Sunteți pe pagina 1din 46

ELECTROMED 2 LMF

Manuale d’uso / User’s Manual / Mode d’emploi / Gebrauchsanleitungen


Manual de instrucciones / Руководство по эксплуатации
Rev.0 – 27/08/2014 ELECTROMED 2 LMF

INDEX How to load a sequence........................................................................................... 17


How to load and I/T curve........................................................................................ 18
INFORMATION ON THIS MANUAL ......................................................................... 1 How to load values of rheobase and chronaxie of a selected program .................. 18
WRITING CONVENTIONS ............................................................................................ 1
CREATE PROGRAMS ............................................................................................ 19
WARRANTY .......................................................................................................... 2 How to create a program ......................................................................................... 19
NOTES .................................................................................................................. 2 How to create a sequence ....................................................................................... 20
How to create an I/t curve ....................................................................................... 21
CAUTIONS ............................................................................................................ 2
How to extract rheobase and chronaxie values....................................................... 22
! WARNING ! ........................................................................................................ 3
COMBINED USE .................................................................................................. 23
INTRODUCTION OF THE TECNOLOGY ..................................................................... 5
MAINTENANCE ................................................................................................... 23
IONOPHORESIS ........................................................................................................... 6
TECHNICAL PROBLEMS ........................................................................................ 24
IN GENERAL .......................................................................................................... 7
ELECTROMAGNETIC INTERFERENCES ................................................................... 24
INTENDED USE...................................................................................................... 7
INDICATIONS............................................................................................................... 7 TROUBLESHOOTING CHART................................................................................. 25
CONTRA-INDICATIONS................................................................................................ 8 TECHNICAL FEATURES ......................................................................................... 26
PRELIMINARY NOTES ............................................................................................ 8 APPENDICES ....................................................................................................... 27
UNPACKING ................................................................................................................ 8 Appendix A - ENVIRONMENTAL CONSIDERATIONS ................................................. 27
SETTING UP................................................................................................................. 9 Appendix B – LABELS ................................................................................................ 27
ACCESSORIES .............................................................................................................. 9 Appendix C – LIST OF PROTOCOLs ........................................................................... 28
CONNECTIONS ............................................................................................................ 9 Appendix D – WAVEFORMS ..................................................................................... 38
DESCRIPTION OF THE EQUIPMENT .......................................................................10 Appendix E – ELECTRO-MAGNETIC COMPATIBILITY TABLES ................................... 42
FRONTAL PANEL ....................................................................................................... 11
ANTERIOR PANEL ...................................................................................................... 11
REAR PANEL .............................................................................................................. 11
HOW TO USE OF THE DEVICE ................................................................................12
BEST USE ................................................................................................................... 12
SETTINGS.............................................................................................................12
VARIOUS ................................................................................................................... 13
LANGUAGE................................................................................................................ 13
ELECTRODE TEST....................................................................................................... 14
FREE PROCEDURE ................................................................................................14
LOADING PROGRAMS FROM MAIN MEMORY ......................................................... 15
HOW TO LOAD A PROGRAM FROM THE USER MEMORY......................................... 16
HOW TO LOAD A PROGRAM..................................................................................... 16
EMB1010 ELECTROMED 2 LMF

INFORMATION ON THIS MANUAL the right to update products and manuals without necessarily updating preceding
products or manuals, unless these have no implications for the safety of the device.
This manual is addressed to: The company will not assume any responsibility for any major cases:
- user of the machine; - improper use of the machine;
- owner; - use against to specific national regulations;
- responsible; - incorrect installation;
- people in charge of moving; - defects in power;
- installers; - serious shortcomings in maintenance;
- users; - changes and unauthorized interventions;
- people in charge of maintenance. - use of parts or materials not specific to the model;
This document provides valuable information regarding the installation, set up and - total or partial non-observance of the instructions;
use of ELECTROMED 2 LMF equipment.
- exceptional events.
It is a useful and essential reference guide for the user: read the contents of the
manual carefully before installing the equipment and keep it on hand at all times for If you would like any further information, please get directly in touch with the
future reference. company EME srl, to stay up to date on the best ways to use these machines and to
receive the necessary assistance.
It is of vital importance that you strictly adhere to the recommendations contained
within the manual in order to avoid malfunction, which may cause damage to the NOTE: The Therapy Application Manual is available upon request.
equipment and consequent annulment of the validity of the warranty. WRITING CONVENTIONS
Furthermore, in order to obtain the highly efficient technical service available from Certain sections of the manual have been underlined in order to highlight their
the manufacturer, it is essential that any handling of the equipment be in accordance importance.
with the instructions provided.
NOTE
The limits of this manual are:
These contain important information and useful tips for operating the equipment.
- the user manual cannot replace proper experience;
CAUTIONS
- the user manual, for particularly difficult operations, can only be a reminder of the The CAUTION message appears before operations, which, if not correctly performed,
main operations. may cause damage to the machine and/or its accessories..
The manual is to be considered part of the equipment and must be preserved for
! WARNING !
future reference until the decommissioning of equipment. The operating instructions
This signals operations or situations, which, if unknown to the operator, or incorrectly
must be available for consultation in the vicinity of the machine and properly stored.
carried out, may harm the operator.
This manual reflects the state of the art at the time of sale and cannot be considered
inadequate because later updated based on new information. The manufacturer has

FT05MI1 1
EMB1010 ELECTROMED 2 LMF

WARRANTY 2. carefully clean and disinfect all parts of the machine and accessories which have
been in contact with patients.
EME srl guarantees the quality of its products for a period of 24 months from the date Any equipment which the technical department does not consider hygienic (Italian
of purchase, when information contained in this manual regarding installation, use law T.U.S. 81/2008 on safety in the workplace) will not be accepted;
and maintenance is strictly adhered to and the warranty coupon is returned within 15
days of purchase. 3. disassemble accessories and any mechanical supports;

The guarantee covers the replacement of faulty parts. 4. use original box and packing materials;

The warranty does not however, include the replacement of the equipment. 5. enclose detailed information regarding the nature of the problem in order to
facilitate the technical department’s intervention and save time on repair.
The warranty does not cover any malfunction or damage caused by:
- incorrect connection and installation; NOTES
PRELIMINARY NOTES
- incorrect use due to non-compliance with instructions contained in this manual; − The installation of the device does not require any special care, is therefore simple and
- use of the machine in environmental conditions which do not conform with immediate.
USE
those specified for the product;
− Each time you click the START button or the STOP button the machine will emit a long
- improper or inadequate maintenance; confirmation beep.
MAINTENANCE
- unauthorized opening of the outer casing; - For an optimal use of the device and to guarantee its maximum performance, it is
recommended to perform maintenance at the correct time and suggested ways.
- tampering or unauthorized modifications; - It’s advisable to switch periodically the polarity according to the way it is connected to the
- use of non-original accessories. applicator plates : the exchange will increase its durability.

EME srl registered offices provide the warranty. CAUTIONS


Should you need to return the goods then please note the packing instructions as PRELIMINARY NOTES
follows. Enclose a copy of the purchasing receipt. - The customer is liable for all damage caused by inadequate packaging of the material.
Keep the original packaging of the unit: it will be needed if the unit is returned to the
You should insure the postal package. company.
- Do not use the equipment in places where it might get wet.
Before sending the machine back for suspected malfunction, we recommend that first - Before operating the machine carefully check the correctness of the connections
you carefully consult sections regarding MAINTENANCE and TROUBLESHOOTING of according to the instructions.
the manual, as a large part of the problems and faults are usually due to inadequate - Do not use accessories other than the ones provided: they might damage the unit, causing
maintenance or small technical problems which can often be easily solved by the user the warranty to become void. In case you have any problems or difficulties with
himself. installation, contact EME srl technical support.
- If using the same extension for the unit and other units, make sure that the total current
A simple call to EME srl technical department may prove to be the solution to the being absorbed by the connected units, does not exceed the max current allowed for that
problem . type of cable and that, however, it does not exceed 15 A.
- The therapeutic suggestions are stored in the permanent memory of the machine. These
When re-packing the equipment for return to the manufacturer, proceed as follows: protocols can be edited but not possible to save any changes.
- The protocols of therapeutic suggestion preloaded on the machine cannot be deleted.
1. unplug the machine and any connections, devices, applicators etc;

FT05MI1 2
EMB1010 ELECTROMED 2 LMF

- It is not possible to define a number of sessions suggested to evaluate the effectiveness of - For optimal use of the apparatus and to ensure its optimum performances it is
the treatment, since they are related to the power delivered to the patient undergoing recommended to perform properly within the time and in the manner recommended
treatment. E 'task of the physician to decide the number of therapy sessions which subject maintenance actions.
the patient according to the specific requirements of the case, in order to ensure to the - For a correct replacement of the installed fuses, observe the following indications:
patient himself the execution of an effective treatment in time and place in conditions of 1. open the fuse box using a screwdriver, making sure you insert the screwdriver in the
absolute safety. slot on the fuse box and levering up outwards;
- Always control sometimes the integrity of the cable and of the probe/applicator. 2. remove the fuse holder making it slide along the track;
USE 3. remove the installed fuses and replace them with new ones;
-The selection of programs to be loaded takes place by default in the user memory, that is 4. replace the fuse holder into the fuse box, making it slide along the track;
the only support of available memory to load customized programs. 5. close the fuse box plastic door.
- For the safety issue relating to the generation of the waveform will occur only if the self- - It is recommended to perform periodic maintenance every year, in order to check:
test safety is positively exceeded o the intensity of any leakage currents;
- In the "CURVE I / t" the electrodes must be placed in the same position in which they are o the continuity and thus the integrity, of the ground conductor;
applied for the normal delivery of electrotherapy o the correctness of the value of insulation resistance;
- Before proceeding with the curve I / t is necessary to format the secondary storage in order to ensure the electrical safety of the device, ensure that it is operating in a safe
(Smart-card user memory) where you want to store guaranteed. For this kind of intervention you should contact a qualified service technician
- For a successful login to the curve I / t is essential that all channels, including the channel or alternatively EME srl or one of its authorized service centers.
No. 1, they are all free. WORKING PROBLEMS
- Use only the power cord Link that is included with the machine for a combined use. It is - Only technicians authorized by the manufacturer may access the interior of the unit.
strictly forbidden to connect to the ends of Link cable other machines than those - You should contact EME srl or its authorized service centers for any repair work or further
specified. EME srl will not be responsible for any damage if used in a different manner information.
from that stated.
- On request we can provide the user manual in electronic form. ! WARNING !
- Because of security reasons, the only specific software must be loaded into each machine.
In case of exchange of software, the machine may immediately stop all its functions, PRELIMINARY NOTES
requiring the intervention of EME srl technical assistance. - The correct position while moving the machine: the apparatus has to be moved exclusively
- It’s a class B machine in terms of emission. The EM device is suitable to be used in every by gripping it with both hands on the curved profiles of the lid.
environment, including domestic ones and those directly connected to a public network - The perfect functionality of the device is guaranteed in accordance with the rules of
power of low voltage which supplies buildings used for domestic purposes. installation and of use included, only with original accessories and spare parts.
MAINTENANCE - If there are problems or installation difficulties, please contact the EME srl technical
- Use the probes/applicators with care: any misuse may affect their performance and assistance department.
features. - Before connecting the cable to the mains plug, check that the equipment wasn’t damaged
- Under no circumstances technicians not authorized by EME srl are allowed to open and/or during transport. Ensure that the power supply specifications on the mains socket
disassemble the probe/applicator: such tampering, besides damaging its characteristics, correspond with the information on the label attached to the back of the unit.
immediately invalidate the right to warranty. - The electric current that powers the unit is VERY DANGEROUS. Before connecting or
- The equipment should never be disassembled for cleaning or inspection purposes: the disconnecting the power cable from the connecter on the unit, make sure it is plugged out
units does not have to be cleaned internally, and if for some reason the unit must be from the mains socket.
opened, it should only be done by specialized technicians authorized by EME srl. - The power cable has an earthed plug for safety reasons.
- Do not use thinners, detergents, acid solutions, aggressive solutions or flammable liquids - Only use with a mains socket suitable for use with earthed systems
to clean the external parts of the unit and accessories. Using these substances, or - The equipment should only be connected to electrical systems that fully comply with
misusing the accessories, will cause the immediate voiding of all warranty rights, as well as regulations.
irreparably damaging the unit and the accessories. - If using extension cords verify the presence and the integrity of the protective conductor
to earth.

