Operating manual

Operating with computer


or terminal

Measuring amplifiers
for desktop housing

SCOUT55

A0241-5.3 en
SCOUT55 3

Content Page

Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

1 Introduction to operation with computer or terminal . . . . . . . . . . . . . 5

2 RS-232-C serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6


2.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Characteristics of the serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . 6

3 Communication with the SCOUT 55 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7


3.1 Connecting the computer to the SCOUT 55 . . . . . . . . . . . . . . . . . . . . 7
3.2 Activation of the RS-232-C interface . . . . . . . . . . . . . . . . . . . . . . . . . . 8

4 Instruction set for the SCOUT 55 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8


4.1 Important conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.1.1 Instruction structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1.2 Structure of data output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.2 Description of the individual instructions . . . . . . . . . . . . . . . . . . . . . . 11
4.2.1 Adjusting parameters of the group Additional functions . . . . 13
4.2.1.1 Adjusting parameters of the RS-232-C
serial interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.2.1.2 Output of the amplifier identification
(version number) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2.1.3 Define printout format . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2.2 Adjusting parameters of the group data set . . . . . . . . . . . . . . 16
4.2.3 Output format, output of measured values . . . . . . . . . . . . . . . 19
4.2.3.1 Measurement output format . . . . . . . . . . . . . . . . . . . . 19
4.2.3.2 Output measured value . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.4 Adjusting parameters of the group adaptation . . . . . . . . . . . . 23
4.2.4.1 Adjusting amplifier input . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.4.2 Change filter setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2.4.3 Adjust the autocalibration . . . . . . . . . . . . . . . . . . . . . . 28

A0241-5.3 en HBM
4 SCOUT55

4.2.5 Adjusting parameters of the Group Calibration . . . . . . . . . . . 30


4.2.5.1 Change engineering unit . . . . . . . . . . . . . . . . . . . . . . . 30
4.2.5.2 Setting indication adaptation . . . . . . . . . . . . . . . . . . . 32
4.2.5.3 Adjust zero balance . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.2.5.4 Adjust Measuring range . . . . . . . . . . . . . . . . . . . . . . . 35
4.2.5.5 Taring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2.6 Adjust the parameters of the group Limit value 1 ... 4 . . . . . 37
4.2.7 Adjust the parameters of the group peak value . . . . . . . . . . . 39
4.2.8 Adjust the parameter of the group Inputs/Outputs . . . . . . . . 41
4.2.8.1 Select amplifier input-signal . . . . . . . . . . . . . . . . . . . . 41
4.2.8.2 Adjust the analogue output . . . . . . . . . . . . . . . . . . . . . 42
4.2.8.3 Adjust remote control . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.2.8.4 Setting the control contact assignment . . . . . . . . . . 45

5 Keywords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

6 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

HBM A0241-5.3 en
SCOUT55 5

Safety instructions

The device complies with the safety requirements of DIN EN 61010-part1


(VDE 0411-part1); protection class I.
Before connecting the device, please note the mains voltage adjusted at the
factory; it is specified on the device rear (230 V / 115 V, 48 ... 60 Hz).
Please also note the safety information in the separate documentation
”Operating manual” Part 1 for the SCOUT 55.

1 Introduction to operation with computer or terminal


The SCOUT 55 amplifier enables measurements from passive transducers.

Main characteristics:
D Transducers that may be connected: S/G full and half bridges, inductive full
and half bridges, piezoresistive and potentiometric transducers, LVDTs
D 10-digit alphanumeric display
D Operation via membrane-type keyboard; single keys can be locked
D 2 peak-value memories for maximum and minimum value, and envelope
curve and current value
D 4 limit switches
D RS-232 serial interface for connection of a computer or printer
D Parameter memory for storage of up to 4 data sets
D Control inputs and outputs (electrically isolated by optocouplers)

This manual explains operation of the SCOUT 55 with a terminal or computer.


The connection of a computer via the RS-232 is described for setting all the
parameters and for obtaining the measurements.
All information necessary to adapt the amplifier to your measurement task via
Terminal are described in a separate operation manual ”Operating manual −
Part 1, Amplifier for panel mounting SCOUT 55”.

A0241-5.3 en HBM
6 SCOUT55

2 RS-232-C serial interface


2.1 General notes
The RS-232-C serial interface is called serial interface, because data is trans-
mitted successively bit by bit. General notes on the interface’s properties are:
D relatively ”low” transmission speed
D requires a 3-core cable for transmission in both directions (duplex or bi-
directional) as minimum arrangement.
D only one device can be connected (point-to-point connection)

START 0 1 2 3 4 5 6 PARITY /STOP


Logic 1

Logic 0

Fig. 1: Line level of character Y with negative logic


A Start bit is set before each character (data byte). Then follow the data bits
and a STOP bit. Due to the fact that data is transmitted successively, the
transmitting speed must correspond to the receiving speed.
The number of bits per second is called baud rate. The exact receiver baud
rate is synchronized with the START bit with each byte transmitted. Then fol-
low the data bits that are all of the same length. When the STOP bit is
reached, the receiver changes to wait state until reactivated by the next
START bit.
Data transmission is controlled with the X-ON (DC1) and X-OFF (DC3) soft-
ware handshake.
If the device is ready for data transmission, it issues the X-ON (DC1) control
character. If it cannot receive any data, e.g. if there is no free memory space,
the X-OFF (DC3) control character is issued.

2.2 Characteristics of the serial interface

Measuring rate 10 Measurements/s


Number of data bits 8 Bit
Stop-Bit 1*; 2
Parity odd, even* and none
Baud rate 300; 600; 1200; 2400; 4800; 9600*
* Factory set up

HBM A0241-5.3 en
SCOUT55 7

3 Communication with the SCOUT 55

3.1 Connecting the computer to the SCOUT 55


On the device rear there is a serial RS-232 interface for connecting a com-
puter or terminal. For RS−232 interfacing, a cable (1.5 m length) with free
ends and a 9−pole subminiature jack* (Order no.:3−3301.0104) is supplied.
Refer to the below figure for the wiring and pin assignment.

2 TxD
3 RxD
Computer RS-232 5 GND RS-232 SCOUT 55

9−pole D-subminiature jack


SCOUT 55 inter- green
RxD 2 TX
face jack brown
PC TxD 3 RX SCOUT 55
yellow
GND 5
DCD 1
DTR 4
DSR 6
CTS 8
7 not assigned
9 not assigned

Abb. 1: Computer/SCOUT 55 interfacing

For connecting the SCOUT 55 to a computer please proceed as follows:


D Connect both systems to the mains, but do not yet switch them on.
D Connect the interface as shown in the figure.
D Make sure that the computer’s interface configuration (baud rate, data for-
mat) does in any case match the SCOUT 55 default settings. Otherwise,
the interface configuration has to be adapted via the keyboard (see SCOUT
55 operating manual, part 1).
D Finally switch on both systems.
It is sufficient to connect a simple line printer with a printing speed of 4 sec-
onds/line. 12 columns are printed corresponding to a line length of 132 char-
acters.

