Documente Academic
Documente Profesional
Documente Cultură
COMTRAXX® MK2430
MK2430_D00129_00_M_XXEN/12.2014
Bender GmbH & Co. KG © Bender GmbH & Co.KG
P.O. Box 1161 • 35301 Grünberg • Germany All rights reserved.
Londorfer Straße 65 • 35305 Grünberg • Germany Reprinting only with permission
Tel.: +49 6401 807-0 • Fax: +49 6401 807-259 of the publisher.
E-Mail: info@bender.de • www.bender.de Subject to change!
Photos: Bender archives.
Table of Contents
MK2430_D00129_00_M_XXEN/12.2014 3
Table of contents
6. Troubleshooting ................................................................................... 43
6.1 MK2430 error messages ...................................................................................... 43
6.2 Malfunctions ............................................................................................................ 44
7. Operation ............................................................................................... 45
7.1 Operator control and display elements ........................................................ 45
7.2 Quick reference guide .......................................................................................... 47
7.2.1 Display in fault-free operating condition ...................................................... 47
4 MK2430_D00129_00_M_XXEN/12.2014
Table of contents
MK2430_D00129_00_M_XXEN/12.2014 5
Table of contents
MK2430_D00129_00_M_XXEN/12.2014 6
1. How to get the most out of this manual
Chapter "Operation" on page 45 can also be used as a quick reference guide by medical
personnel.
Before using the devices, please read this operating manual, the supplement entitled
"Important safety instructions for Bender Products" and the instruction leaflets supplied
with the individual system components. Keep this document in an easily accessible
location near to the devices.
Should you have any further questions, please contact our Technical Sales department.
We are also happy to provide on-site service. Contact our Service Department for more
information.
Although great care has been taken in the drafting of this operating manual, it may nev-
ertheless contain errors and mistakes. Bender cannot accept any liability for injury to
persons or damage to property resulting from errors or mistakes in this manual.
MK2430_D00129_00_M_XXEN/12.2014 7
How to get the most out of this manual
This signal word indicates that there is a high risk of danger that will result
in death or serious injury if not avoided.
DANGER
This signal word indicates a medium risk of danger that can lead to death
or serious injury if not avoided.
WARNING
This signal word indicates a low level risk that can result in minor or
moderate injury or damage to property if not avoided.
CAUTION
This symbol denotes information intended to assist the user to make opti-
mum use of the product.
8 MK2430_D00129_00_M_XXEN/12.2014
2. Safety instructions
Insulation fault
Overload
Overtemperature
Messages from EDS… insulation fault location systems and RCMS… residual cur-
TER®
Supply line failure
Device failure
Test results
Measured values
The clear text display makes information easy to understand. The connection between
the MK… and the transfer switching and monitoring modules is implemented with bus
technology. During normal operation, the MK2430 indicates the readiness for opera-
tion of the system.
Version MK2430-11 features 12 digital inputs allowing messages from other technical
equipment to be recorded and displayed on the MK2430, for example from medical
gases or battery supported safety power supply systems (BSV systems).
MK2430_D00129_00_M_XXEN/12.2014 9
Safety instructions
Public buildings
Please heed the limits of the area of application indicated in the technical specifications.
Any other use than that described in this manual is regarded as improper.
ditions.
The observation of all information in the operating manual.
10 MK2430_D00129_00_M_XXEN/12.2014
Safety instructions
– As intended
– In perfect working order
– In compliance with the accident prevention regulations and the guidelines
applicable at the location of use.
Immediately rectify any faults that may endanger safety.
Do not make any unauthorised changes and only purchase spare parts and
immediately.
Make sure that the dimensions of the BSV (battery-supported safety power sup-
ply), the generator set and the whole wiring are adequate. The applicable
national and international standards must be observed here. Only in this way
selective operation of safety devices can be achieved and a high degree of safety
in case of overload and short circuit can be ensured.
MK2430_D00129_00_M_XXEN/12.2014 11
Safety instructions
12 MK2430_D00129_00_M_XXEN/12.2014
3. System description
3.1 MEDICS®
The MK2430 alarm indicator and test combinations are integral components of the
MEDICS® system. MEDICS® is an intelligent system that guarantees safe power supply in
medical locations.
MK2430_D00129_00_M_XXEN/12.2014 13
System description
Legend to example
MK2430 Alarm indicator and test combination
RCMS… Residual current monitoring systems for TN-S systems
SMI472 Signal converter for third-party systems (e.g. med. gases, BSV systems)
TM… Alarm indicator and operator panel
UFC107E… Changeover and monitoring module for IT systems with insulation
fault location system EDS…
UMC107E… Changeover and monitoring module for IT systems
UMC710D… Changeover module for main distribution boards
USC710D… Control module for changeover modules (preferably in main distri-
bution boards)
ZLT Central Control Technology
GLT Building Control Systems
MEDICS® includes:
Display and operating units such as TM… alarm indicator and operator panels or
MEDICS® system are: UMC…, USC…, UFC… and/or EDS… insulation fault location
systems.