FT05MI1 3
EMB1010 ELECTROMED 2 LMF

- Connect the equipment directly to the wall socket without using extensions. Failure to - It is absolutely forbidden to cover the ventilation slots: such an action may not allow the
comply with these warnings may result in dangerous electrical discharges that could cause machine to work in safe conditions. In case of non-compliance with this indication, EME srl
injury operators and compromise the functioning of the unit. will not be responsible for any accidents.
- Patients with implanted electronic devices (i.e.pacemakers should not be subjected to - It’s important to pay the attention of the operator to the necessity to verify the
electrotherapy without he consent of the physician. correctness of the electric installation of device before activating the supply switch.
- The application of electrodes in the chest area may increase the risk of cardiac fibrillation. - It is advisable to suspend the therapeutic treatment if it were to appear some
- Avoid use in areas where there are ongoing inflammatory processes. disturbances during its emission.
- Simultaneous connection of a patient to a unit of high frequency electro-surgery can lead MAINTENANCE
to burns and damage the unit of electrotherapy. - The electrode is considered a commodity . It is necessary to replace electrodes
- Working in the neightbour of a short- wave or a microwave equipment may produce periodically every 2-3 months in conditions of normal use : an electrode has a duration of
instability in the output of electrotherapy. hundreds of milliamperes/hour.
USE - The use of depleted electrodes reduces the performance of the machine and can cause
- In order to avoid the contamination of the environment where the machines are working burns.
and/or the people involved in its use , do not apply to patients electrodes already used in - For safety reasons before carrying out any maintenance or cleaning the unit, YOU MUST
other appliances. turn off the equipment with the power switch at the back and unplug the socket
- In order to avoid increases in the level of current delivered out of control, the software connected to the mains.
monitors the current circulation, the patient circuit is then monitored and in the event of - Before every treatment, it is recommended to clean with caution all of the accessories
removal from the electrodes’ connection, the therapeutic treatment is interrupted. and the parts of the equipment that have been to contact with the patient.
- The electro-therapy treatments must be provided under the strict control of the operator - The operator must pay attention to the necessity of a periodic maintenance (every 2
to “conscious” patient who are able to interact with the operator concerning the electrical years) of the probes/applicators. EME srl authorized personnel should carry out such
stress produced by the machine. operations.
- Some electro-therapy treatments have a high average value current , which make the - The cleaning and disinfection must be done systematically before the therapeutic
treatment inherently dangerous as it may occur temporary redness and muscle blocks. treatment which subject the patient.
- The advanced termination of a therapy session should be done only by the key Do not - Do not use thinners, detergents, acid solutions, harsh solutions or flammable liquids to
remove the plug from the wall outlet 230V, do not unplug the power cord or using the clean the outside of the unit and its accessories. The use of these substances, with the
bipolar switch On/OFF. improper use of accessories, irreparably damages the equipment and the warranty will
- Moreover it is necessary to remove first the electrodes from the patient’s body and than lapse.
you can proceed with the turning off. - Always control sometimes the integrity of the cable and of the probe/applicator
- Use special care in the arrangement of electrodes and in the delivered current when connector: they must not be damaged or worn.
associated to a continuous component ( Ionophoresis ). - It is advised that personnel with technical preparation substitute the fuses, to perform the
- Use particular attention in the arrangement of electrodes with a current density superior operation in safety conditions.
2
to 2mA/cm right value for efficacy. When using galvanic currents ( Ionophoresis) do not - Do not open the device: inside there are high voltages that may be hazardous.
2
exceeded for any electrode current density of 0,2mA/cm . - Only personnel authorized by the manufacturer may access the internal components. For
- Do not use electrodes in direct contact with the skin, use special moistened sponges. repairs and further information please contact EME srl or its authorized service centers.
- Communicate to the patient to alert the operator in case of the intensity of the supplied WORKING PROBLEMS
current causes discomfort. − Do NOT OPEN the unity, as HIGH VOLTAGE ELECTRICITY is present and may prove VERY
- In order to ensure the functioning of the machine in conditions of absolute safety for the DANGEROUS.
patient, the operator must pay attention to the necessity of a periodic maintenance
(every 2 years) of the equipment. EME srl authorized personnel should carry out such
operations.
- It absolutely forbidden the use of the device in the presence of a flammable anesthetic
mixture and oxygen-rich environments. In case of non-compliance with this indication,
EME srl will not be responsible for any accidents.

FT05MI1 4
EMB1010 ELECTROMED 2 LMF

INTRODUCTION OF THE TECNOLOGY This phenomenon regards the resting electric potential of the membrane that
surrounds the nervous fibre. The external surface of the membrane is positively
Electrotherapy is used in order to stimulate denervated muscle, to reduce the charged with respect to the internal surface.
hypotrophy due to the non-use of the muscle normal-innervated, to maintain the Contact with the negative pole leads to depolarisation of the membrane, while the
muscular trophism of the denervated muscle or partially innervated, to strengthen positive pole is hyperpolarised.
the innervated muscle.
Even the conductibility of the nerve is changed during passage of the direct current.
Electrotherapy uses the biological effects obtained by electrical Energy for Temporary neuropraxia may occur at the positive pole.
therapeutic purposes and consists of electric currents which are passed through the
part of the body concerned, taking care to choose them with certain characteristics It is quite difficult to correctly interpret the effects caused by the direct current
suitable to the aims to be achieved. The electric currents can be continuous or however since the mechanisms that determine them are not fully understood.
variable. This kind of current serves best as a transcutaneous carrier of medicines or
The continuous current (or Galvanic current) is generated by the uniform movement cosmetically active substances .
of electric charges in the same direction. Therefore it has constant intensity over Variable currents are generally meant to denote all those currents where the
time. intensity is not constant over time.
Any action produced on an organism by this type of current is effectively linked to the Alternating current belongs to this category: it varies in intensity, and also periodically
electrochemical effect. If you immerse the electrodes of a galvanic current generator alternates in direction: the classic sinusoidal current belongs to the category and the
into an electrolyte solution, the ions will flow towards the poles of the opposite sign. time that the current takes to make a complete sinusoid is called the period. The
From the point of view of electricity, the human body behaves like an electrolyte space covered during the period is the wavelength.
solution: ions naturally present in the organism ‘migrate’ in the presence of an The number of periods made by the current in a second defines the frequency and its
electric field. unit of measurement is the Hertz [Hz]. We can distinguish between low (0 - 800Hz),
Their distribution in cellular and extracellular spaces is variable, and so the potential medium (800 - 60.000Hz), and high frequency currents (above the 100.000Hz), and
and permeability of cell walls change. for physiotherapy purposes.

These phenomenon give rise to a series of effects that can be summarised as follows : All these currents have some properties in common as regards their application
effects: the most important biological effect is definitely the motor neuronal
- vasomotor: strong hyperthermia may be felt in the area where the current is excitability effect.
applied, even after a brief treatment session;
In order to avoid the risk of possible muscular adjustment during contraction, some
- trophic: improvement in cellular ‘breathing’ due to greater blood flow that equipment can supply currents that automatically vary their frequency at regular
stimulates the metabolic processes; intervals .
- anti-oedematigenous and anti-inflammatory: linked to the reabsorption of The rectangular currents that operate at 50 Hz frequency with 1msec pulse duration
exudatives and inflammatory substances; are of special importance as they produce analgesic effects.
- nervous: the excitability is increased near the negative pole, and lessened near the The so called “ionic acceleration” is also beneficial as it involves an overlapping of
positive pole ('polar' effect of the current). galvanic current with a rectangular current. This device allows us to obtain parallel
muscular stimulation during the iontophoretic delivery (normally done by applying
the current with constant intensity).

FT05MI1 5
EMB1010 ELECTROMED 2 LMF

Treatment time is markedly reduced when this method is used. The substances that can be used are all those with the following properties: a
constant percentage of ionic disassociation in water, stability in solution and in
The different types of currents can be summarised according to their specific actions
contact with electric currents .
(or biological effects) :
However, certain basic rules must be remembered :
- Currents at IONTOPHORETIC action = continuous current (at constant intensity);
- any water used to put the substance into solution must be distilled to avoid
- Currents at EXCITO-MOTOR action = sinusoidal currents, rectangular exponential
preferential transport of parasite ions;
currents,pulse train, triangular
- sponges and electrodes must always be kept perfectly clean and they must
- Currents at ANTALGIC action = rectangular pulses of low frequency (50Hz) and pulse
be washed well in distilled water;
time 1msec.
- if an active substance solution must be prepared on the spot, the
- Currents at TROFIC and VASCULARISING = continuous currents; straightened
concentration of solution must be correctly chosen and measured
currents.
(remember that it is unnecessary to use high concentrations: generally 1% in
IONOPHORESIS weight is easily enough for most substances);
Current thinking confirms that ionophoresis is an effective way of administering - if products contained in vials are used (for parenteral use) ensure that there
different substances when localised action is required. are no products incompatible with the technique among the excipients. If it is
a freeze-dried product, ensure that the solvent is not a physiological solution:
Actually, the percutaneous application of medicinal substances is notably affected by
in that case use distilled water if possible;
the barrier function of the corneal stratum. Methods generally used to neutralise this
effect are not always effective. - avoid including other substances unless you are sure that they are perfectly
compatible: if you consider the substance to be essential, ensure that it has
Overcoming the cutaneous “barrier” stratum is not easy even when direct current is
the same polarity;
used, since substances generally prefer to enter via the glandular ducts and hair
channels. - the correct electrode arrangement is essential for diffusion of the substance,
especially if the molecular weight is low;
In addition, the substances can only reach a few millimetres depth.
If you have to introduce a positive ion, it must be put with a positive electrode and
Excluding the cases where a localised surface effect only is required, the effectiveness
vice versa for negative ions (in the case of complex molecules on the other hand, it
and distinctiveness of ionophoresis lies in the fact that the substances introduced in
seems that electro-osmosis (and also electrophoresis) takes priority: therefore the
this fashion seem to bind more stably with proteins that normally form part of
concept of polarity loses meaning, and penetration is more effective if done at anode
surface tissue. The substance is therefore reabsorbed into the general cycle more
level; the polarities of the most commonly used medicinal substances are described in
slowly than it would have been if administered hypodermically.
another section).
The general effect is however linked with the type of substance used: direct current
Even though the dosage of medicine to introduce is significant from a theoretical
only functions as a medium.
point of view, it is actually related to too many factors (skin resistance, ion size,
Therefore medicinal ionophoresis refers to substances introduced that have electrode placement) to make even an approximate calculation.
pharmalogical effects.
The use of galvanic current to the introduction of substances, when the parameters
of current intensity and duration of application are respected ,does not give rise to
undesirable effects on the skin.

FT05MI1 6
EMB1010 ELECTROMED 2 LMF

Localized redness may occur occasionally, and even burns, but only when these The operator, in fact, should be qualified to be able to use such equipment, and he
parameters are not met, or there is a lack of maintenance or improvement of should have passed an adapted training, or should operate under the control of a
electrodes. Chemical burns at the points of contact of the skin with the electrodes, medical adequately qualified to the use of the equipment, in order to guarantee
due to concentrations of chlorine ions and/or sodium (for formation of caustics safety conditions to the patient
substances with water ) may occur when using water source.
The ME equipment is suitable for use in all establishments including domestic
There is the possibility of local allergic reactions, although relatively uncommon : in establishments and those directly connected to the public low voltage power supply
any case, events are mild. Systemic allergic reactions must be excluded. network that supplies buildings used for domestic purposes.

