Patient Monitor Quick Guide


Thank you for purchasing Mindray Patient Monitor.
We have included this convenient Quick Guide to help you with the basic operations. For more details, please refer to the
Operator’s Manual.

Note
 Before attempting to install, operate or adjust this product, please save and read the Quick Guide completely.
 This Quick Guide is based on the maximum configuration and therefore some contents may not apply to your
monitor. If you have any question, please contact us.

© 2011-2014 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
Content in this guide is subject to change without prior notice.
T1/BeneView T1 Overview
1
Alarm lamp

Display screen
Ambient light sensor
External power indicator
Battery indicator

Power on/off switch Sliding switch

Connector forTemp probe

DC power connector

Connector for
NIBP cuff
Main unit multi-pin connector
Connector for Connector for
IBP sensor ECG cable
Infrared filter
Multifunctional Connector for
connector SpO2 sensor
Contact
Speaker
In Use with the T1 Handle and T1 Docking Station
2

T1 docking station rear view


1. AC power input
2. Equipotential grounding terminal
3. VGA connector
4. External device connector
5. USB connector
6. Network connector

1 2 3 4 5 6
Display Screen
3

1 2 3 1. Alarm Symbols
2. Patient Information/Technical Alarm Area
3. Physiological Alarm Area
4. Waveform Area and Parameter Area A
5. Parameter Area B
6. QuickKeys Area

4 7. Prompt Message Area


6
The Quickkeys allows you to fast access the fol-
lowing functions:
Start or stop NIBP measurements
5
Enter alarm silenced status

Enter alarm paused status

Enter the main menu


7
Basic Operations
4

Turning on the Monitor Turning off the Monitor


1. Before you start to make measurements, check the patient Before turning off the monitor,
monitor for any mechanical damage and make sure that all 1. Confirm that the patient monitoring is finished.
external cables, plug-ins and accessories are properly con-
2. Disconnect the patient cables and sensors from the patient
nected.
monitor.
2. Connect the monitor with the DC adapter or T1 docking sta-
3. Make sure to save or clear the patient monitoring data as re-
tion. If you run the patient monitor on battery power, ensure
quired.
that the battery is sufficiently charged.
Then press and hold the power on/off switch to turn off the
3. Press the power on/off switch on the monitor’s front.
monitor.

Starting Monitoring
Note
1. Decide which measurements you want to make.
2. Check that the patient cables and sensors are correctly con-  Make sure that the operating environment of the patient
nected. monitor meets the specific requirements. Otherwise
3. Check that the patient settings such as [Patient Cat.], [Paced], unexpected consequences, e.g. damage to the equipment,
etc, are appropriate for your patient. could result.
 Do not use the patient monitor for any monitoring
Refer to the Operator's Manual for details of how to perform the procedure on a patient if you suspect it is not working
measurements you require. properly, or if it is mechanically damaged. Contact your
service personnel or us.
Basic Operations
5

Using the Touchscreen Using the External Display


You can select screen items by touching them directly on the T1 can be connected to a external display through the VGA con-
patient monitor’s screen. To avoid misoperation, you can lock the nector of the T1 docking station. When the external display is
touchscreen. The touchscreen is locked automatically if no opera- connected, you can monitor a patient either through the T1 or
tion is detected within 60 seconds. Making the sliding switch through the external display. You can switch between the T1 and
rightward can unlock or lock the screen. external display with the sliding switch on the front panel of T1.

The external display displays differently with T1, for its screen size
Calibrating the Touchscreen is larger. The following screens or functions can only be viewed
and operated on the external display:
The touchscreen should be calibrated if you cannot select screen
items. To calibrate the touchscreeen:  Minitrends Screen
1. Select [Main Menu]→[Maintenance >>]→[User Main-  OxyCRG Screen
tenance >>]→enter the required password→[Cal. Touch-  View Others Screen
 ECG 7-Lead Half-Screen
screen]. The symbol will, in turn, appear at different
 PAWP Screen
positions of the screen.
 Calculations
2. Select each as it appears on the screen. After the calibra-  Spider view
tion is completed, the message [Screen Calibration Com-  Hemodynamic parameters
pleted!] is displayed.  Department change
3. Select [Ok] to confirm the completion of the calibration.
Using the Mouse and Keyboard
When connected to the external display, T1 can connect a mouse
and a keyboard through the USB connector of the T1 docking sta-
tion. When a mouse is in use, only the left mouse-button can be
used. The right mouse-button is disabled.
Basic Operations
6

Setting up a Monitor Setting the Date and Time


In situations where you install a patient monitor or change the 1. Select [Main Menu]→[Maintenance >>]→[System Time
patient monitor’s application site, you need to setup the patient >>].
monitor as follows: 2. Set the date and time.
1. Select [Main Menu]→[Maintenance >>]→[User Mainte- 3. Select [Date Format] and toggle between [yyyy-mm-dd],
nance >>]→enter the required password. [mm-dd-yyyy] and [dd-mm-yyyy].
2. In the [User Maintenance] menu, select [Monitor Name], 4. Select [Time Format] and toggle between [24h] and [12h].
[Department] and [Bed No.] and change their settings.

