Sunteți pe pagina 1din 19

Electronic Gas Volume Corrector

Basics and Installation Examples


Introduction

Electronic gas volume correctors are generally


battery powered, microprocessor based instruments
used in the natural gas industry to convert the
volume of natural gas as measured from gas meters,
to corrected volume. During the conversion
process, the electronic corrector obtains
measurements of the meter’s gas pressure and
temperature so pressure and temperature factors can Figure 1
be computed and applied to the uncorrected volume. Output shaft and drive-dog of I.D type meter
Another factor, called the “Supercompressibility
Factor” is also computed and applied. Other In many correctors, the input shaft also drives a
functions such as scaling various parameters to the magnet disc assembly. As the magnet on the disc
desired units of measure, alarm detection, alarm rotates past a stationary, magnetically actuated reed
notification, pulse output, digital readouts, and user switch, an electronic pulse is created. Other input
interface, are some of the necessary functions of assemblies may incorporate hall-effect, opto
today’s electronic correctors. sensors, or some other detector, but the purpose
remains the same, to detect a complete (or partial)
Detecting Uncorrected Volume from a Gas meter rotation.
Meter
When compared to the older mechanical
Most volume correctors are designed for
correctors, electronic correctors generally offer very
installation on meters having an instrument drive
little torque loading to the meter. This is a desirable
(I.D.) facility. The correctors are bolted to the meter
feature since an additional load to the meter’s
mounting plate using appropriate bolts, gaskets and
output will reduce its accuracy.
adapter plates where required. The mounting holes
in the meters are pre-drilled and tapped to provide
correct alignment to the meter’s rotating output
shaft.

Through gears, shafts, cables and other


mechanical components, the rotation of the meter’s
output shaft is coupled to the corrector’s input shaft,
which in-turn drives the input to a mechanical
index. Since the mechanical index counts upward
in one direction only, provisions for a reversing
mechanism is needed to provide means of selecting
either clockwise or counter-clockwise rotation.
When properly configured, forward flow through
the gas meter results in an increasing count on the
mechanical index.
Figure 2
Mechanical index for recording meter revolutions,
scaled to uncorrected volume
1
Recording Uncorrected Volume

Often, the input assembly makes use of two


stationary reed switches. The dual input switches
allow for continued operation, even if one switch
should fail. In the two-switch arrangement, even
though both switches are activated during one
complete meter revolution, it only provides a count
of one for that revolution.

Figure 4
Door-mounted LCD showing the electronic
uncorrected volume reading

Obtaining Analog Measurements

Electronic volume correctors are generally self-


contained, battery-powered instruments, unless high
current-consumption functions are added. These
additional functions may then require the use of an
Figure 3 external power supply. Pressure and temperature
Corrector’s input shaft with magnet disc assembly transducers are integral components of the
and input switchboard electronic volume corrector. The circuitry
associated with pressure and temperature
When meter rotation is detected, the electronic transducers is normally in an un-powered, standby
pulse provided needs to be assigned a certain mode. When the magnet disc rotates, and actuates
amount of volume per rotation. Generally, each the input switches, the analog circuits are energized,
pulse represents the volume of that particular which begins a correction cycle. However, some
meter’s drive rate, which is usually stamped onto its correctors ‘wake-up’ based on time, not meter
serial plate. This value is then programmed into the revolutions. In such correctors, the time between
appropriate parameter during corrector setup. correction cycles is a parameter programmed by the
Values for this parameter usually range from 1, 5, user, with 30 minutes being a typical value.
10, 100 or 1000 cubic feet, or 1 or 10 cubic meters.
Once the corrector 'wakes-up', the operating
With the electronic pulse now identified as code in firmware instructs the microprocessor to
volume, each incremental value will be added to a obtain analog measurements. Typical analog
memory register recording the totalized uncorrected signals are: gas pressure, gas temperature, battery
volume. However, the volume may be re-scaled to voltage and case temperature. These voltages are
a different unit of measure and may be displayed up multiplexed through an Analog to Digital (A/D)
to the maximum number of digits. For example, the converter and sent to the microprocessor. The
uncorrected volume from a 10-cubic foot meter may microprocessor converts the digitized analog
be displayed as “CCF” (cubic feet X 100) with a voltages to an equivalent numeric value and stores
maximum of six digits. Generally, the setup this digital information in memory. The values are
parameter that controls the maximum numbers of then scaled to proper units of measure.
uncorrected volume digits will range from four to
eight.
2
reading is then scaled and ready for computation of
the Pressure Correction Factor

