check and cannot
NOTE be used in replacement of calibration tools dedicated to
quality control.
In case of occurrence of SW problems, the Technical Assistance may ask the operator to
send some files which may be useful for their analysis and troubleshooting.
Using the available function “Export Logs”, the operator can export the requested files on
a removable drive.
Clicking on “OK”, the process starts and may take several minutes.
Only when recording succeed, the following message appears and the media can be
removed.
11. CALIBRATE
Depending of the specific detector mounted on the Mammo Unit, the SW may automatically
enables the “Calibrate” push button of the “Menu Panel” as shown in the following figure:
In this case, daily and before starting with the exams, the operator must perform the
following check through a specific tool achievable on the SW interface.
To perform this procedure is necessary to use the phantom for periodical system calibration
without compression (see Section 8: accessories) in place of Potter-Bucky and protective
screen. Moreover, it is necessary select the anode filter/combination (W/Rh or W/Ag) and
perform the calibrate procedure only with large Focus
8. At the end of X-ray exposure, the following interface is displayed and the test
automatically runs showing the “Wait” message:
9. After few seconds, the message “Done” appears in place of “Wait” and the window
automatically disappears at the end of the test:
SECTION 8:
MAINTENANCE
GENERAL CONCEPTS
The expected service life of mammography unit is defined and based on Service and
Maintenance Instructions contained in the Technical Manual and on applicable official QC
protocols for image quality.
Maintenance procedures and checklist are included in Technical Manual and can be carried
out only by qualified personnel.
Image quality inspection will be established according to locally applicable standards and QA
protocols but periodicity must be less than six months.
User should be daily (or before every switching ON of Mammo unit) check:
External damages to shield or unit
Check of mains and earth connections, cables and corrugated tube
The Potter-Bucky (24x30 cm format) is provided with an anti-scatter vibrating grid for limiting
the amount of radiation scatter created in a radiographic exposure reaching the detector.
The operator can verify the functionality of grid moving it, without exposure, pressing the
pushbuttons of ACR panel until the grid starts moving for 4 seconds.
MAMMOGRAPHY UNIT
Regular cleaning of the compression paddle and Potter-Bucky (or magnification device) is
necessary prior to proceed with examination.
Use only a soft cloth and distilled water to avoid damage polycarbonate plate of compression
paddle getting it hard with risks of cracks. For other surfaces use cleaning agents that do not
damage plastics, aluminum and carbon fiber.
Wipe any parts that contact the patient using a clean lint-free or pad.
To avoid electrostatic dust collection over the plastic covers, use anti-static dashboard wet-
wipes.
LCD SCREEN
To clean the LCD screen use a microfiber cloth to remove any obvious dust and gently brush
off any dirt particulars.
Although use a 50/50 solution of distilled water/isopropyl alcohol, using a clean soft, lint-free
cloth. Very, very gently wipe the screen in a circular motion. Do not press hard on the cloth.
Let the cleaning solution do its job.
It is recommended acquire a proper cleaning kit (with an alcohol and ammonia-free solution)
to clean and keep LCD screen in a crisp clear pristine condition.
KEYBOARD
If wireless keyboard doesn’t work check ON/OFF switch at rear side GREEN=ON RED=OFF
and internal batteries subject to periodic replacement.
Use only two AA alkaline long life batteries taking care of polarity.
Keyboard Switch ON
To clean the keyboard first hold it upside down to release any debris from the spaces in-
between the keys (pressing the keys is a good way to release it). Use a can of compressed
air to blow any debris from around and under the keys, then use the hose of a vacuum
cleaner to remove it. Finally, clean the keys surface using a wipe made humid with a neutral
detergent.
To clean the transparent anti-X protective barrier use any specific glass cleaner and a clean
soft, lint-free cloth. Then wipe carefully the surface with another dry cloth.
Dust affects reliability because a thick layer of dust can exhibit a certain amount of electrical
capacitance and conductivity. Just as seriously, it will reduce the working life of fans and
increase the risk of overheating.
The Foam density has designed to be high enough to catch dust, but low enough not to
reduce airflow.
In particularly dusty places, every month, it’s advisable to verify filters cleaning to avoid a dust
accretion. If filters are dirty must be replaced quickly with new others.
In any case, the filters must be changed periodically with a frequency depending by the
installation environment.