IN GENERAL The U.S. Food and Drug Administration (FDA) has issued several guide documents for
medical equipment and has provided the following guidelines for the use of
EME srl has recently developed a complete series of apparatus, accessories and equipment for muscle stimulation:
equipment, designed and manufactured according to the highest standards of quality, - relaxation of spastic muscles;
making use of the latest technology and fully adhering to current directives and
norms. - prevention of disuse atrophy;

Particular attention has been paid to the design, easy operation, function and safety - increase local blood circulation;
of the equipment and the final result is this modern, compact unit, which offers an - provide for muscle reeducation;
extremely logical operative sequence supported by a clearly legible display .
- provide immediate post-surgical stimulation of calf muscles to prevent
A wide range of therapeutic applications, and guaranteed patient and therapist safety venous thrombosis;
ensure that equipment is of the highest quality.
- maintain or increase range of motion
The equipment were planned and built in manner that their use, if it happens at the
conditions indicated, doesn’t compromise the health and safety of the patients, of INDICATIONS
the users and of third, taking into consideration the benefit to the patient. Main effects of electrostimulation:
Such equipment are not bound to diagnosis, prevention, monitoring, compensation of 1. Training of the neuromuscular system to respond appropriately during the
injury or handicap, substitution or modification of the anatomy, control of the voluntary and involuntary effort providing an active contraction (isometric,
conception, support/vital support of functions but allow to treat special pathologies concentric and eccentric ) and the resulting joint movements allowing a
and to reduce the illness. proprioceptive feedback;
A special intervention is not required in the event of failure of the medical device, but 2. Modulation of pain by the gate control mechanism or mechanisms of descending
just a normal maintenance/repair. inhibition;
INTENDED USE 3. Control or reduction of spasticity by agonist stimulation (contraction/relaxation),
antagonists (reciprocal inhibition) or motor inhibition sense;
ELECTROMED 2 LMF is an electro-medical device that delivers electrotherapy
4. provide to the trans-dermal release of medicinal substances into the skin
treatments, with the application to the patients of conductive electrodes.
(Ionophoresis );
The use of these equipments is reserved for operators that, by their training, provide
assurance of proper use and safe for the patient.

FT05MI1 7
EMB1010 ELECTROMED 2 LMF

5. Improve or maintain joint mobility through mechanical stretching of muscles or 13. on the phrenic nerve region or on urinary bladder as the stimulus can interfere
connective tissue or reduction of the impediment of movement caused by with the normal function of these structures;
neuromuscular dysfunction, pain or swelling;
14. on scars because the scar tissue has an increased electrical resistance ;
6. Promote wound healing by increasing local circulation, providing a bactericidal
15. The current acts preferencely around the scar causing an increase of current
effect or alternating electric charges in the injured area.
density at the scars’ edges with possible creation of burnings.
1. Delay or resolving edema through muscle pump or the effect of electric charges
on the interstitial proteins (phenomenon of magnetic fields). PRELIMINARY NOTES
CONTRA-INDICATIONS UNPACKING
1. Applications on thoracic region in patients with : arrhythmias, congestive heart The equipment ELECTROMED 2 LMF is specially packaged for transport in a single
failure, recent myocardial infarction or other cardiac abnormalities; pack complete with filling which has been specifically studied for safe transportation
2. Application in any region of the body in people with cardiac pacemakers; and storage.

3. Application on the area of carotid sinus ( the bifurcation of the common carotid To remove the equipment from the pack, place the box on a smooth, flat surface.
artery);, as it may interfere with the normal regulation of the pressure blood Open the top of the box and remove the polystyrene filling. Be very careful when
flow and cardiac contractily; removing the contents of the pack.

4. Trans brain applications, because it can affect neural function (however in some The unit and accessories are wrapped in transparent sheets of polyethylene
situations the micro-currents are today applied in the trans-brain way ) protection and contains the following:

5. Application on pregnant uterus; -the User Manual;

6. Application on cancer tissues ( malignant); - n.1 mains power supply cable;

7. High width application directly above areas where it is localized the bone tissue - n.2 spare fuses (see technical specifications);
on surface , a sit can cause periosteal pain; - n.1 Output cables for electrotherapy (2 channels);
8. Application in damaged area or irritated skin because the current preferentially - n.4 conductive rubber electrodes 60x85 mm;
penetrates through the irritated area causing discomfort ( however some types
of E-stim are used to promote the healing of wounds); - n. conductive rubber electrodes 50x50 mm;

9. Application with electrodes near or touching protruding metal, such as surgical - n.4 sponge for electrodes 60x85 mm;
staples or sutures; - n.4 sponge for electrodes 50x50 mm;
10. Application on patient who reacts negatively to the procedure; - n.2 elastic bands length 1000 mm;
11. Application on patients who cannot provide with a suitable reaction on - n.2 elastic bands length 600 mm.
stimulation level ( Children, children with mental disorders ).
Check the contents of the package and should any of the items be missing then
12. On areas of excessive fatty tissue when the high level of stimulation, required to contact your local authorized EME srl dealer.
activate the deep structures can cause pain or independent reactions;

FT05MI1 8
EMB1010 ELECTROMED 2 LMF

SETTING UP The accessory assembling is simple and intuitive : each cable for the therapy, that
allows the connection of the output channels, is equipped with a multi-polar
Installation of the electrotherapy equipment ELECTROMED 2 LMF is fast and simple. connector to be inserted in the plug on the front panel of the device and with two
The following environmental conditions are ideal when installing the equipment: pairs of pins ( red for the positive electrode and black for the negative one) for
connection to the plates of conductive rubber of each channel.
− room temperature: from +10° to +40°C;
The plate holder sponge’s pockets have different sizes, each suitable to accommodate
− humidity level: from 10% to 80% without condensation; the corresponding electrode of conductive rubber.
- avoid direct exposure to sunlight, chemical products and vibrations. Be sure of the proper insertion of the electrode into the sponge’s pocket.
ACCESSORIES Along the side of each electrode in conductive rubber, there is a coupling plug in
The device can be used with the following accessories: which the banana cable for electrotherapy coming from the equipment must be
connected.
Description Included Optional
Power cable plug shuko 1 Contact authorised dealears EME srl for problems or difficulty installation.
Spare Fuses (see thecnical specification) 1
User manual in DVD 1
CONNECTIONS
Output cables for electrotherapy with banana cables The power entry module can be found on the back of the unit and consists of a three-
1
of 2mm pole socket for the cable set, an extractible fuse box with two fuses (see technical
conductive rubber electrodes 60x85mm 4 specifications) and the main switch.
conductive rubber electrodes 50x50mm 4
Sponge for electrodes 60x85mm 4 Plug the power supply cable three-pin plug into the integrated board and ensure that
Sponge for electrodes 50x50mm 4 it is correctly plugged into the connector.
Short elastic bands (600x50mm) 2
When using an extension lead, make sure that it has been earthed.
Long elastic bands (1000x50mm) 2
Probe for manual stimulation X Failure to comply with the above instructions may lead to dangerous electrical
Uro-gynaecology kit + tecnical specification X discharge causing machine damage and harm to persons.
Link cable for combined use of US/ET ultrasound devices X
Output cables for electrotherapy X Once you have checked that installation and assembly have been carried out
Carrying case TNT X according to instructions provided up to this point in the manual, switch on the
3-shelf trolley (60x37x86cm) X machine making sure that the display screen is turned on correctly.
Conductive rubber electrode 50x50mm (2mm) X
Conductive rubber electrode 60x85mm (2mm) X
Conductive rubber electrode 80x120mm (2mm) X
Sponge for electrodes 50x50mm X
Sponge for electrodes 60x85mm X
Sponge for electrodes 80x120mm X
Short elastic bands (600x50mm) X
Long elastic bands (1000x50mm) X

FT05MI1 9
EMB1010 ELECTROMED 2 LMF

DESCRIPTION OF THE EQUIPMENT

Rear panel Graphic colour display

Conductive rubber Encoder


electrodes include sponges

4 elettrodi per il modello


ELECTROMED 2 LMF Output connectors

Elastic bands

ELECTROMED 2 LMF
25 waveforms(low and medium
Number of waveforms
frequency )
Therapeutic suggestions 126 preloaded protocols
2 independent channels
Delivery channels
with banana plugs 2mm
Output connectors for serial
1 connector CH 1+2
connection
Possibility of combined use MEDISON 1-3

FT05MI1 10
EMB1010 ELECTROMED 2 LMF

FRONTAL PANEL ANTERIOR PANEL

Touch-In colour graphic


display

Output connector
CH1+2

REAR PANEL

Power cable supply

Earthing LINK connector for


terminal combined use
Fuses box

General switch ON/OFF


Encoder

FT05MI1 11
EMB1010 ELECTROMED 2 LMF

HOW TO USE OF THE DEVICE


This section provides important information and instructions on how to make the
best use of the equipment for electrotherapy of ELECTROMED 2 LMF.
All the control functions and the machine itself are handled and co-ordinated by a
microprocessor: apart from making pre-memorised programmes available for
application, the microprocessor ensures that the machine can be personalised and
operated in a highly safe and efficient manner.
Interfacing allows for the operator to communicate with the unit by means of a large,
clear graphic backlit liquid crystal display screen (LCD) through which all operational Fig.1
messages required by the operator, work status during operation, and errors are
visualised. ELECTROMED 2 LMF
The following paragraphs illustrate the procedures to be carried out and the technical
specifications of the unit.
They also deal with the different options available, from the selection of a pre-
memorised programme for use in specific treatments as well as how to determine the
correct working parameters for “personalised” applications.
BEST USE Load programs Settings

After having installed and correctly positioned the machine as per the instructions
described in the previous sections and the electrode applicator connection cable (or Create programs Free procedure
cables) has been connected with the right connectors, plug the machine into a Fig.2
230Vac wall socket and switch on using the ON/OFF main switch on the back panel of
the unit. SETTINGS
Once turned on, the LCD display lights up and the unit is ready for use. Allows you to edit and save to the internal memory the basic settings to be recalled
With the first turn on the device, you can set the language between the six available. automatically every time .
Then turn the encoder to select the desired language and press the knob to confirm At the screen in fig.2 press SETTINGS: window appears fig.3.
the selection. Press the SAVE button to save the changes. A confirmation message will
inform you of any such modification. Turn the encoder knob (which by default is located on the OTHER menu) select the
feature you want to change, and then press the knob to confirm the selection.
The LCD display will illuminate showing the logo (see Fig1). Appears the main screen
that allows to select between four operating modes (Fig.2) by pressing the On the page also appear information about the version of software and firmware of
corresponding button. each module of the power board installed on the device.

FT05MI1 12
EMB1010 ELECTROMED 2 LMF

VARIOUS PROVA SETTING


VARIOUS
Various
By pressing the encoder knob at the menu VARIOUS: appear fig.4. Language Format User Memory
v Buzzer on
You can perform the following operations: Electrode test
Control electrodes
• formatting user memory of the machine which is equipped; Synchronizes CH 1+2
Software v20.1
• turn on or turn off the horn, Firmware v3.7 v3.7
• enable or disable the feature of electrodes;
• enable or disable the functionality of synchronization of the output channels, Exit Save Exit
enabling the machine to provide the same therapeutic treatment on the
output channels selected for synchronization (CH 1+2).
Fig.3 Fig.4
Selecting the option FORMAT USER MEMORY through the rotation of the rotary
encoder is possible to formatting the user memory. The control of the electrodes is not executable in the following conditions :
User confirmation is requested to avoid accidental formatting. • for current intensity not exceeding to 10mA
• for too low pulse frequency
Pressing the FORMAT button, the formatting operation of the selected support
• for too short pulses duration
memory is executed. When formatting is completed, a screen will appear showing
• in constant voltage modality
that the operation is completed. After a few seconds you return to the main screen of
this section (fig.4). To enable synchronization of output channels of electrotherapy treatments ( ie,
enable the machine to provide the same treatment on the output channels, selected
To enable or disable the acoustic signal, turn the encoder knob to select the menu
for synchronization), rotate the encoder knob to select the menu SYNCHRONIZES
BUZZER ON, then press the knob. When there is a check the horn is working.
CH1+2, then press the knob. When there is a check mark , the machine performs the
To enable the control of electrodes, rotate the encoder knob to select the menu action of synchronization.
CONTROL ELECTRODES, then press the knob. When there is a check mark, the
The configuration set can be saved by pressing the function key on the SAVE button
machine will control the electrodes.
(see Figure 4)
This function allows you to check if the electrodes are properly connected to the
Pressing the EXIT function for the button (see Figure 4) to return to the main screen
machine and the patient. This check is performed during the therapy.
of the SETUP menu (Figure 3).
This check is performed to therapy delivery in progress, and fails if one of the two
conditions is not met.The operator is informed of the outcome of the inspection by LANGUAGE
the appearance of a message on the screen. To choose the language in which you write all messages and commands of the
machine, select the desired language by turning the encoder knob and press the knob
to activate the new language choice.
The language can be selected from those available in Italian, English, French, German,
Spanish, Russian and Romanian.

