Documente Academic
Documente Profesional
Documente Cultură
2D621-092EN*B
SERVICE MANUAL
FOR
X-RAY MAMMOGRAPHY UNIT
MODEL MGU-1000A
(2D621-092EN*B)
IMPORTANT!
REVISION RECORD
3
No. 2D621-092EN*B
Safety Precautions
In this manual, the signal wordsDANGER, WARNING, and CAUTION are used to indicate
safety and other important instructions. The signal words and their meanings are defined as follows.
Please understand their meanings clearly before reading this manual.
DANGER Indicates an imminently hazardous situation which, if not avoided, will result
in death or serious injury.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may result
in minor or moderate injury.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in
property damage.
In addition to the above, the signal wordNOTE is used to indicate other important information.
NOTE Indicates reference information that enables more efficient use of the equipment.
4
No. 2D621-092EN*B
2. Safety Precautions
When servicing this system, observe the precautions described below to ensure safety of the operator
and service personnel.
DANGER: 1. This unit is not explosion-proof. Never use any flammable gases or liquids
near the unit. Doing so may result in a fire or an explosion.
If such gases or liquids must be used during work, open the windows of the
2. Before connecting the power cable, besure to turn OFF the breaker of the
power distribution board ofthe facility. Otherwise, death or serious injury
caused by an electric shock may result.
3. When turning the power supply ON/OFF during work, besure to inform all
co-workers and receive confirmation from them.
8. Lead and beryllium are usedin the X-ray tube unit in order to reduce
radiation exposure andto improve image quality. These materials are
highly toxic. Therefore, when disposing of the X-raytube, be sure to
observe the precautions described in the document provided with the
X-ray tube unit.
9. Wait at least 10 minutes after turning OFF the power supply before
touching internal components. Otherwise, the residual charge in the
capacitor unit may cause an electric shock.
5
No. 2D621-092EN*B
WARNING: 10. Take special care to avoid electric shock when performing high-voltage
cable checks, high-voltage generator checks, X-ray control section
checks, test adjustments, etc.
12. Before cleaning or disinfecting the unit, besure to turn OFF the power of
the unit. If the power of the unit is ON, inadvertent operation of theunit
may cause an accident.
13. Before X-ray generation, confirm thatthere are no persons except the
required personnel in the examination room.
15. When work inside the operating consoleis performed, be sure to turn
OFF the breaker at the power distribution board of the facility. Failure to
do so may lead to death or serious injury due to electric shock.
16. Do not use fuses other than those specified. Using fuses that do not
meet the specifications may cause a fire.
6
No. 2D621-092EN*B
2. Before starting work, clarifythe contents and procedures of the checks with
co-workers in order to enhance safety and efficiency.
3. During work, post a sign indicating that work isbeing performed and arrange
an appropriate location for storing removed covers etc.
5. Exercise extreme careto ensure safety during the handling of heavy units.
6. Do not subject the C-arm to excessive force in the rotation direction. Do not
pull cables forcibly or bend themexcessively. The cables may be
disconnected, damaged, etc.
7. To reduce the risk of burn injury, wait for partsto cool before touching
sections near hot lamps.
8. Handle the lamp and lamp holder carefully,because they become extremely
hot if the lamp has been lit for a prolonged period. When handling the lamp,
wear gloves and grasp the base of the lamp; do not touch the lamp with your
fingers directly. If the lamp is contaminated with fingerprints or oil, wipe it
clean with a cloth moistened with alcohol before lighting the lamp.
9. Confirm that the securing points of the important securing sections are
firmly tightened.
10. Before detaching/attaching the Buckydevice, be sure to turn OFF the power
of the control panel.
11. When unpacking the units, donot place the C-arm stand or control uniton its
side.
12. When placing the C-arm stand upright,without support, on thefloor, pay
attention to the following points.
(a) Do not hold the C-arm when moving the C-arm stand. If the C-arm is
rotated, the rotation limit switch may be actuated, making it impossible
to perform motor-driven rotation.
(d) Do not allow the cables to get caught between thefloor and the C-arm
stand base. If the cables get caught, they may break.
13. Before the power ofthe unit is turned ON, measure the voltageat the power
distribution board and use the appropriate tap of the power transformer of
the X-ray high-voltage generator and the C-arm stand according to the
measured voltage.
7
No. 2D621-092EN*B
CAUTION: 14. When removing and reinserting CNN1, be sureto turn OFF the power of the
unit. Otherwise, electric shock may occur.
15. If scaling for the compression plate isperformed with the face guard
mounted, interference between the face guard and the compression plate
may occur, resulting in damage to the unit or a serious accident.
8
No. 2D621-092EN*B
CAUTION: 1. When handling the covers and components, be extremely careful to avoid damaging
them.
2. When carrying in the unit, usea plate ramp at steps in the carrying-in route toprevent the
unit from being subjected to shocks. In addition, if the floor is made of a fragile material
or if the floor is too soft for the use of a cart, use plywood squares on the floor along the
route.
4. When the PC board mounting plate is remounted at its srcinal position after cable
connection of the X-ray high-voltage generator is completed, do not allow the cables to
get caught. If the cables get caught, they may break.
5. After the cables are installed in the cable pit, cover the pit with a metal plate to prevent
the cables from being damaged. To resist deformation, the metal plate should beat least
1.5 mm thick.
6. To improve noise resistance, be sure to route the signal cables and power cables
separately.
7. If the base cover is equipped with a display panel, cables are connected to this panel.
When removing the cover, remove the connectors and ensure that the cables are not
pulled. When remounting the cover, be sure to reconnect the cables.
8. When removing the front cover, it is recommended that the C-armbe raised slightly from
the lowest position to facilitate the removal work.
9. An emergency stop switch is provided onthe outer cover fixing frame, and a cable is
connected to this switch. When removing the cover, open the cover slightly to prevent
the cable from being pulled, and disconnect the connector. Then, remove the cover.
When remounting the cover, be sure to reconnect the connector.
(a) C-arm operation switches are provided on the side covers. When removing the
covers, open the covers slightly to prevent the cables from being pulled, and
disconnect the cableconnectors. Then, remove the covers.
(b) A retaining screw isinstalled at the top of each of side covers <2> and <3> in figure
5-5. Do not remove these two screws at the same time. If both screws are removed
together, the retaining bracket will fall.
11. Before executing list output, be sure to connect a USB storage device. If a USB storage
device is not connected, the operation will be cancelled.
12. Be sure to fully confirm safety before X-rays aregenerated in X-ray tube aging mode. In
addition, ensure that there is an interval of at least 30 seconds between successive X-ray
exposures.
13. When generating X-rays in quality control mode,exercise extreme care to ensure safety.
In addition, be sure to wait at least 30 seconds after the last X-ray generation before
generating X-rays again.
9
No. 2D621-092EN*B
CAUTION: 14. Be sure to confirm that thereare no objects such as peripheralunits within the
movement range of theC-arm stand. Failure to do so may cause interference between
the object and the C-arm stand, resulting in a serious accident.
15. The changes to the AEC data acquisition items do not affect the set radiographic
conditions. To change the filter, enter X-ray tube aging mode. Note that when the
focus size is changed to large, theBucky device should be mounted. When the focus
size is changed to small, the magnification radiography unit should be mounted.
16. the
Theactual
beam aperture
limiting device
isnot blade aperture
changed setting isIt stored
immediately. when to exit from
is necessary is beam
clicked, but
limiting device adjustment mode to make the setting effective. Click to exit
from beam limiting device adjustment mode andreturn to the utility main window. When
this is done, the actual aperture changes to the set aperture.
17. Be sure to change the tube current andthe main heating in increments of one step
while observing the waveform. Large changes to these values may damage the X-ray
tube.
18. Before executing log file output from the statistical information window, be sure to
connect a USB storage device. If a USB storage device is not connected, the operation
will be cancelled.
19. For adjustment, be sureto use the compression plate for 18cm × 24 cm contact
radiography.
20. If the compression force display doesnot correspond to the actual compression force,
adjustment cannot be performed properly. If "Adjustment of theCompressed Breast
Thickness Display" described in subsection 8.2 cannot be performed satisfactorily,
perform "Adjustment of the Compression Force" described in subsection 8.1.
21. The compression plate is made of soft material. Do not compress objects with corners,
such as a phantom, because doing so may scratch the compression plate.
22. When the compression plate is replaced, do not tighten the screws excessively. The
compression plate is made of a soft material and may be damaged by excessive
tightening.
23. VR adjustment alone does not change the value displayed on the LED. X-ray
generation must be performed each time adjustment is performed.
24. In AEC data acquisition mode, it is only possible to change the tube voltage and the
mAs value. All other operations are disabled. If it is necessary to change the filter,
move the compression plate, or perform any other operation during this work, switch to
X-ray tube aging mode from AEC data acquisition mode.
25. For data acquisition with small focus, be sure to mount the adaptor for magnification
radiography to the C-arm stand. Otherwise, small focus cannot be selected.
10
No. 2D621-092EN*B
27. In subsection 8.6.3 "Calculation of the AEC data table", the parameters differ for each
sheet. Be sure to select the correct sheet.
28. When the exposure field is adjusted, besure to use the maximum exposure fieldthat can
be set.
29. Before the adjustment parameters aretransferred, close all theother individual
adjustment windows. If an individual adjustment window isopen, the data from the PWB
may be loaded, leading to overwriting of the existing data in the individual adjustment
window.
If data is lost, restore the backed-up ini folder data and transfer the adjustment
parameters again.
30. To prevent damage to the system, do not use disinfectants of the types specified below.
The performance and safety of systems damaged by use of an improper disinfectant
cannot be ensured.
(b) Disinfectants that are not intended for use on metal, plastic, rubber, or coatings.
