Documente Academic
Documente Profesional
Documente Cultură
Service Manual
1
Click Here for Table of Contents Click Here for Navigation Help
LIFEPAK 1000 Defibrillator/Monitor Table of Contents
Click a Topic
Preface This service manual describes how to maintain, test, troubleshoot, and repair the
LIFEPAK 1000 defibrillator. A separate publication, the LIFEPAK 1000
Defibrillator Operating Instructions, is for use by physicians, clinicians, and
emergency care providers. The operating instructions provide step-by-step
instructions as well as operator-level testing and maintenance.
Click any of the links below to jump to a topic in this section.
Trademarks
Using Adobe Reader
Navigating Through the Manual
Service Personnel Qualifications
Contacting Medtronic
Responsibility for Information
Device Tracking
Service Information
Warranty Information
Glossary
Acronyms
(Continued on next page)
Trademarks
LIFEPAK and FAST-PATCH are registered trademarks of Medtronic Emergency
Response Systems, Inc.
ADAPTIV, CODE SUMMARY, QUIK-COMBO, REDI-PAK, Shock Advisory
System, CODE-STAT and PARTLINE are trademarks of Medtronic Emergency
Response Systems, Inc.
Medtronic is a registered trademark of Medtronic, Inc.
Microsoft and Windows are registered trademarks of Microsoft Corporation in
the US and/or other countries.
Adobe and Acrobat are registered trademarks of Adobe Systems Incorporated.
Tektronix is a registered trademark of Tektronix Incorporated.
QED-6 and Dale601 are trademarks of Fluke Biomedical Corporation.
© 2006 Medtronic Emergency Response Systems, Inc. All rights reserved.
MIN 3206072-000 / CAT. ??????
Accessing Adobe This service manual opens in Adobe® Reader, which is included on this
Reader Help documentation CD. For additional assistance using the Adobe Reader program,
access ADOBE READER HELP in the HELP menu.
Using Bookmarks Bookmarks appear in a column on the left side of the screen. They enable you to
easily navigate to main sections of the manual, similar to a table of contents.
To view or hide the bookmarks column, click the BOOKMARKS tab located along
the left side of the screen.
To jump to a bookmark topic, click the desired topic.
Note: A plus sign to the left of a bookmark topic indicates additional topics exist
under that bookmark level. Click the plus sign to expand or collapse the
bookmarks.
Using Page View Click the PAGES tab located to the far left of the screen to view miniature images
of each page in the document. Scroll through the pages and click an image to
jump quickly to that page.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Blue text indicates a hyperlink. Click a link to jump to that topic or page. Click
Back in the navigation bar to return to your previous location. The pointer
changes to a pointing finger when positioned over a link. A navigation bar at
the bottom of each page also provides helpful links.
The navigation bar includes:
■ Table of Contents Click to jump to the main table of contents for the
manual.
■ Section Contents Click to jump to the table of contents for the section
you are currently viewing.
■ Index Click to jump to the manual’s index.
■ Back Click to retrace your steps in a document, returning to each
page in the reverse order visited.
■ Next Page Click to jump to the next page of the manual.
■ Previous Page Click to jump to the previous page of the manual.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Service technicians must be properly qualified and thoroughly familiar with the
operation of the LIFEPAK 1000 defibrillator. They must meet at least one of the
following requirements (or the equivalent):
■ Associate of Applied Science, with an emphasis in biomedical electronics.
■ Certificate of Technical Training, with an emphasis in biomedical electronics.
■ Equivalent biomedical electronics experience.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
This service manual describes the methods required to maintain, test, and repair
the LIFEPAK 1000 defibrillator. This manual does not address the operation of
the LIFEPAK 1000 defibrillator. Qualified service personnel must consult both
the LIFEPAK 1000 Defibrillator Operating Instructions and this service
manual to obtain a complete understanding of the use and maintenance of the
device.
It is the responsibility of our customers to ensure that the appropriate person(s)
within their organization has access to the information in this service manual,
including any warnings and cautions used throughout the manual.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
General Device It is important to maintain accurate records of defibrillator location within your
Tracking Information facility or system. Maintenance of such records eases the process of locating
defibrillators should it be necessary to modify them. Defibrillators should be
tracked by both the manufacturer’s part number and serial number. Internal
asset or tracking numbers may also be useful in maintaining adequate control of
defibrillators.
Device Tracking The U.S. Food and Drug Administration requires defibrillator manufacturers and
!USA
distributors to track the location of their defibrillators. If your device has been
sold, donated, lost, stolen, exported, or destroyed, or if it was not obtained
directly from Medtronic, please notify the device-tracking coordinator at
1.800.426.4448.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Refer to the product warranty statement included with your LIFEPAK 1000
defibrillator. For duplicate copies, contact your local Medtronic representative.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Glossary 1-13
The following are definitions of terms used throughout this service manual.
■ Automated external defibrillator (AED) — The defibrillator uses an ECG
analysis Shock Advisory System™ (SAS) to advise the device operator if
itdetects a shockable or nonshockable rhythm. Refer to Appendix B of the
operating instructions for additional information. Refer to the operating
instructions – Appendix B.
■ ADAPTIV™ Biphasic — Property of the shock waveform generated by the
LIFEPAK 1000 defibrillator. The energy is delivered in the form of a biphasic
truncated exponential (BTE) defibrillation waveform. LIFEPAK biphasic
defibrillators measure the patient’s transthoracic impedance and
automatically adjust the defibrillation waveform current, duration, and
voltage to meet the needs of the individual patient. Patient impedance is
measured whenever defibrillation electrodes are in contact with the patient.
■ CODE SUMMARY™ Report — A summary report that includes the ECG
segments associated with key events such as analysis or shock. For
additional information, refer to the operating instructions – Data
Management.
■ QUIK-COMBO™ ECG electrodes — An electrode system that allows
delivery of defibrillation therapy to the patient.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Acronyms 1-15
Term Description
AAMI Association for the Advancement of Medical Instrumentation
ADC Analog-to-digital conversion
AED Automated external defibrillator
Ah Ampere hour
AHA American Heart Association
ANSI American National Standards Institute
BTE Biphasic truncated exponential
BF Electrically isolated, external body connection
BPM Beats per minute
CPR Cardiopulmonary resuscitation
DMM Digital Multimeter
ECG Electrocardiogram
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Term Description
EMS Emergency Medical Service
ESD Electrostatic discharge
IEC International Electrotechnical Commission
IrDA Infrared Data Association
LCD Liquid crystal display
LED Light-emitting diode
Li/MnO2 Lithium Manganese Dioxide (battery)
NSR Normal sinus rhythm
NVRAM Non-volatile random-access memory
PC Personal computer
PCB Printed circuit board
PIP Performance inspection procedure
RISC Reduced instruction set computer
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preface
Term Description
RTC Real-time clock
SAS Shock Advisory System
SSD Static-sensitive device
TCP Test and calibration procedure
VF Ventricular fibrillation
VT Ventricular Tachycardia
μA MicroAmpere
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Section Contents
Safety This section describes the general safety conventions, terms, and symbols used
22
Safety 2-2
Terms The following terms are used in this service manual or on the various
configurations of the LIFEPAK 1000 defibrillator. Familiarize yourself with their
definitions and significance.
Danger: Immediate hazards that will result in serious personal injury or death.
Warning: Hazards or unsafe practices that could result in serious personal
injury or death.
Caution: Hazards or unsafe practices that could result in device or property
damage.
Note: Points of particular interest for more efficient or convenient device
operation; additional information or explanation concerning the
subject under discussion.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-3
General Warnings and The following are general warning and caution statements. Keep these warnings
Cautions and cautions in mind when working with the LIFEPAK 1000 defibrillator.
Additional specific warnings and cautions appear throughout this service manual
and the LIFEPAK 1000 Defibrillator Operating Instructions.
WARNINGS!
Shock hazard. The defibrillator delivers up to 360 joules of electrical
energy. Unless properly used as described in these operating instructions,
this electrical energy may cause serious injury or death. Do not attempt to
operate this device unless thoroughly familiar with these operating
instructions, and the function of all controls, indicators, connections, and
accessories.
Shock hazard. Do not disassemble the defibrillator. It contains no
responder-serviceable components and dangerous high voltages may be
present. Contact authorized service personnel for repair.
Shock or fire hazard. Do not immerse any portion of this device in water or
other fluids. Avoid spilling any fluids on device or accessories. Do not
clean with ketones or other flammable agents. Do not autoclave or sterilize
this device or accessories unless otherwise specified.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-4
General Warnings and Possible fire or explosion. Do not use this device in the presence of
Cautions (continued) flammable gases or anesthetics. Use care when operating this device
close to oxygen sources (such as bag-valve-mask devices or ventilator
tubing). Turn off gas source or move source away from patient during
defibrillation.