Determining Pressure Factor


(Boyle’s Law)

Since natural gas is a compressible fluid, the


pressure applied to the gas is an important
consideration in determining the number of gas
molecules (volume) contained in a closed vessel.
Boyles’s Law defines the relationship between gas
pressure and metered volume, and is expressed as:
Fp = Pm  Pb
Figure 5
Various strain gauge pressure transducers
Where:
Fp = Pressure Factor
Pressure Measurement System
Pm = Absolute Meter Pressure (psia)
The pressure measuring system commonly Pb = Absolute Base Pressure (psia)
incorporates a strain gauge pressure transducer. The
transducer is located on the inside of the corrector’s Since Boyle’s Law requires the use of absolute
housing with a 1/4" NPT female connector pressures, pressure readings from an absolute
somewhere on the outside of the housing. A multi- transducer can be used directly, however, readings
conductor cable from the transducer plugs into, or for a gauge-type transducer must be converted to
attaches to, the main circuit board to provide the absolute pressure. Below is the formula to convert
analog pressure signal. gauge pressure to absolute pressure.
P(abs) = P(g) + P(atmos)
The analog signal from the transducer is
proportional to the applied pressure. As an Where:
example, with a 0-100 psig transducer that produces P(abs) = Absolute Pressure (psia)
a 0 - 1.0-volt output, when 100.0 pounds of pressure P(g) = Pressure reading from transducer (psig)
is applied, the output voltage will be 1.0 volt. If P(atmos) = Atmospheric Pressure (psia)
50.0 pounds of pressure is applied, then the output
will be 0.5 volts. Of course, with no pressure Once calculated, the Pressure Factor is used as
applied, the output will be zero volts. a multiplier on the uncorrected volume to produce
Pressure Corrected Volume as defined below.
As with most transducers, ambient temperature Pcorvol = UncVol  Fp
can affect the accuracy of the output. For that
reason, the pressure measuring system includes Where:
some type temperature compensation. A common Pcorvol = Pressure Corrected Volume
method is to store a temperature response table in UncVol = Uncorrected Volume
the corrector’s memory for a specific transducer. Fp = Pressure Correction Factor
Temperature readings from the Case Temperature
Sensor are correlated to the temperature response Gas Temperature System
table, which is then used to offset the temperature
affects. The result is an accurate pressure The temperature system usually consists of a
measurement over the entire operating temperature temperature sensor located inside of ¼” diameter,
range. The temperature-compensated pressure stainless-steel probe. Probe lengths of 4 to 12
inches are common, but there are some applications
3
requiring shorter, smaller diameter probes. The temperature effects metered volume, and is
sensor itself may be either a solid-state thermistor or expressed as:
an RTD. The main characteristics of the two Ft = Tb  Tm
sensors are:
Thermistor, Where:
 is a 2-wire, semiconductor device Ft = Temperature Factor
 has a negative temperature coefficient Tb = Absolute Base Temperature
 produces a non-linear output Tm = Absolute Gas Temperature
 typically provides 10,000 ohms @ 25 C
RTD (Resistive Temperature Device), Since Charles’ Law requires the use of absolute
 is a 3 or 4-wire platinum wire device temperatures, temperature readings from the
 has a positive temperature coefficient corrector must be converted to absolute values.
 produces a linear output Below is the formula to convert Fahrenheit to
 typically provides 100 ohms @ 25 C absolute values.
T(abs) = T F + 459.67

Where:
T(abs) = Absolute Temperature
T F = Gas Temperature in F
459.67 = Conversion Factor

