Sunteți pe pagina 1din 10

Arabian Journal of Chemistry (2019) 12, 388–397

King Saud University

Arabian Journal of Chemistry


www.ksu.edu.sa
www.sciencedirect.com

ORIGINAL ARTICLE

Lanthanum(III) potentiometric sensors based on


ethyl benzoyl acetateq
Eman Y.Z. Frag a,*, Refat F. Aglan b, Hala A. Mohamed a

a
Chemistry Department, Faculty of Science, Cairo University, Gamaa Str., 12613 Giza, Egypt
b
Hot Laboratory Center, Atomic Energy Authority, Cairo, Egypt

Received 25 July 2016; accepted 26 November 2016


Available online 5 December 2016

KEYWORDS Abstract Modified carbon paste electrode (CPE) and screen printed electrode (SPE) based on
SPE; ethyl benzoyl acetate (EBA) were prepared and investigated as lanthanum ion selective electrodes.
CPE; Effect of various plasticizers (o-NPOE, TCP, DBP, DOS, and DOP) and ionophore content was
Lanthanum(III); studied. The reaction mechanism between La(III) and b-diketone ionophore at the electrode surface
b-diketone; was studied through scanning electron microscopy (SEM), energy dispersive X-ray analysis (EDX)
SEM; and IR spectra measurement. The best performance was obtained using modified SPE and CPE
EDX electrodes with o-NPOE and TCP plasticizers. These electrodes showed potentiometric response
with a Nernstian slope of 21.0, 19.5 and 20.5 mV decade1 over a concentration range from
1.0  106 to 1.0  102 mol L1 with a fast response time of 9, 10 and 13 s over the linear concen-
tration range for modified SPE (electrode I; TCP plasticizer) and two modified CPE plasticized with
o-NPOE (electrode II) and TCP (electrode III), respectively. The electrodes exhibited constant
potentiometric response in pH range 4–8, 5–8 and 4–7 for electrodes I, II and III, respectively. They
showed satisfactory good sensitivity toward lanthanum ions with regard to most common transition
metal ions. The sensors were applied for determination of La(III) ion in different water samples (tap
water and wastewater) with satisfactory and good reproducibility results.
Ó 2016 The Authors. Production and hosting by Elsevier B.V. on behalf of King Saud University. This is
an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

1. Introduction

q The reaction mechanism between La(III) and ethyl benzoyl


Lanthanides were considered due to the unique physical and chemical
acetate ionophore at the electrode surface was studied through properties of the rare earth elements (REEs). Lanthanides were used
scanning electron microscopy (SEM) energy dispersive X ray (EDX) widely in metallurgy, medicine, chemical engineering, electronics and
and IR spectra measurement. electrooptics, biomedicine, for manufacturing of magnetostriction
* Corresponding author. materials, lasers, storage batteries with long service life for electric
E-mail address: e_uossry@yahoo.com (E.Y.Z. Frag). vehicles, etc. (Tadjarodi et al., 2015). Lanthanum was one of lan-
Peer review under responsibility of King Saud University. thanide elements found in rare earth minerals such as cerite, monazite,
allanite and bastnasite (Khan et al., 2014). Lanthanum(III) was consid-
ered as fission products. It may be leaked to the environment and cause
risks. So they were considered as hazardous materials (Besharati-
Production and hosting by Elsevier Seidani and Shamsipur, 2015). The lanthanum alloy LaCo5 was used

http://dx.doi.org/10.1016/j.arabjc.2016.11.015
1878-5352 Ó 2016 The Authors. Production and hosting by Elsevier B.V. on behalf of King Saud University.
This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
Lanthanum(III) potentiometric sensors 389

