EQUIPMENT LEGEND ADDITIONAL ACCESSORIES

ELECTRICAL LEGEND: ELECTRICAL USE REQUIREMENTS


ITEM DESCRIPTION BTU/HR WT/LBS WIDTH DEPTH HEIGHT
3. DEHUMIDIFIER
ITEM DESCRIPTION DESCRIPTION REQUIREMENTS GANTRY ACQUISITION WORKSTATION 360 39.8" 23.8" 74.6"
>OASIS DEHUMIDIFIER (D-165HG GST)
HOLOGIC SELENIA -Dimension:
1 MAX 101.2CM 60.5 CM 189.4 CM
ACQUISITION WORKSTATION SHIELD Weight: 51.8 lbs
TYPICAL
Height: 23.8 in
RADIATION LAMP(RELAY CONTACTS RATED: 10A, 250 VAC) GMS/CONTRACTOR INPUT LINE VOLTAGE 200, 208, 220, 230, 240VAC 100, 120 VAC/220, 230, 240 VAC RANGE
Width: 15 in
1700-2500 670 26" 43.7" 85" Depth: 14.4 in
2 HOLOGIC SELENIA GANTRY WITH C-ARM 66 CM 111 CM 214 CM
220 V, 20 A, 3P-2G OUTLET (FOR SERVICE EQUIPMENT) GMS/CONTRACTOR INPUT CURRENT 80 AMP BREAKER (15KVA UPS) 3.50A/2.0A

MAMMO DISTRIBUTION PANEL, 125A MAIN, BRANCHES 90 A, 20 A SINGLE PHASE GMS/CONTRACTOR 50-60 HZ, SINGLE PHASE
NOTE:
DEDICATED OUTSIDE PHONE LINE (OPTIONAL) CUSTOMER SUPPLY NUMBER OF PHASES GANTRY IS TO BE PERMANENTLY WIRED, HOSPITAL GRADE RECEPTACLE
SITE CONDITION FOR SELENIA DIMENSION SYSTEM ADDITIONAL ACCESSORIES
USE OF PLUG AND RECEPTACLE ARE IS REQUIRED MAXIMUM FLOOR LOADING 1050 LBS
NOT ALLOWED MINIMUM RECOMMENDED ROOM SIZE 9'-0" X 12'-0" (257 CM X 365 CM) 1.SecurView Dx Workstation
C5
CAT5 NETWORK CONNECTION RECOMMENDED VPN, STATIC IP READY CUSTOMER SUPPLY MINIMUM CEILING HEIGHT 7'-5" A.F.F (228CM)
(HOSPITAL NETWORK CONNECTION)
>Barco Monitor (Nio 5MP LED)
PHYSICAL AND ENVIRONMENTAL REQUIREMENTS AMBIENT ROOM CONDITION
Specification s:
D 68-86' F (20-30'C) -Active screen size (H x V) NOTE:
DEDICATED 240 V, 20 A DUPLEX OUTLET, 3 PRONG (CONSOLE) GMS/CONTRACTOR 68-86 F ( 20-30 C)/77 F-ROOM TEMPERATURE 77'F ROOM TEMP FOR OPTIMAL 422.4mm x 337.9mm ( 16.5 x 13.3")
OPERATING TEMPERATURE RANGE PERFORMANCE -Aspect ratio (H:V) 5:4
A
OPERATING TEMPERATURE (MAINTAIN TEMPERATURE RANGE MAXIMUM ALLOWABLE TEMPERATURE -Resolution 5MP (2560 x 2048) CUSTOMER IS RESPONSIBLE FOR ENSURING THAT THE FLOOR IS CAPABLE OF
DEDICATED 240 V, 20 A DUPLEX OUTLET, 3 PRONG (AIR-CONDITIONING) CUSTOMER SUPPLY RANGE (MAINTAIN TEMPERATURE MAXIMUM ALLOWABLE TEMPERATURE 24HRS/DAY, 7DAYS/WEEK) CHANGE SUPPORTING THE EQUIPMENT FLOOR LOADING AND ANCHORAGE REQUIREMENTS.
RATE OF 10' C PER HOUR -Dimension with stand (W x H x D) ADD'L BLOCKING OR BRACING MAY BE REQUIRED WHEN WOOD FLOORS ARE
RANGE 24 HRS/DAY, 7DAYS/WK) Portrait: (w x hmax x d) 407 x 623 x 235 mm PRESENT. THE SUPPLIER IS NOT LIABLE FOR ANY DAMAGE OR INJURY CAUSED BY INSTALLATION
(w x hmin x d) 407 x 523 x 235 mm
1 220-240 VAC INCOMING POWER CUSTOMER SUPPLY CHANGE OF RATE OF 10 C PER HR Landscape: (w x hmax x d) 493.5 x 579 x 235 mm
IN ANY LOCATION WITH INADEQUATE FLOOR LOADING CONDITIONS.
OPERATING HUMIDITY RANGE 20-80% NON CONDENSING
(w x hmin x d) 493.5 x 479 x 235 mm

