Documente Academic
Documente Profesional
Documente Cultură
Error Analysis
Objectives
Introduction
The major problem with any measuring instrument is the error. Hence, it is necessary to
select the appropriate measuring system which minimizes the error. So, error analysis in any
measurement system is very critical.
Calibration
The set of criteria defined for instruments which are used to measure quantities that are
slowly varying with time or are mostly constant i.e. do not vary with time are called as Static
Characteristics.
But when quantity changes rapidly with time we define the relations between input and
output in terms of differential equations which are called as Dynamic Characteristics.
Precision is the degree of closeness with which reading is produced again and again for
same value of input quantity. This has no relation with the true as the reading must be
consistent even though it may be inaccurate.
Sensitivity
Charge in output quantity per unit charge in input quantity. Sensitivity must be high
because even for a small change in the value of Input quantity the output produced by the
instrument must be high so that it is observable to the naked eye.
qo
S
qi
Resolution
It is defined as smallest charge in input which can be measured by an instrument. It can
be improved by calibration but not sensitivity. As for example, the resolution of the cm scale
commonly used is 1mm, so resolution refers to the minimum marking on the instrument.
Threshold
Minimum input required to get measurable output by an instrument. If the Input
quantity is less than Threshold value, then the instrument may give an output but the will be
very small or negligible so the instrument will be considered as unresponsive. Threshold
must be as small as possible so that we are able to measure even small values of Input
Quantity.
Drift
Variation in output of an instrument from the desired value for a particular value of input.
Perfect Reproducibility means no Drift as no drift means measured values do not vary with
time. Drift can be classified in the following categories:
Zero drift
If the whole calibration shifts gradually due to permanent set or due to warning up of
electronic tube circuits.
Zonal drift
When drift occurs over a portion of span of an instrument, it is called Zonal drift. It
can be considered as a combination of Zero drift and Span drift.
Drift is mainly due to environmental factors like electric and magnetic fields.
Reproducibility
Degree of closeness in which a given value is measured with an instrument over a
specified period of time. Environmental conditions may not be same.
Repeatability
Degree of closeness in which input quantity is measured again and again for a given set of
reading of an instrument. Environmental conditions are kept same.
Note: Quantities like accuracy, sensitivity and reproducibility are desirable but drift, static
error and dead zone are undesirable.
a. Gross error: Completely due to human negligence i.e. during loose connections, reading
the values, taking the values, during calculations and memorizing the values.
b. Systematic error: These errors are common for all observers and constant with respect to
same conditions.
E.g: Due to shadow of pointer we may not be able to take the reading correctly and this is
common for all observers.
Systematic Error
Instrumental Error: These errors are introduced due to Instruments being faulty. The
error in an Instrument may be induced due to ageing effect and hysteresis.
Environmental Error: These errors are introduced when certain system’s parameters
change due to variable environmental conditions like Temperature, Pressure and
Humidity.
Observational Error: These errors are introduced when user is not able to observe
the position of the pointer accurately which may be due to Parallax Error.
c. Random errors: These are due to unidentified causes and they may be positive or
negative, charges with respect to time but effect is very less and can be neglected.
When accuracy is measured in terms of error, then it is called as guaranteed accuracy error
(GAE) or Limiting error (LE).
GAE is always specified by the manufacturer with respect to the Full Scale Deflection and it is
always constant in value whereas Limiting Error is calculated from GAE by the customer
based on True Value of the measurement and it varies as the true value of the measurement
varies.
Solved Examples
Problem: A (0-250) V voltmeter has a guaranteed accuracy of 2 per cent of full scale reading.
The voltage measured by this voltmeter is 150V. The limiting error in percentage is?
Loading effect
When a voltmeter is connected across a resistance of, whose voltage is to be measured, then
voltmeter draws a current for its own winding, so it acts as a load on the system and due to
this there is an error in the reading of Voltmeter. This is termed as Loading Effect.
So measured value of resistance is less than its actual value. To eliminate Loading Effect,
Voltmeter must be ideal that is its Resistance must be Infinite. So, higher the resistance and
hence sensitivity of Voltmeter, lesser is the Loading Effect.
Hysteresis error
For an instrument, for ascending value of quantity, the meter shows less than actual value
and for descending value it shows more than actual value, so for same input then it is called
Hysteresis error.
Loading Curve is when we traverse the curve in positive x-direction and Unloading Curve is
when we traverse the curve in negative x-direction.
Electrical Hysteresis
During loading and unloading, B is not same.
Mechanical Hysteresis
During loading and unloading, stress is not same.
Quantification of Error
Absolute Error
Absolute error is the difference between measured value of a quantity and true value of
same quantity. This error has the same units as that of measured quantity.