as a permanent magnet and LaNi5 for hydrogen storage. Its high index benzoyl acetate (EBA) were supplied from Aldrich as elec-
of refraction made very pure La2O3 a useful additive to optical glass troactive materials. Lanthanum chloride was supplied from
for camera lenses (Singh et al., 2012). Aldrich. Graphite fine powder (extra pure) was of analytical
Because of the discharge of lanthanum ions in the environment and grade and purchased from Merck. High molecular weight
its useful and harmful biological activity, the La(III) ion determination
polyvinyl chloride (PVC) powder was supplied from Aldrich,
was carried out using different analytical methods. These methods
included flame photometry, atomic absorption spectrometry, induc-
while o-nitrophenyl octyl ether (o-NPOE) was supplied from
tively coupled plasma-mass spectrometry (ICP-MS) Islamnezhad Fluka. Dioctylphthalate (DOP), tricresyl phosphate (TCP),
et al. (2011), ion chromatography (Al-Shawi and Dahl, 1996), atomic dibutyl phthalate (DBP) and dioctylsebacate (DOS) were sup-
emission spectrometry (Jia et al., 2008), inductively coupled plasma- plied from BDH. Acetone and cyclohexanone were supplied
optical emission spectrometry (ICP-OES) Waqar et al. (2009), sector from Fluka (Switzerland).
field inductively coupled plasma mass spectrometry (Chung, 2009)
and electrothermal vaporization inductively coupled plasma mass 2.2. Apparatus and emf measurements
spectrometry (Zhang et al., 2007), etc.
During last two decades, a general interest in metal b-diketonates
Digital Hanna pH/mV meter (model 8417) was used for all
(Mehrotra et al., 1978) was reflected in the chemistry of lanthanides.
potentiometric measurements at 25 °C. Spectrophotometer
This resulted in detailed investigation unfolding some lesser known
facts of general lanthanide coordination chemistry. Early methods (U-2001, model 121-0032 Hitachi, Tokyo, Japan) and scanning
used for preparation of lanthanide tris-b-diketonates were examined electron microscope (National Research Center Quanta
previously (Moeller et al., 1961). It was established that the products FEG250) were used and microanalysis was completed using
from aqueous solutions were invariably hydrates while sometimes the energy dispersive X-ray analyzer (EDX) (National research
hydroxy derivatives were also formed. A large number of center, Egypt). FT-IR spectra were recorded on a Perkin-
b-diketonates of lanthanides were prepared (Dutt et al., 1971; Dutt Elmer 1650 spectrometer (4000–400 cm1) in KBr pellets at
and Rahut, 1971; Dutt and Samyal, 1971; Stites et al., 1948). the Microanalytical Center, Cairo University, Egypt. Ag/AgCl
One of the most important analytical techniques capable of deter- reference electrode contains 10% (w/v) potassium chloride.
mining both organic and inorganic materials in medico-biological
practice (Mohamed et al., 2013) was ion selective electrodes (ISE).
2.3. Sensors preparation (screen printed and carbon paste)
There was a constant progress in the number of electrodes capable
of selectively determining large numbers of metal ions such as carbon
paste electrodes (CPEs). Although the CPEs had an important role in CPEs were prepared using Teflon holder for carbon paste
the electrochemical analysis, the prepared pastes were soft, non- filling as the electrode body, and the carbon paste was formed
compactable and had to be packed into a special piston shaped holder. by mixing graphite powder (250 mg), a suitable liquid
However, in the case of determinations in flowing streams or field binder (DOP, TCP, DBP, DOS or o-NPOE) (100 lL) and
monitoring with portable analyzers, the shapes and designs of such
2.5–12.5 mg of ionophore. This matrix was thoroughly mixed
sensors were not suitable for every purpose as where the respective
in the mortar and the resulted paste was used to fill the
detection units need sensors of special characters (Svancara et al.,
2005). electrode body. A new carbon-paste surface was obtained by
While screen printing electrodes were reproducible, inexpensive pushing gently the stainless-steel screw forward.
and sensitive disposable electrochemical electrodes were used for the SPEs were printed in arrays of six couples consisting of the
determination of trace levels of substances in environmental, pharma- working electrodes (each 5  35 mm) following the procedures
ceutical and biomedical samples (Fanjul-Bolado et al., 2008). previously described (Nour El-Dien et al., 2012; Frag et al.,
The use of modified chemical sensors in determination of metal 2011a–c, 2012; Akl et al., 2013; Mohamed et al., 2010). A poly-
ions was developed in recent years as they provide fast, accurate, sen- vinyl chloride flexible sheet (0.2 mm) was used as a substrate
sitive, reliable and low cost method rather than the methods mentioned which was not affected by the curing temperature or the ink
before. So the aim of the present paper was to fabricate modified
solvent and easily cut by scissors. The working electrodes were
screen printed electrode (electrode I) and carbon paste electrodes
printed using homemade carbon ink (prepared by mixing
(II and III) with ethyl benzoyl acetate (EBA) as electroactive material
for selective determination of lanthanum (III) ion in different water 1–15 mg of EBA electroactive material, 450 mg TCP, 1.25 g
samples. of polyvinyl chloride (8% wnv) and 0.75 g carbon powder).
The type and content of ionophore, suitable pH, plasticizer type, They were printed and cured at 50 °C for 2 h. A layer of an
effect of interfering ions and temperature effect on the performance insulator was then placed onto the printed electrodes, leaving
of the electrodes were also studied. The reaction mechanism at the elec- a defined rectangular shaped (5  5 mm) working area and a
trode surface was studied using IR spectra. The change in the electrode similar area (for the electrical contact) on the other side. Fabri-
surface as the results of complex formation between EBA ionophore cated electrodes were stored at 4 °C and used directly in the
and La(III) ion was studied using SEM, EDX and IR spectra. The potentiometric measurements without any preconditioning.
method was successfully applied for determination of lanthanum(III)
ion in different real water samples using spiking technique. Method
2.4. EMF measurements
validation parameters were optimized according to ICH guidelines
and the proposed potentiometric method can be applied successfully
for the determination of La(III) ions in routine samples. The emf measurement with the plasticized SPE and CPEs was
carried out with the following cell assemblies:
2. Experimental AgjAgCljsatd: KClksample solutionjsensor ðCPE and SPEÞ:

2.1. Materials The detection limit was taken at the point of intersection of
the extrapolated linear segments of the calibration curve which
Reagent grade acetyl acetone (AA), b-diketone 1-benzoyl-2- was plotted with emf as a function of the negative logarithm of
nonanone (Lix 54), ethylacetoacetate (EAA), and ethyl lanthanum ion concentration, and the selectivity of the
390 E.Y.Z. Frag et al.

electrodes is obtained by measuring the selectivity coefficient ion determination. The potential response is shown in Figs. 1
1
(Kpot
La,J) using the separate solution method using 0.001 mol L and 2 for modified SPE and CPEs electrodes, respectively.
La(III) and interfering ions. It was found from the figures that the sensors modified with
2.5 mg and 10 mg of EBA ionophore which was the most effec-
2.5. Determination of La(III) ion in different water samples tive amount gave the best sensitivity with the best Nernstian
slope of 21.0, 19.9 and 20.5 mV decade1, over a wide concen-
Tap water and wastewater samples were collected from differ- tration range from 1.0  106 to 1.0  102 mol L1 with a
ent environment filtered and analyzed for La(III) ion by the good correlation coefficient and very low detection limit for
proposed method. Tap water (sample 1) and wastewater (sam- electrodes I, II and III, respectively. The limits of detection
ple 2) samples were supplied from Sandoub, Mansoura, were determined from the intersection of the two extrapolated
Dakahliya, Egypt. segments of the calibration plots and were found to be
The water samples were spiked with known amounts of lan- 1.0  106 mol L1.
thanum(III) ion, the pH was adjusted with acetate buffer
(pH = 4) and they were analyzed by the proposed general 3.1.3. Effect of plasticizers type in CPE
procedure. The nature of plasticizer was expected to play a key role in
For spectrophotometric measurements at 430 nm a given determining the ion-selective characteristics as it influences
amount of alizarin Red S was added to 5 beakers containing the dielectric constant of the paste of the electrode, the mobil-
10 ml spiked water samples with acetate buffer solution and ity of the ionophore molecule and state of ligand (Gupta et al.,
the pH of the solution was adjusted to 4.60 with hydrochloric 2003). It also caused selective extraction of the target ion which
acid or sodium hydroxide and then diluted to 50 ml. This was created the electrochemical phase boundary potential due to
followed by absorbance measurements. The molecular ratio of thermodynamic equilibria at the interface. So, various plasti-
lanthanum:alizarin Red S is 1:2 (Kawashima et al., 1961). cizers were used namely DBP, DOP, o-NPOE, DOS and
TCP in order to study their influence on the performance of
2.6. Surface analysis the sensors.
Among these plasticizers, o-NPOE and TCP provided fas-
Scanning electron microscope (SEM) and energy dispersive ter, more stable and sensitive response in the concentration
X-ray (EDX) analyzer were used for morphological analysis range from 1.0  106 to 1.0  102 mol L1 as shown from
of the sensors surface at 4000 magnifications for electrodes the graph. The slope was 19.9 and 20.5 mV per decade for
I and III before and after the interaction with lanthanum CPE electrodes II and III, respectively. This can be explained
(III) ion. due to its high dielectric constant and relatively high molecular
weight.
According to the Lubricating Theory of plasticization
3. Result and discussion (Wilkes et al., 2005), the plasticizer molecules diffuse into the
polymer and act as shields to reduce polymer–polymer interac-
3.1. Electrode composition tive forces (van der Waals forces) and hence prevent the forma-
tion of a rigid network. This lowers the PVC Tg and allows the
3.1.1. Type of ionophore polymer chains to move rapidly each other, resulting in
An important requirement for the preparation of an ion selec-
tive sensor was the electroactive material (ionophore), which
was used in the paste. It should exhibit high lipophilicity and
strong affinity for a particular ion to be determined and poor
affinity for others. They should have rapid exchange kinetics
and adequate complex formation constants in the paste and
should be well soluble in the paste matrix. It should have a suf-
ficient lipophilicity to prevent leaching from the paste into the
sample solution.
The lanthanum ion showed high affinity toward b-diketone
compounds due to the higher oxidation-reduction potential,
higher bond energy with oxygen and larger atomic radius
(Bian et al., 2006). Different b-diketone compounds namely
acetyl acetone (AA), Lix 54, ethyl acetoacetate (EAA), and
ethyl benzoyl acetate (EBA) were used to study the selectivity
toward lanthanum(III) ion showing potentiometric response
with slope values of 23.0, 18.0, 12.5 and 19.9 mV decade1,
respectively. It was obvious that the EBA ionophore was the
best electroactive material toward lanthanum(III) ion since it
gave the best slope within the studied concentration range.