Communication between these components via the BMS bus (two-wire connec-
tion).
The connection of third-party technical systems by means of protocol converters
3.2 Characteristics
On its backlit display (4 x 20 characters), the MK2430 displays messages from all BMS
bus devices assigned via alarm addresses. As well as being used as a standalone
indicator, the MK2430 also supports parallel indication of several MK2430 in different
rooms.
In the event of an alarm message, the yellow LED "WARNING" resp. the red LED "ALARM"
lights up and the message appears on the LC display in plain text format. At the same
time there is an audible signal (acknowledgeable/can be muted. If a second message is
14 MK2430_D00129_00_M_XXEN/12.2014
System description
received whilst the first is still pending, the audible signal will sound again and the
messages appear alternately on the LC display. The address of the device triggering the
alarm can also be requested. The audible signal sounds again once a configurable
period of time has elapsed (repetition can be deactivated).
Internal device parameters (alarm addresses, test addresses…) and the parameteri-
sation of EDS- and RCMS systems can be accessed via the menu system.. MK2430 can be
used as a master device in installations with several IT and EDS systems.
The test button can be used to check the function of the associated devices such as
insulation monitoring devices, LIM (Line Isolation Monitors) or GFCI (Ground Fault
Circuit interrupters). The message is available on the MK2430 on which the test button
"TEST" was pressed. The test and its individual evaluations are carried out sequentially.
Finally, a message appears that the test was successful or a fault message will be output.
MK2430 features:
Display of operating status, warning and alarm messages in accordance with
and via BMS bus: EDS46x, EDS47x, EDS49x, RCMS460, RCMS470, RCMS490
Types available for flush and surface mounting
History memory with real-time clock to store 250 warnings and alarms
MK2430_D00129_00_M_XXEN/12.2014 15
System description
3.3 Function
3.3.1 Display/operating elements
The backlit display features four lines of 20 characters each. It supplies medical and
technical personnel with clear and unambiguous information, to help them make
decisions. Every alarm message comprises three lines which appear spontaneously and
three additional lines which can be displayed at the touch of a button. The fourth line
contains status information (i.e. number of messages, test procedures, menu infor-
mation). Three LEDs are located above the text display. They indicate:
Normal operation (green), warnings (yellow) or alarms (red).
Five buttons are available to acknowledge or to mute alarms and warnings and for
testing the assigned devices and for the menu system.
16 MK2430_D00129_00_M_XXEN/12.2014
System description
3.4 Versions
3.4.1 MK2430-12
The MK2430-12 is used for visual and audible indication of alarms from Bender's EDS,
RCMS and MEDICS systems® and for testing assigned devices (e.g. insulation monitoring
devices, LIM, GFCI). Furthermore, the MK2430-12 can be used as parallel indication in
conjunction with MK2430-11 and SMI472-12. The programmed message texts are
displayed on the LCD in the selected national language.
3.4.2 MK2430-11
The MK2430-11 features all the functions of the MK2430-12. In addition, the MK2430-11
provides 12 digital inputs and a programmable relay output. All digital inputs, divided
into three groups of 4, are electrically isolated from each other. The input voltage is
LOW = 0…2 V). In practice, these digital inputs (IN1… IN12) are controlled via potential-
free contacts (N/C or N/O operation configurable). The voltage required for these inputs
is provided via the power supply unit, which also supplies power to the MK2430. Any
message text can be assigned to the inputs.
3.4.3 MK2430P-…
The MK2430P-… contains the programming of the standard display and up to 20
factory-set alarm addresses can be activated. A line containing a note about the
assigned equipment can be programmed individually for each alarm address. This line
appears as the first line in an alarm.
For many applications with medical IT systems, this programming scope is sufficient. A
programming template in the shape of a form is available for customer-specific
programming.
MK2430_D00129_00_M_XXEN/12.2014 17
System description
3.4.4 Interfaces
MK2430 feature a BMS bus and USB interface.
Only the connected MK2430 can be read out and set via the USB interface.
18 MK2430_D00129_00_M_XXEN/12.2014
4. Installation and connection
4.1 Installation
4.1.1 Overview of enclosure variants
– The MK2430 is installed in a flush-mounting enclosure. With a special mount-
ing kit it is also suitable for installation in a panel or cavity wall.
– MK2430A in surface-mounting enclosure
MK2430_D00129_00_M_XXEN/12.2014 19
Installation and connection
4.1.2.1 Flush-mounting
The flush-mounting enclosure is already enclosed in the scope of delivery of MK2430.
1. Insert the cardboard that has been supplied into the flush-mounting enclo-
sure to stabilise the enclosure and to provide protection against pollution.
2. Insert the enclosure so that it is flush with the wall surface.
1 2
The flush-mounting enclosure must not be installed lopsidedly or warped, and must
not be installed too deep below the surface.
