Documente Academic
Documente Profesional
Documente Cultură
The RF attenuator is one of devices to calibrate receivers such as spectrum analyzers to evaluate
qualities of radio waves which are radiated from RF transmitters. The linearity, one of the receiver
performance indices to show adequately the signal in accordance with the received power, can be
evaluated using the RF attenuator, which can reduce the intensity of radio waves transmitting
through a coaxial cable or a waveguide. The RF attenuator is also used to evaluate the RF output
level of the signal generator. NICT performs calibration service of the RF attenuators as a device
for calibration of receivers and signal generators and so on. This report describes the calibration
method of RF attenuators of which operating frequency is from 10 MHz up to 18 GHz and the
estimation method of measurement uncertainty.
Spectrum analyzer
57
2 Research and Development of Calibration Technology
Receiver
Signal generator
P0
G 0 L 0
(a)
Receiver
Signal generator Attenuator
P1
A
G 0 L 0
(b)
FFig. 2 Definition of RF attenuation
Step attenuator
ATT
Switch driver
SW 1 SW 2 SW 3 SW 4
Step Attenuator
10 dB 20 dB 20 dB 40 dB
Also, it is usually expressed using dB. market, called “variable attenuators”; these have the conve-
P nience of providing multiple attenuation amounts without
AdB 10 log10 0 [dB] (2)
P1 0 the need to remove/attach/replace attenuators, so they are
G L
58 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)
2-4 Calibration of RF Attenuators
Receiver
Signal generator Step attenuator
ATT
Pb
Ab
G 0 L 0
(a)
Receiver
Signal generator Step attenuator
ATT
Pe
Ae
G 0 L 0
(b)
FFig. 4 Definition of incremental attenuation
Receiver
ratio of (a) reception power Pb before attenuation is changed obtain accurate values. As described later, this problem is
(attenuation Ab), divided by (b) reception power Pe after treated as “uncertainty” associated with calibration results.
attenuation is changed (attenuation Ae) (here, Ae > Ab). This
attenuation ratio is provided by the following equation, 3 Calibration method
which is called “incremental attenuation” to distinguish it
from the usual attenuation. Here, we describe calibration of a step attenuator used
A P when validating linearity of a receiver (determination of
Ai e b (3)
Ab 0 Pe 0 incremental attenuation), as shown in Fig. 1. Calibration
G L G L
59
2 Research and Development of Calibration Technology
for converting the input signal to 30 MHz. To suppress When the calibration system is confirmed as operating
reflection, fixed attenuators with 6 dB attenuation (Weinchel normally, then the calibration procedure is as follows[2].
44–6) are attached to the ends of the two coaxial cables (1) Connect the device being calibrated (“device under
that connect to the input/output terminals of the step at- test (DUT)”) to the coaxial cables to which the fixed
tenuator. attenuators are attached, with a 0 dB setting for the
10 MHz to 18 GHz frequency signals output from the DUT (Step 1). At this time, read the AbdB value
signal generator pass through the step attenuator, which is displayed on the receiving instrument.
between the fixed attenuators connected by coaxial cables. (2) Set the DUT to the attenuation you want to calibrate
The signal that passes through is input into the receiving (for example, 30 dB), and read the AedB value dis-
instrument. Specifically, the signal passes through a fixed played on the receiving instrument again.
attenuator with 3 dB attenuation (Weinchel 44–3) inserted (3) Use the following equation to determine incremen-
to prevent saturation and for impedance matching, then it tal attenuation.
is input into the 8852 and converted to 30 MHz frequency, dB dB dB
Ai Ae Ab [dB] (5)
and the VM-7 measures the attenuation as a 30 MHz fre-
quency signal. This method of converting frequency and In the actual calibration work, operate the receiving
using the change in signal strength of the intermediate instrument so the value measured in (1) is displayed as 0
frequency (now 30 MHz) to calibrate attenuation of an dB, and arrange so the value measured and displayed in
attenuator is called the “Intermediate Frequency Substitution (2) becomes the incremental attenuation, then measure.
