Documente Academic
Documente Profesional
Documente Cultură
RADIOGRAPHY TESTING
COURSE NO. : SA-M-QC-3.5
Approved by : N. Nagaich
Training Superintendent, RAPS 1 to 4
I express my sincere thanks to Shri Prashant Puri STO for giving his
valuable suggestion during preparation and Shri C.M. Mishra and Shri A.K.
Singh for his guidance and keep me cheerful to complete the task. I owe my
sincere gratitude to Shri N. Nagaich, Training Superintendent RAPS 1-4 for
his kind guidance, cooperation and encouragement to me to complete this
manual.
V. K. Jain
SO/D, NTC.
CONTENTS
DESCRIPTION Page No.
2.1 Introduction 8
2.2 Some Important Properties of Alpha, Beta and Gamma rays 12
2.3 Production Of X-rays 14
2.4 Interaction of Radiation with Matter 16
2.5 Radiation Quantities & Units 22
2.6 Biological Effects of Radiation 25
2.7 Operational Limits 28
Chapter - 3 X-RAY TECHNOLOGY 32
3.1 Introduction 32
3.2 X and Gamma Rays 32
3.3 Production Of X-rays 33
3.4 Requirements of an Industrial X-Ray Tube 35
3.5.1 Fluoroscopy 35
3.5.2 Fine Focus Tube 36
3.5.3 Rod Anode Tube 36
3.5.4 Crawler X-ray Units 36
3.6 X-Ray Generator Circuits 36
3.7 Linear Accelerator 37
4.1 Introduction 38
4.2 Production of Radioisotopes 38
147
4.2.1 Production by activation process 38
4.2.2 Production by nuclear fission 39
4.2.3 Particle accelerators are also used for production of some radioisotopes 39
4.3 Sources for Industrial Radiography 39
4.3.1 Specific Activity 39
4.4 Fabrication of Radioisotopes 39
5.1. Introduction 41
5.2. Types of Cameras 41
5.3 Design 42
5.3.1 Source Assembly 43
5.3.2 Shielding Material 43
5.3.3 Source Exposure 44
5.3.4 Safety Devices 44
5.3.5 Fabrication 45
5.3.6 Maintenance 45
5.3.7 Transportation 45
5.3.8 Installation 45
5.4 Quality Control During Production 45
5.5 Equipment Development 46
5.5.1 Lead Shielded, Low Cast Equipment 46
5.6 Type Approval of Gamma Radiography Equipment 48
6.1 Introduction 50
6.2 Construction of Film 50
6.2.1 Film Base 50
6.2.2 Emulsion 50
6.2.3 Binding Layer 50
6.2.4 Protective Layer 50
6.3 Characteristics of Films 51
6.3.1 Film Density 51
6.3.2 Characteristic Curve 51
6.3.3 Effects of Radiation Energy 53
6.3.4 Various Types of Films 53
7.1 Introduction 54
7.2 Types of radiography screens 54
7.3 Metallic foil screens 54
148
7.3.1 Reduction of scattered radiation 55
7.3.2 Intensifying Action 55
7.4 Salt or fluorescent screen 55
9.1 Introduction 68
9.2 Factors Affecting Sensitivity 68
9.3 Image Quality Indicators (IQI) 68
9.3.1 Commonly Used Image Quality Indicators 69
9.4 Identifications 72
9.5 Control of Radiographic Quality 72
9.5.1 Technique Classification 73
9.5.2 Material and its Thickness 73
9.5.3 Film and Screen 73
149
9.5.4 Film Processing 73
9.5.5 Density of the Radiograph 73
9.5.6 Identification Marking System 73
9.5.7 Source, its active dimensions and calibration 73
9.5.8 SFD 73
9.5.9 Exposure geometry 73
9.5.10 IQI, its location 73
9.5.11 Exposure Time 73
9.6 Unsatisfactory Radiographs 73
9.6.1 High Density 74
9.6.2 Low Density 74
9.6.3 High Radiography Contrast 74
9.6.4 Low Radiography Contrast 74
9.6.5 Poor Definition 74
9.6.6 Fog 75
9.6.7 Spurious Indications 75
9.7. Viewing of Radiographs 75
9.7.1 Brightness Conditions 75
9.7.2 Illuminators 75
9.7.3 Viewing Conditions 76
9.7.4 Effects of Viewing Conditions on Image Quality 77
9.7.5 Optimum film Density 77
Chapter-10 DEFECTOLOGY 78
10.1 Introduction 78
10.2 Object Image Relationship 78
10.3 Defect Detection 81
10.4 Defect Indications 81
11.1 Introduction 85
11.2 Technique parameter selection 86
11.2.1 Radiation Source 86
11.2.2 Type of Film 86
11.2.3 Geometrical Factors 88
11.2.4 Sensitivity 88
11.3 Latitude Technique 89
11.4 Weld Inspection Techniques 89
11.4.1 Technique for Fusion Welds on Flat Plates 89
11.4.2 Techniques for welded joints on pipes or Cylindrical objects 93
150
Chapter-12 RADIOGRAPHIC STANDARDS SELECTION 95
OF TEST METHODS
12.1 Introduction 95
12.2 Inspection Standards and Codes 95
12.2.1 Standards for Image Quality Indicators (IQI) (Penetrameters) 95
12.2.2 Standards for Techniques 95
12.2.3 Standards for Specific Structures and Products 96
12.2.4 Reference Radiographs 96
12.3 Standards for Welds in Steel Plates 97
12.3.1 Classification of Techniques 97
12.3.2 Films and Screens 97
12.3.3 Source to Film Distance (SFD) 98
12.3.4 Length of Specimen 98
12.3.5 Film Density 98
12.6 Circumferential welds on pipes and cylindrical objects 100
12.7 Radiography of Welds of Boilers and Pressure Vessels 101
12.8 Radiography of Castings 101
12.9 Bureau of Indian Standards - Codes and Standards 102
12.10 Acceptance Limits 104
12.10.1 Acceptable and Unacceptable Flaws 104
12.10.2 Classification of Defects 104
12.10.3 Acceptance / Rejection 104
12.10.4 Typical Acceptance Standards 104
151
Chapter-14 RADIATION ACCIDENTS IN INDUSTRIAL RADIOGRAPHY 131
152
CHAPTER -1
Selection of a material for a given job depends upon its physical and mechanical
properties. Most structural materials are subjected to external forces, which generate
mechanical stresses. The reaction of the part to these stresses can be critical to its
continued functioning. Hence, it is important for the NDT personal to know the
normal material properties, as well as effect of discontinuity upon the material
serviceability.
The application for which a material is used, determines which property is most
important.
The chemical properties (reaction with other materials) are of interest, mainly
because of the need for resistance to corrosion.
Physical properties of materials are associated with their atomic structures eg.,
density, crystalline type, atomic spacing, specific heat, melting point, etc.
1
b. Plastic Region at Low Stresses :indicates that at a certain stress level an
abrupt increase in strain occurs and the material is said to yield.
Strains beyond the elastic limit, which result in residual strains on unloading are
called inelastic or plastic strains. Materials which undergo relatively large plastic
strain to rupture are referred to as "ductile". Those which undergo little or no plastic
strain, prior to rapture, are referred to as "brittle".
(C) Creep
Creep is the flow of material over a period of time, when under a load too small to
produce any measurable plastic deformation at the time of application. The simplest
type of creep test is made by just hanging a weight on the test specimen and
observing its elongation, as a function of time by using a microscope of other
sensitive detector of strain.
(D) Fatigue
(E) Hardness
The Hardness of the material is measured by hardness tester. Three types of hardness
tests are the scratch, rebound and penetration tests. Hardness measurements are
extremely useful as a quick and rough indication of the mechanical properties of a
metal.
2
inherent discontinuity. If caused during further processing, fabrication or finishing, it
is called processing discontinuity. Finally, if it arises during the use of the end
product either due to environment, load or both, it is called service discontinuity.
Casting is the process of causing liquid metal to fill a cavity and solidify into a
useful shape.
The discontinuity that can occur during casting process are given below ;
c. Pipe : The molten material, after pouring into the mold, starts to cool and it
solidifies. The solidification process starts from the surface and travels
towards the centre of the ingot. On solidification, the molten metal contracts.
Since the centre of the ingot is the last to cool and solidify, most of the
shrinkage is observed in the centre. This results in a cavity called "PIPE". It
may extend from the top towards the interior of the ingot along the axis.
d. Cold-shut : Cold shut is formed when molten metal is poured over solidified
metal. When the metal is poured, it hits the mold too hard and spatters
small drops of metal. When these drops of metal hit higher up on mold, they
stick and solidify. When the rising molten metal reaches and covers the
solidified drops of metal and a crack like discontinuity is formed. Cold shuts
can also be formed by the lack of fusion between two intercepting surface of
molten material of different temperatures.
e. Hot tear (Shrink crack) : Hot tear is caused by unequal shrinking of light
and heavy sections of a casting as the metal cools. In a casting having light
and heavy sections, the light sections, being smaller, solidify faster ; they
shrink faster pulling the heavier sections towards them, as they are hotter and
do not shrink as fast.
3
unlike the "pipe" in the ingot, which always occurs only at the top portion of
the ingot.
g. Micro shrinkage : Shrinkage can also occur in the casting at the mold gate,
i.e., at the entrance to the mold through which the molten metal is poured.
Shrinkage occurs, if metal at the gate solidifier or is blocked off while some
of the metal beneath is still molten. Shrinkage which occurs at the gate
appear as a number of small holes called "micro shrinkage". Micro shrinkage
can also occur deeper within the metal, if the mold is improperly designed.
h. Blow holes : Blow holes are small holes on the surface of the casting and
are caused by external gas emanating from the mold itself.
Welding is the process of joining metals. Two basic types of welding are used,
fusion welding and pressure welding.
B Pressure welding : In this process, pressure and heat are applied to the
pieces to be welded. The pressure produces plastic deformations, the heat
then produces re-crystallization across the boundary, resulting in the
formation of new crystals which are an integral part of both pieces and thus
the welding takes place.
4
All Critical welds require NDT for assurance of quality or as a means to enable
repairs to be made.
The possible defects (Fig. 1.1) that can occur in fusion weldments are listed below :
a. Overlaps : A fin of surplus metal on the end of a billet may be caught and
folded into the surface during subsequent rolling. The bent-over metal, while
forced tightly against the main stock, will not bond to it. The resulting lap is
a possible initiation point for a fatigue crack.
b. Crater cracks : Crater cracks are caused in the weld bead by improper use
of heat source, either when a weld is started or stopped. A crater crack can
also occur at the temporary stop of the weld. Crater cracks are primarily of
three types-transverse, longitudinal and multiple star-shaped.
c. Stress cracks : The cracking of weld metal and base metal in or near the
weld zone and it is usually caused by high stresses set up by localised
dimensional changes. Stress cracks are most likely to occur when weldments
are of heavy sections. These cracks usually occur transverse to the weld in a
single pass weld and longitudinal in a multiple pass weld.
d. Porosity : Porosity is the term for the pockets or voids free of any solid
material that are frequently found in weld metals. Porosity can come from
gases released by the cooling weld metal from gases formed by chemical
reaction in the weld metal. Porosity may be scattered uniformly throughout
entire weld, isolated in small areas or concentrated at the root. Gas pores are
usually spherical in shape, although they may also occur as non-spherical
pockets along grain boundaries. Most welds contain some amount of porosity
which may be micro of macro in size.
e. Slag inclusions : This term is used to describe the oxides and other non-
metallic solid materials that are entrapped in the weld metal of between weld
metal and base metal. Slag inclusions may be caused by contamination of the
weld metal by the atmosphere, but in most cases, they are generally derived
from electrode covering materials or fluxes, employed in arc welding
operations. In multi layered welding operations, failure to remove the slag
between passes will result in slag inclusions in these zones. Slag inclusions,
are generally linear and may occur either as short particles or long bands.
5
g. Lack of fusion : Lack of fusion or incomplete fusion, as it is frequently
termed, describes the failure of adjacent weld metal and base metal or
interweld passes to fuse together completely. This failure to obtain fusion
may occur at side wall or in the interpass region. Lack of fusion is usually
elongated in the directions of welding and may have either rounded or sharp
edges depending on how it is formed.
i. Undercut : During welding of the final or cover pass, the exposed upper
edges of the weld preparation tend to melt or run down into the deposited
metal in the weld groove. Undercutting occurs when insufficient filler metal
is deposited to fill the resultant, at the edge of the weld bead. The result is a
groove that may be intermittent or continuous and parallel to the weld bead.
Undercutting may be caused by excessive welding current, incorrect arc
length, high speed, incorrect electrode manipulation, etc.
a. Stationary load,
6
b. Dynamic load, unidirectional or multidirectional, multi-directional being more
serious,
c. High temperature,
d. Pressure creating stress above a material's elastic limit
e. Corrosive environment,
f. Vibrations,
g. Excess loading,
h. Improper maintenance and
i. Ageing.
There are two generally accepted types of material failure in service : one is the
easily recognized 'FEATURE' or separation into two or more parts ; the second is
the less easily recognized "EXCESSIVE PLASTIC DEFORMATION" or change of
shape and/or position.
a) Corrosion
b) Fatigue
c) Wear
Wear is probably the most important factor in the deterioration of machinery with
moving components, after limiting both the life and performance of such equipment.
Wear is the loss of material from the surface. Wear is affected by a variety of
conditions, such as the type lubrication, loading, Speed, temperature, materials,
surface finish and hardness
d) Overstress
It may happen when a part is accidentally exposed to a load which is much greater
than its design load. When this happens, the component may undergo plastic
deformation of fracture, to relieve the high stress within the part.
