Notice

The notice for installation


 Do not hang anything on the power wire and be treaded by people.
 The machine needs good ground cable. The electric resistance is not more than 4. Do not
use the heat pipe or water pipe as the temporary ground cable.
 All the holes at the back of the machine should not be covered. Otherwise, the machine
will be damaged because it is too heat.
The notice for application
 The power voltage and frequency should be 220V and 50Hz.
 Do not cover anything on the holes at the back of the machine.
 Do not put the machine in a humid environment. When cleaning the machine, please take
the plug away from the socket. Do not use the liquid cleaner or the alcohol to wash it. Please use
the wet cloth to wipe it. When the machine meets thunderstorm or it stays for a long time, make
sure that the plug is taken away from the socket.
 The cables should be used in a correct way.
 In the cold winter, make the machine heat in advance and then use: in the hot summer, Do
not turn on the machine for a long time. For example, when the temperature in the room is over
35℃, the machine should be turned off after used for over 2 hours, and then turn it on again.
The notice for protection
 Do not open the machine without authorization. The machine should be opened by the
experienced technicians.

I
 Please take the plug away from the socket when meeting the following conditions:
 The power wire and the plug are broken.
 Something drops into the machine.
 The machine can not work.

II
Preface
This operation manual introduces you the total function of the digital brain electric activity
mapping and the detail operation procedure.
Please read the manual carefully before using the machine, which makes you easier to use it.

III
IV
Table of Contents
Chapter 1 Introduction..................................................................................................................... 2
Chapter 2 The Main Technical Index...............................................................................................3
Chapter 3 Technical Device............................................................................................................. 5
Chapter 4 Software and Hardware Setup Wizard............................................................................ 6
4.1. Software Wizard....................................................................................................................... 6
4.2. Hardware Wizard...................................................................................................................... 9
Chapter 5 The Operation Regulation............................................................................................. 11
5.1 The Place of Electrode............................................................................................................. 11
5.2 The System Operation..............................................................................................................11
5.3 The System Operation..............................................................................................................13
5.4 The Acquisition Control Panel.................................................................................................14
5.5 Display View and Operation....................................................................................................17
Function 1: Divide Time................................................................................................................ 18
Function 2: Adjust by Right Menu................................................................................................ 19
Function 3: Montages Per Screen.................................................................................................. 19
Function 4: One channel/Screen.................................................................................................... 19
Function 5: Mark/Event List..........................................................................................................20
Function 6: Select Segments..........................................................................................................21
Function 7: Save Segment............................................................................................................. 21
Function 8: Cancel Segment.......................................................................................................... 22

V
Function 9: Data Analysis..............................................................................................................22
Function 10: Analysis Views......................................................................................................... 22
Function 11: Sleep Analysis...........................................................................................................29
Function 12: Simple Surveyor's Staff............................................................................................ 30
Function 13: Scale......................................................................................................................... 31
Function 14: Automatic Measure...................................................................................................31
Function 15: Band Division...........................................................................................................32
Function 16: Montage Ways.......................................................................................................... 32
Function 17: Design Filter............................................................................................................. 33
Function 18: System Setting.......................................................................................................... 34
Function 19: Auto Calibration....................................................................................................... 40
Function 20: Report Design and Print........................................................................................... 40
Function 21: Input Office Word.....................................................................................................42
Function 22: Select Patient And Extend Case............................................................................... 42
Function 23: Patient List Management..........................................................................................43
Function 24: Output\Input Cases................................................................................................... 44
Function 25: Pathologic Analyse................................................................................................... 45
Function 26: SpO2 trend................................................................................................................ 46
Function 27: Multilingual Interface...............................................................................................46
5.6 Exits......................................................................................................................................... 46
Chapter 6 Maintenance and Troubleshooting................................................................................ 48

VI
Chapter 7 Appendix....................................................................................................................... 50

VII
1
Chapter 1 Introduction
The features of digital brain electric activity mapping:
The digital brain electric activity mapping transit head bioelectrical activity into marked
intuitionistic image, it becomes an advanced, no wound technical instrument for skull sickness,
quantitative analysis for head enginery. Characters for this equipment are as follows:
 Help to bulge and recognize EEG partial features.
 Recognize the subtle features which was easy to be ignored ago.
 Help to academic communication, especially for the amateurs.
 Evaluate the unusual conditions in the traditional EEG and the unexplained situations by
eye measure.

 Refer to instruction manual/booklet.