Method”[1]. Repeat the measurement five times to reduce the effects
Validity of the attenuation value shown on the receiving due to reproducibility of switching in the step attenuator.
instrument is confirmed by measuring a step attenuator
called a STD, calibrated by a superior calibration organiza- 4 Calibration results
tion. NICT ensures traceability to the national standard by
the traceability system shown in Fig. 6. The STD is cali- Table 1 shows a list of calibration points that NICT
brated periodically by the national standard organization,
so the accuracy of the calibration system is assured.
TTable 1 Calibration range (10 MHz to 18 GHz)
(a)
60 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)
2-4 Calibration of RF Attenuators
TTable 2 Results of NICT’s validation of attenuation calibration system (for 30 dB, 60 dB, 90 dB)
61
2 Research and Development of Calibration Technology
62 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)
2-4 Calibration of RF Attenuators
M idB
20 1 S11e G 1 S 22e L S 21e S12e G L occurs at 18 GHz in Table 2.
ln
ln 10 1 S11b G 1 S 22 b L S 21b S12 b G L
8.686 ln
1 S11e G 1 S 22e L S 21e S12e G L [dB] (5) Changes Due to Differences vs. Measurement
1 S11b G 1 S 22 b L S 21b S12 b G L Environment of the Superior Calibration Organization
(9)
For the step attenuator used as the STD, if there
Here, when x is sufficiently smaller than 1, using the are changes in the temperature of the environment
fact that the approximation ln(1+x)≈x is true, we can ex- used, then resistance values of the internal resistance
press this as parts can change and attenuation can vary. It is
M 8 .686 S S [dB] 2 2
documented that calibration of the STD is to be
dB 2 2 2 2 2 2 2
i S S
G S S
11 b 11 e L 22 b 22 e G L 21 b 21 e
63
2 Research and Development of Calibration Technology
90
80 measured points and the regression line. The
70 maximum differences obtained at nine points from
60 10 dB to 90 dB are recorded as causes of uncertain-
50 ties. At 1 GHz, the difference is a maximum 0.014
40 dB when incremental attenuation is 90 dB, so 0.014
30 dB is recorded as a cause of uncertainty. On the
20 other hand, at 10 MHz, differences of measured
10 MHz
10 1 GHz values rapidly increase at 80 dB and 90 dB.
0 Therefore, at 10 MHz frequency, looking at perfor-
0 10 20 30 40 50 60 70 80 90 100 mance of uncertainty evaluations for 10 dB to 70
Attenuation measured by NMIJ [dB] dB separately from those for 80 dB and 90 dB, the
(a)
maximum difference from the regression line ob-
0.25 tained from seven points 10 dB to 70 dB, and the
0.2 10 MHz differences between measured values and calibrated
Difference [dB]
1 GHz
0.15 values from the superior calibration organization at
0.1
80 dB and 90 dB, are each recorded as uncertainty.
0.05
The trend at 30 MHz was similar to that at 10 MHz,
0
-0.05
so we assessed uncertainty for 10 dB to 70 dB
-0.1 separately from uncertainty for 80 dB and 90 dB.
0 10 20 30 40 50 60 70 80 90 100 Differences from the regression line can occur
Attenuation [dB] because of the saturation of the frequency conver-
(b)
sion device, internal noise and signal leaks; the
FFig. 10 Linearity of receiving instrument
numbers of this uncertainty can include the effects
of noise and leaks. One may think that noise and
(6) Changes over Time leaks are usually recorded separately as causes of
We validate the degree that the incremental uncertainty of the receiving instrument, but they
attenuation of the STD changes with time. We can be recorded in duplicate, so we do not record
conducted a study using results from calibration them separately.
system during the past nine years. We did not find Deviations from this regression line could be
certain expected trends such as a gradual increase “systematic bias” that always occurs, so there is no
in attenuation, but we decided to record the maxi- probability density distribution. Therefore, we re-
mum value of change amount in 1 year as uncer- cord the same values obtained as standard uncer-
tainty. Assuming that the chance of the true value tainty[6].