Once calculated, the Temperature Factor is


used as a multiplier on the uncorrected volume to
produce Temperature Corrected Volume as defined
below.
Figure 6 Tcorvol = UncVol  Ft
Temperature probes of varying lengths and
diameters Where:
Tcorvol = Temperature Corrected Volume
Regardless of which sensor is used, when UncVol = Uncorrected Volume
connected to the temperature circuitry, both provide Ft = Temperature Correction Factor
a DC voltage that varies with gas temperature. As
with the pressure signal, the temperature signal Determining Supercompressibility Factor (per
must also be converted to it’s digital equivalent for NX-19 and AGA8)
further processing. This includes scaling the values
to the selected unit of measure (such as Fahrenheit Boyle’s Law and Charles’ Law define the
or Celsius) and applying linearization correction relationship of pressure and temperature to gas
where required. After the temperature signal is volume. These definitions assume an ideal gas, but
measured, digitized and scaled, it is ready for in the ‘real world’, natural gas actually compresses
computation of the Temperature Correction Factor more than these laws state. To accommodate for
this extra bit of gas compression, an additional
Determining Temperature Factor factor is implemented in volume correctors. This
(Charles’ Law) third factor is referred to as Supercompressibility.

Since the temperature of the gas also affects gas


volume calculations, it’s important to understand
this relationship. Charles’ Law defines how gas

4
Supercompressibility compensation in volume system) or provided from a gas chromatograph (gas
correctors are usually based on one of three analyzer) from a sample of the supply gas. Since
methods: the percentages of the individual composition
 AGA3 (using the NX-19 Table) parameters can very, it’s common to update the
 AGA8 (Gross) values periodically.
 AGA8 (Detailed)
AGA3 (American Gas Association Publication Corrected Volume
#3) is by far the most common method of providing
Supercompressibility correction in volume In the previous sections, it was explained how
correctors. To interpolate values from the AGA3, volume correctors detect and record uncorrected
NX-19 tables, gas pressure, gas temperature, volume. Also explained was how correctors
specific gravity, percent Nitrogen and percent measure gas pressure and gas temperature, and how
Carbon Dioxide are entered into a complex to compute pressure factor, temperature factor and
algorithm. The resulting value must then be the Supercompressibility factor. To compute
squared to produce the Supercompressibility Factor, Corrected Volume, it is a simple matter of
(Fpv). multiplying the Uncorrected Volume (detected on
each meter revolution) by the Pressure, Temperature
Some volume correctors offer AGA8 (Gross and Supercompressibility Factors, calculated on the
method) Supercompressibility. This newer method same meter revolution. The result is the amount of
is considered to be more accurate than the NX-19 corrected volume for that meter revolution as
tables, but is limited down to only 32F. defined below.
CorVol = UncVol  Fp Ft Fpv
Two computational methods are available when
configuring the correctors for AGA8 (Gross), Where:
Relative Density and Heating Value. When setting CorVol = Corrected Volume
up the corrector for the Relative Density method, UncVol = Uncorrected Volume
specific gravity, percent Nitrogen and percent Fp = Pressure Correction Factor
Carbon Dioxide are entered. In the Heating Value Ft = Temperature Correction Factor
method, specific gravity, percent Carbon Dioxide Fpv = Supercompressibility Correction Factor
and the gas Heat Value (BTU) are entered.
The corrected volume value (per revolution) is
Very few volume correctors offer the third then added to the memory register that keeps a
method of computing Supercompressibility, AGA8 running total of all corrected volumes computed
(Detailed). The Detailed method has long been from all previous meter revolutions.
used in Orifice meter applications, but is relatively
new to volume correctors. In this method, many
more gas composition parameters must be entered
into the corrector. The parameters usually consist
of; % Methane, % Nitrogen, % Carbon Dioxide, %
Ethane, % Propane, % Water, % Hydrogen Sulfide,
% Hydrogen, % Carbon Monoxide, % Oxygen, %
Iso-Butane, % N-Butane, % Iso-Pentane, % N-
Pentane, % N-Hexane, % N-Heptane, % N-Octane,
% N-Nonane, % N-Decane, % Helium and %
Argon.
As with all Supercompressibility methods, the
individual gas composition values are either Figure 7
assumed (based on history or knowledge of the gas Corrected Volume displayed on LCD
5
Since most volume correctors incorporate some
type of digital display (an LCD being the most
common), the totalized corrected volume is readily
viewable. Corrector configuration allows users to
‘customize’ the corrected volume reading by
offering choices for the total number of digits,
usually ranging from four to eight digits, and to
allow selection of the unit of measure.