(c) Disinfectants that may enter the system, such as formalin gas or spray-type agents.
31. Do not clean the system using organic solvents (suchas paint thinner) or abrasive
cleansers, because they may cause discoloration or damage the surface finish.
32. When using a measuring instrument, fully understand thecontents of the measurement
and be extremely careful when handlingthe system and the instrument. Failure to do so
may result in incorrect measurement results or malfunction of the system or the
measuring instrument.
11
No. 2D621-092EN*B
CONTENTS
Page
1. SCOPE----------------------------------------------------------------------------------------------------------------------
15
3. DETECTION OF MALFUNCTIONS----------------------------------------------------------------------------------
18
6.2 Replacement of the Light Field Lamp and Adjustment of the Light Field --------------------------44
7. PC ----------------------------------------------------------------------------------------------------------------------------
57
12
No. 2D621-092EN*B
CONTENTS - continued
Page
8.3 Scaling--------------------------------------------------------------------------------------------------------------
94
13
No. 2D621-092EN*B
CONTENTS - continued
Page
8.3.3 Adjustment of the compression plate vertical movement position data ----------------96
8.5.2 Adjustment of the inverter pack parameters (C-arm stand vertical movement)---- 100
8.5.3 Adjustment of the inverter pack parameters (C-arm rotation speed) -----------------101
8.10 Replacement of the Compression Section Cover and Load Cell ----------------------------------119
14
No. 2D621-092EN*B
1. SCOPE
Operation
• manual 2B621-657EN
Installation
• manual 2C621-120EN
•Service connection diagram list 2K621-348EN
15
No. 2D621-092EN*B
The electrical wiring diagrams for the standard configuration (X-ray high-voltage generator, C-arm
stand, X-ray tube assembly, and operating console) are shown in figures 2-1 and 2-2.
Figure 2-1
16
No. 2D621-092EN*B
Figure 2-2
17
No. 2D621-092EN*B
3. DETECTION OF MALFUNCTIONS
When an error occurs, the corresponding error code is displayed in theoperation window. An error
message and a 5-digit error code aredisplayed together. The error code configuration is shown in
figure 3-1.
01020
Major Intermediate
Error classification Error type
category category
1 Initialization errors 01 Key shorting
02 EEP range error
2 C-arm stand errors 02 Sensor error (clockwise rotation)
03 Sensor error (counterclockwise rotation)
18
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
19
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
20
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
switch (right
switches on switch in the
the X-ray second
control row and
panel) of the AEC
the positionfor
connector
abnormalities. If the error cannot be cleared by the check,
replace the switch.
14 AEC position/center (2) switch error
(Cause) The AEC position switch (center switch in the second row of
the AEC position switches on the X-ray control panel) is in
the pressed status when the power of the unit is turned ON.
(Corrective action) Turn OFF the power of the unit and check the AEC position
switch (center switch in the second row of the AEC position
switches on the X-ray control panel) and the connector for
abnormalities. If the error cannot be cleared by the check,
replace the switch.
15 AEC position/left (2) switch error
(Cause) The AEC position switch (left switch in the second row of
the AEC position switches on the X-ray control panel) is in
the pressed status when the power of the unit is turned ON.
(Corrective action) Turn OFF the power of the unit and check the AEC position
switch (left switch in the second row of the AEC position
switches on the X-ray control panel) and the connector for
abnormalities. If the error cannot be cleared by the check,
replace the switch.
21
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
(Corrective action) Turn OFF the power of the unit and check the AEC size
switch (large) and the connector for abnormalities. If the
error cannot be cleared by the check, replace the switch.
21 AEC detection field (small) switch error
(Cause) The AEC size switch (small) is in the pressed status when
the power of the unit is turned ON.
(Corrective action) Turn OFF the power of the unit and check the AEC size
switch (small) and the connector for abnormalities. If the
error cannot be cleared by the check, replace the switch.
22
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
23
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
24
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
25
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
26
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
04 More than one AEC X direction sensor is actuated at the same time.
05 More than one AEC Y direction sensor is actuated at the same time.
06 More than one collimator blade limit sensor (chest-wall-end) is actuated at the
same time.
07 More than one collimator blade limit sensor (nipple-end) is actuated at the same
time.
08 More than one collimator blade limit sensor (left/right sides) is actuated at the
same time.
10 Grid photosensors 1 and 2 are ON at the same time during initialization.
11 More than one collimator switching mirror limit sensor is actuated at the same
time.
05 01 Position shift when C-arm rotation is stopped
(Cause) When C-arm rotation is stopped, there is a mismatch of 0.5°
or more between the potentiometer data and encoder data.
(Corrective action) Check the drive system and the inverter.
27
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
28
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
(Cause) The position data is not correct when the sensor actuates.
(Corrective action) Check the sensor position.
02 Lifter limit sensor error
(Cause) The position data is not correct when the sensor actuates.
(Corrective action) Check the sensor position.
03 Compression plate sensor error
(Cause) The position data is not correct when the sensor actuates.
(Corrective action) Check the sensor position.
04 AEC X direction sensor error
(Cause) The sensor does not actuate even though the position data
corresponds to the stop position.
(Corrective action) Check the sensor position.
05 AEC Y direction sensor error
(Cause) The sensor does not actuate even though the position data
corresponds to the stop position.
(Corrective action) Check the sensor position.
10 Grid sensor error
29
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
2 14 01 Position data mismatch between the encoder and potentiometer of the C-arm
(Cause) The encoder data does not correspond to the potentiometer
data.
(Corrective action) Perform scaling.
02 Position data mismatch between the encoder and potentiometer of the lifter
(Cause)The encoder data does not correspond to the potentiometer data.
(Corrective action) Perform scaling.
03 Position data mismatch between the encoder and potentiometer of the
compression plate
(Cause) The encoder data does not correspond to the potentiometer
data.
(Corrective action) Perform scaling.
30
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
3 01 01 X-ray CONT fan error
(Corrective action) Check the voltage between the fan and CNN8. If there are
no abnormalities, replace the fan.
02 X-ray tube fan error
(Corrective action) Check the voltages between the fan and the C ARM I/F
PWB and between the C ARM I/F PWB and the STAND
CONT PWB. If there are no abnormalities, replace thefan.
03 X-ray tube thermal switch ON (thermal error)
(Corrective action) Check the thermal circuit.
02 01 Overcurrent in the primary circuit of the heating inverter
(Corrective action) TR1, TR2, TR3, and TR4 of the H.V. INV PWB are
damaged. Replace the PWB.
02 Excessive heating current
(Corrective action) Adjust the heating current. If the error cannot be cleared,
replace the X-RAY CONT PWB.
03 Insufficient heating current
(Cause) Blown heating fuse or error in the heating control circuit.
(Corrective action) Check the heating fuse. Adjust the heating current. Check
for contact failure of CNN6 of the H.V. INV PWB.
03 01 Detection of overcurrent in the starter IGBT driver
(Cause) The IGBT of the STRTR INV. PWB is damaged.
(Corrective action) Replace the IGBT.
02 The stator does not start rotating even 5 ms after the system reaches the
READY status.
03 Current still flows to the coil 10 seconds after the system has left the READY
status.
(Cause) The stator rotates even though the system is not in the
READY status or the sensor detects a current despite no
rotation of the stator.
(Corrective action) Check the stator circuit and the current sensor.
04 01 The tube voltage is 10 kV or less
(Cause) Failure in the X-ray control inverter circuit or in the high-
voltage circuit
02 The tube voltage is 47 kV or more
(Cause) Failure in the X-ray control inverter circuit or in the tube
voltage feedback circuit
05 01 Overcurrent detected in the high-voltage IGBT driver
(Cause) Damage to the IGBT for X-ray control
02 The tube current is 20% above the set value.
(Cause) Failure in the tube current feedback circuit
03 The tube current is insufficient.
The tube current is 5 mA or less even 25 ms after the start of exposure.
(Cause) Failure in the tube current feedback circuit. Contact failure
of CNN6 of the H.V. INV PWB.
31
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
3 06 01 The SETUP signal goes OFF one or more times during X-ray generation.
07 01 High-speed link disconnected error (X-ray system)
02 High-speed link communication error (X-ray system, grid system)
03 High-speed link read error
04 High-speed link disconnected error (grid system)
08 01 The Ready SW ON (32) command has been sent, but the pre-exposure
condition setting (D1) command or exposure condition setting (D3) command is
not received.
02 The pre-exposure condition setting (D1) command has been received, but the
pre-exposure ready (D2) command is not received.
03 The PREP signal is ON, but the ON trigger of the EXP_WIN signals is not
detected.
04 In radiography with a moving grid, the grid is not ready for X-ray exposure
05 Pre-exposure has been completed, but the exposure condition setting (D3)
command is not received after the falling edge of the EXP_WIN signal.
06 The system is in auto-radiography modewith a moving grid, but the grid is not
ready for X-ray exposure after the falling edge of the pre-exposure EXP_WIN
signal.
07 Eight seconds has elapsed after the falling edge of the pre-exposure EXP_WIN
signal, but the ON trigger of the EXP_WIN signal is not detected.
09 01 The radiographic
outside condition
the specified rangenotification
(22 kV to 39(D4) command includes a tube voltage
kV).
02 The exposure condition setting (D1, D3) commands include a tube voltage
outside the specified range.
• Tube voltage setting range Large focus, contact Buckyradiography, and
radiography without the Bucky device:
22 kV to 39 kV
Small focus (magnification radiography):
22 kV to 35 kV
03 The pre-exposure condition setting (D1, D3) commands include an mAs value
outside the specified range.