The dirty filters can be washed (only with water) and reused totally dry in the next
replacement. The filters replacement procedure is described on Technical Manual and must
be carried out only by Qualified Personnel.
If periodic inspection is not carried out and filters gaps are obstructed the cooling fans can
periodically oscillate with a rate of about 2/3 s generating a typical audible noise that can
recall the operator attention.
Filters obstruction cause a Detector overheating. In this case the temperature can go out of
calibration range and a red message is visible on Status Bar of Acquisition Software (see
also the Section 5-Paragraph 2.1 of this Manual).
LIST OF PARTS
Description
Standard parts
LIST OF OPTIONS
Description
Options
VARIAN M113T biangular tube with tungsten anode (low speed only)
IAE XM12THigh speed voltage generator with wide range [20-40] Kv
Motorized rotation
High speed starter
115Vac Power Supply
Digital Flat Panel Detector (a-Se)
Description
Optional parts
Acquisition Workstation (Transparent Anti-X Protective Barrier with support for monitor + Keyboard
supporting surface), see configurations available
24x30 cm device for geometric magnification with variable (1,5x and 2x) factor
9x21 cm format compression paddle
18x24 cm format spot compression paddle
18x24 cm format shifted compression paddle for 2D biopsy
Bearing for compression paddle (up to n°3 paddles)
Phantom for daily check with compression
SBD 3D FFDM Stereotactic biopsy device (only for MELODY III D C 24x30 cm format)
Kit for digital mammographic unit (with not isocentric C-Arm) installation on VAN
Kit for digital mammographic unit (with isocentric C-Arm) installation on VAN
Transparent Anti-X Protective Barrier
10 kVA Uninterruptible Power Supply (UPS) system
Patient chair (height adjustable)
Patient bed/chair
CONFIGURATIONS AVAILABLE
Control monitor and wireless keyboard are provided with the unit
NOTE
and can be placed as shown in typical configurations below:
BASE
Control monitor and wireless keyboard
are placed on the unit stand.
AA
Control monitor and wireless keyboard
are available on a stand-alone half-
transparent anti-X barrier for operator
protection.
This is equipped with:
2 Mpx diagnostic color monitor
(21,3’’)
wireless keyboard
X-ray pushbutton
AB
Double management configuration.
Mammo unit is equipped on the stand
with:
2 Mpx diagnostic color monitor
(21,3’’)
Wireless keyboard
X-ray pushbutton
AC
Control monitor, a wireless keyboard
and an X-ray pushbutton are provided
with a dedicated support to be
mounted on wall as shown in typical
configuration
AD
Double management configuration.
Mammo unit is equipped on the stand
with:
2 Mpx diagnostic color monitor
(21,3’’)
Wireless keyboard
X-ray pushbutton
Control monitor, a wireless keyboard
and an X-ray pushbutton are provided
with a dedicated support to be
mounted on wall as shown in typical
configuration workstation for
AE
Control monitor and an X-ray
pushbutton are provided with specific
support for desktop placement. A
wireless keyboard is provided
COMPRESSION PADDLES
FOOT-CONTROLS
PROTECTIVE SCREEN
24x30 cm format
SBD FFDM
Half-Transparent Full-Transparent
ACQUISITION WORKSTATION
MODM10C-U02 Page 181 of 226
OPERATOR’S MANUAL
SECTION 10:
SPECIAL HINTS – IMAGE
RECOVERY STRATEGY –
TROUBLESHOOTING
The Acquisition Software provides the customer with different strategies for images recovery
which for various reasons were acquired improperly, avoiding to expose again the patient.
PATIENT POSITIONING
Symmetry
Symmetry
Symmetry
Symmetry
Symmetry
Symmetry
Symmetry
Skinfolds
Skinfolds
Skinfolds
Compression
In case of incorrect size of images acquired with 18x24 compression paddle, it is possible
downsizing to 18x24 cm format with the right click on image. Clicking on “Downsize to 18x24”
button a 18x24 cm rectangle appears: moving it with the mouse select the region of image
under study in order to crop it.
I In case of incorrect size of images acquired with 24x30 compression paddle, it is possible
recovering 24x30 cm format with the right click on image. The “Recover 24x30” function is
available only for the mammography unit with 24x30 cm detector, when an images is
incorrectly downsized to 18x24 cm format.
