Service Manual for DRX Plus and DRX Core

Detectors


*
Important:
Qualified service personnel must perform these procedures.
• See the Carestream Service Portal for possible updates before you begin.
• When doing the procedures in this document, you must use safe work practices and wear the correct personal
protective equipment (for example, safety eyewear) according to your company’s standard operating procedures.
• When servicing this equipment, be sure to read and follow the Manufacturer’s Safety Data Sheets for any chemical,
cleaner, solvent, paint, lubricant, adhesive, or replacement battery to understand proper use and disposal
precautions.

Confidential
Publication No. 6K8894
2019-05-02
Supersedes:
2017-09-29
All rights reserved. No part of this manual may be reproduced or copied in any form by
any means—graphic, electronic, or mechanical, including photocopying, typing, or
information retrieval systems—without written permission.
Contents

1 Safety

2 Replacements
Remove the Bumper...................................................................................................................... 2-1
Install the Bumper ......................................................................................................................... 2-2
Remove the Access Cover Seal Assembly and/or Backup Battery ............................................... 2-3
Install the Access Cover Seal Assembly and/or Backup Battery .................................................. 2-4
Remove the Battery Latch Assembly ............................................................................................ 2-6
Install the Battery latch Assembly ................................................................................................ 2-7
Remove the Battery Contacts ....................................................................................................... 2-8
Install the Battery Contact .......................................................................................................... 2-10

3 Diagnostics
LED Display Matrix......................................................................................................................... 3-1
Reset Button .................................................................................................................................. 3-4
DRX Diagnostic Tool...................................................................................................................... 3-5

Publication History

6K8894 | 2019-05-02 i
Contents

ii 6K8894 | 2019-05-02
1 Safety
Safety
This information defines the safety and information icons used in this publication.
This icon is used for conditions that could cause personal injury from radiation.

CAUTION:
Caution notices identify potentially hazardous situations which, if not avoided, might cause minor
or moderate injury.

ESD: Possible damage from electrostatic discharge


This icon is used for conditions that could cause damage to the equipment.
Electrostatic discharge (ESD) is a primary source of:
• Equipment failure
• Equipment repairs
Electrical components in this equipment can be damaged by a low level electrical discharge
imperceptible to a human being.
Preventive Measures
• Check for an ESD warning label before doing any procedure with ESD-sensitive components.
All sensitive components have graphic labels that frequently include instructions. Use all label
instructions.
• Wear a grounding strap when you touch ESD-sensitive components. Check that the clip
remains fastened to a ground that has a clean surface with no paint.
• Repair components in an ESD-protection area or use a portable grounding mat.
• When moving ESD-sensitive components from area to area, insert and transport the
components in the special material made for the transport of these components.

*
Important:
This icon is used for important information.

Note:
This icon is used for additional information.

6K8894 | 2019-05-02 1–1


Safety

See the Safety and Regulatory Information for additional safety information.

1–2 6K8894 | 2019-05-02


2 Replacements

Remove the Bumper


Prerequisites:
None.

1. Place the detector on a flat surface with the tube side (1) facing up.
2. Locate the 4 corner bumpers (2).
3. Peel back the trim label (3).
Save the trim label for re-use, unless damaged.
4. Use fine-tip needle nose pliers to remove the 2 pins (4) holding the bumper (5).
Pull the pins straight up to remove.
5. Remove the bumper.

2 4

1 5
H224_6512_C1

6K8894 | 2019-05-02 2–1


Replacements

Install the Bumper


1. Install the bumper (5) with the lip end (6) near the tube side of the detector (1).
2. While holding the bumper tightly against the detector housing, use a T-hex drive to
push the 2 pins (4) one at time and straight down into a fully seated position.
When the pin is fully seated it will be flush or below the housing top surface.
3. Install the trim label (3).
If the trim label was damaged when it was removed, install a new trim label.

2 4 6
3

1 5
H224_6512_C2

2–2 6K8894 | 2019-05-02


Replacements

Remove the Access Cover Seal Assembly and/or


Backup Battery
Prerequisites:
None.

