NOTE:
Upon turning ON the analyzer, the operator must first perform CUVETTE ROTOR BLANK.
c. To Calibrate Test
i. click on sample type
ii. select calibrate
iii. select test to calibrate with the check on the box towards the left of the test’s
name
iv. click F8 (New Sample)
d. To perform QC
i. click on sample type
ii. select control
iii. select tests to include in QC list
iv. click F8 (New Sample)
e. To run samples
i. click on sample type
ii. select normal
iii. enter sample / lab ID number & sample name. Include sex, date of birth &
name of Doctor if needed
iv. select tests to perform
v. click F8 (New Sample)
III. LOADING SAMPLES
a. after test requisition, click on F9 (Load Sample) to confirm test sample positions
b. double click on samples one by one or select test and press enter key for each
sample. An indication number noting position will appear on the left of the table
c. click on F3 to confirm load. This will start analysis of tests. Or F2 to select samples
you wish to unload. Or, if a previous batch of tests is still indicated on the sample
rotor and you wish to reset the rotors position, click on F4 to confirm unload of all
previous sample positions.
d. To print current worklist (Samples/Tests to be perform), click on F1 (Print worklist)
e. Make sure to run reagent info before confirming load to make sure there is sufficient
reagent for all requests.
IV. TO VIEW/EVALUATE RESULTS
NOTE: When the analyzer is turned off, the cooling unitis also shut down. Thus all reagents that
require refrigeration must be stored at the same temperature after turning off the instrument.