FT05MI1 13
EMB1010 ELECTROMED 2 LMF

After a short pause for the loading of the new dictionary, the main screen of the The cursor of the encoder knob is located on the default menu of the waveform. To
SETUP menu with the new language. change the type of waveform, press the encoder knob and turn until you find the
desired shape, then press the encoder knob to confirm the selection.
To select the desired language, turn the encoder knob until you reach the appropriate
line and press the encoder knob to confirm your selection, then press the SAVE Before starting treatment, you can customize the program by changing the time of
button to enable the machine to work with the selected language. duration of treatment and the values of its characteristic parameters: polarity
reversal at mid-term and type of emission:
Pressing the EXIT function for the button instead of the encoder knob to select the
language, it goes back to the main screen of the SETTINGS menu (fig. 3) • the changes to the waveform are not storable
To change the language back you can repeat the procedure at any time • changes to the treatment plan cannot be stored in main memory, but custom
protocols can only be saved in the storage media secondary (user memory or
ELECTRODE TEST
smart cards
The electrodes’ test is used to verify that the output boards are working properly.
Pressing the CHANNEL 1, you can select the channel at which to load the desired
Follow the instructions on the screen. After a short time in which the machine is therapeutic treatment.
performing, you will see a video message on the state of the electrodes:
The button CHANNEL 1 indicates that the program has been uploaded refers precisely
• ELECTRODES OK to channel 1 and then pressing the button toggles the selection of output channels
• ELECTRODES OUT available.
It is possible to change the display of the output channels, by pressing the button If synchronization channel electrotherapy is active, the changes made to the
corresponding to the function CHANNEL 1. parameters of treatment at the CHANNEL 1 will be played also at the CHANNEL 2 (or
vice versa) at the time of treatment. In this case, in fact, the same therapeutic
The procedure is started by pressing the START button.
treatment will be provided on both the dispensing channels selected with the
If you press the function key for the button EXIT, the screen returns to Figure 3. synchronization.

FREE PROCEDURE Instead, if the synchronization channel is not active one has the possibility to choose
the dispensing channel between the two available and to modify the treatment
In this section you can quickly change the parameters of treatment and arbitrary parameters independently.
waveform so loaded and use them for your own program. The channel in which the emission is already active is obviously not available for a
After pressing the key button PROCEDURE FREE a screen containing a waveform with new program until the previous one has not ended.
the relevant technical data.
In this screen there are three buttons on the screen:
• button START to start loaded therapy
• CHANNEL 1 button, which allows you to select channels between the two
treatment delivery available;
• the EXIT button, to return to the screen in fig.2

FT05MI1 14
EMB1010 ELECTROMED 2 LMF

LOADING PROGRAMS - type of emission


- the pulse duration
Pressing the LOAD PROGRAMS directly on the screen you can upload therapeutic
programs contained in the machine or possibly those saved in the smart-card. - the type of pulse (variable between S, A and R, where S stands for Symmetric,
A for Asymmetric and R for Rectangular
The main screen of the menu contains 3 buttons:
- The frequency
- button STANDARD PROGRAMS
- button USER PROGRAMS - the rise time
- button SMART-CARD PROGRAMS - The time of descent
- button EXIT
- The time of permanence
and the list of protocols contained in the main memory of the machine, which is
loaded by default when you select this menu. - pause time.

The programs can be customized by the operator is stored on smart-card that the To change the value of one of the characterizing parameters among selected program
user memory, unlike what happens instead for the main memory. follow the instructions below:

Offered stored programs are the result of years of operational experience in user 1. Turn the encoder knob into position of the cursor of parameter you want to
support professional experts. Appendix C contains a list of available programs. change;

The channel in which the emission is already active is obviously not available for a 2. press the encoder knob ;
new program until it is finished. Press the EXIT button (directly on the screen), you 3. turn the encoder knob (clockwise for increasing values in an anti-clockwise to
return to the screen in fig. 2. decreasing values) to select the desired value for the parameter that you
want to customize;
LOADING PROGRAMS FROM MAIN MEMORY
4. press the encoder knob to confirm the assignment of the selected value to
Pressing the STANDARD PROGRAMS, it can be load programs stored on the main
the custom parameter.
memory of the machine (such programs are editable, and then customized by the
operator, but cannot be stored on that media). Repeat steps 1) to 4) for each of the parameters that you wish to modify.
To select one of the treatments available, turn the key encoder to select the desired Push the button START to start emission of the program.
treatment, and then press the encoder knob to confirm the loading of the desired
Push the button EXIT to exit the menu: the software will get back to the main screen
treatment
of the menu LOAD PROGRAMS.
Once you choose the treatment that you want to deliver, a screen appears that
If you push the button CHANNEL 1 you can select the channel where the treatment
contains the parameters that characterize the selected treatment
will be emitted to the patient. Repeatedly pressing this button, you can alternate the
It could be changed: CHANNEL 1, 2.
- delivery time of treatment
- polarity
- mid-term inversion

FT05MI1 15
EMB1010 ELECTROMED 2 LMF

HOW TO LOAD A PROGRAM FROM THE USER MEMORY


SELECT
The user memory is one of the slave memory slots of the device. There you can store Program
therapeutic programs, sequences of therapeutic programs, I/T curves and values of Sequence
Reobase and Cronassia. Curve I/t
Reobase/Cronas.
Before choosing one of the four options, you need to format the user memory,
otherwise the device (when choosing an option) will show a message and emit a an
acoustic signal, requesting to format the memory slot.
Follow the instructions to load a menu:
1. Push the button that reads “USER PROGRAMS (fig.2); Exit
2. Turn the encoder knob and choose the desired menu; Fig.5

3. Push the encoder knob to confirm the choice.


TEST
HOW TO LOAD A PROGRAM
Turn the encoder knob to select the menu PROGRAM, then push it to confirm the
operation.
If the user memory does not contain programs, a message appears on the screen ,
with an acoustic signal, stating that there are no stored programs in the user
memory. The system goes back to the screenshot as in fig.2.
If the user memory contains programs, the list of the stored programs appears on the
screen. You will see a screen like in fig.6. Exit
Fig.6
Push the button EXIT to cancel the loading procedure. The system will go back to the
screen as in fig. 2. Otherwise push the encoder knob after highlighting the menu of
TEST
the program you want to load; the screen as in fig.7 shall appear.
The stop loading push the button EXIT. The system will go back to the screen in fig.6.
To cancel the selected program push the button CANCEL. The list of the programs
stored in the user memory will be updated. Instead, in order to load the selected
program push the button LOAD.
After choosing the desired therapeutic treatment, the screen with all the parameters Cancel
of the treatment chosen will appear (see fig.8). You can modify the treatment
emission time, polarity, inversion halfway the period, type of emission, following the
Load Exit
procedure described in HOW TO LOAD PROGRAMS FROM THE MAIN MEMORY.
Fig.7

FT05MI1 16
EMB1010 ELECTROMED 2 LMF

You can choose the output channel, between the channels available, by pushing the TEST
Polarity: R+
button CHANNEL 1 in succession until the button reading the desired channel will
Inversion to 1/2T: OFF
show up.
CC/CV: CC
To abort the execution of a program, push the button EXIT. The system will go back
to the screen in fig.6.
To start a program push the button START: the screen in fig.9 shall appear. TIME: 15 min

The treatment emission is indicated by the counter that shows remaining time until Start
the program comes to an end. The channel occupied by the emission is indicated by
the number inside the button EXIT.
Channel 1 Exit
During the delivery of electrotherapy treatment can increase the intensity of the
Fig.8
current supplied on the output channels selected: then turn the encoder knob (to the
right or to the left for increasing values for decreasing values) to update the progress TEST
bar relating to 'intensity of the current supplied. CURRENT 0.0 mA
To pause the program push the button PAUSE (see fig.10): the counter stops and the
screen will read PAUSE, the emission channel will be shown until you push again the
button STOP in order to completely stop the emission of the desired program. TIME: 12:36
To resume emission, from the same channel, push the button START and reset the
value of current intensity.
Pause
To terminate the emission push the button STOP.
You can also pause the treatment in one channel and start its emission in another 1
Channel 1 Exit
available channel by pushing the button CHANNEL 1, that will be substituted by the Fig.9
button CHANNEL 2: then push the button START. Inside the button EXIT another
number indicating the channel occupied will appear. TEST
CURRENT 0.0 mA
How to load a sequence
Turn the encoder knob to select the menu SEQUENCE, then push it to confirm the
operation.
PAUSE
If there are no sequences in the user memory, a message stating this will show up on
the screen and an acoustic signal will sound. Otherwise the list of the stored
sequences will be loaded onto the screen. Start Stop

You can abort the loading operation by pushing the button EXIT and the screen in 1
fig.2 will show up again. You can read the details of each sequence by pushing the Channel 1 Exit
button DETAILS. Fig.10

FT05MI1 17
EMB1010 ELECTROMED 2 LMF

To load a sequence, select it with the encoder knob and then push the knob. triangular) push the button CHANGE. You can read further details in the paragraph
related to the creation of an I/t curve.
If you want to abort the loading of the sequence push the button EXIT.
Push the button SAVE to store the I/t curve you just created with the free procedure.
To cancel the program sequence selected push the button CANCEL. The list of the
While you can push the button RESET to clean off the I/t curve chart on the screen.
sequences stored in the user memory will be updated. Whereas to load the selected
sequence push the button LOAD. If you want to abort the storing of a new I/t curve push the button EXIT.
After choosing the program sequence, the screen with all the parameters of the Push the button EXIT some times to go back to screen in fig.2.
treatment chosen will appear. You can modify the treatment emission time, polarity,
inversion halfway the period, type of emission and another parameters, following the How to load values of rheobase and chronaxie of a selected program
procedure described in HOW TO LOAD PROGRAMS FROM THE MAIN MEMORY. Turn the encoder knob to select the menu RHEOBASE/CHRONAX, then push it to
You can choose the output channel, between the channels available, by pushing the confirm the operation.
button CHANNEL 1 in succession until the button reading the desired channel will If there are no programs in the user memory, a message stating this will show up on
show up. the screen and an acoustic signal will sound. Otherwise the list of the stored
To start a program push the button START, while to abort the execution of a program, programs will be loaded onto the screen.
push the button EXIT. You can abort the loading operation by pushing the button EXIT and the screen in
Push the button EXIT some times to go back to screen in fig.2. fig.2 will show up again. To load the program with the selected values of rheobase
and chronaxie, select it and push the encoder knob.
How to load and I/T curve
If you want to abort the loading of the selected program push the button EXIT.
Turn the encoder knob to select the menu I/T CURVE, then push it to confirm the
To cancel the selected program push the button CANCEL. The list of the programs
operation.
with the values of rheobase and chronaxie stored in the user memory will be
If there are no I/t curves in the user memory, a message stating this will show up on updated.
the screen and an acoustic signal will sound. Otherwise the list of the stored I/t curves
Instead, to load the selected program push the button LOAD.
will be loaded onto the screen.
By default the encoder point will be placed on the impulse type parameter; choose
You can abort the loading operation by pushing the button EXIT and the screen in
the pulse type between rectangular and triangular shapes (see fig.11).
fig.2 will show up again.
Push the button EXIT to abort the loading operation and you will get back to pic.5.
To load an I/t curve, select it with the encoder knob and then push the knob.
Instead, to start calculating the rheobase and chronaxie push the button START (see
You can abort the loading operation by pushing the button EXIT. fig.12).
To cancel the I/t curve selected push the button CANCEL. The list of the I/t curves Push the encoder knob after selecting the menu INTENSITY to modify the current
stored in the user memory will be updated. Whereas to load the selected I/t curve intensity.
push the button LOAD.
Turn the encoder knob clockwise for increasing values and anticlockwise for
To clean off the I/t curve chart on the screen push the button RESET. While to start to decreasing values until you choose the desired intensity value.
capture data of a new I/t curve and to alternate pulse type (rectangular and
Push the encoder knob to select the intensity value, then push the button CONFIRM
and the rheobase value will be determined (see fig.13).