• Contact Bucky radiography : 2 mAs to 450 mAs
• Magnification Bucky radiography : 2 mAs to 200 mAs
04 The exposure condition setting (D3) command includes a tube voltage outside
the specified range.
• Tube voltage setting range Large focus, contact Buckyradiography, and
radiography without the Bucky device:
22 kV to 39 kV
Small focus (magnification radiography):
22 kV to 25 kV
05 The pre-exposure condition setting (D3) command includes an mAs value
outside the specified range.
• Contact Bucky radiography : 2 mAs to 450 mAs
• Magnification Bucky radiography : 2 mAs to 200 mAs
32
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
3 09 06 • When the grid is not used (radiography without the Bucky device,
magnification Bucky radiography):
• A pre-exposure condition setting (D1)command or exposure condition setting
(D3) command including an instruction to move the grid (radiography without
the Bucky device: FPD calibration in progress) is received.
33
No. 2D621-092EN*B
Category
Interme- Error details
Major Minor
diate
34
No. 2D621-092EN*B
When an error occurs in the inverter in the C-arm stand, a protective function is operated, an alarm
sounds, system operation stops, and the inverter unit display automatically switches to error
display mode. The errors are described below.
35
No. 2D621-092EN*B
(2) Warnings
36
No. 2D621-092EN*B
4. LIST OF INTERLOCKS
(5) Automatic exposure control (AEC) is selected for the following techniques.
(AEC is not available in the following techniques)
NOTE: In contact radiography, onlythe large focus can be used when generating X-rays,
while in magnification radiography, only the small focus can be used.
(8) The light field display mirror does not retract out of the light field when the system is in the
READY status.
(13) The cassette is not pulled out and inserted after X-ray generation.
(14) In AEC mode, the compressive force is less than 40 N (not compressed status).
37
No. 2D621-092EN*B
When removing or remounting the cover of the X-ray high-voltage generator, refer to figure 5-1.
38
No. 2D621-092EN*B
Be sure to remove the covers from the C-arm stand in the following order: (1) base covers, (2)
outer rear covers, (3) outerfront covers, and (4) inner covers. When remounting the covers,
reverse the removal order.
When removing the base covers or the inner covers, refer to figure 5-2.
Figure 5-2 C-arm stand cover (base covers and inner covers)
39
No. 2D621-092EN*B
Remove and remount the outer rear covers referring to figure 5-3.
Remove and remount thecovers referring to figure 5-4. When removing the covers, be sureto
remove (1) the outer front covers and (2) the outer cover fixing frame in that order.
40
No. 2D621-092EN*B
CAUTION: An emergency stop switch is provided on the outer cover fixing frame, and a cable
is connected to this switch. When removing the cover, open the cover slightly to
prevent the cable from being pulled, and disconnect theconnector. Then, remove
the cover. When remounting the cover, be sure to reconnect the connector.
CAUTION: 1. If the base cover is equipped with a display panel, cables are connected to this
panel. When removing the cover, remove the connectors and ensure thatthe
cables are not pulled. When remounting the cover, besure to reconnect the
cables.
2. When removing the front cover, it is recommended that the C-arm be raised
slightly from the lowest position to facilitate the removal work.
41
No. 2D621-092EN*B
(1) Removal
(a) Remove front cover <1> from the C-arm referring to figure 5-5 (two screws).
(b) Remove left and right side covers <2> and <3> (7 screws).
(c) Remove top cover <4> from the breast support (4 screws).
(d) Remove bottom cover <5> from the breast support (6 screws).
CAUTION: 1. C-arm operation switches are provided on the side covers. When
removing the covers, open the covers slightly to prevent the cables from
being pulled, and disconnect the cable connectors. Then, remove the
covers.
2. A retaining screw is installed at the top of each of side covers <2> and
<3>. Do not remove these two screws at the same time. If both screws
are removed together, the retaining bracket will fall.
(2) Remounting
42
No. 2D621-092EN*B
Figure 6-1
(1) Removal
(b) Remove the four M3 screws and remove the collimator cover (wholeunit) as shown in
figure 6-1.
(2) Reinstallation
Reinstall the collimator cover by reversing the steps described in 6.1 (1).
43
No. 2D621-092EN*B
6.2 Replacement of the Light Field Lamp and Adjustment of the Light Field
(b) Remove the protective plate for irradiated heat and replace the lamp referring to figure
6-2.
• Lamp: BSX58-4045
CAUTION: Handle the lamp and lamp holder carefully, because they
become extremely hot if the lamp has been lit for a prolonged
period. When handling the lamp, wear gloves and grasp the
base of the lamp; do not touch the lamp with your fingers
directly. If the lamp is contaminated with fingerprints or oil,
wipe it clean with a cloth moistened with alcohol before lighting
the lamp.
Figure 6-2
(d) Press the light field lamp ON switch and confirm that the lamp is lit.
44
No. 2D621-092EN*B
<1> Mount the Bucky device onthe C-arm stand and insert a 18 cm× 24 cm cassette
(film not required).
Coin
Bucky
Cassette
D D
A B
Figure 6-3
<2> Place two 18-cm x 24-cm cassettes (with film) on the Bucky device faceplate as
shown in figure 6-3.
<3> Turn ON the lamp and place a coin or other marker at eachedge of the light field.
At this time, place a graph paper on the cassettes to make it easier to see the
edges of the light field.
<4> Generate X-rays with thefollowing conditions: large focus, 22 kV, and 5 mAs.
45
No. 2D621-092EN*B
<6> Adjust the position of the lamp mounting plate until each measurement satisfies
the following conditions.
Figure 6-4
46
No. 2D621-092EN*B
WARNING: Lead and beryllium are used in the X-ray tube unit to prevent
radiation hazards and lossof image quality. Since these materials
are harmful, follow the instructions in the product documents for the
X-ray tube unit when disposing off the unit.
Figure 6-5
47
No. 2D621-092EN*B
Figure 6-6
(e) Remove the four M6 flat-head screws fromthe X-ray tube (refer to figure 6-6).
(f) Replace the X-ray tube. Be sure to tighten the tube fixing screws (4-M6 flat-head
screws).
(g) Reconnect all the cables removed in step (d). At this time, apply silicone grease to the
high-voltage cable and reconnect it.
(j) Replace the bias resistor in the X-ray high-voltage generator(refer to subsection 9.3).
The PWB to be replaced is "L FOCUS BIAS" for a large focus and "S FOCUS BIAS" for
a small focus.
48
No. 2D621-092EN*B
NOTE: Be sure to follow the procedure outlined below to remove the X-ray tube. If the
C-arm rotation locking force is insufficient, the C-arm may rotate accidentally.
49
No. 2D621-092EN*B
The dust test is performedonly after the X-ray tube unit hasbeen replaced. This test is not
performed when the X-ray field is shifted in normal operation status.
(a) Select the large focus and set the Bucky device and cassette (with film).
(b) Select the molybdenum filter (Mo), select a kV and mAs that give a film density of 1.2 to
1.6, and then perform X-ray exposure.
(c) Develop the film and confirm that the film is free of dust.
If dust is found, remove the collimator and clean off the dust.
(d) Follow the same procedures to confirm that the rhodium (Rh) filter and the magnification
radiography adaptor are also free of dust.
CAUTION: Checks must be performed using the maximum combinable X-ray field size.
X-ray tube
Small focus
Large focus
Filter Mo, Rh
Figure 6-7
50
No. 2D621-092EN*B
(a) Place an intensifying screen on the breast support and turn ON the light field lamp.
Tape
Intensifying screen
Light field
Figure 6-8
(b) Provide markings to clarify each edge of the light field. Toshiba recommends the use of
black vinyl tape.
(c) Perform X-ray exposure and confirm the displacement of the light field (large focus,
30 kV, and 60 mAs).
51
No. 2D621-092EN*B
(d) Loosen the collimator fixing bolt (4-M5 bolts) and adjust the collimator position so that
the X-ray field is aligned with the light field. Repeat steps (b) and (c) until adjustment is
completed.
Small focus
Large focus
4-M5 bolt
Figure 6-9
(e) When the displacement between the X-ray field and the light field is almost eliminated,
securely tighten the collimator fixing bolts (4-M5 bolts).
52
No. 2D621-092EN*B
(f) Use the X-ray tube adjustment setscrews (fourM4 and two M6) and adjust them so that
the light field is equally divided at the center on the breast support.
2-M6 4-M4
Figure 6-10
53
No. 2D621-092EN*B
<1> Mount the Bucky device onthe C-arm stand and insert an 18 cm× 24 cm cassette
(with film). (Refer to figure 6-11.)
<2> Place two 18 cm × 24 cm cassettes (with film) on the Bucky device faceplate, turn
ON the lamp, place a coin at each edge of the light field, set the compression plate
for 18 cm × 24 cm contact radiography close to the cassette (almost in contact),
place a coin at the chest wall edge of the compression plate, and then perform
X-ray exposure (large focus, 25 kV, and 20 mAs).
m
m
0
3
ly
e
t
a
Compression m
i
plate x
o
r
p
p
A
Coin
Coin Bucky
Cassete
Y Z
A B
Figure 6-11
<4> Criteria
A+B < 14 mm, |A-B| ≤ 3 mm, C+D < 14 mm, 0 < F < 5 mm, 0≥ Y, and 0 ≥ Z when
the values obtained by increasing each measurement by a factor of 1.05 are used
for these calculations
• The measurements are acceptable when theall above conditions are met.
54
No. 2D621-092EN*B
<5> When adjustment of the large focus is completed, tighten the screws that were
loosened for adjustment and proceed to adjustment of the small focus.
2-M3
Figure 6-12
<8> Set an 18-cm × 24-cm cassette (with film) in place and generate X-rays with the
following conditions: small focus, 22 kV, and 2 mAs.
