Documente Academic
Documente Profesional
Documente Cultură
Radiological Health
An FDA Perspective on
Medical Device EMC and
Wireless
Jeffrey L. Silberberg
US Food and Drug Administration (FDA)
Center for Devices and Radiological
Health (CDRH)
Jeffrey L. Silberberg, MSEE
devices
Defibrillators
Pacemakers
Neurostimulators
Infusion pumps
first time!
Enforcement actions
To the FDA
Medical-device-related deaths
to both the FDA and the manufacturers
Medical-device-related serious injuries
only to the manufacturer
Neurostimulators
Infusion pumps
1V
Pacemaker Sensing Curve
100mV
Sense
10mV ICD Sensing Curve
100uV
10uV
1 10 100 1000
Slide courtesy Seth Seidman Frequency (Hz)
Tulsa
C63.27 Round robin
C63.16, ESD
IEC standards
IEC TR 60601-4-2, Medical electrical
IEC TR 60601-4-2
Scope: The IMMUNITY of medical electrical
equipment and medical electrical systems
Immunity test levels
-1-2: reasonably foreseeable maximum
http://www.accessdata.fda.gov/scripts/cdrh/cfdoc
s/cfStandards/search.cfm
Enter “Standard Designation Number”
Click “Search”
By manufacturers
Methods
Risk-based approach
EMC
Because IEC 60601-1-2:2007 exempts RF
receivers from immunity testing in the
exclusion band, the medical device’s wireless
communication(s) should be actively
transmitting during EMC immunity testing.
Risk management
d = 1,2 P
Conducted RF 3 Vrms 3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz
(Ed 3)
May 8, 2015 Wireless Electronics and Electrical Medical Devices 82
Conclusions
Medical device EMI continues to be reported
Most reports involve active implanted
devices
Defibrillators
Pacemakers
Neurostimulators
Infusion pumps
first time!
Questions?
Appendix
Supplemental material
An EMC recall case study:
EMC conformity gone wrong
Date Recall Initiated: March 17, 2008
Date Posted: July 23, 2008
Recall Number: Z-1902-2008
Product: Model X extracorporeal blood
circulation system
Reason for Recall: Stops pumping:
Interruption of Model X support may occur
when using a Model Y Electrocautery Unit
Manufacturer said:
Pump stopping and alarming was OK
Performed as intended
Reboot
– Ultrasound Artifacts
Spectrum Analysis of a noisy power
supply from a patient monitor
Actual
Interference began at the same time as installation of this new HVAC system,
and continued for 11 months before mitigation was achieved.
Normally Normally
Open Closed
Due to relays in HVAC circuitry that were either not rated for the current through them, or
incorrect for the circuit design, the plastic body melted and caused the normally closed condition
to be slightly open, resulting in arcing. There were two main hallways with an HVAC relay circuit
for each room, more than 60 relays that any one of them could produce broadband interference
of up to 15 minutes in length.
In-room hair dryers found during a pre-installation RF Survey. Not all hair dryers emitted noise.
Defective hair dryers functioned [in]correctly. The proposed wireless system would not have
functioned properly due to this noise.
Broadband noise from automated door into the NICU. These solenoids had degraded over time.
Replacement was recommended.
A broad frequency span shows the Only the weak signals are affected at
arcing characteristics of the dimmers. the time the noise occurs.
Recommendations: Install shielding on the dimmer system control cabinets.
Slide courtesy Dara McLain
Two wireless systems programmed to the same frequencies in use, interfering with each
other.
Unknown to Clinical Engineering, the offending system field service tech reprogrammed
the system transmitters. The affected system company resolved the issue at the
customer’s expense.