Facility Growth
The use of small-scale bioreactors in development and commercial
manufacturing facilities has grown from just a few to over several hundred instruments,
Similarly, correlation coefficients were greater than 0.97 for both glutamine and
glutamate. In addition, the instrument exhibited a high level of comparability with a
standard deviation of 0.13 and 0.07 mmol/L and a coefficient of variability of 6.0% and
5.2% for glutamine and glutamate, respectively.
It should be noted that the linearity and variability for glutamine are not as strong
as the other parameters tested. Though the samples were controlled in the best
possible manner (samples stored on ice prior to analysis), the instability of glutamine
often results in higher variability in large experiments such as this.
Figure 2 illustrates the linearity profiles for ammonium NH4+, Na+, and K+. NH4+,
Na+, and K+ also exhibited very strong linearity correlation coefficients of greater than
0.99 for all three analytes. Standard deviation was 0.09, 2.23 and 0.09 mmol/L and
coefficient of variability was 5.3%, 2.5% and 2.8% for NH4+, Na+ and K+, respectively.
Figure 2 also includes the total cell density linearity profile for the BioProfile FLEX
analyzers. In the figure, results obtained on the BioProfile CDV module are compared to
results obtained on the Cedex cell counter, the reference instrument used in this
laboratory.
The data was generated from bioreactor cultures seeded at varying densities,
providing a range of total cell density and viability samples.
Due to intellectual property concerns, the viability data is not shown. However, a
range of cultures were analyzed with viabilities ranging from approximately 45% to 99%
viability. BioProfile FLEX viabilities were within 3% compared to the Cedex analyzers.
Nova Biomedical provides published specifications for all parameters (excluding
CDV) as measured on a single analyzer for both within run and day-to-day precision.
This study incorporated nine BioProfile instruments, with samples analyzed over the
course of several weeks.
The results reported in this study demonstrate that not only can one expect a
single BioProfile instrument to meet published specifications for precision, but multiple
instruments can also fall well within the published specifications of a single instrument.
This data further supports the BioProfile analyzers as extremely robust analytical tools
for cell culture development and manufacturing process monitoring. GEN