TEST REPORT

STFWT20205970

Product Name Daily Protective Mask

Trust Unit Shenzhen DreamCan Technology Co., Ltd.

Manufacturer Shenzhen DreamCan Technology Co., Ltd.

Test Category Commissioned Sample Delivery Inspection

JIANGSU QUALITY SUPERVISION AND INSPECTION CENTER FOR SPECIAL SAFETY PROTECTION PRODUCTS
Test Report
STFWT20205970 Page 1 of 5

Specification Type DP-A-21L-K


Daily Protective Mask
Product Name
Trademark

Trust Unit Shenzhen DreamCan Co., Ltd. Tel 0755-89890518

Manufacturer Shenzhen DreamCan Co., Ltd. Sample Grade KN95

Sample Receiving
Sample Quantity 40 pieces March 23, 2020
Date

Commissioned Sample Delivery


Test Category Serial Number
Inspection

Samples
Meet the testing requirements
Conditions

Document and GB 2626-2006 Respiratory Protective Articles Self-suction Filtering Type Particulate Matter
Decide
Accordance Prevention Respirator

The samples were tested, and the tested items met the requirements of KN95 specified in GB 2626-
Test 2006.
Conclusion
Issuing Date: April 1, 2020

The test items marked “----” indicate that there is no single evaluation or that there is no single
evaluation condition. The test conclusion of this report is only obtained for the tested items, which does
Remarks
not mean that the items or functions that have not been tested meet the requirements. This report is
only responsible for incoming samples.

Approver: Chen Min Examiner: Wu Liangliang Majortester: Yao Yongwei


(With the special seal for inspection and test of Jiangsu Quality Supervision and Inspection Center
for Special Safety Protection Products)
Testing Results
STFWT20205970 Page 2 of 5
Serial Individual
Test Items Unit Requirement Results
No. Judgment
Materials:
a) materials in direct contact with the face
should be harmless;
b) filter material should be harmless to human
body;
c) the materials used shall have sufficient
strength and shall not be damaged or
Structure:
deformed during normal service life.
a) It is not easy to cause structural damage, and
Structure:
the design, composition and installation of
a) structural damage shall not occur easily,
components do not cause any danger to users;
and the design, composition and installation
General b) The headband is adjustable in design,
1 ----- of components shall not cause any danger to -----
Requirements convenient to wear and remove, and can firmly
users;
fix the mask on the face without obvious
b) the design of the headband should be
compression and pain during wearing.
adjustable to facilitate wearing and removal,
c) It has smaller dead space and larger field of
the mask should be firmly fixed on the face,
vision.
and there should be no obvious compression
or pain during wearing.
c) should be as small as possible dead space
and larger field of vision;
d) the structure of disposable face mask shall
be able to ensure close fit with face and shall
not deform during its service life.
The sample surface should not be damaged,
deformed or have obvious other defects. The
material and structure of the components
Appearance
2 ----- should be able to withstand normal use Meet the requirements Pass
Check
conditions and possible temperature,
humidity and mechanical impact. The
headband should be adjustable.
Measured
Sample No.
Value
Not 21# 92.1
Total
The total inspiratory resistance of each Pretreated 22# 90.4
Inspiratory
Breathing sample should not be greater than 350 and the 25# 89.4
3 Pa Resistance Pretreated Pass
Resistance total expiratory resistance should not be 26# 87.6
greater than 250. Not 23# 72.6
Total
Pretreated 24# 71.7
Expiratory
27# 70.4
Resistance Pretreated
28# 70.6
Field of
4 ---- Lower visual field≥60° 62° Pass
Vision
The volume fraction of carbon dioxide shall
5 Dead Space ---- The volume fraction of carbon dioxide is 0.6% Pass
not exceed 1%
Testing Results
STFWT20205970 Page 3 of 5
Serial Individual
Test Items Unit Requirement Results
No. Judgment
Sample No. Measured Value
Initial 99.0
1#
Additional 98.7
Initial 99.2
2#
Additional 99.0
Initial 99.4
3#
Additional 99.2
Initial 99.2
4#
Additional 99.0
Initial 99.0
Not 5#
Additional 98.7
pretreated
Initial 99.4
6#
Additional 99.1
Filtration Initial 99.2
7#
Efficiency /% Additional 98.9
6 (NaCl ----- KN95:≥95.0 Initial 99.4 Pass
8#
Particulate Additional 99.1
Matter) Initial 99.1
9#
Additional 98.7
Initial 99.2
10#
Additional 99.0
Initial 98.9
11#
Additional 98.7
Initial 98.8
12#
Temperature Additional 98.6
and Initial 99.0
13#
humidity Additional 98.6
treatment Initial 98.7
14#
Additional 98.5
Initial 98.8
15#
Additional 98.6
Not pretreated:
Each headband, buckle and other adjusting
Each headband, buckle and other adjusting
components of 31# mask did not slip or break after
components of the mask shall not slip or
7 Headband ---- 10s. Pass
break when subjected to 10N tension for
Each headband, buckle and other adjusting
10s.
components of the 32# mask did not slip or break
after withstanding 10N tension for 10s.
Testing Results
STFWT20205970 Page 4 of 5
Serial Individual
Test Items Unit Requirement Results
No. Judgment
Sample No. Measured Value
47 actions’ TIL is
Based on the TIL of each 16#
Filter less than 11
action (i.e. 10 people X5
Material 47 actions’ TIL is
actions), at least 46 of the 17#
Grade less than 11
50 actions
Not pretreated 47 actions’ TIL is
18#
less than 11