20 MK2430_D00129_00_M_XXEN/12.2014
Installation and connection
9...50 mm 2...50 mm
MK2430_D00129_00_M_XXEN/12.2014 21
Installation and connection
MK2430A MK2430
A smooth and even surface is a precondition for installation. Only the fas-
tening screws specified below should be used. Failure to observe this can
result in deformation or damage to the enclosure.
Use the empty enclosure as a template for marking the drilling holes.
Maximum diameter of the screws:
Thread of screw: 3 mm, bolt head: 7 mm
22 MK2430_D00129_00_M_XXEN/12.2014
Installation and connection
4.2 Connection
Risk of electric shock!
Before fitting the enclosure and working on the device connections, make
sure that the power supply has been disconnected. .
Failure to comply with this requirement increases the risk of exposing the
DANGER personnel to an electric shock. Furthermore, the electrical installation may
be damaged and the device may be destroyed beyond repair.
– If you are connecting the MK2430 to a DC 24 V supply: Take the line voltage
drop into account if you are using long supply cables for the supply voltage.
– Consider the maximum permissible cable lengths for the supply voltage U2/V2
when supplied by AN450 or AN410 (see "Technical data").
The connection to the internal BMS bus is made via the terminals A/B. Connect
the BMS bus as described in the "BMS bus" instruction leaflet. Use a shielded cable
of at least 0.8 mm cross section for the interface line (e.g. J-Y(St)Y nx2x0.8). Con-
nect the shield to PE (earth) at one end. When routing the supply voltage Us
please note that a 4-wire cable (2 x BUS, 2 x Us) with suitable cross section is
required.
Set the terminating resistor for the BMS bus on the "Term" DIP switch. Factory set-
ting: Off.
2
MK2430-11 only: Use cables with a cross section of at least 0.75 mm when con-
necting the digital inputs and the relay output. The maximum cable length per
connection is 500 m.
MK2430_D00129_00_M_XXEN/12.2014 23
Installation and connection
6
7
1
56
4
3
isoMED. . .
2
ATICS. . .
1
24 MK2430_D00129_00_M_XXEN/12.2014
Installation and connection
Make sure that the MK2430 power supply is isolated from PE. If this is not
taken into consideration and if a PC is connected to the USB interface, the
CAUTION MK 2430 device and the PC may be damaged.
Terminating resistor:
If two or more devices are connected to each other via the BMS bus, the bus
1 line must be terminated at both ends with a resistor (R = 120 Ω). (The resistors
can be activated via DIP switch on the MK2430 and alarm indicator and opera-
tor panel).
Power supply unit in the MEDICS® module, sufficient for supplying power to at
3
least three MK2430.
Cable between the MEDICS® module and MK2430. Please observe the permissi-
4 ble cable lengths and cross sections when using the AN450 power supply unit
in MEDICS® modules.
Digital inputs: The digital inputs may be activated either via internal or external
voltage or potential-free contacts. If the inputs are controlled via an external
5 voltage, the common 0(-) is applied to terminal "0" and the 1(+) signal to the
relevant input IN1 … IN12. In this case, the connections between terminals 0
and V2, and the common connections and U2 are not required.
MK2430-11 only: Relay output. Programmable contact for device errors, test of
7
assigned devices*, device failure and common alarm message.
* Devices without BMS bus connection, but with test input (e.g. ISOMETER®)
MK2430_D00129_00_M_XXEN/12.2014 25
Installation and connection
4.2.3.1 MK2430-12
The MK2430-12 features only one terminal strip as it receives all messages via the BMS
bus. These messages can be received from an 107TD47, an MK2430-11, a signal
converter SMI47x, an EDS… or RCMS…, for example.
Supply voltage US :
looped through connection for supply voltage US.
U2, V2 Note: Make sure that the power supply of the MK2430 is isolated from PE.
If this is not taken into consideration and if a PC is connected to the USB
interface, the MK 2430 device and the PC may be damaged.
A, B BMS bus
26 MK2430_D00129_00_M_XXEN/12.2014
Installation and connection
4.2.3.2 MK2430-11
The MK2430-11 alarm indicator and test combination provides additional terminal
strips for the 12 digital inputs and one optional relay output.
Also refer to "Settings menu 5: Digital inputs (MK2430-11 only)" on page 60.
MK2430_D00129_00_M_XXEN/12.2014 27
Installation and connection
Example 1: Operating theatre or intensive care unit with two IT systems and
three rooms
med. gases
Device Parameters Address settings
28 MK2430_D00129_00_M_XXEN/12.2014
Installation and connection
Example 2: Intensive care unit with two IT systems and four rooms
UFC107E UFC107E
107TD47 107TD47 MK MK MK MK
PRC487 PRC487 2430 2430 2430 2430
PGH474 PGH474
EDS474 EDS474
MK2430_D00129_00_M_XXEN/12.2014 29
Installation and connection
30 MK2430_D00129_00_M_XXEN/12.2014
Installation and connection
Example 3: Intensive care unit with two IT systems and four rooms
„A“ „B“
ATICS ATICS
EDS151 EDS151
MK2430_D00129_00_M_XXEN/12.2014 31
Installation and connection
messages
up to 16
Device Parameters Address settings
32 MK2430_D00129_00_M_XXEN/12.2014
5. Commissioning and testing
The following flow chart explains the commissioning process:
1. Tests before switching on
2. Tests after switching on
3. Parameter setting
– Settings at the MK2430
– Settings in the TMK-SET software
4. Tests after parameter setting
Please note down all settings and attach these to the equipment or instal-
lation documentation.
