Sunteți pe pagina 1din 10

PRACTICE REPORTS  Barcode technology

PRACTICE REPORTS

Effect of barcode technology


with electronic medication administration
record on medication accuracy rates
Heather H. Seibert, Ray R. Maddox, Elizabeth A. Flynn, and Carolyn K. Williams

M
edication errors and con-
sequent deaths continue to Purpose. The effect of barcode-assisted wrong-time errors are excluded, the accu-
escalate, costing health care medication administration (BCMA) with racy rate improved from 92% in phase 1 to
electronic medication administration 96% in phase 3 (p = 0.000008). The overall
systems billions of dollars each year.1
record (eMAR) technology on the occur- accuracy rate did not change significantly
It is estimated that at least 1 medica- rence of medication administration errors from phase 1 to phase 3 at hospital 2; when
tion error occurs per day per hospi- was evaluated. wrong-time errors were excluded from
talized patient. An estimated 450,000 Methods. A pretest–posttest nonequiva- consideration, the accuracy rate improved
adverse drug events—medication lent comparison group was used to in- from 93% in phase 1 to 96% in phase 3
errors that result in patient harm— vestigate the effect of BCMA-eMAR on (p = 0.015).
occur annually, approximately 25% the medication administration accuracy Conclusion. Implementation of BCMA-
rates at two community-based hospitals. eMAR in two hospitals was associated with
of which are preventable.2
Patient care units included three matched significant increases in total medication
Various technologies have been pairs in the two hospitals—two medical– accuracy rates in most study units and did
introduced to help improve the ac- surgical, two telemetry, and two rehabilita- not introduce new types of error into the
curacy of medication administration, tion units—plus a medical–surgical inten- medication administration process. Accu-
including automated dispensing cab- sive care unit, an emergency department, racy rates further improved when wrong-
inets, computerized prescriber order and both an inpatient oncology unit and time errors were excluded from analysis.
entry (CPOE), “smart” (computer- an outpatient oncology service at one of The frequency of errors preventable by
the hospitals. Medication administration BCMA-eMAR decreased significantly in
ized) i.v. infusion pumps, barcode-
accuracy rates were observed and recorded both hospitals after implementation of that
assisted medication administration before (phase 1) and approximately 6 and technology. BCMA-eMAR and direct obser-
(BCMA) systems, electronic medica- 12 months after (phases 2 and 3, respec- vation were more effective than voluntary
tion administration records (eMARs), tively) the implementation of BCMA-eMAR. reporting programs at intercepting and
and wireless connectivity and inte- Results. The overall accuracy rate at hos- recording errors and preventing them from
gration with hospital information pital 1 increased significantly from phase reaching patients.
technology. A recent report on the 1 (89%) to phase 3 (90%) (p = 0.0015); if Am J Health-Syst Pharm. 2014; 71:209-18
adoption rates of medication-safety
technologies revealed that health
care organizations have implemented commonly implemented technol- Medication errors can be identi-
smart pumps more than other tech- ogy, followed by CPOE with deci- fied and quantified through at least
nologies.3 BCMA is the second most sion support. four different processes. The most

Heather H. Seibert, Pharm.D., M.B.A., is Manager and Clinical Address correspondence to Dr. Seibert (hugenerh@sjchs.org).
Pharmacy Specialist, Centers for Medication Management; and The authors have declared no potential conflicts of interest.
Ray R. Maddox, Pharm.D., FASHP, is Director, Clinical Pharmacy,
Research and Pulmonary Medicine, St. Joseph’s/Candler Health Copyright © 2014, American Society of Health-System Pharma-
System, Savannah, GA. Elizabeth A. Flynn, Ph.D., is Independent cists, Inc. All rights reserved. 1079-2082/14/0201-0209$06.00.
Research Consultant, Artesia, NM. Carolyn Williams, B.S.Pharm., DOI 10.2146/ajhp130332
is Medication Safety Specialist, Clinical Pharmacy, St. Joseph’s/
Candler Health System.