7
CHAPTER - 2
2.1 INTRODUCTION
All materials in the universe are made of different elements. Hydrogen, oxygen, iron,
iridium, gold, lead are some of the elements we are familiar with.
Atom is the smallest part of an element. An atom has a positively charged central
portion called 'nucleus'. It also has negatively charged electrons going around the
nucleus, in different orbits. The nucleus has two types of particles - protons which
are positively charged and neutrons which do not carry any charge. An atom, is
electrically neutral as the number of protons in its nucleus is equal to the number of
electrons in the orbits. Structures of some atoms are shown in figure 2.1. The
number of protons in an atom is known as its atomic number (Z). Every atom of a
particular element has definite number of protons eg., aluminium has 13, cobalt has
27, iridium has 77 protons.
A A - Mass number
Z Z- Atomic number
8
Fig. 2.2 Isotopes of Hydrogen.
77
Ir 191 77
Ir 192 77
Ir 193 are isotopes of iridium. Similarly,
27
Co58 27
Co59 27
Co60 are isotopes of cobalt.
Some isotopes are stable, some other are unstable. The stability of an isotope,
depends upon the relative number of neutrons and protons in its nucleus. An
unstable nucleus becomes stable by emitting radiations such as alpha particle, beta
particle. This process is known as radioactivity. Therefore unstable isotopes are
known as radioactive isotopes (or radioisotopes). In the above examples,
77
Ir 192 27
Co58 27
Co60 are radioisotopes.
9
Radioactivity exist in nature generally, among heavier elements. They are called
naturally occurring radioisotopes. Potassium-40, Carbon-14, Uranium-238 are some
of the naturally occurring radioisotopes. Many other elements can be made
radioactive by bombarding them with charged particles and neutrons. These are
called artificially made radiosotopes. Cobalt-60 Irridium-192, Cesium-137 are some
of the artificially made radioisotopes.
As the time elapses, a given radioisotope continues to disintegrate and hence, its
activity reduces.
The 'half-life' of a radioisotope is the duration in which its activity reduces to one
half of its initial value. It the original activity is A Ci, then the activity after one half
life will be A/2Ci. Thus, if T is the half life, after a time.
10
1 A
T, activity = A × = ;
2 2
1 1 A A
2T, activity = A × × = = ; and
2 2 22 4
1 1 1 1 1 A A
5T, activity = A × × × × × = =
2 2 2 2 2 25 32
The reduction in activity of radioisotope with time are shown in figure 2.4.
11
2.2 SOME IMPORTANT PROPERTIES OF ALPHA, BETA AND
GAMMA RAYS
226 222
Ra Rn
88 66
Beta radiations from a radioisotope have different energies from zero upto a
maximum value. The beta radiations are more penetrating than alpha radiations, but
beta radiations could be stopped by a few millimeters of aluminum.
If after the emission of a alpha or beta radiation, the nucleus has some extra energy
(excited state), gamma radiation is emitted. Gamma radiation is highly penetrating
and is similar to X-ray in its properties. Hence, gamma radiations are generally used
for industrial radiography. In some special circumstances, neutrons are also used.
Some properties of X and gamma rays are given below.
8. They can also cause ionization and excitation in the medium through which
they travel.
Relative absorption of alpha, beta and gamma radiations are shown in figure 2.5.
12
Fig. 2.5 Relative Absorption of different Types of Radiation
Visible light, X-rays and gamma rays are electromagnetic radiations. They are
similar in nature. They travel at the same speed (300,000 km/sec.) in air. They differ
in energy and origin of production. They are all called photons.
Table 2.1
RADIOISOTOPES COMMONLY USED IN INDUSTRIAL RADIOGRAPHY
ENERGY OF RADIATIONS
All radiations have energy. Energies of various radiations ( αβγ ) are usually
specified in electron volts.
One electron volt is the energy given to a single electron as it moves across a
potential difference of one volt.
1eV = 1 electron volt = 1.6 x 10-12 erg
13
1 keV = 1 kilo electron volt = 1.6 x 10-9 erg
1 MeV= 1 million electron volt = 1.6 x 10-6 erg
Data regarding the most commonly used radioisotopes in industrial radiography is
given in the Table 2.1
X-rays and gamma radiations are similar in nature, biological effects and action on
photographic film. Gamma rays are emitted by the nucleus of radioisotope, whereas,
X-rays are produced outside the nucleus when high energy electrons interact with
atoms (target). When these electrons interact with the field near the nucleus, they are
stopped or deflected with lesser energy. The difference in energy between the
incident electron and the deflected electron is given out in the form X-rays. The
energy of X-rays depends upon the closeness of the electron path to the target
nucleus. Hence, X-rays are produced with continuous energy, with a maximum equal
to the energy of the incident electrons. X-rays production increases with increase in
atomic number of the target atom. It also increases with increase in the incident
electron energy.
In an X-ray machine, the electrons, emitted by the heated filament, are made to
strike a tungsten target placed, at higher voltage, with respect to the filament. As a
result of interaction between the electrons and target atoms X-rays are produced.
14
Summary
2. Atom consists of NUCLEUS at the centre and electrons, revolving around the centre.
5. The number of protons in the nucleus of an atom is called its ATOMIC NUMBER
(Z).
6. The number of protons plus the number of neutrons in the nucleus of an atom called
it MASS NUMBER (A).
7. If the atomic number is constant and the mass number differs, then such atoms are
called ISOTOPES.
11. Some isotopes emit gamma radiations, after the emission of alpha of beta radiations.
13. ALPHA RADIATION consists of two protons and two neutrons. It is positively
charged. It is least penetrating.
15. GAMMA RADIATION does not carry any charge. It does not have any mass.
Hence, it is highly penetrating
16. HALF LIFE of radiations source is the period in which its activity reduces of half of
its initial activity.
15
2.4 INTERACTION OF RADIATION WITH MATTER
When X, B, ... radiations fall on a material, interaction take place between these
radiations and the electrons in the material. This leads to absorption and scattering
(moving in a different direction of radiation. A part of the incident radiation may
also get transmitted (Pass through).
Beta radiations can be completely absorbed in any medium. For example, beta
radiations with Emax (maximum energy) 1.7 MeV can be absorbed completely in 3
millimeters of aluminum. But in the case of X and gamma radiations complete
absorption is not possible. For example, one percent of cobalt-60 gamma radiations
still get transmitted through 8 centimetres of lead.
There are three main types of interaction for X and gamma radiations in any medium.
These are
a. Photoelectric effect
b. Compton effect
c. Pair production
But, for energies we normally come across in industrial radiography, only the first
two interactions are important.
In Compton effect, gamma radiation interacts with a free electron. The gamma
radiation gives a part of its energy to the electron and travels in different direction
(get scattered) with decreased energy. The electron moves in another direction. This
effect is predominant at medium energies and it is not dependent on the atomic
number of the material.
16
These two interactions are shown in figure 3.1
When and electron is removed from its orbit, the atom looses one negative charge.
In other words, the atom gets positively charged or ionized. Hence, this process
known as ionization.
When an electron is raised from an inner orbit to an outer orbit, the electron gets
extra energy. In other words, the atom gets excited. Hence, this process is called
excitation. Alpha, beta and gamma radiations can directly or indirectly cause
ionization and excitation in material with which they interact. Hence, these radiations
are ionizing radiations.
The intensity of radiation reduces while passing through any material. In case of X
and gamma radiations, the attenuation follows and exponential law :
I = I0e-µx
Where Io is the intensity of the incident radiation beam Io is the intensity of the
beam emerging after traversing a thickness x of the material and µ is known as
linear attenuation coefficient. It is defined as the fractional decrease in intensity per
unit thickness of the material. If the energy region of 100keV to 1.5 MeV is
considered, the attenuation of efficient decreases with increase in energy and it
increases with increase in atomic number of the absorber material. Hence, lower
energy radiations are better absorbed and higher atomic number material are go
absorbers.
17
Fig. 2.7 Reduction in Radiation Intensity with Thickness.
18
TABLE 2.2
HVT AND TVT VALUES FOR X AND GAMMA RADIATIONS
Material
Concrete Steel Lead Uranium
HVT TVT HVT TVT HVT TVT HVT TVT
Radioisotope cm cm cm cm cm cm cm cm
The thickness of the material, usually called absorber, (eg., steel, lead, depleted
uranium, etc.) needed to reduce the intensity of radiation to half its initial value is
known as the half value thickness (HVT)
The HVT depends on the energy of radiation and the nature of the absorber.
In the case of X-ray, the HVT value for any kV depends to some extent on the
amount of inherent filtration of the X-ray tube. However, the value of HVT for a
given material can be taken as constant for heavily filtered beam.
As higher atomic number materials reduces the intensity to half of the original
intensity. Two HVT reduces the intensity to 1/2x1/2 = (1/2)2, ie., 1/4 of the original
intensity. In general, 'n' HVT reduces the intensity to (1/2)n (Recall the decrease of
radioactivity with time).
The reduction in radiation intensity with absorbers of different half value thickness is
shown in figure 2.7
19
Tenth Value Thickness (TVT)
Tenth value thickness is the thickness of any material needed to reduce the radiation
intensity to one tenth of its initial value. Similarly, Two TVT reduces the intensity to
1/10 x 1/10 = (1/10)ie., 1/100 of the original intensity. Generally 3.3 HVT reduces
the intensity by a factor of 10, hence.
HVT and TVT values in different materials, for X and gamma radiation, are given in
Table 2.2
Example 2.1
The radiation level at a place, due to an iridium-192 source is 10 mR/h. What is the
thickness of steel required to reduce the level to 2.5 m/Rh?
Example 2.2
What is the thickness of lead required to reduce the radiation intensity at a place,
due to cobalt-60 source from 1000 mR/h. to 5 mR/h ?
Example 3.3
The radiation intensity at a place from a 200 kV X-ray beam is 5000 mR/h. This
has to be reduced to 2.5 mR/h. What is the thickness of concrete required to achieve
this ?
20
Summary
1. Beta radiations can be absorbed completely. But X and gamma radiations cannot be
absorbed completely.
9. Alpha, beta and gamma radiations can cause ionization or excitation. Hence, these
radiations are called IONIZING RADIATIONS.
10. Higher atomic number material, like lead, uranium, absorb X and gamma radiations
better than lower atomic number material, like plastic, aluminum
11. HALF VALUE THICKNESS (HVT) of any material reduces the radiation intensity
at a place of half of the original intensity.
12. TENTH VALUE THICKNESS (TVT) of any material reduces the radiation intensity
at a place to one tenth of the original intensity.
21
2.5 RADIATION QUANTITIES & UNITS
a. Radioactivity
b. Exposure
c. Dose
d. Equivalent Dose.
One becquerel corresponds to one disintegration per second. Formerly, the unit of
activity was Curie (Ci) corresponding to 3.7 x 1010 disintegration per second.
= 3.7 x 1010 Bq
= 37 GBq = 37 x 109 Bq
Submultiples of Curie viz., 1 mCi (1/1000Ci) and 1 U Ci (1/106 Ci) are occasionally
used.
EXPOSURE
The unit of exposure is coulomb per kg (C/kg), which is defined as the quantity of
charge produced by ionizing radiation in 1 kg or air.
The earlier unit of exposure was Roentgen (R) which was used for X or gamma rays
upto MeV. This was defined the amount of exposure due to X or gamma radiation
of energy upto 3 MeV which would cause ionization resulting 1 electrostatic unit of
charge of either sign in 1 cc of air at STP (Standard Temperature and Pressure).
22
DOSE
The dose from ionizing radiation is the energy imparted, to matter by the ionizing by
the ionizing radiation, per units mass of the irradiated material at the place of
interest.
If one joule of energy is imparted by any radiation in one kilogram of any material,
the dose is 1 gray.
The concept of dose applies to any medium, all types of radiations and all types of
radiations and all energies. The dose is a measure. If energy deposited by radiation
in the irradiated material. All effects of radiation - physical, chemical and biological
depend upon the dose.
Normally, the unit of dose was rad. One rad corresponds to 100 ergs/g of irradiated
material
Therefore,
Hence,
Dose rate interns of Gy/h, Gy/year, rad/year, etc. are also occasionally used.
In the case of exposure, for practical convenience, the concept of air kerma is used
to specify the radiation quality instead of exposure.
Kerma is the total kinetic energy of all the charged particles released in 1 kg of the
medium by the ionizing radiation.
Air kerma gives an idea about the energy transferred to air. The unit of air kerma is
Joules per kilogram (Gy)
The usefulness of air kerma can be extended to the determination of the radiation
output at a specified distance from a radiation source in terms of mGy per hour.
23
EQUIVALENT DOSE
The biological damage suffered by the tissue exposed to different radiations may be
different. That is, The biological damage due to 1 Gy of alpha dose would be
different from the damage due to 1 Gy of gamma dose. This is because the energy
loss per unit path length for different types of radiation is different. Alpha particles
because of their larger charge and mass, cause greater ionization per unit path length
than gamma radiations which mediate through singly charged electrons, Hence, One
Gy of alpha dose is approximately 20 time more effective in causing biological
damage, compared to one Gy of gamma of X-ray dose. To account for this variation
among different types of radiations, a terms known as the Radiation weighting
Factor (WR) is used to modify the doses due to each type of radiation. Thus,
Equivalent Dose is obtained by multiplying 'dose' by the radiation weighting factor
corresponding to the radiation of interest.
The unit of equivalent dose is also Joules Per kilogram. The special name for the
unit of equivalent dos is Sievert (Sv) Equivalent Dose in Sv = Dose in Gy X WR.
Formerly, the equivalent dose was called Dose Equivalent and its unit of was rem
and WR was called Quality Factor (QF)
1 Sv = 100 rem or
1 rem = 0.01 Sv.
