1. The device must be accessible to nurses so that they may be able to remove or apply the
device in less than 5 seconds.
2. The IVPD must be comfortable so that the blood pressure is not affected and less than
2/100 users complains of discomfort or irritation after 5 hours.
3. The IVPD must stay in place such that it does not move more than 1 cm in either
direction during use.
4. The band must be adjustable so that it fits an arm with a circumference between 5 and
10 inches.
5. The band must be entertaining enough to hold the child’s attention for 20 minutes
and/or have an 80% approval rating from a survey taken of children and parents.
6. The IVPD must be sanitary by meeting Class 1 Medical Device cleanliness specifications.
7. The IVPD must block the IV from the child’s sight by letting less than 10% of light
through during a light test.
8. The device must be inexpensive to produce by keeping production costs under $5.
9. The device must not interfere with the function of the IV by maintaining less than a 0.1%
difference in flow rate.
10. The IVPD must comply with regulations by gaining FDA approval and the approval from
agencies in 10 other countries.