FT05MI1 18
EMB1010 ELECTROMED 2 LMF

Then rotate the encoder knob until to position the cursor on the PULSE menu , then Push the button SAVE in the screen of fig.14 to continue saving the rheobase and
press the encoder knob for to enter in the routine of modification of the pulse chronaxie values of the selected program, then you will go back to fig.6. Otherwise
duration. push the button EXIT to abort the saving procedure and you will go back nevertheless
to fig.6.
Rotate the encoder knob (in clockwise way for increasing values , in counterclockwise
way for decreasing values) until to select the desired value to assign at the pulse Push the button EXIT some times to go back to screen in fig.2.
duration.
CREATE PROGRAMS
Press the encoder knob for to confirm the assignation of the selected value at the
pulse duration , then press the CONFIRM button and the chronaxie value will be This feature allows you to save therapeutic custom programs on user memory, which
determined (see fig.14). is the only available memories to save the new programs.
Waveform: Rectangular Waveform: Rectangular To create a program, press the button CREATE PROGRAMS: a new screen will appear
Intensity: 0.0mA Intensity: 0.0mA to create a program, create a sequence of programs (no more than 4), create an I/t
Pulse: 1000.0mS Pulse: 1000.0mS
Pause: 1 sec Pause: 1 sec curve and to calculate the values of rheobase and chronaxie from a new program.

Reobase: 0.0 mA
Turn the encoder knob to position the cursor on the desired function, then press the
>>Reobase: 0.0mA
same knob to confirm the execution of the selected operation.
Cronaxya: 0.0 mS Cronaxya: 0.0mS

Start
How to create a program
Stop
In order to create a new program follow these steps:
Exit Confirm 1. Press the CREATE PROGRAMS button (fig.2);

Fig.11 Fig.12
2. Turn the encoder knob till the cursor it’s positioned on PROGRAM menu;
3. Press the encoder knob to confirm your choise: the screen on fig.15 will
Waveforms: Rectangular Waveforms: Rectangular
appear.
Intensity: 1.0mA Intensity: 1.0mA
Pulse: 0.1mS Pulse: 0.5mS You can customize the program, by changing the selected waveform and / or the
Pause: 1 sec Pause: 1 sec values of its characterizing parameters: duration of treatment, polarity, reversing half
Reobase: 0.5mA Reobase: 0.5mA period and type of emission and another parameters.
>>Cronaxya: 0.0mS Cronaxya: 0.5mS
To cancel the creation of the program, press the EXIT button: you’ll return to the
Stop Save screen shown in Figure 5.
Otherwise, press the SAVE button to save the program on a storage media available
Confirm Exit
(user memory or smart-card): screen appears in Figure 16.

Fig.13 Fig.14 To assign a name to your custom program use the following procedure:

You can abort this operations of extraction of the reobase / chronaxie values by 1. press the encoder knob to enable the insertion of the name for the program
pushing the button STOP in the screen of fig.12 or fig.13, then you will go back to the (see fig.17);
screen in fig.11.

FT05MI1 19
EMB1010 ELECTROMED 2 LMF

2. Turn the encoder knob to select the desired alphanumeric character (turning MONOPHASE Enter name:
the encoder knob, you can select the entire set of alphanumeric characters), Polarity: R+
Inversion to 1/2T: OFF _
see the screen in Figure 18;
CC/CV: CC
3. press the encoder knob to confirm the entry of the alphanumeric character
selected (see fig.19); Time: 15 min

4. Turn the encoder knob to position the cursor on a new space in order to
proceed with entering the next alphanumeric character; OK
Repeat steps 1) to 4) to fully enter the name you want to assign to the program (see
fig.20). Save Exit

Press the OK button to confirm the assignment of the name you typed in your
program. The customized program will be automatically saved in the user memory. Fig.15 Fig.16
Enter name: Enter name:
After a few seconds a screen will appear that informs you of the completion of the
requested operation, then automatically returns to the screen fig.5. _ T

How to create a sequence


In order to create a new sequence follow these steps:
1. press the CREATE PROGRAMS button (fig.2);
OK OK
2. turn the encoder knob till the cursor il positioned on SEQUENCE menu;
3. press the encoder knob to confirm your choice.
The screen of Figure 21 appears in which you can store a sequence of up to 4
programs. Fig.17 Fig.18

The cursor is by default set to STEP 1, then press the encoder knob to enter the first Enter name: Enter name:
program in the sequence. T TEST_

The screen in Figure 15 appears in which you can customize the program by changing
the selected waveform and / or the values of its characterizing parameters.
To cancel the creation of the new program, press the EXIT button, it returns to the
screen in Figure 21. Otherwise, press the SAVE button.
OK OK
You will find the stored program in the sequence you are creating (see an example in
fig.22).

Fig.19 Fig.20

FT05MI1 20
EMB1010 ELECTROMED 2 LMF

Premere il pulsante OK per confermare l’assegnazione del nome digitato alla


SELECT SELECT sequenza personalizzata. The custom sequence will be automatically saved in the
STEP 1 [EMPTY] STEP 1 [MONOPHASE] user memory.
STEP 2 [EMPTY] STEP 2 [EMPTY]
STEP 3 [EMPTY] STEP 3 [EMPTY] After a few seconds a screen will appear that informs the operator of the completion
STEP 4 [EMPTY] STEP 4 [EMPTY] of the operation and automatically returns to the screen fig.5.
The idea that underlies the creation of a sequence that is at the end of a delivered
program the machine steps to dispense the next program, depending on the channel
on which it is loaded; this is repeated until the sequence has completed.
Save Exit Save Exit In the creation of a sequence is not possible to save programs in not consecutive
positions.
Fig.21 Fig.22 How to create an I/t curve
SELECT The I/t CURVE is the graphical representation of the perceptible thresholds of muscle
STEP 1 [MONOPHASE] stimulation and is useful for controlling the improvement resulting from a therapy or
STEP 2 [DIPHASE] a surgery.
STEP 3 [SINC.MONOPHASE]
STEP 4 [SINC.DIPHASE]
Rectangular or triangular waveforms are used on the same application points where
the electrodes are placed for regular electrotherapy.
The electrotherapy channel enabled for the creation of the curve I / t is always the
CHANNEL 1.

Save Exit Before proceeding with the curve I / t is necessary to format the secondary storage
(smart card or user memory) where you want to store it.
Fig.23 To create a I/t curve follow the instructions below:
Rotate again the encoder knob to position the cursor on the menu STEP 2, then press 1. place the electrodes (positive and negative) of ELECTROTHERAPY CHANNEL 1
the encoder knob to confirm and follow the above steps to save a new program in the on the ends of the muscle of interest;
sequence. 2. press the CREATE PROGRAMS button (see Figure 2);
You can follow the same procedure for the other two empty positions, for a 3. Turn the encoder knob to position the cursor on the menu I/t CURVE;
maximum of 4 programs in a sequence (See an example in fig.23). 4. then press the encoder knob to confirm the selection.

To cancel the creation of the sequence, press the EXIT button, it returns to the screen Figure 24 shows the screen where you can:
shown in Figure 5. Otherwise, press the SAVE button to save the newly created - Press the EXIT button to cancel the creation of the I/t curve and you will
sequence. return to the screen of Figure 5;
The screen in Figure 16 appears, allowing you to assign a name to the custom - Press the RESET button to clear the screen from any loaded graphic
sequence in the same way described in the previous paragraph.. waveform;

FT05MI1 21
EMB1010 ELECTROMED 2 LMF

- Press CHANGE to select the type of pulse you want to use (rectangular or At the end of the acquisition process, it displays a screen where you can:
triangular).
- Press the EXIT button to cancel the acquisition of I/t curve; returns to the
The only currently adjustable parameter is the duration of the pause time, which can screen fig.5;
be varied by means of a simple rotation of the encoder knob (clockwise for increasing - Press the RESET button to clear the screen from any loaded graphic
values, counter-clockwise for decreasing values). waveform;
- Press the SAVE button to save the I/t curve just got into secondary storage
The acquisition of the curve requires you to press the START button to allow it to
media.
operate: fig.25 screen appears.
Pressing the SAVE button Figure 16 screen appears, which allows you to assign a
name to the custom sequence, following the instructions in the previous paragraph
CURRENT: 0.0mA CURRENT: 0.0mA
Press the OK button to confirm the assignment of the name you typed in the
customized curve I/t. The customized I/t a will be automatically saved in the user
1000.0 mS 1000.0 mS
memory.
Pause: 1Sec Pause: 1Sec
After a few seconds a screen will appear that informs the operator of the completion
of the requested operation, then automatically returns to the screen fig.5.
Start FreeChange
Procedure Stop
How to extract rheobase and chronaxie values
Reset Exit Confirm Once you have chosen the type of pulse (rectangular or triangular) and the pause
time, it is possible to trace the values of rheobase and chronaxie.
Fig.24 Fig.25 The rheobase is the intensity of the stress felt by the patient after delivering a pulse 1
Rotate the encoder knob to define the value of the current intensity (clockwise for second long.
increasing values and counter-clockwise for decreasing values). The chronaxie is the time taken to measure muscle contraction after giving a signal
When the diagnostic operator note a reaction of muscle to the stimulus press the intensity of the double value set arbitrarily by the user.
CONFIRM button to confirm the entry of the selected value and the acquisition of the While the first parameter is detected directly by the software, the second parameter
same curve in the graph of I/t curve. is detected during the delivery of the therapeutic treatment: in fact it can be
Automatically, with each storage, the intensity of the pulse is reset to zero and measured from the time that is allowed to vary the pulse width, capable of producing
prepared the new pulse time; diagnostic operator has the task of detecting the the muscle contraction.
intensity that causes a contraction perceptible. The electrotherapy channel enabled for extract the of the values of rheobase and
Press the STOP button to stop the sequence at any time. chronaxie is always the CHANNEL 1.
If the sequence is stopped with the STOP button, it can be restarted at the same point In order to extract the values of rheobase and chronaxie follow these instructions:
with the START button. 1. place the electrodes (positive and negative) of ELECTROTHERAPY CHANNEL 1
The curve I / t can be stored always but to have a significant curve I / t it is necessary on the ends of the muscle of interest;
to acquire all of the 11 steps, then repeat the above procedure for all steps that 2. press the CREATE PROGRAMS button (see Figure 2);
comprise the creation of the curve I / t.

FT05MI1 22
EMB1010 ELECTROMED 2 LMF

3. turn the encoder knob to position the cursor on the menu COMBINED USE
RHEOBASE/CHRONAXIE;
4. press the encoder knob to confirm the selection: Figure 11 screen is You can use the “LINK” cable (see ACCESSORIES section) to connect a ELECTROMED 2
displayed. LMF unit to MEDISON 1-3 unit; this union allows to realize an hardware connection
The cursor is positioned by default on the type of impulse; then choose the type of between the devices, and it allows to emit combined treatments of electro-therapy
impulse between the rectangular and the triangular waveform. and ultrasound-therapy .

Press the START button to begin the rheobase and chronaxie calculation routine (see For this scope, you must to apply on the patient the positive electrode of the channel
fig.12). 1 of electrotherapy and the ultrasound probe, because the ultrasound probe due to
LINK cable represents the negative electrode of the channel 1 of the electrotherapy.
Rotate and the press the encoder knob on the INTENSITY menu to enter the routine This connection allows to substitute to an hardware level the negative electrode of
for changing the current value. the channel 1 of the electrotherapy with the ultrasound probe
Rotate the encoder knob clockwise for increasing values or counter-clockwise for This means that :
decreasing values, to select the desired current value.
− pushing the START button on ELECTROMED 2 LMF but not on MEDISON 1-3,
Press the encoder knob to confirm the selected current value, then press the the electrotherapy treatment will be emitted using the electrode pair
CONFIRM button to show the extracted rheobase value: appears the screen of fig.13. constituted by the positive electrode of the channel 1 of electrotherapy and
Then rotate the encoder knob until to position the cursor on the PULSE menu , then the ultrasound probe.
press the encoder knob for to enter in the routine of modification of the pulse - pushing the START button on ELECTROMED 2 LMF and on MEDISON 1-3, the
duration. ultrasound probe becomes the emission terminal both for the ultrasound
Rotate the encoder knob (in clockwise way for increasing values , in counterclockwise therapy treatment and the electrotherapy treatment.
way for decreasing values) until to select the desired value to assign at the pulse The remaining channels are still available for the treatment of electrotherapy.
duration.
Press the encoder knob for to confirm the assignation of the selected value at the MAINTENANCE
pulse duration , then press the CONFIRM button and the chronaxie value will be
The ELECTROMED 2 LMF device for the electrotherapy do not require any particular
determined: appears the screen of fig.14.
maintenance operations.
You can abort this operations of extraction of the reobase / chronaxie values by
When cleaning the outer part of the equipment, make sure to use a soft, clean cloth
pushing the button STOP in the screen of pic.12 or pic.13, then you will go back to the
dampened with luke-warm water or very mild non inflammable detergents.
screen in pic.11.
The front panel can be cleaned in the same way.
Press the SAVE button in the screen of fig.14 to continue and save rheobase and
chronaxie values into secondary storage media following the instructions described in The electrodes in conductive rubber must be regularly cleaned with water and
"CREATE PROGRAMS" at the "How to create a program" section. denatured alcohol.
Replace the cables and electrodes with care at the end of each treatment session. The
electrodes in conductive rubber must be periodically replaced as they lose their
conductivity with use