During the setting of MK2430 using the configuration software TMK-SET, a
project file is being created. Save this file. Create a backup copy of this file
and keep it in a safe place.
MK2430_D00129_00_M_XXEN/12.2014 33
Commissioning and testing
34 MK2430_D00129_00_M_XXEN/12.2014
Commissioning and testing
All settings can be carried out via the TMK-SET software. Alternatively,
some settings can be carried out via the MK2430 menu (see diagrams).
MK2430_D00129_00_M_XXEN/12.2014 35
Commissioning and testing
36 MK2430_D00129_00_M_XXEN/12.2014
Commissioning and testing
BSV
MK2430_D00129_00_M_XXEN/12.2014 37
Commissioning and testing
no
Select "Test mode" from the make corrections
menu "External devices".
Are all external device
settings OK?
yes
no
Simulate fault messages make corrections
(e.g. device failure,
insulation fault). Generate
a fault directly or select the
"Test communication"
function. Is the fault
correctly signalled by the
MK2430?
no
Does the MK2430 react make corrections
properly to signals at the
digital inputs?
yes
MK2430 is ready for
operation.
38 MK2430_D00129_00_M_XXEN/12.2014
Commissioning and testing
Carried
Action Interval
out by
Functional test of the IT system monitoring (insulation, load cur- Once
rent, transformer temperature and connection monitoring) by Medical every
pressing the "TEST" button on the alarm indicator and test combi- personnel working
nation or on the alarm indicator and operator panel. day
Functional test of the transfer switching device*: Electri-
Functional test of the automatic transfer switching device. The cally Every six
instructions in chapter "Test of the switching device" (manual skilled months
"Transfer switching device") have to be observed! person
Electri-
Functional test of the IT system monitoring (insulation, load cur-
cally Every six
rent, transformer temperature and connection monitoring) on the
skilled months
insulation monitoring device.
person
Electri-
cally Every 12
Checking the setting values and changeover periods.
skilled months
person
Test the transfer switching device, the IT system monitoring, and Bender Every 24
the connection to the SCADA system (Supervisory Control and Data months
Acquisition) (if applicable) and the interaction between the compo-
nents in the system.
The test encompasses the following:
- Inspection: Marking, display elements, mechanical components,
wiring, parameterisation, connection of third-party systems, eval-
uation of fault memory.
- Measurement: Internal/external supply voltages/potentials,
bus voltage, bus protocol, bus scan.
- Testing: Device function, device communication.
- Documentation: Test results, recommendations for eliminating
defects.
* This test must only be performed by an authorised electrically skilled person in agree-
ment with the medical locations concerned.
Before carrying out the tests, please refer to the instructions relating to the functional
tests in the check list. If no national directives apply, you should perform the tests
recommended by IEC 60364-7-710 :2002-11, section 710.62 and DIN VDE 0100-710
(VDE 0100 Part 710):
MK2430_D00129_00_M_XXEN/12.2014 39
Commissioning and testing
Commissioning
Troubleshooting
Repair Service
Repair, calibration, update and replacement service for all Bender products.
Repair, calibration, testing and analysis of Bender products.
Extended warranty for Bender devices with in-house repair service resp. replace-
40 MK2430_D00129_00_M_XXEN/12.2014
Commissioning and testing
Field Service
On-site service for all Bender products.
Commissioning, parameter setting, maintenance, troubleshooting for Bender
products.
Analysis of the electrical installation in the building (power quality test, EMC test,
thermography).
Practical training courses for customers
*Available from 7.00 a.m. to 8.00 p.m. on 365 days of the year (CET/UTC+1)
**Mo-Thu 7.00 a.m. - 8.00 p.m., Fr 7.00 a.m. - 13.00 p.m.
Internet: www.bender.de
5.4.3 Maintenance
MK2430 does not contain any parts that require maintenance. Despite this, the intervals
specified for periodic verification should be adhered to.
MK2430_D00129_00_M_XXEN/12.2014 41
Commissioning and testing
42 MK2430_D00129_00_M_XXEN/12.2014
6. Troubleshooting
MK2430_D00129_00_M_XXEN/12.2014 43
Troubleshooting
6.2 Malfunctions
List of possible errors and proposals for the correction of the faults. This error list does
not claim to be exhaustive.
Possible error codes occurring during a test are listed in the chapter "Test function" on
page 49f.