Radiation weighting factor for X-rays being one, equivalent dose in Sv can be
considered as equivalent to dose in Gy. Therefore, for X and gamma rays,
1 Sv = 1 Gy
1 rem = 1 rad
KILOVOLTAGE (kV)
The potential difference applied between the cathode and the anode of an X-ray
machine is expressed in terms of kilovoltage (kV.) It determines the maximum
energy of X-rays produced by the equipment. For example, when the potential
difference applied is 300 kV, energy of X-rays produced varies from 0 to 300 keV.
Higher the applied potential difference, higher is the energy of the x-rays emitted
and higher is the penetration. In an X-rays equipment, the energy of X-rays can be
varied, depending upon the thickness of the object to be radiographed, by altering
the applied potential difference.
24
MILLIAMPERE (mA)
Milliampere (mA) of an X-ray unit indicates the current flowing through the filament
circuit. The intensity of X-rays at a given applied voltage is proportional to the
current flowing through the filament. The intensity of X-rays at 10 mA is twice the
intensity at 5 mA. The exposure to be given for a radiograph, at a given potential is
indicated in terms of mAs, ie., the product of filament current and duration of
exposure (sec). The exposure due to 10 mA current for 2 minutes is equal to that
due to 5 mA current for 4 minutes (1200 mAs.) machine.
CELL
Cells are the units of living organisms. All tissues are made of cells. Adult human
body consists of about 1014 cells. Cells of different organs carry out the functions
specific for those organs. For example, nerve cells are responsible for transmitting
electrical impulses from one part of the body to the other; liver cells are responsible
for metabolising the food we take and preparing nutrition for all other tissues in the
body; red blood cells carry oxygen and nutrition to all tissues.
Cells of different tissues have different sizes and shapes. However, in general, they
all contain a central nucleus (with the exception of red blood cell), surrounded by a
viscous fluid called cytoplasm. The nucleus contains chromosomes which constitute
an array of several million genes. The chromosomes control all the functions of the
cell and hence, of the tissue. The cytoplasm, contains various membrane systems and
other components that are necessary for the function of the cell.
Cell Division
Cells originate of multiply from preexisting cells by the process of cell division. For
example, a human being, like any other animal, develops from a single cell, which
is formed by the fusion of two germ cells, one from the father and the other from
the mother. During the nine-month period of pregnancy, this cell undergoes a
number off cell divisions. The new cell in turn specializes deform various organs of
the fetus (unborn baby). After birth, cell division continues until the organs attain
adult proportions. Cells in many tissues of the body have life span shorter than that
of the body, as a whole. For example, blood cells, skin cells, intestinal cells etc.,
have life span ranging from a few weeks to a few days. They wear out and die.
They are continuously replaced for the proper function of this tissue. Hence, even in
an adult, cell division continues in some tissues. Tissues undergoing cell division are
more sensitive to radiation than others.
25
From the point of view of radiation hazard, cells can be classified into two groups :
a) Somatic Cells, b) Germ Cells. Germ cells are the ones involved in reproductive
process. They are sperms in the male and eggs in the female. All other cells in the
body are somatic cells.
When radiation passes through body, it transfers some of its energy to the cells in the
form of ionizations and excitations, which in turn lead to a number of chemical changes.
Generally, these chemical changes are harmful to the cells. Depending upon the
seriousness of the harm, a cell either dies or gets modified. All the biological effects
of radiation arise from these two effects on cell - cells Death or Cell modification.
Radiation can cause breaks in chromosomes. Majority of these breaks get repaired,
but certain breaks may lead to loss or rearragement of genetic material which can be
seen under a microscope. Such events are called chromosome aberrations.
Biological effects of radiation be classified into two groups, depending upon the type
of cells damaged. They are somatic effects and hereditary effects. As the name itself
suggests, somatic effects arise from damage to somatic cells and they occur in the
tissues of the exposed person (eg., effects on skin, blood, lung bone, thyroid, etc.)
Hereditary effects arise from damage caused to the germ cells and occur in the
progeny of the irradiated person.
Somatic effects can be further divided into early and late effects.
Whenever the body is exposes to radiation, some cells are killed and some are
modified. At low doses, since only a few cells are killed, the body can cope up with
the loss and hence no immediate effects are seen. However, as dose increases, more
and more cells get killed and this results in radiation sickness, nausea (feeling like
vomiting), vomiting, fatigue, etc., within a few hours or irradiation. However, these
effects are transient and disappear after a few hours. As dose increase further, the
effects appear quicker, are more server and last longer. At sufficiently high doses,
the damage caused to important organs such as blood forming organs intenstine, etc.,
is so severe, the patient will not be able to recover. Anemia, infection and high fever
would occur, leading to death. Fifty percent of the persons exposed to a dose in
range of 3-5 Gy, die within a period of 60 days. This is known as LD50/60.At still
higher doses, death occurs earlier.
Instead of the whole body, if only a part of the body exposed the, damage the confined
to the exposed part. Depending upon the part of the body exposed, different local effects
at different doses are produced Skin is the most frequently exposed organ. Doses less
26
than 5 Gy can cause only a transient (temporary) erythema which lasts for 1-2 days.
Higher doses of the order of 10 Gy can cause depilation, fixed crythema and very high
doses cause dry or wet desquamation depending upon the severity of loss of epidermal
basal cells, within 3-6 weeks (10-25 Gy.) Doses higher than 25 Gy leads to leads to late
phase of erythema during 8-16 weeks, finally leading to necrosis. When large areas of
the skin are exposed as in the case of beta radiation burns, death can occur in a manner
similar to that by thermal burns. For example, if skin is exposed as in the case of beta
radiation burns, death 6 Gy of Z or gamma radiation, reddening of the skin takes place
within a few days. This is known as skin in addition to erythema, hair falls off, leading
to epilation. Another sensitive organ is the reproductive organs-testes in male and ovaries
in female. Due to death of the germ cells sterilization results. This may be temporary or
permanent depending upon the dose. When eyes are exposed, the cells in the lens are
killed leading to the opacity of the lens, which is called cataract. Unlike the other effects
which appear within a few days of weeks after exposure, cataract formation takes a
mean latent period of 2 to 3 years.
In brief, radiation exposure may lead to various types of early somatic effects
depending upon whole or only part of the body is exposed. The characteristic of the
early somatic effects are :
Table 5.1 shows some of the early somatic effects and their threshold doses. It is
generally assumed that the exposure takes place within a short time - minutes to
hours (acute exposure). If the doses are protracter (chronic exposure), then the
threshold doses will be much higher the effectiveness of radiation decrease due to
the recovery process.
The threshold doses for early effects are generally, much higher than doses received
by persons during normal working conditions. Hence, when proper working
conditions are practiced these early effects of radiation do not occur. However,
accidents such as holding sources by bare hands, hiding stolen sources in pockets,
etc. leading to serve skin burn and tissue damage have been reported.
The most important late somatic effect is cancer. When the irradiated cell is modified
rather than killed it may develop into cancer, after a prolonged delay. The delay may
very from 5 years (blood cancer) to 30 years or more (lung cancer). Unlike the case of
early effects, the probability of cancer resulting from radiation increases with every
increment of dose, probably without any threshold. Radiation is not the only agent which
27
induces cancer. A number of chemical agents (such as tobacco and its fumes, as in beedi
and cigarette smoke and biological processes can also induce cancer. Compared to these,
radiation is a week carcinogen. However cancers induced by radiation are
indistinguishable from those induced by other agents.
HEREDITY EFFECTS
Hereditary effects may result when the irradiated germ cell is modified rather than killed
and if it also participates in the reproductive process. under such circumstances, the
damage caused to the genetic material in the modified germ cell will be transmitted to
the subsequent generations. As in the case of cancer, there is probably no threshold
dose for hereditary effects. Even though radiation is found to induce hereditary effects
in experimental animals, there is no conclusive evidence of the same in man.
Furthermore, nature incidence of hereditary defects in man is quite high. Mulations in
germ cells could lead to a variety of skeletal abnormalities leading to malformations,
neonatal cataract and a variety of genetic diseases associated with mental retardation
(eg., Downs Syndrome). Fetus and children are generally known to be more sensitive
than adults to all effects of radiation In normal radiation work, it is necessary to ensure
that the risk of radiation induced cancer and hereditary defects are kept at acceptable
limits.
Every profession has its own risks and benefits. While radiation can be applied to a
number of beneficial purpose, it can produce early and late harmful effects in the
exposed persons. Hence, it is necessary to set up safety standards in the use of
radiation. The international Commission Radiological Protection (ICRP) sets
guidelines in this respects in the form of annual dose limits of individuals. Since the
early effects have dose thresholds, they can be easily prevented by setting the dose
limits below the threshold. On the other hand late effects such as cancer and
hereditary effects have no threshold, which means that there is no dose below which
they can be completely prevented. Hence, in setting up the dose limits, it should be
ensured the risk of cancer and hereditary defects do not exceed certain acceptable
limits.
28
TABLE 2.3
DOSE LIMITS
Dose limit
Application Occupational Public
Accordingly, the ICRP has suggested the annual occupational dose limits to radiation
workers (including pregnant women workers), as shown in Table 2.3, for whole
body exposure, as well as for some specific organs. The average effective dose
(whole body) should not exceed 20 millisievert per year limited to 30 milliservert for
a particular year. Since, it is desirable to keep the risks well below the acceptable
limits, the ICRP recommends that the dose limits should be considered as the upper
limits and the actual doses received must be kept As Low As Reasonably
Achievable (ALARA). In addition to workers who handle the radiation sources,
member of the public may also get exposed to radiation. Hence, separate limits for
general public have also been suggested. These are also shown in Table 2.3. The
limits for general public as a group is lower than that for workers, since it a)
consists of more sensitive section of population (children), b) does not derive the
direct benefits from the use of radiation and c) no individual monitoring for
assessing exposure is done. The ALARA principle is also applicable in the
protection of public. The dose limits given in Table 2.3 do not include medical
exposure and natural background radiation.
Radiation has always been a part of our daily life. We are constantly exposed to
cosmic radiation, the earth's minerals (uranium, radium) and even from those
radioisotopes within our own body (Potassium-40, Carbon-14). As a result, the food
we eat, the air we breathe, the water we drink and the house we live in, contain
traces of radioactivity. These constitute Natural Background Radioactivity. The
average annual dose resulting from natural radioactivity is estimated to be about 2-
3mSv.
The dose limits for radiation workers is about 10 times higher than the dose due to
normal natural background radiation. But, the average occupational dose is of the
same order as the natural background. This corresponds to about 100 mSv over ones
29
lifetime. If 1000 persons receive 100 mSv over their lifetime, 4 persons may suffer
from cancer.
It should be noted that care, rather than fear, is needed while handling radiation
sources and the associated risk should be viewed in relation with other risks. The
procedures available to control exposures to ionizing radiations are sufficient, if sed
properly, to make sure that the risk due to radiation is small compared to many
other risks to which we are all exposed.
30
Summary
1. Any profession has some risk associated with it.
3. In radiation work we are concerned with both acute exposures and chronic exposure and
the associate hazard.
6. While setting operational limits, the radiation worker and his future generation are all
considered.
7. There are separate operational limits for the whole body and the individual organs
(see Table 6.1).
8. These limits do not include medical exposure and exposures due to natural
background radiation.
10. The annual operational limit for whole body is 20 mSv (2rem), for radiation workers
and 1 mSv (100 mrem) for non-radiation workers.
31
CHAPTER - 3
X-RAY TECHNOLOGY
3.1 INTRODUCTION
X-rays were discovered by a German scientist, Prof. Wilhelm Conrad Roentgen in 1895.
Some of the properties of X-rays are given below.
1. They are electromagnetic radiations, similar to visible light, but they have higher
energy.
2. They can pass through matter and get absorbed/scattered in the process.
4. They can excite and ionize atoms of the medium, through which they pass.
X and gamma rays have similar properties. Gamma rays are emitted by the nucleus,
whereas X-rays are generated outside the nucleus when high speed electrons interact
with atoms. Gamma rays have definite, discrete energies, whereas, X-rays have
continuous energies. The maximum energy of X-rays depends on the incident
electron energy (Fig. 3.1)
a. Advantages
2. They have small focal spot size, which gave sharper images.
3. The use of X-rays results in better image contrast, due to its continuous
spectrum
4. X-ray units ensure complete radiation safety, when the unit is switched
"OFF".
32
b. Disadvantages
= 200
( )
X-rays are produced when a beam of high energy electrons collides with any
material.
X-rays production increases with increase in atomic number of the target atom and
with increases the incident electron energy. In an X-ray machine, if the potential
difference between the filament and the target or the applied kilovolts is 200kV, then
the energy of the electrons hitting the target is 200 keV and the maximum energy of
the X-ray would be 200keV. A typical X-ray spectrum is shown in Fig 3.1. The
continuous X-rays spectrum will also contain one or more sharp peaks. These peaks
are of definite energies, dependent on the target element, .(hence are called
characteristic X-rays).
The quality of and X-ray beam can be described by its Half Value Thickness (HVT.)
The HVT is a function of the effective energy of the X-ray beam, which is
approximately 1/3rd of the applied kilovoltage. It also depends upon the nature of
the power supply and the added filtration.
33
The essential requirements for the production of X-rays are
c. a target, usually tungsten, to stop the electrons and to convert them to X-rays.
The target is usually of small dimension, say 2-3 mm. When the electrons hit the
target, much of the energy appears in the form of heat and it has to be rapidly
removed. Copper, is used for the purpose of heat removal. Certain anodes are hollow
in construction, so that primary coolants can be circulated through the same to
remove the generated heat. Mineral oil is also sometimes used around the X-ray unit,
to remove heat and serve as electrical insulator.