Figure 6-13
55
No. 2D621-092EN*B
<10> Criterion
0 < a < 5 mm
• If the above requirement isnot satisfied, adjust the mounting position of the light
field control plate using the M3 setscrews (2 locations) and repeat measurement.
(h) When all requirements for both the large focus and the small focus are satisfied,
retighten the loosened screws and remount the cover.
(i) Measure the exposure field on the developed film and perform adjustment of the beam
limiting device blades (left and right blades and nipple-end blade) so that the exposure
field becomes 138 mm x 138 mm in the adjustment window.
138 mm
m
m
8
3
1
Figure 6-14
56
No. 2D621-092EN*B
7. PC
When the button is selected from the examination window during normal operation,
the utility main window is displayed. In this window, settings can be performed by selecting the
buttons. The displayed buttons differ according tothe access rights of the logged in user.
(4) (7)
(1) (9)
(2) (8)
(3)
(5)
(6)
NOTE: To perform setting, be sure to log in to the system using an operator ID with service
engineer rights. The items that can be set differ according to the access rights.
57
No. 2D621-092EN*B
No.
(1) Date setting button The date setting window is displayed.
(5) Radiography result output button The radiography result output window is
displayed.
(6) Edit examinee information button The edit examinee information window
is displayed.
(7) X-ray tube aging button The X-ray tube aging window is
displayed.
58
No. 2D621-092EN*B
The date can be set in the window displayed by pressing this setting button.
Item Details
Date setting The date and time can be set using the and buttons.
NOTE: Data input is only possible using the spin buttons ( and ). The value cannot be
input directly in the field.
Click to apply the changes and return to the utility main window. Click to
cancel the changes.
59
No. 2D621-092EN*B
The compulsory entry items and the optional entry items in the examinee registration window can
be specified. Select for each item that is a compulsory entry item and select
for each item that is an optional entry item.
Item Details
Examinee ID The examinee ID cannot be set to [Optional].
Name Either or can be set.
Date of birth Either or can be set.
Comments Either or can be set.
Click to apply the changes and return to the utility window. Click to cancel
the changes and return to the utility window.
60
No. 2D621-092EN*B
Item Details
Set MLO angle The initial value of the MLO angle can be set (5° to 85°).
Set first compressive force The first compressive force level can be set in increments of 10 N
(80 N to 150 N).
Set second compressive force The second compressive force level can be set in increments of
10 N (80 N to 200 N).
AEC detector switch mode Either [Fixed to small] or [Switch between large/small] can be
selected.
Compression plate retraction The automatic retraction distance of the compression plate can
distance be set in increments of 10 mm (150 mm to 200 mm).
Click to apply the changes and return to the utility window. Click to cancel
the changes and return to the utility window.
NOTE: 1. If the MLO angle is changed during an examination, the changed MLOangle
becomes the default angle until the powerof the unit is turned OFF. When the power
of the unit is turned ON again, the value set here will be the initial MLO angle.
2. When the compressive force reaches thefirst compressive force levelset here, the
compression plate stops. Compression can be resumed until the compression force
reaches the second compression force level set here. Therefore, the value setfor the
second compression force level must be larger than that set for the first compression
force level.
61
No. 2D621-092EN*B
The list of operators is displayed. In this window, operators can be registered and deleted. Up to
20 operators can be displayed in the window. The displayed list differs according to the access
rights of the logged in user.
(2)
(3)
(1)
(4)
Item Details
Operator ID The operator ID is displayed (max. 20 bytes).
Operator name The operator name is displayed (max. 20 bytes).
Account The operator's access rights are displayed.
The page number in the operator list is displayed. The current page numberand the total
number of pages are displayed, separated by a stroke.
Used to move to the previous or next pageof the operator list. The operator selections that
are made are maintained even when the displayed page is changed.
62
No. 2D621-092EN*B
Click in the operator list window to display the operator name registration window.
If an operator is selected in the operator list window and is clicked, the selected
operator's information is displayed.
Item Details
Technologist ID Enter the technologist ID here (max. 20 bytes).
Clicking clears the entries. To set the entries and return to the operator list window,
click . At this time, if there is a mismatch between the new password and the new
password entered again for confirmation, the entries do not become effective and an error
message is displayed. To cancel the entries and return to the operator list window, click
.
63
No. 2D621-092EN*B
(3)
(1) (4)
(5)
(2)
The target data can be searched for in the displayed list by entering the search conditions as
follows.
Item Details
All The information for all examinees who have undergone
radiography is displayed.
Start date for search The start date of the search range is displayed. The current date
is displayed as the default.
Click to set the date. The calendar is displayed and the
date selected in the calendar is set as the start date of the search
range.
End date for search The end date of the search range is displayed. The current date
is displayed as the default.
Click to set the date. The calendar is displayed and the
date selected in the calendar is set as the end date of the search
range.
64
No. 2D621-092EN*B
The list of examinees who haveundergone radiography is displayed. More than one
examinee can be selected. The displayed items are as follows.
Item Details
Examinee The examinee ID is displayed (max. 15 bytes).
Examinee name The examinee name is displayed (max. 32 bytes).
Date of examination The date of examination is displayed as YY/MM/DD HH:mm.
The current page number in the examinee list is displayed together with the total number of
pages, separated by a stroke.
The examinee selections that are made are maintained even when the displayed page is
changed.
When an examinee is selected in the list and is clicked, the data is output to the
specified folder with the file name "Actual_YYYYMMDDHHMMSS.csv". To select all the
examinees in the list, click . Click again to clear the selection.
Click to return to the utility main window.
CAUTION: Before executing list output, be sure to connect a USB storage device. If a
USB storage device is not connected, the operation will be cancelled.
65
No. 2D621-092EN*B
Item Contents
Examinee ID Up to 15 bytes
Examinee name Up to 32 bytes
Comment Up to 64 bytes
Date of birth YY/MM/DD
Age Age at the time of examination
Sex Female/Male/Other
Radiography number
Examination date YY/MM/DD HH:mm
Facility name
Examination name
Name of photosensitive material
SF/CR classification 0: SF 1: CR
Contact spot 1: Contact 2: Magnification
Radiography mode 0: Full-automatic 1: Semi-automatic
2: Manual
Tube voltage [kV]
mAs (× 10)
Density
Filter 0: Mo 1: Rh
AEC 0: Small 1: Large
AEC position
Right/left 0: Right 1: Left
Surface dose [mGy]
Breast tissue dose [mGy]
C-arm angle
Breast thickness [mm]
Compression force [N]
Magnification ratio (× 100)
Film size 1: 18×24 2: 24×30
Operator name
66
No. 2D621-092EN*B
The examinee information can be edited in the window displayed by pressing this button.
Select the desired examinee and click to delete the examinee. Click to
display the edit window. The examinee information can be edited in the displayed window. To
select all examinees, click . Click to return to the utility window.
(3)
(1) (4)
(5)
(2)
A search can be performed by selecting the search conditions in the displayed list. The
following items can be set for the search conditions.
Item Details
All The information for all examinees who have undergone
radiography is displayed.
Start date for search The start date of the search range is displayed. The current date
is displayed as the default.
Click to set the date. The calendar is displayed and the
date selected in the calendar is set as the start date of the search
range.
End date for search The end date of the search range is displayed. The current date
is displayed as the default.
Click to set the date. The calendar is displayed and the
date selected in the calendar is set as the end date of the search
range.
67
No. 2D621-092EN*B
The list of examinees who haveundergone radiography is displayed. More than one
examinee can be selected. The displayed items are as follows.
Item Details
Examinee ID The examinee ID is displayed (max. 15 bytes).
Examine name The examinee name is displayed (max. 32 bytes).
Date of examination The date of examination is displayed as DD-MM-YY.
The current page number in the examinee list is displayed together with the total number of
pages, separated by a stroke.
Used to display the previous ornext page of the examinee list. The examinee selections that
are made are maintained even when the displayed page is changed.
68
No. 2D621-092EN*B
The information for the examinee selected in the examinee list can be corrected.
Item Details
Clicking clears the entries. To set the entries and display the examinee information
correction window, click . To cancel the entries and display the examinee information
69
No. 2D621-092EN*B
The X-ray tube aging can be performed from the window displayed by pressing this button.
During X-ray tube aging, various interlocks are disabled and X-ray exposure can be performed
continuously. Click to return to the utility window.
CAUTION: Be sure to fully confirm safety before X-rays are generated in X-ray tube aging mode.
In addition, ensure that there is an interval of at least 30 seconds between
successive X-ray exposures.
70
No. 2D621-092EN*B
Operations related to quality control of the unit can be performed from the window displayed by
pressing this button.
In the same manner as during X-ray tube aging, various interlocks are disabled in this mode,
allowing X-ray exposure tobe performed continuously. The X-ray exposure history is displayed.
Click to return to the utility window.
CAUTION: When generating X-rays in quality control mode, exercise extreme care to ensure
safety. In addition, be sure towait at least 30 seconds after thelast X-ray generation
before generating X-rays again.
71
No. 2D621-092EN*B
Scaling can be performed in the window displayed by pressing this button. Click to
return to the utility main window.
(1) Click the movement for which scaling is to be performed in the window shown in figure 7-12.
Figure 7-12
CAUTION: Be sure to confirm that there are no objects such as peripheral units within the
movement range of theC-arm stand. Failure to do so may cause interference
between the object and the C-arm stand, resulting in a serious accident.
72
No. 2D621-092EN*B
Use the C-arm operating footswitch or C-arm operating switch toperform operation. The
operation being performed is shown in blue.
Figure 7-13
Figure 7-14
73
No. 2D621-092EN*B
Perform operation using the C-arm operating footswitch or the C-arm operating switch.
