Controller Troubleshooting
Garment Models
Garment Troubleshooting
FRONT VIEW
Handle
Control Panel
REAR VIEW
Set:
1. Place CritiCool unit in the desired location and lock front wheels.
2. Pour cold tap water into water tank up to maximum level (between the two lines).
3. Turn power ON.
• Following a short Self Test, the system automatically cools the water
• Avoid further connections until patient arrival to the ICU
• Place appropriate garment on bed or underneath patient
(see Instructions of Use).
• Connect Water tubes to garment and to CritiCool unit.
The garment automatically fills up! (Lower water level in the reservoir is normal)
• Check that clamps on garment are open and a clicking sound is heard.
Wrap:
4. Secure on patient
Cool:
• Plug Gray sensor into Core socket and Green sensor into Surface socket of
CritiCool unit.
• Once sensors are inserted to patient and valid feedback is obtained,
algorithm is activated.
• The CritiCool unit will not Thermo regulate if core sensor is not properly
inserted into patient
• Ensure no interference from other factors influence correct continuous
temperature feedback readings.
• When feedback is unavailable predefined cooling program is set, an alarm
sounds and a message appears.
Re-warming:
Gradual re-warming is advised (do not exceed more than 2 °C above actual temperature)
* For Accidental hypothermia or fast re-warming: switch to Normothermia management mode
CONTROL PANEL
Cooling\Heating
Core
Status
Temperature
System
Surface Function Status
Temperature
Energy Transfer
to/from Patient
Mode Key
Set Point
Temperature
Alarm Mute
Arrow Keys
Step-by-Step Operation
Treatment Mode
Selection Screen
An alarm will sound to inform about any problem. The system will
assist you in solving the problem.
Controller Troubleshooting
connect again
Water leak between
connecting tubes and Damage to connecting tubes Replace connecting
CritiCool unit tubes
Damage to quick coupling con- Call Biomedical depart-
nector ment
Special Male
Connector for Draining
Water Tank Special Male Connector
for Draining Water Tank
Figure 1 Figure 2
Controller Troubleshooting
HALT - PLEASE Error during normal Turn OFF system for three seconds
RESTART function and then ON again. If problem
persists, turn OFF CritiCool unit and
contact your local service
representative.
XX Note the number that appears on
screen
ATTENTION
CHECK CORE Misplacement of sensor Connect core sensor to appropriate
SENSOR in core socket socket
Controller Troubleshooting
Turn OFF system for three
ATTENTION seconds and turn ON
WATER TEMP TOO LOW again. If problem persists,
Water Temperature
turn OFF CritiCool unit
< 10oC (50.0F)
and contact your local
16 service
representative
ATTENTION Water Temperature Turn OFF system for three
WATER TEMP > 42oC (107.6F) seconds and turn ON
TOO HIGH again. If problem persists,
turn OFF CritiCool unit
and contact your local
17 service representative
** Pressing Alarm Mute silences buzzer for this present condition and any
re-occurrence of this condition for thirty minutes.
The message will
ATTENTION disappear after ten
The CritiCool unit detects
ELECTRO-MAGNETIC seconds If message
electromagnetic field higher
INTERFERENCE appears for more than
than allowable standard
22 ten seconds, replace
Core Sensor
ATTENTION
No Surface Sensor inserted Connect Surface Sensor
CONNECT SURFACE
into socket to socket
SENSOR
Garment Models
°CureWrap
Pediatric
ThermoWrap
Infant ThermoWrap
Infant °CureWrap
Garment Models
°CureWrap
Pediatric Infant
ThermoWrap ThermoWrap
•The garment should be filled with water prior to wrapping and securing to patient.
•Do not use garment to lift or move patient.
•Avoid handling garment with sharp objects.
•Make sure garment is not folded or clamped.
•Before moving patient from the operating table, ensure garment is fully released.
Garment Troubleshooting
Garment leak Accidental puncture of gar- Turn OFF CritiCool unit and
ment during course of op- allow water to return to
eration reservoir
Water leak from connector Connecting tubes are not Close garment clamps.
between garment and connected properly Disconnect connecting
connecting tube tubes and connect again
10