LOGIQ™ S6 Software Option Installation
Instructions

Operating Documentation

Copyright© 2006 by General Electric


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

Important Precautions
Language

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN
ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
AVERTISSEMENT • NE PAS TENTER D’INTERVENTION SUR LES é QUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS é Té CONSULTé ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAîNER CHEZ LE
TECHNICIEN, L’OPé RATEUR OU LE PATIENT DES BLESSURES DUES à DES
DANGERS é LECTRIQUES, Mé CANIQUES OU AUTRES.

• DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENö TIGT, IST
ES AUFGABE DES KUNDEN Fü R EINE ENTSPRECHENDE ÜBERSETZUNG ZU
SORGEN.
• VERSUCHEN SIE NICHT, DAS GERä T ZU REPARIEREN, BEVOR DIESES
WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

- 1
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

• ESTE MANUAL DE SERVICIO Só LO EXISTE EN INGLé S.


• SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLé S, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIó N.
• NO SE DEBERá DAR SERVICIO Té CNICO AL EQUIPO, SIN HABER
AVISO CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELé CTRICAS, MECá NICAS O DE OTRA
NATURALEZA.

• ESTE MANUAL DE ASSISTê NCIA Té CNICA Só SE ENCONTRA DISPONíVEL EM


INGLê S.
• SE QUALQUER OUTRO SERVIç O DE ASSISTê NCIA Té CNICA, QUE Nã O A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, é DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIç OS DE TRADUç ã O.
ATENÇÃO • Nã O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTê NCIA Té CNICA.
• O Nã O CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANç A
DO Té CNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELé TRICOS,
MECâ NICOS OU OUTROS.

• IL PRESENTE MANUALE DI MANUTENZIONE è DISPONIBILE SOLTANTO IN


INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE è TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVVERTENZA AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

2 -
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

- 3
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (262) 827-3468 or 8*285-3468 immediately after
damage is found. At this time be ready to supply name of carrier, delivery date, consignee name, freight
or express bill number, item damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy And Procedures
Bulletins.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be performed by qualified GE Medical Systems
personnel. In performing all electrical work on these products, GE will use its own specially trained field
engineers. All of GE’s electrical work on these products will comply with the requirements of the
applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.

OMISSIONS & ERRORS


If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Medical Systems Global Documentation Group with specific information listing the system type,
manual title, part number, revision number, page number and suggestion details. Mail the information
to: Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226.

GE Medical Systems employees should use the iTrak System to report all documentation errors or
omissions.

4 -
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

Revision History

Revision Date Reason for change


1 April 10, 2006 Draft Release

List of Effected Pages

PAGES REVISION PAGES REVISION PAGES REVISION


Title Page 1 1 to 16 1 Back Cover N/A
Warning
1
(pages i to iv)
Revision History
1
(pages v to vi)
List of Effective
1
(pages v to vi)

- 5
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

6 -
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

LOGIQ™ S6 Software Option Installation


Instructions

Section 1-1
Introduction

This Service instructions describe the LOGIQ™ S6 software option installation as follows:

NOTICE The Stress Echo, 4D, Vocal-volume calculation, and VCI-Volume contrast imaging require 512MB
memory option.

R.6.x.x or
Option Part Number HCAT# later
Advanced 3D 5171754 H41932LF Yes
B-Flow 5169730 H41932LD Yes
Coded Contrast 5173023 H41932LG Yes
DICOM 5173166 H41932LH Yes
LOGIQ View 5166914 H41932LE Yes
CrossXBeam 5173238 H41932LJ Yes
Report Writer 5154930 H41932LK Yes
Stress Echo 5172058 H41932LL Yes
PFD 5168447 H41932LM Yes
TVI 5173189 H41932LA Yes
SRI 5172730 H41932LN Yes
Vocal-Volume
Calculation 5173079 H41932LB Yes
VCI-Volume Contrast
Imaging 5173017 H41932LC Yes
LS6 M12L Enable Option 5175356 Yes
LS6 M7C Enable Option 5175187 Yes
LS6 M3S Enable Option 5175946 Yes

NOTICE Vocal-volume calculation and VCI-Volume contrast imaging require the realtime 4D option.