The penetration of X-ray beam depends on the applied kilovoltage, whereas, the
intensity is decided by the current flowing through the filament (Millamperage).
3. It should possess high thermal conductivity (to dissipate the heat quickly).
Tungsten, having and atomic number 74, and melting point 3400oC is the most
preferred target material.
34
X-rays tube is contained in a suitably shaped steel for ruggedness. The power
ratings, viz., kilovoltage (kVp), tube current (mA), besides the cooling pattern,
decides the structure of an X-ray unit.
3. It should possess the smallest possible focal area. For maximum radiographic
definition, a point source of radiation is one of the requirements. Modern X-
ray tubes have exceptionally small focal areas.
4. The design safety should include sufficient shielding material (say lead or
equivalent steel) so that the leakage radiation level at every rating
combination (kV, mA) does not exceed 1 R/h at 1 meter from the target.
X-ray units can also be used as cabinet installations with incorporated lead shielding
and safety interlocks (eg., the unit becomes operable, by actuation of certain
microswitches, only when the object occupies a preset position in front of the beam
port).
3.5.1 Fluoroscopy
35
3.5.2 FINE FOCUS TUBE
The examination of confined spaces, like the pipes of a steam boiler or the cylinder
heads of an internal combustion engine, has given rise to a tube with the target at
the end of a long tube. The target and therefore, the whole anode is earthed, so that
the source of radiation can be pushed into the cavities mentioned above. In X-ray
units used for circumferential radiography, the target is placed at right angle to the
tube axis and as a result, the radiation emerges all round in the form of a disc. For
unidirectional beam, the target is at 45o inclination (Fig. 3.3).
- -
Fig. 3.3 Rod Anode System
These units are useful for cross-country pipe line inspection, with automatic
movement from joint to joint and are becoming increasingly popular. The power
input is obtained from diesel generators.
The power supply required for the operation of an X-ray tube are
The filament of an X-ray tube is normally operated at 6-12 voles with 5-6 amps of
current. This is derived from the mains line using a step down transformer. The high
voltage is usually obtained from a step-up transformer.
To maintain the target at positive potential with respect to the filament, different
types of rectification circuits are used, viz., half-wave rectification, full-wave
rectification and constant potential units. There are various advantages of using a
36
constant potential X-ray unit. It yields better X-ray output than that produced by a
pulsating potential having the same peak kilovoltages. It gives a more penetrating
beam, as required in industrial radiography.
To obtain high energy X-rays in the MeV range, for inspection of very thick objects,
linear accelerators are used. In these, the X-ray output can be of the order of few
hundred R/minute.
37
CHAPTER - 4
RADIATION SOURCES
4.1 INTRODUCTION
b. processing the fission products from spent Uranium fuel rods from a nuclear
reactor,.
When a target element is bombarded with neutrons in a reactor, activation may occur
mainly by one of the following processes depending on the energy of the neutrons.
a. (n....) reaction eg 27
Co59 (n,c) 27
Co60
c. (n,a) reaction 13
Al27 (n,a) 11
Na24
38
4.2.2 Production by nuclear fission
4.2.3 Particle accelerators are also used for production of some radioisotopes.
Board of Radiation and Isotope Technology (Brit), Bombay fabricates and supplies
cobalt 60 and iridium-192 radiography sources for use in a variety of equipment and
in a range of activity values. Although there are differences in shapes and sizes of
source assemblies, the inner source capsule, loaded in these, are similar in shape and
dimensions. It is a stainless steel capsule with dimensions : 8mm diameter. Each
iridium-192 pellet is of 2.5mm diameter and 0.3 mm thickness. The number of
pellets in a capsule depends on the activity required. A 1000 GB capsule may have
four or five such pellets.
39
TABLE 4.1
Months->00 02 04 06 08 10
Years
TABLE 4.2
DECAY CHART FOR IRIDIUM-192
Days 0 5 10 15 20 25 30 35 40 45
050 0.63 0.60 0.57 0.54 0.52 0.50 0.47 0.45 0.43 0.41
100 0.39 0.37 0.36 0.34 0.32 0.31 0.30 0.28 0.27 0.26
150 0.25 0.23 0.22 0.21 0.20 0.19 0.19 0.18 0.17 0.16
200 0.15 0.15 0.14 0.13 0.13 0.12 0.12 0.11 0.11 0.10
40
CHAPTER - 5
1. A source housing, which serves as a shielded container for the gamma source
during its storage and it has an arrangement to give a controlled radiation
beam, when needed.
1. Optimum shielding,
With these features in mind, various shapes of radiography cameras have been
designed (fig.2.4) Size, shape and weight of a radiography camera depend upon the
energy and activity of gamma source and also on the shielding materials used. In
41
general, the cameras can be broadly classified on the basis of their weight.
Trolley and manually handled units can be moved on the site for exposures.
Classifications given in brackets are as per ISO 3999
a) Source held in the housing and the shutter is displaced to expose the source
(Fig.5.1a). It can be used only in portable cameras because of practical
limitations on size and weight of the shutter.
b) Source held on a cylindrical drum which is rotated to expose the source (Fig.
5.1b). It provide wide radiation beam.
d) Source assembly attached to the tip of a flexible cable and driven out of the
source housing for panoramic exposure (fig. 5.1d)
5.3 DESIGN
42
1. Type of source assembly,
2. Shielding material,
5. Feasibility of fabrication,
7. Installation and
8. Transportation.
5.3.1 Source Assembly : The radiation source in a radiography camera is always housed
in a secondary container called 'source assembly'. The source assembly can be of
two types, right or flexible. Rigid source holder is preferred for source housing
shown in fig.5.1a to 5.1c. Flexible source assembly is used, when the source is
required to travel distance to and from the source housing, through bends and at
different planes.
43
TABLE 5.1
5.3.3 Source exposure : Directional exposures with collimated beam it source fixed in the
source housing are best from the radiation safety point. However, such exposures
greatly restrict their flexibility of operation of operation. Therefore, most of the
modern units are designed for panoramic exposure with provisions to attach
collimators for directional exposures.
Panoramic exposures are made by moving the source assembly out of the source
housing, using remote driving devices, operated by pneumatic, electrical or
mechanical systems. With flexible source assembly, only mechanical systems using
teleflex cable are favoured.
5.3.4 Safety Devices : All radiography equipment should be provided with safety devices
such as interlocks, source couplings source holder locking, immobilizing device for
the source, etc.
In-house facilities are planned with more elaborate safety devices, such as door
interlocks, audio-visual signals, search operations in exposure areas, and other full-
proof safety features.
44
In addition to the above, radiation monitoring instruments should also be
incorporated in the design
5.3.5 Fabrication : Design of the unit should be done keeping in mind the limitations
during the fabrication of components. These components are evaluated for soundness
and reliability to ensure fool proof performance over the designed life of the
equipment.
5.3.6 Maintenance : The equipment should be designed in such a way that minimum
maintenance required during service and defective part is easily replaceable.
5.3.7 Transportation : The design of the camera, when loaded with source should meet
the prescribe transport regulations.
5.3.8 Installation : Installation of units with kilocurie activity needs special gadgets. These
gadgets should be designed in conjuction with the unit to be installed.
TABLE 5.2
45
Prototype units with remote control system are specially tested for operational
reliability under simulated conditions to assess the life of components and associated
accessories.
As is known, iridium-192 and cobalt-60 source together can cover and inspection
rage of thickness 10-200mm, steel equivalent. Thus, efforts for equipment
development have been mainly directed towards these sources. Multipurpose
standardised equipment are commercially available, with source housings, shielded
either with lead, heavy alloy or depleted uranium. Main features of these equipment
are discussed here briefly.
a Iridium-192 Units : ROLI-1 is a remote operated lead shielded camera designed for
35 Ci iridium-192 source (Fig. 5.3) It has 'S' conduit to house a flexible source
assembly.
46
Fig. 5.4 CRC-2 Radiography Camera
c Light Weight High Activity Equipment : These have depleted uranium or heavy
alloy material for shielding. Fabrication of these source housings is not as easy as
lead source housings. The cost of the material is also very high.
47
Most of the imported standerdised radiography equipment available commercially are
depleted uranium shielded ones. Some of these are shown in figures 5.5 to 5.8
The built-in safety of gamma radiography equipment, both from radiation and
operational safety point of view, is very essential because majority of the
radiography work is carried out in workshop areas and construction sites which call
for a strong and rugged equipment.
48
immersion test and endurance test, as per the specification laid down by International
Standards organisation (ISO 3999). At present, only those equipment which conform
to all the provisions of ISO 3999 are type approved and permitted to be used for
industrial radiography work.
49
CHAPTER - 6
6.1 INTRODUCTION
The type of X-rays film used in radiography, an important role in the detection of
flaws. It is therefore, important to have detailed knowledge about construction and
properties of films.
6.2.2 Emulsion
The protective layer consists only gelatin. It serves to protect the lower layers from
physical damage, abrasion and stress marks.
50
6.3 CHARACTERISTICS OF FILMS
Io
D = Log10
It
a. type of film
b. energy of radiation
d. processing conditions
If number of different exposures are given to various of an X-ray film and densities
obtained these areas after processing are plotted as a function of Log10 (exposure),
the resulting curve is called 'Characteristic Curve' of the film (Fig. 6.2). A
characteristic curve given information on :
51
a. speed of film and
b. film contrast
a Speed
Speed is defined as the density recorded on a film due to a given radiation input. It
is measured in terms of inverse of exposure required to produce a radiograph of a
particular density, under given conditions. Keeping other factors constant, a film
which requires less exposure is faster. Figure 6.2 shows characteristics of two
industrial X-ray films in which A is faster than film B.
TABLE 6.1
NDT-55 3.8 R
NDT-65 1.4 R
NDT-70 1.0 R
Cobalt-60 Agfa D-2 19.6 R
Agfa D-4 8.0 R
Agfa D-7 2.5 R
NDT-55 7.5 R
NDT-65 3.0 R
NDT-70 2.0 R
b Contrast
Film contrast of gradient is defined as the change in density recorded on a film for
a given change in radiation intensity.
52
The gradient of a film depends on the size of its crystals. It is practically
independent of the radiation energy. Gradient of industrial X-ray films goes on
increasing with optical density, as shown in figure 6.3. In general, fine grain films
have lower speeds, but, higher gradient at a given density, than coarse grain films.
There is very little effect on the shape of the characteristic curve due to energy of
radiation, However the film speed depends on radiation energy, can be seen in table
6.1
1. The Films for use with salt screens, also known as salt screen films.
2. Films for use with metal screens or without screens (also called 'direct
films'). This group covers a large range of industrial X-ray films. The films
are classified depending upon the grain size of the films ;
53
CHAPTER - 7
7.1 INTRODUCTION
When an on object is radiographed without using screens, there are two disturbing
factors.
2. Low energy scattered radiation is absorbed more, which spoils the image
quality.
There are two types of radiography screen, commonly used in industrial radiography
practice : metallic foil screens and fluorescent screen.
These screens differ in their basic characteristics e.g., speed, contrast and elimination
of scattered radiation.
Metallic screens help in reduction of scattered radiation from the object. The X-ray
film is sandwiched between a pair of lead screens, as shown in figure 7.1.
Fig. 7.1 X-ray Films Sandwiched between pair of lead intensifying screens.
54
7.3.1 Reduction of scattered radiation
In this definition, it is assumed that same film and radiation source are used for both
the exposures. 'IF' due to double lead screen exposure with iridium-192 source is
about 4.5.
a. foil element,
b. thickness of foil
d. specimen thickness.
These are also called 'Salt Screens'. These screens consist of a card board of plastic
material on which certain inorganic crystalline substances, which glow when X or
55
gamma rays fall on them, are coated on one side. These screens are used in pairs, so
that a double coated X-ray film can be sand-witched between them as shown in
figure 7.2
During X or gamma ray exposure, each crystal in salt screen emits light and this
light diverges in all directions and gives image unsharpness
a. density of radiograph
b. thickness of screen
d. energy of radiation
In gamma radiography, the intensification factors with salt screens, are not very
much greater than that obtained with lead screens. Secondly, these screens give poor
quality image. Hence, these screens are not much used in industrial radiography.
56
CHAPTER - 8
8.1 INTRODUCTION
In radiography, the work parameters and conditions play an equally important role as
the applied radiography techniques. The handling of X-ray films before and after
exposure and film processing are important in this respect.
X-ray films should be handled carefully to avoid physical strains such as pressure
creasing, buckling and friction on it. The normal pressure applied on a cassette to
provide good contact with the object does not damage the films. Whenever the films
are loaded in flexible cassettes and external clamping devices are used, it should be
seen that pressure is applied uniformly. Localised pressure on film can produce
densitised areas.
Films should be held by edges in order to prevent finger or pressure marks. Also the
precaution should be taken to avoid handling films in a manner which would cause
friction and give confusing black marks on the radiograph.
The expiry date on a film pack means that the film should be used before the given
date. Expiry dates of the films depend on the storage conditions.
X-ray film should be protected from X-rays and gamma rays to avoid fogging
Film should not be kept in a chemical storage room or in any location, where there
is leakage of illuminating gas. Chemical vapours, high humidity and temperatures
also contribute to film fogging. Film should be stored in as cool a place as possible,
and in tropical regions, temperatures control in the store room is recommended. Film
may be stored for long periods in a refrigerator.
57
8.4 FILM PROCESSING
After exposure to the radiation, a latent image of the object is formed in the X-ray
film. The purpose of the film processing is to convert this invisible image to visible
and permanent image. There are four main stages involved in film processing
Ag+ + e- Ag
2. Liquid concentrate, which require appropriate dilution with water to give the
working solution.