It allows to restore the original 24x30 cm format of acquired image. Clicking on “Recover
24x30” button, the image is restored at original acquisition format.
MODM10C-U02 Page 193 of 226
OPERATOR’S MANUAL
Images with metallic clips, surgical marker and cluster of micro calcification
In case of image with metallic clips, surgical marker and numerous cluster of micro
calcification, in which image contrast could be not adequate for radiologist physician, you
have to right click on image and select Process → OTSU → Marker, as shown in following
image.
To recover the original image contrast, you have to right click on image and select Process →
Marker
In case of image with surgical anatomical parts, in which image contrast could be not
adequate for radiologist physician, you have to right click on image and select Process →
OTSU → Surgery, as shown in following image
To recover the original image contrast, you have to right click on image and select Process
→ Surgery
In case of image with breast specimen (i.e. vacuum assisted biopsy, trucut biopsy) , in which
image contrast could be not adequate for radiologist physician, you have to right click on
image and select Process → OTSU → Specimen, as shown in following image
To recover the original image contrast, you have to right click on image and select Process→
Specimen
In case of wrong ACR settings, click right on image and select Change ACR settings and
modify laterality, as shown in following image
In the following we are considering the use of the so called spot shifted compression paddle,
shown in the following figure, designed to be used in contact mode (not for magnification).
In its current conformation, this spot compression device is made by an almost round
compression plate in polycarbonate, supported by a metal structure. This structure is heavily
radio-opaque and can be considered an extraneous part clearly visible in a wide standard
image of which only a little ROI is usable.
Furthermore, an important number of gray levels is sacrificed in favor of the metal part.
See for example following figures. On the left, the image of a TOR MAM phantom is shown
with its resulting histogram. On the right, you can see the same phantom imaged with the
spot compression device. The large part of metal in the image field causes a loss of dynamic,
noticeable as a compressed histogram of the part of the image corresponding to the
phantom, highlighted in red. The blue indicates the part of dynamic stolen by the metal from
the relevant object.
Operative indications
The image presentation algorithm has been revised to face this particular situation and to
offer the best result in all the possible configurations. Nevertheless, since this feature is not
much used anymore in our clinical context, it has been tested on phantoms only.
Whenever a loss of contrast in the spot region is perceived, the operator can enhance the
image further by means of tools like Window/Level or Histo in the MELODY III D C software.
An operative example is illustrated below.
Once an exposure is executed and the image is shown in the image window, right click on it:
a menu as shown in following image will appear.
From here, select Histo: a new window will appear, where the operator can lower the
maximum displayed value (the “white point”) to increase contrast in the region of interest, as
shown in following image.
New White
Click on Preview and evaluate the result. Adjust further, if needed. Please note that also the
lowest displayed level can be adjusted. Once the result is satisfying, click on Apply.
We must also advise you that the exposure modality “Automatic – Pre” cannot be used when
imaging with the spot compression device inserted, since this AEC algorithm, due to the
presence of metal parts in a relevant area of the image, will erroneously recognize a high
breast density delivering too much dose to the patient.
When using the spot compression device, please select Manual or Automatic – Fast
mode.
The Manufacturer will remain on hand in case the image quality needs further improvements.
Clinical cases with comments, suggestions and eventual requests are highly welcome.
Fine tunings and/or modifications are possible on the presentation algorithm and we do not
exclude to design a new spot compression device with less invasive arm.
