Bell Prover Gas Meter Test Bench is an equipment to test domestic gas
Feature:
1. The device can test 4pcs gas meters simultaneously, the device is an
ideal testing equipment for gas meter factory and testing institute.
2. The standard meter is bell prover, accuracy 0.5%, it can test Qmax.
1.2.1 Among the pipe system of test bench, the testing range of absolute
The device consists of bell prover and testing table. The effective
volume inside the bell prover is regarded as standard volume, when bell
prover falls, the air inside bell prover flows to tested meters, the fall
flowed.
3
the button.
This button is to control the scanning heads. All scanning heads can drop
down to test meter by pushing this button, and scanning heads can go
Noted: this button can not ensure each scanning head can catch testing
This button is to stop the existing test and start new tests.
1.5.5.1 The top LED (6 digits) displays flow rate during test, and
1.5.5.3 When test bench is switched on or test bench is ready for work
indicator is on.
in the pipes due to loose clip, and devise stop testing automatically.
X X X X X X
○ ○ ○ ○ ○
○ ○ ○ ○ ○
F1 F2 F3 F4 F5
○ ○
1 2 3
Start Stop
○
4 5 6 7
Enter
Reset 8 9 0 OK
1.6.2 ‘left/right’ key is for choosing meter type. Press this key, the
indicator is on to test meter with left air input, on the contrary, to test
1.6.6 Key F3 is to record bell prover pulse coefficient. Press the key,
Qmin.
1.6.10 Key ‘Stop’, press this key to force stop of all tests during testing;
press this key to save setup of parameter and logout during setup.
1.6.11 Key ‘air input’, press this key to raise bell prover.
1.6.14 Key ‘confirmation’ is for confirm data and save during setup.
1.6.15 Press F1, the indicator is on, LED displays 000000, the test
bench is ready for tests, please open the outlet valve in the bell prover
Basic pulse×100L
Basic volume(L)
After test, press F1, the indicator is off and LED displays – – – – –, test
7
stops.
1.7.1.1 Press ‘set’ once, the indicator is on, LED displays 1– – – –L,
semi-automatic test, please use figure keys to set value and press
‘confirmation’ to save.
please use figure keys to set value and press ‘confirmation’ to save.
L, it means to set start-delay time for testing Qmax, 0.2Qmax and Qmin
XX(range 00-99s), meantime please set the time for Qmax, 0.2Qmax
L, no function.
set the time with figure keys then press ‘confirmation’ to save the data.
time with figure keys then press ‘confirmation’ to save the data.
semi-automatic test, meantime please set with figure keys then press
the first place from right, 1 means that bell prover raises automatically
9
after tests, 0 means that bell prover raises manually after tests, please set
XXX(range 00-99), meantime please set with figure keys then press
1.7.2.3 (G4——36;G2.5——24;G1.6——10)
means to set system amendment value for test pressure loss at Qmax,
pressure loss for gas meter G1.6 at Qmax with figure keys, and press
‘confirmation’ to save the data, the same to set for gas meter G2.5 and
1.8.1 The test bench has a function shift key F2, when F2 is pressed, the
meantime record the initial counter, when leakage tests over, test bench
open Qmin flow points automatically, test bench will stop automatically
1.9.1 When test bench is switched on, the test bench is at initial state and
LED displays
– –b–
A: to test gas meter with left air input, B: to test gas meter with right air
input.
Qmax
0.2Qmax
Qmin
11
X X X X X
X X X X X
X X X X X
4-1
RS232 port is for connecting PC. Before performance, the test bench
should be connected with PC. The function of pins: pin ②— TXD, pin
③— RXD、pin ⑤— GND.
This port is for connecting scanning head. The function of pins: pin ①
For testing impulse meter, this port is for connecting the impulse cable.
12
1.11.1 Put tested meters at testing position and clip tightly, adjust the
bell prover, then start to test after the bell prover is stable for 1 minute.
1.11.3 Choose the correct meter type (right/left air input), then choose
1.12.2 Notes:
The test bench should work indoor with stable temperature 20°C±2°
C,humidity: 30~80%RH。
between air inside bell prover, water or oil and room temperature.
Every time test bench stop work, please raise the bell prover and switch
power supply.
regularly, 1 time/year.
13
2.3.1 Set up
14
Operator can input data through data list easily, moreover data can be
b. Set parameter
For the performance, the system data should be set as test bench status,
Balance time: when airtightness test, the period for pressure balance in
Testing time: when airtightness test, the testing period after the pressure
Testing pressure: when airtightness test, the pressure which the test
Leakage rate: this is a basic value to judge the airtightness test is past or
not.
Pressure loss: this is a basic value to judge the pressure loss test is past
There are different testing scheme for different meter, meantime testing
setting testing points, testing times, testing volume, error limit, testing
Before testing, the testing data should be chosen as tested meters such as
Noted: when the tested meter is changed, the testing scheme should be
changed as well.
b. Testing operation
At testing stage, first step is to drop clamp, if for all clamps, please
choose meter No. 0, the click ‘drop clamp’ button.; if for separate meter,
please choose the meter No, then click ‘drop clamp’ button;
c. Test Airtightness
If click ‘single’, after finishing airtightness test, the system enter waiting
state.
Click ‘pressure loss’ to have pressure loss test, the menu is as following:
The testing result will be compared with the set max. and mini. limit and
If click ‘single’, after finishing pressure loss test, the system enter
waiting state.
19
If click ‘continuous’, after finishing pressure loss test, then system enter
e.You can click ‘single’ or ‘continuous’ to have error tests, the main
function of menu:
If click ‘single’, after finishing a testing points test, the system enter
enter the next tests automatically till all the testing points are finished,
Note: if the testing results is within max. and mini. limit, the figure is
buttons.
20
moreover system can draw error curve as per errors of all the testing
21
points. The curve can be enlarged and reduced or removed left and right.
Statistic can be as per daily, monthly, yearly and range. The menu is as
following:
The statistic displays the quantity and testing data of tested meters and
meter number.
First please confirm the test bench is at working condition, and connected
with PC, once the system can remember the interface, once searching (3-10s),