Am J Health-Syst Pharm—Vol 71 Feb 1, 2014 209


PRACTICE REPORTS  Barcode technology

common way of identifying a medi- Errors recognized the potential value with wireless scanners in the criti-
cation error is through the volun- of bedside BCMA verification but cal care areas and negative-pressure
tary reporting of events recognized noted that data from observational isolation rooms. The change in
during the medication-use process studies of medication administration hardware resulted in significant im-
by clinicians caring for the patient. are needed.1 There is also a need for provements in staff acceptance, the
However, this method vastly under- evidence comparing the accuracy of scanning rate, and system use.
reports the number, frequency, and such electronic systems to the recog- BCMA-eMAR implementation.
outcomes of medication errors. 4 nized gold-standard error-detection In 2006, the medication safety team
Other commonly used methods in- method (direct observation).4 evaluated the state of BCMA tech-
volve computerized monitoring and This study evaluated the effects of nology at both hospitals, completed
manual chart review, though com- a BCMA-eMAR system on the rate the American Hospital Association
puterized monitoring may identify of medication administration errors (AHA) and Institute for Safe Medica-
fewer errors than chart surveillance.5 in both regular and special care units tion Practices (ISMP) Readiness Plan
A more-thorough method of iden- in two community, nonteaching hos- for BCMA,16 and identified infra-
tifying medication errors is direct pitals and analyzed the differences in structure improvements that would
observation of medication adminis- event rates from voluntary reporting, be needed before implementation
tration by caregivers in patient care interventions caught by BCMA tech- could proceed. Improvements iden-
areas. Direct observation is resource nology, and interventions identified tified as a result of the AHA/ISMP
intensive and seldom used but is through direct observation. assessment included the need for
recognized as the gold standard of all medications to have a machine-
identifying and documenting medi- Methods readable barcode, new patient wrist-
cation errors.4 Lastly, data from vari- Setting. St. Joseph’s/Candler bands that contain machine-readable
ous medication-safety technologies Health System comprises two tertiary barcodes, significant changes in the
(smart pumps, BCMA, CPOE) that care, community hospitals totaling computer database, and equipment
intervene and prevent errors from 644 beds, with an annual patient and training for pharmacy, nursing,
occurring present a picture of the volume of 22,807. The hospital staff and respiratory staff. Completion of
“potential” medication error rate in includes 455 community-based, pri- the project was divided over three
health systems where these technolo- vate practice physicians, 1,245 nurs- years, allowing the health system to
gies are deployed.6-8 es, and 53 pharmacists. Although spread the cost of equipment over
Studies have shown that the per- most patients in these hospitals are several budget cycles.
centage of medication misadventures adults, one hospital is a high-volume Implementation of BCMA-eMAR
attributable to errors in drug admin- provider of obstetric services. began in fall 2007. Before the
istration ranges from 2.4% to 11.1% Technology used for BCMA- BCMA-eMAR system went live, the
but may be as high as 34–49%, ac- eMAR. At the time of BCMA-eMAR decision was made to conduct a
cording to some international evalu- implementation, the health system research study on implementation
ations.9 Few of these errors are inter- used health information technol- to assess whether medication errors
cepted before reaching the patient. ogy that was integrated across both were prevented or new errors were
For this reason, the use of BCMA to hospitals (Meditech Magic 5.61, introduced through the implementa-
help improve the accuracy of medi- Westwood, MA). During the course tion of the BCMA-eMAR system.
cation administration at the point of of the BCMA-eMAR implementa- Data collection. A pretest–posttest
care has seemed particularly promis- tion, an upgrade to a higher version nonequivalent comparison group
ing. However, the implementation of of the software (Magic 5.64) was design was used to investigate the
BCMA has proved challenging, with completed, but no significant chang- effect of BCMA used in conjunction
fewer than half of nonfederal hospi- es to the BCMA-eMAR component with an eMAR on the rate of medica-
tals having adopted this technology.3 were made. tion administration errors. Observa-
Published research substantiating Mobile medication carts with tions of medication administration
the efficacy of BCMA in decreasing thin client computers and tethered errors were made before (phase 1)
the frequency of medication er- scanners were initially provided for and approximately 6 and 12 months
rors is limited.8,10-15 An Institute of each staff member who administered after (phases 2 and 3, respectively)
Medicine (IOM) report cited a lack medications. Within a few months, implementation of the BCMA-eMAR
of solid evidence demonstrating the the tethered scanners were replaced system. Postimplementation data
effect of technology on medication with wireless scanners. Shortly after were collected via direct observation
errors. 1 The IOM Committee on the system went live, the decision after the study unit staff were fully
Identifying and Preventing Medication was made to install inroom terminals trained, the system was operational