MODM10C-U02 Page 200 of 226
OPERATOR’S MANUAL
SECTION 11:
TECHNICAL SPECIFICATIONS
TECHNICAL SPECIFICATIONS
POWER SUPPLY
Line voltage 220/230/240 Vac 10% 50/60 Hz
115 Vac 10% 50/60 Hz (optional)
Power 6.6 kVA (<0.2 kVA stand-by)
Current absorption 30 A peak (at 230 Vac)
Number of phases 1 or 2 configurable
Connection Permanently installed (IEC 60601-1)
Wall connection 20 A Thermal-magnetic circuit breaker
(40 A Thermal-magnetic circuit breaker in 115 Vac
option)
ENVIRONMENTAL CONDITIONS
Mammography Unit Temperature: -20°C÷+70°C
Transit/Delivery and Storage Conditions Relative humidity: 10%÷90%
Barometric pressure: 500 hPa÷1060 hPa
Detector Conditions during: Temperature for a-Si Detector:
Transfer to Transport Terminal -15°C÷+65°C
Destination Terminal Custom Temperature for a-Se Detector:
Clearance Period +5°C÷+40°C
Door Delivery
Note: Detector packed in the original crate Relative humidity for a-Si detector:
10%÷85%
Relative humidity for a-Se detector:
10%÷90%
Barometric pressure: 700÷1060 hPa
Detector Conditions during:
International Air Transport a-Si Detector
Note: Detector packed in the original crate Temperature: -18°C÷+70°C (under 36h)
Relative humidity: 10%÷85%
Barometric pressure: 700 hPa÷1060 hPa
a-Se Detector
Temperature: -20°C÷+60°C (24h)
Relative humidity: 10%÷90%
Barometric pressure: 700 hPa÷1060 hPa
Operating Conditions Temperature for a-Si Detector:
Mammo unit switched on +5°C÷+40°C
Relative humidity: 30%÷ 85%
MAMMOGRAPHY UNIT
FILTER PROPERTIES
50 m Silver 0.55 mm Al eq. @ 28 kV,
measured with W target
50 m Rhodium 0.51 mm Al eq @ 28 kV,
measured with W target
FILTER PROPERTIES
50 m Silver 0.55 mm Al eq. @ 28 kV,
measured with W target
50 m Rhodium 0.51 mm Al eq @ 28 kV,
measured with W target
IMAGE QUALITY
Spatial resolution Conformity with:
“European Guidelines for quality assurance in
mammography screening”, third edition,
and with
”Recommended specifications” for Quality
assurance in mammography of American College of
Radiology
DOSE CALCULATOR
Method of Calculation Average Glandular Dose (AGD) according to: “D.R.
Dance et al.”
Data visualization (mGy) On display
Method of recording Image Header (DICOM)
AGD with a 4 cm PMMA phantom 1,216 mGy
Dose Rate (28 kV-80 mAs) 36,63 R/min without Compression Paddle
29,80 R/min with Compression Paddle
C-ARM
F.F.D. (Focus Detector Distance)/S.I.D. 66 cm
(Source to image receptor Distance)
Motorized Movements Vertical and rotation (optional)
Range of Vertical movement (from floor) From 43 to 128 cm (travel of 85 cm)
Range of C-arm rotation +/-180°
(CW, CCW continuous to any position)
Projection preset position N° 5 Programmable projections
(LAT, OBL, CC, OBL, LAT)
Speed of C-arm Rotation 90°/7,2 s with acceleration and deceleration ramp
for smooth operation
Display of angle rotation On auxiliary Display (ACR)
On Display of controllers for manual compression
ISOCENTRIC C-ARM
F.F.D. (Focus Detector Distance)/S.I.D. 66 cm
(Source to image receptor Distance)
Motorized Movements Vertical and rotation
+/- 15° Rotation (only with BYM 3D FFDM)
Range of Vertical movement (from floor) From 75 to 160 cm (travel of 85 cm)
Range of C-arm rotation +/-130°
(CW, CCW continuous to any position)
MAG. X 2 = 97,5 mm
Compression Force Adjustable from 70 to 200 N
Compression Force Display Effective applied force with 1 N resolution
Compression Paddle Descent Speed Proportionally decreasing compressing the breast
Maximum Compression Force Safety Device Triple safety device:
electronic, electro-mechanical, mechanical
Soft Compression paddle release after Selectable from Acquisition Software
exposure
Compression Thickness Display Displayed in mm
Compression paddle aluminum equivalence Less than 0.2 mm Al (0.135 mm Al≈30 kV)
Special function for 0 compression force for oncology procedures
FOOT-CONTROLS
Number and type One with two pedals and one push-button for
motor driven compression unlock
Two with two pedals and one push-button for
motor driven compression unlock (optional)
Control Actions Vertical movement of Compression Paddle
Motor driven compression unlock
EMERGENCY STOPS
Number and Type Two red push-buttons on both sides of
mammography unit
PERSONAL COMPUTER
Operating Systems Windows 7 Professional 64-bit
CPU Intel Core i7 2600 3,4 GHz
RAM 8 GB (optionally 16 GB available)
HDD 1 TB SATA for Operating System, Software, studies
storage (~25.000 images)
DVD recorder 48x SATA DVD +/-RW DL
Power Pack 400 W
UPS (uninterruptible Power Supply) 650 VA
AirFlow 178 m³/h
Medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the accompanying
documents.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Information: Fixed equipment or system cabling, which cannot be removed by the user, is
not listed. This cabling is part of the system and was present at all EMC-measurements.