Commercial Developers
Some of the commercially available developers for manual processing of X-ray films
are given in Table 8.1
58
TABLE 8.1
After developing, the image comes into existence and remaining processes make it
permanent and enable to view it conveniently. If no action is taken, the developer
will continue to work producing enhanced density, fog or irregular developer stains.
The unwanted developer with stop bath. There are two types of stop baths generally
employed.
8.7.1 Water Rinse : The excess unwanted developer is removed from the film simply by
immersing it in a running water bath. Static water rinse involves the risk of solution
becoming alkaline with use and cease to be of any use in stopping development.
8.7.2 Acetic Acid Rinse : This bath is static and 2 to 3 percent solution of acetic acid is
employed. This is more efficient in action and required films to be immersed in it
for about 5 seconds with agitation. However, unlike flow water rinse, acetic acid will
become continuously exhausted with use. A throughput of 100 FT2/gallon is
reasonable suggested life for this.
59
8.7.3 Faults arising during rinsing : The only fault which can occur at this stage is due
to delay in putting the developer laden film in stop bath. The oxidation of thin layer
of developer on film can lead to production of brown developer stains.
The fixer reacts with undeveloped silver bromide in the film and transforms it into
water soluble compound which gets dissolved in washing operation. In addition, it
enables the image to become permanent. Since the silver halide is still sensitive to
light to a small extent, and so will slowly darken if it is allowed to remain on the
film.
8.8.1 Fixing Agent : The most widely used agent is sodium thiosulphate, commonly
known as 'hypo'. Its solution in water rapidly dissolves and bromide.
8.8.2 Use of Fixing Bath : Generally, the film is kept for 2-3 minutes in rapid fixing bath
and about 10 minutes in ordinary hypo bath is found sufficient.
2. Temperature : The fixation process is less critical than development process and a
temperature range of ±2.50C to standard 200C is acceptable.
8.8.3 Faults arising due to fixation : Most of the defects arise due to introduction of
developer into fixer through careless use of stop bath stage. Faults may also arise if
films are inserted without agitation. Streaks may be produced, due to uneven
removal of developer.
Washing and drying is carried out to eliminate unwanted products from the
emulation after the fixing operation. The excess water is then removed, since a wet
radiograph is not convenient to handle and easily liable to mechanical damage.
8.9.1 Washing : In the washing stage, various silver compounds in the gelatine, and
excess fixer are removed. For this purpose, film is washed in running water for
about 20 minutes and then taken for drying. It is important to use flowing water, so
that on its removal from bath is taken out from uncontaminated water.
Faults arising during washing : Usually twenty minutes are allotted for this stage,
but the period may prove insufficient if,
60
3. tank is too small for film through out and
4. film is inserted in a haphazard manner.
8.9.2 Drying : Usually the wet films are dried by blowing a current of hot air over the
films, and various types of drying cabinets exist for this purpose. These cabinets
after include filter at their air intake to free the from dust, which may stick to the
wet film.
Black streaks or blotches Light leaks due to faulty film holder or cassette.
Drying spots and Drops of water on semi dried film or drops of water
running streaks down semi-dried surface.
61
Brown stains Inadequate fixing or exhausted fixing solution,
prolonged development in old developer, inadequate
rinsing.
deposits
(light spots/patches)
crack-like line)
White spots and areas Pitted or worn screens. Dirt on film screens.
62
8.11 GEOMETRY OF IMAGE FORMATION
The radiographic image on the film is governed mainly by basic principles of the
shadow formation. A sharp and undistorted image is obtained by controlling various
geometrical factors such as source size, source to film distance, film to object
distance, alignment of radiation beam with respect to the plane of the object and
film. Different exposure geometries are shown in Fig. 4.1
From these exposure geometries, it could be seen that the radiographic images
always have an element of magnification which will depend upon the closeness of
the film with the specimen. Distortion in the image would depend on the relative
planes of the object and the film. Attempts are made to keep both of them in the
same plane, however, with specimen of odd geometry image distortions are
observed.
Both X and gama rays produce a certain unsharpness on the radiograph due to the
finite size of the focal spot or the source. This unsharpness is called geometrical
unsharpness (Ug) which is calculated as
dxt
Ug = or SFD = t ( I + d / U g )
SFD − t
It is clear from this expression, that Ug can never be zero. The magnitude of Ug can
only be minimum by proper selection of exposure variables. Considering this fact,
various codes permit a certain Ug value related to object thickness.
1. Source of focal spot size should be as small as practicable. Ideal source will
be a point source,
4. Source location should be such, that radiation pass normal through object
thickness.
63
Fig. 8.1 Geometric Representation of Radiographic Image.
64
X-RAYS
Film : Structurix
FFD : 70 cm
Screen Thickness : 0.025
mm
Density : 2
65
8.12 EXPOSURE TIME CALCULATIONS
For calculating the exposure time, the following steps should be followed:
1. Calculate the present source activity in curies (Ci) from the decay charts.
3. Find HVT (cm) of object material, for the given source, from the HVT tables.
TABLE 8.2
F x 2 x / T 12 x ( SFD ) 2 x 60
Exposure time (minutes) =
C x RHM x (100) 2
It may be noted that this formula is not very accurate for higher thicknesses,
particularly when using Cobalt - 60. Exposure charts time, with the following date :
66
. x 2 25/125 x 50 −2 60
14
Exposure Time (minutes) =
10 x 0.5 x 1002
14
. x 4 x 2500 x 60
= 50 x 10000
= 16.8 minutes
From exposure chart (Fig. 4.2) we can see that for 25 mm. steel and SFD 50 cm.,
we need around 2.8 curie x hr. It means the product of source activity and time in
hours should be 2.8.
= 17 minutes
67
CHAPTER - 9
9.1 INTRODUCTION
Film contrast (G) refers to the slope of characteristic curve of the film at a
given density. It depends upon the type of film, processing conditions, and
optical density, It is independent of radiation quality. Mathematically, it can
be expressed as :
D
G=
LogE
68
1. The material of construction should, preferably, be the same as that being
inspected.
Table 9.1
Various types of designs of IQI based on wires, holes, slits, etc. have been used.
Some designs are shown in figure 9.1
69
a Plaque/Hole (ASTM) Type This type of IQI is a constant thickness plate
having three drilled holes of different diameters. If the plate thickness is T,
the hole diameters are T, 2T and 4T with minimum hole diameters are 0.010,
0.020 and 0.040 inches respectively. The plaque must be of the same material
as specimen. The same penetrameters are accepted by ASME Boiler and
Pressure Vessel Code.
Identification numbers and notches, specifying the material and IQI thickness
are fixed on each plate. The number indicates thickness in thousands of an
inch. Three quality levels can be determined with a given ASTM IQI, as
given in table 9.2
Table 9.2
1-1T 1 1T 0.7
1-2T 1 2T 1
1-4T 1 4T 1.4
2-1T 2 1T 1.4
2.2T 2 2T 2.0
2-4T 2 4T 2.8
4-1T 4 1T 2.8
4-2T 4 2T 4
4-4T 4 4T 5.6
Where,
T = thickness of penetrameter.
70
SE is defined as the thickness of penetrameter expressed in percentage in which 2T
hole would be visible under the same condition.
b Wire Type This type of IQI consists of a series of wires of minimum length 25 mm
mounted side by side parallel with a distance between the axis of wires of not less
than three times the wire diameter and not less than 5 mm and arranged in order of
increasing diameter.
Wire type of IQIs have been adopted internationally. Most commonly used are DIN
type and ASTM type IQIs.
Sensitivity =
n
Q
2T
71
9.3.2 Placement of IQI
As a general practice, IQI must be placed on the source side of the specimen. Film
side IQI are also permitted in some codes only in special circumstances.
In case of plaques type IQI, it is generally placed parallel to the weld at a location
of poorest sensitivity for a given technique. Wire IQI is placed across the weld (Fig
9.2). When the weld with the reinforcement is examined, a shim of metal should be
placed under the IQI the match the total thickness
When the objects are too small or have a complicated structure for IQI placement,
IQI can be placed on a uniform block of the same material
9.4 IDENTIFICATIONS
Every radiograph must have permanent identification marks which identify the job,
the region examined and the orientation of the film. Except on very thick steel
sections, thin lead letters and characters can be used for placement on the specimen
or film to be visible on the radiograph. They should be fixed on the specimen when
it is casting or weld and left in position whenever used.
However, most of the specimen will have permanent marking either by stamping,
engraving or painting according to what is most appropriate for the service condition
of the specimen.
As given in table 9.1, there are many factors which contribute to the resultant quality
of a radiograph. There factors should be carefully controlled by following a well
recognised national/international practice usually specified in codes. Information on
the following points is essential.
72
9.5.1 Technique Classification and general requirements such as very high sensitivity,
high sensitivity and medium sensitivity are followed as per code.
9.5.2 Material and its Thickness data is essential to select the source, SFD and other
important radiography parameters. Whenever possible, fabrication procedure such as
welding, casting, etc. should made available.
9.5.3 Film and Screen used are checked up with the recommendations of the code with
respect to the technique classification.
9.5.4 Film Processing should be standard as per the recommendations of manufacturer of the
respect to the chemicals. Shortened and prolonged development time must be avoided.
9.5.5 Density of the Radiograph should be between 2-3. Higher may be employed
provided high intensity illuminator is available for viewing the radiographs. Fog
density and expiry date of the film should be checked. Fog density should not
exceed 0.2.
9.5.6 Identification Marking System has to be explicit to enable finding the exact area
of the specimen without any ambiguity. Lead markers impression must appear on the
radiograph.
9.5.7 Source, its active dimensions and calibration data should be verified with
recommendations of the code.
9.5.8 SFD is directly linked with specified Ug value It also determines the diagnostic
area coverage of the specimen.
9.5.9 Exposure geometry with relative position of source, specimen and films shown be
given in a sketch for verification.
9.5.10 IQI, its location bears the most important information leading to assessment of the
sensitivity of a radiograph. Observed sensitivity is calculated in terms of thickness of
plaque/hole or diameter of the wire discernible by eye.
Besides the above points, the radiograph is carefully examined on both sides with
reflected light to reveal superficial marks such as scratches, crimps and other
processing marks, Use of a magnifying glass would be helpful. Familiarity with film
artifacts and their indication is also essential to determine the genuine discontinuity
as different from the spurious indication on the film.
73
9.6.1 High Density
It can be due to excessive subject contrast or high film contrast. High subject
contrast can be rectified by increasing the kV, use of a filter at the tube or use of
thickness compensation methods.
Low contrast film can also be used. Check also for wrongly mixed developer, as
well as for prolonged development times to compensate for under-exposed film.
It can be due to low subject and film contrast and under-development. It can be
corrected by decreasing energy of radiation, use of higher contrast film and
processing the film under stand
Possible causes can be geometrical exposure factors, poor contact between film and
screen, graininess of screens and graininess of film. The geometrical factors should
be verified and readjusted to meet the specified Ug value. Film and screen graininess
can be controlled by using and improving the contact between the film and the
screen.
74
9.6.6 Fog
It can result from excessive exposure to light (including safe light), inadequate film
storage facing or improper film processing.
Film loading and processing area should be examined for leakage of light and
brightness of safe light. The filter on safe light should also be examined for fading
of coating dyes due to use of high wattage bulb.
Faulty film processing and careless handling of films - before, during and after the
exposure - would result in various types of spurious indications on the film.
Conventional black and white radiographs are viewed against light source. The equipment
providing the light is called as 'illuminator'. Various types of illuminators are commercially
available However, the most suitable illuminator is the one which has high intensity
light source, sufficient penetrate the film densities upto 4, with provision as for intensity
control and mechanical arrangement for masking the lighted area.
Viewing the radiographs on the illuminator is generally carried out in a room fitted
with dim light free from glare, dazzle and reflection. This helps in quick dark
adaptation of eyes, reduces eye fatigue and improves visuals acuity.
9.7.2 Illuminators
75
1. Colour of light shall be normally white, but colours between orange and pale
green are acceptable.
2. Diffusion of light through one sheet of thin opal glass is found to be adequate.
a Ambient light : The radiograph is examined in a darkened room. Care should taken
so that as little is reflected as possible from the film surface, directly towards the
observer.
Generally, the light transmitted through the radiograph on the illuminated screen is
sufficient for writing etc., but if felt necessary, the ambient lighting can be increased,
provided it does not produce any specular reflections of the surface e of the
radiograph.
b Adaptation : Eye adaptation depends upon the light condition to which the
observer is subjected before viewing. As a guidance, an observer coming
from full sunlight should allow at least 10 minutes in subdued lighting before
commencing viewing; from ordinary artificial room light about 30 seconds
adaptation is necessary. The adaptation period should be one continuous
period of time. If eyes are subjected to full brightness of the illuminator
during the changing of radiographs, at least 30 seconds readaptation is
necessary.
e Eye Sight : All film readers shall check their eye sight annually for ability to
see small details at normal film viewing distance. These readers should be
able to read the letters and numbers of the second line on the attached chart
at their normal viewing distance.
76
9.7.4 Effects of Viewing Conditions on Image Quality
77
CHAPTER - 10
DEFECTOLOGY
10.1 INTRODUCTION
As mentioned above, the image of discontinuity in the object is formed due to the
differentials absorption of radiation in the object. Intensity of radiation transmitted
through 'x' having no discontinuity, can be graphically shown as in figure 5.1 (a).
Mathematically, it can be expressed by equation
Depending upon the type of the material of the discontinuity, two possibilities, exist,
either µ is less than or grater than µ'. Graphical attenuation in both cases is shown
in figure 5.1 (c). The equation 3 can be rewritten as :
78
(A) Attenuation of Radiation (B) Influence of radiation quality on
radiographic contrast
79
This is not true in radiography process with X-ray (Continuous spectrum) and
radioisotopes having more than one energy. The values of µ change with increase in
thickness of material, as softer components of the radiation beam are filtered off by
initial thickness.
