Documente Academic
Documente Profesional
Documente Cultură
Taylar Crist
Date: 9/25/2019
Abstract
During week 1 the purpose of the lab was to determine the correction factors for
sequential 10 mL aliquots of water delivered via buret into a 250 mL flask in order to generate a
volume correction factor calibration curve to be used for two titrations in the second week. Each
10 mL aliquot of deionized water delivered to the flask was massed out and the mass of the water
delivered was compared to the recorded volume on the buret to determine the correction factor.
The purpose of part two was to perform a standardization titration on KHP to determine the
molarity of the unknown NaOH solution. Further another standardization titration was performed
determine its molarity. Both titrations were performed using the buret to deliver the titrant NaOH
or sodium thiosulfate to a 250 mL beaker holding a known amount of KHP or triiodide anion
respectively. The average concentration of NaOH was 0.0814 ± 0.0001 M with an relative
standard deviation of 0.2%. The average concentration of sodium thiosulfate was 0.0453 ±
0.0001 with a relative standard deviation of 0.3%. The relative standard deviations for both
titrations were less than 0.8% indicating that both titration methods were successful for
determining the concentration of the standard solutions of NaOH and sodium thiosulfate.
Introduction
Determining the correction factor for the recorded buret volume reading and generating a
correction factor plot was the objective for week 1. For week 2, performing a successful titration
of KHP and triiodide anion to determine the molarity of the unknown concentration of NaOH
and sodium thiosulfate respectively, further ensuring the titration molarities were within 0.8%
relative standard deviation were the desired outcomes. The procedure from week 1 and week 2 is
important because it outlines the proper technique for creating standard solutions. Creating
standard solutions is important due to their high purity and accurately determined concentration.
This makes them useful in analytical chemistry for various techniques such as titrations and
dilutions.
Calibration is an important technique that is used to minimize method error and random
error in a procedure. In addition, calibration improves the accuracy and precision of each
measurement. In terms of the experiment, calibrating the buret improves accuracy and precision
of the actual amount of titrant added to the analyte. Calibrating a buret is important because the
ability of the buret to deliver an accurate volume degrades over time. The glass of the buret can
warp overtime due to the temperature fluctuations the tip of the buret can build up with
particulates, and the solution put in the buret can succumb to thermal expansion. More
specifically to the experiment, calibrating the burets improves the accuracy of the actual amount
of sodium thiosulfate or NaOH delivered during the titration, which makes the objective of
achieving a relative standard deviation of less than 0.8% more attainable. The calibration
experiment from the first week provided a correction factor for any volume of substance
delivered via the buret. Once the correction factors were graphed, it was used in the second part
of the experiment to determine the more accurate volume of NaOH and sodium thiosulfate
delivered which in turn increases the accuracy of the calculated molarity of the unknown titrants.
The correction factor calculation is displayed below in the results section labeled sample
calculation 3.
standardization titration was used to determine the concentration of NaOH with high accuracy
using a KHP as the primary standard. KHP was titrated with NaOH of an unknown concentration
and a triiodide anion was titrated with an unknown concentration of sodium thiosulfate. A
primary standard is a reference chemical of a known concentration that is used to determine the
concentration of an unknown solution. In this experiment the NaOH and sodium thiosulfate
solution concentrations were unknown. The titration of KHP was performed to the endpoint
which was measured by the color change in the solution caused by the indicator phenolphthalein.
NaOH reacts with KHP in a 1:1 molar fashion that results in the formation of water molecule and
a base salt, as shown in the reaction labeled KHP reaction with NaOH. This is important to take
O O
- + -
OK O Na+
+ Na+OH- - +
+ H2O
OH O K
O O
The titration for the triiodide anion with sodium thiosulfate is not as straight forward. In
the first reaction of this titration, the triiodide anion is formed by reacting KIO3 and KI with
sulfuric acid.
because it allows us to determine the moles of I3- in solution. Next sodium thiosulfate is oxidized
The common form of thiosulfate, 𝑁𝑎2 𝑆2 𝑂3 ∗ 5𝐻2 𝑂, is not pure enough. Solutions of
𝑁𝑎2 𝑆2 𝑂3 had to be freshly prepared by dissolving thiosulfate in boiling water to promote the
disproportionation of 𝑆2 𝑂32− ,
Methods
Buret Calibration
The 50 mL buret washed according to the washing instructions printed out at the
glassware washing station. The 50 mL buret was determined to be clean after distilled water had
been flushed through the buret and there were not any distilled water droplets sticking to the
walls of the buret. The 50 mL buret was filled with deionized water a few mL over the 0.00 mL
mark. The water was drained into a waste beaker until the air bubble below the stop-cock
disappeared. Once the stop-cock was retightened the tip of the buret was gently grazed with the
side of the beaker to remove that last drop of water. The initial volume was noted and after 5
minutes the volume of the buret did not change. A 250 mL flask and glass stopper was massed
out. 10 mL portions of deionized water from the buret were drained into the 250 ml flask at a rate
of about 0.5 mL/sec. After each 10 ml draining the 250 ml flask was massed out, in addition the
deionized water in the buret was allowed to settle for about a minute after each 10 mL draining
in order to get an accurate and precise reading. After 50 ml of deionized water was drained, the
remaining water was drained and the buret was rewashed. The procedure was repeated a second
time.