* Accessories

A0241-5.3 en HBM
8 SCOUT55

3.2 Activation of the RS-232-C interface


Use the below instruction to activate the HBM interpreter:
* CTRL R (DC2) - Computer operation without echo
Enter the control character to set the device to the remote control state and
lock all functions except for the display functions.
Computer operation without echo means that no control characters are trans-
mitted back to the MVD 2555 but only the generated data. With the RS-232-C
interface any information that has been generated will be output immediately
after it has been transmitted completely to the output buffer.
Use the below instruction to deactivate the remote−control state:
CTRL A (SOH); see also DCL instruction on page 11.

4 Instruction set for the SCOUT 55

4.1 Important conventions


These conventions and general notes are to facilitate work with the
HBM-interpreter instructions.

Notation
D All instructions can be entered in lower-case or upper-case letters.

Instruction abbreviation
D The instruction abbreviations comprise 3 or 5 characters and, depending
on the instruction, a list of parameters that are separated by commas.

e.g. BDR6,2,1 (x)

Blanks
D With parameters, preceding and subsequent blanks are suppressed.

Types of instruction
D Interrogation instructions are marked by a question mark (?), they are used
to read out information.

e.g. BDR?

Responses
D The device responses given in the examples are printed in italics.

HBM A0241-5.3 en
SCOUT55 9

End of instruction
with input instructions:
D The end-of-instruction character is marked with (x). Permissible end-of-in-
structions are:

”;”, LF, LFCR, CRLF


with output instructions:
D The end-of-instruction character is marked with (y). The end-of-instruction
is always CRLF.

Input / output numbers


D The numbers that have been entered are converted into the numerical type
used with the respective parameter.
D Numbers are always output as fixed-point numbers.

Interfaces − serial
D With the RS-232-C-interface, computer communication starts with the per-
missible control characters.

’ CTRL R’ or ’CTRL B’ and is terminated by ’CTRL A’ or the DCL instruction.

D With serial interfaces, every instruction generates an output (response).

Response behavior
D Output instructions − marked by ? − always generate output data.
D Change parameters.
D When parameters effecting the measurement are changed, a calibration
will be carried out after input; the calibration may take 1..3s.

Instruction format
All instructions used have a specific format. Generally, there are two types of
instructions:

Setup instructions:
The SCOUT 55 is adjusted via the computer.
Example: BDR6,2,1 (x)
0 (y)
The interface has been adjusted to 9600 Baud, even parity and 1 STOP-Bit.

A0241-5.3 en HBM
10 SCOUT55

Interrogation instructions:
Measured values or device setups are read from the SCOUT 55
and displayed on the screen.
Example: BDR? (x)
6,2,1 (y)
The serial interface has been set to 9600 Baud, even parity and 1 STOP-Bit.

4.1.1 Instruction structure


Instruction abbrev. Parameter Endmark
TTT? p1,p2, ... p n (x)
Example: BDR? (x)
BDR Instruction abbreviation as alpha character (a ... z)
? only with interrogation instructions
p1,p2..pn Parameter values including sign (+/−) and numbers (0 ... 9)
or character strings (always in inverted commas” ”).
A positive sign may also be omitted.
, Separator
(x) End of instruction
Line Feed (LF), Semicolon (;) Carriage Return/Line Feed (CRLF)
or Line Feed/ Carriage Return (LFCR)
CR ASCII-character: Carriage Return = decimal 13
LF ASCII-character: Line Feed = decimal 10
; ASCII-character: Semicolon = decimal 59
If an additional parameter (e.g. parameter 2) is omitted, at least the separator
has to be entered. e.g. ASA1,,0(x)
If all additional parameters from a specific position are omitted, input may be
terminated by the end-of-instruction mark.

4.1.2 Structure of data output


q1,q2..qn (y)
Example 1:IDN? (x)
HBM,SCOUT 55,0,P10 (y)
In this manual, the responses transmitted by the SCOUT 55 are printed in ital-
ics (second line in the examples).
Output values:
q1,q2..qn Numerical values with sign, character strings (always in ” ” )
or ’?’ ” as error message
; Separator
(y) End sequence (CRLF)

HBM A0241-5.3 en
SCOUT55 11

4.2 Description of the individual instructions


In the next section, every instruction is quoted, its structure analyzed and ex-
plained by an example.
Instruction
The character string you have to enter to operate the SCOUT 55.
e.g. BDR
Syntax
The notation that has to be observed when entering an instruction:
e.g. BDR p1,p2,p3 (x)
Parameter
The meaning of possible parameters is explained:
e.g. if with the ASA instruction the parameter p1=1, this means:
1 V bridge excitation voltage
Effect
e.g. explanation of the SCOUT 55 setup.
Response
The SCOUT 55 responds to your entry. With terminal mode, this
response will be displayed on the screen (always with
output instructions).
Example
The example illustrates the instruction that has been entered and
the SCOUT 55’s response. The response is always printed in
italics. In the appendix, we have listed the individual instructions in
alphabetical order.
Instruction
DCL Device Clear
Terminate communication
Syntax: DCL (x) or with RS-232-C / RS-485 control code
CTRL A (ASCII code 01 decimal).
Parameter: None
Effect: Remote control is terminated.
Response: None
Example: DCL(x)
Interpreter is no longer active.
Note:
After this instruction, a new instruction can only be en-
tered after about 3 s.

A0241-5.3 en HBM
12 SCOUT55

ESR? Standard Event Status Register


Output error status register
Syntax: ESR? (x)
Parameter: None
Effect: Output decimal equivalent of Standard Event Status
Register (ESR).
The Standard Event Status Register (ESR) is set in the
case of communication errors. Different error causes
set different bits such that errors can be specified ex-
actly.
Response: q1(y)
q1 8, 16 or 32

Bit: 7 6 5 4 3 2 1 0
Device Dependent Error:
e.g. instruction not permitted
with this device

Execution Error:
parameter error
e.g. too many parameters
Command Error:
unknown instruction
(syntax error)

All other bits have not been assigned.

HBM A0241-5.3 en
SCOUT55 13

4.2.1 Adjusting parameters of the group Additional functions


4.2.1.1 Adjusting parameters of the RS-232-C serial interface

Instruction
BDR Baud Rate
Setting the parameters for the serial interface
Syntax: BDR p1,p2,p3 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P1 Baud rate

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 300

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 600

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 1200

ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
5
2400
4800
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
6 9600

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P2 Parity

ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
none
odd
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 even
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P3 Stop bits

ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1

ÁÁÁÁÁÁÁÁÁÁÁÁÁ
2
1 Stop bit
2 Stop bit

The character length of data transmission is always 8


bits.
Effect: Sets baud rate, parity bit and the number of stop bits.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

Example: Operation of the SCOUT 55 via RS-232-C interface:


BDR6,2,1 (x)
0 (y)
The RS-232-C interface has been adjusted to 9600
bauds, even parity and 1 stop bit.

A0241-5.3 en HBM
14 SCOUT55

Instruction
BDR? Baud Rate Query
Output of all parameters for the serial interface
Syntax: BDR?(x)
Parameter: None
Effect: Output of the adjusted baud rate, parity bit and number
of stop bits.
Response: q1,q2,q3 (y)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Baud rate *

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3
Parity
Stop bits

Example: BDR? (x)


6,2,1 (y)
The serial interface has been set to 9600 baud, even
parity and 1 stop bit.

4.2.1.2 Output of the amplifier identification (version number)


Instruction
SNR? Serial Number
Output serial number of device
Syntax: SNR?
Parameter: None
Effect: Output of the serial number of the device
Response: Character string (10 characters)
Example: SNR? (x)
4021837410
Instruction
AID? Amplifier Identification Query
Output amplifier identification
Syntax: AID?(x)
Parameter: None
Effect: Output of the amplifier identification (version number)
Response: Character string (20 characters )
Example: AID? (x)
HBM,SCOUT 55,0,P10 (y)
Company, device name, 0, version number.

HBM A0241-5.3 en
SCOUT55 15

4.2.1.3 Define printout format


Instruction
PFS Print Format Select
Define printout format
Syntax: PFS p1 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Signal to be printed
Measured value on display
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Gross

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Nett

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 Peak value 1 (max. value)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
8
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
16
Peak value 2 (min. value)
Peak value 3 (peak-peak value)
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
63 All signal combinations and status limits

All signal combinations are set by summing up the


code numbers.
Effect: Defines the signal to be printed.
The setup is effective for output via the external start of
printout (control input, key element).

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response:

ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Meaning
Instruction executed
ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Example:ÁÁÁÁÁÁÁÁ
Gross and nett ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error
are to be printed
p1 = 1+2
PFS 3 (x)
0 (y)
Instruction
PFS? Print Format Select Query
Query printout format
Syntax: PFS?(x)
Parameter: None
Effect: Output the signal to be printed.
Response: q1 (y)
Signal or signal combination set with the PFS instruc-
tion (coding, see PFS instruction).

A0241-5.3 en HBM
16 SCOUT55

Example: PFS? (x)


1 (y)
Gross signal is to be printed. (Initiation via control con-
tact or key).
4.2.2 Adjusting parameters of the group data set

Instruction
MDD Memory Device Data
Input data for amplifier setup
Syntax: MDD p1 (x)
Parameter: p1
Data for amplifier set up recalled from the amplifier us-
ing the instruction MDD? (as hexadecimal string
” ___ ”, approx. 100 Byte = 200 Characters).
Effect: Use this instruction to save and recall the complete
setup.
To change individual parameters, please use the
appropriate instruction (e.g. IMR).

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response:

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
0
?
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Instruction executed
Error

Example:ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
MDD ” ____(Hexadecimal string) ____” (x)
0 (y)
Amplifier is adjusted.

Instruction
MDD? Memory Device Data Query
Output amplifier setup data
Syntax: MDD? (x)
Parameter: None
Effect: Output of the amplifier setup-parameters
Response: ” ___ (Hexadecimal string) ___” (y) ”ca. 100 Byte = 200
characters.
Example: MDD? (x)
”0a00ff.......” (y)
Output of all setup-parameters.

HBM A0241-5.3 en
SCOUT55 17

Instruction
TDD Transmit Device Data
Store amplifier setup
Syntax: TDD p1,p2 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Amplifier settings

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Factory settings (set-up)
RECALL from parameter set 1 ... 8
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 SAVE from parameter set 1 ... 8

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 Automatic saving of zero/tare values

if p1=0 (factory setup); p2 no. effect


if p1=1 or p1=2 ; p2=no. of data set

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p2 Number of data set (if p1=1 or p1=2)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ... 4 Data set 1 ... 8

if p1=2 ; p2= automat. storage of zero balance value or


tare value is on/off

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p2 Status of automatic storage (Zero/Tare value)
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ (if p1=3)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
OFF

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ON

Effect: The amplifier setup is saved or backed up. The auto-


matical storage of the zero balance value or tare value
in EEPROM is on/off.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Meaning
Instruction executed
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

A0241-5.3 en HBM
18 SCOUT55

Example 1:TDD2,4 (x)


0 (y)
The actual amplifier setup has been saved in data
set 4.
Note: This instruction starts a calibration, thus allowing com-
munication to be continued only after 1 ... 3 s.

Example 2:TDD3,1 (x)


0 (y)
The automatic zero/tare value storage is switched on.
The zero setting value is saved in the current set of pa-
rameters each time the zero is set. The tare value is
saved in the current set of parameters with each taring.
Instruction
TDD? Transmit Device Data Query
Request for origin of amplifier setup
Syntax: TDD?p1 (x)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Parameter:

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Source of amplifier setup
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 Status of automatic storage (Zero/tare value)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: The source of the currently effective amplifier settings
are output or the status of the zero/tare storage is
given.
Response: if p1 = 0, q1 shows the origin of amplifier setup

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Source of amplifier setup
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ... 4 Data set 1 ... 4

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

if p1 = 3, q1 shows the status of automatic storage


(Zero/tare value)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Status of automatic storage / (Zero/tare value)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁ
0
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
OFF

ÁÁÁÁÁÁ (x)ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ON
Example 1:TDD?0

HBM A0241-5.3 en
SCOUT55 19

2 (y)
The origin of the currently effective amplifier setup is
data set 2.
Example 2:TDD?3 (x)
1 (y)
Automatic storage of zero point value and tare value is
on.

4.2.3 Output format, output of measured values


4.2.3.1 Measurement output format

Instruction
COF Change Output Format
Change measurement output format
Syntax: COF p1 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Measurement output format
Measured value, status (ASCII format)
ÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Measured value (ASCII format)
ÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Binary output of measured values 4 Byte

ÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
(MSB XXXX LSB)

ÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Binary output of measured values 4 Byte
(LSB XXXX MSB)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 Binary output of measured values 2 Byte (MSB LSB)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
5
6
Binary output of measured values 2 Byte (LSB MSB)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ Output of measured value (BCD-format)

4-byte-Output:

MSB LSB
1 2 3 4

3 meas. bytes 1 status byte

A0241-5.3 en HBM
20 SCOUT55

Bit: 7 6 5 4 3 2 1 0 status byte

1 Limit 1 On
2 Limit 2 On
4 Limit 3 On
8 Limit 4 On
16 Gross overflow
32 Net overflow
64 calibration error
128 changed setup
(e.g. by remote control signals)

2-byte-output: 1=MSB, 2=LSB

MSB LSB
1 2

2 measurement byte

BCD output:
1 2 3

1 byte sign 3 measurement bytes 1 status byte

The measurement values are scaled to the indicated


full-scale value. With binary format measured values
are output in ADC units.
Effect: With the subsequent MSV instruction, measured val-
ues will be output in the desired format.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error

Example: COF0 (x)


0 (y)
Output measured value and status in ASCII format.

HBM A0241-5.3 en
SCOUT55 21

Instruction
COF? Change Output Format Query
Query measurement output format
Syntax: COF?(x)
Parameter: None
Effect: Output code for the measurement output format .
Response: q1 (y)
Example: COF?(x)
0 (y)
Output of measured values and status in ASCII format.

4.2.3.2 Output measured value

Instruction
MSV? Measuring Signal Value Query
Output measured value
Syntax: MSV p1,p2 (x)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Parameter:

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
BRU
Signal
Gross (with display filtering)
ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 NET Net (with display filtering)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 SPWT1 Peak value1 (maximum)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 SPWT2 Peak value 2 (minimum)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
5

ÁÁÁÁÁÁ
6
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
SPWT3

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
GW1
Peak value 3 (peak-peak)
Level
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
7 GW1 Hysteresis
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
8 GW2 Level

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
9 GW2 Hysteresis

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
10 GW3 Level

ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
11 GW3
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
12 GW4
Hysteresis
Level
ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
13 GW4 Hysteresis

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
14

ÁÁÁÁÁÁÁÁÁÁÁÁ
BRU Gross (dyn., without filter)

ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
15 NET

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p2
Net (dyn., without filter)
Number of measured values
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Endless transmission
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ... 65535 Default = 1

A0241-5.3 en HBM
22 SCOUT55

Effect: Measurement output for the desired signal p1. The out-
put format depends on the COF instr. last entered.
Response: Measured value (output format see COF Instruction).
Example 1:Output in full-format ASCII
COF0 (x)
0 (y)
Get 1 measured value (gross signal).
MSV?1 (x)
9.998,0 (y)
Status byte
Measured value (e.g. 9,998 kN)

Get 3 measured values (Net signal).


MSV?2,3 (x)
9.998,0 CRLF
9.998,0 CRLF
9.998,0 CRLF (y)
Status byte
Measured value (e.g. 9,998 kN)

Example 2:Output in 4-Byte binary format


4-byte binary format
COF2 (x)
0 (y)
Get 1 measured value (gross signal)
MSV?1 (x)
#0ffeedd00CRLF(y)
1 status byte
3 measurement bytes
code for binary output

Example 3:Endless output


Endless output of measured values (gross signal).
MSV?1,0 (x)
#0ffeedd00CRLF
#0ffeedd00CRLF
#0ffeedd00CRLF
STP(x) Terminate output

HBM A0241-5.3 en
SCOUT55 23

Instruction
STP Stop
Stop measurement output
Syntax: STP (x)
Parameter: None
Effect: The measurement output which was started with MSV?
ceases.
Response: None
Example: STP (x)

4.2.4 Adjusting parameters of the group adaptation


4.2.4.1 Adjusting amplifier input

Instruction
ASA Amplifier Sensor Adaptation
Input bridge excitation voltage, transducer type and
input signal range
Syntax: ASA p1,p2,p3 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P1 Bridge excitation voltage

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 1V

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 2.5 V

ÁÁÁÁÁ
ÁÁÁÁÁ
P2
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Transducer
Full bridge
ÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Half bridge

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 LVDT

ÁÁÁÁÁ
ÁÁÁÁÁ
P3
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Input signal range (at Ub)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 mV/V (Ub=2.5 V) / 10 mV/V (Ub=1 V)
ÁÁÁÁÁ 2 ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
40 mV/V / 100 mV/V
ÁÁÁÁÁ
ÁÁÁÁÁ 3 ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
400 mV/V / 1000 mV/V
Effect: Bridge excitation voltage, transducer type and input
signal range have been set.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

A0241-5.3 en HBM
24 SCOUT55

Example: The SCOUT 55 is adjusted:


ASA1,2,2 (x)
0 (y)
The SCOUT 55 has been adjusted to 1 V bridge ex-
citation voltage, strain-gauge half bridge and input sig-
nal range 100 mV/V.

Instruction
ASA? Amplifier Sensor Adaptation Query
Output bridge excitation voltage, transducer type and
input signal range.
Syntax: ASA?(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Output the setting for the bridge excitation voltage,

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
the type of transducer and the input signal range

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Output the table of possible settings for the bridge
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
excitation voltage, the type of transducer and the

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ input signal range

Effect: The amplifier outputs the bridge excitation voltage, the


transducer type and the input signal range.
Response: ASA?0 (x)
q1,q2,q3 (y)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Bridge excitation range

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3
Transducer type
Input voltage range

Example: ASA?0 (x)


1,2,2 (y)
The SCOUT 55 has been adjusted to 1 V bridge ex-
citation voltage, strain-gauge half bridge and input sig-
nal range 1000 mV/V.
Response: ASA?1 (x)
q1,q2,q3 (y)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Table containing the possible setups
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Bridge excitation range

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2 Transducer type

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3 Input voltage range

HBM A0241-5.3 en
SCOUT55 25

Example: ASA?1 (x)


Amplifier response:
”01.002.50”, ”123”, ”123”(y)

4.2.4.2 Change filter setup

Instruction
ASF Amplifier Signal Filtering
Input cutoff frequency and filter characteristics
Syntax: ASF p1,p2(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ... n
Filter code
Code for cutoff frequency
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
(corresponds to the index of the frequency table

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
that can be output using ASF?0 instruction)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p2

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Filter characteristic
Bessel
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Butterworth

Effect: A frequency value and a filter characteristic is assigned


to the low-pass filter.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error

Example: Input of frequency value and the filter characteristic:


ASF 10,1(x)
0 (y)
The filter is adjusted to 40 Hz cutoff frequency and
Bessel characteristic.
Instruction
ASF? Amplifier Signal Filtering Query
Output cutoff frequency and filter characteristic.
Syntax: ASF?p1(x)
Parameter:
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Filter index
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Currently adjusted cutoff frequency

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Frequency table (Bessel and Butterworth)

A0241-5.3 en HBM
26 SCOUT55

Effect: Output off the cutoff frequency and filter characteristic.


Response: if p1 =0
q1,q2 (y)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Index of the cutoff frequency

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2 Filter characteristic (1=Bessel, 2= Butterworth)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Table containing the possible filter frequencies (Bessel/
Butterworth).
Example: Table containing the filter frequencies
ASF?1 (x)
”0.050 0.100 0.200 0.500 1.250 2.500 5.000 10.00
20.00 40.00 100.0 200.0 400.0”,
”5.000 10.00 20.00 40.00 80.00 200.0 500.0” (y)
The following table contains the possible cutoff fre-
quencies with Bessel and Butterworth characteristics
(every item in the table is 5 characters long).
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁ
p1
ÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁ
Bessel frequencies Butterworth frequencies

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
(Hz) (Hz)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 0.050 5.000

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁ
3
ÁÁÁÁÁÁÁÁÁÁÁÁ
0.100
ÁÁÁÁÁÁÁÁÁÁÁÁ
0.200
10.00
20.00
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 0.500 40.00

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
5 1.250 80.00

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
6 2.500 200.0

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
7
ÁÁÁÁÁÁÁÁÁ
8
ÁÁÁÁÁÁÁÁÁÁÁÁ
5.000
ÁÁÁÁÁÁÁÁÁÁÁÁ
10.00
500.0

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
9 20.00

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
10 40.00

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
11 100.0

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
12
ÁÁÁÁÁÁÁÁÁ
13
ÁÁÁÁÁÁÁÁÁÁÁÁ
200.0
ÁÁÁÁÁÁÁÁÁÁÁÁ
400.0
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Instruction
MTC Motion Control
Define steady-state display (measurements/tolerance
band/output).
Syntax: MTC p1,p2,p3 (x)

HBM A0241-5.3 en
SCOUT55 27

ÁÁÁÁÁÁ
Parameters:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Number of measurements
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Steady-state display switched off

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ... 255 Number of measurements; in conjunction with the

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
sample rate gives the maximum time span (see
table of filter frequencies)
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p2 Tolerance band

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
In Digits referred to the display units (full scale)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p3 Output status of steady-state display

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 No status output via ”WARNING”
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Status output via ”WARNING”
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: The steady-state display function is set.
Example: MTC 200,10,1 (x)
0 (y)

Assuming:
Filter setting: f<2.5 Hz = Sample rate 1200
measurements/s.
Display FSD: 100.00 N
The steady-state display is set:
If 200 measurements are located within a tolerance
band of 0.1 N (10 digits from 10000), the steady-
state display is activated. In addition the status is
output via ”WARNING”.
Instruction
MTC? Motion Control Query
Steady-state display output
Syntax: MTC?p1(x)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Parameter:

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Steady-state display settings
Steady-state display status
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: Output of steady-state display settings
Response: If p1=0; output of steady-state display settings
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1,q2,q3 (y)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Number of measurements
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2 Tolerance band in display units

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3 Status output ”WARNING”

A0241-5.3 en HBM
28 SCOUT55

ÁÁÁÁÁÁ
ÁÁÁÁÁÁ q1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
If p1=1; q1 indicates the steady-state display status
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Steady-state display status
ÁÁÁÁÁÁ q2 ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
No steady-state; conditions not fulfilled
ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3 Steady-state; conditions fulfilled

Example: MTC?0 (x)


0,0,0 (y)
The steady-state display is not activated. The steady-
state display status is not output via ”WARNING”.

4.2.4.3 Adjust the autocalibration

Instruction
ACL Autocal
Switches the autocalibration on/off
Syntax: ACL p1 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Automatical calibration

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Switch off
Switch on

Effect: Switches the autocalibration on/off.


Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Meaning
Instruction executed
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Example: ACL1 (x)
0 (y)
A calibration is started and cyclic autocalibration is
switched on. This means that at intervals of 5 minutes,
measurement is interrupted and the amplifier cali-
brated. If such an interruption is not desired during
measurement, automatic calibration must be switched
off.

Instruction
ACL? Autocal Query
Switch status of autocalibration
Syntax: ACL ? (x)

HBM A0241-5.3 en
SCOUT55 29

Parameter: None
Effect: Output of the autocalibration status.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Status

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Autocalibration is off
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Autocalibration is on

Example: ACL? (x)


1 (y)
Autocalibration is switched on.
Instruction
CAL Calibrate
Calibration
Syntax: CAL (x)
Parameter: None
Effect: A single calibration procedure is started.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error

Example: CAL (x)


0 (y)
One single calibration procedure is started.
Note: This instruction starts a calibration, thus allowing com-
munication to be continued only after 1 ... 3 s.

A0241-5.3 en HBM
30 SCOUT55

4.2.5 Adjusting parameters of the Group Calibration


4.2.5.1 Change engineering unit

Instruction
ENU Engineering Unit
Input unit
Syntax: ENU p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ... n
Engineering unit
Code for engineering unit
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
(see table)

Effect: The unit is set.


Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Example: ENU11(x)
0 (y)
kN has been set.

Instruction
ENU? Engineering Unit Query
Output of unit.
Syntax: ENU?p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Output engineering unit

ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Output of the currently set unit
Table of possible units
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: Output of the currently set engineering unit.
Response: q1 (y)

Example 1:ENU?0 (x)


11 (y)
kN has been set.

HBM A0241-5.3 en
SCOUT55 31

Example 2:ENU?1 (x)


” mV/V, V, g, kg, T, kT, TON, LB, oz, N, kN, bar, mbar,
Pa, PAS, HPas, KPas
PSI, µm, mm, cm, m, Inch, Nm, kNm, FTLB, INLB,
µm/m, m/s, m/ss, %, ‰, PPM s, MP, MN, A, mA ” (y)
Table of possible units and numbers.

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 p1 p1 p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 mV/V 13 mbar 25 kNm 37 MN

ÁÁÁ
2
ÁÁÁ
3
ÁÁÁÁÁÁÁÁ
V
ÁÁÁÁÁÁÁÁ
ÁÁÁ
g
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
14
ÁÁÁ
Pa
ÁÁÁÁÁÁÁÁ
15
ÁÁÁÁ
ÁÁÁÁÁÁÁ
PASÁÁÁÁ
26
27
FTLB
INLB
38
39
A
mA
ÁÁÁ
ÁÁÁÁÁ ÁÁÁÁÁ ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 kg 16 HPas 28 µm/m

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
5 T 17 kPas 29 m/s

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
6 kT 18 PSI 30 m/ss

ÁÁÁ
7
ÁÁÁ
8
ÁÁÁÁÁÁÁÁ
TON
ÁÁÁÁÁÁÁÁ
ÁÁÁ
LB
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ
19
ÁÁÁ
µm
ÁÁÁÁÁÁÁÁ
20
ÁÁÁÁ
ÁÁÁÁÁÁÁ
mmÁÁÁÁ
31
32
%

ÁÁÁ
ÁÁÁÁÁ ÁÁÁÁÁ ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
9 oz 21 cm 33 PPM

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
10 N 22 m 34 s

ÁÁÁ
11
12
ÁÁÁÁÁÁÁÁ
kN
bar
ÁÁÁÁÁÁÁÁÁÁÁÁÁ
23
24
ÁÁÁÁÁÁÁ
Inch

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Nm
35
36
”empty”
Mp

A0241-5.3 en HBM
32 SCOUT55

4.2.5.2 Setting indication adaptation


Instruction
IAD Indication Adaption
Input signal range, full scale val., dec. point, step width.
Syntax: IAD p1,p2,p3 (x)
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Full-scale without decimal point

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ (up to 200000)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
p2

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Decimal point
(Number of decimal digits 0 ... 5)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁ
p3

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Step width
(=minimum digital value, see table)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P3 Step width

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 2

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4
5
10
ÁÁÁÁÁÁÁÁÁ 5 20
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
6 50