The most common corrected volume units are


“CCF” and “MCF” for Imperial units, or “m3” and
“m3 x 10” for metric.

Where:
CCF represents Cubic Feet X 100
MCF represents Cubic Feet X 1000
m3 represents Cubic Meters Figure 8
m3 X 10 represents Cubic Meters X 10 Various battery packs used electronic volume
correctors
Depending on the particular corrector, other
volume units may also be available for selection. Lithium cells provide more stored energy than
an alkaline cell of the same physical size, but they
Instrument Power (battery, solar, AC, etc) are also more expensive. Lithium cells also output
over twice the voltage of an alkaline cell, so it’s
In general, today’s volume correctors are usually not possible to simply insert lithium cells
designed to be battery powered. Especially since into a battery pack designed for alkaline batteries.
the newer circuit boards make use of low-power Regardless of the battery type, be sure to use
consumption components. There are a variety of only those battery packs specified by the corrector
battery options that include, but are not limited to manufacturer. Since most volume correctors have
alkaline, lithium, gel-cell, lead-acid and nickel- some type of safety approval, battery packs with
cadium. The last three are rechargeable batteries, part numbers listed on the manufacturer’s control
usually requiring an external power source such as drawings must be used to conform to the approved
an AC Power Supply or a Solar Panel charging installations into hazardous locations.
system.
Instrument Configuration
For the most part, an alkaline battery pack is (programming of items)
provided as the corrector’s standard power supply.
The reliability, longevity and shelf life of alkaline Electronic volume correctors are dedicated
batteries have increased over the years to the point computers, and like standard computers, correctors
where, a multi-cell alkaline battery pack can provide can be programmed to fit the needs of the user.
years of service. However, if options such as Electronic correctors have at least one and probably
modems are installed, it might be necessary to several types of memory. Certain types of memory
substitute a lithium battery pack to provide longer are better suited to contain the operating system,
service. sometimes referred to as firmware. Firmware is the
operating code that makes this dedicated computer a
volume corrector. In some correctors, the firmware
may be replaced (upgraded) to a newer version,
while others incorporate OTP (One Time Program)
6
microprocessors. In these cases, firmware upgrades Where possible, it’s a common practice for gas
are not possible and any feature enhancements or companies to configure most (or all) correctors to be
bug fixes require the replacement of the processor nearly identical. Some items, such as Meter Index
board. Rate and Pressure Alarm Limits need to be site
specific, but others, like Corrected Volume Units
While the corrector is operational, i.e. running may be the same for all correctors. It’s really up to
the operating code (firmware), it becomes necessary the individual gas companies to determine the best
to store information, either temporarily or long- configuration of the corrector. The goal of course,
term. RAM (Random Access Memory) or is to provide the best information in the format that
EEPROM (Electronically Erasable Programmable best fits their particular needs.
Read Only Memory) are often used for this type of
storage. Firmware creates memory locations for The specific manner in which correctors are
storing various types of information. These configured can vary from one manufacturer to the
locations may be referred to as Registers, Memory next. Some older correctors provided an integral or
Registers, ID Codes, Item Codes or some other detachable keypad for on-site configuration and
similar name. calibration. Others use DIP switches, trim-pots or
toggle switches to select or adjust certain functions.
A group of memory registers can then be set However, most current day correctors provide
aside to receive information from the user on how to instrument configuration via a serial data port.
customize a particular corrector for the needs of a Laptop computers or some other hand-held devices
specific installation. This group of items is often are now the most common method of accessing and
referred to as Configuration Items. changing instrument configuration.