2
Chapter 2 The Main Technical Index
1. Channels: 19 leads EEG (Option: 1 lead ECG, 1 lead BR, 1 lead EMG, 2 leads EOG).
2. The normal working condition
a) The range of environment temperature: 5℃~40℃
b) Relative humidity range: ≤85%
c) Voltage: USB DC 5V
d) Atmospheric pressure range: 700hPa~1060hPa
3. Calibration voltage: sinusoidal signal 5mm / 50μV; error: ±5%
4. Scan speed: 30mm / s; error: ±5%
5. Noise level: ≤5μV
6. Common-mode rejection ratio: ≥90dB
7. Input impedance: >10MΩ
8. 50 Hz interference suppression: ≥30dB
9. Frequency characteristics: Instrument Input 0.8Hz~30Hz, -3dB, 50μV sinusoidal AC signal
should not be distorted.
10. Resistance to polarization voltage: Add ± 300mV DC polarization voltage deviation is less
than ±5%
11. Transport and storage
a) The range of environment temperature: -20℃~+55℃
b) Relative humidity range: ≤90%
c) Atmospheric pressure range: 700hPa~1060hPa

3
12. Product safety type: II Class BF-based application part.

4
Chapter 3 Technical Device
CPU: Pentium, more than 4, 2.0G (Dual Core Processor if have the function of video)
Memory Card: More than 512M (more than 1G DDR2 if have the function of video)
Motherboard: Support USB2.0, recommend to use Intel chipset motherboard
Hard drive: More than 80G
Display card: More than 64M
Printer: 300dpi Color printers
Screen: ≥17 inch
CD-ROM: More than 24X CD-ROM (or choose CD-ROM system)
Operation System: Microsoft Windows XP / 7 (32bit)
Resolution ratio: More than 1024*768
Color: More than 24 colors
Font: Natural
Interface: USB 2.0
This operation manual is used for EEG 24 Channel series digital brain electric activity mapping.
Because of the different setup in every series of machines, this manual is accordance to the
largest configuration.

5
Chapter 4 Software and Hardware Setup Wizard
4.1. Software Wizard
First of all, put the CD-ROM contains software into drive, and then install the setup. The
following will pop-up the dialog box as below.

Figure 4-1 Figure 4-2

To continue, click "Next", if you would like to select a different folder, click "Browse". Click
"Next" 3 times.

6
Figure 4-3 Figure 4-4
There will be shown "Ready to Install", click to install.

Figure 4-5

7
There next will be shown "Hardware Installation", click "Continue Anyway".

Figure 4-6 Figure 4-7


And then click to "Finish", the installation is complete.
Uninstalling the software:
Method 1
Click "Start" menu in the "All Programs" to find the installation package to install software by
clicking the following icon to uninstall the software.
Method 2
Find software installation directory (usually installed by default under the C drive), open the
installation directory to find and double-click the icon below to unload

8
4.2. Hardware Wizard
After installing the software, link the EEG amplifier and the USB of computer main frame, then
the computer will be prompted to find a new hardware.

Figure 4-8 Figure 4-9


Choose the third one: "No, not this time", click "Next" to continue:
Click "Next" to continue:
Then will show a notice, click "Continue Anyway" to continue.

9
Figure 4-10 Figure 4-11

There will be shown "Completing the Wizard", click "Finish".


Notice: After click "Finish", the computer will be prompted to find a new hardware again.
And you can setup the hardware according to the steps above again. The hardware will
finish the setup when setup is complete.

10
Chapter 5 The Operation Regulation
5.1 The Place of Electrode
This system use silver chloride channel electrode, drip in brine liquor at 5% at work. Wash down
with clean water after use and airing everyday. Cover it to avoid exposal to protect the
chlorination on electrode.
Place the electrode according to international system electrode emplacement.
Notice: In the process of measurement, all of the measuring lead lines should be connected
with the corresponding leads on the scalp without hanging in the air. If parts of leads need
to be connected, make sure to connect all unused leads with GND to avoid the interference
caused by the lead hanging in the air. Simultaneously, close the leads on the software which
correspond with the unused leads (click the characters at the right of the corresponding
leads) to make the waveform beeline, in order to avoid the influence on other leads.
5.2 The System Operation
Follow is a list of function buttons, it may help you operate the software easily:

: New patient : Select patient

: Display EEG wave view : Play the EEG View


: When the EEG is playing, stop it. : Data analysis.

: View of Power Spectrum Graph : View of BEAM

11
: View of Compressed Spectrum Graph : View of Monochromatic BEAM

: View of Three-dimensional rotary BEAM : View of Side BEAM

: View of Numerical BEAM : View of Spectrum Rectangle Graph

: Sleep Analysis : Sleep Report

: Sleep Process Chart : Video play

: The BEAM's relative form or absolute form : Cancel one selected segment

: All montages are displayed : 1/2 montages are displayed

: 1/4montages are displayed : Version information.