2 POWER CABLE FROM GANTRY GMS/CONTRACTOR OPERATING HUMIDITY RANGE 20-80% NON CONDENSING -13'- 140'F-TUBESTAND, 0-95% HUMIDITY-
NON CONDENSING
-Net weight with stand: 13.25kg
STORAGE TEMPERATURE (SYSTEM -Net weight w/o stand: 8.23kg
POWERED DOWN OR IN STORAGE) 50-104'F- IMAGE RECEPTOR, 10%-80%
-Net weight packaged with stand: 19.61 kg NETWORKING NOTES
HUMIDITY-NON CONDENSING
3 WIRINGS FOR FROM RADIATION LAMP WITH 1.5METERS EXCESS FROM THE WALL GMS/CONTRACTOR 50-104 F (10-40C)-GANTRY, EXCLUDING DETECTOR ASSEMBLY.
>Eaton 1kVA UPS
STORAGE TEMPERATURE -Dimension: 230mm x 160mm x 380mm (HXWXD) SUPPLYING AND INSTALLING NETWORK CONNECTION BETWEEN DIAGNOSTIC IMAGING

4 (SYSTEM POWERED DOWN, OR EQUIPMENT ENTRY REQUIREMENTS Weight: 14.5kg EQUIPMENT AND OTHER NODES CONNECTION SHALL BE THE CUSTOMERS RESPONSIBILITY,
WIRINGS FROM MAMMO DISTRIBUTION PANEL TO SECURVIEW PC GMS/CONTRACTOR 59-86 F (15-30 C)-DETECTOR ASSEMBLY CUSTOMERS MUST SPECIFY, SUPPLY AND INSTALL PROPER RECEPTACLES, AUI TRANSCEIVERS,
IN STORAGE) GANTRY SHIPPING CRATE 38" X 66" X 88" (97CM X 168CM X 224CM) >CPU (DELL PRECISION TOWER 5810) CABLING, REPEATERS, ROUTERS, MODEMS, ETC. WHICH ARE CONSISTENT WITH THE SITE NEEDS
-Height: 16.in AND SOUND NETWORK IMPLEMENTATION PRACTICES. ALSO INCLUDED, BUT NOT LIMITED TO,
5 LAN CONNECTION BETWEEN CONSOLE AND SECURVIEW PC THROUGH CAT5 GMS/CONTRACTOR 14-104 F (-10-40 C )-ACQUISITION WORKSTATION ACQUISITION WORKSTATION 45" X 48" 47" (114CM X 122CM X 119CM) -width: 6.8 in THE CUSTOMER'S RESPONSIBILITY IS THE PURCHASE AND INSTALLATION OF CONDUITS
SHIPPING CRATE -Depth: 18.5 in AND/ OR RACEWAYS OR CONTAINMENT AND ROUTING OF NETWORKING CABLES, AS DETERMINED
BY SITE CONDITIONS, OR REQUIRED BY PREVAILING CODES.
CO1 CUSTOMER SUPPLY
MAXIMUM SEPARATION N/A N/A MINIMUM ENTRANCE 2'-5" X 6'-6" (76CM X 203CM)

FROM IMAGE
HORIZONTAL SURFACE WIREWAY 1 34 " X 4 34" WIREMOLD 4000 SERIES OR EQUAL,
NOTE:
WH1 DIVIDED, WITH REMOVABLE COVERS, BOTTOM OF WIREWAY 12" A.F.F. ACCESS TO GMS/CONTRACTOR
26"W X 35"D X 88" *C-ARM SPANS 56" W X 24" D X 80" H (142 CM X NOTE:
WV1 AND 'WV2'
OVERALL DIMENSIONS 68"W*(66 CM X 89 CM X 224 CM) 61 CM X 203 CM) A WARNING NOTICE SHALL BE PUT OUTSIDE THE X-RAY EXAMINATION ROOM DOOR. IT SHALL BE
ALL INDICATED WALLS MUST BE MADE OF ANY OF THE FF. MATERIALS TO A HEIGHT OF AT LEAST 2 METERS FROM THE FLOOR OR GROUND OUTSIDE THE RAY ROOM MADE UP OF THE SOLID YELLOW EQUILATERAL TRIANGLE 180MM LONG ON EACH SIDE. AT THE
CENTER OF THE TRIANGLE IS A BLACK TRE-FOIL SIGM FOR RADIATION. UNDER THE TRIANGLE
A.) AT LEAST SIX (6) INCHES POURED CONCRETE OR AT LEAST SIX 9S) INCHES CONCRETE HOLLOW BLOCK WITH ALL CAVITIES FILLED AND PACKED WITH CONCRETE ARE THE WORDS "X-RAY ROOM: DO NOT ENTER WHEN THE RED LIGHT IS ON". THE WARNING
AND PLASTERED WITH AT LEAST 12 INCH CEMENT PLASTER ON EACH SIDE OF THE BLOCKS, OR, NOTICE SHALL BE ON A 180 MM X 270 MM WHITE BACKGROUND.