Possible electrical interference with device performance. Equipment
operating in close proximity may emit strong electromagnetic or radio
frequency interference (RFI) which could affect the performance of this
device. RFI may result in improper device operation, distorted ECG or
failure to detect a shockable rhythm. Avoid operating the device near
cauterizers, diathermy equipment, cellular phones, or other portable and
mobile RF communications equipment. Maintain equipment separation of
at least 1.2 m (4 ft) and do not rapidly key EMS radios on and off. Contact a
technical support representative if assistance is required.
Possible electrical interference. Using cables, electrodes, or accessories
not specified for use with this device may result in increased emissions or
decreased resistance to electromagnetic interference which could affect
the performance of this device or of equipment in close proximity. Use
only parts and accessories specified in these operating instructions.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-5
General Warnings and Possible electrical interference. This defibrillator may cause
Cautions (continued) electromagnetic interference (EMI) especially during charge and energy
transfers. EMI may affect the performance of equipment operating in close
proximity. Verify the effects of defibrillator discharge on other equipment
prior to using defibrillator in an emergency situation, if possible.
Possible device shutdown. Always have access to a spare, fully-charged,
properly maintained battery. Replace the battery when the device displays
a low battery warning.
Possible improper device performance. Using other manufacturers’
cables, electrodes, or batteries may cause the device to perform
improperly and invalidates the safety agency certification and may
invalidate the warranty. Use only the accessories specified in these
operating instructions.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-6
General Warnings and Safety risk and possible equipment damage. Monitors, defibrillators, and
Cautions (continued) their accessories (including electrodes and cables) contain ferromagnetic
materials. As with all ferromagnetic equipment, these products must not
be used in the presence of the high magnetic field created by a Magnetic
Resonance Imaging (MRI) device. The high magnetic field created by an
MRI device will attract the equipment with a force sufficient to cause death
or serious personal injury to persons between the equipment and the MRI
device. This magnetic attraction may also damage the equipment. Skin
burns will also occur due to heating of electrically conductive materials,
such as patient leads. Consult the MRI manufacturer for more information.
CAUTION!
Possible equipment damage. This device may be damaged by mechanical
or physical abuse such as immersion in water or by dropping the device. If
the device has been abused, remove it from use and contact qualified
service personnel.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-7
Symbols The following symbols may be found in this service manual or on the LIFEPAK
1000 defibrillator and its accessories. Some symbols may not be relevant to your
device or used in every country.
Menu button
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-8
Symbols (continued)
Service symbol
This end up
Fragile/breakable
Handle with care
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-9
Symbols (continued)
Single use only
Cable Connector
!USA
For USA audiences only
Power On/Off
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-10
Symbols (continued)
Shock button
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Safety
Safety 2-11
Symbols (continued)
Nonrechargeable battery
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Section Contents
Device
3
Introduction 3-2
About the Device The LIFEPAK 1000 defibrillator (device) is a small, lightweight, portable, battery-
powered therapeutic medical device that provides both semiautomatic and
manual defibrillation operation. When clinically indicated, the device enables the
operator to deliver a brief, high-energy pulse of electricity to the patient’s heart.
Operator interaction is required to defibrillate the patient.
The LIFEPAK 1000 defibrillator uses the patented Medtronic Shock Advisory
System (SAS) to analyze the patient's electrocardiographic (ECG) rhythm. A
suspect rhythm alerts the operator with a priority tone and screen message. For
more information about SAS, refer to the operating instructions –
Appendix B.
The LIFEPAK 1000 defibrillator generates a biphasic truncated exponential
(BTE) shock pulse for defibrillation. The standard method of energy delivery is
through self-adhesive QUIK-COMBO ECG electrodes. When using these
disposable defibrillation electrodes (DDEs), internal circuitry continuously
measures the impedance between the electrodes and allows defibrillation only
when the defibrillation electrodes are attached to the patient.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Operation Modes The device operates in one of three modes: manual mode, automated external
defibrillation (AED) mode, and ECG mode. For additional information refer to
Modes of Operation.
Device Primary The device has four primary functions:
Functions ■ Semiautomatic or Manual (AED) defibrillation.
■ Storage of both patient and service data at each event.
■ Heart Rhythm Analysis
– Displays one nondiagnostic ECG waveform.
– Displays a continuous heart rate readout.
– Analyzes the patient’s ECG rhythm for a potentially shockable ryhthm.
■ Service features including calibration and diagnostic functions.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Controls and Indicators For a description of each feature, click the number displayed in the illustration
below.
9 10
8 1
7 2
6 4
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
ON/OFF key
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
SHOCK key
5 Used to select operating modes (manual or AED) and to enter
information in setup mode.
MENU key
6 Battery Accommodates a single battery pack.
compartment
7 Two softkeys work in conjunction with the screen, providing a
method of making selections while using the defibrillator.
Softkey Softkey functions vary, depending on the task you are
performing at the time. The current function of each softkey is
identified by the label displayed above it on the screen.
8 IrDA port The Infrared Data Association port provides wireless
communication for transferring data from the defibrillator to a
PC.
9 LCD display Displays pertinent information for use during all modes of
operation.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
ECG waveform
Softkeys
Heart rate indicator — The heart rate indicator displays heart rates between 20
and 300 BPM. This indicator is present only during AED monitoring, in manual
mode, or when the 3-wire ECG cable is used.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Battery status symbol — When the defibrillator is ON, the battery symbol
appears on the screen indicating the relative level of charge. One bar indicates
the battery is low. When the battery is very low, the symbol is blank and a
REPLACE BATTERY message appears on the screen.
ECG — The ECG appearing on the screen is a nondiagnostic ECG, obtained by
means of the therapy electrodes or the Lead II ECG cable. The presence of an
ECG does not ensure that the patient has a pulse.
Softkey labels — These labels are displayed on the LCD display above the left
and right softkeys and define the function that can be activated by pressing that
softkey (for example, ANALYZE and DISARM).
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
The table below summarizes supplies and accessories that are available. For
ordering instructions, refer to Ordering Parts.
Item Description MIN CAT.
QUIK-COMBO™ Electrodes with 3202674 11996-000017
REDI-PAK™ preconnect system
Infant/child reduced energy 3202380 11101-000016
defibrillation electrodes (not
compatible with QUIK-COMBO
defibrillation cable)
Infant/child electrodes starter kit 3202784-004 11101-000017
(English, Dutch, French, German,
Spanish, Italian, Danish, Norwegian,
Finnish, Swedish)
Infant/child electrodes starter kit 3202784-005 11101-000018
(English, Hungarian, Polish, Brazilian
Portuguese, Iberian Portuguese,
Spanish, Korean, Japanese,
Mandarin Chinese)
LIFEPAK 1000 nonrechargeable 3205379-002 11141-000101
lithium manganese dioxide battery
pak
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Device Description
Component Description
W01 QUIK-COMBO 7 conductor cable that connects the A03 upper enclosure to the
Harness A01 Analog PCB.
W02 Speaker Used to annunciate device warnings, alarms, tones, and voice
prompts.
W03 Battery Harness Connects the Battery Pack to the A01 Analog PCB
W05 Readiness Provides the user with status information at a glance through
Display the display of symbols.
W06 Header 48 pin header connector that connects the digital signals
between the A01 Analog PCB and the A02 Digital PCB.
W07 High-Energy A metallized film capacitor used for energy storage.
Capacitor
W08 Inductive Dissipates energy from the W07 High-Energy Capacitor.
Resistor
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Operating Instructions
Operating The LIFEPAK 1000 Defibrillator Operating Instructions familiarize the operator
4
with basic device functions and identify controls, indicators, and connectors.
Instructions Qualified service personnel must consult both the operating instructions and this
service manual for a complete understanding of the use and maintenance of the
device.
WARNINGS!
Possible improper device performance. Use only Medtronic QUIK-COMBO
electrodes and LIFEPAK 1000 battery pak described in this service manual.
Substitution may cause the device to operate improperly.
Possible loss of power during patient care. Using an improperly
maintained battery pak to power the LIFEPAK 1000 defibrillator may cause
premature power loss.
■ Data Management
– Managing Defibrillator Data
■ Caring for the LIFEPAK 1000 Defibrillator
– Maintenance and Testing Schedule
– Self-Test Performance
– Inspection
– Cleaning
– Battery Maintenance
– Electrode Storage
– Service and Repair
– Product Recycling Information
– Supplies, Accessories, and Training Tools
– Warranty Information
■ Appendix A Specifications
■ Appendix B Shock Advisory System
■ Appendix C cprMAX Technology
■ Appendix D Changing Setup Options
■ Appendix E User’s Checklist
Modes of The LIFEPAK 1000 defibrillator can be turned ON in AED mode (default),
5
manual mode, or setup mode. ECG mode is an option that can also be
Operation displayed. Service mode is used for testing and inspection procedures. Choose
from the following links to learn more about a particular mode.
AED Mode
Manual Mode
ECG Mode
Setup Mode
Service Mode
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Modes of Operation
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Modes of Operation
The ECG mode option displays a nondiagnostic ECG waveform and heart rate
indicator when the ECG cable is connected and electrodes are applied. In ECG
mode, the defibrillator’s shock capability is disabled; however, the device
continues to analyze the patient’s ECG for a potentially shockable rhythm.