ESD: Possible damage from electrostatic discharge


Remember to wear a grounded ESD wrist strap.
1. Remove the main battery.
2. From the area for the backup battery, remove the label (1).
3. Using a T5 TORX driver, remove 12 screws (2) and access cover (3).
4. Remove the seal assembly (4) and discard.
Clean all adhesive from housing—do not use cleaners or solvents.
5. For Backup Battery replacement only. Disconnect the Battery wire (5) from circuit
board.
6. For Backup Battery replacement only. Remove and discard the backup battery (6).
Dispose of used batteries according to the manufacturer’s instructions. At the end of
its useful life, properly dispose of the battery following all local regulations.

3
2

H224_6510_C1

6K8894 | 2019-05-02 2–3


Replacements

Install the Access Cover Seal Assembly and/or Backup


Battery
Prerequisites:
None.

ESD: Possible damage from electrostatic discharge


Remember to wear a grounded ESD wrist strap.

3
2

H224_6510_C1

1. For Backup Battery replacement only. Connect the backup battery wire (5) to the
circuit board.
2. For Backup Battery replacement only. Install the backup battery (6).

*
Important:
If there is currently no tape affixed inside the detector to hold wires in place, affix the two
pieces of Kapton tape that are provided as described below.
3. Remove the release paper from the back of one piece of Kapton tape.

2–4 6K8894 | 2019-05-02


Replacements

4. Affix the tape across the three bosses (posts) to prevent the wires from moving onto
the bosses.

5. Cut off a small piece of the second Kapton tape.


6. Affix this tape over the short antenna wire to keep it in place.

7. Remove the release paper from the foam pad on the bottom side of the seal
assembly.
8. Install:
a. New seal assembly (4)—One corner of the seal assembly is notched for proper
orientation.
b. Access cover (3).
c. 12 screws (2)—use a T5 TORX driver.
Torque the screws to 40 in oz.
d. Label (1).

6K8894 | 2019-05-02 2–5


Replacements

Remove the Battery Latch Assembly


Prerequisites:
None.

1. Remove the main battery from the detector (1).


2. Remove:
a. 2 Screws (2)—use a T6 TORX driver.
b. Battery latch assembly (3).

3 2

H224_6513_C1

2–6 6K8894 | 2019-05-02


Replacements

Install the Battery latch Assembly


Prerequisites:
None.

1. Install:
a. Battery latch assembly (3).
b. 2 Screws (2)—use a T6 TORX driver.
Tighten the screws to 40 oz-in.
2. Install the battery into the detector.

3 2

H224_6513_C1

6K8894 | 2019-05-02 2–7


Replacements

Remove the Battery Contacts


Prerequisites:
None.

1. Remove the main battery from the detector.

ESD: Possible damage from electrostatic discharge


Remember to wear a grounded ESD wrist strap.
2. Remove:
a. 4 screws (1)—use a T6 TORX driver.
b. Plate (2).
2 1

H224_6505_C1

*
Important:
The battery contact is not symmetrical. Observe the orientation of the battery contact before
removing.

2–8 6K8894 | 2019-05-02


Replacements

3. Remove:
a. 2 screws (1)—use a T6 TORX driver or 1.5 mm ALLEN wrench.
b. Battery contact assembly (2)—carefully observe the orientation of the battery
contact before removing.
1

H224_6506_C1

4. Clean all the black RTV from around the housing for the battery contact.
Do not use cleaners or solvents.

6K8894 | 2019-05-02 2–9


Replacements

Install the Battery Contact


Prerequisites:
None.