: To adjust every montage's Amp : Reset the six frequency segments' range

: Select or design montage connect ways : Design the filter's parameter


: Save the configuration of system in order to apply it later

12
: Print : Preview

: EEG report : Divide one minute into five average parts

: Setup different categories color : Pathologic waveform analyse

: Add mark and information at somewhere : SpO2 trend

: List of marks and events could search someone mark and event position at EEG wave.
5.3 The System Operation
Turn on the computer, it enter the digital brain electric activity mapping operational system
automatically.

Figure 5-1

13
Select the menu of "Case Management" -> "New Patient" or click button , the physician
input parameters accord to the patients’ condition. The name item could input English or Chinese.
Press "Ctrl+Space" in keyboard would switch into Chinese input mode, and then input Chinese
name. The physician can choose the acquired method (full display or part display) by mouse's
arrowhead. Click "Gather" and enter into gather view.

Figure 5-2 Figure 5-2


5.4 The Acquisition Control Panel
In the acquisition interface, the left is the acquired wave to display the patient’s EEG at real time
and the right is the acquisition control panel to display various of premiums events and control
the acquired methods. Shown as Figure 5-3, Figure 5-4 or Figure 5-5:

14
Figure 5-4 (version of video) Figure 5-5 (version of SpO2)
Evoked Event:
Select evoked event by left button down, include the one added by user. EEG color will display
at defined color and show the event's name at evoked. The list of evoked events are disabled if
user hasn't added names.
Spark Control:
Join the flashlight with USB line, select "Sparking" evoked event and choose one of spark types:
"Auto", system control frequency from low to high automatically. "Manual", control frequency
by user; "Spark Scheme", select one of scheme from the list which designed in "System Setting".
Click "Spark" button to start sparking and click "Stop" button to stop sparking.

15
Speed, Plus:
Adjust display speed and amplitude of EEG wave at gather time.
Filter, Band:
The "Filter" would get rid of AC work frequency disturb which is recommended to use for batter
images; physician could choose the Band stop filter of EEG in practice.
Gather, End, Freeze:
"Gather", begin to record the EEG data into file; "Freeze", freeze image but continue gather and
save EEG data. "End", finish gather EEG data and return the play view automatically.
Montage Type:
Choose one of montage ways from 9 types according to practical use.
Gather Mode:
Physician could chose reference electrode at will when gather EEG according to practice.
Time:
Show system time before clicking "Gather" button; Show total time of gather after clicking
"Gather" button.
SpO2:
If the acquisition settings in the "System Settings", select "Show SpO2 Channel", connect the
oxygen probe, the control panel to the top will show the current oxygen value and pulse rate
values, click on "SpO2TrendMap" on the control panel will draw blood oxygen and pulse rate
trend Graph. In the meantime, collecting the interface will draw oxygen channel. If the oxygen
probe is off, the wave will use a different color.

16
5.5 Display View and Operation.
Click "End" to stop EEG acquisition and return display view. Refer to Figure 5-6 (Figure 5-7 is
video version, click button to review the video):

Figure 5-6 Figure 5-7


"Play": play EEG wave by order. "Stop": stop play wave. "Start": at begin of the EEG wave;
"End" at the end of wave. "Ascend": augment the EEG plus; "Descent": minish the EEG plus.
This interface is for user to replay EEG in sequence.
Select menu of "Review" -> "Start Play" or click button to play EEG wave from front to
end, and click button to stop.
Function 1: Divide Time

17
Select menu: "Review" -> "Divide Time" or click button could divide one second into five
parts to display, refer to Figure 5-8:

Figure 5-8 Figure 5-9


Function 2: Adjust by Right Menu
Right button down will pop up menu to adjust play types, amplitude and so on. And most of
information will be shown at left-bottom screen. Refer to Figure 5-9:
Function 3: Montages Per Screen
button is used to display 19-channel EEG in the same screen.
button is used to display 10-channel EEG in the same screen.