After all of the data had been collected, a correction factor was computed for the volume
of deionized water dispensed by taking the volume per gram that the deionized water occupies at
24°C and multiplying it by the mass of water the was dispensed by the buret. A reproducibility
factor was calculated for between each trial for each aliquot of deionized water delivered. A
correction graph was made by plotting the recorded volume of deionized water from the buret
versus the average correction factor calculated for each aliquot of water delivered.
An NaOH solution was prepared by measuring out about 1000 mL of boiled deionized
water. 5.3 mL of NaOH was added to the flask once cooled. Following, about 0.5 grams of KHP
was added to each of the four 250 mL flasks, and each flask was filled with 30 mL of deionized
water in order to dissolve the KHP. A buret was washed according to the washing protocol
posted in the lab. The buret wash additionally flushed with 10 mL of the NaOH before filling the
buret just over the 0.00 mL mark. The excess NaOH was drained into a waste beaker to get the
initial recording. The first titration was performed quickly, while the following three titrations of
A 0.01 M KIO3 solution was made by dissolving 0.4 g of KIO3 in a 200 mL volumetric
flask. In addition, a 0.044 M Na2S2O3 solution was made by dissolving 2.77 g of Na2S2O3 ·5H2O
in 250 mL of boiled deionized water containing 0.025 g of Na2CO3. The 0.044 M Na2S2O3 was
made up in a tightly capped bottle. Next 20.00 mL of KIO3 was pipetted using a volumetric pipet
into and Erlenmeyer flask. 0.8 g of KI solid and 4 mL of 0.5 M H2SO4 was added to the
Erlenmeyer flask, and the sodium thiosulfate was immediately titrated following the addition of
H2SO4. The sodium thiosulfate solution eventually turned a pale yellow color and at that point 1
mL of starch indicator was added and the solution was titrated until the deep blue color of the
Results
From the recorded masses of water delivered from the buret, the actual volume of water
was calculated by taking the mass of water delivered and multiplying it by the density of water at
Sample Calculation 1
𝒎𝑳
𝟗. 𝟗𝟕𝟖𝟖 𝒈 (𝟏. 𝟎𝟎𝟑𝟖 ) = 𝟏𝟎. 𝟎𝟏𝟕 𝒎𝑳 𝑯𝟐 𝑶
𝒈
The buret readings were determined by taking the final buret reading for that aliquot of
water and subtracting it from the initial reading displayed in sample calculation 2.
Sample Calculation 2
𝟏𝟗. 𝟖𝟎 𝒎𝑳 − 𝟎. 𝟎𝟎 𝒎𝑳 = 𝟏𝟗. 𝟖𝟎 𝒎𝑳
The correction factor was calculated by taking the actual amount of water delivered
determined from its mass and subtracting it from the observed amount delivered read from the
buret as shown in sample calculation 3. This was done for both titrations.
Sample Calculation 3
Once the correction factors were calculated from the first and second trial, the
reproducibility of the correction factors were calculated by taking the difference between the
correction factors from each trial with respect to the aliquot of water delivered. Meaning the
correction factor produced for the 1 aliquot of water in trial 1 was subtracted by the correction
factor calculated for the first aliquot delivered in trial 2. The reproducibility factor calculation is
Sample Calculation 4
The reproducibility factor was less than 0.04 mL for all trials, which means that the
correction factors were significantly reproducible. Once the correction factors were confirmed by
the reproducibility factor, the average of the correction factors was plotted on the y-axis, while
the observed volume delivered was plotted on the x-axis. For the NaOH titration, the graph in
figure 1 was used to correct for the volume of NaOH delivered while the graph in figure 2 was
used to correct for the sodium thiosulfate delivered. The correction factor graphs were applied to
the procedure in week 2 to record an accurate and precise volume of titrant delivered. Based on
the amount of NaOH or sodium thiosulfate delivered determines the correction factor that is
applied to that volume as seen in sample calculation 4. Once the three solution samples were
titrated, the concentration of the NaOH solution was determined by taking the moles of KHP in
solution and converting moles of KHP to moles of NaOH, and then dividing the moles of NaOH
by the corrected volume of NaOH to determine the molarity of the NaOH solution.