Character matrix visible on display, Processor does not start; replace MK2430
but firmware doesn't run.
Error during transfer of the assigned MK2430 address not properly set (menu); MK2430
addresses or basic settings via the address does not match setting of TMK-SET configu-
USB interface. ration software; USB cable defective; wrong serial
interface (com port) set in TMK-SET.
Error on internal BMS bus. Device addresses not properly set on the internal
bus; interface cables A/B mixed up; network incor-
rectly terminated or not at all; incorrect parameter
setting with TMK-SET.
Digital input function error. Digital inputs not correctly set with TMK-SET. Defec-
tive connection (does not match pre-assignment).
Incorrect setting "neutral/medical".
44 MK2430_D00129_00_M_XXEN/12.2014
7. Operation
This chapter can also be used as a quick reference guide by technical operating
personnel.
2
1 3
5 8
6
9
7
4 LCD: Display of operating status and alarm messages as well as menu functions.
MK2430_D00129_00_M_XXEN/12.2014 45
Operation
"ESC" button
" " Mute button Exit function (without saving)
5 Set buzzer to mute following an alarm/ or go up one menu level.
acknowledge alarm When the buzzer is activated, the ESC
button will mute the buzzer.
"TEST" button
Press and release: LED test
8 Press and hold down: Initiate testing of No function
assigned devices (insulation monitoring
devices, LIM, GFCI)
"MENU" button
Starts menu mode for setting the "↵“ (Enter button)
9
MK2430; To confirm menu selection.
for display and control functions
46 MK2430_D00129_00_M_XXEN/12.2014
Operation
Power supply:OK
Med. gases: OK
09:50
"Alarm" will light up. The green LED "Normal" goes out.
At the same time the buzzer sounds. If the cause of the message cannot be reme-
died immediately, the buzzer can be deactivated by pressing the mute button
" " (5).
The LC display shows details about the message.
MK2430_D00129_00_M_XXEN/12.2014 47
Operation
zzz Meaning
noMA No master
48 MK2430_D00129_00_M_XXEN/12.2014
Operation
1. Test all devices Tests all devices set in the "test addresses" menu.
2. Test single device Select a device using the arrow buttons and press the
"↵" button.
During the test the status line displays "TEST". The message "TEST" flashes if the
message being displayed has been caused by the test.
The assigned devices are tested one after another. The MK2430 automatically evaluates
the messages that appear. Once the process is complete, either a common message
about a successful test or an error message is displayed.
If more than one device has been tested, a separate error code will be displayed for each
device failing the test.
MK2430_D00129_00_M_XXEN/12.2014 49
Operation
The following error codes are displayed in the event of an Isometer® failing the
test:
Error Error code description for the Error code description for the
Note
code 107TD47 ISOMETER® (hospital) IRDH… ISOMETER® (industry)
No messages received from the No messages received from the
ISOMETER® although the test ISOMETER® although the test
0
command was confirmed by the command was confirmed by the
ISOMETER®. ISOMETER®.
Only insulation fault message Only insulation fault message
1 Channel 1
received. from channel 1 received.
Only insulation fault message
2 Only overload message received. Channel 2
from channel 2 received.
Only insulation fault message and Channel 1
3
overload message received and 2
Only overtemperature message
4 Channel 3
received.
Only insulation fault message and
Channel 1
5 overtemperature message
and 3
received.
Only overload message and over-
6 Channel
temperature message received.
Test command could not be sent Test command could not be sent
14 because no query was made because no query was made Slave only
(slave). (slave).
ISOMETER® did not confirm the ISOMETER® did not confirm the
15
test command (no answer). test command (no answer).
mand, however, cannot be sent because MK2430 was not queried. This can be the
case if the address gap upstream of the MK2430 is so large that the master does
not scan the MK2430. (refer to "BMS bus" operating manual). This error code is not
so much an indication of a faulty ISOMETER® but that the BMS bus system is
faulty.
50 MK2430_D00129_00_M_XXEN/12.2014
8. Menu mode: Operation and setting
MK2430-11 Addr.:001
Software 4.02 D228
Date: 02.05.12
www.bender.de
If the MK2430 has not been turned on for several days, a longer time may
be required for start-up (approx. 30 seconds). In this case, enter the time
and date again.
SYSTEM READY!
09:50
The TMK-SET software allows you to change the standard display and the message
texts.
MK2430_D00129_00_M_XXEN/12.2014 51
Menu mode: Operation and setting
To open the main menu, press the "Menu" button for approx. two seconds.
1.Exit
2.Values
3.History
4.Settings
5.Control
6.External devices
7.Info
If no buttons have been pressed in a menu for more than five minutes,
menu mode is exited automatically (exception: "test communication" in
the "Control" menu and "position mode" in the "External devices" menu).
Once a valid password has been entered, full access will be given to all
menus (except the Service menu) until menu mode is exited.