A Am At o
A absolute static error of quantity ' A '
Am measured value of quantity ' A '
A t True value of quantity ' A '
Relative error
Relative error is the ratio of absolute error in the measurement of a quantity to the true value
of same quantity. This error is unit-less.
AbsoluteError A o Am A T
r
Truevalue AT AT AT
Am
AT
1 er
Am A T (1 r ); A T Am 1 r
Percentage Error
When absolute error is expressed as a percentage of true value of a quantity then it is
termed as Percentage Error. This error is also unit-less.
Am A t
% r 100 A 100 r 100
At At
Composite Errors
When a quantity is expressed as function of other independent quantities like Power can be
expressed as product of Voltage and Current, then error in measurement of dependent
quantity can be calculated from the errors in reading of independent quantities. This is
termed as Composite Error. The errors computed in this section are Limiting Errors.
Some of the basic algebraic operations are illustrated below:
Sum of quantities
n
X Xi
i1
X (X1 X 2 ..... Xn )
X X X X X
%error x 100% 1 . 1 2 . 2 ..... x100%
X X1 X X2 X
Hence, absolute error is sum of absolute errors in all quantities.
Difference of Quantities
X X1 X 2
X X1 X2
For maximum error or Limiting Error
X X1 X2
X X X X X X X
1 2 1 1 2 2
X X X X1 X X2 X
Hence, absolute error is sum of absolute errors in all quantities.
Product of Quantities
X X1 .X2 .X3
log X log X1 log X2 log X3
1 1 X1 1 X 2 1 X3
. . .
X X1 X X 2 X X3 X
X X X X
1 2 3
X X1 X2 X3
Hence, relative error is sum of relative errors in all quantities.
Division of Quantities
X1
X
X2
logX logX1 logX2
X 1 X1 1 X2
X X1 X X2 X
For maximum error or Limiting Error
X X X
1 2
X X1 X2
Hence, relative error is sum of relative errors in all quantities.
Note: To find Limiting Errors in addition and subtraction operation, we add absolute errors
whereas Relative Errors or Percentage Errors are added in multiplication and division
operation.
Composite factor
If a quantity X is expressed as shown below:
X1m Xn2
X
Xp3
Then, relative error in X is given by,
X X X X
m 1 n 2 p 3
X X1 X2 X3
Statistical Analysis
Certain statistical parameters can also be defined when we are taking multiple readings for
the same measurement. Some of the statistical parameters are listed below:
1) Mean Value
X
X1 X 2 ...... Xn
X i
n n
Where, Xi ith measurement
n = number of readings
2) Deviation
Deviation for a particular reading is defined as departure of that reading from the mean
value.
di (Xi X)
3) Mean Deviation
n di
di
i1 n
It indicates precision of instrument, as for more precise measurement the mean deviation
must be small in magnitude so that readings are clustered together.
4) Standard Deviation ( )
Standard Deviation for a set of data signifies the spread of data points around the mean
value. Less standard deviation signifies less spread and a sharper distribution.
2
n di
For n>20; i 1 n
2
n di
For n<20; (n 1)
i 1
5) Variance
V= 2
6) Precision Index
Higher the standard deviation, more is the data spread around the mean value and lesser is
the precision of the measurement. So precision index is proportional to the reciprocal of
standard deviation.
1
h
1
h
2
7) Probable Error
It is the possible error in the measurement of a quantity.
r= 0.6745σ
0.4769
r=
h
2 2 2
x 2 x 2 x 2
rx rx rx .... r
x xn
x 1 x 2
1 2 n
Solved Examples
Given w xy z ,
Problem: The resistance of a circuit is found by measuring current flowing and the power
fed into the circuit. If the limiting errors in the measurement of power and current are 1.5%
and 1.0% respectively, the limiting error in the measurement of resistance will be?
Problem: The total current I = I1 + I2 in a circuit is measured as I1 = 150 1A, I2 = 250 2A,
where the limits of error are given as standard deviations. I is measured as?
Problem: Four ammeters M1, M2, M3 and M4 with the following specifications are available
M1 20
M2 PMMC 10 0.20
Electro-dynamic
M3 5 0.50
Moving-iron
M4 1 1.00
0.2
M2: Error x10 0.02A
100
0.5
M3: Error x5 0.025A
100
1
M4: Error x1 0.01A
100
Problem: Resistances R1 and R2 have, respectively, nominal values of 10 and 5, and
tolerances of 5% and 10%. The range of values for the parallel combination of R1 and R2
is?
R2 Ranges 5 10
100 0.5 R 2 5.5 4.5
R1 xR 2
For Parallel Combination, R eq
R1 R 2
10.5x5.5
Maximum Value = Req(max) 3.60
10.5 5.5
9.5x 4.5
Minimum Value = Req(min) 3.05
9.5 4.5