Without this cabling there is no complete functionality of the system.
2 kV common mode
Voltage dips, short 0% Un for 0.5 cycle( at IEC 60601-1-2 Mains power quality should be that of a
interruptions and voltage 0°,45°,90°,135°,180°,225°,270°, test level typical commercial or hospital environment.
variations on power supply and 315°) If the user of the MELODY III D C requires
input lines 0% Un for 1 cycle continued operation during power mains
IEC 61000-4-11 interruptions, it is recommended that the
70% Un for 25 cycle MELODY III D C be powered from an
0% Un for 5 s uninterruptible power supply or battery.
Power frequency 30 A/m IEC 60601-1-2 Power frequency magnetic fields should be
(50/60 Hz) test level at levels characteristic of a typical location
magnetic field in a typical commercial or hospital
environment.
IEC 61000-4-8
Conducted RF 3V 3V d 1 .2 P
IEC 61000-4-6 150 kHz to 80 MHz*
Where P is the maximum output rating of the transmitter
in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in meters
(m).
* 6 V RMS in the ISM band (6,765 to 6,795, 13.553 to 13.567, 26.957 to 27.283 and 40.66 to 40.70 MHz)
that is present on the device, means that within the European Union the product must be
taken to separate collection at the product end-of life. Therefore, at the end of the life-cycle of
the device, the user should deliver the device to the proper collection facilities of the Electric
and Electronic Equipments. Alternatively, the user can return the device to the seller, on a
one-to-one basis, as long as he is buying a new one of equivalent type and that fulfills the
same functions as the old one.
Disposing of the device separately avoids possible negative consequences for the
environment and health deriving from inappropriate disposal and enables the constituent
materials to be recovered to obtain significant savings in energy and resources.
Who disposes any Electric and Electronic Equipment, reporting the above symbol, as
unsorted municipal waste, instead of collecting it separately, incurs the administrative
sanctions in accordance with law.
SECTION 12:
WARRANTY CONDITIONS
WARRANTY CONDITIONS
The Manufacturer is committed to replacing, free of charge, any part of the machine that
proves to be faulty, for a period of twelve months from the date of installation.
THIS WARRANTY DOES NOT PROVIDE FOR DEFECTS DUE TO:
damage caused during transportation
damage for storage/operation outside temperature and humidity limits
damage caused by the incorrect installation of the unit, if not carried out by personnel
trained by the Manufacturer
damage caused by inappropriate connection to other units
damage caused by improper use, negligence, carelessness or inability to use unit.
use of not original spare parts or accessories
THE WARRANTY DOES NOT APPLY TO MATERIAL SUBJECT TO WEAR AND TEAR .
X-Ray tube is covered by pro rata temporis warranty
The aforesaid terms are to be considered valid except as otherwise set forth in the contract.
To:
FAILURE REPORT
IMPORTANT NOTICE:
This report shall be submitted whenever a failure occurs. Information given through this form
will expedite manufacturer’s technical assistance.
In case of warranty claim this report is mandatory: warranty will not be processed without
receiving this form properly filled in.
DISTRIBUTOR: NAME
ADDRESS
USER: NAME
ADDRESS
SERIAL NUMBER
DATE OF PURCHASE
DATE OF INSTALLATION
SOFTWARE RELEASE
NUMBER OF EXPOSURES
FAILURE: DESCRIPTION
MECHANICAL
ELECTRICAL
ELECTRONIC
To:
SAFETY FEEDBACK
In order to continuously improve quality and safety of our medical products, we kindly ask you
to quickly signal us, filling in this form, any potential cause of risk bound to our medical
devices.
DISTRIBUTOR: NAME
ADDRESS
USER: NAME
ADDRESS
EQUIPMENT: TYPE
SERIAL NUMBER
PURCHASE DATE
INSTALLATION DATE
SOFTWARE REVISION
NUMBER OF EXPOSURES