52 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
1.Exit
2. Values
1.Exit
3.History
2.Alarm addresses
4.Value addresses
5.Digital inputs
6.Buzzer
7.Common reset
8.Clock
9.Language
10.Interface
11.Relay
12.Password
13.Service menu
1.Exit
2.Reset (AlarmClear)
5.Control
3.EDS Start/Stop
4.Test communication
6.External devices
7.Info
MK2430_D00129_00_M_XXEN/12.2014 53
Menu mode: Operation and setting
2. Values No function
1.Exit
2.Values
3.History
4.Settings
54 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
2. Press the "↵" button to select the message text of the selected entry. The last
line shows the address the message was sent to. In this case an insulation
fault was received via the internal BMS bus from address 003, channel 01.
The table on the following page contains information about other possible
displays.
System: 01
Insulation fault
Meas. value:
Address 003/01
3. If analogue values are displayed as shown in the example above you can
press the "▲" button to display the min. and max. values.
4. Press the "↵", button again to return to the entry selection.
Repeat these operating steps for all messages you need. Then press "ESC" to exit the
menu.
MK2430_D00129_00_M_XXEN/12.2014 55
Menu mode: Operation and setting
Possible displays in the last line of the history memory message text display:
Text Meaning
Address: Address of the device triggering the message (aaa= device address, kk=
aaa/kk channel no. of the message)
Digital Inp Digital input number (kk) of the digital input that has triggered the mes-
No.: kk sage on the device.
Note that the latest history entry was made due to a test initiated at this
TESTex
MK.
Note that the latest history entry was made due to a test initiated on the
TESTin
device assigned to it (insulation monitoring device, LIM, GFCI).
56 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
MK2430_D00129_00_M_XXEN/12.2014 57
Menu mode: Operation and setting
8.4.4.1 Exit
Exit menu mode.
1.Exit
2.Alarm addresses
3.Test addresses
4.Value addresses
1.Exit 1.Exit
2.Edit table 2.Address 002 On
3.Clear table System No.: nn
---Alarm addresses---
58 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
nn Meaning
1.Exit 1.Exit
2.Edit table 2.Address: 002
3.Clear table Device type: MED-Iso
---Test addresses----
MK2430_D00129_00_M_XXEN/12.2014 59
Menu mode: Operation and setting
1.Exit 1.Exit
2.Operation mode 2.Input: IN01: 24V
3.Function: neutral Alarm at 24V
---Digital inputs---
The TMK-SET PC software can be used to assign other messages to individual or all
digital inputs.
60 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
Alarm messages for medical gases are signalled by the red "ALARM" LED
and the buzzer. The buzzer can be set to mute (acknowledged). The buzzer
sounds again after 15 minutes as a reminder.
Different settings can only be made using the TMK-SET software.
MK2430_D00129_00_M_XXEN/12.2014 61
Menu mode: Operation and setting
1.Exit
2.Warning: 5
3.Alarm: 2
1.Exit
2.Common ack: On
62 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
1.Exit (CEST)
2.Time: 17:45
3.Date: 21.09.11
4.Format: dd.mm.yy
Time and date for the system can be set on any MK… or TM… panel. The
settings are transferred via the BMS bus to the master device (address 1),
which transfers the settings to all other MK… or TM…panels.
MK2430_D00129_00_M_XXEN/12.2014 63
Menu mode: Operation and setting
Portuguese
Spanish Portuguese Dutch
(Brazil)
64 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
1.Exit
2.Addr. RS485: 001
3.Baud RS485: 09600
Change the corresponding device address if several MK2430 are connected to one
internal BMS bus. One MK2430 must have the address 1 (master). All other MK2430 are
addressed in consecutive order: 2,3,4… There should be no gaps between addresses.
Only in this way the functionality of the system can be ensured (also refer to the "BMS
bus" manual).
Data exchange will only work between devices with the same baud rate.
MK2430_D00129_00_M_XXEN/12.2014 65
Menu mode: Operation and setting
1.Exit
2.Relay mode: N/O
3.Function:
Device error
Setting Description
66 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
1.Exit
2.Password: X X X
3.Status: off
Once a valid password has been entered, full access will be given to all
menus (except the Service menu) until menu mode is exited.
MK2430_D00129_00_M_XXEN/12.2014 67
Menu mode: Operation and setting
2. Reset
Resetting all fault messages pending on the BMS bus. 68
(AlarmClear)
8.4.5.1 Exit
Exit menu mode.
1.Exit
2.Reset (AlarmClear)
3.EDS start/stop
4.Test Communication
1.Exit
2.Reset (AlarmClear)
3.EDS start/stop
Reset done!
This reset command is sent via the BMS bus. Note that some devices (e.g. PRC470) do
not respond to this reset.
68 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
1.Exit 1.Exit
2.Reset (AlarmClear) 2.Reset (AlarmClear)
3.EDS start/stop 3.EDS start/stop
EDS system running EDS system stop
EDS system runningOnce started, the EDS4xx-12 and PGH47x run continuously. If you
exit the menu, the standard display "EDSp" will appear in
the last line in order to indicate continuous running*.
