HEALTHCARE Enclosure

Imaging Services

Document No: DD+DIS127.11E


This number is needed for production line only:
A603739.0 CR 30-X 2nd
Type 5175 / 200 / 205 / 220 / 225

Installation Instructions for


Digitizer Software CIR_1013* and higher Versions
*CIR_1013 = Version number of the software.
Always use the installation instructions delivered with the Software.
Download the latest published software from:
Agfa HealthCare Library => Computed Radiography => CR Digitizer => CR 30-X 2nd => Software

► Purpose of this Document

This enclosure provides the installation instructions for:


 Software upgrade
 Software installation on a new CF Card

► Referenced Documents

Document Title
Not applicable

NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).

CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0
07-2010 printed in Germany Document Node ID: 33663293
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2011 Agfa HealthCare N.V.
DD+DIS127.11E Enclosure

► Manufacturer
Agfa HealthCare N.V.

Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright  2011 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
Read the "Generic Safety Directions" document
(see Agfa HealthCare Library => General Info => Agfa HealthCare => Publications
=> Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
Strictly observe all safety directions within the "Generic Safety Directions" and on the
product.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 2 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

LIST OF CONTENTS

1 INTRODUCTION ......................................................................................................................4
2 SOFTWARE UPGRADE VIA USB STICK................................................................................5
2.1 Checking current Software Version ..........................................................................................5

2.2 Taking out the Digitizer USB Stick ............................................................................................6

2.3 Preparing the USB Stick ...........................................................................................................7

2.4 Performing the Software Installation.........................................................................................8

2.5 Creating a Backup ....................................................................................................................9

2.6 Verifying the Software Upgrade ................................................................................................9

3 APPENDIX..............................................................................................................................10
3.1 Software Upgrade via Browser ...............................................................................................10

3.2 Software Installation on a new CF Card .................................................................................11

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 3 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

Subject Software Installation Instructions for Software CIR_1013 and higher Versions

SCOPE OF DELIVERY:
ZIP File CIR_1013*
* or higher version

REQUIRED TOOLS:
 Service PC with USB port
 USB Stick with at least 10 MB free disk space.

Recommendation: As the digitizer is not able to read all kinds of USB sticks, use
the USB stick delivered with the digitizer.
The USB stick including software can also be ordered via order number*:
CM+6027410

* The last digit in the spare part number indicates the spare part revision at release of this
document. When ordering, the actual revision of the spare part is delivered.

REQUIRED TIME:
Approximately 15 minutes for the installation via USB Stick

NOTE:
Register a software upgrade in the local service management system as modification,
e.g. in SMS as “MOD”.

1 Introduction

This document provides following instructions:

# Type of software installation When to be used Section


1 Software upgrade via USB Stick Recommended proceeding 2
for a software upgrade.
2 Software upgrade via browser Exceptional proceeding, if 3.1
e.g. no USB stick is available.
3 Software installation on a new Software installation including 3.2
CF card formatting after CF card
replacement.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 4 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

2 Software Upgrade via USB Stick

The software upgrade via USB stick is performed in following steps:

# Type of software installation Section


1 Check current software version. 2.1
2 Take out USB stick. 2.2
3 Prepare the USB stick. 2.3
4 Perform the software upgrade. 2.4
5 Create a backup. 2.5
6 Verify the software upgrade. 2.6

2.1 Checking current Software Version

(1) Open the browser on the NX workstation.


(2) Enter the IP address of the digitizer in the address window.

NOTE:
If the IP address is not known: Press 'Ctrl + Shift + H' to get the browser history list.
For Internet Explorer Version 6 press the "History" button in the toolbar.
The IP address of the digitizer should be in the list.

(3) Enter username and password.


(4) Check the displayed software
version to ensure that the software
to be installed is higher than the
displayed software.

Figure 1

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 5 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

2.2 Taking out the Digitizer USB Stick

(1) Press both buttons at the feed A


table (A).
(2) Pull out the cassette unit (B).

B
5175_c35_00A.cdr

Figure 2

CAUTION:
Parts underneath the top cover can be damaged.
Carefully remove the top cover. Follow the instructions as shown in next step.

(3) Remove two screws (A).


(4) Take off the top cover (B, C).

5175_c35_004.cdr

Figure 3

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 6 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

(5) Remove screw (A).


(6) Take out the USB stick (B). B

5175_c35_backup1.cd

Figure 4

(7) Mount the Top Cover.


Do not mount the screws yet. A

5175_c35_404.cdr

Figure 5

2.3 Preparing the USB Stick

(1) Extract all files from the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
 Software package CIR_1013.zip*
 This document
 Depending on the software version a readme.txt

(2) Extract all files from CIR_1013.zip*.


Folder CIR_1013* is created automatically.
* or higher version

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 7 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

(3) Copy the folder “CIR_1013*“ and its contents to the


USB Memory Stick root directory.
(4) Go to folder “CIR_1013*” on the USB Memory Stick.
* or higher version

(5) Double click on: <Prepare_Installation.cmd>


 A DOS Box opens for a short time.
 The file startup.com is written to the root of the
USB stick
(6) Release the USB Memory Stick from the Service PC. 5175_ENC_401.cdr

Figure 6

2.4 Performing the Software Installation

IMPORTANT:
Do not switch off the digitizer during software installation as the Cassette Handling
Board is re-programmed during the software installation procedure.

(1) Pull out the cassette unit.