47 actions’ TIL is
19#
less than 11
47 actions’ TIL is
20#
less than 11
47 actions’ TIL is
21#
less than 11
KN95 <11 47 actions’ TIL is
22#
less than 11
Temperature
47 actions’ TIL is
and humidity 23#
less than 11
treatment
47 actions’ TIL is
24#
Leakage/% less than 11
(Disposable 47 actions’ TIL is
8 ----- 25# Pass
Mask less than 11
TIL) 9 people’s TIL is less
Filter Based on the total TIL of 16#
Not pretreated than 8
Material people, at least 8 of 10
9 people’s TIL is less
Grade subjects 17#
than 8
9 people’s TIL is less
18#
than 8
9 people’s TIL is less
19#
than 8
9 people’s TIL is less
KN95 <8 20#
than 8
9 people’s TIL is less
Temperature 21#
than 8
and humidity
9 people’s TIL is less
treatment 22#
than 8
9 people’s TIL is less
23#
than 8
9 people’s TIL is less
24#
than 8
9 people’s TIL is less
25#
than 8
Not pretreated:
Components of 1# and 2# samples exposed to
Components exposed to the flame shall not flame did not burn after being removed from the
burn after being removed from the flame; If flame.
9 Flammability ----- Pass
burning, the duration of continuous burning After pretreatment:
shall not exceed 5s. Components of 3# and 4# samples exposed to
flame did not burn after being removed from the
flame.
Testing Results

STFWT20205970 Page 5 of 5

Picture of Sample

----blank below----
Points For Attention
1. This Report Is Invalid If Without “The Text Report Special Seal” Or The Official
Seal Of The Institute.
2. The Reproduction Is Invalid If Without Being Confirmed By “The Text Report
Special Seal” Or The Official Seal Of The Institute.
3. This Report Is Invalid If Without Signature Of The Major Tester And The Examiner
And The Approver.
4. This Report Is Invalid If In Any Form By Any Means Altered.

The Institute Address: Lingang Road 166, Lingang Economic Park, Gaogang,
Taizhou.Jiangsu
The Institute Complain Tel:0523-86989901
The Institute Businese Tel:0523-86989959
The Institute Fax:0523-86989939
The Institute Post:225300
The Institute Website: www.jstfzx.com
The Institute E-mail: 1735889887@qq.com
TEST REPORT

STFWT20205971

Product Name Daily Protective Mask

Trust Unit Shenzhen DreamCan Co., Ltd.

Manufacturer Shenzhen DreamCan Co., Ltd.

Test Category Commissioned Sample Delivery Inspection

JIANGSU QUALITY SUPERVISION AND INSPECTION CENTER FOR SPECIAL SAFETY PROTECTION PRODUCTS

Test Report
STFWT20205971 Page 1 of 4
Daily Protective Mask Specification Type DP-A-21L-K
Product Name
Trademark -----

Trust Unit Shenzhen DreamCan Co., Ltd. Tel 0755-89890518

Manufacturer Shenzhen DreamCan Co., Ltd. Sample Grade Grade C

Sample Receiving
Sample Quantity 50 pieces March 23, 2020
Date

Commissioned Sample Delivery


Test Category Serial Number -------
Inspection

Samples
Meet the testing requirements
Conditions

Document and
Decide GB/T 32610-2016 Technical Specification for Daily Protective Mask
Accordance

The samples were tested, and the tested items meet the requirements of grade C protection effect
Test specified in GB/T32610-2016.
Conclusion
Issuing Date: April 1, 2020

Microbiological index test data are provided by Taizhou Product Quality Supervision and Inspection
Institute. CMA: 151020110122 The test conclusion of this report is only obtained for the tested
Remarks
items, which does not mean that the items or functions not tested meet the requirements.
This report is only responsible for incoming samples.