Note: If the ECG waveform is not visible in AED mode, the device is either not
configured for the ECG mode option or ECG DISPLAY isn’t set to ON
(default) in the SETUP/AED MODE menu.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Modes of Operation
Accessing the Setup Setup mode is used to configure default settings and view service options.
Menu (continued) To access the SETUP menu:
1. The device should be OFF, and no cable should be connected.
2. Press and hold both SOFTKEYS, and then press the ON/OFF key. When the
SETUP menu appears, release the SOFTKEYS.
3. Use the SOFTKEYS to navigate and make selections on the SETUP menu.
4. To exit the SETUP menu, turn the defibrillator OFF.
The following table lists the SETUP menu options. Refer to the operating
ON/OFF
instructions – Appendix D for compete description of setup options.
SOFTKEYS MENU key key
Option Description
General Set up general device options
AED Mode Set up AED mode defaults
Manual Mode Set up manual mode defaults
Service Mode View service options
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Modes of Operation
Accessing the Service Service mode is a password-protected menu. In the Service mode, the operator
Menu can perform an energy calibration on the device or execute LCD pixel display
test, as well as view and clear device logs.
To access the SERVICE menu:
1. Access the SETUP menu.
2. Select SERVICE MODE from the SETUP menu.
3. Select YES to enter the password.
4. To enter the four-digit password, 5433, press INCREASE or DECREASE to
select each digit, and then press the MENU key to set each entry. When the
last number is entered, the SERVICE menu appears.
Note: When entering service mode, if you wait longer than five seconds to
make an entry, the screen will revert to the SETUP menu.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Modes of Operation
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Section Contents
TABLE OF CONTENTS
PIP – Scope and Applicability
PIP – Resource Requirements
PIP – Test Equipment Requirements
PIP – Instructions
PIP – Checklist
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
This section describes the requirements for PIP equipment, PIP test equipment
verification, PIP workstation power, and PIP personnel qualifications.
PIP – Equipment To perform the PIP, use the equipment listed in the PIP – Test Equipment
Requirements table. Although the table lists specific test equipment by
manufacturer, test equipment with equivalent specifications may be substituted.
PIP – Test Equipment All test equipment used to perform the PIP must have a current calibration label.
Verification The calibration label must be issued by a certified calibration facility.
PIP – Workstation If ac power if required for testing, the ac line power to the workstation must be
Power connected to a grounded power source.
PIP – Personnel Personnel who perform the PIP must be familiar with the operation of the
LIFEPAK 1000 Defibrillator and must meet at least one of the following
requirements (or the equivalent):
• ASSOCIATE OF APPLIED SCIENCE, WITH AN EMPHASIS IN BIOMEDICAL ELECTRONICS
• CERTIFICATE OF TECHNICAL TRAINING, WITH AN EMPHASIS IN BIOMEDICAL ELECTRONICS
• EQUIVALENT BIOMEDICAL ELECTRONICS EXPERIENCE
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
Safety Analyzer 110 or 220 Vac line voltage Dale Model 600/600E, 601/601E or
Current range: 0-1999 μA equivalent*
Current accuracy: +/-5% of reading +/-1
digit
Electrode test Post Adapter MERS -P/N 3205979-000
(2 each)
Cable Assembly, 3 wire ECG, MERS -P/N 3204273-002, -003
LIFEPAK1000
*Equivalent equipment is required to meet the specifications listed in the specifications column.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
PIP – General Instructions This section lists the general instructions for performing the Performance
Inspection Procedure (PIP).
• All PIP tests applicable to the LIFEPAK 1000 Defibrillator configuration under
test must be performed.
• See Troubleshooting section of the Service Manual to correct failures
and Service Log diagnostic codes 4000 and above, then repeat the PIP.
Note: All diagnostic codes below 4000 are informational and can be cleared.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
PIP – Physical Inspection Perform the exterior physical inspection, described as follows
1. Remove device from softcase. (if applicable)
2. Inspect all exterior surfaces of the Device for damage.
3. Lift and turn over the Device and listen for loose or rattling internal hardware.
If loose hardware is suspected, investigate and correct or contact MERS
before performing this procedure.
4. Inspect the case for separation (integrity of case seal).
5. Check for bent, broken, or corroded therapy connector sockets or therapy
cable pins. Clean or replace damaged components as required.
6. Check date sensitive materials (electrodes, batteries) to ensure they have
not exceeded their shelf life date.
Note: Whenever possible, use the customer owned cables and accessories
when performing the PIP.
Note: Damage to softcase or date sensitive materials exceeding shelf life are
not reasons for failing the PIP.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
1 - Battery Installation Test 1. Verify no battery installed into the Device, no electrodes connected to the
device.
Readiness Indicator 2. Wait 60 seconds before installing the battery into the Device.
3. Install battery into the Device. Verify the Device completes a brief power
initialization cycle, which includes the following:
– The Readiness Indicator graphics toggle from a unpowered state to the
final powered state as shown below:
OK INDICATOR
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
2 - Power Up Initialization Test 1. Verify Battery Pack is installed into the Device, no electrodes connected to
the device and Service indicator is OFF, OK indicator is ON.
2. Turn the device ON to initiate the device initialization self- test.
– Verify the OK INDICATOR flash momentarily ON and back OFF
– Verify the SHOCK LEDs flash momentarily ON and back OFF.
– Verify ON indicator glows steadily while Device is on.
– Verify the LCD display does not show “Replace Battery message”.
3. Verify the Voice prompt is heard “Connect Electrodes”.
4. Turn the Device OFF.
Initalization Display Screen
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
3 - Service Alert test / Audio- 1. Verify Battery Pack is installed into the Device, no electrodes connected to
Voice prompt tests the device and Service indicator is OFF, OK indicator is ON
2. Hold down L SOFTKEY and R SOFTKEY controls and press ON (green) button
on the Device. Hold the controls until SETUP menu appears.
3. Select GENERAL from the Setup menu. Select Audio from the Setup/General
menu. Select Service Alert from the Setup/General/ Audio menu
4. Select Service Alert ON.
L SOFTKEY R SOFTKEY MENU KEY
5. Verify that several audio beeps are clearly audible.
Note: As required, reset the Service Alert to the customer original setting.
6. Select Prompt Volume from the Setup/General/ Audio menu.
7. Verify that the Prompt Volume is set to High. If not already set, press Next
and Select to set Prompt Volume to High.
8. Turn the Device off then on, and verify the sound prompt produced by
Device has no audible distortion.
Note: As required, reset the Prompt Volume to the customer original setting.
9. Turn the Device OFF.
L SOFTKEY R SOFTKEY
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
4 - Defibrillator Energy Output 1. Verify Battery Pack is installed into the Device, no electrodes connected to
/ Charge Time / Buttons and the device and Service indicator is OFF, OK indicator is ON.
Keys test / Defibrillator Energy
2. This test will exercise all keys, verify that all control keys on the Device are
Dump test.
functioning properly during performance of this test.
3. Place the Device in Service Mode using the following steps:
a) Hold down L SOFT KEY and R SOFT KEY controls and press ON (green)
button on the Device. Hold the controls until SETUP menu appears.
b) Press the Next key until Service Mode is highlighted.
c) Select Service Mode from the Setup menu.
d) Select YES to enter Service Mode - Password screen.
e) Enter 4 digit password (5433) to enter SERVICE MODE.
L SOFTKEY R SOFTKEY MENU KEY f) Select DEFIB CAL from the on-screen SERVICE menu to access the Service/
Defib Cal Sub-Menu
Therapy
Cable
Defib. +
Analyzer
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
4 - Defibrillator Energy Output Note: While entering Service Mode, if you wait longer than 5 seconds to make
/ Charge Time / Buttons and an entry, the screen will revert to the General Setup screen. In this case restart
Keys test / Defibrillator Energy the test from step b.
Dump test (continued).
4. Connect the therapy cable into the Defibrillator Analyzer.
5. Set the Defibrillator Analyzer to the highest joule scale setting, select Energy.
6. Select START on the Device to initiate the first calibration energy level (100J).
7. Press the SHOCK button to discharge the device as soon as it is fully
charged.
8. Verify that the Defibrillator Analyzer indicates the delivered energy at 100J is
within 95 J and 105 J. Perform the TCP-Energy Calibration if the delivered
energy falls outside of the acceptable output range.
9. Press Menu key four times to retain the current calibration value if the
delivered energy falls within the acceptable output range.
Warning: Do not touch the Increase or Decrease buttons on the Device during
the verification. Failure to do so may lead to the Device energy
miscalculations.
10. Press OK and start Stopwatch simultaneously (use Stopwatch for first 360 J
charge time).The Device will proceed to the next calibration level (360 J).
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
4 - Defibrillator Energy Output 11. Stop the Stopwatch as soon as the Device’s SHOCK button starts to flash.
/ Charge Time / Buttons and 12. Verify that the Device charges to 360 J in 14 seconds or less.
