CALIBRATION CERTIFICATE

ULR No. CC225319000003218F Calibration Field -


Page 1 of 2
Certificate No. LES-CCL/MECH/VG/140 Mechanical
Calibration Date 11.06.2019 Suggested Date of Next Calibration 10.06.2020
Customer Name :- M/s ENVIRON ACTION CONSULTANCY
Address :- 23,KAVERI PLAZA,OPP . SARDAR
PATEL BHAVAN, NADIAD, DISTT - KHEDA ( GUJARAT)

Reference :- S.R.F. No.


2019/669 Date :- 10.06.2019 Date of Issue:- 11.06.2020

01. DUC Fitted in Instrument


Name Make Model Sl.No.
V.Stack Sampler Vaibhav VES - 5 1C3P19

02. Details of DUC


Name Vacuum Gauge (GAS) / (SPM) Temperature(oC) 25 ± 3
Make VIGA Relative Humidity (%) 45 -75
Sl.No. SVG - 1C3P19/ GVG - 1C3P19 B. Pressure (mmHg) 743.45
Cal. Range 0 - 480 mmHg

03. Standard Equipment used for calibration


Standard Equipment Name Range Sl.No. Traceability
1. Digital Vacuum Gauge -1.0 to 2 bar VEM1503079 / BELZ, Faridabad
LES-CCL/R/2504
Certificate No. Calibration Date Valid Up to
7138966 19.06.2018 18.06.2019

04. Calibration Procedure :- LES-CCL/WI/31/MECH/02


Remark : Refer page 2 of 2 for Calibration Results

Notes :-
Checked By Authorized By
1. Reference used are directly traceable to national standard through
2. Results reported are valid at the time of and under the stated conditions of measurement
3. This Certificate refers only to the particular item calibrated.
4 .This certificate shall not be reproduced, except in full without the written DEVENDRA SINGH SHIV SHANKAR SINGH
permisson of LES-CCL. Greater Noida (U.P.) ( Technical Manager ) ( Chief Executive Officer )
LES - Centre For Calibration Laboratory

ULR No. CC225319000003218F Page 2 of 2


Calibration Date 11.06.2019 Suggested Date of Next Calibration 10.06.2020
Certificate No. LES-CCL/MECH/VG/140

05.Calibration Results for Vacuum Gauge (Sl. No. SVG- 1C3P19)

Set DUC CYCLE 1 CYCLE 2 CYCLE 3 Standard Error Ue.


Value UP DOWN UP DOWN UP DOWN Avg.Reading % ±
(mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) FS (bar)
80 77.6 78.4 78.2 77.9 75.8 75.8 78.1 0.38 0.0180
100 112.5 81.7 82.1 82.4 82.5 83.5 81.6 3.67 0.0180
200 177.8 182.3 181.2 184.2 181.9 186.7 182.4 3.53 0.0180
300 301.7 301.8 302.7 184.2 302.4 302.5 302.1 -0.42 0.0180
400 455.8 455.3 456.0 455.7 456.0 456.8 455.9 -11.19 0.0180
480 476.5 478.4 482.5 480.4 480.4 478.2 479.4 0.12 0.0180

06.Calibration Results for Vacuum Gauge (Sl. No. GVG- 1C3P19)

Set DUC CYCLE 1 CYCLE 2 CYCLE 3 Standard Error Ue.


Value UP DOWN UP DOWN UP DOWN Avg.Reading % ±
(mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) FS (bar)
80 75.7 74.6 76.0 75.9 75.8 75.8 75.6 0.87 0.0180
100 112.5 116.4 116.5 116.0 115.7 116.4 115.6 -3.12 0.0180
200 184.4 184.4 185.3 185.4 186.0 185.4 185.2 2.97 0.0180
300 303.9 304.4 304.9 305.5 306.6 306.5 302.1 -0.42 0.0180
400 425.7 425.8 426.0 426.0 425.8 426.0 455.9 -11.19 0.0180
480 479.3 477.8 478.2 478.4 481.4 480.1 479.2 0.16 0.0180

Remark : The reading of vacuum represents the Average of five reading


Uncertainty Contributing Factor 1. Repeatability (based on 3 Cycle) 2.Uncertainty of master instruments
3.Uncertainty due to resolution of DUC 4.Uncertainty due to hystersis,
The evaluated Expanded Uncertainty in calibration at a coverage factor k = 2 , for degrees of freedom =∞ and
confidence level is 95 % for Normal distribution.
Notes :-
Checked By Authorized By
1. Reference used are directly traceable to national standard through
unbroken chain of calibration .
2. Results reported are valid at the time of and under the stated conditions of measurement
3. This Certificate refers only to the particular item calibrated.