18
button is used to display 5-channel EEG in the same screen.
Function 4: One channel/Screen
Double click one pole name on Replay view, the full page will show the channel which has been
double clicked, Refer to Figure 5-10:

Figure 5-10 Figure 5-11


Function 5: Mark/Event List
1. Mark List
Physicians can add tag information at where it is needed to, specifically as follows: Select menu
of "Review" -> "Add Marks" or click the toolbar button , then click left mouse button down
at where you want to add a tag, pop-up shown as in Figure 5-11 window, the corresponding
location displays the flag icon .
After inputting information in "Remark" bank and clicking "Modify", the operation of adding

19
marks is finished.
Modify the existent marks: select menu "Review" -> "Mark/Event List" or click button , pop
up the same box as Figure 5-11, select the marks you want to modify, and change the contents of
"Remark", finally click "Modify" to finish modification.
Click to find the mark location, click and to select marks and trace back to where the
selected mark is.
When open a patient case, click button to display or hide marks.
2. Event List
Select menu "Review" -> "Mark/Event List" or click button , pop-up shown as in Figure
5-11 window. Choose "Event List" in left window. Window will be switched shown as in Figure
5-12. The window shows the information (name, start time, duration) of every evoked event.
Click and to trace back to where the selected mark is.

Figure 5-12 Figure 5-13

20
Function 6: Select Segments
The method is as follow: remove the arrowhead of the mouse to the beginning of the EEG which
is needed to analyze, and remove it to the end point. Press "Shift" on the keyboard, and Click the
left key of the mouse, so as to choose the EEG segment. The chosen EEG segment turns into
opposite color. Refer to Figure 5-13:
Function 7: Save Segment
If you want to save some segments of waveform, you can select menu: "Patient Management" ->
"Save Segment", look at the Figure 5-14, input the name of the segment, click "OK" to save it.

Figure 5-14
Function 8: Cancel Segment
If you want to cancel the latest segment, you should select menu "Review" -> "Cancel One
Segment" or click tool button .

21
Function 9: Data Analysis
After selecting one or more segments, select menu "Data Analysis" -> "Data Analysis" or click
button to dispose the chosen EEG segment to FFT exchange and insert value operation.
When the data disposal is over, the EEG is to be analyzed.
Function 10: Analysis Views
Power Spectrum Graph
Select menu "Data Analysis" -> "Power Spectrum Graph" or click button to enter the
following picture. This menu is to observe power chart distributing map. Click left button of
mouse on any power chart map could magnify this special lead power chart map.
Click right button of mouse on the enlarged power chart map could show frequency and value
where you have clicked.

Figure 5-15

22
BEAM
Select menu "Data Analysis" -> "BEAM" or click button to enter the following picture. The
left of this menu is to display the patient’s EEG, and the right is the look down relief map.
button is used to switch in the absolute relief map and comparative relief map.

Figure 5-16
Compressed Spectrum Graph
Select menu "Data Analysis" -> "Compressed Spectrum Graph" or click button to enter the
following picture

23
Figure 5-17 Figure 5-18
The physician can design compressed chart map according to menu tips. After design, click "Ok"
and enter into Figure 5-18 as follows:
This menu is used to display compressed chart map of any two lead of the patient.
X means frequency, the unit is Hz.
Y means power, the unit is μV2 /Hz.
Z means dimension of the compressed chart map, it can be decided by the physician.
Monochromatic View BEAM
Select menu "Data Analysis" -> "Monochromatic View BEAM" or click button to enter the
following picture. The left of this menu is the patient’s EEG. The right is the patient’s
Monochromatic View BEAM. Click button to exchange absolute EEG relief map and
comparative EEG relief map. Shown as Figure 5-19.

24
Figure 5-19 Figure 5-20
Three-dimensional Rotary BEAM
Select menu "Data Analysis" -> "Three-dimensional rotary BEAM" or click button to enter
the following picture. The right of the menu is the patient’s three-dimension stereoscopic EEG.
Press every EEG and turn it, which is easy for the physician to definite. Click to exchange
absolute EEG and comparative EEG.
Side View BEAM
Select menu "Data Analysis" -> "Side View BEAM" or click button and enter the following
picture. Click button to observe the patient’s side-view EEG from left or right. Click
button to exchange absolute EEG and comparative EEG.

25
Figure 5-21
Numerical BEAM
Select menu "Data Analysis" -> "Numerical BEAM" or click button and enter the following
picture. This menu is Numerical BEAM

26
Figure 5-22
Spectrum Rectangle Graph
Select menu "Data Analysis" -> "Spectrum Rectangle Graph" or click button to enter the
following picture. Click any lead, you could magnify it.

Figure 5-23
Function 11: Sleep Analysis
Click button to come into Figure 5-24. This picture is the dialog box for sleep analysis setting

27
(only the time of waveform is more than 5 minutes, the data can be used for sleep analysis).