- 1-1
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

Section 1-2
Tools Required
Option strings

Section 1-3
Time Required
Approximately 15 minutes plus travel time

Section 1-4
Parts Required
The software options listed in Introduction contain the same parts in the kit.

ITEM PART NAME QTY PART# REMARKS


1 Software Certification Sheet 1 2137225-100
2 Option Label 1 2111577
3 Option Label 1 2111567
4 Installation Instructions 1 5173830 This manual
>
5 PDF key top label L6 1 5167351 For LS6 PDF Option ONLY

1-2 Section 1-2 - Tools Required


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

Section 1-5
Procedure
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation.
Perform each step in sequence and check it off when completed. If a problem occurs after completion,
go back through the procedures and check for implementation errors before contacting your Ultrasound
Region Engineer.

1-5-1 Preparations
• Option strings (This is a option key with a sequence of letters and numbers. Consult your local GE
personnel.)

1-5-2 Installing Option Key Strings


1.) Push the Power switch approximately 2 seconds.
2.) The SYSTEM EXIT window appears. Click on Logoff.

Figure 1-1 System EXIT window

3.) Press UTILITY > Admin on the touch panel.


4.) The OPERATOR LOGIN window appears.
Change the User level to Admin, then enter Password. Then click on Log on.

Figure 1-2 Operator LOGIN window

5.) Select Utility > Admin > System Admin.

- 1-3
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
6.) Enter Option string into New Key dialogue box. Click on Add, then Save.

Add

Key String
Entry
Disable

• If an option fails to install, delete it as follows:


a.) Click on the option key string to be deleted from the SW Option Key list.
b.) Click on Remove.
c.) Click on Utility to confirm it.

1-4 1-5-2 - Installing Option Key Strings


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3 Functional Check-out

1-5-3-1 Advanced3D
1.) Press 3D key on the keyboard.
2.) Touch 3D soft key on the touch panel.

3.) Verify that Adv3D tab appears.

1-5-3-2 B-Flow

NOTICE You can use 10L, M12L, 7L, 12L, 3.5C, 3.5CS, 4C, or T739 probe for this check-out.

1.) Connect any of probe listed above into a probe connector.

- 1-5
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
2.) Select the related probe on the touch panel.
3.) Verify that “Option not available” message does NOT appear when the B-Flow button is pressed.

1-5-3-3 Coded Contrast

NOTICE You can use 10L, 7L, 3.5C, 3.5CS, 4C, M3S or 3S probe for this check-out.

Verify that “Option not available” message does NOT appear when the Contrast button is pressed.

1-5-3-4 DICOM

DICOM Setting
After installing DICOM option software, you must set the parameters to establish communications
between the scanner and DICOM device(s).

1.) Touch the Utility button on the Touch Panel.


2.) Touch the Connectivity button on the Touch Panel.

Figure 1-3 Touch Panel - Utility

3.) Click on Device tab. Device setting screen appears.


4.) Click on Add and enter Name and IP Address of the destination device in Properties.

1-6 1-5-3 - Functional Check-out


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
5.) Click on Save.

Item Descriptions

Select the device name to which the scanner is connected via network.
Name
When the DICOM device is not listed in this menu, add a new device using Add button.

IP address Enter IP Address of the DICOM device to which the scanner is connected via network.

6.) Click on Service tab. Service setting screen appears.


7.) Select DICOM device type (ex. Dicom Image Storage) then click on Add.
8.) Set the parameters below. Set other parameters if necessary.
9.) Click on Save.

Item Descriptions

Name You can enter the name which you desire.

AE Title DICOM Application Entity Title of the DICOM Server

Port No. Port Number of the DICOM Server

Figure 1-4 Device and Service

- 1-7
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS

10.)Click on Buttons tab. Set the parameters below:

Item Descriptions

Button Select the button assigned for DICOM device.