Listed below are common Configuration Items Instrument Calibration


available in most correctors.
 Corrected volume units As mentioned, keypads, DIP switches, trim-
 Number of Corrected volume digits pots and toggle switches were often used to
 Uncorrected volume units configure and calibrate older correctors. Today,
 Number of Uncorrected volume digits pressure and temperature transducers are calibrated
 Pressure units using a laptop computer, running companion
 Temperature units software provided by the corrector manufacturer.
 Base pressure The corrector’s serial port is connected to the
 Base temperature laptop’s serial port to provide a data link. The
software is written to provide a very simple two-
 Atmospheric pressure
step calibration process. Step-one (Zero
 Specific gravity
Calibration) for both pressure and temperature
 % Nitrogen
calibration, is to apply a low-reading condition to
 % Carbon Dioxide the transducer. For pressure calibration, the most
 Pressure Low alarm limit convenient zero reference is to just open the
 Pressure High alarm limit transducer inlet to atmosphere. For temperature
 Temperature Low alarm limit systems, a common practice is to insert the
 Temperature High alarm limit temperature probe into an agitated ice bath that
 Battery Low alarm limit produces a reading close to 32F (0C).
 Meter index rate

7
Data Logging (audit trail, event log, etc.),
purpose and examples

The primary purpose of an electronic volume


corrector is to calculate and display an accurate
running total of corrected volume. However, in
many correctors, a secondary function is to record
log-data of the meter site. The resulting information
has various names, including, Load Profile, Load
Survey, Archive Data, Historical Data, Audit Trail
Data, to name a few. The logged data is often
Figure 9 stored in RAM or EEPROM memory and the
Example of one manufacturer’s keypad used to amount of stored data is limited by the total amount
configure and calibrate the corrector of memory made available for this function.
Storage capacities vary, but one to many months of
When the reading has stabilized, the software hourly data is common.
obtains several sample readings via the serial port.
The samples are averaged and then compared to the Log-data is the accumulation of flow
zero-reference value to determine if an adjustment information recorded at regular time intervals. The
to the zero reading is required. The software sends time intervals are usually every one-hour or every
new calibration information through the serial port one-day (or both), however, other intervals such as
to specific memory registers in the corrector to 5-minute, 15-minute and 30-minute intervals may
make the offset adjustments. After the offset also be offered.
correction is applied, the corrector will then provide
correct readings for ‘zero’ conditions. When configured for hourly logging, the time
intervals start at the beginning of each new hour and
The corrector provides a linear output for both end at the beginning of the next hour. Therefore,
pressure and temperature measurements. Since it each day of audit trail data is comprised of 24
takes two points to define a straight line, a records. Conversely, each day of daily logging will
minimum of two calibration points is required. The only produce one record.
first calibration point was obtained during Zero The number of log-data fields, and the data
Calibration in the previous paragraph. The second contained within the data fields can vary, depending
point is obtained during Span Calibration. During on the corrector’s manufacturer. However, it’s
Span Calibration, the transducer is subjected to an quite common to provide at least six data fields
upscale reading. After the reading has stabilized, containing the following information.
the software obtains a few sample readings. The
samples are averaged and compared to an accurate Data Field #1: Date
calibration reference. Span adjustments are then Data Field #2: Time
sent to the corrector via the serial port and stored in Data Field #3 Interval Corrector Volume
specific memory registers. After Zero and Span Data Field #4: Interval Uncorrected Volume
calibrations are completed, the corrector will Data Field #5: Interval Average Pressure
provide accurate readings across the entire Data Field #6: Interval Average Temperature
transducer range.

8
Some volume correctors allow as many as Below is a sample of an abbreviated Event Log
twelve data fields per record and even permit the
user to select the type of data to be recorded in each 01/01/03 08:11:27 0 Base Press 14.73 14.65
data field. In general, the first six data fields can be 01/01/03 08:11:29 1 Sp. Gr. 0.600 0.585
considered as ‘billing data’ while the remaining 01/01/03 08:11:34 2 % N2 0.000 1.500
fields may be used for troubleshooting or other 01/01/03 08:11:37 3 % CO2 0.000 2.700
specific functions. Below is an example of hourly 01/01/03 08:11:40 4 CorVol units CCF MCF
log-data in a tabular listing. 01/01/03 08:11:42 5 UncVol units CCF CCF

#1 #2 #3 #4 #5 #6 After reviewing the above information, an


01/01/03 08:00:00 21 4 61.91 47.32 obvious use of this data would be to determine the
01/01/03 09:00:00 26 5 61.78 47.12 exact data and time an item value was changed.
01/01/03 10:00:00 21 4 62.04 47.44 This could be useful in billing disputes if the Event
01/01/03 11:00:00 22 4 62.22 47.37 Log contains the change that caused the dispute.
01/01/03 12:00:00 16 3 63.55 47.32 For example, if a corrector is installed with
01/01/03 13:00:00 27 5 61.55 47.82 incorrect volume units (such as CCF) and was then
later changed to the correct units (MCF), a billing
The following is an example of daily log-data in a adjustment can be justified since the time of day (to
tabular listing. the second) the change was made is recorded.

#1 #2 #3 #4 #5 #6 Alarm Detection
01/01/03 09:00:00 84 16 61.91 47.32
01/02/03 09:00:00 104 20 61.78 47.12 As described in the section “Obtaining Analog
01/03/03 09:00:00 85 16 62.04 47.44 Measurements” the corrector reads pressure,
01/04/03 09:00:00 88 16 62.22 47.37 temperature and battery voltages. These analog
01/05/03 09:00:00 64 12 63.55 47.32 measurements are multiplexed through an Analog to
01/06/03 09:00:00 108 20 61.55 47.82 Digital (A/D) converter and sent to the
microprocessor. The microprocessor converts the
A second type of log-data is the Event Log. digitized analog voltages to an equivalent numeric
The Event Log is usually a separate memory value and stores this digital information in memory.
location, apart from the log-data described above The values are then scaled to proper units of
but may also reside in RAM or EEPROM. The measure that can be compared to stored alarm limits
purpose of the Event Log is to record changes made to produce a variety of instrument alarms.
to the corrector’s configuration. An Event Log
record is created when a configuration change is Typical alarms from analog data may include,
made. Date and time fields are included in each but not limited to:
Event Log record to time-stamp when the changes
were made. In addition to date and time, other  Pressure High Alarm
parameters included with each record are: a  Pressure Low Alarm
sequential event number (starting at zero), the name  Temperature High Alarm
of the changed item, the item’s old value and the  Temperature Low Alarm
item’s new value.  Battery Voltage Low Alarm
 Back-up Battery Voltage Low Alarm