3.1.2. Ionophore content Fig. 1 Effect of ionophore content on potential response


Different amounts of EBA ionophore as electroactive material of SPE (electrode I) for determination of La(III) in acetate buffer
were used to prepare the modified SPE and CPEs for La(III) (pH = 4).
Lanthanum(III) potentiometric sensors 391

the spectra is observed. The bands due to t(C‚O) of benzoyl


and ester groups were shifted to lower wave numbers in the
spectrum (1729 and 1644 for electrode I and 1741 and
1613 cm1 for electrode III) suggesting complex formation
via coordination of the EBA ionophore to La(III) ion through
the C‚O of benzoyl and C‚O of ester groups .

3.2. The effect of pH

The pH dependence of the electrode potential was investigated


over the pH range of 2.0–10.0 in a 1.0  103 and 1.0  105
mol L1 solution of La(III) ion. The pH was adjusted by using
very small volumes of HCl and NaOH solutions. The results
indicated that the potential was independent of pH in the
range of 4–8, 5–8 and 4–7 for electrodes I, II and III, respec-
tively. Outside this range, the potential changed significantly.
The increase of potential below pH 4.0 may be ascribed to
the competitive binding of protons to the ligands on the elec-
trode surface rather than La(III) ion. The decrease of potential
above pH 8.0 can be accounted for to the formation of hydro-
Fig. 2 Effect of ionophore content on potential response of CPE
xyl complexes of La(III) such as La(OH)+2, La(OH)+ 2 and La
(electrode III) for determination of La(III) in acetate buffer
(OH)3 which monished its ability to combine with the carrier
(pH = 4).
(Yuan et al., 2013; Shamsipur et al., 2002).

increased flexibility, softness, and elongation. Different plasti- 3.3. Response time
cizers yield different plasticization effects because of the differ-
ences in the strengths of the plasticizer–polymer and The time it takes for the electrode potential of an ion selective
plasticizer–plasticizer interactions. At low plasticizer levels, electrode to become stable within a range of variation of
the plasticizer–PVC interactions were the dominant interac- ±1 mV depends on electrode type and structure, ion type,
tions, while plasticizer–plasticizer interactions can become concentration, and ionic strength. The response time when
more significant at high plasticizer concentrations. The polar measuring a high ion concentration after measuring low ion
portion of the molecule must be able to bind reversibly with concentrations was relatively short, while a longer response
the PVC polymer, thus softening the PVC, while the non- time was obtained for the reverse process. Additionally, at
polar portion of the molecule allows the PVC interaction to around the minimum measurement limit, the response time
be controlled so it is not so powerful a solvator as to destroy was generally relatively long, being of the order of several
the PVC crystallinity. minutes. Hence, when using the ion electrode method, it was
necessary to wait until the potentiometer indicated a stable
3.1.4. Surface characterization value before taking a reading (Eric et al., 1997).
In order to characterize the morphology of the CPE and SPE, One of the most important factors for any ion-selective
scanning electron microscopy (SEM) and energy dispersive electrode is response time. The practical response time of the
X-ray (EDX) analyzer were used as shown in Figs. 3–6. All sensor was recorded by observing the potential change over
electrodes surfaces showed a similar configuration with ran- a concentration range from 1.0  106 to 1.0  102 mol L1
domly directed graphite particles ensemble into an insulator of La(III). The potentials versus time traces are shown in
matrix (PVC or plasticizer for SPE and CPE, respectively) Fig. 9.
where the graphite particle size is longer for CPE than for As can be shown in Fig. 9, electrodes reached their equilib-
SPE, so this distribution of the electrode material in case of rium responses in a very short time of 9, 10 and 13 s over the
SPE (microelectrodes) could lead to a highly packed structure. linear concentration range for electrodes I, II and III, respec-
SPE showed a more uniform surface topography rather than tively. This may be due to the fast exchange kinetics of com
CPE. Incorporation of a suitable ionophore in the paste fol- plexation–decomplexation of La(III) ion with the ionophore
lowed by soaking of the sensors in La(III) ion solution led to at the paste of the electrodes (Ganjali et al., 2004).
the complex formation between the La(III) ions and EBA
ionophore which was observed in filling the space between gra- 3.4. Effect of temperature
phite surface after the interaction with La(III) ion. Formation
of ionophore-La3+ complex at electrode surface which was One of the important parameters which has effect on the per-
subsequently extracted by the plasticizer into the paste is formance of the sensors is the change of temperature at test
shown in Figs. 4 and 6. These data are supported by IR spectra solution. The stability of the electrodes was studied within
as shown in Figs. 7 and 8, where the ligand EBA exhibited the temperatures 10, 20, 30, 40, 50 and 60 °C. The electrode
medium intensity bands at 1740 and 1685 cm1 due to exhibited good Nernstian behavior in the temperature range
t(C‚O) of benzoyl and ester groups for electrodes I and III, (10–50 °C). The slope of electrode did not show a good
respectively. After soaking electrodes I and III in La(III) ion Nernstian behavior at higher temperature. The standard cell
solution for 1 h and carrying out IR spectra, a difference in potentials (E°Cell), were determined at different temperatures
392 E.Y.Z. Frag et al.