210 Am J Health-Syst Pharm—Vol 71 Feb 1, 2014


PRACTICE REPORTS  Barcode technology

for at least 6 months, and study medication administration error rate These error types are defined in the
unit nurses achieved an electronic was used to estimate the sample size appendix. All doses observed during
scanning rate of at least 80%. Post- needed for 90% confidence. the study period were entered into
implementation data were not col- The direct observation method the AU MEDS software program for
lected at the same time for all study of Barker et al.18 was used to collect analysis. Data collected by all observ-
units. Study units were reevaluated data. Direct observation is a scientifi- ers were maintained in a secure mas-
approximately 12 months after cally validated technique for measur- ter database.
BCMA-eMAR implementation. ing medication errors and provides Observers did not count doses as
The study population comprised an accurate description of how many opportunities for error if a drug was
randomly observed nurses who ad- errors occur and insight into how er- left at the patient’s bedside for self-
ministered at least one medication to rors may be prevented. This method administration and the administra-
adult patients on the selected units of collecting errors is nonjudgmental tion was not actually witnessed by
during the observation times. Nurse and nonpunitive. the observer, if a dose was associated
subjects were those working in the Errors were documented and with an uninterpretable written or-
selected patient care units when tabulated in the AU MEDS system der, or if the observer did not witness
observation occurred. In order to (MedAccuracy LLC, Lenexa, KS). the entire process of administration.
maintain nurse confidentiality and This system is a nationally standard- Any deviation between the order and
due to the complexity of trying to ized method for monitoring medica- what was observed was recorded as
collect sufficient doses, no attempt tion error rates based on the direct an error. After examining all doses
was made to match the preimple- observation of nurses preparing and witnessed, the observer tallied all
mentation and postimplementation administering medications.19 Data omitted doses. The medication
observations for individual nurses. were collected by 15 licensed health administration error rate was calcu-
Instead, observations were made by care professionals (7 pharmacists, 8 lated by dividing the number of ob-
patient care unit. nurses) and the research pharma- served errors by the sum of all doses
To facilitate extrapolation of the cist, all of whom were employed by witnessed and all omitted doses.
results to other institutions, data the health system. These individuals Definition of terms. A medication
were collected on patient care units were certified medication observers administration error was any discrep-
that had a diversified adult patient trained by the AU MEDS specialist. ancy between a prescriber’s interpre-
population to whom a wide variety Observers witnessed nurses in the table medication order and what was
of medications was administered. selected units providing standard administered to a patient. Drug ad-
Patient care units were selected patient care; observers did not inter- ministration more than 60 minutes
for study inclusion if their use of a fere with any activities and were not before or after the scheduled time
medication distribution system was involved with patient care in any way. was considered a wrong-time error.
deemed either typical (i.e., similar to On the other hand, if at any time an For routine doses, the administra-
like units in other hospitals, such as error could potentially result in pa- tion schedule listed in each hospital’s
internal medicine and critical care) tient harm, the research pharmacist medication administration record
or included special interest units (i.e., intervened as discreetly as possible to was used to determine whether a
emergency department and outpa- correct the issue. dose was administered within the ac-
tient chemotherapy infusion center). The observer introduced the study ceptable time frame.
Patient care units included three to the nurses, recorded drug prepara- Target errors were errors that
matched pairs in the two hospitals— tion and administration, reviewed should have been prevented by the
two medical–surgical, two telemetry, the original medication orders, and BCMA system—wrong dose, wrong
and two rehabilitation units—plus a compared the orders with what was form, extra dose, unauthorized drug,
medical–surgical intensive care unit administered to determine if any er- and omission.
(ICU), an emergency department, rors occurred. Observers’ notes were An opportunity for error is a mea-
and both an inpatient oncology unit reconciled with the original physi- sure used as the basic unit of data
and outpatient oncology service at cians’ orders to identify any discrep- in observational error studies.10 In
one of the hospitals. ancies between what was written and this study, opportunity for error was
Investigators estimated the sample what was observed. defined as any dose that was ordered
size (number of doses to be observed) Discrepancies were classified by and either administered or omitted.
needed to achieve 90% confidence the type of error that had occurred: Any administered dose was desig-
that the true medication administra- wrong time, wrong route, wrong nated as either correct or incorrect
tion error rate was being measured. technique, omission, wrong form, (error or no error), which meant the
Based on similar studies,17 a 10% extra dose, and unauthorized drug. error rate could not exceed 100%.

Am J Health-Syst Pharm—Vol 71 Feb 1, 2014 211


PRACTICE REPORTS  Barcode technology

The total opportunities for error phases 1 and 3 to determine whether reccurrence. The software provided a
was the combined total of the num- the BCMA-eMAR system was as- number of search and reporting op-
ber of doses given plus the number of sociated with accurate medication tions that permitted the tabulation of
omissions. administration in each patient care recorded data.
Total errors was defined as the unit.
total of all errors that were observed, The a priori level of significance Results
including omission, unauthorized was 0.05. The level of power achieved Medication administration ac-
drug, extra dose, and wrong route, was calculated, with a goal of 0.80; curacy. The total opportunities for
form, technique, dose, or time. small-effect sample size20 was used error and the accuracy rates by pa-
Accuracy rate was the percentage to conservatively assess the effect of tient care unit during phases 1 and
of doses administered correctly and the BCMA-eMAR system. Confi- 3 are listed in Table 1. Electronic
calculated as follows: (total opportu- dence intervals were calculated for scanning percentages exceeded 90%
nities for error – total errors)/(total all means. during phase 3. The occurrence rates
opportunities for error × 100). Data not collected by direct ob- for all error types are shown in Table
Observed doses were those for servation. When a nurse scanned 2. The overall accuracy rate at hos-
which the observer witnessed both a medication before administering pital 1 changed significantly, from
nursing preparation (e.g., drawing it, one of the following warnings 89% in phase 1 to 90% in phase
up medication from vial) and ad- could have appeared on the com- 3 (p = 0.0015). If wrong-time er-
ministration of the medication to the puter screen: medication is not on rors are excluded, the accuracy rate
patient.10 current eMAR, medication is for a improved from 92% in phase 1 to
Electronic scanning was the proc- different patient, abnormal labora- 96% in phase 3 (p = 0.000008). The
ess of using an electronic device to tory test results, or allergy. At this overall accuracy rate did not change
read a medication barcode and inter- point, the nurse can decide not to significantly from phase 1 to phase
pret data using software to validate administer the medication. When the 3 at hospital 2; when wrong-time
medication accuracy against defined nurse was warned before medication errors were excluded from consid-
information such as patient identi- administration and decided not to eration, the accuracy rate improved
fication, medication identification, administer the medication, this was from 93% in phase 1 to 96% in phase
and medication strength while docu- considered an averted event. When 3 (p = 0.015).
menting the actual administration a potential error was intercepted by Target-error analysis. The results
and defined variables such as nurse the system after the nurse scanned of the target-error analysis in study
name, date, and time. the medication or the patient’s phases 1 and 3 were compared (Table
An averted event was defined as a wristband, the BCMA-eMAR system 2). The number of target errors at
near-miss event, meaning that an er- recorded the event and tabulated the both facilities did decrease from
ror occurred but was intercepted, pre- number of times the nurse decided to phase 1 to phase 3; for hospital 1,
venting it from reaching the patient. administer or not administer a dose this analysis was performed with
Data analysis. Observer reports of medication. the emergency department and out-
were reviewed by the research phar- Voluntarily reported data were patient oncology units excluded, as
macist to ensure that error defini- recorded for five time periods: three these were considered special interest
tions had been applied accurately. before BCMA-eMAR implementa- units.
Any questionable error was discussed tion and two after BCMA-eMAR Special care units. The medica-
by the observer and the research implementation. The voluntary tion accuracy rate decreased signifi-
pharmacist or with the AU MEDS reporting process was performed cantly after BCMA implementation
specialist to determine if the error using a Web-based reporting and in the ICU at hospital 2 (accuracy
was actually an error and the issue analysis software system (Quantros rate, 94% and 83% in phases 1 and 3,
was resolved. Data from the emer- Safety Event Manager, version 5.13, respectively; p = 0.004). The analysis
gency department and outpatient Milpitas, CA). This system allowed showed a large increase in technique
oncology unit of hospital 1 were the reporter to provide a narra- errors (e.g., failure to use a filter
not included in the calculation of tive description of the event and straw to remove medication contents
changes in target errors because of included specific fields for record- from an ampul, failure to have the
important differences between these ing the date and time of the event, patient rinse his or her mouth after
and other units in the drug distribu- patient-specific demographics, loca- using an inhaler)—1 in phase 1 and
tion system. tion, type of error or event, cause, 13 in phase 3.
Chi-square analysis with Yates name of medication, severity of error The accuracy rate for hospital 1’s
correction was used to compare or harm, and strategies to prevent outpatient oncology unit remained

212 Am J Health-Syst Pharm—Vol 71 Feb 1, 2014


PRACTICE REPORTS  Barcode technology

Table 1.
Total Opportunities for Error and Accuracy Rate Before (Phase 1) and After (Phase 3) Implementation
of BCMA with Electronic MARa
Accuracy Rate
Total Opportunities Accuracy Excluding Wrong-Time
Unit and Study Phase for Error Rate (%) Errors (%)
Hospital 1
 Medical–surgical
  Phase 1 534 88 93
  Phase 3 101 93 96
 Telemetry
  Phase 1 310 94 94
  Phase 3 161 88b 94
 Rehabilitation
  Phase 1 531 86 90
  Phase 3 744 85 94c
 Emergency
  Phase 1 205 86 87
  Phase 3 237 95d 99e
  Inpatient oncology
  Phase 1 310 89 94
  Phase 3 87 94 98
  Outpatient oncology
  Phase 1 202 97 97
  Phase 3 247 97 98g
   Total
    Phase 1 2092 89 92
    Phase 3 1577 90d 96f
Hospital 2
 Medical–surgical
  Phase 1 569 89 94
  Phase 3 129 92 93
 Telemetry
  Phase 1 465 89 93
  Phase 3 254 89 98g
 Rehabilitation
  Phase 1 692 87 92
  Phase 3 325 94h 97i
  Intensive care
  Phase 1 335 94 96
  Phase 3 65 83c 88j
   Total
    Phase 1 2061 89 93
    Phase 3 773 91 96k
a
Chi-square analysis with Yates correction was conducted for comparisons between phases 1 and 3 in the same study units. BCMA = barcode-assisted medication
administration, MAR = medication administration record.
b
p = 0.04.
c
p = 0.004.
d
p = 0.0015.
e
p = 0.000002.
f
p = 0.000008.
g
p = 0.006.
h
p = 0.0005.
i
p = 0.002.
j
p = 0.003.
k
p = 0.015.

Am J Health-Syst Pharm—Vol 71 Feb 1, 2014 213


PRACTICE REPORTS  Barcode technology

fairly consistent, even when wrong-

Excludes the emergency department and outpatient oncology unit because of differences in their drug distribution systems. The exclusion was for target errors only. With the exclusions, there were 115 total target errors and
time errors were excluded. Because

Wrong
Route

0
9

0
9

0
1
most chemotherapy and support
Types of Errors Observed in Study Hospitals Before (Phase 1) and After (Phase 3) Implementation of Barcode-Assisted Medication medications administered are for a
single dose, wrong-time errors are
Other Errors
not as relevant in this unit as in other

Technique
Wrong
practice settings. The accuracy rate

36
13

13
11

45
13
increased from 97% in phase 1 to

There were 92 total target errors and 1969 no-error doses during phase 1 and 17 total target errors and 756 no-error doses during phase 3 (p = 0.002, chi-square analysis with Yates correction).
98% in phase 3. In this unit, there
were a total of 4 errors in phase 3, in-
cluding 2 technique errors involving
the administration rate of support
Wrong
Time

67
94

66
83

86
36
medications (not chemotherapy), 1
unauthorized-drug error involving a
support medication that was not in-
dicated on the physician’s order, and

1570 no-target error doses during phase 1 and 37 total target errors and 1056 no-error doses during phase 3 (p = 0.0001, chi-square analysis with Yates correction).
Omission

1 wrong dose involving a support


71
22

69
19

54
6
medication.
In the emergency department at
hospital 1, the accuracy rate increased
from 86% to 95% (p = 0.0015) from
Unauthorized

phase 1 to phase 3. When wrong-


time errors were excluded, the ac-
Drug

6
3

6
2

7
2

curacy rate improved—from 87% to


99% (p = 0.000002). The number of
wrong-dose errors decreased from 8
to 0 in phase 3.
Target Errors

Errors averted by BCMA-eMAR


and voluntarily reported errors. As
Extra
Dose
Administration with Electronic Medication Administration Record

4
1

4
1

7
0

demonstrated in the observational


Hospital 2—all unitsb

data, some of the medication er-


rors identified were not detected
or prevented by BCMA-eMAR.
Electronic data collected from the
Wrong
Form

BCMA-eMAR system and volun-


4
0

4
0

4
0
Hospital 1­—all units

tarily reported data were compared


with the data obtained via direct
observation.
Table 3 shows the number of
Wrong

doses of medication administered


Dose

40
16

32
15

20
9

in the study hospitals, the number


of errors reported in the voluntary
reporting system, the number of ob-
served errors during the study for all
observation periods, and the number
of events averted by BCMA-eMAR
Hospital 1—units excludeda

on the observational units. Medica-


tion errors consistently declined after
Study Phase

BCMA-eMAR was implemented,


with the number of doses adminis-
tered showing little change (Figure
  Phase 1
  Phase 3

  Phase 1
  Phase 3

  Phase 1
  Phase 3

1). The number of averted events far


Table 2.

exceeded both voluntarily reported


and directly observed medication er-
b
a

214 Am J Health-Syst Pharm—Vol 71 Feb 1, 2014


PRACTICE REPORTS  Barcode technology

Table 3.
Errors Reported Voluntarily and Detected by Direct Observation Before and After Implementation
of Barcode-Assisted Medication Administration (BCMA)
No. No. Errors No. Errors
Total No. Voluntarily Observed Averted by
Doses Reported No. BCMA Directly in BCMA in Study
Date Range Givena Errorsa Units Studied Study Units Units
Before BCMA
  Dec 2006–Mar 2007 663,785 101 4 451 NA
  Dec 2007–Mar 2008 697,830 81 2 87 NA
  Jan–May 2008 907,441 73 4 152 NA
After BCMA
  Feb–May 2009 724,068 26 8 182 1,121
  Sep 2010–Dec 2010 671,834 33 4 58 1,234
a
For all units (not just study units) of both study hospitals combined. NA = not applicable.

Figure 1. Number of medication errors occurring in hospital 1 (A) and hospital 2 (B) after implementation of barcode-assisted medication
administration with electronic medication administration record. Bars represent doses administered.

A 250000 35
31
29 30
No. Doses Administered

200000
24 25
22

No. Errors
150000 20
18
15 16 16
15
16 16 15
100000 12
16 11 11 12 11 11
14 13
12 10
50000 10 9 9 10
8 7 8 8
7 5

0 0
Dec
Jan 07
Feb
Mar
Dec
Jan
Feb
Mar
Apr
May
Feb
Mar 09
Apr
May
Sep
Oct
Nov
Dec
11 Jan
Feb
Mar
Apr
May
Jun
Jul
Aug

Oct

Dec
Sep

Nov

B 250000
37
40

33 35
No. Doses Administered

200000
30

25
No. Errors

150000
24 19 20
20 17
16
100000 20 14 15 14 14 15
11 12 12 11 12
15
9 9 10
8
50000 9
8 7 5
6 6
3 4 4 3
0 0
Dec
Jan 07
Feb
Mar
Dec
Jan 08
Feb
Mar
Apr
May
Feb
Mar 09
Apr
May
Sep
Oct
Nov
Dec
11 Jan
Feb
Mar
Apr
May
Jun
Jul
Aug

Oct

Dec
Sep

Nov

No. errors
Trend line

Am J Health-Syst Pharm—Vol 71 Feb 1, 2014 215


PRACTICE REPORTS  Barcode technology

rors. The majority of averted events uated BCMA-eMAR in three special Some medication errors may not
were related to abnormal labora- care environments. The number of be prevented by the use of BCMA-
tory results, in which case the nurse medication errors in the emergency eMAR, such as prescribing the
chose to withhold the dose based department, most of which were wrong medication for a condition,
on the laboratory test results. There technique errors, significantly de- failing to appropriately tailor the
was a general decline in voluntarily creased from phase 1 to phase 3 (p = dosage for the patient, prescribing
reported errors over the entire study 0.0015; when excluding wrong-time the wrong route of administration,
period even though the number of errors, p = 0.000002). However, there failing to implement an order, and
administered doses remained fairly were no reductions in accuracy rates failing to change an i.v. fluid or initi-
constant. in the outpatient oncology infusion ate new medications when an order
clinic (hospital 1) and the ICU (hos- is changed. If a medication order
Discussion pital 2). The outpatient oncology is incorrectly entered on a patient’s
The results of this study demon- infusion clinic incorporates multiple profile by a pharmacist, a nurse re-
strated that BCMA-eMAR was as- double checks with nurses and phar- viewing the order must detect the
sociated with significant reductions macists, which likely explains the error or the BCMA-eMAR system
in target errors. In addition, many high accuracy rates before and after will allow the wrong medication to
adverse events associated with giving BCMA-eMAR implementation. The be administered. BCMA-eMAR does
a medication when a patient had a increase in observed errors in the ICU not prevent a medication that should
contraindicating laboratory test re- was the result of multiple medication be administered intravenously from
sult may have been averted by using technique misadventures. Technique being administered orally. I.V. pump
BCMA-eMAR. errors are not typically identified and programming errors will not be
Improvements in medication prevented using BCMA. In this case, prevented by BCMA-eMAR unless
accuracy rates were seen in adult education of the nursing staff is key barcode technology is used to man-
inpatient units. These changes were to ensuring that staff members un- age i.v. pump manipulations.
greater when wrong-time errors were derstand the relationships between The observational data collected
eliminated from the comparisons. All medication dosage forms, clinical revealed a large number of wrong-
of the patient care units studied were effects, and routes of administra- time medication administration
high-volume medication administra- tion. An institution may implement errors despite the liberal two-hour
tion environments. Results from the pop-up reminders on the eMAR to administration window. While these
medical–surgical and rehabilitation remind nurses of proper medication- errors may have less importance in
units from both hospitals were simi- handling techniques. some cases, they may be critical in
lar. However, the accuracy rate, when The use of BCMA-eMAR was certain patients, such as those who
wrong-time errors were excluded, accompanied by a reduction in the have diabetes mellitus and may have
in the telemetry unit of hospital 2 number of voluntarily reported had their meal but whose insulin was
improved significantly, but there was medication errors. BCMA-eMAR delayed or given too early, result-
no change in the telemetry unit of and direct observation are more ef- ing in blood sugar abnormalities.
hospital 1. There was a difference in fective than voluntary reporting pro- BCMA-eMAR made little difference
the number of beds in these units (38 grams for intercepting and recording in improving the timeliness of medi-
beds versus 14 beds, respectively). In errors and preventing them from cation administration in the two
addition, patients in the telemetry reaching patients. Voluntary report- hospitals studied. In fact, wrong-time
unit of hospital 1 were predomi- ing is typically rich in detail and often errors increased in hospital 1 but
nantly medical–cardiology patients, allows a more thorough analysis of decreased in hospital 2 after BCMA-
while patients in the telemetry unit events, possibly resulting in substan- eMAR implementation.
of hospital 2 were a mixture of tive improvement opportunities. Importantly, this study found that
postprocedure interventional pa- Direct observation of medication BCMA-eMAR did not introduce new
tients, including patients who had administration successfully identifies types of errors into the medication
undergone angioplasty or coronary medication errors but requires more administration process. Although
artery bypass graft surgery as well as resources to perform. BCMA-eMAR nurses initially believed that a re-
medical–cardiology patients. Final- systems electronically capture events quirement to scan barcodes on medi-
ly, patients in the telemetry unit in and provide reports that can be cations, patients, and themselves
hospital 1 had a longer mean length used to develop improvement plans. would dramatically slow the medica-
of stay than did those in hospital 2. When used together, these systems tion administration process, this ef-
In addition to observations in provide rich information for process fect was not observed as they became
adult inpatient units, this study eval- improvements. efficient with the use of the system.

216 Am J Health-Syst Pharm—Vol 71 Feb 1, 2014


PRACTICE REPORTS  Barcode technology

The results of this study expand 6. Maddox R, Danello S, Williams GK, a c c u r a c y. c o m / AU % 2 0 M E D S . h t m


Fields M. Intravenous infusion safety (accessed 2013 Oct 21).
and further support the positive initiative: collaboration, evidence-based 20. Cohen J. Statistical power analysis for the
effects and limitations of BCMA best practices and “smart” technol- behavioral sciences, 2nd ed. Hillsdale, NJ:
on reducing medication errors. In ogy help avert high-risk adverse drug Lawrence Erlbaum; 1988:235.
events and improve patient outcomes.
addition, this study included obser- www.ncbi.nlm.nih.gov/books/NBK43752
vations from patient care units for Appendix—Error category definitions
(accessed 2013 Oct 21).
Unauthorized drug: Administration of a dose
which there are no other data in the 7. Helmons PJ, Wargel LN, Daniels CE.
Effect of bar-code-assisted medication of medication that was not ordered for the
literature (emergency department, administration on medication admin- patient; may also be referred to as an unau-
ICU, and oncology) and provided istration errors and accuracy in mul- thorized or wrong-drug error.
comparative data from different tiple patient care areas. Am J Health-Syst Extra dose: Any dose given in excess of the total
Pharm. 2009; 66:1202-10. number of times ordered by the physician,
methodologies of gathering medi- 8. Ammenwerth E, Schnell-Inderst P, Machan such as a dose given on the basis of an ex-
cation errors (voluntarily reported C et al. The effect of electronic prescrib- pired order, after a drug has been discontin-
versus directly observed). ing on medication errors and adverse ued, or after a drug has been put on hold. If
drug events: a systematic review. J Am a physician ordered a drug to be given every
Med Inform Assoc. 2008; 15:585-600.
Conclusion 9. Medication errors: incidence and cost.
morning and the nurse gave it in the evening,
the error is placed in this category.
Implementation of BCMA- In: Aspden P, Wolcott JA, Bootman JL, Omission: Failure to give an ordered dose. If the
Cronen Wett LR, eds. Preventing medica-
eMAR in two hospitals was associat- tion errors: quality chasm series. Wash-
patient refuses the medication, an opportu-
ed with significant increases in total nity for error is not counted if the nurse re-
ington, DC: National Academies Press;
sponsible for administering the dose tried to
medication accuracy rates in most 2007:105-42.
give it. If no attempt was made to administer
10. Patterson ES, Rogers ML, Chapman RJ et
study units and did not introduce al. Compliance with intended use of bar the dose, then an omission error is counted.
new types of error into the medica- code medication administration in acute Doses withheld according to policies calling
tion administration process. Accu- and long-term care: an observational for the withholding of medication doses,
study. Hum Factors. 2006; 48:15-22. such as nothing by mouth before surgery, are
racy rates further improved when 11. Poon EG, Cina JL, Churchill W et al. not counted as errors or opportunities for
wrong-time errors were excluded Medication dispensing errors and po- errors. The observer will detect omissions by
from analysis. The frequency of er- tential adverse drug events before and comparing the medications administered at a
after implementing bar code technol- given time with doses that should have been
rors preventable by BCMA-eMAR ogy in the pharmacy. Ann Int Med. 2006; given at that time based on the prescriber’s
decreased significantly in both hos- 145:426-34. written orders.
pitals after implementation of that 12. Young J, Slebodnik M, Sands L. Bar code Wrong route: Medication administered to a
technology and medication administra- patient using a different route than ordered
technology. BCMA-eMAR and di- tion error. J Patient Saf. 2010; 6:115-20. (e.g., oral administration of a drug ordered
rect observation were more effective 13. Poon EG, Keohane CA, Yoon CS et al. Ef- for intramuscular use). Also included in this
than voluntary reporting programs fect of bar-code technology on the safety category are doses given in the wrong site,
of medication administration. N Engl J
at intercepting and recording errors Med. 2010; 362:1698-707
such as the right eye instead of the left eye.
and preventing them from reaching Wrong form: The administration of a dose in a