Figure 5-24
After setting the time of sleep, click "OK", you can do the sleep analysis. After the analysis,
select menu "Data Analysis" -> "Sleep Process Graph" or click button to view the process
graph, as shown Figure 5-25. Select menu "Data Analysis" -> "Sleep Analysis Report" or click
button to view the report of sleep analysis, as shown Figure 5-26.

Figure 5-25 Figure 5-26


Function 12: Simple Surveyor's Staff

28
When you review case, please press the Ctrl key and left key of mouse, then pull the mouse. The
area where the mouse goes through could be displayed. Including the distance (mm) horizontal
and vertical, the horizontal coordinate means voltage (μV), the vertical means frequency (Hz). A
ruler distributed regularize unit, which can set up different distance range from 1mm to 100mm.
The effect can be shown at Figure 5-27.

Figure 5-27 Figure 5-28


Function 13: Scale
In replayed picture, here is a proportional rule in the left down area when you preview one (or
more) piece (s) of waveform map. Because the printer's frequency is not fixed, so the millimetre
shown at computer will be different from the paper. This ruler presents 5mm and 20μV in
realization. It provides reference to the doctors who like observe in paper. As Figure 5-28.
Function 14: Automatic Measure

29
Double click the left mouse at one of leads in the EEG waveform area, dialog box will appear as
been shown in the Figure 5-29. Select any one of the three crosses in dialog, drag the mouse and
the gain of waveform, period and frequency will be changed. You can also roll the wheel on the
mouse to zoom in and out.

Figure 5-29 Figure 5-30


Function 15: Band Division
Click button and enter the following picture. The physician could divide frequency segment
according to this menu. Click "Ok" button to save and return or "Cancel" button to cancel and
return after partition.
Function 16: Montage Ways
Select menu "System Configure" -> "Montage Ways" or click button to enter the following
picture. In this menu, the physician can choose the channel method (mono-channel or
double-channel). After the choice, Click "Ok" and display the accordant EEG.

30
Figure 5-31
Add or remove leads:
Click right of the mouse to customize any lead from 3 to 9 lead (the first two are default leads
which can not be modified), click the left of the mouse to drag any lead out of the channel list
box, and this operation finishes the deletion of one lead. If you need to add lead, just use the
mouse to click any two electrodes in the dialog box to complete the operation of add setting, then
click the button of "OK" after finishing 19 leads setting.
Change names of montage way:
Double click left mouse and pop up dialog box. Edit a name and click "OK" to save your
operation. Montage ways' name will appear at where montage ways changed in gather or replay

31
view.
Attention:
Ways 1 and 2 are default, they could not be modified. When create a new patient case, the
current montage ways are saved with case. If you open a case and want to modify the montage
way, it will prompt you whether to change the case's montage way.
Function 17: Design Filter
Click button and enter the following picture. In this menu, the physician can design filter
according to the patient’s EEG to clear EEG’s disturbance and artifact. Click "OK" button and
appear EEG which has being filtered.

Figure 5-32 Figure 5-33


Function 18: System Setting
In this part, we can set hospital information, review parameters, gather parameters, color
parameters, events, spark event and glossary.
Hospital Setting

32
After setting up the hospital information encountered need to fill such information, software
content in accordance with the set to fill in these items automatically, eliminating the need to
re-enter these items.
Select menu "System Configure" -> "System Setting" -> "Information Setting" and Figure 5-33
dialog box will appear.
These information could be printed according as user's need.
Review Setting
The playback interface in accordance with the parameters set in this tab, whether it is the overall
viewing or careful observation of the parameter settings according to their preferences.
Select menu "System Configure" -> "System Setting" -> "Review Setting" and Figure 5-34
dialog box will appear.

Figure 5-34 Figure 5-35


Review parameters show as Figure 5-35.

33
Show Speed: Control the waveform and display the delay.
Mtg./Scr.: Control the number of montages in full screen. "ALL EGG "means display
completely, "1/2 EEG" display one half of montages, "1/4 EGG" means display all screen's 1/4.
Play Speed: Can choose the mode and speed when played. The parameters are also available on
the replay interface right-click menu choice of more quickly, but not save.
Show Event: The display style of evoked events. Including unshown, show at changed, show at
per second.
Shield Mtg.: Whether to display and united display style of the montages which were shield at
the gathered time, is divided into: no shield (normal), show line, unshown.
Show Time: Into that displays full time, and the second time with absolute and relative
switching between the display style.
Show Filter Info.: After using the filtering, the filter information will be showed on the lower
left
corner of playback interface and the lower right corner of the preview one.
Show Plus: Adjust plus of EEG, ECG, BR, EMG, EOG. The parameters are also available on
the replay interface right-click menu choice of more quickly, but not save.
Scale Distance: Function of 11 and parameter settings.
Windows XP: When the waveform is played back, they are still in the screen, no matter opening
other attaching windows or taking correlative operations.
Gather Setting
You can set the line width in the acquisition, the number of leads, breathing the tone set, you can

34
also set the acquisition time limited gather, gathering a fixed length of waveform, the program
will automatically end.Meanwhile, you can also choose which reference montages to show, and
the corresponding waveform will be drawn at gather time.
Select menu "System Configure" -> "System Setting" -> "Gather Setting" and Figure 5-36
dialog box will appear.