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
7 100

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
8

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
9
200
500
ÁÁÁÁÁÁÁÁÁ 10 1000
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: The indication adaptation values are input with this in-
struction.
Note: For ”V” and ”mV/V” the indication adaptation is fixed.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

HBM A0241-5.3 en
SCOUT55 33

Example: IAD 10000,3,4 (x)


0 (y)
The indication adaptation is adjusted to:
Indication adaptation 10.000 and step width 0.010
Instruction
IAD? Indication Adaptation Query
Output signal range, full scale val., dec. point,
step width.
Syntax: IAD?(x)
Parameter: None
Effect: Output of currently set indication adaptation.
Response: q1,q2,q3 (y)
Parameter: see IAD instruction
Example: IAD? (x)
10000,3,4 (y)
The indication adaptation is set to:
Indication adaptation 10.000 and step width 0.010

4.2.5.3 Adjust zero balance


Instruction
CDW Calibration Dead Weight
Start zeroing/Input zero value (balance)
Syntax: CDW (x) or CDW p1(x)
Parameter: p1 (optional)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Zero value in mV/V

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Zero value in mV/V; related to the selected input
signal range (e.g. 2 mV/V)
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: The input value is stored in the amplifier zero-point
memory.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ

A0241-5.3 en HBM
34 SCOUT55

Example 1:Start zeroing


CDW (x)
0 (y)
The currently applied measuring value is zeroed.

Example 2:Input zero value 2.0000 mV/V (actual input signal


range 4 mV/V).
CDW 2.0000(x)
0 (y)
If the value read with CDW? 1 is transmitted for p1, the
currently applied measuring signal is zeroed.

Instruction
CDW? Calibration Dead Weight Query
Output zero value
Syntax: CDW?p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Zero value

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
1
Currently adjusted zero value

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁCurrently measured value

Effect: Use this instruction to output the zero value from the
memory or the currently applied measuring signal.
Response: q1 (y)

Example 1:CDW?0 (x)


3.256 (y)
Zero value of 3.256 mV/V is set.

Example 2:CDW?1 (x)


2.001 (y)
Currently applied measuring signal is output.
This signal is set to zero by CDW2.001.

HBM A0241-5.3 en
SCOUT55 35

4.2.5.4 Adjust Measuring range


Instruction
IMR Input Measuring Range
Input measuring range
Syntax: IMR p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Full scale value in mV/V

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Value in mV/V; related to input signal range

Effect: The measurement range is set.


Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

Example: IMR 2.0 (x)


0 (y)
Measuring range is adjusted to 2.0 mV/V.
Instruction
IMR? Input Measuring Range Query
Output full-scale measuring range
Syntax: IMR?p1(x)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Parameter:

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Full scale value
Currently adjusted measuring range in mV/V
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Current measurement signal in mV/V
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Maximum and minimum adjustable full scale val-
2

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ues

Effect: Output the adjusted measuring range.


Response: q1,q2 (y)
Example1: IMR?0 (x)
1.987 (y)
Measuring range has been set to 1.987 mV/V.

Example2: IMR?2 (x)


”4.0”,”0.2” (y)
With a selected input signal range of 2 mV/V the
maximum value is 4.0 mV/V and the minimum value is
0.2 mV/V.

A0241-5.3 en HBM
36 SCOUT55

4.2.5.5 Taring
Instruction
TAR Tare Instruction
Start taring/Input tare value
Syntax: TAR (x) or TAR p1(x)
Parameter: p1 (optional) or tare value in indication units
Effect: The signal is tared using this instruction. If no value is
sent in engineering units, the measurement being cur-
rently taken is tared.

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response:

ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
Meaning
Instruction executed
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

Example1: Start taring


TAR (x)
0 (y)
The signal present on the amplifier input is accepted as
the tare value.
Note: Taring is carried out by calculation and not by balanc-
ing of the input signal.
Example2: TAR200.0 (x) (Indication value 2000.0 kg)
0 (y)
Input value is written into the tare memory.

Instruction
TAR? Tare Value Query
Output tare value
Syntax: TAR?(x)
Parameter: None
Effect: Output of the tare value in indication units.
Response: q1 (y)
Tare value in indication units.
Example: TAR? (x)
200.0 (y)
The indication adaptation is adjusted to 2000.0 kN. The
tare value is 200.0 kN.

HBM A0241-5.3 en
SCOUT55 37

4.2.6 Adjust the parameters of the group Limit value 1 ... 4


Instruction
LIV Limit Value
Input limit switch settings
Syntax: LIV p1,p2,p3,p4,p5,p6,p7 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P1 Limit switch

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 1

ÁÁÁÁÁ
ÁÁÁÁÁ
2
3
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2
3
ÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 4

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P2 Limit monitoring

ÁÁÁÁÁ 0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
OFF
ON
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P3 Source input signal

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Gross value

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Net value

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 Peak value 1 (Maximum)

ÁÁÁÁÁ
ÁÁÁÁÁ
4
5
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Peak value 2 (Minimum)
Peak value 3 (Peak-Peak value)
ÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P4 Switch direction

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Switches if input signal is higher than the set level

ÁÁÁÁÁ
ÁÁÁÁÁ
2
p5
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Switches if input signal is lower than the set level

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Limit level in engineering units
ÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Value is stated in engineering units (e.g. kN)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁ
ÁÁÁÁÁ
p6
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Hysteresis value in engineering units

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Value is stated in engineering units (e.g. 100 kN);

ÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ always positive

ÁÁÁÁÁ
ÁÁÁÁÁ
p7
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Limit switch output logic
Active corresponds to ON
ÁÁÁÁÁ 2 ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Active corresponds to OFF

Effect:
ÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Use this instruction to adjust the limit value p1 to limit-
value monitoring, to input source p3, to switch direction
p4, to level value p5, the hysteresis value p6 and the
output logic p7.
Response:
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

A0241-5.3 en HBM
38 SCOUT55

Example: LIV1,1,3,1,100,10,1,1 (x)


0 (y)
Limit switch 1 is activated and the input signal SPWT
Max (maximum value) is assigned. The limit switch
switches when the 100 kN level is exceeded (corre-
sponding to the switch-on threshold) and has a hyster-
esis of 10 kN (switch-off threshold 90 kN). The control
output is active when the level is exceeded.