14. DeYoung JL, Vanderkooi ME, Barletta
different form than ordered by the physician
patients. JF. Effect of bar-code-assisted medica-
when the physician specified or implied a
tion administration on medication error
specific dose form. Giving a tablet when
References rates in an adult medical intensive care
unit. Am J Health-Syst Pharm. 2009; a suspension was ordered is an example.
1. Aspden P, Wolcott JA, Bootman JL, If one of the following dosage forms is
Cronenwett LR, eds. Preventing medica- 66:1110-5.
15. Hassink JJ, Essenberg MD, Roukema JA, crushed, a wrong-form error is counted:
tion errors: quality chasm series. Wash-
ington, DC: National Academies Press; van den Bemt PM. Effect of bar-code- extended-release products, enteric-coated
2007. assisted medication administration on drugs, sublingual medications, and efferves-
2. Bates DW, Cullen DJ, Laird N et al. medication administration errors. Am cent tablets.
Incidence of adverse drug events and J Health-Syst Pharm. 2013; 70:572-3. Wrong technique: An incorrect or omitted ac-
potential adverse events. Implications for Letter. tion by the nurse during the preparation or
prevention. JAMA. 1995; 274:29-34. 16. American Hospital Association, Health administration of a dose that does not result
3. Pedersen CA, Schneider PJ, Scheckelhoff Research and Educational Trust, Institute in another type of error. For example, if the
DJ. ASHP national survey of pharmacy for Safe Medication Practices. Pathways wrong rate of infusion is used and the patient
practice in hospital settings: monitor- for medication safety, assessing bedside receives the correct dose, a wrong-technique
ing and patient education—2012. Am J bar-code readiness. www.ismp.org/tools/ error has occurred. If a heart rate is not
Health-Syst Pharm. 2013; 70:787-803. pathwaysection3.pdf (accessed 2013 Oct measured prior to drug administration, a
4. Flynn EA, Barker KN, Pepper GA et al. 21).
wrong-technique error has occurred. If the
Comparison of methods for detecting 17. Barker KN, Flynn EA, Pepper GA et al.
Medication errors observed in 36 health heart rate is measured and the rate is too low
medication errors in 36 hospitals and for the dose to be given but the nurse still
skilled-nursing facilities. Am J Health-Syst care facilities. Arch Intern Med. 2002;
162:1897-903. administers the drug, an extra-dose error
Pharm. 2002; 59:436-46.
5. Jha AK, Kuperman GJ, Teich JM et al. 18. Barker KN, Flynn EA, Pepper GA. Obser- has occurred
Identifying adverse drug events: develop- vation method of detecting medication Wrong dose: Any dose that contained the wrong
ment of a computer-based monitor and errors. Am J Health-Syst Pharm. 2002; number of preformed dosage units (such as
comparison with chart review and stimu- 59:2314-6. tablets) or is, in the judgment of the observer,
lated voluntary report. J Am Med Inform 19. MedAccuracy. AU MEDS: how it was ±17% the correct oral dosage. In judging
Assoc. 1998; 5:305-14. developed and how it works. www.med dosage, measuring devices and graduations

Am J Health-Syst Pharm—Vol 71 Feb 1, 2014 217


PRACTICE REPORTS  Barcode technology

are those provided for routine use by the


institution: graduations on the syringe for
injections, on medicine cups for oral liquids,
and drops for the dropper provided. Any
dose of an injectable product that is ±10%
or more of the correct dosage is considered
an error of this type. Wrong-dose errors are
counted for ointments, topical solutions, and
similar medications when the dose is quan-
titatively specified by the physician (e.g., in
inches of ointment).
Wrong time: Administration of a dose more than
60 minutes before or after the scheduled ad-
ministration time, unless there is a valid rea-
son. Valid reasons include situations where
the physician has ordered that the patient
not consume anything by mouth, the patient
is off the floor for a diagnostic test, or in
surgery. As-needed doses should be adminis-
tered only as frequently as ordered—the time
of the previous dose’s administration should
be determined from the medication adminis-
tration record. The first dose given according
to the standard administration schedule is
considered to establish that the schedule and
subsequent doses on the same day can then
be examined for wrong-time errors.

218 Am J Health-Syst Pharm—Vol 71 Feb 1, 2014

S-ar putea să vă placă și