The above discussion also to the defect-image contrast which depends upon the
difference of attenuation co-efficient of the defect and the sound material. More the
difference, higher would be the contrast and hence, easily detectable.
Table 10.1
µ Cm-1)
LINEAR ABSORPTION CO-EFFICIENT (µ
Source Material
Lead Steel Aluminum
From the above, the subject contrast in percentage for the cavity of 5 mm can be
calculated as given in Table 10.2. It can be seen from the values that the image of a
cavity with lead will be seen as positive image, whereas with aluminium, it will be
seen as negative image. On an X-ray film, the positive image would be brighter and
negative image would be darker.
80
TABLE 10.2
The example discussed above shows that radiation from iridium-192 source produced a
better subject contrast, as compared to that from cobalt - 60 source. Greater contrast
helps in detection of smaller defects.
A defect will be visible when the radiographic contrast -^D is equal to or greater
than minimum perceptible density difference (-^Dmin), which depends upon ;
a. image size,
b. density distribution,
c. radiation energy
d. film processing
Examples of welding and casting defects and their radiographic appearances are
given in Table 10.3 and 10.4.
81
TABLE 10.3
WELDING DEFECTS
82
Defect Cause Radiographic detection
83
TABLE 10.4
CASTING DEFECTS
84
CHAPTER - 11
1. Radiation source
2. Films used
A technique is selected on the basis of the knowledge about the following factors :
2. Fabrication Process
85
11.2 TECHNIQUE PARAMETER SELECTION
Table 11.1 gives the source selection for various steel equivalent thicknesses.
TABLE 11.1
150 keV 15 50
200 keV 25 75
Choice of film from class 1 to class IV, is made as in table 11.2 a & b. Generally,
for detection of finer defects, fine grain, high contrast films are used.
86
RECOMMENDED AGFA-STRUCTURIX X-RAY FILM
0-06 D2, D4
D5, D7
Over 100 D7 D7 D7 D7
Magnesium
0-06 D2, D4 D2 D2 D2
50-100 D7 D4,D5,D7 D4
100-200 D7 D4
Aluminium
87
06-12 D5, D7 D4 D2, D4 D2
12-25 D7 D4,D5 D2
50-100 D7 D7 D4 D5
100-200 D7 D7
Steel
50-100 D7 D7
a Dimensions of focus and source are important for determining the value
of geometric unsharpness (Ug.)
SFD = t (1 + d/Ug)
Where,
SFD also controls the area of the specimen covered in a single exposure.
11.2.4 Sensitivity
88
11.3 LATITUDE TECHNIQUE
Some of the recommended practices for latitude technique are given below :
a Use of Low Contrast Film : Using a low contrast film, the thinnest portion of the
specimen is exposed for highest acceptable film density. Depending upon the film gradient,
higher thickness, represented by the lower limit of acceptable film density, can be covered
in the inspection.
c Use of Filter : A filter placed near the X-ray tube removes the softer
component from the radiation beam, thus, it improves latitude.
Fusion welds can be on flat plates or cylindrical objects, such as pipes, The weld
can also be of butt or fillet type. Radiographic inspection of these welds with
sensitive techniques reveals flaws, such as cracks, lack of root penetration, lack of
fusion porosity, inclusions, etc. Occurence of such defects in weld can generally be
identified with joint preparations, welding procedure, heat treatment etc. Information
on all these basic factors, guides in anticipating the occurrence of flaw at a location
and thus, helps in selecting a particular exposure geometry.
a Butt Welds : Butt joints on flat plates are usually made with edge
preparation of single V, double V or square. Welds with single and double V edge
preparation are examined, with the possible exposure geometries as shown in figure
11.1. Technique (a) & (c) with source located perpendicular to the weld is ideal for
detection of root flaws such as cracks, lack of root penetration. Angular exposure (b)
& (d) is best for detection of lack of fusion between weld and parent metal.
Inspection techniques for butt welds with ASTM image quality indicators are shown in
figure 11.2. These are applicable for :
89
Fig. 11.1 Radiography techniques for inspection of Butt Welds on flat plates.
b Fillet Welds : Fillet welds are generally made with square or bevel edge
preparation.
The exposure set up for joints having 100 percent corner penetration and 100
percent standing leg penetration are shown in figure 6.3 (a&b). The source is
positioned at an angle to enable detection of lack of side fusion and also lack
of root fusion.
90
FIg. 11.2 Radiography of Butt-Welds with use of ASTM penetrameters.
91
Square joints with four fillets and lap joints are examined for root penetration
with exposure set up as shown in figure 11. 3 c&d).
92
Corner joints may have configuration as shown in figure 11.4. Square joint
with 100 percent corner penetration is radiographed as shown in figure 11.4a
for revealment of side fusion, if any. Corner joints with double bevel are
examined with exposure set up as shown in Fig. 11.4b A plane corner joint,
having 100 percent joint penetration is examined for root cracks with film
and joint in a parallel position (Fig. 11.4c)
Depending upon the size and accessibility of both sides of the pipe, the following
techniques are adopted.
In this technique, several methods of exposure are possible (Fig. 11.5). The
ideal position to locate the source would be the centre of the pipe of cylinder
as in (a) & (d). However, if the centre is not accessible or if pipe diameter is
small, the source can be off the centre and the exposure scan be taken as in
(b). The welds can be divided into several segments and the exposures are
carried out.
93
In the second case, the film is placed inside the pipe and the source is placed
externally as in (c). This technique is more sensitive for detecting root defects.
However, the area accessible for interpretation will be small and the entire weld is
covered by taking several exposures.
1. Source 2. Film
If there is no access to the inner side of the pipe, to keep either the film or
the source, this technique may be adopted (Fig. 11.6). This is mostly suitable for the
pipes of diameter above 80mm. Source can be kept on the pipe or away from the
pipe, depending on the diameter of the pipe, SFD required and accessibility.
This technique is adopted for smaller size pipes of diameter upto 80mm. The
source is kept at one angle with respect to the weld, so as to avoid overlap of the
top and bottom weld. Generally, 10o - 12o offset angle is given for the exposure.
Two exposure are taken by rotating the pipe through 90o.
94
CHAPTER - 12
12.1 INTRODUCTION
The purpose of inspection standards and codes is to prescribe a test procedure for
correct assessment of the quality of a product. The inspection standards and codes
prescribe optimum conditions, which could produce test sensitivity closest to the
theoretical values.
There are several national and international standards and codes practice in industrial
radiography. Some of the prominent ones are ASTM, ASME, B.S, DIN and ISO.
These are broadly classified into four groups.
Relevant standards
These standards specify the type, material, size and shape of IQI and method for
calculation of sensitivity.
4. Radiograph density.
5. Source selection.
95
8. Procedure requirements.
ASME Boiler and Pressure Vessel code Sec. III & VIII
BS 5500 (1515)
Sets of different types of radiographs are available for welds and castings.
Indications revealed on the radiographs are classified as per type of defects and also
graded according to their severity levels. Principle collection of reference radiographs
include.
Weld
Castings (Steel)
96
12.3 STANDARDS FOR WELDS IN STEEL PLATES
Class A- General technique for X-rays applicable to mild or low alloy steels
Class B- More sensitive for X-ray techniques for important applications needing
higher sensitive.
Class C - General techniques for gamma rays for general applications when shape,
thickness and accessibility renders use X-rays impossible.
For plate thickness in the range 50-200 mm, ISO 2405 is applicable.
Table 12.1
Energy Thickness in mm
Iridium-192
ISO R 1106
97
According to ISO 2405, only direct types of films with medium speed fine grain and
very fine grain films can be used. Recommendations for screens, as per BS 2600,
are given in table 12.1
Minimum SFD depends upon the permissible Ug values. It also decides the field of
coverage in a single unidirectional exposure.
In ISO 1106, the resulting Ug value should not be more than 0.4 mm for techniques
A and C and 0.2 mm for technique B.
Table 12.2
Material Maximum Ug
thickness
(mm) (mm)
Upto 50 0.5
50-75 0.75
75-100 1.0
ISO/R 1106 - 10% of the material thickness for Class A & B techniques.
The limits of acceptable film densities and maximum fog density levels are given at
table 12.3 Higher film densities are permissible, if adequate viewing facility is
available.
98
TABLE 12.3
TABLE 12.4
ISO R.1106 10 25 38
BS 2600 Part I 10 25 50
ASME 19 38
99
TABLE 12.5
Shapes and types of IQI generally used in radiography testing are shown in chapter
4B. Types of IQI specified under different standards are given in table 12.6.
TABLE 12.6
API 1104 - Standard for welding pipelines and related facility by American
petroleum Institute
BS 2633 - Specification for Class I Arc welding of ferrite steel pipe work for
carrying fluids.
100
12.7 RADIOGRAPHY OF WELDS OF BOILERS AND PRESSURE
VESSELS
ASME Boiler and Pressure Vessels Code Section V exclusively deals with NDT
requirements.
The code prescribe minimum qualification of the personnel as well as their visual
activity.
Though ASTM and ASME do not specify separate codes for radiographic
examination of castings, the specifications cover important aspects of radiography
procedures in various designations. Some of the important ones are
Besides these, there are number of sets of casting reference radiographs, covering
materials such as steel, copper-nickel alloy, tin, bronze, aluminum and magnesium,
described in different designations. BS 4080 gives procedure for radiographic
examination of all types of steel castings. important point to be considered, besides
other general recommendations, are given below.
a. Marking of area
b. IQI-Use of wire or step type IQI given in BS 3971 and their acceptable
sensitivity levels for different thickness are specified.
c. Only direct type X-ray films to be used in conjuction with metal foils made
of lead, copper, tantalum and tungsten, the thickness and the material
depending upon radiation energy employed.
101
Table 12.7
While using double film technique to cover thick and thin sections in a casting, the
kilovoltage energy of the radiation will be determined by the thicker section
TABLE 12.8
CODE TITLE
102
Except IS 3657, all other standards cover radiography practices in a general way.
Highlights of these specifications are as under :
c IS 3657 : The standard gives the specification for fabrication of image quality
indicators.
For standard radiation sources, the source to film distance can be read off
directly from the graphs. The acceptable film density is 1.7 to 3.0 for
techniques (a) and (b) and 2.0 to 3.0 for techniques (c).
103
12.10 ACCEPTANCE LIMITS
12.10.3 Acceptance/Rejection
b) Measuring dimensions of the flaws and tallying their sum with the acceptable
norms, given in the standard.
a) Welds
104
ASTM-E 390 has 3 volumes of reference radiographs, based on seven nominal weld
thicknesses. There are graded flaws shown in five severity levels. Upgraded flaws
are given for information.
b) Cracks
Any type of crack is unacceptable, under the codes for pressure vessels, and
pipelines, exception being crater crack upto 4mm permissible under API-1104
Not acceptable, under all codes applicable to pressure vessels. However, codes
applicable to pipelines carrying fluids such as BS 2654 Part 2, API 1104, BS 2633,
BS 4677, the welds are rejectable when the length exceeds the specified limit.
d) Inclusion
In case of tungsten inclusion, only a small size less than 3 mm or 1/4 of thickness
of weld, which ever is less, is permitted.
e) Porosity
f) Root Cavities
Root cavities resulting in thickness of welds lesser than that of parent metal are
unacceptable.
g) Burn Through
h) Undercuts
Undercuts should not exceed certain percent of the pipe wall thickness, maximum
permitted being 0.8 mm
i) Castings
ASME section VIII Provides guidelines for acceptance/rejection limits for steel
castings, upto thickness of 305 mm, using following ASTM standards :
105
E-466 - Upto 51 mm thickness
Casting defects in these standards have been divided in seven groups, such as gas
porosity, sand and slag inclusions, shrinkages hot tears, cracks, linear discontinuity,
inserts and mottling. These defects have been presented in 5 classes representing
increasing severity levels. Maximum acceptable severity limits under ASME section
VIII are given in table 12.9.
TABLE 12.9
Gas porosity 1 1 2 2
Inclusions
Shrinkages 1 3
Type 1 1 3
Type 2 1 2
Type 3 3 2
Hot tears 0 0 0
Cracks 0 0 0
Linear 0 0 0 0
discontinuity 0 0 0 0
Inserts 0 0 0 0
Mottling 0 0 - -
For ferrous castings, weighing 50 kg. and above, suggested acceptance guidelines are
given in table 12.10
106
TABLE 12.10
2. Sponginess Unacceptable
107
TABLE 12.11
DETAILS OF STANDARDS
ISO R 1106 Recommended practice for radiographic inspection of fusion welded butt
joints, for steel upto 50 mm (2 inches) thick.
ISO 2405 Recommended practice for radiographic inspection of fusion welded butt
joints, for steel plates 50 to 200 mm thick.
Part - 1973 examination of fusion weld circumferential butt joints in steel pipes.
B.S. 4677 Specifications for Class I arc welding of 1971 austenitic stainless
steel pipe work, used for carrying fluids.
108
CHAPTER - 13
SAFETY IN RADIOGRAPHY
13.1 RADIATION HAZARD CONTROL
TABLE 13.1
Cobalt-60 source has gamma radiation of two energies 1.17 Mev and 1.33 MeV.
Gamma radiation energies from a given radioisotope (say cobalt-60) are the same
whether it's activity 1 Ci, 1000 Ci or of any other value. Gamma radiation of higher
energy has higher penetrating power.
The radiation level or the exposure rate at a particular location increases with
increase in activity of source.