Figure 7-15
74
No. 2D621-092EN*B
Before scaling the compression plate, be sure to remove the face guard.
Perform operation using theC-arm operating footswitch. The operation being performed is
shown in blue.
Figure 7-16
CAUTION: If scaling for the compression plate is performed with the face guard
mounted, interference between the face guard and the compression
plate may occur, resulting in damage to the unit or a serious
accident.
75
No. 2D621-092EN*B
The system information can be set in the window displayed by pressing this button.
Item Details
Systems used in combination Select the systems used in combination.
•CR : Only for systems installed in Japan
• Combination none : Only for systems installed outside Japan
Setting is performed at the time of shipment. It is therefore not
necessary to perform setting during installation. If a setting is
changed, it becomes necessary to restart the unit.
CR interlock mode When the CR is used in combination, the communication mode
with the CR is set. Set the CR interlock mode (enabled/disabled)
to correspond with the CR communication mode.
Auto-kV function Select whether the Auto-kV function is used. If [Auto-kV ON] is
set, the tube voltage is automatically raised (by 1 kV to 2 kV)
when the breast tissue density is high, obviating the need for
increasing the mAs value. In addition, when this function is
selected, the "Full auto transfer" function is disabled.
Sound setting for X-ray Select whether the buzzer sounds when X-ray generation is
generation completed or during X-ray generation.
Offset for adjusting the height Set the offset value for adjusting the height of the magnification
of the magnification radiography unit (-10 mm to 10 mm).
radiography unit
To set the entries and return to the utility window, click . To cancel the entries and
return to the utility window, click .
76
No. 2D621-092EN*B
7.12 Adjustment
Item Details
AEC data acquisition The AEC data acquisition window is displayed.
Beam limiting device The beam limiting device adjustment window is displayed.
adjustment
Tube current/main heating The tube current/main heating adjustment window is displayed.
Adjustment parameter transfer The adjustment parameter transfer window is displayed.
Full-auto transfer The full-auto transfer window is displayed.
NOTE: If the communication link with the C-arm stand breaks during display of the adjustment
window, the display returns to the utility window.
77
No. 2D621-092EN*B
Acquisition and setting of the AEC data are performed. Radiography can be performed from
this window.
(1)
(2)
The radiographic conditions that are currently set can be displayed and changed.
Item Details
Filter The selected filter type is displayed.
The filter can be changed in X-ray tube aging mode.
Tube voltage The tube voltage can be set.
mAs The tube current – time product can be set in increments of
1 mAs (2 mAs to 600 mAs).
78
No. 2D621-092EN*B
Item Details
Filter Mo (molybdenum), Rh (rhodium), or Al (aluminum) can be
selected.
Focus size switch Large or small can be selected.
CAUTION: The changes to the AEC data acquisition items do not affect the set
radiographic conditions. To change the filter, enter X-ray tubeaging mode.
Note that when the focus size is changed to large, the Bucky device should
be mounted. When the focus size is changedto small, the magnification
radiography unit should be mounted.
Clicking starts the AEC data acquisition application. To set the entries, click
. Click to return to the utility main window.
79
No. 2D621-092EN*B
The adjustment values for the beam limiting device can be set.
Item Details
Contact left/right Sets the size of the opening between the left/right blades of
the beam limiting device during contact radiography
(-10 to 10).
Contact nipple Sets the size of the opening between the nipple-end blade
and the chest-wall end of the beam limiting device during
contact radiography (-10 to 10).
Magnification left/right Sets the size of the opening between the left/right blades of
the beam limiting device during magnification radiography
(-10 to 10).
Magnification nipple Sets the size of the opening between the nipple-end blade
and the chest-wall end of the beam limiting device during
magnification radiography (-10 to 10).
To set the entries, click . Click to return to the utility main window.
CAUTION: The blade aperture of the beam limiting device is not set by clicking . It
is necessary to exit from beam limiting device adjustment mode to make the
settings effective. Click to exit from beam limiting device adjustment
mode and return to the utility main window.
80
No. 2D621-092EN*B
Item Details
Filter The filter type that is currently selected is displayed.
To change the filter, perform setting in X-ray tube aging mode.
Tube voltage The currently set tube voltage is displayed. Any value
between 22 kV and 35 kV can be set for contact radiography
and any value between 22 kV and 39 kV can be set for
magnification radiography.
mAs The currently set tube current - time product is displayed. Any
value between 2 mAs and 600 mAs can be set for contact
radiography and any value between 2 mAs and 200 mAs can
be set for magnification radiography.
Tube current The tube current setting can be adjusted between -100% and
100%.
Main heating The main heating setting can be adjusted between -100% and
100%.
Focus size The focus size can be switched between large and small.
To set the entries, click . Click to return to the utility main window.
CAUTION: Be sure to change the tube current and the main heating in increments of one step
while observing the waveform. Large changes to these values may damage the
X-ray tube.
81
No. 2D621-092EN*B
Adjustment parameter transfer is performed using the parameters stored in the PC. Select the
items to be transferred and click . Click to return to the utility main
window without executing the transfer operation.
82
No. 2D621-092EN*B
The tube voltage and filter type for different compressed breast thickness ranges can be set. To
set the entries, click . Click to cancel the entries and return to the utility
main window.
NOTE: 1. When setting the entries, confirm that largerkV values and appropriate filters are
set for higher compressed breast thickness values. When incorrect values are set,
an error message is displayed.
2. The full-auto transferfunction is enabled only when the auto-kV function is not
used. When the auto-kV function is used, the settings will be as follows.
21 to 30 26 Mo +1 +1
31 to 40 28 Mo +1 +1
41 to 50 30 Mo +1 +1
51 to 60 31 Rh +2 +2
61 to 70 32 Rh +2 +2
71 to 80 33 Rh +2 +2
81 to 90 34 Rh +2 +1
91 to 35 Rh +2 0
83
No. 2D621-092EN*B
Item Details
Error log The error log is displayed here.
Clicking clears the error log.
Clicking outputs the error log as a log file.
Operating time The operating time is displayed here.
Clicking resets the time.
Number of exposures The number of exposures is displayed here.
Clicking resets the data.
CAUTION: Before executing log file output from the statistical information window, be sure to
connect a USB storage device. If a USB storage device is not connected, the
operation will be cancelled.
84
No. 2D621-092EN*B
After setting is completed, back up the set data. The backed up data facilitates quick setting when
the system software is upgraded.
(1) Press [Ctrl] + [Alt] + [Delete] on the keyboard. When the "Windows Security" window
opens, select "Task Manager".
(2) When "Task Manager" opens, select "MainProcess" and click [End Task].
Figure 7-25
85
No. 2D621-092EN*B
(1) Click "Start" in the taskbar. In the displayed menu, select "My Computer".
(2) Open "C:\MGU1000A\" and copy the "ini" folder in this directory to the USB storage device.
(3) When the system software is updated, the system settings can be restored by copying the
backed up "ini" folder from the USB storage to the "C:\MGU1000A\" directory, overwriting
the existing folder in the directory.
(4) To restart the system software, double click the shortcut to "WatchProcess.exe" on the
desktop.
86
No. 2D621-092EN*B
7.14 Other
The information common to the windows is displayed at the top and bottom of each window.
Figure 7-26
87
No. 2D621-092EN*B
Figure 7-27
<d> X-ray source unit small focus indicator This indicator is displayed when the small
focus is selected for the X-ray source unit.
X-ray source unit large focus indicator This indicator is displayed when the large
focus is selected for the X-ray source unit.
88
No. 2D621-092EN*B
(3) Administrator
(5) Superuser
Examination
Date setting
Examinee information
item setting
Operator name list General operators General operators General operators All
display Administrators Administrators
Service engineers
Operator registration × General operators General operators All
Administrators Administrators
Service engineers
Operator name Own name only General operators General operators All operators in
correction Administrators Administrators the list
Service engineers
Operator deletion × General operators General operators General operators
Administrators Administrators Administrators
Service engineers Service engineers
89
No. 2D621-092EN*B
Quality control × ×
Scaling × ×
Statistical information × ×
: Permitted
× : Not permitted
90
No. 2D621-092EN*B
CAUTION: For adjustment, be sure to use the compression plate for 18 cm x 24 cm contact
radiography.
(1) Set S9 and S10 of the STAND CONT PWB to 6 and 4, respectively. The voltage of the load
cell is displayed on the LED.
(2) Connect the pressure gauge to the C-arm stand as shown in figure 8-1.
Compression plate
Chest-wall-end
Measurement point of the pressure gauge
(3) Lower the compression plate until the pressure gauge reads 5 kg.
(4) While maintaining the status in step (3), perform adjustment with VR1 so that the LED of the
STAND CONT PWB reads 1980.
(5) Lower the compression plate again until the pressure gauge reads 20 kg.
(6) While maintaining the status in step (5), perform adjustment with VR2 so that the LED of the
STAND CONT PWB reads 6930.
91
No. 2D621-092EN*B
CAUTION: For adjustment, be sure to use the compression plate for 18 cm x 24 cm contact
radiography.
(1) Set S9 and S10 of the STAND CONT PWB to 1 and 0, respectively.
(2) Turn ON the power of the unit. The system starts up and enters compression plate
adjustment mode. At this time, only the compression force isdisplayed on the C-arm stand
control panel.
Figure 8-2
(3) Place a 10-cm x 10-cm acrylic plate (5 cm thick) on the C-arm stand.
Acrylic plate
Figure 8-3
NOTE: Be sure to use a 10-cm x 10-cm acrylic plate and set it at the center of the chest-
wall-end.
