Sunteți pe pagina 1din 1

This uncertainty budget applies to the following type of measurement: Mettler Toledo XP 56 20 grams

Applicable range of measurement: 0-52 grams

Following calibration procedure no. and rev.: N/A

All uncertainties are expressed in units of: Millligrams

Number of significant figures for reporting of expanded uncertainty: 2 Date prepared: 01/30/09

Uncertainty budget prepared by: Brad Masters

Uncertainty,
i Component of Uncertainty Distribution Divisor Std Unc, u(xi)
U(xi)
1 Uncertainty of the weight used 0.005 Normal, 1s 1.00 0.005 Millligrams
2 Drift of the Standard 0.005 Rectangular 1.73 0.00289 Millligrams
3 Balance repeatability 0.0015 Normal, 1s 1.00 0.0015 Millligrams
4 Rounding value at zero reading 0.00005 Rectangular 1.73 0.0000289 Millligrams
5 Rounding value of indicated value 0.0005 Rectangular 1.73 0.000289 Millligrams
6 Air buoyancy 0.02 Rectangular 1.73 0.0115470054 Millligrams
7 Repeatability of indication 0.001 Normal, 1s 1.00 0.001 Millligrams
8 Millligrams
9
10
combined standard uncertainty, uc 0.013 Millligrams
coverage factor, k 2
expanded uncertainty, Uc 0.026 Millligrams
Expanded uncertainty rounded UP to 2 significant figures 0.026 Millligrams

i Notes that document the basis for the above uncertainty estimates.
1 Ascertained from the calibration certificate for the standard used

2 Not calculated at this time. Set equal to the uncertainty from 1

3 Obtained from the manufacturer's specifications

4 From M3003 obtaining uncertainties

5 From M3003 obtaining uncertainties

6 From M3003 obtaining uncertainties; mass standards are stainless steel

7 One operator, one weight. Standard deviation calculated among ten readings at 20 grams divided by the square root of 10.

10

This template for a simplified uncertainty budget is provided with the compliments of the Calibration Laboratory Assessment Program (CLAS)
of the National Research Council Canada who assumes no responsibility or liability for any unintentional errors. It remains the user's responsibility
to validate the calculations and to confirm suitability for the particular purpose.
The template is designed for simple cases when:
- the components of uncertainty can be treated as independent of each other;
- dominant components of uncertainty can be estimated with reasonably high confidence; and
- the measurement result is measured directly or is computed through a linear equation involving only multiplication and/or division (e.g., W = V x I).

S-ar putea să vă placă și