Figure 5-36 Figure 5-37


Gather Time: System will stop gather automatically when gather time achieve this value. Select
this item otherwise it doesn't work.
Line width: Control the width of wave line both at gather and review time.
Video Mode: Control the mode of Video gathering. There are three modes to choose.
Show montage: Control display different kinds of channels at gather and review time.
Breath Prompt Sound Settings: To set the interval of breath, we can hear sound of respiratory

35
frequency prompt which controlled by system when gather waves.
Spark frequency channel: Show the waveform of spark frequency channel both at gather and
review time.
Show SpO2 channel: Show the waveform of SpO2 channel both at gather and review time.
Color Setting
The color here, including: background color, text color, color of all kinds of tag lines, and color
of the corresponding waveform induced events and the reference lead, by setting to control
playback and acquisition interface controls all colors, convenient color unified management.
Select menu "System Configure" -> "System Setting" -> "Color Setting" and Figure 5-37 dialog
box will appear.
In this page, you can set background color, text color and so on. The operation is as follows:
click the left button down on the color button which you want to change, then it will pop up
color palette, choose one color and click "Ok".
Notice: To distinguish all kinds of lines, please do not set the same color or similar color.
The software completed. The software completed from the old version to new version
upgrades. If the software is opened for the first time after installation, it will automatically
detect and update the old version case. If old version cases are imported and used, the case
will not apply to previous versions of the software, so you are advised to keep the backup of
the old cases if necessary.
Add Events Setting
Induced events added will be used when gathering. Induced events can use the software that

36
comes with the acquisition, if necessary, you can use custom-induced events acquisition,
waveform color, in accordance with the color settings in this dialog box displays.
Select menu "System Configure" -> "System Setting" -> "Add Events Setting" and Figure 5-38
dialog box will appear.

Figure 5-38 Figure 5-39


Click the left of mouse to the edit frame after "name" and input the name of the adding event,
click the color button below, click the "Add" button to finish the adding of the user-defined
induced event; if you need to delete the induced event, you can select the induced event item
with the left of the mouse and click "Delete" button, in this way the delete is completed.
Spark Event Setting
Flash stimulus package for the acquisition setup, when gathering, select event: Sparking, and
select the program number, click on "Spark", the flash will work in accordance with the set of
this dialog box.
Select menu "System Configure" -> "System Setting" -> "Spark Event Setting" and Figure 5-39

37
dialog box will appear.
When change frequency value, user should make the list pitched on state (viz.the list turn blue)
Scheme 1 is default, so you can not change this one. After change other schemes, click "Save"
button to save it.
Glossary Editing
Mainly used to store some of the commonly used terminology, when editing reports, these terms
are out of tune, click the items to input, eliminating the need to re-enter commonly used
sentences in trouble. (Specific operations, see Function 20: Report Design and Print).
Select menu "System Configure" -> "System Setting" -> "Glossary Editing" and Figure 5-40
dialog box will appear.

Figure 5-40
Input nomenclature in the edit box, and click the button " " on left, then the nomenclature
have been added in glossary list. Select one nomenclature and Click button " " to delete.
After editing, click "Ok"to save your setting.

38
Click "Default" setting, all the information will be defaulted and shown in this window. Click
"Ok" to save it. If you open the case again, all the information is the same as the defaulted
information. (each time you set the parameters, it will come into effect the next time you
open it).
Function 19: Auto Calibration
Click to come into the calibrate interface which only used for manufacture's checking.
Function 20: Report Design and Print
At any relief maps or wave replay views, user can print it out, the operation is follows: click
button will appear dialog as Figure 5-41, you can set the printing area range and you can
input the number of the first page and the last page, if you want the current page, set the range to
1-1, finally click "Ok" to print.