ESD: Possible damage from electrostatic discharge


Remember to wear a grounded ESD wrist strap.
1. Carefully clean the surfaces (2) on the detector that mate with the battery contact.
Do not use cleaners or solvents.
2. Apply a continuous bead of RTV (3) approximately 1.00 mm (0.03 in.) in diameter to
the front edge (3) of the housing and to the left and right edge (1) of the battery
contacts.
Make sure that the RTV does not get on the plated contacts of the circuit board.
1

2 3

H224_6507_C1

Note:
You must install the new battery contact in the same orientation as the old one was installed.
3. Install:
a. Battery Contact (2).
b. 2 screws (1)—use a T6 TORX driver and tighten to 40 oz-in.
4. Apply black RTV to the cracks on the left, right, and top edges of the new battery
contact.
5. Apply a continuous bead of RTV (3) approximately 1 mm (0.03 in.) in diameter in a
closed loop that flows along the front edge of the battery contacts, then crosses over

2–10 6K8894 | 2019-05-02


Replacements

the caulked top cracks, and follows the overlap area in the housing for the battery
retainer plate.
1 2 3

H224_6508_C1

6. Install:
a. Plate (2)
b. 4 Screws (1)
7. Use foam-tipped swabs to clean up any squeezed-out RTV.
Do not use cleaners or solvents.
With practice, it is possible to do this operation with no squeeze-out and no clean up.
1
2

H224_6509 C1

6K8894 | 2019-05-02 2–11


Replacements

8. Install the main battery into the detector.

Postrequisites:
None.

2–12 6K8894 | 2019-05-02


3 Diagnostics

LED Display Matrix

Key

The drawing above locates the Reset button (1), and the LEDs that indicate Error (2),
Communication (3), and Power (4).

LED Description

Blue

Green

Magenta

Orange

Yellow

6K8894 | 2019-05-02 3–1


Diagnostics

LED Description

White

Blank, off, clear

Double circle (any colors) indicates blinking

Error Status

LED Indicator Condition Meaning

Blinking The detector has a fault.

Blinking Calibration has failed.

On Shock is recorded and calibration is recommended.


Below firmware version 86

On Shock is recorded and calibration is recommended.


Firmware version 86 and above

3–2 6K8894 | 2019-05-02


Diagnostics

Communication Status

LED Indicator Condition Meaning

AP Mode

Blinking Wireless associating and authenticating.

Non AP Mode

AP Mode

On Authenticated on wireless network or Ethernet on tether.

Non AP Mode

AP Mode On The detector is connected but the console is not ready to


Non AP Mode expose.

Power Status

LED Indicator Condition Meaning

On The detector is Ready, and the battery is good.

On The detector is Ready, and the battery is low. Limited expo-


sures are remaining.

On The power is on, and the battery is below imaging level.

6K8894 | 2019-05-02 3–3


Diagnostics

Reset Button
The Reset button is located on the side of the detector. It is intended for a quick restart
of the detector software, similar to the action of removing and reinserting a battery.

Note:
When the reset button is pressed the LEDs will brighten as an immediate feedback response.
Pressing and holding the Reset button for more than 5 seconds disrupts the detector and causes
an immediate hard shutdown. A hard shutdown should only be used when directed by a Service
provider or as a part of troubleshooting a serious condition.
Pressing and releasing the reset button 3 times in 3 seconds is only used in the most serious
conditions. It is intended for Service use only.
Pressing and releasing the reset button 4 times in 3 seconds is only used for detector AP mode.

Reset Button Action

Press once Detector software restart

Press and hold ≥ 5 sec Hard shutdown and reboot

Press and release 3 times in 3 seconds Resets the detector to private default settings.
(For service use only) • IP Address: 10.0.1.150
• SSID: CSH_DRX_FACTORY
• Passphrase: sundance2009

Press and release 4 times in 3 seconds Places the detector in AP mode and sets/restores the
detector's public default settings:
• IP Address: 192.168.1.150
• SSID: Carestream-serial # of detector
• Passphrase: sundance2009
The country code and channel are not changed

3–4 6K8894 | 2019-05-02


Diagnostics

DRX Diagnostic Tool

*
Important:

• The diagnostic tool can only be used with the DRX Plus and DRX Core detectors. It cannot be
used on previous detectors such as DRX-1, DRX-1C, DRX 2530C or TDR 3543.
• The detector must be powered on and connected to the console in either wireless or tether
mode.
1. From the DR console or Image Suite Application, exit to desktop to shut down the
Directview or Image Suite software.
2. For Image Suite only, it is necessary to ensure that the Detector Integration
Package (DIP) is not running on the system.

