Safe use of return electrodes
CONTENTS
INTRODUCTION03
CHECKLIST07
U▻ Important information
02
Electrosurgery is the surgical technique most frequently deployed in the
operating room. In almost every operating room, throughout all disciplines,
an electrosurgery unit is used.
03
The current flow in
monopolar electrosurgery
HF
HF
NE
01 02
Closed electric circuit in monopolar electrosurgery. ↑ This orientation reduces the leading edge effect
Current distribution over the entire surface of the return electrode
with lower current density ↓ Higher leading edge effect with asymmetrical orientation of the return electrode
Current with a high current density is emitted at the active electrode in A higher current density arises on the side of the return electrode closer
order to create an electrosurgical cut or coagulation. At the exit site, the to surgical site, as the current seeks the shortest path with the lowest
return electrode, no appreciable thermal effects arise if the current is resistance. The resulting uneven temperature distribution on the return
conducted over the entire large-surface return electrode at a low current electrode is also known as the “leading edge” effect.
density.
If the short side of the return electrode faces the operating field, the cur-
rent density is higher and the heat distribution is more inhomogeneous.
For this reason, the return electrode should always be oriented with the
longest side towards the operating field. This long edge orientation can
reduce heat peaks (Fig. 02↑and↓).
04
Integrated
safety system
Our safety tip:
Avoid the leading edge problem
by using a NESSY Ω® return electrode.
03 04
Even distribution of current density and heat generation on both surfaces ↑ Red traffic light: Activation is stopped if the return electrode
of the return electrode has not been applied correctly
NESSY® stands for Neutral Electrode Safety System (neutral electrode Activation is possible with a low deviation. In the event of a large dif-
= return electrode). NESSY®, integrated in the VIO® series, checks ference, there is the risk of thermal tissue necrosis.
whether the split-pad return electrode is correctly applied over the
entire area. The current flowing through the two contact surfaces are NESSY® stops the activation and outputs a warning signal (display: red
continuously equalized. If the return electrode is not applied properly traffic light ↑). In order to avoid thermal necrosis, the electrosurgery
there is a current asymmetry. The risk: partial thermal peaks due to function can only be reactivated once the application of the return
increased current density. electrode has been corrected.
05
The NESSY Ω ®
safety concept
01
Thermal comparison between a conventional return electrode and the NESSY Ω® NESSY Ω® with the outer, uncontacted equipotential ring
(on the right): homogeneous current distribution (application on the thigh r. and l.)
06
Information
on safe use of electrosurgery
Select a suitable return electrode Attach the return electrode free of creases The return electrode can be placed under the
antithrombosis stocking
1. PLACE THE PATIENT IN AN ELEC- 3. SELECTING THE POSITION FOR 4. PREPARING THE APPLICATION SIDE
TRICALLY INSULATED POSITION THE RETURN ELECTRODE
• Do not attach the return electrode to
• Place the patient on a dry and electri- • is possible to position the return elec-
It areas with hair. Shave the return elect-
cally insulated OR table pad. trode on the patient’s thigh, upper arm rode application site beforehand.
• Ensure that the arm support cover is or flank. • The return electrode application site
insulated. • Attach the return electrode as close as must be dry and free of skin oil.
• Remove any jewelry (piercings, rings, possible to the operating field while
chains, watches, bracelets) and/or re- maintaining a minimum distance of 15 cm.
movable dental prostheses; taping over • T he monopolar current should not pass 5. A
TTACHING THE RETURN
jewelry does not suffice. through constricted areas of the body ELECTRODE CORRECTLY
• Bend the patient’s arms and legs or (e.g. elbows, knees).
wrap them in towels so as to prevent • P osition the return electrode above • not trim the return electrode.
Do
contact with the body; avoid skin-to- electrically conducting tissue (muscle • Always attach oblong return electrodes
skin contact in areas of loose skin or tissue) whenever possible. with the long side directed towards the
breast folds (by covering the areas with • Do not place the return electrode on top operating field.
dry surgical gauze). of fat tissue, bones/joints, skin folds or • Make sure the entire surface of the
• The patient must not touch any electri- on the patient’s head. return electrode is attached to the
cally conducting objects (drip stands, • Attach the return electrode to healthy patient’s skin and is wrinkle-free; avoid
hoses). tissue if possible. Avoid scars, hemor- air bubbles.
rhages, tattoos. • For patients wearing antithrombosis
• T he patient should not lie on top of the stockings, the return electrode can be
2. S
ELECTING A SUITABLE RETURN return electrode, cables or the cable attached underneath one of them with
ELECTRODE connector. the connector and cable exposed.
• W hen lifting and moving the patient • Self-adhesive return electrodes can only
• elf-adhesive split return electrodes are
S make sure the return electrode and the be used once.
preferable to single return electrodes cable do not detach and the patient is
and silicone electrodes. not laid down on top of them.
• A suitable return electrode has to be 6. AVOIDING IGNITION OF
used for infants. Patients with active and passive implants: COMBUSTIBLE SUBSTANCES
• We recommend NESSY Ω®, which can be • In cases of patients with a cardiac pace-
placed without orientation towards the maker or anyother kind of conductive • Disinfectants must not flow underneath
operation field. implant use bipolar instruments where- the patient.
• Please only use split return electrodes if ver possible. • Wait until the disinfectant has been
possible, as only these can be monitored • When monopolar instruments are used, dried off and the combustible gas has
by the safety system. attach the return electrode far enough evaporated before covering the patient.
away from the implant to prevent the • Avoid inflammable and oxidizing gases
current path of the return electrode in the surgical field (e.g. anesthetic or
from passing over the implant. Minimize endogenous gases).
CAUTION: This checklist does not replace
the effect number (voltage) and output
the instructions for use.
limitation.