Fig. 3 SEM images for SPE surface, electrode I (a) before and (b) after soaking in 1  103 mol L1 La(III) ion.

Fig. 4 EDX of SPE surface, electrode I (a) before and (b) after soaking in 1.0  103 mol L1 La(III) ion.

Fig. 5 SEM images for CPE surface, electrode III (a) before and (b) after soaking in 1  103 mol L1 La(III) ion.
Lanthanum(III) potentiometric sensors 393

Fig. 6 EDX of CPE surface, electrode III (a) before and (b) after soaking in 1.0  103 mol L1 La(III) ion.

Fig. 8 IR spectra for electrode I (a) before and (b) after soaking
in 1.0  103 mol L1 La(III) ion.
Fig. 7 IR spectra for electrode III (a) before and (b) after
soaking in 1.0  103 mol L1 La(III) ion.
3.5. Selectivity

Clearly, the selectivity was one of the most important charac-


from the respective calibration plots as the intercepts of these
teristics of a sensor. It often determines whether a reliable mea-
plots at p La(III) = 0, and were used to determine the isother-
surement in the target sample was possible or not. In clinical
mal temperature coefficient (dE0/dt) of the cell with the aid of
applications, it was critical where the allowed emf deviation
the following equation (Khalil and El-Aliem, 2002;
(error) may sometimes not be larger than 0.1 mV for whole
Electrochemistry, 1972):
blood or serum measurements. Theoretically, selectivity
E0cell þ E0reference ¼ E0electrode description allowed researchers to identify the key parameters
for optimizing the performance of potentiometric sensors, e.g.,
Plot of E0 electrode versus (t–25) gave a straight line. The by adjusting weighing parameters (i.e., absolute membrane
slope of the line was taken as the isothermal temperature coef- concentrations) or choosing different plasticizers or matrices
ficient of the La (III) electrode. It amounted to 0.121  102, (Eric et al., 1997). Ion selective electrodes have the most impor-
0.55  103 and 0.245  102 V/°C for electrodes I, II and tant characteristic that is its relative response to other ions pre-
III, respectively. The small values of (dE0/dt)cell and sent in solution, which expressed in terms of selectivity
(dE0/dt)electrode revealed the high thermal stability of the coefficients (KPotA,B). Different methods can be used for measur-
electrodes within the investigated temperature range. ing the selectivity coefficients of the electrode such as separate
394 E.Y.Z. Frag et al.

Fig. 9 Dynamic response time of (a) electrode I, (b) electrode II and (c) electrode III for La(III) in acetate buffer (pH = 4).

solution method (SSM) Bakker et al. (2000) and matched selectivity coefficient was smaller than 1.0, it showed that the
potential method (MPM) Lindner and Umezawa (2008). In sensor was selective to the primary ion over the interfering
the present study SSM was used which depends on measuring ions. The data given in Table 1 indicate that the electrodes
the potential of a cell comprising an ISE and a reference elec- (I and III) have good selectivity toward the lanthanum ion
trode with two separate solutions. One contains the ion of than other metal ions. But for electrode II, it had good selec-
interest A at the activity aA (but no B) and the other contains tivity toward primary lanthanum ion except in the presence
the interfering ion B at the same activity aB = aA (but no A). of Al(III), Ca(II) and Mg(II) ions, and it showed interfering
In this method, the values of the selectivity coefficient can be ion with those metal ions.
derived from the following equation:
3.6. Life time
A;B ¼ ½ðEB  EA Þ=S þ ð1  ðZB =ZA ÞÞ log aA
log Kpot
EA and EB mean the potentiometric response of aA and aB, Electrodes (I–III) were tested at different intervals to follow up
respectively. aA is the primary ion activity and aB is the activity its life time and reproducibility. It was clear from the figure
of an interfering ion. The single ion activities were calculated that modified SPE had the longest life time up to three weeks.
by the extended Debye-Hückel equation (Kamata et al., 1998). The surface of CPEs was continuously polished using a fil-
In this work, aA (1.0  103 mol L1 La(III) ion) and aB ter paper during calibration process to obtain new working
(1.0  103 mol L1 interfering ion) were used to measure surface and rinsed carefully with double distilled water to
the selectivity coefficients. S, Nernstian slope, ZA and ZB are remove the memory effect of electrode. This indicated that
the charge of the primary and interfering ions, respectively. the SPE had high mechanical durability and good adherent
The selectivity coefficient values are listed in Table 1. The value properties. Longer stability test was also investigated and the
of selectivity coefficient reached to 1.0 indicated equal response electrodes were successfully used for at least 50 consecutive
to both primary ion and interfering ions. If the value of measurements without any preconditioning before use.
Lanthanum(III) potentiometric sensors 395