9
Serial Communications
As mentioned earlier, since the volume corrector
is really a dedicated computer, a variety of Transferring data to or from the corrector is a
parameters are also calculated and stored at their very important and necessary function. Most
specific memory register. These calculated electronic volume correctors provide this function
parameters could also be compared to user-defined via a serial data port. The data port is usually
alarm limits to produce their own alarms. externally mounted for ease of access, but some
correctors may have the connection inside the
Typical alarms from calculated data may corrector housing for security or other reasons.
include, but not be limited to:
Serial data communications in volume
 Daily CorVol Limit Exceeded correctors usually implement the RS-232C protocol
 Flow Rate High Limit Exceeded standards. This protocol deals with parameters such
 Flow Rate Low Limit Exceeded as: voltage levels of the data signals, number of data
bits, number of stop bits, parity, baud rate, etc. Of
Volume correctors can also produce a variety of these, baud rate is the most common parameter the
other alarms that would be specific to that model of user will configure (either in the corrector or in the
corrector. These types of alarms may include laptop) to obtain a communication link.
problems (or faults) with input pulsers, output
pulsers, A/D circuitry, memory, and others. While the laptop is communicating with the
corrector, another type of protocol is implemented
Pulse Output and may be referred to as the corrector’s ‘native’
protocol. The instrument’s native protocol is a
Most electronic volume correctors offer as a collection of rules that must be followed by the
standard feature, the ability to send corrected and corrector and the laptop (via host software) to
uncorrected volume pulses to a remote receiver. properly exchange data through the serial link.
The purpose of remote receivers is to collect and These ’rules’ are a collection of commands, control
store volume data for an AMR (Automatic Meter characters, alphanumeric characters, etc. that must
Reading) system or to display a remote corrected be assembled in proper formats. The corrector’s
volume reading. In both cases, the pulses are native protocol is generally used to configure,
configured (through instrument item codes) to a calibrate and extract log-data from the instrument.
specific amount of volume per pulse. In general,
the value of the pulse is the same volume unit as the Modbus is probably the most familiar serial
corrected volume on the LCD. However, pulse protocol used in the natural gas industry and is
scaling is usually adjustable to provide higher generally used by the host computer system
resolution (more pulses) if required. (SCADA software) to obtain remote readings. But
even Modbus ‘rules’ can vary from one system to
The transmission of Form-A pulses (2-wire another since there are many variants. Two major
connection) is more common than Form-C pulses variations within Modbus are ASCII and RTU
(3-wire connection). Some correctors offer the (binary).
ability to use both but either type requires a
‘wetting’ voltage provided by the receiving device. Regardless of the serial protocol used, the
When transmitting pulses to a pulse receiver, factors corrector must receive properly formatted
such as type of cable (shielded or not shielded), wire commands to return proper responses. Most serial
size, length of cable, number of conductors and use protocols incorporate some type of error checking to
of safety barriers must be considered. prevent the acceptance of corrupted data. Error
checking is a necessity when transmitting data over
long distances, especially over radio and modem
10
links. In those cases, where corruption is detected, a Traditional Meter-Mounted Correctors,
retry scheme is attempted. If data corruption is Examples
severe enough, the serial session (data link) is
terminated and a call-back can be tried at a later Photographs in Figures 11-19 are representative
time. of meter installations using the traditional
corrector-to-meter mounting. At these installations,
The protocols used in volume correctors have the input signal for uncorrected volume is
varying degrees of effective error checking. Some developed from a reed switch that produces one
protocols have next to no error checking while some switch closure per meter revolution. A rotating
have very robust error checking. Two common magnet, driven by the meter’s output shaft, actuates
types of error checking implemented in the the reed switch. The corrector’s input shaft also
corrector’s native protocol are LRC (Longitudinal drives a mechanical uncorrected volume index, as
Redundancy Check) and CRC-16 (Cyclical illustrated in the simplified drawing of figure 10.
Redundancy Check).

Corrector Mounting and Interfacing


(traditional and non-traditional)

Over the past few decades, most volume


correctors have been installed on I.D. type meters,
i.e., those meters with a rotating output shaft.
For the most part, these are still the most
common or traditional type installations. However,
in recent years, non-traditional types of meter
mounting and meter interfacing have become more
visible.

Figure 10
Simplified diagram of a volume corrector for
traditional meter installations. These correctors
generally include a mechanical index and an
internally generated, uncorrected volume signal.

11
Figure 11
Dresser 5M LMMA rotary meter with meter-
mounted corrector
Figure 13
Rockwell 10,000 diaphragm meter with meter-
mounted corrector

Figure 12 Figure 14
Two Dresser 11M rotary meters with meter- Three Dresser 102M rotary meters with meter-
mounted correctors mounted correctors

12
Figure 16
American AL-5000 diaphragm meter with meter-
mounted pulse transmitter and corrector.
Figure 15
American AL-800 diaphragm meter with meter-
mounted corrector

Figure 17
Two Equimeter 4” turbine meters with meter-
mounted correctors, both wired for Internet access.

13
Figure 18
American 6” turbine meter with meter-mounted
corrector

Figure 19
Equimeter 6” turbine meter with meter-mounted
corrector. The corrector contains an internal
modem and AC-to-DC power supply

14
Non-traditional Corrector Mounting and
Interfacing, Examples

Photographs in Figures 21–36 are


representative of meter installations using non-
traditional meter-to-corrector interfacing.
Generally, the correctors are remote-mounted but in
close proximity to the meters. Uncorrected volume
pulses or switch closures are connected to the
instruments via signal cables. A simplified diagram
of each example (Figures, 20, 25, 28 and 34) is
provided to compare the different interfaces.