Keys test / Defibrillator Energy
Dump test (continued). 13. Press the SHOCK button to discharge the Device as soon as it is fully
charged.
14. Verify that the Defibrillator Analyzer indicates the delivered energy at 360 J is
within 342 J and 378 J. Perform the TCP-Energy Calibration if the delivered
energy falls outside of the acceptable output range.
15. Press Menu key four times to retain the current calibration value if the
delivered energy falls within the acceptable output range. Press OK.
16. Repeat steps 6,7,8,9,10,13,14 and 15 to verify Defibrillator Energy Output
for a second time are within the acceptable output limits.
17. This time repeat steps 6,7,9, and 10.
18. When full 360 J charge is reached, turn the Device OFF and then turn the
Device back ON.
19. Verify Energy dump test doesn’t result in service wrench (Service indicator is
OFF).
20. Turn the device OFF.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
5 - Date & Time / Contrast / 1. Verify Battery Pack is installed into the Device, no electrodes connected to
Full Monthly Auto test the device and Service indicator is OFF, OK indicator is ON.
2. Hold down L SOFTKEY and R SOFTKEY controls and press ON (green) button
on the Device. Hold the controls until SETUP menu appears.
3. Select GENERAL from the Setup menu.
4. Verify that the Date/Time is correct and contrast of characters displayed
on LCD are legible from the Setup/General.
5. Select Date/Time from the Setup/General menu.
6. Set Date to 2004 / Jan / 05, which is the first Monday of that month.
L SOFTKEY R SOFTKEY
7. Set Time to 02:59.
8. Immediately turn the Device off. Then observe the Device performing Full
(Monthly) Auto test at 03:00.
9. Wait for 80 seconds; look for the power ON LED and SHOCK button to blink.
10. Verify OK returns to the Readiness Indicator.
11. Reset date and time to correct time from the Setup/General/Date/Time menu.
12. Verify no diagnostic codes (except codes below 4000) in Service Log.
L SOFTKEY R SOFTKEY
(Service Log is in the Service Mode menu) See PIP - General Instructions
13. Turn the device OFF.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
6 - Electrode Impedance test 1. Verify Battery Pack is installed into the Device, electrodes connected to the
device and Service indicator is OFF, OK indicator is ON.
2. Connect the Therapy Cable to the Decade Resistance Box.
3. Set the Decade Resistance box to 355 Ω
4. Using DMM to verify the Decade Resistance box is 355Ω+/−5 Ω.
5. Turn the device ON. Verify that the Device voice prompts:
• CONNECT ELECTRODES
Note: Actual voice prompts will vary depending on the device configuration.
Therapy Cable Note: Do not discharge into the Decade Resistance Box.
6. Set the Decade Resistance Box to 246 Ω .
7. Using DMM to verify the Decade Resistance box is 246 Ω +/−4 Ω.
8. Verify that the Device voice prompts:
Test Post
Adapters (2 ea.) • STAND CLEAR OR PUSH ANALYZE
9. Set the Decade Resistance Box to 5 Ω .
Decade
Resistance Box 10. Using DMM to verify the Decade Resistance box is 5 Ω +/−0.2 Ω.
11. Verify that the Device voice prompts:
• CONNECT ELECTRODES.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
7 - Warm Boot Detection Test 1. Verify Battery Pack is installed into the Device, no electrodes connected to
the device and Service indicator is off. If Device is OFF, OK indicator is ON.
2. If OFF, Turn the Device ON.
3. Verify Readiness indicator OK and Wrench are displayed momentarily.
4. Verify Readiness display is updated.
5. Note the current count on the elapse timer, found at top of LCD display.
6. Turn the Device OFF.
7. Turn the Device ON between 5 - 49 seconds after Device shutdown.
8. Verify that the Device returns to normal ON operation and the elapse timer
has continued to increment counter during the Device shutdown period (The
Device does not restart time count from 00:00).
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
8 - ECG Leads Characteristic N/A - PIP Checklist test step if 3 Lead ECG feature isn’t installed in Device.
test 1. Verify Battery Pack is installed into the Device, no electrodes connected to
the device and Service indicator is off. If Device is OFF, OK indicator is ON.
2. If OFF, Turn the Device ON.
3. Set the device to AED mode or verify the device is on AED mode.
4. Verify the device responds with a voice prompt: “Connect electrodes”
5. Connect the 3 lead ECG cable to the device.
6. Verify that the Device displays ECG mode and “Connect ECG leads”
message.
7. Connect the 3 lead ECG cable to Defibrillator Analyzer.
ECG 8. Verify that the Device displays “ECG mode” message and a scrolling ECG
Leads
Cable RA LA LL
trace.
9. Set Defibrillator Analyzer to 1mV, 10Hz Sinewave.
10. Use the graticule to measure the sine-wave amplitude* of 10 mm +/- 10%
Defib. + peak-to-peak on the LCD display.
Analyzer
Note*: The output of QED 6 is fixed at 1mV Lead II (RA-LL)
Note: If no ECG waveform is displayed, then Device is configured with ECG
Monitoring - Off.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
8 - ECG Leads Characteristic 11. Set Defibrillator Analyzer to ECG wave and 120 BPM.
test (continued) 12. Verify that the Device displays the Heart Rate on LCD Display is within 117
to 123 BPM.
13. Remove RA lead from Defibrillator Analyzer.
14. Verify that the Device displays “Connect ECG Leads” message within 1
second.
Note: Do not touch lead contacts with hand, this could interrupt leads off
condition.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
9 - Therapy ECG 1. Verify Battery Pack is installed into the Device, no electrodes connected to
Characteristic test the device and Service indicator is off. If Device is OFF, OK indicator is ON
2. Set Defibrillator Analyzer to ECG Wave and 120 BPM.
3. Connect Therapy cable to Device and Defibrillator Analyzer as shown.
4. If OFF, turn the Device ON.
5. Use the graticule to measure the ECG R -wave amplitude* of 10 mm +/- 10%
peak-to-peak on the LCD display
Note*: The output of QED 6 is fixed at 1.1 mV (Apex-Sternum)
Note: If no ECG waveform is displayed, then device may be configured with ECG
L SOFTKEY R SOFTKEY MENU KEY
Display - Off. Go to Setup/General/AED mode menu set ECG Display to ON. As
required, reset the ECG Display to the customer original setting
Therapy
Cable
10 mm
Defib. +
Analyzer
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
10 - Leakage Current Check to ensure the leakage current meets the following industry standards:
• AAMI/ANSI (Association for the Advancement of Medical Instrumentation/
American National Standards Institute) DF2-1989, DF39-1993
• IEC (International Electrotechnical Commission) 601-1 and 601-2-4
WARNING
Shock Hazard. Failure to properly perform these tests could result in a
failure to detect excessive leakage current. Make sure you are familiar with
your test equipment and these test performance procedures.
Note: Because of the variety of safety analyzers that may be used for these
tests, this electronic service manual provides only general instructions. For
information about configuration and testing methods, refer to your safety
analyzer operating instructions.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
10a - Therapy Leakage 1. Verify Battery Pack is installed into the Device, no electrodes connected to
Current the device and Service indicator is off. If Device is OFF, OK indicator is ON.
2. Connect the Quik-combo to ECG Snap Cable into the Device.
3. Connect the Apex/LL of Quik-combo to ECG snap cable to the Safety
Analyzer at LL snap.
4. Connect the Sternum/RA of Quik-combo to ECG snap cable to the Safety
Analyzer at RA snap.
5. Set the Safety Analyzer Lead setting to RA or LL.
Therapy
6. If off, turn the device ON.
Leakage
Test cable 7. Verify that the Lead- Lead Leakage Current (RA-LL) is between 1μA to 9
μA.
8. Remove the Quik-combo to ECG snap cable from the Device and Safety
Analyzer.
Safety
Analyzer
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
10b- ECG Leads Leakage N/A - PIP Checklist test step if 3 Lead ECG feature isn’t installed in Device.
Current Test 1. Verify Battery Pack is installed into the Device, no electrodes connected to
the device and Service indicator is off. If Device is OFF, OK indicator is ON.
2. Connect the 3 Lead ECG Test Cable into the Device.
3. Connect the LL Lead of 3 Lead ECG Test Cable to the Safety Analyzer at LL
snap.
4. Connect the RA Lead of 3 Lead ECG Test Cable to the Safety Analyzer at
RA snap.
ECG 5. Set the Safety Analyzer Lead setting to RA or LL.
Leads
cable 6. If off, turn the device ON.
7. Verify that the Lead - Lead Leakage Current (RA - LL) is between 0μA to 9
μA
8. Remove the LL Lead of 3 Lead ECG Test Cable from the Safety Analyzer.
Safety 9. Connect the LA Lead of 3 Lead ECG Test Cable to the Safety Analyzer at
Analyzer
LA snap.