92
No. 2D621-092EN*B
(4) Compress the acrylic plate with the compression plate until the compression force reaches
100 N.
(5) While maintaining the status in step (4), press S6 of the STAND CONT PWB.
(6) Remove the acrylic plate and move the compression plate so that the LED of the STAND
CONT PWB reads 3.0.
(7) While maintaining the status in step (6), adjust the sensor position for detecting the 3-mm
position. Move the sensor upward and fix it at the position where LED 7 lights.
(8) Move the compression plate so that the LED of the STAND CONT PWB reads 200.0.
(9) While maintaining the status in step (8), adjust the sensor position for detecting the 200-mm
position. Move the sensor upward andfix it at the position where LED8 lights.
(10) Move the compression plate so that the LED of the STAND CONT PWB reads 250.0.
(11) While maintaining the status in step (10), adjust the sensor position for detecting the 250-mm
position. Move the sensor upward and fix it at the position where LED 9 lights.
(12) Turn OFF the power of the unit and return the settings of S9 and S10 to 0.
Figure 8-4
CAUTION: If the compression force display does not correspond to the actual compression
force, adjustment cannot beperformed properly. In this case, perform compression
force adjustment as described in subsection 8.1.
93
No. 2D621-092EN*B
8.3 Scaling
NOTE: Scaling can be performed from the control console. Refer to subsection 7.10 "Scaling
Button".
(1) Set S9 and S10 of the STAND CONT PWB to 2 and 0, respectively.
(2) Turn ON the power of the unit. The system starts up and enters adjustment mode. The
AEC size selection (large and small) and AEC position (nipple-end) LEDs light (refer to
figure 8-5).
Lit
Figure 8-5
(4) Hold down the C-arm clockwise rotation switch until the C-arm stops at -150°. Note that
the C-arm counterclockwise rotation switch is disabled until the C-arm reaches -150°.
(5) Hold down the C-arm counterclockwise rotation switch until the C-arm stops at 180°. Note
that the C-arm clockwise rotation switch is disabled until the C-arm reaches 180°.
(6) Hold down the C-arm clockwise rotation switch until the C-arm stops at 0°. Note that the
C-arm counterclockwise rotation switch is disabled until the C-arm reaches 0°.
(7) The C-arm position data is now saved. Turn OFF the power of the unit and set both S9
and S10 to 0.
NOTE: 1. When adjusting the C-armrotation, the C-arm hasto be rotated from -150° to 180°.
Before starting the work, raise the C-arm stand so that the C-arm does not interfere
with the floor. In addition, confirm that there are no obstacles within the C-arm
rotation range.
2. When adjustment work isto be continued, it is not necessary to turn OFF the power
of the unit.
94
No. 2D621-092EN*B
The C-arm stand vertical movement position data can be set as follows.
(1) Set S9 and S10 of the STAND CONT PWB to 2 and 1, respectively.
(2) Turn ON the power of the unit. The system starts up and enters adjustment mode. The
AEC size selection (large and small) and AEC position (nipple-end) LEDs light (refer to
figure 8-5).
(4) Hold down the C-arm stand DOWN switch until the C-arm stand stops at the lowest
position. Note that the C-arm stand UP switch is disabled until the C-arm stand reaches
the lowest position.
(5) Hold down the C-arm stand UP switch until the C-arm stand stops at the highest position.
Note that the C-arm stand DOWN switch is disabled until the C-arm stand reaches the
highest position.
(6) Hold down the C-arm stand DOWN switch until the C-arm stand stops at the lowest
position. Note that the C-arm stand UP switch is disabled until the C-arm stand reaches
the lowest position.
(7) The C-arm stand vertical movement position data is now saved. Turn OFF the power of
the unit and set both S9 and S10 to 0.
NOTE: 1. Before starting the C-arm stand vertical movement adjustment procedure, setthe
C-arm to 0°. In addition, confirm that there are no obstacles within the C-arm
movement range.
2. When adjustment work isto be continued, it is not necessary to turn OFF the power
of the unit.
95
No. 2D621-092EN*B
The compression plate vertical movement position data can be set as follows.
(2) Turn ON the power of the unit. The system starts up and enters adjustment mode. The
AEC size selection (large and small) and AEC position (nipple-end) LEDs light (refer to
figure 8-5).
(4) Hold down the compression plate DOWN switch until the compression plate stops at the
lowest position. Note that the compression plate UP switch is disabled until the
compression plate reaches the lowest position.
(5) Hold down the compression plate UP switch until the compression plate stops at the
highest position. Note that the compression plate DOWN switch is disabled until the
compression plate reaches the highest position.
(6) Hold down the compression plate DOWN switch until the compression plate stops at the
lowest position. Note that the compression plate UP switch is disabled until the
compression plate reaches the lowest position.
(7) The compression plate vertical movement position data is now saved. Turn OFF the
power of the unit and set both S9 and S10 to 0.
NOTE: 1. Before starting the compression plate vertical movementadjustment procedure, set
the C-arm to 0°. In addition, be sure to confirm that there are no obstacles within
the compression plate movement range.
2. When adjustment work isto be continued, it is not necessary to turn OFF the power
of the unit.
96
No. 2D621-092EN*B
To replace the lithium battery for the backup RAM on the PWB, follow the procedures below.
(1) Battery
Incorrect
Correct
97
No. 2D621-092EN*B
When the inverter packs are replaced, set the following parameters.
NOTE: When setting the inverter pack parameters, be sure to set parameter 79 at the end.
98
No. 2D621-092EN*B
Frequency setting
volume
[RUN] key [STOP/RESET] key [STOP/RESET] key [UP/DOWN] key
Figure 8-6
99
No. 2D621-092EN*B
8.5.2 Adjustment of the inverter pack parameters (C-arm stand vertical movement)
When the inverter packs (INV1) are replaced, strong vibration may occur during C-arm stand
vertical movement. In this case, perform adjustment as follows.
(1) Remove the covers from the C-arm stand and turn ON the power of the unit.
(b) Press the MODE key once. The system enters parameter setting mode.
(c) Press the SET key. The most significant digit blinks. Press the UP/DOWN key to set
it to "0".
(d) Press the SET key again. The middle digit blinks. Press the UP/DOWN key to set it
to "7".
(e) Press the SET key again. The least significant digit blinks. Press the UP/DOWN key
to set it to "9".
(f) Press the SET key again. The current setting is displayed. Press the UP/DOWN key
to set it to "1".
(g) Hold down the SET key for about 1.5 seconds. The set value blinks and is written,
and the system enters PU operation mode.
(h) Press the MODE key four times to return to parameter setting mode.
(i) Press the SET key. The most significant digit blinks. Press the UP/DOWN key to set
it to "0".
(j) Press the SET key again. The middle digit blinks. Press the UP/DOWN key to set it
to "3".
(k) Press the SET key again. The least significant digit blinks. Press the UP/DOWN key
to set it to "8".
(l) Press the SET key again. The current setting is displayed. Press the UP/DOWN key
to lower the value by 1. Lowering the setting value reducesvibrations.
(m) Hold down the SET key for about 1.5 seconds. The set value blinks and is written.
(n) Press the MODE key four times to return to parameter setting mode.
(o) Repeat steps (c) to (g) to return the setting value of parameter 79 to "2".
(p) Press the MODE key three times to return to monitor mode.
(3) Move the C-arm stand up and down to confirm that there is no significant vibration. If
necessary, repeat the steps described in (2) to eliminate vibrations.
(4) Turn OFF the power of the unit and remount the cover.
(5) Turn ON the power of the unit and perform the final checks.
100
No. 2D621-092EN*B
After the inverter packs (INV2) are replaced, rotate the C-arm to check the rotation speed.
(1) Remove the cover from the C-arm and turn ON the power of the unit.
(2) Rotate the C-arm and confirm that the value displayed on the inverter pack display is
between 43 Hz and 45 Hz. If this condition is satisfied, skip steps (3) and (4) and proceed
directly to step (5).
(b) Press the MODE key once to enter parameter setting mode.
(c) Press the SET key. The most significant digit blinks. Press the UP/DOWN key to set
it to "0".
(d) Press the SET key again. The middle digit blinks. Press the UP/DOWN key to set it
to "7".
(e) Press the SET key again. The least significant digit blinks. Press the UP/DOWN key
to set it to "9".
(f) Press the SET key again. The current setting is displayed. Press the UP/DOWN key
to set it to "1".
(g) Hold down the SET key for about 1.5 seconds. The set value blinks and is written,
and the system enters PU operation mode.
(h) Press the MODE key four times to return to parameter setting mode.
(i) Press the SET key. The most significant digit blinks. Press the UP/DOWN key to set
it to "0".
(j) Press the SET key again. The middle digit blinks. Press the UP/DOWN key to set it
to "3".
(k) Press the SET key again. The least significant digit blinks. Press the UP/DOWN key
to set it to "8".
(l) Press the SET key again. The current setting is displayed. Press the UP/DOWN key
to change the value. When the set value is changed by 2 Hz, the rotation speed is
changed by 1.5 rps.
(m) Hold down the SET key for about 1.5 seconds. The set value blinks and is written.
(n) Press the MODE key four times to return to parameter setting mode.
(o) Repeat steps (c) to (g) to return the set value of parameter 79 to "2".
(p) Press the MODE key three times to return to monitor mode.
(4) Rotate the C-arm and confirm that the displayed value is between 43 Hz and 45 Hz. If
necessary, repeat the steps described in (3) until the displayed value is within the specified
range.
101
No. 2D621-092EN*B
(5) Turn OFF the power of the unit and remount the cover.
(6) Turn ON the power of the unit and perform the final checks.
The film development conditions and cassettes differ according to the site. It is therefore
necessary to perform gain adjustment at the time of installation.