Instruction
LIV? Limit Value Query
Output limit switch parameters
Syntax: LIV? p1,p2 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Limit switch

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Output measured value of parameter p2 (in engi-
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ neering units)

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Currently adjusted settings GR1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Currently adjusted settings GR2
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 Currently adjusted settings GR3

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 Currently adjusted settings GR4

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P2

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Signal code, if p1=0
Currently applied gross signal in indication units
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Currently applied net signal in indication units
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 Currently measured max. value in indication units

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 Currently measured min. value in indication units

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
5 Currently measured peak-peak value in indication

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ units

Effect: The setting of limit switch p1 is output with this instruc-


tion.
Response: q1,q2,q3,q4,q5,q6,q7 (y)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Number of limit switch
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2 Limit monitoring ON/OFF

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3 Input signal for the limit switch

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q4 Switch direction positive/negative

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q5 Switching level of limit switch
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q6 Hysteresis value
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q7 Logic of control output

HBM A0241-5.3 en
SCOUT55 39

Example1: LIV?2 (x)


2,1,3,1,100,10,1,1 (y)
Limit switch 2 is activated and the input signal SPWT
Max (maximum value) is assigned. The limit switch
switches when the level exceeds 100 kN (corresponds
to the switch-on threshold). The hysteresis is 10 kN
(switch-off threshold 90 kN). The control output is ac-
tive when the threshold is exceeded.
Example2: LIV?0,3 (x)
200 (y)
The current value of peak value store PEAKV1Max is
200 kN.

4.2.7 Adjust the parameters of the group peak value


Instruction
PVS Peak Value Select
Input peak-value store setup
Syntax: PVS p1,p2,p3,p4 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P1 Peak-value store

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2
Maximum value
Minimum value
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 Peak-Peak value
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P2 Peak value determination (applies for all stores)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 OFF

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ON
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P3
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Input signal for peak value

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Gross

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Net

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P4 Envelope-curve function (applies for all stores)

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Envelope-curve function is switched off
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
00100 ... Time constant for envelope-curve function in ms
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁ
60000
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: Use this instruction to adjust the functions for peak-
value store p1.
Response:
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning
ÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

A0241-5.3 en HBM
40 SCOUT55

Example: PVS1,1,1,0 (x)


0 (y)
Peak value store 1 (Maximum) has been set to gross
signal. All peak-value stores are activated; the enve-
lope-curve function is deactivated.

Instruction
PVS? Peak Value Select Query
Output peak-value store setup
Syntax: PVS?p1(x)
Parameter: p1
Code for the peak-value store (see PVS instruction).
Effect: Use this instruction for output of the peak-value store
p1 setup.

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response: q1,q2,q3,q4 (y)

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Code of the peak-value store
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q2 Peak-value determination on/off
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q3 Code for the input signal
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q4 Time constant for envelope-curve function in ms

Example: PVS?1 (x)


1,1,1,0 (y)
Peak value store 1 (Maximum) has been set to gross
signal. All peak-value stores have been activated; the
envelope-curve function has been deactivated.

Instruction
CPV Clear Peak Value
Clear peak-value store
Syntax: CPV (x)
Parameter: None
Effect: Use this instruction to clear the peak-value stores.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

HBM A0241-5.3 en
SCOUT55 41

Example: CPV (x)


0 (y)
Peak value store 1 (Max), Peak value store 2 (Min) and
Peak value store 3 (MinMax) have been cleared.
Note: After the peak-value stores have been cleared, the out-
put signal from memories 1 and 2 corresponds to the
output signal. Memory 3 (Min Max) has the value zero.

4.2.8 Adjust the parameter of the group Inputs/Outputs


4.2.8.1 Select amplifier input-signal
Instruction
ASS Amplifier Signal Select
Select amplifier input-signal
Syntax: ASS p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Input source

ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Internal zero signal
Internal calibration signal
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Measuring signal

Effect: Specification of the amplifier input-signal.


Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error

Example: The SCOUT 55 is adjusted to:


ASS 0(x)
0 (y)
Amplifier input is switched to internal zero signal.
Note: This instruction starts a calibration, thus allowing com-
munication to be continued only after 1 ... 3 s.
To continue measurement: enter p 1=2

A0241-5.3 en HBM
42 SCOUT55

Instruction
ASS? Amplifier Signal Select
Output type of input signal.
Syntax: ASS?(x)
Parameter: None
Effect: Output of the type of amplifier input signal.

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response: q1 (y)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Amplifier input-signal source

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Amplifier input set to ZERO

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Amplifier input set to CAL
ÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Amplifier input set to MEAS

Example: ASS? (x)


2 (y)
Amplifier input is set to MEAS (measuring signal).

4.2.8.2 Adjust the analogue output


Instruction
OPS Output Path Select
Assign analogue output and select operating mode
Syntax: OPS p1,p2 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P1 Output signal

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Gross signal assigned to analogue output

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 Net signal assigned to analogue output

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 PVS1 (Maximum-val.) assigned to analogue output

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
4 PVS2 (Minimum-val.) assigned to analogue output
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
5 PVS3 (peak-peak val.) assigned to analogue
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
output

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
P2 Mode of analogue output (U / I)*

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Analogue output switched off
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 "10 V (U) / "20 mA (I)
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 No Fct. (U) / 4 ... 20 mA (I)

Effect: A signal is assigned to the analogue output and the op-


erating mode is set.
Note: The selection of voltage or current for the analogue
output is made using jumpers. The setting is described
in Part 1 of the operating manual.

HBM A0241-5.3 en
SCOUT55 43

Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error

Example: OPS1,1 (x)


0 (y)
The gross signal is assigned to the analogue output.
"10 V is set as the operating mode.
Assumption: Voltage is defined as the analogue output
signal using jumpers.)
Instruction
OPS? Output Path Select Query
The input signal for the analogue output and the oper-
ating mode are output.
Syntax: OPS?p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Analogue output: Signal and mode

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Currently assigned input signal

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 Operation mode of analogue output:

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
voltage or current set via jumpers

Effect: The currently assigned input signal of the analogue


output or the selected operating mode is output.
Response: q1,q2 (y)
q2 corresponds to parameter p2 (see OPS instruction )
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
q1 Operating mode (via jumper)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2
Voltage set
Current set

Example: OPS?0 (x)


2 (y)
Net signal is assigned to analogue output.
Example: OPS?1 (x)
2,2 (y)
Current output has been set (with jumpers on
PCB); mode 4 ... 20 mA has been set.

A0241-5.3 en HBM
44 SCOUT55

4.2.8.3 Adjust remote control


Instruction
LOR Local / Remote
Local-to-Remote switching
Syntax: LOR p1 (x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1 Status

ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1
Remote
Local
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Effect: Switching to remote control for specific amplifier func-
tions via remote-control inputs.
Response:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
?
Instruction executed
Error

Example: LOR1 (x)


0 (y)
Local control is switched on, i.e. all setup functions for
amplifier parameters via remote control inputs are
blocked.

Instruction
LOR? Local / Remote Query
Query local / remote status
Syntax: LOR? (x)
Parameter: None
Effect: Output of local / remote-control status.
Response: q1 (y) corresponds to p1 (see LOR instruction)
Example: LOR? (x)
0 (y)
Remote control is switched on, i.e. all setup functions
for amplifier parameters via remote control inputs are
possible.