Figure 5-41 Figure 5-42 Figure 5-43


After analyzing, select menu "Print" -> "Print Report" or click button to edit report, shown as

39
Figure 5-42. Or you can select "Print" -> "Report Setting" to edit report name and select
information to print out, refer to Figure 5-43:
Click the edit area, you can input words; When editing, double click this area, you can use the
glossary which has been saved.Choose glossary (After choosing, you can edit the glossary).
Click edit area again, the glossary is written on the area. Press Enter or click other area, editing is
finished.
You can choose template, text align pattern, analysis view of report, edit the title and choose
patient information to print out.
Notice: Advising you set report template first, and then edit contents of report; Otherwise
the system renovation will lead to the contents lost!
Function 21: Input Office Word
Select the menu: "Print" -> "Output Office WORD", it can load template 1 of report into office
word. But before using this function, please ensure you have installed the software of Microsoft
Office.
Function 22: Select Patient And Extend Case
1.Select menu
"Case Management" -> "Select Patient" or click button to open a case. By clicking the left
mouse on one case, and clicking the "Open" button (or double click button down on one case) to
enter into replay view.
2.Extend Case
"Case Management" -> "Select Patient" or click button to open a case. By clicking the left

40
mouse on one case, and clicking the "Extend Case" button to enter into gather view.

Figure 5-44 Figure 5-45


As shown in Figure 5-44, click any one case with one click of the left mouse, the information of
the patient will be shown in the right side of the column; if there are many cases you need to
choose, you can also do conditional fast search to find the case you need.
The search operation is following:
Click small rectangular box before "Search" to select search function. Then select condition and
input condition content, finally click button "Search" down to begin searching for case. The
search function can find the case dimly, for example, if you select the name search condition, the
list will be shown to you all cases which theirs' names contains the words you have inputted. And
you could also set range of case number when you select case number search condition.
The content that accords with the case condition will be displayed in the list on the left, and for
those cases do not meet the conditions of the case, the system will point out, Click the "Clear up",

41
the content of the user input will be deleted.
Attention: In this function, patients' information only could be shown, not be modified.
Function 23: Patient List Management
Select menu "Case Management" -> "Patient list management" to enter picture Figure 5-45.
As shown in the menu of Figure 5-45, you can choose cases (You can also make use of the
function of research to find the case you need.), and modify it (them) in the right side of the
column. After that, click button "Save" and the prompt will appear. Then click "yes", you can
save the modified case. You can also delete the cases you want: select one of the cases you need
to delete, and click button "Delete"to delete the case.
Function 24: Output\Input Cases
You can select menu "Case management" -> "Case Input \\Output" to save the case files in folder
named "EEG 24 ChannelCaseData" under appointed path. See the picture as Figure 5-46.

Figure 5-46 Figure 5-47


Select the cases that you want to output and click "Output"button, then pop-up prompt window

42
shown as Figure 5-47, the meaning of clicking "Yes" button is to cut cases, the "No" is to copy
cases and the "Cancel" is to give up outputting cases.
"Open File" will load one patient case into software.
"Open Folder" will load all patient cases in the same folder which you selected into software,
and update the information in list.
Click "Output EDF", then pop-up prompt window shown as Figure 5-48, the meaning is output
EDF successfully. (BDF is the same).

Figure 5-48 Figure 5-49


Function 25: Pathologic Analyse
Select menu "Data Analysis" -> "Pathologic Analyse" or click button , will pop up dialog
shown as Figure 5-49, set up parameters of wave then click the "OK" to detect waveform, use
the appointed color to mark the pathologic.
Select menu "System Configure" -> "Inspect Setting" or click , will pop up dialog shown as
Figure 5-50, can set up category and color of different pathologies.

43
Figure 5-50 Figure 5-51
Choose appropriate color by clicking menu combo of every waveform. After setting up color, the
"Pathologic Analyse" is enabled again. And then inspect and color the waveform again also can
display the assigned wave view.
Function 26: SpO2 trend
When there is data of SpO2 in the case of patient, click button , you can observe the trend of
the oxygen values and pulse values recorded in the gathering process, As Figure 5-51 shows.
Press the Esc key on your keyboard to exit.

44
Function 27: Multilingual Interface
In the Chinese interface, choose "System Control" -> "Language Select" -> "English" switch to
English interface state; if system is in English interface state, choose the "System Configure" ->
"Language the Select" -> "Chinese" switch the system to Chinese interface state. The same
operation is also applicable to other languages.
5.6 Exits
Select menu "Case Management" -> "Exit" or click on top right corner of the menu to exit
from system of software.