Radiation level (R/h) at 1 metre from a 1 Ci (37 GBq) radioactive source is known
as the RHM (Roentgen/Hour at 1 metre from the source). It is a constant for a
particular source.
109
RHM = 0.52 E R/h (See Table 13.1)
Similarly, of the activity of the source is given in Si units i.e. in GBq, then the air
kerma rate at 1 meter for 1 GBq activity, for any particular source can be considered
as a constant. This value would be = 0.125 mGy/h (air kerma rate)
Thus, the radiation level at 1 metre from a 7 Ci (259 GBq) iridium-192 source is
Similarly, at 1 metre from a 185 GBq (5Ci) cobalt-60 source, the radiation level will
be
If the radiation level at 1 meter is known, the activity of the source can be
determined. Say, if the radiation level at 1 meter from an iridium- 192 source is 1.5
R/h, then
Similarly if the source is cobalt-60, and the radiation level is 0.66 R/h at 1 meter
from the source, then 1.32 x Ci = 0.66 R/h
0.66
Activity of the cobalt-60 source = = 0.5 Ci = 18.5 GBq
1.32
b. X-ray output.
X-RAY OUTPUT (R/H) depends not only on the kV and mA of the X-ray
equipment, but also on various other factors such as construction of the tube, size of
the target, type of rectification, amount of inherent filtration of X-ray tube, etc. The
radiation output from an X-ray equipment is very high as compared to the radiation
output from gamma sources, normally used in industrial radiography. Table 13.2
gives and idea about the radiation output of some common industrial X-ray
equipment.
110
The hazard from and X-ray equipment depends not only on the output, but also on
the applied voltage, since the later determines the penetrating power of the radiation.
TABLE 13.2
Make & Type Max kVv/mA Filter Focal Spot Radiation output
mm mm2 R/min at 1 m
The three basic factors by which external radiation hazard can be controlled are
a. time
b. distance
c. shielding
Time :
It is essential, from radiation protection point of view, that the time spent near
gamma or X-ray equipment during radiographic exposure should be minimum. The
radiation dose received by a radiographer depends upon the total time spent by him
near a source.
In view of high radiation output from an X-ray equipment, even a few seconds of
exposure to direct X-ray beam will result in very high dose to the exposed person.
111
If an operator takes 1 minute while working with a source and is exposed to 10mR,
another operator who takes 2 minutes to do the same job would be exposed to
2 x 10 = 20 mR. If we were to take 1/4 minute, he would have been exposed to
only 1/4 x 10 = 2.5 mR.
Example 13.1 :
Example 13.2 :
112
Distance :
Larger the distance, lesser is the radiation level. The most effective and economic
method of reducing external radiation is to increase the distance between the
radiation source and persons.
b. Use long cables between X-ray unit and the control panel.
The radiation intensity or radiation level from a given source decreases as we move
away from the source. It is governed by the following inverse square law.
I1D12 = I2D22
If means, if we double the distance, the radiation level would be one fourth of the
x˚
ˆ 2 =
ˆ
earlier level.
Thus, at 2 m distance (D1) from a gamma radiography source, if the radiation level
is 125 m R/h (I1), at 5 m distance, (D2) the radiation level (I2) the radiation level (I2)
would be 20 mR/h, since
125 x 22 = 12 x 52
= 20 mR/h
Example 13.3 :
113
RHM for iridium-192 = 0.5 R/h at 1 m
Radiation level at 1 m (D1) from a 74 GBq (2Ci) source (I1) 2 x 0.5 = 1 R/h
Since, 1 x 1 = I2 x 52
Example 13.4 :
The radiation level at 4 metres from a gamma source is 500 mR/h. What is the
distance required to reduce the radiation level to 20 mR/h ?
500 x 42 = 20 D22
or
D2 = V400 = 20 m = 20 metres
Example 13.5 :
What is the distance in metres required the radiation level from a 148 GBq (4 Ci)
iridium-192 source to 0.20 mR/h ?
2000 20000
D22 = = = 10000
0.20 2
D2 = v10000 = 100 m
114
Example 13.6 :
I1 = 20 R/min. D1 = 50 cm I2 = ? D2 = 10 m
I1D12 = I2D22
20 x 0.5 x 0.5 = I2 x 10 x 10
20 x 0.5 x 0.5
I2 =
10 x 10 = 0.05 R/h = 50 mR/h\
Example 13.7 :
What is the distance in metres required to reduce the radiation level from 40 R/hour
at 1 metre to 100 mR/hour ?
I1D12 = I2D22
40000
D2 2 = = 400
100
D2 2 = 400 = 20 metres
13.1.4 Shielding :
When maximum distance and minimum time do not bring down the radiation level
to an acceptably low value, materials like concrete, steel or lead are used as a shield
between the source and the operator.
Shielding is provided in order to reduce the radiation intensity to the required value.
The shielding thickness required can be calculated from transmission graphs (Fig.
13.2). Transmission is the ratio of the addition level with the shield to that without
shield. Radiation level with shield.
115
The required shielding thickness can also be calculated by making of half value
thickness or tenth value thickness
The shielding properties of any material depend upon the atomic number of the
material and the energy of the incident radiation.
Example 13.8 :
Calculate the lead shielding thickness required to reduce the radiation level from a
14 Ci ( 518 GBq) iridium-192 source to 3.5 mR/h at 10 metres.
116
14 x 0.5 14 x 0.5
= 10 x 10 R/h = 10 x 10 x 1000 mR/h = 70 mR/h.
This is the radiation level at 10 metres from the source without any shielding. One
TVT of shielding will reduce the radiation level to 70-10 = 7 mR/h. Additional
thickness of 1 HVT of shielding will reduce the radiation level to 7-2 = 3.5 mR/h.
Therefore, the total thickness required is
= 1 TVT + 1 HVT
Example 13.9 :
The radiation level at 1 metre from a lead container housing an Irridium-192 source
is 10 mR/h. If 1 cm lead is removed from the container, what will be the new
radiation level at the same distance ?
1
No, of HVT in 1 cm of lead = = 2 (approx)
+
0.48+ +
ˆ ++ ˆ ++
Removal of 1 HVT of shielding will increase the radiation level by 2 times.
Hence, Radiation level at 1 metre from the container after removal of 1 cm lead =
22 x 10 = 40 mR/h.
Example 13.10 :
Calculate the required shielding thickness in lead which will reduce the radiation
level from an X-ray equipment (300kv 10mA) to 3mR/h at one metre. [Given-
radiation output per mA. min (for 300 kV X-ray equipment) at 50 cm from the
target is 20 R]
117
3000
Reduction factor = = 1000 = 103
3
= 18mm = 1.8 cm
Example 13.11
An X-ray machine is operated at 150 kv and 5 kv and 5 mA. What is the thickness
of concrete required to reduce the radiation level at 5 meter to 1 mR/h?
(See Figure 13.2)
the radiation level at 4 m should not exceed (0.001-5 ) R/h i.e., 0.0002R/h
13.2.1 General
Radiography installations are those areas where X and gamma radiations are used for
purpose of radiographic inspection.
Areas which are occupied only by radiation workers and are under direct supervision
of Radiological Safety Officer are called controlled areas and the radiation levels in
these areas shall not exceed the specified whole body does limits for radiation
workers (40 mR in on week.)
118
Areas which are occupied only by radiation workers and are called uncontrolled
areas. The radiation levels in these areas shall not exceed the whole body dose limits
recommended for members of public (2 mR in one week).
In the case of enclosed installations, the radiation levels outside are controlled by
providing barriers (shielding) of concrete and brick (for walls) and steel or lead (for
entrance doors). The thickness of the barrier depends upon many parameters, such as
radiation source used and its activity, in case of radioisotopes, the maximum ratings
(kV & mA) of X-ray equipment, X-ray output in case of X-rays, weekly workload,
distance between the barrier and the radiography equipment, occupancy factor, use
factor and dose limits. The HVT and TVT values, can be made use of for
calculating shielding thickness values.
While selecting the area for use as X-ray installation, an area of comparatively low
occupancy is preferred. The area chosen shall be such that it should facilitate easy
and covenant movement of radiography objects.
Enclosed installations are those area specifically earmarked for radiography purpose
with walls of adequate thickness all around, so that the radiation levels outside the
walls are well below the operational limits. The objects to be radiographed are
always brought to the enclosure for radiography examination. Radiography work can
be carried out safety inside the enclosure without causing hindrance or radiation
hazard to other workers outside. As far as practicable, all radiography exposures
should be given, only in such radiography enclosures, the nature and activity of the
source used, the size of the objects to be radiographed, the total work-load etc.
(figure 13.3 and 13.4)
119
Wherever the objects are huge in size and also heavy, a open-to enclosure or a pit
type enclose (fig. 13.5) may be provided so that the objects for radiography may be
lowered into the exposure room by means of overhead cranes. In such radiography
enclosure, the radiation dose to the cranes operator should be well within the
specified limits. Further, in such installations, the sky shine radiation level (the
contribution due to the radiation scattered by air and other intervening materials
above, such as beam, roof etc.) around the enclosure in the workshop should be kept
below the limits. The maximum activity of the source that can be handled in such
installation is mainly decide by the sky shine considerations. The construction
material used for such installations may be either brick or concrete. These installation
should preferably be located in the least occupied areas. Expert advice on the
planning of such radiography installations is given by the Radiation Protection
Services (RPSD) BARC, Bombay 400 085
Figure 13.4 - Enclosed Installation (Exposure Room with Maze Entry System.)
120
Fig. 13.5 Enclose Installation (Pit type- Open top)
The following are some of the salient points which need to be considered while
planning radiography enclosures as well as during the use of radiography sources
inside such enclosed installations.
2. All walls, floors ceilings and doors shall have sufficient shielding so as to
minimize radiation levels, below the recommended dose limits. The Shielding
material chosen should be structurally strong and should be properly
supported.
3. All exposures should be given from a separate control room, situated outside
the exposure room.
6. The entrance door to the exposure room should be locked during exposures.
A suitable electrical or mechanical interlocking system for the door may be
incorporated so that the radiation beam cannot be made 'ON' when the door
is open.
121
7. Underground conduits should be provided, for cables, between the control
room and the exposure room. There shall not be any through and through
opening or hole in any of the walls of exposure room.
8. If there is a lead glass viewing window in the wall between control room and
exposure room, it is essential to confirm its lead equivalence, so that it
provides adequate protection. It is also necessary to confirm that there is no
radiation streaming through the fixing frame of the viewing window.
9. There should not be any windows on the walls of the exposure room. Exhaust
openings should be provided on walls adjoining unoccupied areas at a height
of at least 2.5 metres above the floor level. These ventilators or exhausts or
any such openings in the walls should be provided with suitable baffles.
10. A red warning light shall be provided above the entrance door, as well as
inside the exposure room and this light shall be made 'NO' whenever the
radiography equipment is energized. Radiation warning signs with legends
"RADIATION-DANGER-KEEP AWAY" SHALL BE conspicuously painted
on the entrance door.
12. In the case of open top enclosures, red light should be provided on the top of
the walls and should be made 'ON' during exposures as a warning to the care
operator.
13. All operations with the equipment should be done from the control room.
14. An emergency switch to turn the X-ray equipment 'OFF' and a means of
communication shall be provided inside the exposure room, which can be
made use of by persons who are trapped there by mistake. It should be
possible to energise the X-ray equipment from the control panel, only after
pressing a reset switch provided inside the exposure room.
15. Setting up of objects, films, etc. for radiographic inspection should be duly
completed , before starting the exposure.
16. Wherever possible, the radiation beam should be directed towards areas of
minimum occupancy. The beam should never be pointed towards doors,
windows or the control panel. Any restrictions on the beam directions which
are assumed while planning the installation should be strictly followed. This
information should also be prominently displayed in a poster in the
radiography area.
122
17. All safety accessories, tools and handling devices required for routine or
emergency use in the exposure room must be readily available in good
working condition in the control room.
19. If more than one radiation machine is used in the same room, it should be
ensured that only one of them is operated at a time.
In some cases, it may not be possible to restrict the radiography work within the
exposure room, eg., huge objects, radiography at erection sites etc. In such cases
radiography work in the field/plan area is permitted subject to the approval of the
site by the RPSD, BARC.
3. Certain area around the source, should be cordoned off using ropes and
radiation warning symbols. The radiation outside the cordoned area should be
within the dose limit for non-radiation workers.
4. The exact area to be cordoned off will depend upon the nature and activity
of the source, type of exposures (collimated or panoramic), the work load and
the nature of occupancy around.
6. Proper collimators should be used to limit the radiation beam to the job to be
radiographed.
123
7. The collimated beam or in case of X-rays, the primary beam, shall be
directed only towards unoccupied areas or adequately shielded areas.
11. Maximum length of the cable, provided between the control panel and
radiography equipment, shall always be made use of.
12. Whenever possible, the control panel of the X-ray equipment should be
positioned behind some temporary shielding barrier. A temporary Shield,
preferably lead, should also be provided behind the object, in the direction of
the direction of the primary beam.
13. The primary X-ray beam shall never be pointed towards the control unit.
14. Properly calibrated radiation monitors, shall always be used, to confirm that
radiation levels, in controlled areas, as well as along the cordon, are below
specified limits.
2. Monitor the guide tube to ensure that the source is not left inside the guide
tube.
3. Lock the camera so as to ensure that the source pigtail would not be released
from the camera.
4. Monitor the camera with a suitable survey meter (eg., MR 4500, SM 140D
or Minirad), in order to determine if there are any hot spots on the camera. If
124
any hot spot is detected the camera should not be despatched without
obtaining specific prior permission from Head, RPSD, BARC, Bombay 400
085.
5. If the camera is free from hot spots, place the camera in a sturdy wooden box,
preferably in the same box in which it was originally received from the supplier or
in an equally sturdy box. Make sure that the box is in good condition.