HBM A0241-5.3 en
SCOUT55 45

4.2.8.4 Setting the control contact assignment


Instruction
RFP Remote Function Programming
Assignment of remote functions
Syntax: RFPp1,p2 (x)
Parameter: p1 corresp. to the contact number (1 ... 6)

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p2 corresp. to the function code (see table)

ÁÁÁÁÁÁ
Code
ÁÁÁÁÁÁ
0
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
NOP
Function
No Function
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
1 ACAL Autocalibration

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
2 TARA Start taring

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
3 CPV1 Instantaneous value/peak value 1

ÁÁÁÁÁÁ
4
ÁÁÁÁÁÁ
5
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
HLD1
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
CPV2
Peak value 1 Hold
Instantaneous value/peak value 2
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
6 HLD2 Peak value 2 Hold

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
7 NULL Start zeroing

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
8 PRNT Start printing

ÁÁÁÁÁÁ
9
ÁÁÁÁÁÁ
10
ÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
PAR1 Bit for binary interrog. of data set 1 ... 4
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
PAR2 Bit for binary interrog. of data set 1 ... 4
ÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
The default assignment after a device ”SET UP” is ”No
function” for all contacts. The four sets of parameters
can be called by binary (00 to 11) using the functions
PAR1 and PAR2.
Effect: The effect of the control contacts of the amplifier is
specified.

ÁÁÁÁÁÁÁÁ
Response:
ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Response Meaning
ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0 Instruction executed

ÁÁÁÁÁÁÁÁ
Example: RFP 2,1 (x)
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
? Error

0 (y)
Remote control 2 has been assigned to Autocal
(ACAL).
Note: Remote-to local switching is always
possible, even if the device is in Local status.

A0241-5.3 en HBM
46 SCOUT55

Instruction
RFP? Remote Function Programming Query
Query remote function assignment
Syntax: RFP?p1(x)
Parameter:
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
p1

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
0
1 ... 6
Output table containing all possible functions

ÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁÁ
Output remote-function assignment

Effect: Output of the assignment of the remote functions


Response: q1 (y)
Example1: RFP?2 (x)
1 (y)
Remote control contact 2 has been set to autocal.
(ACAL).

Example2: RFP?0 (x)


”NOPACALTARACPV1HLD1CPV2HLD2NULLPRNT-
PAR1PAR2” (y)

HBM A0241-5.3 en
SCOUT55 47

5 Keywords

A Input measuring range, 35


Input zero value, 33
Amplifier input−signal, 41
Interface configuration, 7
Analogue output, 42
Autocalibration , 28
L
B Limit switch, 38
Limit−value monitoring, 37
Baudrate, 6 , 14
Local, 44
Bridge excitation voltage, 23
LVDT, 23

C
Calibration, 29
M
Measured value, 21
Cutoff frequency, 25

D P
Parity, 6
Dec. point, 32
Pin assignment, 7
Define printout format, 15
Printer, 7

E
Echo, 8
R
Remote, 44
Engineering unit, 30
Remote−control state, 8
Envelope−curve, 40
RS−232, 7
Envelope−curves, 39
RS−232−C, 8

F
Factory set up, 6
S
Serial interface, 6
Factory settings, 17
Start zeroing, 33
Filter characteristics, 25
START−Bit, 6
Full bridge, 23
Step width, 32
Full scale value, 32
STOP−Bit, 6
Store amplifier setup, 17
H
Subminiature jack, 7
Half bridge, 23

T
I
Taring, 36 , 45
Indication adaptation, 33
Transducer type and input signal range,
Input limit switch settings, 37 23

A0241-5.3 en HBM
48 SCOUT55

6 Index
Code Instruction Page Code Instruction Page
A
ACL Autocal . . . . . . . . . . . . . . . . . . . . . . . . . . 29
O
ACL? Autocal Query . . . . . . . . . . . . . . . . . . . . 29 OPS Output Path Select . . . . . . . . . . . . . . . . 43
AID? Amplifier Identification Query . . . . . . . 15 OPS? Output Path Select Query . . . . . . . . . . 44
ASA Amplifier Sensor Adaption . . . . . . . . . 24
ASA? Amplifier Sensor Adaption Query . . . 25 P
ASF Amplifier Signal Filtering . . . . . . . . . . . 26 PFS Print Format Select . . . . . . . . . . . . . . . 15
ASF? Amplifier Signal Filtering Query . . . . . 26 PFS? Print Format Select Query . . . . . . . . . 16
ASS Amplifier Signal Select . . . . . . . . . . . . 42
PVS Peak Value Select . . . . . . . . . . . . . . . . 40
ASS? Amplifier Signal Select . . . . . . . . . . . . 43
PVS? Peak Value Select Query . . . . . . . . . . 41
B
BDR Baud Rate . . . . . . . . . . . . . . . . . . . . . . . 13 R
BDR? Baud Rate Query . . . . . . . . . . . . . . . . . 14 RFP Remote Function Programming . . . . . 46
C RFP? Remote Function
Programming Query . . . . . . . . . . . . . . 47
CAL Calibrate . . . . . . . . . . . . . . . . . . . . . . . . 30
CDW Calibration Dead Weight . . . . . . . . . . . 34 S
CDW? Calibration Dead Weight Query . . . . . 35
COF Change Output Format . . . . . . . . . . . . 20 SNR? Output serial number of device . . . . . 14
COF? Change Output Format Query . . . . . . 22 STP Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
CPV Clear Peak Value . . . . . . . . . . . . . . . . . 41
T
D TAR Tare Instruction . . . . . . . . . . . . . . . . . . . 37
DCL Device Clear . . . . . . . . . . . . . . . . . . . . . 11
TAR? Tare Value Query . . . . . . . . . . . . . . . . . 37
E TDD Transmit Device Data . . . . . . . . . . . . . 18
ENU Engineering Unit . . . . . . . . . . . . . . . . . . 31 TDD? Transmit Device Data Query . . . . . . . 19
ENU? Engineering Unit Query . . . . . . . . . . . . 31
ESR? Standart Event Status Register . . . . . 12
I
IAD Indication Adaption . . . . . . . . . . . . . . . 33
IAD? Indication Adaption Query . . . . . . . . . 34
IMR Input Measuring Range . . . . . . . . . . . . 36
IMR? Input Measuring Range Query . . . . . . 36
L
LIV Limit Value . . . . . . . . . . . . . . . . . . . . . . . 38
LIV? Limit Value Query . . . . . . . . . . . . . . . . . 39
LOR Local / Remote−Local . . . . . . . . . . . . . 45
LOR? Local / Remote Query Local . . . . . . . 45
M
MDD Memory Device Data . . . . . . . . . . . . . . 16
MDD? Memory Device Data Query . . . . . . . . 17
MSV? Measuring Signal Value Query . . . . . 22
MTC Motion Control . . . . . . . . . . . . . . . . . . . 27
MTC? Motion Control Query . . . . . . . . . . . . . 28

HBM A0241-5.3 en
SCOUT55 49

A0241-5.3 en HBM
50 SCOUT55

HBM A0241-5.3 en
Modifications reserved.
All details describe our products in general form only.They are
not to be understood as express warranty and do not constitute
any liability whatsoever.