45
Chapter 6 Maintenance and Troubleshooting
Problem Cause Correction
1. The power plug is not put in
Put the power plug in the socket.
the socket.
2. The power is not connected
The power light is not Connect the power wire with the
with the mainframe, or they’re
light. mainframe tightly.
not connected well.
Connect with the Technical
3. The power is broken.
department
1.The display plug is not Connect the display plug with the
The display light is not connected with the mainframe. mainframe.
light. 2. The power of the display is not
Turn on the power of the display.
turned on.
1.The display’s wire is not Connect the display’s wire with the
connected with the mainframe. mainframe.
The display’s light is 2.The light-control button is
Adjust the light-control button.
light but no display. turned to the darkest.
Connect with the technical
3.The display is broken.
department
The display’s light is not The display is broken. Connect with the technical

46
light and no display. department
The printer is not Read the related part of the operation
printed. manual.
The print’s color is too
The ink is used up. Change another new ink.
light or not clear.
1.The print line is not connected Connect the print line tightly.
The printer does not 2.The print line is broken. Change another new print line.
work. 3.The procedure of the print is Turn off the computer, and then turn
wrong. on the printer, and the computer.
1.The ground cable is not Place the ground cable in a standard
connected well. way.
2.The electrode is corroded. Change the new electrode.
EEG is disturbed a lot. 3.The cramp of the cable is
Change the new cable.
corroded.
4.The cable is broken. Change the new cable.
5.The hair is too dirty. Wash the hair.
One part is disturbed a The electrode of that place is not
Wear the electrode again.
lot. connected well.
Some channel is The electrode of that place is not
Wear the electrode again.
disturbed a lot. connected well.

47
Chapter 7 Appendix
The Principle of EEG Relief Map:
EEG relief map is a new method to analyze head bioelectricity in accordance with computer
technology and computer picture disposal technology. As a new method around the world right
now, it transforms huge complex but abnormity head bioelectricity to a form which could reflex
quantitative change and orientation by computer. It displays complexion about diversification of
head energy, it becomes an excellence, no wound and scientific method for skull sickness
diagnose and research on head energy quantitative analysis. It is useful to analysis correctly on
head energy using this technology.
Principle and Method:
Placement of electrodes: lay 16 leads on outboard head according to international 10/20 system
method standard, collect head wave signals and make A/D (modules transform) transform.
F-transform:
Transform the collected head wave signals to power chart map form to display the relationship
about head wave frequency flied and power.
Frequency Assort:
δ frequency flied: 0.8~3.8Hz
θ frequency flied: 4.0~7.8Hz
α1 frequency flied: 8.0~9.8Hz
α2 frequency flied: 10.0~12.8Hz
β1 frequency flied: 13.0~19.8Hz

48
β2 frequency flied: 20.0~30.0Hz
Insert Value Operation:
Find out some insert function according to two-dimension insert inside value principle based on
5 known
these 5
m
sin( X2500
sixteen numbers, calculate / 5)
point numbers. 5 (expressions
 sin The Y  n / 5)are as follows:
  (m / 5  n / 5) 
Print:
m 1 n 1 5 ( X  m / 5)  5 (Y  n / 5)
Print the received data to module picture and get power relief picture.EEG sign collection is
amplified by the EEG Mapping, then the signal is input to computer to complete analog to digital
conversion and Fourier transform.
Usual Relief Map:
Some power of δ wave relief map higher than regular value at 8 level and those exceed would
belong to abnormities.
Some power of θ wave relief map exceed normal value at 10 level local or whole head belong to
abnormities.
Evident un-symmetry of α wave relief map about the distributing of the two hemispheres. Those
exceed 3 or 5 belong to abnormities.
Power Chart Relief Map of Patient:
Most of the patient’s relief map would be abnormity and main of those are power exceed of δ
wave and θ wave, especially evident in the forward of head. Followed by the increase of the
power of α2 wave (namelyαwave) and the power of β wave also has trend to increase.
Electrode Locations:

49
10-20 EEG Locations:

Figure 7-1 Figure 7-2 Figure 7-3


EEG Electrode Placement:
 A2 is the reference electrode
 A is the grand electrode, locate between FP1 and FP2.Figure 7-2
ECG Electrode Placement: Figure 7-3
EOG Electrode Placement:
 One Electrode is ROC+ or LOC+, set as the follow picture:
 The other Electrode is ROC- or LOC-, the placement is the same as A2 or A1.

50
Figure 7-4 Figure 7-5 Figure 7-6
EMG Placement:
 Submental (chin)
 Often a third under the chin
 Two surface electrodes placed several centimeters apart:Figure 7-5
Sense organ of respiration placement:Figure 7-6

51