Molarity of NaOH
0.0025404𝑚𝑜𝑙 𝑁𝑎𝑂𝐻
= 0.081214757 𝑀
0.03128 𝐿 𝑁𝑎𝑂𝐻
This was done for the three solution samples for NaOH and sodium thiosulfate. Next the
The standard deviation of the average of the three solutions for NaOH and sodium
Sample Calculation 7
The relative standard deviation was calculated by dividing the standard deviation by the
Sample Calculation 8
𝟎𝟎𝟎𝟏
𝟏𝟎𝟎 ( ) = 𝟎. 𝟏%
𝟎. 𝟎𝟖𝟏𝟒
The standard deviation of the mean was calculated by taking the standard deviation and
𝟎. 𝟎𝟎𝟎𝟏
= 𝟎. 𝟎𝟎𝟎𝟎𝟔
𝟑𝟎.𝟓
Lastly the 95 percent confidence interval was calculated using the CONFIDENCE.T
Sample Calculation 10
RSD 0.2
0.14
0.12
0.1
0.08
0.06
0.04
0.02
0
0 10 20 30 40 50
Volume Delivered (mL)
Figure 2. Sodium Thiosulfate Titration Correction Volume for Observed Buret Reading.
RSD 0.3
The objective during the first experiment was to calibrate the buret, by finding the
correction factor for each aliquot of water delivered. The next step was to take the correction
factors and plot them according versus the volume of water delivered according to the buret. The
object during the second experiment was to perform a standardization titration with known
amounts of KHP and triiodide anion to determine the concentration of NaOH and sodium
thiosulfate respectively. In combination with the experiment from week 1, another objective for
the second experiment was to apply the correction factor curve to the volumes of NaOH and
sodium thiosulfate delivered to accurately and precisely calculate their concentrations. The
relative standard deviation for the NaOH titration was 0.2% and the relative standard deviation
for the sodium thiosulfate titration was 0.3%. Both of the relative standard deviations were less
than 0.8% indicating that our titrations were successful in determining the concentration for the
standardized solutions. The concentration of the NaOH solution was 0.0814 ± 0.0001 M while
the concentration of the sodium thiosulfate solution was 0.0453 ± 0.0001 M. The concentrations
of the solutions of NaOH and sodium thiosulfate can both be accepted as standard solutions
because both relative standard deviations are less than 0.8%. These solutions have been
determined to be reliable to standardize other solutions and make dilutions of these standard
solutions. These results match the expectations for the standardization protocol because the
concentration of the analytes, or primary standards were known and both of the primary
standards were stable, they did not decompose, and both of the reactions between the titrants and
the analytes went to completion. In addition, the expectation that calibrating the buret from the
first week in lab proved to significantly impact the accuracy and precision of the calculating
concentrations of sodium thiosulfate and NaOH because the relative standard deviation was less
than 0.8%.
One limitation in the experiment when titrating the KHP was being able to decipher when
the solution turned pink. There was no certain point where the solution was about to turn pink
gradually, a single drop of NaOH could turn the whole solutions vivid pink. Once the endpoint
was nearly reached all the analyte needed was an incredibly small amount of NaOH, which was
hard to control using the buret. In addition, when performing both titrations it was challenging to
get the last drop of the titrant into the flask. The last drop would run down the side of the flask
and would entirely work its way into solutions leaving traces of it on the side. Further it was
slightly challenging to operate the buret and swirl the flask at the same time, especially when the
titrations were nearing their endpoint. One area that needs to be improved upon is the use of the
95 percent confidence interval and the standard deviation of the mean. Neither of those two
calculated values could be used to support or antagonize the results because there is not another
Conclusion
The purpose of the first week of the experiment was to perform a calibration protocol of a
50 mL buret and generate a calibration curve to improve the accuracy and precision of the
volume delivered by the buret. During the second week of the two-part lab, the purpose was to
apply the calibration curve to determine a more accurate measurement of the aliquot of sodium
thiosulfate and NaOH delivered to the analyte during the titration in order to standardize the
solutions of NaOH and sodium thiosulfate. During week 1 the burets were calibrated by
delivering sequential aliquots of distilled water and massing out each of the aliquots. The volume
recorded by the buret was compared to the actual amount of water delivered into the flask as
determined by calculating the volume of the water mass and density at 24°C. During the second
week of lab two titrations were performed. One of the titrations was a direct titration of a known
concentration of KHP with an unknown concentration of NaOH. The end point of the titration
was determined by the phenolphthalein indicator. The concentration of the standardized NaOH
solution was 0.0814 ± 0.0001 M with a relative standard deviation of 0.2%. The second titration
performed was sort of a back titration of a known concentration triiodide anion with an unknown
sodium thiosulfate. The reason this was not a direct titration was because there was an initial
reation between 𝐾𝐼𝑂3 and 𝐾𝐼 that formed the triiodide anion. The concentration of the sodium
thiosulfate was 0.0453 ± 0.0001. The relative standard deviation of the sodium thiosulfate
solution was 0.3%. Lastly we found that the concentrations of the standardized solutions are
accurate and precise because of the small standard deviations and the relative standard deviations
being less than 0.8%. There is a high confidence that the titrations were successful and the
1. Harris, D.C. Quantitative Chemical Analysis, 8th edition, W. H. Freeman and Co., New
York; 2006.