FT05MI1 23
EMB1010 ELECTROMED 2 LMF

Contact authorised dealers of EME srl for information regarding original spare parts specialised institutes. The customer, upon prior request, may view the reports
or components relative to EMC measures within the company.
Do not immerse the unit in water. The equipment does not generate significant radio frequency energy and is
adequately immune to radiated electromagnetic fields. Therefore it does not
After cleaning the external part of the equipment, make sure to dry it perfectly before
detrimentally interfere with radio-electric communications, electro-medical
turning on the unit.
equipment for monitoring, diagnosis, therapy and surgery, office electronic devices
The unit must under no circumstances be opened or dismantled in order to clean or such as computers, printers, photocopiers, fax machines, etc. or any electric or
check inner parts of equipment does not require cleaning of inner parts and in all electronic equipment used in these environments, as long as said equipment
cases, only specialised technicians or EME srl authorised personnel should carry out complies with the ELECTROMAGNETIC COMPATIBILITY directive .
such operations.
In any case, in order to avoid any interference problems, we recommend that you
The useful life of the device is 10 years. operate the therapy equipment far enough away from critical equipment for
monitoring vital patient functions, and that you be careful when applying therapy to
TECHNICAL PROBLEMS patients with pacemakers..
The equipment for electrotherapy has been designed and manufactured using highly
advanced technology and first class components for reliable and efficient
performance.
However, should you meet with any operational problems, we recommended that
you consult the following guide before contacting any of our authorised service
centres.
If any of the following situations occur, disconnect the machine and contact EME srl
authorised service centres:
− the cable set or rear supply panel show signs of wear and tear or are
damaged;
− the liquid has entered the equipment
• the equipment has been exposed to rain.

ELECTROMAGNETIC INTERFERENCES
The equipment for electrotherapy, ELECTROMED 2 LMF, has been designed and
manufactured according to the ELECTROMAGNETIC COMPATIBILITY DIRECTIVE
2004/108/CE with the aim of providing adequate protection from harmful
interference when installed in homes and health establishments.
All required measurements and tests have been carried out in EME’s internal Testing,
Measurement and Inspection laboratory (LPMC), in addition to other external

FT05MI1 24
EMB1010 ELECTROMED 2 LMF

PROBLEM POSSIBLE CAUSE SOLUTION


TROUBLESHOOTING CHART
Contact with the patient
insufficient, faulty, and/or
Contact the EME srl assistance
PROBLEM POSSIBLE CAUSE SOLUTION uncertain.
centre.
Mains plug not plugged in Check that the mains socket is Current generator electronic
properly working circuit fault
Mains cable not properly Plug in properly and connect Carry out maintenance
The output circuit from the
Front panel LCD connected to the unit cable to the unit operations as described.
patient is not connected
display doesn’t come Mains cable worn or blocked. Replace the mains cable. Install and position the unit as
properly.
on: the unit doesn’t Switch off. Switch on the mains switch. described
work. Fuse or fuses defective or Replace the missing, defective Check the condition of the
The unit works Uncertain or defective cables, the connectors and the
blown. or blown fuses. properly but there is a conductive rubber electrode output circuit electrodes
Electronic control circuit Contact the EME srl assistance notable reduction
malfunction. centre contacts (including the Ensure the electrodes adhere
In the effectiveness of electrode-holder sponge perfectly to the treatment
Front panel LCD the treatment.
Defective components in the Contact the EME srl assistance pockets) with the surface to be surface.
display doesn’t come
electronic control board. centre treated Moisten the electrode-holder
on
sponge pockets.
Some commands on Defective keys or buttons.
the front control Contact the EME srl assistance Possible current generator Contact the EME srl assistance
Electronic control system circuit fault in the equipment. centre
panel are not centre
malfunction.
working properly.
Defective connections in the
patient application output
circuit.
Check that the output
Output cable blocked or not
connection is made properly.
connected properly.
Output cable worn and/or bad
The display signals
connection.
that the unit is
Connect the output cables
unable to supply the Connectors badly inserted,
correctly.
required output defective or not connected to
Replace the cables and/or the
current . the equipment and/or the
defective connectors that show
electrodes.
evident signs of wear.
Ensure that the electrodes
Electrodes in conductive rubber adhere properly to the patient,
worn, need to be replaced replace defective and/or worn
electrodes.

FT05MI1 25
EMB1010 ELECTROMED 2 LMF

TECHNICAL FEATURES Sequences that can be saved in the user memory 10

Curve I/t that can be saved in the user memory 100


230 Vac, 50-60Hz, ±10%
Main voltage: Value of reobase/cronaxya that can be saved in the user
115 Vac, 50-60Hz, ±10% * 100
memory
Max. Power absorption: ELECTROMED 2 LMF 35 VA
Table container in plastic, external size (width x height x
315 mA - T - 5 x 20 mm 39 x H14 x 30 cm
depth)
for alimentation 230Vac
Double delay type (T) mains fuse
protection: ELECTROMED 2 LMF
630 mA - T - 5 x 20 mm
for alimentation 115Vac Unit body weight ELECTROMED 2 LMF 3.80 Kg
Backlit LCD Display, to visualise and control operating Colour graphic 320 x 240 pixel touch
parameters screen + encoder
Programmable treatment time Up to 60 minutes
Class of isolation/parts applied according to the rule EN Room temperature (+10 : +40) °C
I / BF Conditions for use
60601-1
Classification in compliance with the directive 93/42/CEE II B Relative humidity (10 : 80) % without condensation
Degree of protection against the input of liquids
IPX0
according to the UNI EN 60601-1 Room temperature (-40 : +70) °C
All low and medium
frequency currents , Conditions for stocking /
Emission frequency ELECTROMED 2 LMF Relative humidity (10 : 100) % without condensation
transport
25 Waveforms total number
Atmospheric pressure (500 : 1060) hPa
Voltage Costant
Functioning: ELECTROMED 2 LMF Current Costant
* on demand
Pulse current 100 mA
Peak current
Diadynamic current 70 mA
(Load resistance 1KOhm)
Continuous current 50 mA
Pulse current 100 V
Peak voltage
Diadynamic current 70 V
(Load resistance 1KOhm)
Continuous current 50V
Channel output ELECTROMED 2 LMF 2 indipendents

Protocols in memory ELECTROMED 2 LMF 126

Number of protocols that can be saved in the user


200
memory

FT05MI1 26
EMB1010 ELECTROMED 2 LMF

APPENDICES
Device model
Appendix A - ENVIRONMENTAL CONSIDERATIONS Input frequency of the device

ELECTROMED 2 LMF equipment for electrotherapy has been designed and


manufactured to have minimal negative environmental impact, in line with its Serial number Output characteristics of the device
operational and safety requirements.
Rigorous standards were followed in order to minimise the amount of waste, use of
toxic materials, noise, non-required radiation and energy consumption. Output voltage of the device Output current of the device

In accordance with careful research, the unit has been designed to optimise power
consumption in keeping with energy saving principles.
Output frequency of the device
This symbol means that the product should not be disposed of as Tabella 2
domestic waste.
Etichetta Etichetta
Significato Significato
The user must dispose of scrap equipment by taking it to a recognised electrical and
electronic recycling centre.
Appendix B – LABELS
Label placed near the LINK connector ,
Simbolo Simbolo used for a connection towards an MEDISON unit ,
situed on the rear panel of the device Label placed near the output connector.
Significato Significato

This product complies with regulations issued under


the certification from a Notified Body Type BF equipment

Label near the equipotential connection.

The product must be dispose of as “electronic


Attention
waste”, not as “domestic waste
Table 1
Simbolo Simbolo
Significato Significato

input characteristics Input voltage to the device (mains)

Fuses Input power of the device (absorbed power)

FT05MI1 27
EMB1010 ELECTROMED 2 LMF

Appendix C – LIST OF PROTOCOLs


Waveform Modulation (Hz)
Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum

1 Ankle pain 1 Diphase (fixed) 10 100 12

2 Ankle pain 2 TENS S/A/R 15 300 usec 3 3 110 110 12-15

Variable positioning of the


3 Knee pain 1 Bifhasic S/A 15 250 usec 6 2 100 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
4 Knee pain 2 Interpherential 15 3 2 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
5 Knee pain 3 TENS S/A/R 15 300 usec 3 3 110 110 electrodes cannot be 12
defined a priori.
Variable positioning of the
6 Algodistr. refl Interpherential classic 20 3 2 50 120 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
Dermal local
7 Iontophoresis 10 1 0 8000 electrodes cannot be 1
anaesthesia (Fantanyl)
defined a priori.
Dermal local Variable positioning of the
8 anaesthesia (Lydocaine / Iontophoresis 10 1 0 8000 electrodes cannot be 1
epinephrine) defined a priori.
Variable positioning of the
9 Rheumat.Artxit1 Bifhasic S/A 30 200 usec 3 3 4 4 electrodes cannot be 15
defined a priori.
Variable positioning of the
burst burst
10 Rheumat.Artxit2 Tens burst S/A/R 15 150 usec 1 1 40 electrodes cannot be 12
frequency 5Hz frequency5Hz
defined a priori.
Variable positioning of the
11 Hips osteoart.1 Difase (fissa) 12 100 electrodes cannot be 15
defined a priori.
Variable positioning of the
12 Hips osteoart.2 Interferenziale 20 5 3 1 2 2500 electrodes cannot be 12
defined a priori.

13 Ankle osteoart1 Diphase (fixed) 10 100 12

FT05MI1 28
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum

14 Ankle osteoart2 Interpherential 15 5 3 1 2 2500 12

15 Cervi. Ostarth1 Interpherential 12 5 3 1 2 2500 12

16 Cervi. Ostarth2 TENS S/A/R 15 190 usec 5 3 80 80 12

17 Knee osteoartx1 Diphase (fixed) 15 100 12

18 Knee osteoartx2 Galvanic 15 1 0 12

19 Knee osteoartx3 Interpherential 15 5 3 1 2 2500 12

20 Knee osteoartx4 TENS S/A/R 15 300 usec 5 2 110 110 12-15

21 Elbw osteoart.1 Diphase (fixed) 10 100 12

22 Elbw osteoart.2 Galvanic 15 1 0 12

23 Elbw osteoart.3 Interpherential 12 5 3 1 2 2500 12

Variable positioning of the


24 Foot osteoarth1 Short period 5 2 2 50 100 electrodes cannot be 12
defined a priori.

FT05MI1 29
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum
Variable positioning of the
25 Foot osteoarth2 Monophase (fixed) 5 50 electrodes cannot be 12
defined a priori.
Variable positioning of the
26 Hand osteoartx1 Short period 10 2 2 50 100 electrodes cannot be 12-15
defined a priori.
Variable positioning of the
27 Hand osteoartx2 Interpherential 15 5 3 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
28 Hand osteoartx3 Monophase (fixed) 5 50 electrodes cannot be 12
defined a priori.
Variable positioning of the
29 Lumbr osteoart1 Diphase (fixed) 12 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
30 Lumbr osteoart2 Interpherential 15 5 3 1 2 2500 electrodes cannot be 12-15
defined a priori.

31 Mandb.ostoarth1 Interpherential classic 15 3 2 50 120 2500 10-12

32 Mandb.ostoarth2 TENS S/A/R 15 250 usec 3 1 150 150 12

33 Wrist osteoart1 Diphase (fixed) 10 100 12

34 Wrist osteoart2 Interpherential 12 5 3 1 2 2500 12

Variable positioning of the


35 Should osteart1 Interpherential 20 5 3 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
36 Should osteart2 TENS S/A/R 20 300 usec 6 2 110 110 electrodes cannot be 15
defined a priori.
Variable positioning of the
37 Rheumat.Artxrit Interpherential classic 20 6 2 2500 electrodes cannot be 12-15
defined a priori.
Variable positioning of the
38 Bursitis 1 Galvanic 12 1 0 electrodes cannot be 12
defined a priori.