1.0  102 mol L1, and the minimum was 1.0  106
Table 1 Potentiometric selectivity coefficient values of elec-
mol L1. The proposed Lanthanum ion-selective electrodes
trodes I, II and III using SSM.
(I-III) were used for potentiometric determination of lan-
SPE Electrode I CPE thanum(III) ion in different water samples (waste and tape
(2.5 mg EBA) water). As mentioned above in procedure (2.7) the samples
Electrode II Electrode III
Interfering Log KpotLa3+,B Log KpotLa3+,B Log KpotLa3+,B were treated and analyzed. The results obtained were also
Species compared with those from spectrophotometric analysis. The
Fe3+ 11 16 21 results obtained and listed in Table 2 indicated the successful
Mn2+ 9.7 10 5.7 use of the proposed electrodes (I-III) for determination of lan-
Sr2+ 9.61 9 5.8 thanum ion and there was a satisfactory agreement with those
Na+ 13.43 – – obtained by spectrophotometric method.
Ce3+ – 16 15
Ni2+ 9.07 10.1 4.7 3.8. Method validation
Co2+ 8.64 7.6 6.8
Pb2+ 10.78 9.1 7.6
3.8.1. Inter- and intra-day accuracy and precision
Zn2+ 8.40 2.4 5.4
Fe2+ 10.07 3.2 6 Different real water samples spiked with different concentra-
K+ 11.33 10 9 tions of La(III) ion were used to carry out four replicate exper-
Ba2+ 8.73 3.8 7.2 iments to evaluate the validity and applicability of the
Cd2+ 9.40 4.2 5.3 proposed method and reproducibility of the results obtained.
Cr3+ 9.40 6.05 8.6 Tables 3 and 4 show the values of the inter- and intra-day rel-
Mg2+ 7.97 2.5 5
ative standard deviations for different concentrations of the
Al3+ 7.62 2.7 7.7
samples, obtained from experiments carried out over a period
Cu2+ 10.07 4 10
Ca2+ 7.02 2.9 4.7 of four days (inter-day) or within the same day (intra-day). It
Zr4+ 10.01 – – was found that, modified SPE had reproducible response in the
V5+ 5.28 – – intra-day measurements rather than modified CPE which need
to scratch the paste surface to remove the memory effect and
obtain a new surface electrode and this is considered an advan-
tage of SPE over CPE. From the data obtained, the relative
3.7. Analytical application
standard deviations were found to be small indicating reason-
able repeatability of the proposed sensors, so modified CPEs
In general, the maximum La(III) ion concentration that and SPE were successfully applied to determine La(III) in pure
ion selective electrodes can measure was found to be and wastewater samples.

Table 2 Determination of lanthanum(III) ion in water samples using proposed sensors (I–III).
Sample Taken Spectrophotometry Electrode II Electrode III Electrode I
(mg mL1)
Found % Found % Found % Found %
(mg mL1) Recovery (mg mL1) Recovery (mg mL1) Recovery (mg mL1) Recovery
2 3.22 3.22 100.0 3.16 98.13% 3.14 97.52% 3.21 99.69%
2 1.611 1.67 103.6 1.59 98.69% 1.68 104.28% 1.611 100.0%
1 0.322 0.33 102.4 0.322 100.0% 0.322 100.0% 0.331 102.79%
1 1.611 1.66 103.0 1.58 98.08% 1.69 104.90% 1.585 98.39%

Table 3 Intra and inter days precision of the determination of La (III) in pure and spiked water samples using SPE (electrode I).
Taken, mg mL1 Found, mg mL1 Recovery % SD RSD %
Inter day Intra day Inter day Intra day Inter day Intra day Inter day Intra day
Pure 3.249 3.235 3.216 99.57 98.90 1.270 0.577 1.29 0.542
0.325 0.318 0.322 97.80 99.00 0.350 0.955 0.359 0.793
0.033 0.033 0.032 100.0 96.90 0.411 0.478 3.87 0.367
Water sample No. 1 0.322 0.322 0.331 100.0 102.8 0.478 0.478 4.012 1.239
1.611 1.600 1.585 99.32 98.39 0.500 0.500 5.714 0.836
Water sample No. 2 3.220 3.190 3.210 99.07 99.69 0.500 0.478 6.800 0.644
1.611 1.609 1.611 99.88 100.0 0.478 0.500 6.490 1.290
396 E.Y.Z. Frag et al.