7. Act as directed in the 'procedure for making, labelling and forwarding the package
containing radioactive materials' given below.
1. Monitor the radiation level at the external surface of the box using a suitable
survey meter (eg. MR 4500, SM 140D or Minirad) and record the maximum
radiation level at the external surface of the package in mR/h. This should
not exceed the limit of 200 mR/h.
2. Measure the radiation levels at a distance of one meter from the external
surface of the package in mR/h and record the maximum value. This
maximum valve is the transport index (T.I.) of the package.
3. Having thus determined the category of the package, affix the labels of the
determined category on at least two opposite sides, on the external surface of
the package.
TABLE 13.3
5. Write the following details on the external surface of the package clearly and
legibly.
125
a. Complete address of the sender and the receiver of the package,
8. Forward the package to the carrier along with the above transport documents.
1. An appropriate source storage facility necessary for safe storage and proper
security of the source. Not having a proper storage facility may result in loss
of radiography sources. All the radiography sources, cameras and source
containers (lead pot), when not in use, should be stored necessarily in a
separate storage room/shed with a pit under lock and key. Radiography
cameras and leadpots should never be left at the work site unattended.
3. The radiographer should regularly check the proper and safe storage of
radiography sources inside the radiography camera/storage room by
monitoring the radiation levels around, using a suitable radiation monitor.
4. After storing the radiography camera/lead pot inside the storage place, the
storage place, the storage place should be securely locked and the key should
be handed over to a responsible person for safe custody.
5. If more than one radiography camera is to be stored inside the storage room,
individual storage pits may be provided inside the room for easy
identification so as to avoid any possible mix-up of source. While providing
storage pits, care should be taken to prevent water seepage into the pits
during monsoon.
126
Fig. 13.6 Radiography source storage room.
7. Radiation level around the storage room should be less than 0.25 mR/h.
1. The operators of the radiography equipment should ensure that all the safety
instructions, stipulated for the particular radiography equipment, by
Radiological Safety Officer (RSO), are scrupulously followed.
127
6. Unauthorized entry into controlled areas shall be prevented. Particulars care
should be exercised while using panoramic X-ray equipment (rod anode
tubes) since in such units, the primary X-ray beam is emitted in all directions
(3600).
7. All radiographic exposure shall be given only by duly approved operators and
they shall be in immediate attendance when the radiographic equipment is in
use.
9. The radiographic equipment, when not being used, shall be secured in such a
manner as to prevent its operation by unauthorized persons.
11. The operator of the radiographic equipment shall maintain a log book to
record the details of day-to-day use of the equipment.
12. Any emergency situation which might result in higher radiation dose, should
be immediately brought to the notice of Radiological safety Officer.
13. The operators should be aware of their dose records and of periodical
medical examination.
Gloves (not even lead gloves) do not offer any protection. Always use
suitable manipulator rods/tongs for handling the source pencil.
128
3. Never forget to carry the remote handling tongs and other source handling
accessories along with the radiography equipment to the site.
4. Never leave the camera or lead pot unattended at the site. Always store the
camera and the lead pot in a separate storage area under lock and key.
5. Never leave the source unattended at the site during exposures. A person
must be present at the site during exposures to have a constant watch so that
the sources are not picked up by others.
7. Never forget to cordon off the area with ropes. Radiation warning symbols or
red lights should also be provided along the cordon to warn people.
9. Never attempt to take out the source pencil from the camera.
10. Always try to use the minimum required source activity for panoramic
exposures.
11. Try dummy operations prior to giving the actual exposures in order to reduce
the time of operation with radiation source.
12. Always keep maximum possible distance from the source during exposures.
129
16. The control badge provided at each site along with personnel badges must be
kept in a radiation free zone away from the area where the radiography
sources are used/stored.
17. In case of accidents such as damage to the camera or source pencil or loss of
source pencil, cordon-off he area and report the matter, immediately, to the
superiors and subsequently, to Head, Radiation Protection Services Division,
BARC, Bombay 400 085.
130
CHAPTER - 14
1. inadequate training,
2. failure to monitor,
Of these, the last factor, the tendency of the operators to overlook some of the basic
requirements for proper work practice during routine work, has been the major cause
in maximum number of accidents in industrial radiography. The following are some
of the events which may lead to radiation emergencies.
131
14.2 INCIDENTS ON RADIOGRAPHY
The following are some of the accidents which have occurred in our country.
c The drive cable of a cobalt-60 camera was kept coupled to the source pencil.
Once during use, the source pencil got detached from the drive cable and remained
in the guide tube. The operator thought that the source is safe inside the camera
as he could rotate the shutter drum, after retraction of cable. He failed to confirm
this, with a radiation monitor. After sometime, that too after completion of job
for the day, the source pencil fell out of the guide tube, when it was uncoupled.
He made some efforts to put the source pencil back into the camera. He received,
considerably high dose. This accident occurred because the coupling between
the source pencil and the drive cable was not checked periodically no monitoring
was done with survey meters to confirm the position of the source.
132
radiography work. This person, not knowing that he is handling hazardous
material, left the source penicillin the driver's cabin of a lorry and sold the
transport container to a scrap dealer. The source pencil was located only after
two ways. The above incident took place, mainly, because of negligence on
the part of the radiographer in not keeping the radiography camera in the
source storage room, after completion of radiography work.
Cause : Poor design of source capsule. Operator failed to apply his common
sense.
Many more incidents have occurred in industrial radiography. The main cause
is poor preventive maintenance and failure of the operator to monitor. Many
overexposure could be avoided, if radiation monitors are used regularly.
There are about 400 institutions who use radiation sources for radiography.
What is given above are a few typical ones of around 70 accidents, which
have occurred in the country. These accidents occurred mainly because of
ignorance, negligence and carelessness.
The hazard associated with such accidents would be less serious, if only the people
working with radiation sources, use radiation monitors properly, during their routine
work.
The following are a few points which help in minimizing the occurrence of a
radiation accident and associated hazard.
1. Proper training.
2. Awareness to hazard.
It should also be noted that, if, inside of all precautions, any unusual incident occurs,
one should never become panicky. The operator should stop work with the
equipment, cordon the area and report the matter to his higher authorities and
Radiation Protection Services Division (RPSD), Bhabha Atomic Research Centre,
Bombay 400 085.
The site incharge/R.S.O. Should monitor around the place of accident and analyze
the situation. No attempt should be made without calm thicking and proper planing.
If he has necessary tools, radiation monitors, pocket dosimeters and if he is familiar
with the equipment and confident of handling the situation with least exposure to
himself and to others in vicinity, he may do it. In such a case, he should inform
RPSD, BARC, Bomaby 400 085, about the accident and write in detail about the
situation and the procedure adopted in handling it and the dose received by those
who handled it. If he feels that he would not be able to handle it and the situation is
complicated, he should cordon the requisite area, provide necessary shielding around
the source and inform RPSD, for further action.
Madding with the equipment or in other words, grouping in the dark when an
unusual incident occurs, would make the situation worse and complicate the
management of the accident.
134
14.3 REQUIREMENTS HANDLE RADIATION EMERGENCY SITUATIONS
** Your are permitted to work with radiography source because you are a trained
person.
135
CHAPTER - 15
Work with radiations sources may involve exposure of workers and public.
Therefore, work with radiation should be controlled by the Government. Such control
can be effective only if rules and regulations are made by the Government. The
Government of India issued Radiation Protection Rules, 1971. These rules prescribe
that a licence should obtained by any person., prior to handling radioactive materials.
The rules are enforced by the Atomic Energy Regulatory Board (AERB), who is the
Competent Authority. Under these rules, a surveillance order was issued in 1980,
with particular reference to industrial radiography. The salient features of this order
are given below.
3. For each radiography site, the licensee should appoint a qualified, trained site
incharge.
6. Each side should be equipped with suitable area monitoring devices, which is
capable of measuring low dose rates as well high dose rates.
8. The licence should provide adequate facilities for the of the source.
10. The licensee should maintain complete records relating to the details of the
work carried out, personnel dose and medical history.
136
15.2 DUTIES AND RESPONSIBILITIES OR CERTIFIED RADIOGRAPHER
4. After completing each exposure, make sure that the source is safely back in the
camera, by means of a suitable radiation monitor.
5. In the event of an emergency, report the matter to the site incharge without delay.
Unit the emergency situation is brought under control, take suitable measures for
prevention of excessive exposure of other radiation workers and public to radiation.
137
CHAPTER - 16
2. Scintillation media
Radiation produces ionization in a gas medium an the ions so produced are collected
as current of pulses. Gas filled detectors are generally, of cylindrical shape, with two
electrodes, a central electrode and an outer sheath, separated by an insulator (figure
16.1). A positive voltage is applied to the central electrode, with respect to outer
sheath. On exposure to radiation, ionization takes place in the gas medium, ie.,
electrons are knocked off the gas molecules. The ionized gas molecule is called the
positive ion. The ;positive ions and the negative electrons, thus formed, respectively
move to the outer (Negative) and central (Positive) electrodes. These ion pairs are
collected and measured. These gas filled detectors function differently, at different
voltages applied to the central electrode. There are five different regions of response.
Of these, two regions are of interest to us.
138
Fig. 16.1 Gas Filled Detector System
a. Ionization Chamber Region : In this low voltage region (-200 volts), the
number of ion pairs produced and collected are same. For radiation
monitoring, the ionization chamber are filled with air, generally, at
atmospheric pressure. The effective atomic number of the wall material would
be close to that of air. Hence, such an ionization chamber would give energy
independent response. Beta Gamma Exposure Rate Meter (SM 140) of
Electronics Corporation of India Ltd., Hyderabad, Incorporates an ionization
chamber as a radiation detector. Pocket dosimeter is another device which has
an ionization chamber. It has a built-in capacitance which can be charged by
an external voltage and when exposed to radiation, the ionization in the
chamber decreases the voltage across the capacitance. The reduction in the
voltage across the capacitance is a measure of the amount of ionization and
hence the quality of radiation exposure.
b. Gieger Mueller (GM) Region : This occurs at higher voltage (300-1400 volts),
wherein multiple ionization takes place in the medium. As a result, the number
of ion Paris collected are much more than that collected ionizing radiation, because
it has high sensitivity. In a GM counter, the central electrode is fine wire (0.2
mm to 0.1 mm diameter) of tungsten and the outer electrode is usually of metal.
The filling gas is generally argon with quenching gases, like ethyl alcohol or
halogen vapour, suppress secondary discharge. Depending upon the gas used
and the pressure, the voltage difference between the two electrodes may anywhere
between 300 volts to 1400 volts. There are GM detectors of different shapes and
sizes, depending on the purpose of use and radiation level to be monitored. GM
detectors may fail to respond when the radiation level is very high. Many radiation
monitors incorporate GM detectors.
139
produced, multiplied as a large electrical pulse. Thallium activated iodide-NaI
(TI) schitillator is used for gamma radiation detection. These detectors are
used for monitoring very low radiation levels.
I is light intensity through the processed film. Photographic films are used in
industrial radiography for revealing defects in castings, welds, etc. Photographic films
are also used for personnel monitoring, by loading personnel monitoring films in
film cassettes containing different metallic filters, designed for monitoring different
types of radiations.
Since neutrons are uncharged particles, their detection methods are different from
those of charged particles and gamma radiation. Neutrons are classified on the basis
of their energy. Thermal neutrons (energy - 0.025 eV) are generally detected by
using boron trifluride gas (BF3) filled proportional counters. Here, thermal neutrons
interact with boron to release alpha particles and these alpha particles produce ion
pairs in the gas medium an in turn electrical pulses. In personnel monitoring film
badge, a cadmium filter is used for the detection of thermal neutron dose. Cadmium
140
filter gives prompt gamma radiations on absorption of thermal neutrons and these
gamma radiations produce blackening on the film, which become visible after
processing. Fast neutrons are generally detected by first reducing their energy to a
very low value (thermal neutrons) using devices are classified as personnel
monitoring devices and areas monitoring devices.
Using film badges doses from 0.1 mSv to 10 Sv (10 mrem to 1000 rem) of different
types of radiations can be evaluated. The film serves as a permanent record and if
worn on chest, gives the most representative value of the whole body dose under
normal working conditions. Films can be used to assess the radiation dose received
from exposure to beta rays, X-rays, gamma rays or thermal neutrons. Thermal
neutrons cannot affect the film directly. They are detected through emission of
gamma radiations by cadmium subsequent to absorption of thermal neutrons. For fast
neutron monitoring, a special NTA film is used. This film records tracks of recoil
protons formed due to interaction of cast neutrons with hydrogen atoms. The recoil
proton tracks are counted and correlated to fast neutron dose.
The dose received by radiation worker can also be determined by the use the TLD
badges. This consists of three CaSO 4: Dy TLD discs embedded in a metallic
framework and enclosed in a multifilter cassette (Fig. 16.3). The TLD badge can be
used to monitor beta, gamma and X-ray. The TLD badge can cover a wide range of
doses from 0.1 mSv to 10 Sv (10 mrem to 1000 rem). Radiation dose to personnel
can also be assessed by using pocket dosimeters (fig. 16.4). Pocket dosimeter are
very useful in certain operations, where the radiation levels very considerably and
may be quite hazardous. They give an on the-spot information of radiation dose, as
141
the dose record can be read directly by the person. Pocket dosimeters in the range of
200 mR, 5 R, 10 R etc., are available.
142
MINIRAD monitor can measure upto 5R/h and MR 4500 can measure upto 50 R/h
making it useful in radiation emergencies. Both these instruments can be used for
area monitoring as well as for monitoring, leakage radiation level around source
housing, X-ray equipment.
143
TABLE 16.1
144