Approver: Chen Min Examiner: Wu Liangliang Majortester: Yao Yongwei


(With the special seal for inspection and test of Jiangsu Quality Supervision and Inspection Center
for Special Safety Protection Products)
Testing Results
STFWT20205971 Page 2 of 4
Serial Individual
Test Items Unit Requirement Results
No. Judgment
4-Aminobiphenyl Should be forbidden
Benzidine Should be forbidden
4- Chloro-o-methylaniline Should be forbidden
2- Naphthylamine Should be forbidden
Ortho-Aminoazotoluene Should be forbidden
5- Nitro-o-toluidine Should be forbidden
P-chloroaniline Should be forbidden
2,4- Diaminoanisole Should be forbidden
4,4'- Diaminodiphenylmethane Should be forbidden
3,3'- Dichlorobenzidine Should be forbidden
3,3'- Dimethoxybenzidine Should be forbidden
Only the dyeing and
Decomposable 3,3'-Dimethylaniline Should be forbidden
printing parts will
carcinogenic 3,3'- Dimethyl -4,4'-
1 mg/kg Should be forbidden be tested, and this -----
aromatic amine Diaminodiphenylmethane
item will not be
dyes 2- Methoxy -5- Methylaniline Should be forbidden
tested.
4,4'-Methylene-bis -(2-
Should be forbidden
Chloroaniline)
4,4-Diaminodiphenyl ether Should be forbidden
4,4'-Diaminodiphenyl sulfide Should be forbidden
Ortho-toluidine Should be forbidden
2,4-Diaminotoluene Should be forbidden
2,4,5-Trimethylaniline Should be forbidden
O-aminoanisole Should be forbidden
2,6-Dimethylaniline Should be forbidden
2,4- Dimethylaniline Should be forbidden
4-Aminoazobenzene Should be forbidden
Testing Results
STFWT20205971 Page 3 of 4
Serial Individual
Test Items Unit Requirement Results
No. Judgment
Ten masks were selected and
There should be no damage, oil inspected by visual inspection:
2 Appearance Requirements ---- stain, deformation and other obvious the mask surface was free from Pass
defects on the mask surface. damage, oil stain, deformation
and other obvious defects.

Sample No. Measured


Value
3 Inspiratory Resistance Pa ≤175 Not 21# 110.2 Pass
pretreated 22# 104.2
25# 107.4
Pretreated
26# 106.2
Not 23# 82.1
pretreated 24# 81.9
4 Expiratory Resistance Pa ≤145 Pass
27# 74.7
Pretreated
28# 76.1
The breaking strength of the
5 mask belt and the joint between N ≥20 27 Pass
the mask belt and the mask body
6 Mask lower field of vision ---- ≥60° 62° Pass
Salt media:
Class III: ≥90 Salt media: 99.0
7 Filtration Efficiency/% ---- Pass
Oily media: Oily media: 91.0
Class III: ≥80
Grade C: ≥75
When the protective effect level of
the mask is grade A, the filtration
efficiency should reach class II and
8 Protective Effect/% ---- above; When the protective effect 80.4 Pass
level of the mask is grade B, grade
C and grade D, the filtration
efficiency should reach class III and
above.
Slippage, breakage and deformation No exhalation valve, this item
9 Breathing Valve Cover Fastness ---- ----
shall not occur. shall not be detected.
Testing Results
STFWT20205971 Page 4 of 4
Individual
Serial No. Test Items Unit Requirement Results
Judgment

Color Fastness Dry rubbing Only the dyeing and printing


10 to ≥4 parts will be tested, and this item ----
Rubbing/Grade Wet rubbing will not be tested.

11 PH value ---- 4.0~8.5 6.4 Pass


12 Formaldehyde content mg/kg ≤20 Not found Pass
This item is not marked as it has
13 Ethylene oxide residue μg/g ≤10 been treated by ethylene oxide, ----
and is not tested.
Total bacterial colonies CFU/g ≤200 <20
fail to test not found
Coliform group ----
fail to test not found
Staphylococcus
Microbial ---- fail to test not found Pass
14 aureus
index Pathogenic
Pseudomonas
pyogenic ---- fail to test not found
aeruginosa
bacteria
Hemolytic
---- fail to test not found
streptococcus
Total fungal colonies CFU/g ≤100 <20
Remarks Formaldehyde results are less than 20mg/kg, and the test results report “not found”

Picture of sample

---blank below---
Points For Attention
1. This Report Is Invalid If Without “The Text Report Special Seal” Or The Official
Seal Of The Institute.
2. The Reproduction Is Invalid If Without Being Confirmed By “The Text Report
Special Seal” Or The Official Seal Of The Institute.
3. This Report Is Invalid If Without Signature Of The Major Tester And The Examiner
And The Approver.
4. This Report Is Invalid If In Any Form By Any Means Altered.

The Institute Address: Lingang Road 166, Lingang Economic Park, Gaogang,
Taizhou.Jiangsu
The Institute Complain Tel:0523-86989901
The Institute Businese Tel:0523-86989959
The Institute Fax:0523-86989939
The Institute Post:225300
The Institute Website: www.jstfzx.com
The Institute E-mail: 1735889887@qq.com