10. Set the Safety Analyzer Lead setting to RA or LA.
11. Verify that the Lead - Lead Leakage Current (RA- LA) is between 0μA to 9
μA.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Performance Inspection Procedure
10b - ECG Leads Leakage 12. Remove the RA Lead of 3 Lead ECG Test Cable from the Safety Analyzer.
Current Test (continued). 13. Connect the LL Lead of 3 Lead ECG Test Cable to the Safety Analyzer at LL
snap.
14. Set the Safety Analyzer Lead setting to LA or LL.
15. Verify that the Lead - Lead Leakage Current (LA- LL) is between 0μA to 9
μA.
16. Remove the 3 Lead ECG Test cable between the Device and Safety
Analyzer.
17. Turn the Device OFF.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Section Contents
Test and This section covers the Test and Calibration Procedures (TCP). Complete the
relevant TCP to correct out-of-specification conditions detected during the
Calibration Performance Inspection Procedure (PIP). Note: Any time the LIFEPAK 1000
Procedures defibrillator is opened for repair, component replacement, or calibration, the
Device must successfully pass all portions of the closed-case PIP before being
returned to active service. If a Device exhibits any fault code, contact Medtronic-
ERS; do not place it back in service.
The TCP applies to the LIFEPAK 1000 defibrillator exclusively. You may perform
the procedures outlined in this section in any order.
For Display issues, perform the LCD Screen Test.
When the Analog PCBA is replaced, perform Defibrillator Energy Calibration,
and Device Error Log check.
When the Digital PCBA is replaced, perform Device Error Log check.
Any time the LIFEPAK 1000 defibrillator is opened for repair or component
replacement, or after calibration, the Device must successfully pass the entire
Performance Inspection Procedure before it is returned to active service.
Refer to Resource Requirements for necessary equipment, test equipment
verification, workstation power, and TCP personnel qualifications.
Refer to Test Equipment Requirements for a listing of test equipment,
(including specifications) required to complete the TCP.
See Troubleshooting section of the Service Manual to correct failures and
Service Log diagnostic codes 4000 and above, then repeat the PIP.
Note: All diagnostic codes below 4000 are informational and can be cleared.
Note: Throughout the body of this TCP, the LIFEPAK 1000 defibrillator is
referred to as the Device.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
This section describes the requirements for TCP equipment, TCP test
equipment verification, TCP workstation power, and TCP personnel.
TCP – Equipment To perform the TCP, you must use the equipment listed in the Test Equipment
Requirements table. Although the table lists specific test equipment by
manufacturer, test equipment with equivalent specifications may be substituted.
TCP – Test Equipment All test equipment used to perform the TCP must have a current calibration
Verification label. The calibration label must be issued by a certified calibration facility.
TCP – Workstation The ac line power to the workstation must be connected to a grounded power
Power source.
TCP – Personnel Personnel who perform the TCP must be thoroughly familiar with the operation
of the Device and must meet at least one of the following requirements (or the
equivalent):
• Associate of Applied Science, with an emphasis in biomedical electronics
• Certificate of Technical Training, with an emphasis in biomedical electronics
• Equivalent biomedical electronics experience
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Perform this test when display problem is suspected or when LCD display and
display cabling have been replaced.
• Before beginning test, verify the SERVICE INDICATOR is covered by black
rectangle, the OK INDICATOR is displayed, the BATTERY CAPACITY
INDICATOR has one or more segments displayed.
To perform LCD SCREEN TEST:
1. Place the Device into Service Mode by following the steps below:
L SOFTKEY R SOFTKEY
a. Hold down L SOFT KEY and R SOFT KEY controls and turn on the Device. Hold the
controls until SETUP menu appears.
b. Press the Next key until Service Mode is highlighted.
c. Select Service Mode from the Setup menu.
d. Select YES to enter Service Mode- Password screen.
e. Enter 4 digit password (5433) to enter SERVICE MODE.
L SOFTKEY R SOFTKEY
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Shock hazard. Avoid contact with the energy meter. Dangerous voltages
will be present on energy meter electrode plates/posts.
• Before beginning test, verify the SERVICE INDICATOR is covered by black
rectangle, the OK INDICATOR is displayed, the BATTERY CAPACITY
INDICATOR has one or more segments displayed.
Perform Defibrillator Energy Calibration when a component in the energy
delivery path is replaced or if Defib. Energy Output is out of spec:
1. Place the Device into Service Mode by following the steps below:
L SOFTKEY R SOFTKEY
a. Hold down L SOFT KEY and R SOFT KEY controls and turn on the Device. Hold the
controls until SETUP menu appears.
b. Press the Next key until Service Mode is highlighted.
c. Select Service Mode from the Setup menu.
d. Select YES to enter Service Mode - Password screen.
e. Enter 4 digit password (5433) to enter SERVICE MODE.
L SOFTKEY R SOFTKEY
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Note: While entering Service Mode, if you wait longer than 5 seconds to make
an entry, the screen will revert to the General Setup screen. In this case restart
the test from step b.
2. Connect the therapy cable between the Device and Defibrillator Analyzer.
Therapy
Cable
3. Set the Defibrillator Analyzer to the highest joule scale (1000J).
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Current
I1
TPHASE2 Time
TPHASE1
I4
Note 3.
(I1 – I4 )
Tilt = -----------------------
-X100
I1
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
L SOFTKEY R SOFTKEY
2. Select SERVICE LOG on Service menu.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Test and Calibration Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Section Contents
Troubleshooting The Troubleshooting section describes error code usage, interpretation, and
8
Introduction The LIFEPAK 1000 defibrillator logs an error code and exhibits the SERVICE
symbol on the readiness display whenever an internal program or process fails
to execute properly. Errors rarely occur and should be investigated thoroughly
by qualified service personnel before the device is placed back into active use.
Always complete the Performance Inspection Procedure (PIP) after
encountering and clearing any error code(s).
Error codes stored in the Service Log may not necessarily indicate a
permanent error. Error codes can indicate transient electromagnetic interference
(EMI) or electrostatic discharge (ESD). If you suspect transient EMI or ESD as
the source of an error, clear error code(s), turn the device off and then back on.
If the error code does not reoccur, it may be the result of EMI or ESD.
Processing Error Codes To process an error code:
1. Note any problems with the defibrillator and consult the Troubleshooting
Chart.
2. Review error codes by displaying the Service Log. Record any errors,
including the date, time, error, and error extension.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Processing Error Codes 3. Complete the Performance Inspection Procedure (PIP). If completed
(continued) successfully, the device may be returned to regular use.
4. If the device exhibits the SERVICE symbol on the readiness display at any
time during the PIP, stop the PIP, then continue with Step 5.
Error Code Category 5. Compare your PIP failure with the Troubleshooting Chart.
-OR-
4101-4204 Page 8-6
Compare the error codes against the Error Code Categories listed in the
4301- Page 8-7 table on the left, and click the appropriate category to find the correct error
4601- Page 8-9 code in the Error Code Table.
6. Find the error code in the Error Code Table.
4701- Page 8-11 7. Read the corrective action(s). If the corrective action calls for the
4811- Page 8-13 replacement of a part, click on that part caption in the footer at the bottom of
the page to jump to the corrective action process.
4904-4a03 Page 8-14 8. Service the device based on these inputs, and then repeat the PIP.
8101-8902 Page 8-15 9. For persistent error codes, contact your local Medtronic service or sales
representative.
8b01-ffff Page 8-16
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
430a CONTROL NORMAL MODE CHARGE Charging did not complete within 30 Check HV Capacitor connection. If
TIMER EXPIRED seconds. failure persists, replace A01 Analog
PCB.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Troubleshooting
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Section Contents
Preventive Periodic maintenance, inspection, and testing of the LIFEPAK 1000 defibrillator
9
helps detect and prevent possible electrical and mechanical problems. Refer to
Maintenance the operating instructions – Operator Checklist for additional items. For
information about battery-related topics, refer to Battery Maintenance.
Maintenance and Testing Schedule
Guidelines
Cleaning
Device Useful Life
Device Support Policy
Recycling Information
Self-Tests Each time you turn the device ON, a five-second, internal self-test occurs to
ensure proper functioning of the internal electrical components and circuits. If a
problem is detected that requires immediate service, such as a malfunctioning
charging circuit, the CALL SERVICE message appears on the LCD display and the
SERVICE symbol appears on the readiness display. If necessary, the device can
still be used for a cardiac emergency. However, the problem should be corrected
as soon as possible. The SERVICE symbol will remain visible until the problem is
corrected.
When the device is OFF, an automatic self-test occurs daily and monthly at
3:00 a.m. During the automatic self-test, the defibrillator turns ON briefly and
completes a self-test. It then turns OFF. If the device is already ON at 3:00 a.m.,
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Self-Tests (continued) or if a battery is not installed, the automatic self-test will not occur. If the
defibrillator is turned ON while a self-test is in progress, the test is halted and the
defibrillator will turn on normally.
Test and Service Data The results of the self-test and auto tests are stored in a test log. HOW DO
THEY VIEW AND CLEAR THE TEST LOG?