FT05MI1 30
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum
Variable positioning of the
Burst frequency Burst frequency
39 Bursitis 2 Tens burst S/A/R 15 200 usec 6 3 80 electrodes cannot be 12
2 HZ 2 HZ
defined a priori.
Variable positioning of the
40 Bursitis 3 TENS S/A/R 20 180 usec 6 3 80 80 electrodes cannot be 12
defined a priori.
Variable positioning of the
Periarticular
41 Galvanic 20 1 0 electrodes cannot be 15
calcifications
defined a priori.

42 Cervicobrachi.1 Galvanic 15 1 0 12

43 Cervicobrachi.2 TENS S/A/R 12 180 usec 6 3 200 200 12

Variable positioning of the


44 Keloids 1 Galvanic 20 1 0 electrodes cannot be 15
defined a priori.
Variable positioning of the
45 Keloids 2 Iontophoresis 20 5 1 8000 electrodes cannot be 15
defined a priori.
Variable positioning of the
Acute pain after surgery
46 Iontophoresis 10 1 0 8000 electrodes cannot be 1-7 o più
(Fantanyl)
defined a priori.
Variable positioning of the
Acute pain after surgery Burst frequency Burst frequency
47 Tens burst S/A/R 15 200 usec 6 2 80 electrodes cannot be 1 o più
1 2 HZ 2 HZ
defined a priori.
Variable positioning of the
Acute pain after surgery
48 TENS S/A/R 25 180 usec 6 3 250 250 electrodes cannot be 1 o più
2
defined a priori.
Variable positioning of the
Chronic pain after
49 Tens S/A/R 10 50 usec 3 0 150 150 electrodes cannot be 15
surgery 1
defined a priori.
Variable positioning of the
Chronic pain after
50 Tens S/A/R 10 180 usec 6 3 250 250 electrodes cannot be 15
surgery 2
defined a priori.

51 Dorsalgia Tens S/A/R 15 110 usec 6 3 100 100 12-15

Variable positioning of the


52 Seqnt.drain.lgs Kotz 25 10 50 50 50 2500 electrodes cannot be 12-15
defined a priori.

FT05MI1 31
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum
Variable positioning of the
53 Seq.drain.up.am Kotz 20 10 50 50 50 2500 electrodes cannot be 12-15
defined a priori.

54 Epicondylitis Tens S/A/R 15 100 usec 12 2 100 100 12

55 Medl epicdylis Tens S/A/R 15 100 usec 12 2 100 100 12

Variable positioning of the


56 Fibromyalgia 1 Interpherential 20 5 3 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
Burst frequency Burst frequency
57 Fibromyalgia 2 Tens burst S/A/R 10 200 usec 6 3 100 electrodes cannot be 12
2 HZ 2 HZ
defined a priori.
Variable positioning of the
58 Ionto-PatchTest Iontophoresis 10 1 0 8000 electrodes cannot be 2-12
defined a priori.
Variable positioning of the
59 Hyperkeratosis Iontophoresis 15 1 0 8000 electrodes cannot be 12
defined a priori.
Variable positioning of the
Hyperhidrosis of hands
60 Iontophoresis 25 1 0 8000 electrodes cannot be 12
and feet
defined a priori.
Variable positioning of the
61 Post tra.musc.1 Kotz 25 1 1 50 50 2500 electrodes cannot be 12-15
defined a priori.
Variable positioning of the
62 Post tra.musc.2 Rectangular pharadic 15 1 msec 5 2 50 50 electrodes cannot be 12-15
defined a priori.
Variable positioning of the
63 Lymphoedema 1 Iontophoresis 10 1 0 8000 electrodes cannot be 15
defined a priori.
Variable positioning of the
64 Lymphoedema 2 Kotz 25 10 50 50 50 2500 electrodes cannot be 12 o più
defined a priori.

65 Acute lumbago 1 Diphase (fixed) 12 100 15

66 Acute lumbago 2 Interpherential 15 5 3 1 2 2500 12

FT05MI1 32
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum

67 Acute lumbago 3 Biphasic S/A 15 250 usec 3 3 100 100 15

68 Acute lumbago 4 Short period 12 2 2 50 100 12

69 Acute lumbago 5 Galvanic 20 1 0 12

70 Acute lumbago 6 Interpherential classic 15 3 2 50 120 2500 15

71 Acute lumbago 7 Interpherential 15 5 3 1 2 2500 15

72 Acute lumbago 8 Long period 12 12 6 50 100 12

Burst frequency Burst frequency


73 Acute lumbago 9 Tens burst S/A/R 15 50 usec 1 1 100 15
2 HZ 2 HZ

74 Acute lumbago10 Tens S/A/R 15 50 usec 3 3 100 100 15

75 LumbsciaticPain Tens S/A/R 25 200 usec 12 3 80 80 12-15

76 Metatarsalgia 1 Galvanic 15 1 0 12

FT05MI1 33
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum

77 Metatarsalgia 2 Interpherential 15 3 2 1 2 2500 12

78 Metatarsalgia 3 Tens S/A/R 12 320 usec 5 2 110 110 12

79 Metatarsalgia 4 Tens S/A/R 20 50 usec 3 3 100 100 12

Variable positioning of the


80 Neuralgia 1 Galvanic 15 1 0 electrodes cannot be 12
defined a priori.
Variable positioning of the
81 Neuralgia 2 Interpherential classic 15 3 2 50 120 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
burst frequency Burst frequency
82 Trigemns nerlg1 Tens burst S/A/R 10 100 usec 6 3 60 electrodes cannot be 12
2 HZ 2 HZ
defined a priori.
Variable positioning of the
83 Trigemns nerlg2 Tens S/A/R 10 110 usec 6 3 70 100 electrodes cannot be 12
defined a priori.

84 Int-cost neurlg Interpherential classic 20 3 2 50 120 2500 12

Variable positioning of the


85 Radial neuralg1 Interpherential classic 10 4 2 50 120 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
86 Radial neuralg2 Tens S/A/R 12 200 usec 6 3 200 200 electrodes cannot be 12
defined a priori.
Variable positioning of the
87 Ulnar neuralge. Interpherential classic 15 1 0 50 120 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
88 Pst herpt.neu.1 Biphasic S/A 30 80 usec 3 3 100 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
89 Pst herpt.neu.2 Tens S/A/R 10 200 usec 6 3 80 80 electrodes cannot be 12
defined a priori.
Variable positioning of the
90 PEFS 1 Biphasic S/A 35 500 usec 3 3 35 35 electrodes cannot be 12
defined a priori.

FT05MI1 34
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum
Variable positioning of the
91 PEFS 2 Biphasic S/A 15 500 usec 3 0 15 15 electrodes cannot be 15
defined a priori.

92 Should.periart1 Galvanic 15 1 0 12

93 Should.periart2 Interpherential 15 3 2 1 2 2500 12

94 Hips periarthr1 Interpherential 20 3 2 1 2 2500 12

95 Hips periarthr2 Tens S/A/R 25 300 usec 3 3 110 110 12

Variable positioning of the


96 Musclar rheumt1 Biphasic S/A 12 250 usec 3 3 100 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
97 Musclar rheumt2 Galvanic 12 1 0 electrodes cannot be 12
defined a priori.
Variable positioning of the
98 Musclar rheumt3 Interpherential 12 3 2 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
99 Mscl pass wrmup Kotz 12 1 0 100 100 2500 electrodes cannot be 12-15
defined a priori.
Variable positioning of the
100 Rhizoarthrosis1 Diphase (fixed) 10 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
101 Rhizoarthrosis2 Interpherential 12 3 2 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
102 Rhizoarthrosis3 Interpherential 20 3 2 1 2 2500 electrodes cannot be 12
defined a priori.
Variable positioning of the
103 Rhizoarthrosis4 Tens S/A/R 20 300 usec 3 3 110 110 electrodes cannot be 12
defined a priori.

FT05MI1 35
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum

Syndrome of Osgood-
104 Galvanic 20 1 0 12
Schlatter 1

Syndrome of Osgood-
105 Interpherential 15 3 2 1 2 2500 12
Schlatter 2

Variable positioning of the


106 Volkman syndrom Short period 10 2 2 50 100 electrodes cannot be 12-15
defined a priori.
Variable positioning of the
Burst frequency Burst
107 Myofasc pn syn1 Tens burst S/A/R 10 200 usec 6 3 80 electrodes cannot be 12
2 HZ frequency2 HZ
defined a priori.
Variable positioning of the
108 Myofasc pn syn2 Tens S/A/R 10 150 usec 6 3 80 80 electrodes cannot be 12
defined a priori.
Variable positioning of the
Muscle contractions and
109 Galvanic 12 1 0 electrodes cannot be 8-12
spasms
defined a priori.
Variable positioning of the
110 Tendonitis 1 Diphase (fixed) 5 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
111 Tendonitis 2 Long period 7 12 6 50 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
Burst frequency Burst frequency
112 Tendonitis 3 Tens burst S/A/R 12 200 usec 6 3 80 electrodes cannot be 12
2 HZ 2 HZ
defined a priori.
Variable positioning of the
113 Tendonitis 4 Tens S/A/R 15 150 usec 6 3 100 100 electrodes cannot be 12
defined a priori.

114 Biceps tendnits Tens S/A/R 15 310 usec 6 3 150 150 12

115 Achills tendnit Tens S/A/R 15 100 usec 12 2 100 100 12-15

Variable positioning of the


116 Ankl tndonits Tens S/A/R 15 100 usec 12 2 100 100 electrodes cannot be 12
defined a priori.

FT05MI1 36
EMB1010 ELECTROMED 2 LMF

Waveform Modulation (Hz)


Electrotherpy treatments Time length (min) Pulse length Action time (sec) Pause time (sec) Frequency (Hz) Electrodes positioning N°of sessions
associated Minimum Maximum

117 Ansri tendonits Tens S/A/R 15 100 usec 12 2 100 100 12

Variable positioning of the


118 Patl.tndts-jmp Tens S/A/R 15 100 usec 12 2 100 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
119 Tendinosis Tens S/A/R 15 100 usec 12 2 100 100 electrodes cannot be 12
defined a priori.
Variable positioning of the
120 Mscl pass tonif Kotz 20 1 0 50 50 2500 electrodes cannot be 15
defined a priori.
Variable positioning of the
Muscle relaxation
121 Kotz 15 10 50 50 50 2500 electrodes cannot be 3-12
treatment 1
defined a priori.
Variable positioning of the
Muscle relaxation
122 Monophase 15 50 electrodes cannot be 12
treatment 2
defined a priori.

Burst frequency Burst frequency


123 Carpal Tunnel 1 Tens burst S/A/R 10 200 usec 6 3 80 12
2 HZ 2 HZ

124 Carpal Tunnel 2 Modulated pharadic 10 180 usec 6 3 100 200 12

Variable positioning of the


125 Warts Iontophoresi 10 1 0 8000 electrodes cannot be 15
defined a priori.
Variable positioning of the
126 Vitiligo Iontophoresi 10 1 0 8000 electrodes cannot be 10 o più
defined a priori.