Select
This is the same parameter as Service Type.
Destinations

a.) Select the button assigned for DICOM device.


b.) Click on DICOM device type (ex. Dicom Image Storage).
c.) Click on >> button to add Printflow View.
d.) Click on Save.

Figure 1-5 Button

11.)Restart the system.


12.)If the DICOM device requires the scanner setting at the device side, set the parameters to
communicate with the scanner.
- For AE title and IP address of the scanner, check them using Connectivity > TCPIP >
Computer Name and IP address.

Functional Check
Send an image via DICOM to a DICOM device (server, printer.), then verify that it should be received
with no error.

1-5-3-5 LOGIQView
Verify that “Option not available” message does NOT appear when the LOGIQView button is pressed.

1-8 1-5-3 - Functional Check-out


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3-6 CrossXBeam
When using a linear probe, the half-tone dot meshing of the CrossXBeam button disappears.

1-5-3-7 On-Board Report


Click on Report on the touch panel, then verify that the DESIGNER button in Reports can be used.

- 1-9
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3-8 Stress Echo
1.) Select Preset > Card (Cardiac).

2.) Select Scan > Protocol. Verify that the protocol screens appears on the CRT monitor and the
protocol buttons appears on the LCD screen.

1-5-3-9 PFD
1.) Enter the color mode.

1 - 10 1-5-3 - Functional Check-out


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
2.) Press the PFD button. Verify that the following screen appears.

PFD button

When pressing the PFD Button,


the scale display changes.

- 1 - 11
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3-10 TVI
1.) Connect a sector probe to the scanner.
2.) Select a cardiac application.
3.) Press the PFD button.
4.) Verify that TVI tab appears on the touch panel screen.

1-5-3-11 SRI
1.) Select B mode and touch SRI HD to verify that level selection button appears.

1 - 12 1-5-3 - Functional Check-out


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-3-12 Vocal-Volume Calculation
1.) Connect and select 4D probe. Press 3D/4D button. This operation enter pre-4D mode.

2.) Press L button. The 4D screen appears.

3.) Press Freeze button.


4.) Verify that “Vocal” tab appears on the touch panel screen.

- 1 - 13
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
5.) Touch Vocal tab. Verify that the vocal menu appears on the screen.

1-5-3-13 VCI-Volume Contrast Imaging


1.) Connect and select 4D probe. Press 3D/4D button. This operation enter pre-4D mode.

1 - 14 1-5-3 - Functional Check-out


GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
2.) Press L button. The 4D screen appears.

3.) Press Freeze button.


4.) Verify that “Static 3D” tab appears on the touch panel screen.
5.) Touch Static 3D and Visualization, then verify that VCI Static button appears

- 1 - 15
GE MEDICAL SYSTEMS
DIRECTION 5173830, REVISION 1 LOGIQ™ S6 SOFTWARE OPTION INSTALLATION INSTRUCTIONS
1-5-4 Final Procedure
1.) Unscrew four screws and remove the Right Side Cover. Refer to Components Replacement in
Chapter 8, LOGIQ™ 7 Service Manual, 5162630 (included in LOGIQ S6 eDoc CD (5166927-200)).
2.) Attach the Option Label onto the Nest Cover as shown in Figure 1-6.

System Version Label


Software Label

Nest Cover
Option Labels

Figure 1-6 Option Label on Nest Cover

3.) Re-install the Right Side Cover. Refer to Components Replacement in Chapter 8, LOGIQ™ 7
Service Manual, 5162630.
4.) Fill out the proper customer information on the Product Locator Card. Mail the installation card
“Product Locator” to the address corresponding to your region.

Figure 1-7 Product Locator Card

NOTE: The Product Locator Card indicated in Figure 1-7 may be different from the actually provided on.

5.) Properly dispose of excess material.


6.) Keep these option service instructions with LOGIQ™ 7 Service Manual.
7.) This completes the installation of the LOGIQ™ 7 Software Option.

1 - 16 1-5-4 - Final Procedure