If the LIFEPAK 1000 defibrillator requires service as indicated by testing,
troubleshooting, or the readiness display service symbol, contact authorized
service personnel.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Maintenance and Periodic maintenance, inspection, and testing of the device will help prevent
Testing Schedule possible electrical and mechanical problems. Refer to the operating
instructions – Operator Checklist for additional items.
Guidelines The following table shows the schedule for preventive maintenance activities.
For items that should be replaced at regular intervals, refer to scheduled
replacement Items shown below.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Cleaning 9-5
The tools and materials that you will need to perform an external and internal
cleaning of the LIFEPAK 1000 defibrillator are shown below.
Product Description
Static-discharge-protected work Grounded conductive surface and wrist
area strap
Isopropyl alcohol
Quaternary ammonium compounds
Peroxide (peracetic acid) solutions
Cotton swabs
Vacuum cleaner
Soft-bristle brush Nonmetallic
Cloth Clean and lint-free
Compressed air Clean and dry (60 psi, max.)
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
CAUTION!
Possible case damage. Do not clean any part of this device or accessories
with bleach, bleach dilution, or phenolic compounds. Do not use abrasive
or flammable cleaning agents. Do not attempt to sterilize this device or any
accessories unless otherwise specified in accessory operating
instructions.
Clean the exterior of the LIFEPAK 1000 defibrillator by wiping the surface with
any of the following solutions:
■ Soap and water
■ Quaternary ammonium compounds
■ Isopropyl alcohol
■ Peroxide (peracetic acid) solutions
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Cable Cleaning Clean the ECG cable by wiping them with a 70% isopropyl alcohol pad and
Procedure allowing them to dry. Do not soak or immerse the cables in any liquid solution.
Do not attempt to sterilize them.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
CAUTION!
Possible case damage. Do not clean any part of this device or accessories
with bleach, bleach dilution, or phenolic compounds. Do not use abrasive
or flammable cleaning agents. Do not attempt to sterilize this device or any
accessories unless otherwise specified in accessory operating
instructions.
In a static protected work environment, clean the interior of the LIFEPAK 1000
defibrillator as described below.
1. Brush interior surfaces and parts with a nonmetallic, soft-bristle brush.
2. Remove loosened dirt and dust using a dry, low-pressure compressed air
(60 psi) or vacuum cleaner.
3. Wipe metal surfaces with a soft, nonabrasive cloth that has been dampened
with isopropyl alcohol.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Preventive Maintenance
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Section Contents
Battery Follow the guidelines described in this section to help maximize battery life and
10
performance.
Maintenance Introduction
Maintaining the Battery Pak
Battery Shelf Life
Disposal/Recycling
CAUTION!
Possible battery damage. Electrical connection between battery contacts
can blow an internal fuse and permanently disable the battery.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Battery Maintenance
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator/Monitor Battery Maintenance
Battery Shelf Life A new nonrechargeable battery pak (stored outside the defibrillator) has a shelf
life of approximately five years if stored at the proper temperature (20° to 30°C ).
The battery pak (stored outside the defibrillator) self-discharges over time;
therefore, when the battery is eventually placed in the defibrillator, its useful life
will be reduced depending on how long it “sat on the shelf.”
Disposal/Recycling Nonrechargeable batteries should be fully discharged before disposal.
Before disposing of nonrechargeable battery paks, cover the battery terminals
with the plastic cap provided with the new battery.
Follow your national, regional, and local regulations for disposal. Contact a local
Medtronic representative for more information.
In the USA, Environmental Protection Agency and Department of Transportation
regulations allow disposal of nonrechargeable batteries with ordinary household
waste provided they are fully discharged. Be sure to comply with any other local
or regional regulations before disposal. For more information or assistance,
contact your local Medtronic representative or call 1.800.442.1142.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Section Contents
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
The following general warnings and cautions apply to all actions you may
perform during maintenance of the LIFEPAK 1000 defibrillator.
WARNINGS!
Shock hazard. Servicing of this device must be performed by properly
trained individuals. This device may retain potentially lethal charges
accessible inside the device at any time–even when off. Follow the
procedures carefully for discharging the W07 High-Energy Capacitor.
Shock hazard. The W07 High-Energy Capacitor carries high voltage.
Discharge the capacitor before handling.
Possible shock and device damage. It is possible to pinch and damage
wires during reassembly. To avoid pinching wires, carefully follow
reassembly instructions.
CAUTION!
Possible component damage. The PCB assemblies contain static-sensitive
devices (SSDs). To avoid damage, observe the special handling practices
described in Static-Sensitive Devices (SSD).
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
About SSD Handling Many electronic semiconductor devices (such as MOS ICs, FETs, optical
isolators, or film resistors) can be damaged by the discharge of static electricity.
Static charge buildup is very common. Static discharges commonly occur when
the operator wears synthetic clothes and transfers the charge to any object
touched. These discharges can damage or destroy static sensitive devices
(SSDs). In most cases, the discharge is not even perceptible to the person who
causes it.
To prevent static discharge damage to SSDs, observe the following precautions
during any open-case test, maintenance, or repair procedures:
The SSD Symbol SSDs are identified with the following warning symbol:
ATTENTION
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Wear a Wrist Strap Always wear a conductive wrist strap connected to the mat and to ground except
when working on energized equipment or when discharging high-voltage
circuits. The strap must be snug enough to make good contact against bare skin.
WARNING!
Shock hazard. Remove the wrist strap when working on energized
equipment or when discharging high voltage circuits.
Transport and Store Transport and store PCBs in anti-static racks or inside conductive bags. Label
PCBs Properly the package that contains the PCBs as static-sensitive.
Keep Work Area Keep static-generating products, such as styrofoam cups or trays, away from the
Static-Free work area. Connect all electrical equipment, such as soldering irons and test
equipment, to ground with a three-prong plug.
Test Work Area Test all the antistatic parts of the work area (mat, straps, cables) routinely. It is
Routinely recommended that you keep a log of the test results.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Viewing the Device ID Before beginning any repair action, the Device ID configuration must be
Configuration Code reviewed and noted to ensure that it remains the same after completing any
repair procedure.
Note: Refer to the Device Configuration Table table for a description of each
digit.
To view the device configuration code:
1. Access the SETUP menu.
2. Select GENERAL from the SETUP menu.
3. Select DEVICE DATA from the SETUP/GENERAL menu.
4. View the 20-digit Device Configuration code and record it at the top of the
PIP Checklist.
5. When device repairs are complete, check the Device Configuration code on
the SETUP/GENERAL/DEVICE DATA menu to make sure it matches the device
Config. code entered on the PIP Checklist.
6. If the configuration matches the PIP Checklist, turn the device OFF and
continue with the PIP procedure.
7. If the configuration does NOT match the PIP Checklist, refer to Reentering
the Device Configuration.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Building a Capacitor Construct a capacitor discharge tool as illustrated to discharge the W07 High-
Discharge Tool Energy Capacitor. The following materials are required:
■ 10 kΩ, 2 W resistor (ten 1 KΩ, 2 W), high-voltage
neon lamp ■ 5 MΩ, 5 W resistor, high-voltage
RTV silicone
5 MΩ, 5 W rubber sealant ■ Neon lamp, NE76, NE2, or NE2H
resistor ■ 8 AWG copper wire
■ Clear plastic tubing, capable of insulating 10 kV
clear
plastic
10 KΩ, 2 W ■ 10 kV insulation
tubing
resistor
■ RTV, silicone rubber sealant
RTV silicone
rubber sealant
Refer to Using the Capacitor Discharge Tool on the next page to discharge
the high-energy capacitor.
6 inches
10 kV insulation
8 AWG wire
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
CAUTION:
5. Observe the neon lamp inside the capacitor discharge tool. If a charge of
approximately 90 volts is present, the neon lamp will light.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Lens Assembly Note: The A05 Lens Assembly cannot be reused. If removed, install a new
Removal Lens Assembly.
The A05 Lens Assembly should be removed from the upper enclosure first if any
20 of the following components require replacement:
■ A04 LCD Display
■ A02 Digital PCB
■ A03 Upper Enclosure
Note: The membrane switch is not a replaceable item. A damaged membrane
switch must be replaced as part of the A05 Lens Assembly repair kit.
To remove the A05 Lens Assembly from the defibrillator:
1. Remove the battery pak.
2. Remove the keypad from the Lens Assembly and discard (the keypad is
not reusable).
3. Remove the six 4-40 x .25 screws (20) from the Lens Assembly.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Lens Assembly 4. Remove the Lens Assembly from the upper enclosure using the lens
Removal (continued) removal tool (MIN 3206005) as follows.
– Place the tool’s strap around the middle of the case and center the
debris-free plunger over the lens.
– Pull the plunger straight up evenly.
– Remove the tool and carefully rotate the lens assembly to gain access to
the membrane switch flex connector.
5. Disconnect the membrane switch connector from the A02 Digital PCB at J2.
6. If required, continue with the A04 LCD Display replacement procedure.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Lens Assembly To install a new A05 Lens Assembly into the defibrillator (the Lens Assembly is
Installation not reuseable):
Note: Verify that the A04 LCD Display is installed before installing the Lens
Assembly.