*
Important:
When you are done using the tool, it is necessary to restart the DIP services.
a. Click the Windows Start button.
b. Select Task Manager.
c. Click the Services tab.
d. Select Detector Integration Manager from the list of running services.
e. Right-click and select Stop from the drop-down menu.
f. Close the Task Manager dialog by clicking the X.

6K8894 | 2019-05-02 3–5


Diagnostics

3. Launch the diagnostic tool.


DirectView Image Suite

For systems with V5.7 software, if avail- Step


able, click the shortcut icon in the desk-
top or, select C:\Program Files > 1.) Select C:\Program Files > Image
Kodak > CaptureConsole > bin > Suite > PrimaryWS > DIP > bin
DRXDiagnosticTool.exe For systems
with V5.6, V5.5 or V5.3, the diagnostic 2.) Right-click on the application and select
tool must be loaded manually. Run as administrator.

4. To connect the application to the detector type the following into the screen.
a. Serial number
b. IP Address—If the IP Address is unknown, leave the field blank.
5. Click Connect.
The application will search for the IP Address if it was not entered in the previous
step.

6. Navigate through the following tabs on the screen to capture the desired
information.
Tab Description

Provides basic information about the detector, such as serial num-


Summary
ber, firmware version, private IP address, and list of current faults.

Provides more information about the detector, such as the MAC


Additional Detail address or image count. Provides the capability to reboot the
detector or get detector logs for diagnostics or troubleshooting.

Provides the detector IP Address, SSID, and Passphrase and the


Private Network
capability to change the settings or reset to factory defaults.

Enterprise Network Provides the settings for the hospital network.

View shocks for the detector. The same information is available in


Shocks
the DirectView software.

View battery information for the detector. You can click Main
Battery Battery or Backup Battery to view information specific to that
component.

3–6 6K8894 | 2019-05-02


Diagnostics

Tab Description

Clear orphan images from the detector. Note that this just deletes
Orphan Images
the images – they will not be saved anywhere.

Provides the detector calibration history, which is useful for diag-


Calibrations
nostics and troubleshooting.

Used to test network speed by pinging the detector a specified


Diagnostics
number of times. You can export the results to a file for analysis.

7. For additional diagnostics, select Tools > DRX Remote Command.

Note:
For Image Suite software, the DRX Remote Command tool can also be launched separately as
an application from the bin folder (same location as DRX Diagnostic Tool).
8. Type the detector IP Address.

6K8894 | 2019-05-02 3–7


Diagnostics

9. Select the desired toggle button for information regarding the private network and
Enterprise Wireless performance.

Note:
Set Network Mode and the buttons below it are only for Enterprise Wireless configurations.
10. Click Disconnect when you are done using the tool.

*
Important:
For Image Suite, after you are done using the tool, it is necessary to restart the DIP services.
Navigate to the same location in Task Manager, right-click on Detector Integration Manager,
and Start the service.

3–8 6K8894 | 2019-05-02


Publication History

Version Date Changes

A 2015-07-01 First release

B 2016-04-28 • Added the DRX Core detector to the title page


and the DRX Diagnostic Tool section.
• Removed reference to the DRX Plus detector
from the LED Display Matrix section.

C 2016-09-06 Added information for Image Suite software to the


DRX Diagnostic Tool section.

D 2017-08-01 • Updated Backup Removal and Battery Install


sections
• Updated LED tables

E 2017-09-29 • Added instructions and graphics for affixing


Kapton tape in Install the Access Cover Seal
Assembly and/or Backup Battery
• Updated LED tables

F 2019-05-02 Updated the LED Display Matrix section.

6K8894 | 2019-05-02 I
Carestream Health, Inc.
150 Verona Street
Rochester, NY, USA 14608
© Carestream Health, Inc., 2019
Made in USA.
CARESTREAM is a trademark of Carestream Health.
Pub. No. 6K8894
Rev. F