Instruction for Methodinstallation to PGC9000 with Mictalk

Tools: Notebook with Serial Com-Port,


Zero-Modem-Cable,
RMG-Files Mictalk.exe, mic.dat, 4911509N.met

1. Open Switch on Front, change GC-Mode to STOP, Connect PC-Com1 with Zero-Modem-Cable to
GC9000 Backside C4
2. Open DOS-BOX and type in „mictalk 1 0 5“ (1=Com1)

3. After „Enter“ follow screen will shown. Type in “mic.dat”

4. After „Enter“ follow screen will shown. Type in “L” and press Enter

--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
C:\RMG_Server_Hotline\PGC9000_MicTalk.doc Seite 1 von 2
5. After „Enter“ follow screen will shown. Type in “S” press Enter. Then type “4911509N.met” and
press Enter.

The new method will sent to GC9000. When finish Press X for quit the program. At least you must
start a Kalibration-Run.

--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------