Table 4 Inter -days precision of the determination of La (III) in pure and water samples using CPE (electrodes II and III).
Taken, Found, mg mL1 Recovery % SD RSD %
mg mL1
Electrode Electrode Electrode Electrode Electrode Electrode Electrode Electrode
II III II III II III II III
Pure 0.325 0.321 0.316 98.70 97.23 0.500 0.157 1.526 3.125
0.033 0.032 0.032 98.00 96.97 0.478 0.479 2.139 0.450
Water sample 0.322 0.317 0.322 98.44 100.0 0.816 0.525 2.148 0.416
No. 1 1.611 1.610 1.690 99.90 104.9 0.500 0.816 6.870 0.703
Water sample 3.220 3.160 3.270 98.14 101.55 0.478 0.816 7.509 0.595
No. 2 1.611 1.611 1.680 100.0 104.28 0.577 0.645 0.289 0.431

Table 5 Response characteristics of the investigated modified electrodes.


Parameters Electrode I Electrode II Electrode III
Slope, mV decade1 21.0 20.5 19.5
Linear range, mol L1 1.0  106–1.0  102 1.0  106–1.0  102 1.0  106–1.0  102
Limit of detection, mol L1 1.0  106 1.0  106 1.0  106
Limit of quantification, mol L1 3.3  106 3.3  106 3.3  106
Working pH range, 4–8 5–8 4–7
Life time, days 25 8 8
SD 0.879 0.288 0.894
RSD 0.998 0.962 0.999
Intercept 176 199.6 204
Recovery % 101 99.0 100
Isothermal coefficient V/°C 0.121  102 0.55 103 0.24502

3.8.2. Limits of detection and quantification method with high sensitivity, good selectivity, fast response,
The limit of quantification (LOQ) was determined by estab- and low detection limit with Nernstian behavior over a wide
lishing the least concentration that can be measured according concentration range.
to ICH Q2(R1) recommendations, below which the calibration
range was non linear. The results obtained are listed in Table 5. References
The limit of detection (LOD) was determined by evaluating the
lowest concentration of the measuring ions analytes that can Akl, M.A., Frag, E.Y.Z., Mohamed, G.G., Bashanaini, M.S.A., 2013.
be readily detected and were found to be 1.0  106 mol L1. Int. J. Electrochem. Sci. 8, 11546–11563.
The LOQ and LOD were calculated according to the following Al-Shawi, A.W., Dahl, R., 1996. Anal. Chim. Acta 333, 23–30.
Bakker, E., Pretsch, E., Buhlmann, P., 2000. Anal. Chem. 72, 1127.
equations (ICH 2005):
Besharati-Seidani, A., Shamsipur, M., 2015. Micro-Chim. Acta 182
LOQ ¼ 10 Sa =b (9–10), 1747–1755.
Bian, Y.J., Xu-Guang, Y., Hong-Wei, P., Ji-Tai, L., 2006. Synth.
LOD ¼ 3 Sa =b Commun. 36 (17), 2513–2518.
Chung, Ch., Brenner, I., You, Ch., 2009. <http://dx.doi.org/10.1016/j.
where (Sa) is the standard deviation of the intercept of the sab..06.013>.
regression line and (b) is the slope of the calibration curve. Dutt, N.K., Rahut, S., 1971. J. Inorg. Nucl. Chem. 33, 1725.
Dutt, N.K., Samyal, S., 1971. J. Inorg. Nucl. Chem. 33, 651.
3.8.3. Linearity Dutt, N.K., Sur, S., Rahut, S., 1971. J. Inorg. Nucl. Chem. 33, 121.
Theoretical Electrochemistry, 1972. Antropov L. I., Mir, Moscow.
The calibration graphs obtained by plotting the potential val-
Eric, B., Philippe, B., Erno, P., 1997. Carrier-Based Ion-Selective
ues versus the final concentration were found to be rectilinear
Electrodes and Bulk Optodes. 1. General Characteristics. 1997.
over the concentration range cited in Table 5. Chem. Rev. 97, 3083–3132.
Fanjul-Bolado, P., Hernández-Santos, D., Lamas-Ardisana, P.J.,
4. Conclusion Martı́n-Pernı́a, A., Costa-Garcı́a, A., 2008. Electrochim. Acta 53,
3635–3642.
Frag, E.Y.Z., Mohamed, G.G., Nour El-Dien, F.A., Mohamed, M.E.,
In this work, modified ion selective electrodes (carbon paste
2011a. Analyst 136, 332–339.
and screen printed) were fabricated based on EBA as Frag, E.Y.Z., Mohamed, A.M.K., Mohamed, G.G., Alrahmony, E.E.,
ionophore. The proposed electrodes possessed excellent perfor- 2011b. Int. J. Electrochem. Sci. 6, 3508–3524.
mance in determination of La(III) in tap water and wastewater Frag, E.Y.Z., Mohamed, G.G., El-Sayed, W.G., 2011c. Bioelectro-
with high recovery in comparison with spectrophotometric chemistry 82, 79–86.
Lanthanum(III) potentiometric sensors 397