(2) Plug the USB stick into the
USB slot of the digitizer.
See Figure 7.
(3) Close the cassette unit.
(4) Switch on the digitizer. The
digitizer boots from USB stick.
(5) Wait until the digitizer LED is
green. This takes
approximately 3 minutes.
See also NOTE below.
5175_c35_094a.cdr

Figure 7

NOTE:
The digitizer LED activity sequence during software installation is as follows:
 Red blinking: Approx. 0,5 minutes
 Off: Approx. 0,5 minutes
 Red blinking: Approx. 2 minutes
 Green: After approx. 3 minutes

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 8 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

2.5 Creating a Backup

(1) In service menu <Backup>


select <Complete backup>
and <SAVE TO USB>.
(2) Unplug the USB stick when
the message is displayed:
Remove USB stick.
(3) Put the USB stick back in its
cover and mount the screw.
(4) Mount the top cover.
(5) Close the cassette unit.
Figure 8

2.6 Verifying the Software Upgrade

Part 1: Checking the software version in the service user interface:


(1) In the digitizer service user interface go to menu: <Device Info>
(2) Check, that the installed software version is displayed.
(3) Select menu <Logout> and close the browser.

Part 2: Functional check:


(1) Insert cassette with unexposed IP in the digitizer.
(2) At the NX workstation select examination type “System diagnosis – Flatfield”.
(3) Select “ID” to identify the image: Scanning starts.
(4) Confirm that the image arrives at the NX workstation.
(5) Reject the image.
(6) Log-off user crservice.

Result: The digitizer software is updated.

NOTE:
For new features and solved bugs in this software refer to the digitizer chapter 3.8,
Software Releases, Patches, of the digitizer service manual.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 9 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

3 Appendix

This section provides following instructions:

# Type of software installation Purpose Section


1 Software Upgrade via Browser Software upgrade 3.1
2 Software installation on a new Software installation including 3.2
CF card formatting after CF card
replacement

3.1 Software Upgrade via Browser

NOTE:
Software upgrade via browser can be performed from any PC in the network.
It is tested with Firefox 3 and 4 and Internet Explorer 6, 7 and 8.

(1) Extract all files from the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
 Software package CIR_1013.zip*
 This document
 Depending on the software version a readme.txt
(2) Copy the software ZIP file CIR_1013.zip* from the Service PC to the PC in the
network.
(3) Open the browser.
(4) Enter the IP address of the digitizer in the address window.
* or higher version

NOTE:
If the IP address is not known: Press 'Ctrl + Shift + H' to get the browser history list.
For Internet Explorer Version 6 press the "History" button in the toolbar.
The IP address of the digitizer should be in the list.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 10 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

(5) Enter username and


password.
(6) Check the current software
version to ensure that the
software to be installed is
higher than the displayed
software version.
(7) Select menu: <Software
Installation> Figure 9

(8) Browse for the software ZIP file.


(9) Select: <Upload and Install Software>
(10) Wait until the digitizer reboots automatically and the digitizer LED gets green.
This takes approximately 3 minutes.
(11) Take out the digitizer USB stick from its cover. See section 2.2.
(12) Create a backup. See section 2.5.
(13) Verify the software upgrade. See section 2.6.

3.2 Software Installation on a new CF Card

(1) Take out the digitizer USB stick. See section 2.2.
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
 Software package CIR_1013.zip*
 This document
 Depending on the software version a readme.txt

(3) Extract all files from CIR_1013.zip*.


Folder CIR_1013* is created automatically.
* or higher version

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 11 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

(4) Copy the folder “CIR_1013*“ and its contents to the


USB Memory Stick root directory.
(5) Go to folder “CIR_1013” on the USB Memory Stick.
* or higher version

(6) Double click on: <Prepare_Formatting.cmd>


 A DOS Box opens for a short time.
 The file startup.com is written to the root of the
USB stick
(7) Release the USB Memory Stick from the Service PC. 5175_ENC_401.cdr

Figure 10

(8) Perform the software installation. See section 2.4

NOTE:
After software installation on a new CF card the digitizer LED will usually not get
green. Reason: The destination information is not available anymore. This is only
available after restoring the latest backup.

(9) In service menu <Restore> select <Restore from


USB> to restore the latest backup. See also the
IMPORTANT note below.

IMPORTANT:
 The backup is a file with following syntax:
Backup_<DigitizerSN>_<YYYYMMDD>.zip
If the backup is not available, following data are lost:
o Digitizer configuration, except digitizer network settings. The network settings
are stored in an EPROM on the CPMI board.
o Logfiles
o Infocounter
In this case it is required to perform the digitizer configuration. For details refer to
chapter 1, Controls, Connections and Set-up Procedure, of the digitizer service
manual.
 IF CF card and CPMI board are replaced simultaneously, the digitizer network
settings are lost. A restore is not possible in this case, as the browser connection
is not available anymore. To solve this issue the network settings have to be set to
factory default. For instructions refer to chapter 3.3, Troubleshooting, of the
digitizer service manual.

DOCUMENT CONTROL NOTE:


The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.

Revision 0 CR 30-X 2nd Page 12 of 13


07-2010 Type 5175 / 200 / 205 / 220 / 225 Agfa Company Confidential
DD+DIS127.11E Enclosure

(10) Unplug the USB stick when the message is displayed: Remove USB stick.
(11) Put the USB stick back in its cover and mount the screw.
(12) Mount the top cover.
(13) Close the cassette unit.
(14) Verify the software installation. See section 2.6.