(2) Remove the covers from the C-arm stand to adjust the variable resistors (VRs)on the
STAND CONT PWB. For the cover removal procedure, refer to subsection 5.2.4. The
STAND CONT PWB is mounted atthe lower right of theC-arm stand. Set rotary switches
S9 and S10 on the STAND CONT PWB to "6" (this setting permits the AEC data to be
displayed on the 7-segment LED of the PWB).
(3) Mount the Bucky device to the C-arm stand and set a cassette containing a film.
(4) Turn ON the power of the unit. When the login window is displayed on the monitor, log in
using an operator ID with service engineer rights. After login, select the X-ray tube aging
mode in the utility window of the operating screen.
102
No. 2D621-092EN*B
(5) Place a phantom (BR12(50/50) 4 cm) on the C-arm stand as shown in figure 8-8. Set the
AEC position to the center of the area closest to the chest wall and set the AEC size to
large. Lower the compression plateso that it just touches the phantom.
Phantom
CAUTION: The compression plate is made of soft material. Do not compress objects
with corners, such as a phantom, because doing so may scratch the
compression plate.
(6) In the X-ray tube aging mode window, set the radiographic conditions to"Manual", "Mo",
"28 kV", and "20 mAs".
(7) Generate X-rays and check the display of the 7-segment LED of the STAND CONT PWB.
If the 7-segment LED displays "27", gain adjustment is not necessary. If "27" is not
displayed, adjust VR10 of the STAND CONT PWB, generate X-rays again, and check the
display of the 7-segment LED again. Adjust VR10 until "27" is displayed.
CAUTION: VR adjustment alone does not change the value displayed on the LED.
X-ray generation must be performed each time adjustment is performed.
(8) When adjustment for the large AEC size is completed, set the AEC size to small.
(9) Generate X-rays with the same conditions asfor the large AEC size and check the display
of the 7-segment LED ofthe STAND CONT PWB. If the 7-segment LED displays "27",
gain adjustment is not necessary. If "27" is not displayed, adjust VR11 of the STAND
CONT PWB until "27" is displayed.
(10) When the above steps are completed, return thesettings of rotary switches S9 and S10 of
the STAND CONT PWB to "0" (normal operation mode). In the operating window, press
[Return] to return to the utility window. Continue with the work from subsection 8.6.2 "AEC
data acquisition".
103
No. 2D621-092EN*B
(1) In the utility window, select [Adjustment] → [AEC data acquisition]. The AEC data
acquisition window opens.
Select
Select
Figure 8-8
(2) Place the phantom correspondingto the compressed breast thickness to be measured on
the C-arm stand in the same manner as for AEC gain adjustment. Set the AEC position to
the center of the area closest to the chest wall and set the AEC size to large.
CAUTION: In AEC data acquisition mode, it is only possible to change the tube voltage
and the mAs value. All other operations are disabled. If it is necessary to
change the filter, move the compression plate, or perform any other
operation during this work, switch to X-ray tube aging mode from AEC data
acquisition mode.
(3) Set the radiographic conditions(tube voltage and mAs) and generate X-rays.
(5) Measure the density several timesby varying the mAs value and determine the mAs value
that provides the desired s value(or image density). Fill in the determined value inthe
table on the following page.
(6) Repeat the work described above and perform measurement at least six times for each
table, setting a different mAs value each time.
104
No. 2D621-092EN*B
(7) There are four AEC data tables, large focus/Mo, large focus/Rh, small focus/Mo,and small
focus/Rh (only two tables when magnification radiography is not performed: large
focus/Mo and large focus/Rh).
NOTE: 1. In AEC data acquisition mode, the mAs value can be changed in increments
of 1 mAs.
CAUTION: 1. For data acquisition with small focus, be sure to mount the adaptor for
magnification radiography to the C-arm stand. Otherwise, small focus
cannot be selected.
Figure 8-10
105
No. 2D621-092EN*B
• Large focus/Mo
] 25
V
k
[
e 26
g 27
a
tl
o 28
v
e
b 29
u
T
30
31
32
33
• Large focus/Rh
27
] 28
V
k
[ 29
e
g 30
a
tl
o 31
v
e
b 32
u
T
33
34
35
36
37
38
39
106
No. 2D621-092EN*B
• Small focus/Mo
] 25
V
k
[
e 26
g 27
a
tl
o 28
v
e
b 29
u
T
30
31
32
33
• Small focus/Rh
25
] 26
V
[k 27
e
g 28
a
tl
o 29
v
e
b 30
u
T
31
32
33
34
35
107
No. 2D621-092EN*B
(1) Select [Input] in the AEC data acquisition window. Excel starts and the approximate-
calculation sheet opens.
Select
Figure 8-11
CAUTION: The parameters differ for each sheet; therefore, be sure to select the sheet
correctly.
(4) When data input is completed, press the [mAs data calculation] button. Approximate
calculation is performed. Select [OK] in the confirmation window that opens at each step.
108
No. 2D621-092EN*B
(3) data.
Input the measurement
Figure 8-12
(6) Confirm that the actual measurement points are on or close to the approximate curve. If
they deviate significantly from the curve, increase the number of measurement points.
Figure 8-13
109
No. 2D621-092EN*B
(7) After confirmation, press the [Calculate AEC value] button. 10-bit integrated value data is
calculated. Select [OK] in the confirmation window that opens at each step.
Figure 8-14
(9) In the same manner as above, calculate the integratedvalue data for each AEC table.
(10) When calculation for all the data tables is completed, select the [table_data] sheet.
110
No. 2D621-092EN*B
(11) Click the [Input data] button. The calculated integrated data is read out from each sheet
and displayed. Select [OK] in the confirmation window that opens at each step.
(12) Click the [Save data] button. The AEC data tables are saved to the hard disk of the PC in
the csv format. Select [OK] in the confirmation window that opens at each step.
Figure 8-15
(13) When all the work described above is completed, close Excel. When a message asking
whether the changes are to be saved is displayed, click [Yes]. The AEC data acquisition
window is then displayed.
NOTE: The [Save Data] operation saves each AEC data table in
c:\MGU1000A\ini\ folder. The file names are as follows. They can be
overwritten.
111
No. 2D621-092EN*B
When the csv files are saved, transfer the data to the STAND CONT PWB.
(1) In the AEC data acquisition window, select the data type of the file to be transferred.
(2) Click the button. The data is transferred and written into the memory of the
STAND CONT PWB.
Figure 8-16
112
No. 2D621-092EN*B
NOTE: 1. When is executed, only the selected AEC data table is written. In
figure 8-16, only the data for large focus/Mo is written.
Figure 8-17
113
No. 2D621-092EN*B
The signals from each sensor and switch of the C-arm stand can be checked on the LEDs of the
STAND CONT PWB by changing thesetting of S8. The items displayed on the LEDsdepend on
the switch settings as shown in the tables below.
(2) AEC
114
No. 2D621-092EN*B
(4) Collimator
(5) Switch
115
No. 2D621-092EN*B
The main system and display data adjustment modes can be changed by changing the rotary
switch settings (S9 and S10) of the STAND CONT PWB. The items displayed on the LEDs
depend on the switch settings as shown in the table below.
Switch
Mode Item 7-segment LED LED display
S9 S10
Special 1 0 Mechanical adjustment Normalized value LED9: 250 mm
plalte of the encoder LED8: 200 mm
LED7: 3 mm
Adjustment of 2 0 C-arm rotation Normalized value LED9: 180°
the main unit of the LED8: -150°
potentiometer LED7: 0°
1 C-arm stand vertical LED9: 690 mm
movement LED8: 0 mm
2 Compression plate vertical LED9: 250 mm
movement LED8: 200 mm
LED7: 3 mm
Free operation 3 0 C-arm rotation No display LED9: 180°
LED8: -150°
LED7: 0°
1 C-arm stand vertical LED9: 690 mm
movement LED8: 0 mm
2 Compression plate vertical LED9: 250 mm
movement LED8: 200 mm
LED7: 3 mm
Normal mode 5 0 C-arm rotation encoder Raw value LED9: 180°
1 Normalized value LED8: -150°
LED7: 0°
2 C-arm rotation
Raw value
3 potentiometer Normalized value
4 C-arm stand vertical Raw value LED9: 690 mm
5 movement encoder Normalized value LED8: 0 mm
6 C-arm stand
Raw
vertical
value
7 movement potentiometer Normalized value
8 Compression plate vertical Raw value LED9: 250 mm
9 movement encoder Normalized value LED8: 200 mm
LED7: 3 mm
A Compression
Rawplate
value
vertical
B movement potentiometer Normalized value
C AEC X-axis encoder Raw value LED9: Right
D AEC Y-axis encoder LED8: Center
LED7: Left
LED6: 4 (Nipple-end)
LED5: 3
LED4: 2
LED3: 1 (Chest-wall-end)
E Beam limiting device None LED9: OPEN
F encoder (chest-wall-end) LED8: CLOSE
116
No. 2D621-092EN*B
Switch
Mode Item 7-segment LED LED display
S9 S10
Normal mode 6 0 Beam limiting device Raw value LED9: OPEN
1 encoder (nipple-end) Normalized value LED8: CLOSE
2 Beam limiting device Raw value LED9: OPEN
3 encoders (left/right sides) Normalized value LED8: CLOSE
4 Load cell voltage A/D conversion LED9: -1.5-V disconnection
potentiometer
117
No. 2D621-092EN*B
(2) Turn OFF the power of the unit and remove the cover from the C-arm stand.
(3) Replace the STAND CONT PWB, which is mounted on the side of the main unit. Be sure
return all the switches and jumpers on the PWB to their srcinal settings.
(4) Turn ON the power of the unit. Perform the procedures listed below.