Frag, E.Y.Z., Ali, T.A., Mohamed, G.G., Awad, Y.H.H., 2012. Int. J. Mohamed, G.G., Ali, T.A., El-Shahat, M.F., Al-Sabagh, A.M.,
Electrochem. Sci. 7, 4443. Migahed, M.A., Khaled, E., 2010. Anal. Chim. Acta. 673, 79.
Ganjali, M.R., Qomi, M., Daftari, A., Norouzi, P., Salavati-Niasari, Mohamed, G.G., Nour El-Dien, F.A., Frag, E.Y., Mohamed, M.E.B.,
M., Rabbani, M., 2004. Sensor Actuat B – Chem. 98 (1), 92–96. 2013. J. Pharm. Anal. 3, 367–375.
Gupta, V.K., Jain, S., Chandra, S., 2003. Anal. Chim. Acta 486 (2), Nour El-Dien, F.A., Mohamed, G.G., Frag, E.Y.Z., Mohamed, M.E.,
199–207. 2012. Int. J. Electrochem. Sci. 7, 10266–10281.
(a)
Islamnezhad, A., Zanjanchi, M.A., Shariati, S., Abri, A., 2011. Shamsipur, M., Yousefi, M., Hosseini, M., Ganjali, M.R., 2002. Anal.
Acta Chim. Slov 58, 46–52(b)Wilson, S.J., Peter, M.M., . Anal. Chem. 74 (21), 5538–5543.
Commun. 36, 31–33(c)Svehla, G., Slevin, P.J., . Talanta 15 (9), Singh, A., Singh, A., Singh, G., 2012. Asian J. Res. Chem. 5 (7).
978–982. Stites, J.G., McCarty, C.N., Quill, L.L., 1948. J. Am. Chem. Soc. 70,
Jia, Q., Kong, X., Zhou, W., Bi, L., 2008. Microchem. J. 89, 82–87. 3142.
Kamata, S., Bhale, A., Fukunaga, Y., Murata, A., 1998. Anal. Chem. Svancara, I., Kotzian, P., Bartos, M., Vytras, K., 2005. Electrochem.
60, 2464–2467. Commun. 7, 657–662.
Kawashima, T., Ogawa, H., Hamaguchi, H., 1961. Talanta 8 (7), 552– Tadjarodi, A., Jalalat, V., Zare-Dorabei, R., 2015. Mater. Res. Bull.
556. 61, 113–119.
Khalil, S., Abd El-Aliem, S., 2002. J. Pharm. Biomed. Anal. 27, 25–29. Waqar, F., Jan, S., Mohammad, B., Hakim, M., Alam, Sh., Yawar,
Khan, S.B., Rahman, M.M., Marwani, H.M., Asiri, A.M., Alamry, K. W., 2009. J. Chinese Chem. Soc. 56, 335–340.
A., 2014. J. Taiwan Inst. Chem. Eng. 45 (5), 2770–2776. Wilkes, C.E., Daniels, C.A., James, W.S., 2005. PVC Handbook, pp.
Lindner, E., Umezawa, Y., 2008. Pure Appl. Chem. 80, 85. 174–175. ISBN: 3-446-22714-8.
Mehrotra, R.C., Bohra, R., Gaur, D.P., 1978. Metal b-diketonates and Yuan, X., Chai, Y., Yuan, R., Zhao, Q., 2013. Anal. Chim. Acta 779,
Allied Derivatives. Academic Press, London. 35–40.
Moeller, T., Martin, D.F., Thompson, L.C., Ferrus, R., Feistal, G.R., Zhang, Y., Jiang, Z., He, M., Hu, B., 2007. Environ. Pollut. 148, 459–
Randall, W.J., 1961. Chem. Rev. 65, 1. 467.

S-ar putea să vă placă și