existing is uniform across this range, the probabil-
ity distribution is treated as a uniform distribution. 5.2 Uncertainty when a device under test is
(7) Linearity of receiving instrument calibrated
We verify linearity of a receiving instrument by Uncertainty can be caused by the following three items
using an STD and comparing values measured by when calibrating a step attenuator of the DUT using a
the receiving instrument with the values specified calibration system that was calibrated using an STD.
in calibration certification of STD. Figure 10 shows (8) Resolution of a Digital Indication
an example of linearity at 10 MHz and 1 GHz. In Same as (2)
Figure 10 (a), the vertical axis is the measurement (9) Repeatability of Measurements (5 times)
value of incremental attenuation by the calibration Same as (3). However, measurements are taken
system, and the horizontal axis is the calibration 5 times, so experimental standard deviation divided
value at the superior calibration organization (value by the square root of 5 is recorded as the standard
specified in the calibration certification); the ▲ uncertainty of the average value.
64 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)
2-4 Calibration of RF Attenuators
TTable 4 Increase in uncertainty due to frequency ranges extended We explained the calibration techniques that NICT
by interpolation uses, and the way of thinking about uncertainty associated
Extended with calibration results, for an RF attenuator needed to
Extended Expanded
Uncertainty validate performance and calibrate a receiver and signal
Frequency Attenuation Uncertainty
for Frequency generator used to test wireless equipment. By using the
Range by NMIJ
range extension
uncertainty budget table, and closely investigating causes
Over 10 MHz of greater uncertainty, we showed that reflection in the
0.020 dB 0.030 dB
to under 1 GHz 10, 20, 30,
40, 50, 60, calibration system and linearity of the receiving instrument
Over 1 GHz 70, 80, 90 dB are causes that increase uncertainty. To solve this problem,
0.020 dB 0.044 dB
to under 10 GHz
calibration precision could be improved by using an im-
Over 10 GHz 10, 20, 30, 0.010 dB pedance tuner to reduce reflection instead of a fixed at-
0.042 dB
to under 18 GHz 40, 50, 60 dB (18 GHz) tenuator, and by preparing a receiving instrument with
65
2 Research and Development of Calibration Technology
66 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)
2-4 Calibration of RF Attenuators
0.065 dB
80 dB 0.16 dB
10 MHz, 30 MHz, Step 10, 20, 30, 40, 50, 10 MHz
100 MHz, attenuator 60, 70, 80, 90 dB 90 dB 0.48 dB
500 MHz, 80 dB 0.083 dB
1 GHz, 5 GHz, 30 MHz
10 GHz 90 dB 0.19 dB
Step 1, 2, 3, 4, 5, 6, 7,
30 MHz 0.018 dB
attenuator 8, 9 dB
(b)
0.069 dB
80 dB 0.17 dB
Over 10 MHz to
Step 10, 20, 30, 40, 50,
under 30 MHz 90 dB 0.48 dB
attenuator 60, 70, 80, 90 dB
Over 10 MHz
to under 1 GHz 80 dB 0.086 dB
Over 30 MHz to
under 100 MHz 90 dB 0.19 dB
67
2 Research and Development of Calibration Technology
Receiver
Signal generator
P2 _ direct
G L
b2
aG
G L
a2
Attenuator Receiver
Signal generator
S11 S12
S
12 S 22 P2
G L
a1 S 21 b2
aG
S11 S 22
G L
S12
b1 a2
FFig. A.1 Expression using S-parameters of attenuation measurements
substituted with the following matrix equation. definition of attenuation was the value when ΓG = ΓL= 0,
S11 S12 0 1 so consider substituting ΓG=ΓL=0 in Equation (A.8).