1. Clean any dust and debris from the lens prior to installation.
2. While tilting the left side of the lens up, use the membrane switch connector
installation tool (MIN 3205532) to connect the membrane switch connector
to the Digital PCB at J2.
Note: Tool MIN 3205532 provides verification that the membrane switch
connector is fully seated.
3. Make sure the backlight wires and ferrite from the LCD display are in their
holding features in the upper enclosure.
4. Press the Lens Assembly into the upper enclosure, using equal pressure,
until fully seated. Do not press on the dome switches.
5. Install the six new 4-40 x .25 screws (20). Tighten to 7 in. lbs.
6. Install a new keypad into the Lens Assembly.
membrane switch connector
installation tool
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
The lower enclosure must be removed if any of the following components require
replacement:
■ A01 Analog PCB
■ A02 Digital PCB
■ A03 Upper Enclosure
■ W01 QUIK-COMBO to Analog PCB Harness (part of A03 replacment)
■ W02 Speaker Assembly
■ W03 Battery to Analog PCB Harness
■ W05 Readiness Indicator
■ W07 High-Energy Capacitor
■ W08 Inductive Resistor
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Keypad Removal Remove the keypad to gain access to the A05 Lens Assembly.
Note: The keypad is not reuseable after removal.
To remove the keypad from the defibrillator:
1. Peel up a corner of the keypad, (each corner contains a plug inserted into
the screw hole). Carefully peel the keypad away from the lens assembly.
Note: Using sharp tools to remove the keypad is not advised. The
membrane switch can easily be damaged with excess handling.
2. As required, continue to the A05 Lens Assembly replacement procedure.
Keypad Installation To install a new keypad into the defibrillator:
1. Make sure the Lens Assembly is installed in the device and is free of dust
and debris.
Note: Verify six screws are installed into the Lens Assembly. Verify that the
5
double microphone cover labels (5) are present and not damaged.
2. Remove half of the release liner from the keypad (9).
3. Place the keypad onto one side of the lens assembly and roll it across.
Remove the second half of the release liner and press the keypad onto the
lens assembly, using equal pressure, until fully seated.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
LCD Display Removal To remove the A04 LCD Display from the defibrillator:
1. Remove the A05 Lens Assembly.
2. Using needle-nosed pliers, disconnect the backlight cable from J5 of the
A02 Digital PCB. Insert the pliers into the two center holes of the connector
and gently pull upward.
3. Remove the four 4-40 x .25 (20) screws from the LCD display. Discard the
screws.
4. Using the display flex installation/removal tool (MIN 3205524), open
connector J6 of the Digital PCB, and gently guide the display flex from the
J6 connector out through the ferrite.
Note: Leave the ferrite adhered in place if the upper enclosure will be
reused.
5. If the LCD display will be reinstalled as part of the repair process, protect the
display from scratching or other damage.
6. If required, continue with the lower enclosure removal process.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
LCD Display Installation To Install the A04 LCD Display into the defibrillator:
1. Make sure the LCD Display is not damaged or scratched.
Note: Verify the Display EMI Shield (1) is present. Verify the lower enclosure
is installed.
2. Remove the adhered ferrite from the upper enclosure. Slide the LCD Display
flex through the ferrite.
Note: the LCD Display flex lays on top of the Display EMI Shield.
3. Using the display flex install/removal tool (MIN 3205524), guide the display
flex into the J6 connector on the Digital PCB. While holding the cable in
place, press the tool toward the front of the enclosure, rotate the handle
upward, and remove the tool from the slot.
4. Use the display flex verification tool (MIN 3205525) to verify that the display
flex is fully inserted. Silver traces of cable should not be visible when the flex
cable is fully inserted.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
LCD Display Installation 5. If the repair includes a new upper enclosure, remove the liner from the
(continued) adhesive at the ferrite location. Mount the ferrite (18) into the upper
enclosure.
6. Fasten the LCD Display in place using four 4-40 x .25 screws (20). Tighten to
7 in. lbs.
Note: Make sure the backlight wire are not pinched under the fastened
display.
7. Connect the backlight cable from the LCD Display to J5 of the Digital PCB.
8. Place the round ferrite on the backlight cable into the holder area in the
upper enclosure.
9. Continue with the A05 Lens Assembly installation process.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Battery Harness To remove the W03 Battery to Analog PCB Harness from the defibrillator:
Removal 1. Remove the lower enclosure.
2. Disconnect the Battery Harness connector from J504 of the Analog PCB.
3. Lift the Battery Harness PCB, wedged ferrite, and wire harness from the
upper enclosure.
4. Continue with the next repair procedure, as required.
Battery Harness To install the W03 Battery to Analog PCB Harness into the defibrillator:
Installation 1. Inspect the Battery to Analog PCB Harness for damage prior to installation.
2. Install the Battery Harness into the upper enclosure in the following
sequence:
– Install the Battery Harness PCB into the slot in the battery well in the
upper enclosure.
– Connect the Battery Harness connector to J504 of the Analog PCB.
– Route the Battery Harness wires along the groove in the upper
enclosure. Wedge the ferrite in the groove to secure the harness in
place.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Battery Harness Note: Verify all wires are tucked into the groove of the enclosure. Correct
Installation (continued) routing of wires prevents pinching when the lower enclosure is installed.
3. Install the lower enclosure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Speaker Assembly To remove the W02 Speaker Assembly from the defibrillator:
Removal 1. Remove the lower enclosure.
2. Disconnect the W02 Speaker Assembly connector from J507 of the Analog
PCB.
3. Lift and remove the Speaker Assembly from the device.
4. Continue with the next repair procedure, as required.
Speaker Assembly To install the W02 Speaker Assembly into the defibrillator:
Installation Note: Verify that the upper speaker backing (13) is in place before installing
the Speaker Assembly.
1. Inspect the Speaker Assembly for damage prior to installation.
2. With the tab oriented to the bottom of the Speaker Assembly, place the
speaker into the upper enclosure.
3. Connect the Speaker Assembly connector into J507 of the Analog PCB.
4. Route the Speaker Assembly wires along the groove and slot in the upper
enclosure.
5. Install the lower enclosure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Inductive Resistor To remove the W08 Inductive Resistor from the defibrillator:
Removal 1. Remove the lower enclosure.
2. Disconnect the W08 Inductive Resistor connector from J501 of the
W01 A01 Analog PCB.
3. Lift and remove the Inductive Resistor from the device.
4. Continue with the next repair procedure, as required.
Inductive Resistor To install the W08 Inductive Resistor into the defibrillator:
Installation 1. Inspect the Inductive Resistor for damage prior to installation.
2. Place the Inductive Resistor into the case features in the upper enclosure
(short wire at the top with the soldered connection toward the back).
3. Route the Inductive Resistor wires over the top of the W01 QUIK-COMBO
harness and W07 Capacitor wires.
4. Connect the Inductive Resistor connector into J501 of the Analog PCB.
5. Install the lower enclosure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Capacitor Removal To remove the W07 High-Energy Capacitor from the defibrillator:
1. Remove the lower enclosure.
2. Disconnect the W07 Capacitor connector from J502 of the Analog PCB.
3. Lift and remove the W07 Capacitor from the device.
4. Continue with the next procedure, as required.
Capacitor Installation To install the W07 High-Energy Capacitor into the defibrillator:
1. Inspect the W07 Capacitor for damage prior to installation.
Note: Verify the W05 Readiness Display is installed and connected to the
A02 Digital PCB.
2. Place the capacitor into the upper enclosure. Orient the capacitor terminals
vertically with the Warning label up.
3. Connect the capacitor connector into J502 of the Analog PCB. (Route and
connect the capacitor to J502 prior to routing and connecting the Inductive
Resistor to J501.)
4. Install the lower enclosure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Analog PCB Removal To remove the A01 Analog PCB from the defibrillator:
1. Remove the Lower Enclosure.
2. Disconnect the W03 Battery Harness from the J504 connector.
3. Disconnect the W02 Speaker Assembly from the the J507 connector.
4. Disconnect the W08 Inductive Resistor from the the J501 connector.
5. Disconnect the W07 High-Energy Capacitor from the J502 connector.
6. Unhook the cable retainer on the W01 QUIK-COMBO to Therapy harness.
7. Disconnect the W01 QUIK-COMBO to Therapy harness from the J503 and
J506 connectors.
8. Carefully lift the A01 Analog PCB away from the A02 Digital PCB, with most
of the upward pressure concentrated around the P16 connector.
Note: Keep the W06 48-pin Header inserted in the A01 Analog PCB.
Protect the exposed pins.
9. Store the A01 Analog PCB in a protective container.
10. Continue with the next procedure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Analog PCB Installation To install the A01 Analog PCB into the defibrillator:
1. Inspect the A01 Analog PCB and the W06 48-pin Header for damage prior to
installation.