Note
Changing Language
1. Select [Main Menu]→[Maintenance >>]→[User Mainte-  Changing date and time will affect the storage of trends
nance >>]→enter the required password. and events and may cause data missing.
2. In the [User Maintenance] menu, select [Language] and
then select the desired language.
Entering the Parameter Setup Menu
3. Restart the patient monitor.
You can select the waveform area or parameter area to enter cor-
responding parameter setup menu.
Adjusting the Screen Brightness
1. Select [Main Menu]→[Screen Setup >>]→[Brightness]. Entering the Alarm Setup Menu
2. Select the appropriate setting for the screen brightness. 10 is
Select [Main Menu]→[Alarm Setup >>] to enter the Alarm
the brightest, and 1 is the least bright. If you select [Auto],
setup menu.
Screen brightness will be adjusted automatically.
Basic Operations
7

Adjusting Alarm Volume


1. Select [Main Menu]→[Alarm Setup >>]→[Others].
2. Select [Alm Volume] and then select the appropriate vol-
ume.
3. Set [High Alarm Volume].
4. Set [Reminder Vol]. Area A

Adjusting Key Volume


1. Select [Main Menu]→[Screen Setup >>].
2. Select [Key Volume] and then select the appropriate volume. Area B

Choosing the Screen


Adjusting QRS Volume
By selecting [Main Menu]→[Screen Layout >>]→[Choose
The QRS tone is derived from either the HR or PR, depending on Screen], you can choose either:
which is currently selected as the alarm source in [ECG Setup] or
 Normal Screen, or
[SpO2 Setup].
 Big Numerics screen, or
1. Select the the ECG parameter window→[Others >>], or the  ECG 7-Lead Full-Screen if 5-lead or 12-lead ECG is selected, or
SpO2 parameter window.  ECG 12-Lead Full-Screen if 12-lead ECG is selected, or
2. Select [QRS Volume] or [Beat Vol] and then select the ap-  PiCCO Screen if PiCCO module is configured.
propriate volume.
Basic Operations
8

Changing the Screen Layout


You can enter the [Screen Setup] window by selecting the
[Main Menu]→[Screen Layout >>]→[Screen Setup]. In this
window, you can allocate the positions of the parameters and
waveforms.
The ECG parameter and the first ECG waveform always display in
the first row. The configurable areas can be classified as Area A,
and Area B.
 In Area A, you can choose to display the parameters (having
waveforms) and their waveforms. Each parameter and the
associated waveform are displayed in the same row.
 In Area B, you can choose to display all parameters except
ECG. Associated waveforms will not be displayed.

The screen can automatically adjust to ensure the best view


based on your screen setup.

Note

 The parameters whose positions are not allocated in the


[Screen Setup] window will not be displayed. However, the
monitor can still give alarms of these parameters.
Care and Cleaning
9
Use only the substances approved by us and methods listed in this manual to clean and disinfect your equipment. Warranty
does not cover damages caused by unapproved cleaning substances or methods.

Recommended cleaning agents are:


 Sodium hypochlorite bleach (diluted)
 Hydrogen peroxide (3%)
 Ethanol (70%)
 Isopropanol (70%)

To clean your equipment, follow this procedure:


1. Shut down the equipment, and disconnect the power cord and other cables, and then remove the batteries.
2. Clean the display screen using a soft, clean cloth dampened with a glass cleaner.
3. Clean the exterior surface of the equipment using a soft cloth dampened with the cleanser.
4. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
5. Dry your equipment in a ventilated, cool place.

Disinfection may cause damage to the equipment and is therefore not recommended for this patient monitor unless other-
wise indicated in your hospital’s servicing schedule. Cleaning equipment before disinfecting is recommended.

The recommended disinfectants include: ethanol 70%, isopropanol 70%, Perform® classic concentrate OXY (KHSO4 solution).
BeneView T1 T1
Manufacturer: Manufacturer:
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, Shenzhen 518057 P.R. China Nanshan, Shenzhen 518057 P.R. China
Tel: +86 755 81888998 Tel: +86 755 81888998
Fax: +86 755 26582680 Fax: +86 755 26582680
www.mindray.com
Distributor:
Mindray DS USA, Inc.
800 MacArthur Boulevard, Mahwah, New Jersey 07430 USA
Tel: 1.800.288.2121, 1.201.995.8000

Comply with the requirements of the Council


Directive 93/42/EEC.
Comply with the Council Directive 2011/65/EU

Shanghai International Holding Corp. GmbH (Europe)


Eiffestraβe 80, 20537 Hamburg, Germany