PLEASE FIND BELOW THE DETAILS OF PROCEDURE FOLLOWED DURING THE

THERMAL SURVEY. THE BELOW DETAILS ARE THE EXTRACT OF OUR


FURNACE cALrBRArroN pRocEDURE No
pwl - 30s
.ad 6
REF STD : AMS 27s08, /q.7".f
t*i//':
s
1r.0 PRocEDURE \,ei ii
planning and preparation of the on site calibration job is done [#X-2
11.1
\sqsf
lI.2 Select the no. of thermocouples to be used as per the volume of the furna
surveyed as per Table 11 of AMS 27508

1 1.3 Standard thermocouples (survey sensors ) and the temperature monitoring device should
be calibrated internally against Master standards before going for site .

11.4 The engineer coordinates with the site in charge and starl the work.

11.5 Check the specifications of the furnace suryey requirements for any deviation and
feedback to the customer before calibration / survey.

1 1.6 The survey equipments are placed at proper place where the line voltage and the earth are
as required.

11.1 Place thermocouples to monitor time and temperature profiles in the furnace
to evaluate the uniformity of the intended working zone. Place them at the outer
extremities, the center of the proposed working zone, and in areas of suspected variability
in the proposed working zone. Attach suruey test sensors to, or set them in the load or
place them with the Heat sinks.

1 1.8 For proposed working zone less than or equal to 0.085m3 the thermocouples are placed at
the top, bottom and centre or front, middle and rear for surveys.

1 1.9 For proposed working zone more than 3 cubic feet [0.085m'] the thermocouples are
placed at the outer extremities, centre and in suspected areas of variability of the furnace
working zone.

11.10 The thermocouples are connected to the temp. monitoring device with compensating
cable & connectors with proper polarity.

1 1.1 1 The furnace is fired / switched on and allowed to stabilize at the first survey point (temp.)
with the required rate of heating.

11.12 Switch on the temp Indicating device to monitor the temp

11.1 3 Take readings starting when the furnace temp. rises to within 200'f [1 00'c] of the set
point temperature and continue throughout t
11.14 Readings should be taken at least once every 3 minutes to determine the temperature of
the furnace working zone approaches the bottom of the temperature range being surveyed
on data sheet.

1 1.15 Once the furnace temperature has reached the set point temperature, the temperature of
all test locations should be recorded at 2 minutes intervals maximum, for at least 10
minutes. Then readings should be taken at 5 minutes interval maximum, for sufficient
time to determine the recurrent temperature pattern of the furnace working zone for at
least 30 minutes.

1 1.16 Temperature at any point in the working zone shall not vary more than The specified
limits Before the furnace set point temperature is reached.

11.17 Now the furnace is set at the next required calibration / survey point. follow
steps 11.13 to 1i.16.