Presentation on

Guided by: Dr. V.T.Thakkar

APC, S. Patel 1
BROOKFIELD DIGITAL
VISCOMETER
• PHOTOGALLERY
0
• SOP
1
• VALIDATION
2
APC, S.P.Patel 2
APC, S.P.Patel 3
APC, S.P.Patel 4
APC, S.P.Patel 5
APC, S.P.Patel 6
SOP
1) Turn on the viscometer and allow standing,
must be auto zero. After the few seconds the
screen appears which indicates 2 digits.
2) Now press the key. The screen displays the
instruction to remove spindle. After removing
the spindle and pressing the key the
instrument begins. It is auto zeroed.
3) After approximately 15 second the screen
displays the instruction to replace spindle
4) Attach the spindle to viscometer by screwing
them on the lower shaft using left hand
thread.

APC, S.P.Patel 7
5) Press select spindle key and up and down arrow
keys. When the desired code is displayed
release the arrow key.
6) To select spindle, first press either up and down
key which cause the area to show current speed,
press the set speed key for adjusting the speed.
7) Insert center of this spindle in the test material until
the fluid level is at the immersion groove on the
spindle shaft. Tilt the spindle slightly while
immersing to avoid air entrapment.
8) To measure high viscosity choose a small spindle
and to slow speed if the chosen spindle/speed
result in torque above 100% and then reduce the
value.

APC, S.P.Patel 8
9) Allow time for the individual reading to stabilize,
record the value.
10) Press the motor on/off/escape key to turn off motor
11) The time mode allows the user to record the
reading for fixed period of time or until a set torque
value is attained.
12) Then enter/auto turnkey allows determining
maximum calculated viscosity possible with current
spindle/speed.
13) Pressing the up and down arrow keys will allow the
viscometer data to be examine, pressing and other
key ( except enter/ print key) will bring back
normal display.
14) Turn off the mains after use.

APC, S.P.Patel 9
VALIDATION
• INSTALLATION
IQ QUALIFICATION

• OPERATION
OQ QUALIFICATION

• PERFORMANCE
PQ QUALIFICATION

APC, S.P.Patel 10
IQ: INSTALLATION
QUALIFICATION
 Inspect package for shipping damage
 Confirm that all components have been
supplied as detailed in the Operating
Manual and Packing List
 Assemble lab stand as shown in manual.
 Mount viscometer head to assembled lab
stand as shown in manual.
 Verify appropriate spindle set (LV #1-
#4)(RV/HA/HB #2-#7)

APC, S.P.Patel 11
OQ: OPERATION
QUALIFICATION
 Level viscometer using adjustable feet
 Remove shipping cap on spindle
coupling nut
 Power on Viscometer and follow
instructions displayed to complete auto
zero
 Perform oscillation check.

APC, S.P.Patel 12
PQ: PERFORMANCE
QUALIFICATION
 Verify calibration using a minimum of
one calibrated viscosity standard fluid
according to the procedure described
below
 Centipoise reading should be within the
value stated on the fluid ± the instrument
accuracy and the fluid accuracy.

APC, S.P.Patel 13
Oscillation check
 This check verifies the mechanical
condition of the sensing mechanism in the
viscometer.
 Auto-zero the instrument according to
directions in the manual.
 Do not attach the spindle but do press any
key so the main screen will be displayed.
 The display should show 0.0% torque. The
tolerance for this and all tests is ±0.1%

APC, S.P.Patel 14
1) Start the motor rotating at 10 or 20
RPM.
2) The % reading should initially bounce,
then is should stabilize at 0.0% ±0.1%
3) Stop the motor. When the % display
again stabilizes it should return to 0.0%
±0.1%

APC, S.P.Patel 15
Operational and Calibration
Check
 Determining Measurement Accuracy
 All Brookfield instruments are guaranteed to be
accurate to within 1% of the full scale
range(FSR) in use.
 The full scale range (FSR) is simply the
maximum viscosity that can be measured with
any spindle/speed combination.
 Select a specific spindle and rotational speed.
 To determine the resulting FSR (full scale
range), depress and hold the AUTORANGE key.

APC, S.P.Patel 16
 The % display will show 100% and the
cP display will show the FSR (full scale
range) that can be measured under
current choice of spindle and speed.
 1% of this value is the measurement
accuracy under these conditions.

APC, S.P.Patel 17
 Calibration Check
 In principle, all Brookfield rotational
viscometers are calibrated torque meters.
 Brookfield offers a variety of viscosity
calibration fluids that enable one to check
the calibration and linearity of the instrument
over its operating torque range of 10%-
100% torque.

APC, S.P.Patel 18
 Any spindle and any one of several calibration
fluids may be used to perform a calibration
check.
 Experts recommend using one spindle and one
fluid while taking measurements at multiple
speeds.
 This enables multiple calculations of viscosity
throughout the torque range of the instrument;
 one at low torque, one at medium torque and one at
high torque.

 The FSR, and consequently the instrument


accuracy in terms of centipoise, will be different
at each speed.

APC, S.P.Patel 19
 The calibration fluid must be transferred into
a 600ml low form, Griffin beaker (without
creating any bubbles) and,
 on LV and RV models, the instrument Guard
Leg must be used.
 The beaker, fluid, spindle and Guard Leg
must stabilized at 25.0°C (±0.1°C) before the
calibration check can be performed.

APC, S.P.Patel 20
EX:
Actual viscosity is 5,120cP,
 Allowable error = 1% of the FSR + 1% of the
actual value of the calibration fluid
 The lower and upper viscosity limits,
indicating acceptable calibration, equal the
actual value of the fluid,
 in this case 5,120cP, ± the allowable error.

APC, S.P.Patel 21
RPM MIN MAX cP 1% of % cP ALLOWAB
cP (FSR) FSR torque READIN LE
G ERROR

12 1000 10000 100 51.2 5120 ± 151

6 2000 20000 200 25.5 5100 ± 251

3 4000 40000 400 12.7 5080 ± 451

Allowable error = 1% of the FSR + 1% of the actual


value of the calibration fluid

APC, S.P.Patel 22
Calibration Worksheet
A B C D E F G H I
Instrument Spindle
Fluid 1% of RPM FSR 1% of 1% of Visco
Value Fluid Model cP FSR FSR sity
from Value (depre cP cP cP
Certificate
cP ss
cP and
hold
AUTORA
NGE
key)

APC, S.P.Patel 23
 The viscosity reading should equal the cP value on the
fluid standard to within the combined accuracies of
the Viscometer and the viscosity standard as calculated
below
Low Limit Reading High Limit Pass or Fail
= A – (B + G) I = A + (B + G)

APC, S.P.Patel 24
 Any reading outside these limits may
indicate a Viscometer problem.
 Contact the Brookfield technical sales
department or your local Brookfield
dealer/distributor with test results to
determine the nature of the problem.

APC, S.P.Patel 25
Reference:
 Brookfield DV-ii+ pro, programmable
viscometer, operating manual no. M/03-
165-A0404
 “Solution for a viscous world” brookfield
2007 catalog.
 www.brookfieldengineering.com

APC, S.P.Patel 26
APC, S.P.Patel 27