Figure 21
American RPM 3.5M rotary meter with meter-
bracket mounted corrector. An internal meter
switch and probe provide uncorrected volume and
temperature signals to the corrector

Figure 20
Simplified diagram a remote switch input for non-
traditional meter installations. These correctors
may be mounted to the meter with a special
mounting bracket, or may be wall or pipe-mounted
near the meter. A low frequency reed switch,
located inside the meter, provides the uncorrected
volume signal to the corrector. This type of
corrector would not normally include a mechanical
uncorrected volume index, however, one may be Figure 22
provided with the meter. Romet 11000 rotary meter with pipe-mounted
corrector. The uncorrected volume signal is
provided via a switch, internal to the meter’s
mechanical index.

15
Figure 25
Simplified diagram of a low frequency input for
non-traditional meter installations. These
Figure 23 correctors are usually wall or pipe-mounted near the
Instromet 11M rotary meter with pipe-mounted meter. A low frequency pulser provides the
corrector and cellular communications. The uncorrected volume signal to the corrector.
uncorrected volume signal is provided via a switch,
internal to the meter’s mechanical index.

Figure 24 Figure 26
Instromet 2M rotary meter with bracket-mounted RMG turbine meter with bracket-mounted
corrector. The uncorrected volume signal is corrector. The meter’s low-frequency pulse output
provided via a switch, internal to the meter’s assembly provides the uncorrected volume signal.
mechanical index

16
Figure 28
Simplified diagram of a high frequency input for
non-traditional meter installations. These
correctors can be wall-mounted, pipe-mounted or
meter-bracket mounted. The corrector scales the
input signal to provide a compatible low frequency
output.

Figure 27
Schlumberger Dattus fluidic oscillation meter with
wall-mounted corrector. The meter’s low-
frequency pulse output assembly provides the
uncorrected volume signal.

Figure 29
Daniel 8” turbine meter with bracket-mounted
corrector. The meter’s high-frequency blade-tip
sensors provide the uncorrected volume signal.

17
Figure 32
Two Dresser 1.5M LMMA rotary meters with
integral-corrector. The corrector’s high-frequency
magnetic sensors provide the uncorrected volume
Figure 30 signal.
Instromet Ultrasonic meter with pipe-mounted
corrector, providing fully electronic metering and
volume correction, i.e. no moving parts. The
meter’s processor unit provides a high frequency
output as the uncorrected volume signal. Although
the output frequency range is adjustable, this
configuration provides 5000 hertz at max. flow rate.
The corrector scales the input frequency to a
compatible low frequency signal.

Figure 33
Dresser 5M rotary meter with integral-corrector.
This installation is a retrofit of the corrector in
Figure 31 photo 11. The corrector’s high-frequency magnetic
Dresser 3M LMMA rotary meter with integral- sensors provide the uncorrected volume signal.
corrector. The corrector’s high-frequency magnetic
sensors provide the uncorrected volume signal.

18
Figure 34
Simplified diagram of a dual, high frequency input
for non-traditional meter installations. These
correctors receive two high frequency signals,
which are processed using proprietary algorithms
supplied by the meter’s manufacturer. A variety of
outputs are computed including a low frequency
pulse that receives the pressure, temperature and Figure 36
Super corrections. Invensys Auto Adjust Turbo Meters (AAT) with a
meter-mounted corrector. The meter’s rotating
output shaft provides a mechanical uncorrected
volume reading while the dual, high frequency slot-
sensor signals are the inputs used to compute the
complete Self-Checking and Self-Adjusting
computations. These meter/corrector combinations
require an external power supply. At this location, a
thermal electric generator (TEG) is located to the
left of the meter.

Figure 35
Two Invensys 12” Auto Adjust Turbo Meters
(AAT) with meter-mounted correctors. The dual,
high frequency slot-sensor signals are the inputs
used to compute the complete Self-Checking and
Self-Adjusting computations.

19

S-ar putea să vă placă și