S (A.5)
21 S 22 1 0 In this case, we obtain
2
AdB LdB 10 log10 S 21 [dB] (A.9)
Then we obtain the following equation. G L 0
68 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)
2-4 Calibration of RF Attenuators
G 0
Substitution Loss L 0 Incremental Attenuation
×-1 dB dB
dB dB dB S 21b S 21e
L L L
i b e A i
dB
A AdB
b
dB
e
G 0
Insertion Loss L 0 Attenuation S-parameter
×-1
LdB
e AbdB dB
S 21e
Measurement (2nd time)
FFig. A.2 Relationship between attenuation, insertion loss, substitution loss, and S-parameters
S 21b
2
1 S11e G 1 S 22e L S 21e S12e G L 2 RReference
Li
S 21e
2
1 S11b G 1 S 22 b L S 21b S12 b G L 2 1 H. Iida, “A survey on RF attenuation standard,” AIST Bulletin of Metrology,
vol.3, no.4, pp.609–624, NMIJ, AIST, Feb. 2005.
(A.10) 2 JCSS Estimation Examples of Uncertainty (Coaxial step attenuator at a setting
of 30 dB), 6th Edition, JCG211S11, NITE, Aug. 2007.
If expressed as dB, we obtain 3 ISO, Guide to the Expression of Uncertainty in Measurement, 1st edition, 1995.
dB S 21e
2
1 S11b G 1 S 22 b L S 21b S12 b G L
2 4 L. A. Harris and F. L. Warner, “Re-examination of mismatch uncertainty when
Li 10 log10 [dB] measuring microwave power and attenuation,” IEE Proc., vol.128, Pt. H, no.1,
S 21b
2
1 S11e G 1 S 22e L S 21e S12e G L 2 Feb. 1981.
(A.11) 5 Y. Bitou and K. Ehara, “Uncertainty Evaluation Method with Uncorrected
Deflection,” Journal of the Japan Society for Precision Engineering, vol.74, no.6,
This value is called “Substitution Loss”. Moreover, if we
pp.604–610, Aug. 2008.
substitute ΓG=ΓL=0 in this equation, we obtain 6 Y. Bitou, H. Hosoya, K. Mashimo, and K. Ehara, “Uncertainty Evaluation
2
dB dB S Method with Uncorrected Deflection (2nd Report),” Journal of the Japan
Ai Li 10 log10 21e 2 [dB] (A.12) Society for Precision Engineering, vol.76, no.9, pp.1036–1042, March 2010.
G L 0
S 21b 7 JCSS Estimation Examples of Uncertainty, Guide for Expression of Uncertainty
Even in the case of “incremental attenuation”, like attenu- - Interpolation Formula, 1st Edition, JCG200S21-01, July 2011.
ation, we can say the following.
z Incremental attenuation is the value of substitution
loss when the reflection coefficients of the signal Katsumi FUJII, Dr. Eng.
generator and the receiver are 0. Research Manager, Electromagnetic
z Incremental attenuation is the ratio of the inverse of Compatibility Laboratory, Applied
Electromagnetic Research Institute
the square of the amplitude of the S-parameters of Calibration of Measuring Instruments
the attenuator; when expressed in dB, it is the same and Antennas for Radio Equipment,
Electromagnetic Compatibility
as the amount calculated by taking the ratio of the
square value of the amplitude of S21 and assigning a
negative sign to it. Kojiro SAKAI
z Incremental attenuation is a unique value of the step Technical Expert, Electromagnetic
attenuator, not dependent on qualities (reflection Compatibility Laboratory, Applied
Electromagnetic Research Institute
coefficients) of the signal generator and the receiver. Calibration of Measuring Instruments
Figure A.2 shows the relationships between attenuation, and Antennas for Radio Equipment
69
2 Research and Development of Calibration Technology
Tsutomu SUGIYAMA
Senior Researcher, Electromagnetic
Compatibillity Laboratory, Applied
Electromagnetic Research Institute
Calibration of Measuring Instruments
and Antennas for Radio Equipment
Kouichi SEBATA
Senior Researcher, Electromagnetic
Compatibillity Laboratory, Applied
Electromagnetic Research Institute
Calibration of Measuring Instruments
and Antennas for Radio Equipment,
geodesy
Iwao NISHIYAMA
Electromagnetic Compatibillity Laboratory,
Applied Electromagnetic Research Institute
Calibration of Measuring Instruments
and Antennas for Radio Equipment
70 Journal of the National Institute of Information and Communications Technology Vol. 63 No. 1 (2016)