EDS system Stop Stops the continuous running of EDS4xx-12 and PGH47x.
If you exit the menu, the standard display "EDS" will
appear in the last line until the current test run finishes.
*
Other abbreviations that might appear in the last line of the display:
EDSa Automatic mode: Insulation fault location has been
started on the PGH47x by e.g. ISOMETER®, ATICS® or
isoMED427P.
EDSs Single mode: A single run has been started by IN2 on
PGH471.
1.Exit
2.Channel: 003
-Test communication-
MK2430_D00129_00_M_XXEN/12.2014 69
Menu mode: Operation and setting
All devices connected to the BMS bus are indicated. Select the address of the external
device to be displayed (e.g. EDS4xx-12 or RCMS4xx-12).
1. Exit
001: MK2430 V4.02
002: EDS151 V1.00
003: ATICS2-ISO V1.08
Address, type and version of the connected device are indicated. When no device is
detected under this address, the character "?" appears. Use the arrow buttons to select
the address of the external device and confirm your selection with the "↵" button.
If the device has been recognised, the MK2430 will read the current settings of the
connected device. The device type is displayed in the first line of the display.
Use the Up/Down buttons to select the appropriate function or device setting. Confirm
with the "↵" button. Example:
A variety of setting options are available in the EDS or RCMS menu. These are described
in more detail in the operating manuals for the corresponding systems.
1. Exit (EDS470)
2. General
3. Channel
4. Relay
70 MK2430_D00129_00_M_XXEN/12.2014
Menu mode: Operation and setting
MK2430-11 Addr.:001
Software 4.02 D228
Date: 02.09.11
www.bender.de
Information about the device type, the version and the last time the assignments were
transmitted. Assignments are settings carried out via the PC software TMK-SET:
Enter standard texts
Assign texts and functions to the alarm messages and digital inputs of the
MK2430.
Set parameters
Select:
MK2430_D00129_00_M_XXEN/12.2014 71
Parameter Set in
72
Name Range Factory User MK2430 MK2430 TMK-SET Note
setting setting menu Service Parameter
menu menu
History entries -- -- -- -- clear clear Data records
Clear history
Power-down watchdog -- -- -- -- clear clear Reset counter
Time-out int. 42…200 ms 60 ms M13 x x Change only for test
purposes!
ALMI idle time 1…2 s 1s M13 x x Time between two
ALMI queries on the
internal BMS
Backlight autom./cont. autom. M13 x x
S/W time autom./OFF autom. M8 -- x Automat.
summer/winter time
Buzzer ON/OFF for ON -- M13 x -- e.g. for
1...9 hrs commissioning
Password prompt Y/N Y M12 -- x Activate password
prompt
Password 000...999 807 M12 -- x Change password
Language (for menu only) English/ German M9 -- x
German
Language (for message 20 German M9 -- x
only) languages
Time/message 3…8 s 5s M13 x x Message interval
Buzzer settings Duration/ 2 M6 -- x
interval 1…5
frequency
Common acknowledge Y/N Y M7 -- x Response to "buzzer
int. OFF" via BMS bus
Time of day/date M8 x
RS-485 settings address 1…150 1 M7 -- x Int. BMS bus
address
Time display Y/N Y M13 x x Time in the standard
display
Display ATICS Y/N N M13 x x
maintenance messages
Menu mode: Operation and setting
MK2430_D00129_00_M_XXEN/12.2014
9. Technical data
MK2430_D00129_00_M_XXEN/12.2014 73
Technical data
74 MK2430_D00129_00_M_XXEN/12.2014
Technical data
Colours
Front foil........................................................................... RAL 7035 (light grey); RAL 7040 (basalt grey)
Marking........................................................................................................ RAL 5005 (ultramarine blue)
Front plate ............................................................................................................... RAL 7035 (light grey)
Switching elements (MK2430…-11 only)
Number .....................................................................................................................1 changeover contact
Function ............................................................................................................................. programmable
Operating principle ...................................................................... N/C or N/O operation (programmable)
Electrical service life under rated operating conditions...............................10.000 switching operations
Contact data acc. to IEC 60947-5-1
Utilisation category..............................................................AC-13...................... AC-14 .................DC-12
Rated operational voltage ......................................................24 V......................... 24 V ....................24 V
Rated operational current.........................................................5 A........................... 3 A ......................1 A
Minimum contact rating.......................................................................................1 mA at AC / DC > 10 V
Environment/EMC
EMC immunity ........................................................................................................ acc. to EN 61000-6-2
EMC immunity ........................................................................................................ acc. to EN 61000-6-3
Classification of climatic conditions acc. to IEC 60721:
Stationary use ........................................................................................................................................3K5
Transport................................................................................................................................................2K3
Long-term storage.................................................................................................................................1K4
Operating temperature..........................................................................................................-5…+55 °C
Classification of mechanical conditions acc. to IEC 60721:
Stationary use ...................................................................................................................................... 3M4
Transport.............................................................................................................................................. 2M2
Long-term storage............................................................................................................................... 1M3
Connection
Connection pluggable screw terminals
Connection properties (supply voltage, BMS bus):
Connection of single conductors
rigid/flexible/conductor sizes...................................................0.2…2.5 / 0.2…2.5 mm2/AWG 24-12
flexible with ferrules, without/with plastic sleeve .................................. 0.25…2.5 / 0.25…2.5 mm2
Multi-conductor connection (2 conductors with the same cross section):
Rigid/flexible..................................................................................................... 0.2…1 / 0.2…1.5 mm2
flexible with ferrule without plastic sleeve........................................................................ 0.25…1 mm2
flexible with TWIN ferrule with plastic sleeve ................................................................ 0.5 … 1.5 mm2
MK2430_D00129_00_M_XXEN/12.2014 75
Technical data
76 MK2430_D00129_00_M_XXEN/12.2014
Technical data
MK2430-11 Alarm indicator and test combination according to IEC B 9510 0001
60364-7-710:2002-11 (DIN VDE 0100-710),
featuring a BMS bus and a USB interface, 12 digital inputs,
one relay output, alarm texts programmable via interfaces
and PC, standard text display. Type: Flush-mounting enclo-
sure
MK2430-12 Alarm indicator and test combination according to IEC B 9510 0002
60364-7-710:2002-11 (DIN VDE 0100-710),
featuring a BMS bus and a USB interface,
alarm texts can be programmed via interfaces and PC,
standard text display. Type: Flush-mounting enclosure
MK2430PA-11 As MK2430A-11, but programmed in the factory and with B 9510 0007
surface-mounting enclosure.
MK2430PA-12 As MK2430A-12, but programmed in the factory and with B 9510 0008
surface-mounting enclosure.
MK2430S-11 As 2430-11, but front plate is fixed in place with screws. B 9510 0011
MK2430S-12 As 2430-12, but front plate is fixed in place with screws. B 9510 0012
MK2430- Mounting angle kit for flash-mounting enclosure to enable B 9510 1000
mounting kit, snap-on mounting of the MK2430. Mounting kits for cavity
complete wall mounting, panel mounting or DIN rail mounting.
MK2430_D00129_00_M_XXEN/12.2014 77
Technical data
78 MK2430_D00129_00_M_XXEN/12.2014
INDEX
A Date format 63
Acknowledgement 62 Device address 65
Additional text 48 Diagram 53
Address settings 28, 32 Digital inputs 25, 60
Alarm addresses 16, 58 Display 16
Alarm message 16 Drilling holes 19
- neutral 60
- specific 60 E
Alarm relay 66 Enclosure variants 19
alarm relay 43 Error codes 50
Assignments 71 Error list 44
Audio frequency 62 Evaluating devices 70
Automatic mode 69 External devices 70
B F
Baud rate 65 fault
Beep c 43 - condition 47
Bender Service personnel 67 Firmware version 51, 71
BMS bus 18 Flush-mounting enclosure 20
Buzzer 47, 62, 66 Functional test 39
C H
Cable length 23 History memory 17, 55
Cavity wall mounting 21
Changeover periods 39 I
Characteristics 15
Intended use 9
Commissioning 33
Interfaces 18
Common acknowledgement 62
Communication 69
Connection 23 L
Control 68 Language 64
LEDs 16
D
Date 63
MK2430_D00129_00_M_XXEN/12.2014 79
INDEX
M Reset 68
Main menus 52
Master password 67 S
Measured value addresses 60 Selective operation 11
Medical Setting 57
- gases 61 Setting possibilities 71
Medical gases 61 Setting values 39
MEDICS® 13 Single mode 69
MK2430-11 17 Snap-on mounting 21
MK2430-12 17 Software 16, 18
MK2430P-… 17 Standard display 51
Mounting brackets 21 Status line 48, 49
Summertime 63
N Switching on 34, 35
National languages 16 System components 7
Navigation 53
T
O Terminating resistor 23
Operation 45 Terminating resistor
- fault-free 47 25, 26, 27
Ordering information 77 Test 49
Test procedure 69
Text messages 17
P
Time 63
Panel mounting 21 TMK-SET 58, 60, 61
Parallel indication 14, 17 Transmission rate 65
Password 52, 67
Periodic verification 38
U
Personnel 10
Potential free contacts 17 USB interface 18
Programming 17
Q
Quick reference guide 45
R
Regulations for erection and installation 76
Relay mode 66
Relay output 27
80 MK2430_D00129_00_M_XXEN/12.2014
Bender GmbH & Co. KG
P.O. Box 1161 • 35301 Grünberg • Germany
Londorfer Straße 65 • 35305 Grünberg • Germany
Tel.: +49 6401 807-0 • Fax: +49 6401 807-259
E-Mail: info@bender.de • www.bender.de
Photos: Bender archives. BENDER Group