Documente Academic
Documente Profesional
Documente Cultură
Imaging Services
► Referenced Documents
Document Title
Not applicable
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
07-2010 printed in Germany Document Node ID: 33663293
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2011 Agfa HealthCare N.V.
DD+DIS127.11E Enclosure
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2011 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
Read the "Generic Safety Directions" document
(see Agfa HealthCare Library => General Info => Agfa HealthCare => Publications
=> Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
Strictly observe all safety directions within the "Generic Safety Directions" and on the
product.
LIST OF CONTENTS
1 INTRODUCTION ......................................................................................................................4
2 SOFTWARE UPGRADE VIA USB STICK................................................................................5
2.1 Checking current Software Version ..........................................................................................5
3 APPENDIX..............................................................................................................................10
3.1 Software Upgrade via Browser ...............................................................................................10
Subject Software Installation Instructions for Software CIR_1013 and higher Versions
SCOPE OF DELIVERY:
ZIP File CIR_1013*
* or higher version
REQUIRED TOOLS:
Service PC with USB port
USB Stick with at least 10 MB free disk space.
Recommendation: As the digitizer is not able to read all kinds of USB sticks, use
the USB stick delivered with the digitizer.
The USB stick including software can also be ordered via order number*:
CM+6027410
* The last digit in the spare part number indicates the spare part revision at release of this
document. When ordering, the actual revision of the spare part is delivered.
REQUIRED TIME:
Approximately 15 minutes for the installation via USB Stick
NOTE:
Register a software upgrade in the local service management system as modification,
e.g. in SMS as “MOD”.
1 Introduction
NOTE:
If the IP address is not known: Press 'Ctrl + Shift + H' to get the browser history list.
For Internet Explorer Version 6 press the "History" button in the toolbar.
The IP address of the digitizer should be in the list.
Figure 1
B
5175_c35_00A.cdr
Figure 2
CAUTION:
Parts underneath the top cover can be damaged.
Carefully remove the top cover. Follow the instructions as shown in next step.
5175_c35_004.cdr
Figure 3
5175_c35_backup1.cd
Figure 4
5175_c35_404.cdr
Figure 5
(1) Extract all files from the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
Software package CIR_1013.zip*
This document
Depending on the software version a readme.txt
Figure 6
IMPORTANT:
Do not switch off the digitizer during software installation as the Cassette Handling
Board is re-programmed during the software installation procedure.
Figure 7
NOTE:
The digitizer LED activity sequence during software installation is as follows:
Red blinking: Approx. 0,5 minutes
Off: Approx. 0,5 minutes
Red blinking: Approx. 2 minutes
Green: After approx. 3 minutes
NOTE:
For new features and solved bugs in this software refer to the digitizer chapter 3.8,
Software Releases, Patches, of the digitizer service manual.
3 Appendix
NOTE:
Software upgrade via browser can be performed from any PC in the network.
It is tested with Firefox 3 and 4 and Internet Explorer 6, 7 and 8.
(1) Extract all files from the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
Software package CIR_1013.zip*
This document
Depending on the software version a readme.txt
(2) Copy the software ZIP file CIR_1013.zip* from the Service PC to the PC in the
network.
(3) Open the browser.
(4) Enter the IP address of the digitizer in the address window.
* or higher version
NOTE:
If the IP address is not known: Press 'Ctrl + Shift + H' to get the browser history list.
For Internet Explorer Version 6 press the "History" button in the toolbar.
The IP address of the digitizer should be in the list.
(1) Take out the digitizer USB stick. See section 2.2.
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
Software package CIR_1013.zip*
This document
Depending on the software version a readme.txt
Figure 10
NOTE:
After software installation on a new CF card the digitizer LED will usually not get
green. Reason: The destination information is not available anymore. This is only
available after restoring the latest backup.
IMPORTANT:
The backup is a file with following syntax:
Backup_<DigitizerSN>_<YYYYMMDD>.zip
If the backup is not available, following data are lost:
o Digitizer configuration, except digitizer network settings. The network settings
are stored in an EPROM on the CPMI board.
o Logfiles
o Infocounter
In this case it is required to perform the digitizer configuration. For details refer to
chapter 1, Controls, Connections and Set-up Procedure, of the digitizer service
manual.
IF CF card and CPMI board are replaced simultaneously, the digitizer network
settings are lost. A restore is not possible in this case, as the browser connection
is not available anymore. To solve this issue the network settings have to be set to
factory default. For instructions refer to chapter 3.3, Troubleshooting, of the
digitizer service manual.
(10) Unplug the USB stick when the message is displayed: Remove USB stick.
(11) Put the USB stick back in its cover and mount the screw.
(12) Mount the top cover.
(13) Close the cassette unit.
(14) Verify the software installation. See section 2.6.