HANDYSCAN GETTING STARTED & QUICK SCAN REFERENCE

System Connection Part Preparation

Connect the scanner in this order: • Ideally, the part to be scanned will be
white with a matte finish. If the part is
Plug in the power supply shiny or dark in color, the resulting scan
to an outlet
Installation & Licensing may be less accurate and noisy.
• You may apply a foot powder or
Install VXelements by downloading the inspection spray to make the part easier
latest version from the Creaform Support Connect the power to scan.
website. supply to the USB cable
• On parts where the residue from the
http://support.creaform3d.com. powder is undesirable, cyclododecane
may be useful as it should evaporate.
Plug the USB cable
You will need the CD key for the scanner into a USB 3.0 port • If you need to use a powder, apply it
and any modules that you own during this before the part is targeted.
process. • The part should be stable during the scan
Connect the scanner to and you may need to target the
After VXelements is installed, you will need environment for a cleaner and more
to activate the scanner in the Product the USB cable
accurate scan.
Manager

Start Menu > All Programs > VXelements >


Product Manager

Select the scanner, hit activate, and download


Target Application
• Randomly place targets on flat areas
the version specific license and calibration as much as possible
files. • Place targets in nonlinear patterns
• Targets on curved surfaces are acceptable
If you do not have internet access, use the if target is not deformed • Maximize the number of triangular shapes
provided USB stick in the scanner case to • Minimium distance between targets: 0.75” between targets
install the version of the software that came (2cm) • Place targets 0.25” (6mm) from
with the scanner. The license and calibration • Maximum distance between targets: edges or small features
files for the scanner are also on the USB 4” (10cm) • Avoid damaged or unclean targets
stick.

NEED HELP?
ASK JOE
kb.goengineer.com
888.559.6167
HANDYSCAN GETTING STARTED & QUICK SCAN REFERENCE

Scanning Techniques Scan Positioning Targets

• Minimum of 4 targets must been seen by 1 Change the scanning mode to Scan
the scanner at all times Positioning Targets by using the arrow
• Stand-off distance of the scanner from the next to Scan
Scanning Process part should be 12” (30cm). Verify distance 2 Click on Scan
using the meter in VX elements or the 3 Scan the positioning targets. Start in the
light on the scanner center of the part and use a spider web
1 Start VXelements
• Stay as perpendicular to the surface pattern. Make sure that your model
2 Check Calibration of the part as you can and never closes nicely
3 Check Configuration deviate more than 45° 4 Stop the scan and save the
4 Define desired resolution • Use adequate speed when moving the positioning targets if needed:
In the product tree, click on Scan. Change the scanner. Higher resolution = slower speed File > Export > Positioning Targets
resolution under scan parameters depending on 5 Change the scan mode back to
the level of detail that you need for the part. As a
Scan Surface
rule of thumb, the reslolution should be at least 3x
smaller than the smallest radius on the part. Once
the resolution has been entered, click Apply.
5 Select the appropiate target type
In the product tree, click on Positioning. Change the
target type under Positioning Parameters.
Optional: Scan Postioning Targets Calibration Configuration
6 Click on Scan
7 Scan the part 1 Get the calibration plate from the scanner 1 Open the configuration module:
See Scanning Techniques case located under the protective foam Configure > Scanner > Configuration
*To stop the scanning process, click on Scan insert 2 Click on Auto-Adjust
8 Check the result 2 Open the plate’s lid and place on a 3 Push the main button on the scanner to
Make sure all needed surfaces are present flat table away from other targets start the process and move the scanner
9 Redo steps 4-8 as needed 3 Open the configuration module: to a large flat area on the part. Make sure
10 Optimize the Positioning Targets Configure > Scanner > Calibration you are in the recommended range of
Optimize Positioning Model located under 4 Push the main button on the scanner to 8”-12” (20-30cm) from the part
Positioning Parameters in the product tree start the process and move the scanner 4 Wait until the scanner parameters are
11 Save the session file as indicated on the screen to gather the adjusted
File > Save Session 14 readings needed for the calibration 5 Push the main button on the scanner
12 Save the mesh file process.
File > Save Mesh
6 Click Apply
5 Once the 14 readings have been aquired,
Optional: Save the positioning targets click OK. The scanner is now calibrated
File > Save Positioning Targets