FT05MI1 37
EMB1010 ELECTROMED 2 LMF

Appendix D – WAVEFORMS
Rise time Decay time Action time Pause time
Pulse duration Frequency [Hz]
[sec] [sec] [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)
N° Name of the current waveform
Min time Max time Monophase Diphase duration
Frequency A [Hz] Frequency B [Hz]
[sec] [sec] duration [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)

/ / / / 20ms 50
1 MONOPHASE
/ / / / / /

/ / / / 10ms 100
2 DIPHASE
/ / / / / /

1 (0 ÷ 30) 1 (0 ÷ 30) 6 (1 ÷ 120) 5 (0 ÷ 120) 20ms 50


3 SYNCHOPED MONOPHASE
/ / / / / /

1 (0 ÷ 30) 1 (0 ÷ 30) 6 (1 ÷ 120) 5 (0 ÷ 120) 10ms 100


4 SYNCHOPED DIPHASE
/ / / / / /

20ms MONOPHASE 50 MONOPHASE


/ / / 2 (0 ÷ 120)
10ms DIPHASE 100 DIPHASE
5 SHORT PERIOD
/ / 1 (1 ÷ 60) 1 (1 ÷ 60) / /

20ms MONOPHASE 50 MONOPHASE


/ / / 6 (0 ÷ 30)
10ms DIPHASE 100 DIPHASE
6 LONG PERIOD
/ / 6 (1 ÷ 60) 6 (1 ÷ 60) / /

500 us
1 (0 ÷ 30) 1 (0 ÷ 30) / 5 (0 ÷ 120) /
(100us ÷ 9.0ms)
7 RECTANGULAR
50 100
1 (0 ÷ 120) 3 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *
500 us
1 (0 ÷ 30) 1 (0 ÷ 30) / 5 (0 ÷ 120) /
(100us ÷ 9.0ms)
8 ESPONENTIAL
50 100
1 (0 ÷ 120) 3 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *

FT05MI1 38
EMB1010 ELECTROMED 2 LMF

Rise time Decay time Action time Pause time


Pulse duration Frequency [Hz]
[sec] [sec] [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)
N° Name of the current waveform
Min time Max time Monophase Diphase duration
Frequency A [Hz] Frequency B [Hz]
[sec] [sec] duration [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)
500 us
1 (0 ÷ 30) 1 (0 ÷ 30) / 5 (0 ÷ 120) /
(100us ÷ 9.0ms)
9 TRIANGULAR
50 100
1 (0 ÷ 120) 3 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *
Pausa impulso:5.0 ms 2.0ms
/ / / 140
(100 us - 2 s) (100us ÷ 2s)
10 TRAEBERT
/ / / / / /

500 us
1 (0 ÷ 30) 1 (0 ÷ 30) / 5 (0 ÷ 120) /
(100us ÷ 9.0ms)
11 RECTANGULAR PHARADIC
50 100
1 (0 ÷ 120) 3 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *
1.0ms
1 (0 ÷ 30) 1 (0 ÷ 30) 6 (1 ÷ 120) 5 (0 ÷ 120) 100 (1 ÷ 250)
(100us ÷ 9.0ms)
12 MODULATED PHARADIC
/ / / / / /

1 (0 ÷ 30) 1 (0 ÷ 30) / 0 (0 ÷ 120) / /


TRIANGULAR
13
NEODYNAMIC 200 200
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *

1 (0 ÷ 30) 1 (0 ÷ 30) / 0 (0 ÷ 120) / /


14 NEODYNAMIC
200 200
3 (0 ÷ 120) 3 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *
100us
1 (1 ÷ 30) 1 (1 ÷ 30) / 3 (0 ÷ 120) /
( 50us ÷ 1ms)
15 TENS S/A/R
100 100
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(1 ÷ freq.B) * (freq.A ÷ 250) *

/ / 1 (1 ÷ 120) 1 (0 ÷ 120) 50us (50us ÷ 1s) 100 (1 ÷ 200)


16 TENS RANDOM S/A/R
/ / / / / /

FT05MI1 39
EMB1010 ELECTROMED 2 LMF

Rise time Decay time Action time Pause time


Pulse duration Frequency [Hz]
[sec] [sec] [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)
N° Name of the current waveform
Min time Max time Monophase duration Diphase duration
Frequency A [Hz] Frequency B [Hz]
[sec] [sec] [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)

/ / 1 (1 ÷ 120) 1 (0 ÷ 120) 50us (50us ÷ 1s) /


17 TENS BURST S/A/R
100 2
/ / / / 3 3
(1 ÷ 200) * (1 ÷ 10) *
250us
1 (1 ÷ 30) 1 (1 ÷ 30) / 3 (0 ÷ 120) /
(50us ÷ 2ms)
18 BIPHASIC S/A
100 100
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(1 ÷freq.B) * (freq.A ÷ 250) *

/ / 1 (1 ÷ 120) 0 (0 ÷ 120) / /
GALVANIC
19
(IONOPHORESIS)
/ / / / / /

8000
/ / 1 (1 ÷ 120) 0 (0 ÷ 120) /
INTERRUPTED GALVANIC (2000 ÷ 8000)
20
(IONTOPHORESIS)
/ / / / / /

/ / 1 (1 ÷ 120) 0 (0 ÷ 120) / /
21 KOTZ
2500 50
/ / / / 4 4
(1000 ÷ 5000) * (1 ÷ 250) *
2500
1 (0 ÷ 30) 1 (0 ÷ 30) / 2 (0 ÷ 30) / 5
(2500 o 4000) *
22 INTERPHERENTIAL
1 2
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(0 ÷ freq.B) * (freq.A ÷ 250) *
2500
1 (0 ÷ 30) 1 (0 ÷ 30) / 2 (0 ÷ 120) / 5
(2500 o 4000) *
23 INTERPHERENTIAL CLASSIC
50 120
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(0 ÷ freq.B) * (freq.A ÷ 250) *
2500
1 (0 ÷ 30) 1 (0 ÷ 30) / 2 (0 ÷ 120) / 5
INTERPHERENTIAL (2500 o 4000) *
24
ISOPLANAR 50 120
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(0 ÷ freq.B) * (freq.A ÷ 250) *

FT05MI1 40
EMB1010 ELECTROMED 2 LMF

Rise time Decay time Action time Pause time


Pulse duration Frequency [Hz]
[sec] [sec] [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)
N° Name of the current waveform
Min time Max time Monophase duration Diphase duration
Frequency A [Hz] Frequency B [Hz]
[sec] [sec] [sec] [sec]
Default (Range) Default (Range) Default (Range) Default (Range) Default (Range) Default (Range)
2500
1 (0 ÷ 30) 1 (0 ÷ 30) / 2 (0 ÷ 120) / 5
INTERPHERENTIAL (2500 o 4000) *
25
VECTORIAL 50 120
0 (0 ÷ 120) 1 (1 ÷ 120) / / 2 2
(0 ÷ freq.B) * (freq.A ÷ 250) *

Nota: S -> simmetrico, A -> asimmetrico, R -> rettangolare.


2
* il range generale di variazione della frequenza è (0 Hz ÷ 250 Hz).
3
* nella forma d’onda TENS BURST S/A/R la Frequenza A e la Frequenza B rappresentano rispettivamente la Frequenza TENS e quella BURST con i rispettivi range di variazione.
4
* nella forma d’onda KOTZ la Frequenza A e la Frequenza B rappresentano rispettivamente la Frequenza PORTANTE e quella MODULATA con i rispettivi range di variazione.
5
* nelle forme d’onda INTERFERENZIALE, INTERFERENZIALE CLASSICA, INTERFERENZIALE ISOPLANARE, INTERFERENZIALE VETTORIALE la Frequenza rappresenta la Frequenza PORTANTE.

FT05MI1 41
EMB1010 ELECTROMED 2 LMF

Appendix E – ELECTRO-MAGNETIC COMPATIBILITY TABLES


Guidance and manufacturer’s declaration – electromagnetic emissions 13.4.1 Guidance and manufacturer’s declaration – electromagnetic immunity
FOR ALL ME EQUIPMENT FOR ALL ME EQUIPMENT
The ME EQUIPMENT is intended for use in the electromagnetic environment specified below. The ME EQUIPMENT is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME EQUIPMENT should assure that it is used in such an The customer or the user of the ME EQUIPMENT should assure that it is used in such an
environment. environment.
Emissions test Compliance Electromagnetic environment – guidance IEC 60601 Electromagnetic environment –
Immunity test Compliance level
Test level guidance
The ME EQUIPMENT uses RF energy only for its Floors should be wood, concrete
RF Emissions internal function. Therefore, its RF emissions are ± 6kV contact ± 6kV contact or ceramic tile.
Group 1 Electrostatic
CISPR 11 very low and are not likely to cause any If floors are covered with
interference in nearby electronic equipment. discharge (ESD)
IEC 61000-4-2 synthetic material, the relative
± 8kV air ± 8kV air humidity should be at least 30 %.
RF Emissions
Class B
CISPR 11 The ME EQUIPMENT is suitable for use in all ± 2kV per power ± 2kV per power
Electrical fast supply lines supply lines Mains power quality should be
Harmonic emissions establishments including domestic establishments
Class B transient/burst that of a typical commercial or
IEC 61000-3-2 and those directly connected to the public low ± 8kV for
IEC 61000-4-4 NOT APPLICABLE hospital environment.
voltage power supply network that supplies input / output lines
Voltage fluctuations/ buildings used for domestic purposes.
flicker emissions Complies ± 1kV line(s) to ± 1kV line(s) to
Mains power quality should be
IEC 61000-3-3 Surge line(s) line(s)
that of a typical commercial or
IEC 61000-4-5 ± 2kV line(s) to ± 2kV line(s) to
hospital environment.
earth earth
<5% UT (>95% dip
<5% UT for 0.5Mains power quality should be
in UT) for 0.5
cycles that of a typical commercial or
cycles
Voltage dips, short hospital environment. If the user
interruptions and 40% UT (60% dip in
<5% UT for 1 cycle of the ME EQUIPMENT requires
voltage variations on UT) for 5 cycles continued operation during
power supply input power mains interruptions, it is
lines 70% UT (30% dip in 70% UT for 25 recommended that the ME
IEC 61000-4-11 UT) for 25 cycles cycles EQUIPMENT be powered from an
uninterruptible power supply or
<5% UT (>95% dip
<5% UT for 5 sec a battery.
in UT) for 5 sec
Power frequency magnetic fields
Power frequency
should be at levels characteristic
(50/60Hz) magnetic
3A/m 3A/m of a typical location in a typical
field
commercial or hospital
IEC 61000-4-8
environment
NOTE : UT is the a.c. mains voltage prior to application of the test level.

FT05MI1 42
EMB1010 ELECTROMED 2 LMF

13.4.2 Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY – Recommended separation distances between portable and mobile RF communications equipment
FOR ME EQUIPMENT THAT ARE NOT LIFE-SUPPORTING and the ME EQUIPMENT
The ME EQUIPMENT is intended for use in the electromagnetic environment specified below. The The ME EQUIPMENT is intended for use in an electromagnetic environment in which radiated RF
customer or the user of the ME EQUIPMENT should assure that it is used in such an environment. disturbances are controlled.
IEC 60601 Compliance Electromagnetic environment – The customer or the user of the ME EQUIPMENT can help prevent electromagnetic interference by
Immunity test
Test level level guidance maintaining a minimum distance between portable and mobile RF communications equipment
Portable and mobile RF communications equipment should be used no closer to any part of the ME (transmitters) and the ME EQUIPMENT as recommended below, according to the maximum output
EQUIPMENT, including cables, than the recommended separation distance calculated from the power of the communications equipment.
equation applicable to the frequency of the transmitter. Separation distance according to frequency of transmitter ( m )
Recommended separation distance Rated maximum output 150 kHz ÷ 80 MHz 80 MHz ÷ 800 MHz 800 MHz ÷ 2.5 GHz
power of transmitter (W) d=  3,5  d=  3,5  P d=  7  P
  P    
Conducted RF 3Vrms d=  3,5  P V
 1 E
 1  E1 
IEC 61000-4-6 150kHz to 80MHz
3V (V1)  
 V1  0.01 0.12 0.12 0.23
d=  3.5  P
0.1 0.37 0.37 0.74
  1 1.17 1.17 2.33
 E1  10 3.69 3.69 7.38
Radiated RF 3V/m from 80 MHz to 800 MHz
3V/m (E1) 100 11.67 11.67 23.33
d=  7  P
IEC 61000-4-3 80MHz to 2,5GHz For transmitters rated at a maximum output power not listed above, the recommended separation
  distance d in metres (m) can be determined using the equation applicable to the frequency of the
 E1 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
from 80 MHz to 2,5 GHz
the transmitter manufacturer.
where P is the maximum output power rating of the transmitter in watts (W) according to the
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
transmitter manufacturer and d is the recommended separation distance in metres (m). Field
strengths from fixed RF transmitters, as determined by an electromagnetic site survey,*1 should be NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
less than the compliance level in each frequency range.*2 absorption and reflection from structures, objects and people.

Interference may occur in the vicinity of equipment marked with the following symbol : SW: v20.1
NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
*1: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the ME
EQUIPMENT is used exceeds the applicable RF compliance level above, the ME EQUIPMENT should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the ME EQUIPMENT .
*2: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

FT05MI1 43
MEDICAL TECHNOLOGY
EME s.r.l. Via Degli Abeti 88/1 · 61122 Pesaro (PU) Italy
Ph. +39 0721 400791 · Fax +39 0721 26385 ·Mail info@eme-srl.com

S-ar putea să vă placă și