3 3 1
HORIZONTAL SURFACE WIREWAY, 1 4"
X 4 WIREMOLD 4000 SERIES OR EQUAL
4"
B.) AT LEAST 6 INCH (1.5 MM) THICK LEAD SHEET. NOTE THAT CARE MUST BE TAKEN TO AVOID PUNCTURES IN THE LEAD SHEET BE GLUED
GMS/CONTRACTOR ONTO AND SANDWICH BETWEEN WOODEN PANELS.
FD1, FD2 DIVIDED, WITH REMOVABLR COVERS PROVIDE FULL ACCESS TO CORRESPONDING
1
DOOR LEADING TO THE X-RAY ROOM MUST BE LINED WITH AT LEAST 1.5 MM ( INCH) THICK LEAD SHEET TO A HEIGHT OF A AT LEAST 2 METERS FROM THE
VERTICAL DUCT, GROMMETED OPEN END FLOOR/ GROUND OUTSIDE.
6

2 WIRE + GND
HORIZONTAL SURFACE WIREWAY, 1 34 " X 4 34" WIREMOLD 4000 SERIES OR EQUAL
DIVIDED, WITH REMOVABLR COVERS PROVIDE FULL ACCESS TO CORRESPONDING GMS/CONTRACTOR
WV1, WV2 UPS DETAIL
FLOOR DUCT (WV1 TO FD1/ WV2 TO FD2)

2 WIRE + GND 2 WIRE + GND 2 WIRE + GND EATON 9E 15KVA UPS


WALL THICKNESS: 6 INCHES SHEER CONCRETE OR 1.5mm LEAD SHIELDING DEPENDING ON LOCAL REGULATIONS
RECTIFIER
NOTE: ROOM LAYOUT MAY VARY DEPENDING ON THE ACTUAL ROOM DIMENSIONS INCOMING POWER 220V
TO LOAD

125AT
125AT
220V
20AT 100AT

CONVINIENCE OUTLET POWER SUPPLY TO UPS


2-POLE 2-POLE

ECB 20KVA XF 20AT 100AT


L-N/220-0
POWER SUPPLY TO GANTRY
CONVINIENCE OUTLET 2-POLE 2-POLE
(READING ROOM)
20AT

CONVINIENCE OUTLET
2-POLE
(SERVICE EQUIPMENT)

- PROPOSED EATON 9E 15KVA SINGLE LINE DIAGRAM AS-BUILT PLAN


- N T S

PREPARED BY : PROFESSIONAL ELECTRICAL ENGINEER REPUBLIC ACT 9266 ARCHITECTURE ACT OF 2004 ARTCLE IV SECTION 33
DRAWING AND SPECIFICATIONS DULY SIGNED, STAMPED OR SEALEDAS INSTRUMENTS OF SERVICE
PROJECT TITLE: OWNER: REVISIONS: SHEET CONTENT: SHEET NO.
ARE THE INTELLECTUAL PROPERTIES & DOCUMENTS OF THE ARCHITECT, WHETHER THE OBJECT FOR
WHICH THEY ARE MADE IS EXECUTED OR NOT. IT SHALL BE UNLAWFUL FOR ANY PERSON, WITHOUT
THE CONSENT OF THE ARCHITECT OR AUTHOR OF SAID DOCUMENTS TO DUPLICATE OR TO MAKE
DATE: OCT. 2017 GENERAL NOTES
EDGARDO H. LIGAS COPIES OF SAID DOCUMENTS FOR USE IN THE REPETITION OF AND FOR OTHER PROJECTS OR
BUILDINGS, WHETHER EXECUTED PARTLY OR INWHOLE. PROPOSED ROOM LAYOUT
DRAWN BY: MRF ELECTRICAL DETAILS
AND ACCESSORIES HOLOGIC AVIA
GLOBAL MEDICAL SOLUTIONS, INC. PRESIDENTIAL DECREE NO. 49 " PROTECTION OF INTELLECTUAL PROPERTY " NOV. 14 1974
CHINESE GENERAL HOSPITAL
E-1
WORKS OF ARTS, ARCHITECTURE, MODELS AND DESIGNARE CREATIVE WORK AND CONSIDERED AS PART OF
PRC NO.:1936 VALIDITY: THE INTELLECTUAL PROPERTY OF THE AUTHOR DULY PROTECTED UNDER PD 49
DATE: SINGLE DIAGRAM
3/F RTS BLDG. ECONOMIA ST. BAGUMBAYAN, QUEZON CITY AIAPOA NO.: OR NO.: DATE: LOCATION: AND MEDICAL CENTER REVIEWED BY:
REPUBLIC ACT 8293 INTELLECTUAL PROPERTY LAW SECTION 186
WORK OF ARCHITECTURE - COPYRIGHT IN A WORK OF ARCHITECTURE SHALL INCLUDE THE RIGHT TO
PTR NO. ISSUED ON: CONTROL THE ERECTION OF ANY BUILDING WHICH REPRODUCES THE WHOLE OR A SUBTANTIAL PART OF
286 Blumentritt Rd, Sampaloc, Manila, DATE:
THE WORK EITHER IN ITS ORIGINAL FORM OR IN ANY FORM RECOGNIZABLY DERIVED FROM THE ORIGINAL.

ISSUED AT: TIN NO.: 3802750 Metro Manila PLOTTING DATE: