o not use Lucitone FRS on large full cases. Lucitone FRS is indicated for the fabrication of
temporary partial or small to medium size full removable dentures, as well as occlusal splints
and nightguards.
Apply petrolatum to the inside of the Cover the deflasking hole with two Position the “pin half” of the Space
Metal Flask, and place the Flask on pieces of moist paper towel, or with Maintainer in the injection cavity.
the Leveler (side “1” up). wax. POUR #1: Mix Dental Stone, and
embed the cast and wax-up in the
Flask, placing the cast as close as
possible to the back of the Flask.
Remove all excess investment.
Then, fit the top half of the Space
Maintainer securely on the bottom
half.
Apply separator to the investment, POUR #2: Place the Flask on the Apply separator (such as sodium
and place the top half of the Flask on Leveler (side #2 up), mix investment, silicate or another alginate or
the bottom half, ensuring complete, and pour it into the Flask, only petrolatum separator) over pour #2
intimate metal contact and closure up to the occlusal incisal surface once set. (A separator plate may be
of the halves. Secure metal Flask of the teeth. As the investment used as an alternative.)
Brackets to the Flask and tighten. begins to set, wipe away any excess POUR #3: Pour additional investment
investment to keep the occlusal on top of separator, filling the Flask.
incisal surface clear. Remove excess, level, and allow
investment to set completely.
Loosen the bolts and remove the metal Flask Brackets. Place the Flask in Add mechanical retention (diatorics) to
boiling water 4 - 6 minutes. Open the Flask and remove the Space Maintainer. the teeth. Use vent holes if possible.
Discard all wax. Flush and clean thoroughly. Remove only enough tooth material to
Bevel the stone around the mouth of the sprue. Check Flask margin and enhance retention. Excessive removal
cavities, ensuring that both Flask halves make intimate metal contact. of tooth material will weaken the tooth
and may lead to fracture.
Caution: Flask will be hot. Use heat resistant The injection piston should remain Bench cool the Flask assembly for 5
gloves. After heating for 17 minutes, remove engaged for one (1) minute. After minutes before deflasking. Caution
Flask from heat source, and remove power. Injection, immediately remove the should be used when handling the
If heat lamps were used, assemble the Flask assembly from the system. hot parts.
warm Flask halves with brackets and the Immediately disengage the cartridge Remove the investment from the
Injection Insert. Place this Flask assembly assembly from the Flask assembly. If bottom half of the Flask by blowing
in front of the Success Injection System. the cartridge walls remain uncrushed compressed air through the
Remove the Cartridge Sleeve and cartridge after injection, use a knife to bend deflasking hole (see photo).
assembly from the furnace. Position the them down. Expel the used cartridge Remove the investment from the top
cartridge assembly on top of the Flask using the Knock Out Base and Knock half of the Flask.
assembly. Slide the combined assemblies Out Rod and return the Cartridge Divest the denture and cast from the
into the Success Injection System. Make Sleeve to the furnace. The Cartridge bulk of the investment.
sure that the piston head is properly aligned Sleeve must remain in the furnace
with the Cartridge Sleeve. Engage the piston when not in use.
by depressing the rocker switch.