Subsection or manual
Procedure to be performed Description
to be referred to
(a) Scaling 7.10 or 8.3 Perform scaling for the drive
section.
(b) Adjustment of the breast 8.1 Adjust the breast compression
compression force force.
(c) Adjustment of the 8.2 Adjust the compressed breast
compressed breast thickness thickness.
(d) Transferring the adjustment 7.12.4 Transfer all the parameters.
parameters
(e) Operational checks Installation manual There should be no abnormalities.
CAUTION: Before the adjustment parametersare transferred, close all the other individual
adjustment windows. If an individual adjustment window is open, thedata from
the PWB may be loaded, leading to overwriting of the existing data in the
individual adjustment window.
If data is lost, restore the backed-up ini folder data and transfer the adjustment
parameters again.
(5) Perform the operational checks and, if no problems are found, remount the covers.
118
No. 2D621-092EN*B
<2>
<1>
<3> <4>
<5>
<6>
<1>
Figure 8-18
(1) Removal
(a) Remove front/rear cover <1> and panel <2> (five screws).
(b) Remove the cable and load cell connector from the panel.
(d) Remove link plate <4> (four screws). Spacers and a bearing are installed inside the
mounting screw section of the link plate; therefore, be careful when removing the link
plate.
119
No. 2D621-092EN*B
(2) Reinstallation
(c) Set S9 and S10 of the STAND CONT PWB to "6" and "5", respectively.
(d) While in the no-compression status,perform adjustment using the adjustment screw
(figure 8-19) so that the 7-segment LED displays a value between 5 and 25.
Adjustment screw
Figure 8-19
120
No. 2D621-092EN*B
<2> <1>
<1> <1>
<2>
<1>
Figure 8-20
(1) Removal
(b) Loosen set screws <2> (two locations, two screws at each location).
(2) Reinstallation
(a) Mount the plate to the arm and lightly tighten mountingscrews <1> (four locations).
(b) Two set screws each are provided on the outside and the inside of the plate. Adjust the
plate so that the tip of the plate protrudes from the Bucky device within a range from
0 mm to 2.5 mm, and partially secure the plate (using two screws at each of the two
sections).
Figure 8-21
CAUTION: When the compression plate is replaced, do not tighten the screws excessively. The
compression plate is made of a soft material and may be damaged by excessive
tightening.
121
No. 2D621-092EN*B
The resistance values corresponding tothe various positions must beas specified below. When
the potentiometer is replaced, set the resistances accordingly.
(1) Remove the C-arm covers. For the removal procedure, refer to subsection 5.2.4 " C-arm
and C-arm stand covers ".
(2) Lower the compression plate to the lowest position (about 250 mm).
(3) Remove the cable from the potentiometer and remove mounting plate <1> (two screws).
(5) Remove the potentiometer and replace it. Set the resistance value of the potentiometer
(between the blue cable and the white cable) to 1.0 Ω
k.
(6) Install the new potentiometer in the unit by reversing the removal procedure and reconnect
the cables.
<2>
<1>
<3>
Figure 8-22
122
No. 2D621-092EN*B
Perform adjustment with VR1 so that the period of the waveform for TP9 (ST-INV1) of the
X-RAY CONT PWB becomes 5.5 ms. Also perform adjustment with VR2 so that the
waveform width becomes 2.2 ms.
2.2 ms
5.5 ms
Figure 9-1
(a) Turn ON the power of the unit and perform adjustment withVR16 so that the voltage of
TP30 of the X-RAY CONT PWB becomes 0 V before the unit enters the standby state
(about 30 seconds before heatingstarts). (Alternatively, remove CNN1 from the H.V.
INV PWB, turn ON the power of the unit, and perform adjustment with VR16 so that
TP30 of the X-RAY CONT PWB becomes 0 V.)
(b) Set SW2 of the X-RAY CONT PWB to "TEST" (when CNN1 of the H.V. CONT PWB has
been removed, reinsert itin advance). Connect a dummy filament (resistance of about
1 Ω) between L and C of the generator. Perform adjustment with VR9 of the X-RAY
CONT PWB so that a voltage of 1.0 V (1.4 V during preheating) at TP42 of the X-RAY
CONT PWB corresponds to 1.0 A (1.4 A during preheating) of the filament current.
(c) When adjustment is completed, return SW2 of the X-RAY CONT PWB to the srcinal
setting (NOR).
CAUTION: When removing and reinserting CNN1, be sure to turn OFF the power
of the unit. Otherwise, an electric shockmay result.
(a) Set SW2 of the X-RAY CONT PWB to "TEST". In contact radiography with large focus,
manually set the tube voltage to 22 kV. Perform adjustment with VR3 so that the
waveforms for TP21 (kV-ACT) of the X-RAY CONT PWB and the bleeder voltage are
similar.
(b) Perform adjustment with VR4 so that the measurement value for the bleeder becomes
±4% (21.1 to 22.9 kV) of the specified voltage.
(c) When adjustment is completed, return SW2 of the X-ray CONT PWB to the srcinal
setting (NOR).
123
No. 2D621-092EN*B
(a) Set SW2 of the X-RAY CONT PWB to "TEST". In contact radiography with large focus,
manually set the tube voltage to 22 kV. Perform adjustment with VR8 so that the
measurement value for the bleeder becomes 86 mA±5% (81.7 mA to 90.3 mA).
(b) When adjustment is completed, return SW2 of the X-RAY CONT PWB to the srcinal
setting (NOR).
(a) Set SW2 of the X-RAY CONT PWB to "TEST". In contact radiography with large focus,
manually set the tube voltage to22 kV. Perform adjustment with VR9 so that the
waveform for TP28 of the X-RAY CONT PWB is smooth with no overshooting or
undershooting. At this time, the measurement value of thetube voltage should be
86 mA ±5% (81.7 mA to 90.3 mA).
(b) When adjustment is completed, return SW2 of the X-RAY CONT PWB to the srcinal
setting (NOR).
(a) Confirm that tube voltages other than 22 kV are also within ±4% of the set voltage.
When the tube voltage is set to more than 34 kV, it is necessary to change the filter
from Mo to Rh.
(b) Also confirm that the tube current is within ±5% of the set value.
(c) When the value is not within the specified range, perform adjustment again from step
(3) " Adjustment of the tube voltage".
NOTE: 1. Adjustment of the starter has been performed at the time of shipment. It is
therefore not necessary to adjust it at the site.
2. Adjustment of the tubevoltage and tube current have been performed at the
time of shipment. They are also adjusted automatically by feedback control
based on actual detected values ofthe tube voltage and tubecurrent. It is
therefore generally not necessary to adjust them at the site.
3. Adjustment of filament heating is necessary after the X-RAY CONT PWB or H.V.
INV. PWB of the X-ray tube assembly or the X-ray high-voltage generator is
replaced.
124
No. 2D621-092EN*B
*1 *1 *2
Tube-voltage X-ray high-
C-arm stand Tube-current meter bleeder meter bleeder voltage generator
X-ray TUBE SIDE TRANS SIDE ANODE SIDE
tube Generator
assembly
X-RAY
CONT
PWB
T1 T2
(-) (+)
Tube- Tube-
current voltage mAs meter
meter meter
Figure 9-2
(a) Insert the receptacle for mammography into the "TUBE SIDE" receptacle section of the
tube-current meter bleeder.
(b) Insert the receptacle for mammography into the "ANODE SIDE" receptacle section of
the tube-current meter bleeder.
(c) Pull out the high-voltage cable from the X-ray H-high-voltage generator and insert the
cable into the "TUBE SIDE" of the tube-current bleeder.
(d) Connect the normal high-voltage cable between the "TRANS SIDE" of the tube-current
meter bleeder and the "ANODE SIDE" of the tube-voltage generator.
(e) Connect the high-voltage cable for mammography between the tube voltage bleeder
side where the receptacle for mammography has been inserted and the high-voltage
generator.
(f) Connect the shielded line of the high-voltage cable for mammography to the ground
terminal of the tube-voltage meter. Then connect this terminal to theground terminal of
the X-ray high-voltage generator.
(g) Disconnect the cablebetween terminal boards T1 and T2 on the X-RAY CONT PWB.
Then connect the plus side of the mAs meter to T2 and the minus side of the mAs meter
to T1.
125
No. 2D621-092EN*B
(1) If the bias resistor (one each for the large and small focuses) resistance valuesare different
from those of the L and S focus bias resistors on the X-RAY CONT PWB in the X-ray high-
voltage generator when the X-ray tube is changed, replace the bias resistors.
(2) When the X-RAY CONT PWB is replaced, remove the L-focus and S-focus bias resistors
from the old PWB and mount them to the new PWB.
3
N
N
C
4
1
N
1 N
N C
N
C
CNN12
CNN10
126
No. 2D621-092EN*B
(3) Wipe the scale (made of insulating material) with a dry cloth and insert the scale into the oil
inlet carefully up to 40 mm.
(4) Withdraw the scale carefully and read the oil level.
(Reference value from the oil inlet: 30 ±3 mm)
127
No. 2D621-092EN*B
Fuse
Rating Part number Remarks
No.
F1 30 A/250 V NORMAL ASF052*30A X-ray power supply
F3 2 A/250 V NORMAL ASF101-10 Heating
F4 1 A/250 V NORMAL ASF101-08 Power ON
Fuse
Rating Part number Remarks
No.
F1 1 A/250 V TIME-LAG ASF041*1A
F2 1 A/250 V TIME-LAG ASF041*1A
F3 10 A/32 V TIIME-LAG ASF041*10A Field lamp
WARNING: Be sure to use the designated fuse. Failure to do so may leadto fire.
128
MEMO
129
1385, SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN 324-0036, JAPAN
130 E