2. Verfiy the A02 Digital PCB is installed.
3. Verify the high-voltage shield image plane (4) is installed.
4
Note: The A01 Analog PCB replacement kit includes a new W06 48-pin
Header. Install the W06 into the A01 Analog PCB at P16 prior to installing
the A01 Analog PCB into the defibrillator.
Caution!
Contamination. Handle the PCB by the board edges to avoid
contamination, which can affect performance of high impedance circuits.
4. Carefully place the A01 Analog PCB onto the upper enclosure bosses. Hold
the front edge of D up to view the W06 48-pin Header engage J16 of the
Digital PCB.
5. Connect the W01 QUIK-COMBO to Therapy Harness to the J503 and J506
connectors.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Analog PCB Installation 6. Secure the cable retainer around the W01 QUIK-COMBO to Therapy
(continued) Harness.
7. Connect the W07 High-Energy Capacitor to the J502 connector.
8. Route the W08 Inductive Resistor wires over the top of the W01 and W07
10
harnesses. Connect the W08 Inductive Resistor to the J501 connector.
9. Connect the W02 Speaker Assembly to the J507 connector. Check the wire
routing.
10. Connect the W03 Battery Harness to the J504 connector.
Note: Verify all wires are tucked into the groove of the enclosure. Correct
routing of wires prevents pinching when the lower enclosure is installed.
11. Verify the high-voltage caution shield (10) is in place on the Analog PCB.
12. Install the lower enclosure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Digital PCB Removal To remove the A02 Digital PCB from the defibrillator:
4
1. Remove the A05 Lens Assembly.
2. Remove the A04 Display Assembly.
3. Remove the Lower Enclosure.
4. Remove the A01 Analog PCB.
5. Remove the W07 High Energy Capacitor.
6. Remove the high-voltage shield image plane (4) from the bosses in the
upper enclosure. Keep the metal tabs of the shield folded in as in the
installed state.
A02
7. Disconnect the W05 Readiness Display cable from the J3 connector of the
A02 Digital PCB.
Note: Verify that the keypad and display assemblies are disconnected from
the A02 Digital PCB before continuing with the next step.
8. Carefull lift the A02 Digital PCB away from the upper enclosure.
9. Store the A02 Digital PCB in a protective container.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Digital PCB Installation To install the A02 Digital PCB into the defibrillator:
A02 1. Inspect the A02 Digital PCB for damage prior to installation.
2. Place the A02 Digital PCB onto the upper enclosure bosses.
3. Connect the W05 Readiness Display flex to the J3 connector of the
A02 Digital PCB.
4. Install the high-voltage shield image plane (4) onto the bosses in the upper
enclosure. The metal tabs on the shield go over each boss and must make
contact with the A01 Analog PCB. Verify the center metal tab is inserted onto
4 the pin of the enclosure.
5. Install the W07 High-Energy Capacitor into the upper enclosure.
6. Install the A01 Analog PCB.
7. Install the lower enclosure.
8. Install the A04 Display Assembly.
9. Install the A05 Lens Assembly.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Readiness Display To remove the W05 Readiness Display from the defibrillator:
Removal 1. Remove the lower enclosure.
2. Remove the W07 High Energy Capacitor.
3. Lift the rear corner (near speaker) of the A01 Analog PCB and the high-
voltage shield image plane (4) high enough to gain access to the
J3 connector of the A02 Digital PCB.
4. Disconnect the Readiness Display cable from the J3 connector of the A02
Digital PCB.
5. Push three fingers torward the top of the enclosure to access the Readiness
Display. Use the plastic tool to push the Readiness Display out of the rubber
holding features in the upper enclosure.
Note: Use care in removing the Readiness Display from the upper
enclosure. The glass of the Readiness Display can scratch the clear lens of
the upper enclosure.
6. Remove the Readiness Display from the upper enclosure.
7. If the Readiness Display will be reinstalled as part of the repair process,
protect the display from scratching or other damage.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Readiness Display To install the W05 Readiness Display into the defibrillator:
Installation 1. Locate the Readiness Display, unpackage and prepare for installation.
Check the display for damage with no scratches on the display viewing area.
2. Clean the lens and Readiness Display in preparation for installing the
display.
3. Holding the three fingers back in the upper enclosure, insert the bottom edge
of the Readiness Display first and then rotate the top edge into place. Center
the Readiness Display in the bottom two holding features. Release the
fingers to hold the display in place.
4. Connect the Readiness Display cable into the J3 connector of the A02
Digital PCB.
5. Install the W07 High-Energy Capacitor.
6. Install the lower enclosure, as required.
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
J16
EMI shield HV image plane shield W06 Header
J504
W03
A01 Analog PCB J501
Battery Pak Battery W08 Inductive
Harness MIN 3203555 Resistor
MIN 3203730 MIN 3010212
J507 T1 preamp
shield shields J502
W02 Speaker
MIN 3203730 W07 Capacitor
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
17
13
W07 10
A01
W08 W03
W06
4
A03
W02
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Parts Lists
Item Qty MIN CAT. Part Description Notes
A01 1 3203555-002 21330-001079 Analog PCB
A02 1 3203558-006 21330-001078 Digital PCB
A03 1 3203716-002 21300-006594 Upper Enclosure
A04 1 3203723-001 21300-005746 LCD Display, monochrome, 1/4 VGA, LED backlight
A05 1 3205306-000 21330-001100 Lens Assembly
W01 1 3203730-007 21300-006595 Wire Harness - QUIK-COMBO to Therapy PCB
W02 1 3203730-008 21300-006617 Wire Harness - Speaker Assembly
W03 1 3203730-003 21300-005936 Wire Harness - Battery to Analog PCB
W05 1 3203914-002 21300-006576 LCD Readiness Indicator
W06 1 3203984-000 21300-006044 Header Connector, stacking, 2-row, 48-pin
W07 1 3200442-004 21300-006043 High-Energy Capacitor
W08 1 3010212-008 21300-006042 Inductive Resistor, 5 ohm
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Replacement Procedures
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Section 12 Contents
Parts Lists and This section provides assembly diagrams and lists each part to help identify
components needed to repair the LIFEPAK 1000 defibrillator.
Assembly
Interconnect Diagram
Diagrams
Service Replacement Kits
Defibrillator Part Number and Serial Number
Ordering Parts
42
Parts Lists and The Interconnect Diagram provides detailed interconnect information for each
assembly and cable, with reference designations (for example, System PCB,
Assembly W05 Power PCB/Contact PCB Cable). Click a link to view a detailed diagram.
Diagrams The Service Replacement Kits contain all items needed to replace major
(continued) components. Each kit has its own MIN.
The Serial Number and Part Number description is useful for decoding
information on the device label.
Refer to Ordering Parts to obtain replacement parts for the device.
For additional parts lists, including items necessary to keep the device in clinical
service, see Ordering Devices, Supplies, and Accessories.
43
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
Glossary
The following terms appear in this section (including column headings of the
parts list tables):
■ Common Parts are components used in every version of the LIFEPAK 12
defibrillator/monitor, regardless of options. Common parts are divided into
front case, rear case, and System/Memory/Therapy PCBs.
■ Internal Parts are components internal to the case that are specific to your
device.
■ External Parts are components external to the case that are specific to your
device.
■ An Item number is the reference designator for each unique part on the
device. Click an item number on a diagram to view the MIN and description
for that part. Click an item number on a parts list to view a diagram showing
that part. Item numbers that begin with a letter (for example, A01, W21)
designate a main component (such as a PCB or cable) that includes
individual parts.
■ The Quantity identifies how many of the listed part are included in the order.
■ MIN refers to the Medtronic manufacturer’s item number (part number).
■ Part Description provides a brief description of the part.
■ Part Number is the PN number on the device label that identifies the model
of each device.
44
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
Glossary (continued)
■ The catalog number (CAT.) is the reference designator used for ordering
each part.
■ Options are assemblies that are not required on the basic device and can
be specified by the customer when purchased. Parts on these assemblies
may be referred to as optional parts.
45
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
46
(Continued on next page)
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
47
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
48
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
13
W07 10
A01
W08 W03
W06
4
W02
49
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
50
Previous Page Table of Contents Section Contents Back Index Next Page
LIFEPAK 1000 Defibrillator Parts Lists and Assembly Diagrams
51
Previous Page Table of Contents Section Contents Back Index Next Page
A
A Physical Inspection
Testing
Record the result of Service Alert Test and Audio- Voice Prompt Test.
4 Defibrillator Energy Output / Charge time Tests/Buttons and Keys Test/Dump test
Record the result of Time and Date, LCD Contrast and Full Monthly Test.
8 3 ECG Leads Characteristic Test (Note: Complete if the Device is equipped with 3 Lead N/A
ECG option)
a - Record the result of ECG Lead Characteristic Test. 10Hz Sinewave, 1mv = 10mm
b - Record the result of ECG BPM Test. ECG 120 BPM = 120 + 3 BPM
Record the result of Therapy ECG Characteristic Test. QRS Waveform @120 BPM = 11mm