LABORIE
LABORIE OFFICES:
USA CANADA
400 Avenue D, Suite 10 2101 Boul. Lapinière
Williston, Vermont, USA 05495 Brossard, Québec, Canada J4W 1L7
Sales/Marketing: 802.857.1300 Montreal: 450.671.5901
Customer Service: 800.522.6743 Customer Service: 888.522.6743
Technical Support: 800.333.1039 Technical Support: 800.333.1039
Fax: 802.878.1122 Fax: 450.671.7182
service@laborie.com service@laborie.com
EUROPE INTERNATIONAL
Lumonics House, Valley Drive 6415 Northwest Dr., Unit 10
Rugby, Warwickshire CV21 1TQ UK Mississauga, Ontario, Canada L4V 1X1
Tel.: +44.01788.547888 Customer Service: 905.612.1170
Fax: +44.01788.536434 Fax: 905.612.9731
internationalsupport@laborie.com internationalsupport@laborie.com
www.laborie.com
UDS120, GOBY, Urostym, EVOX, AquaVu, and i-LIST are trademarks of LABORIE Medical Technologies Canada
ULC. Windows, Microsoft Office, the Start Button, and Internet Explorer are registered trademarks of Microsoft Corp.
Bluetooth is a registered trademark of Bluetooth SIG, Inc. HP is a registered trademark of the Hewlett Packard
Company. Adobe Acrobat and Adobe Reader are registered trademarks of Adobe Systems Incorporated in the U.S.
and other countries. Norton Antivirus is a trademark of Symantec Corporation or its affiliates in the U.S. and other
countries. PIXMA is a trademark of Canon Inc. SeeMore is a trademark of Interson Corp. TECRA is a trademark of
Toshiba Corporation. All other names are trademarks of their respective owners.
All rights reserved. No part of this publication may be reproduced in any form whatsoever without the prior written
permission of LABORIE.
18 October 23, 2014 Updated installation for recommended Nitro PDF Pro V10
17 May 5, 2014 Updated display resolution settings for PC configuration
16 March 14, 2014 Updated minimum requirements for installation; Updated PC configuration instructions;
Added configuration instructions for recommended Toshiba Tecra Z50 laptop; Added Grid
Connect USB serial adapter installation instructions; Updated EVOX installation section;
Updated i-LIST Office Reporter section to include patch installation instructions; Removed
COM Port Standardization from Appendix
15 December 5, 2013 Updated i-LIST Installation instructions
14 August 20, 2013 Updated StarTech USB to Serial installation instructions to add PER381; Updated laptop
connection diagram for Urostym installation; Added new COM port standardization
information
13 August 12, 2013 Removed previous USB to Serial adapter information and replaced with new StarTech USB
to Serial installation instructions; Added note for StarTech serial adapter use with
installation of UDS120 software with a wired connection; Updated Urostym installation
instructions to reference new StarTech USB to Serial adapter; Removed COM port
standardization information
12 May 17, 2013 Removed configuration for Toshiba Tecra A11 laptop and DELL T3500 PC models; Updated
minimum requirements; Updated configuration information for DELL E6530 laptops;
Updated configuration information for DELL Precision T3600 PC; Updated PDF installation
instructions; Updated Antivirus configuration information; Updated UDS120 software
installation instructions; Updated EVOX installation instructions; Updated iLIST Office
Reporter installation instructions
11 January 15, 2013 Updated minimum requirements; Added configuration information for DELL E6530 laptops;
Added configuration information for DELL Precision T3600 PC; Added Antivirus
configuration information for DELL E6530 laptop and DELL Precision T3600 PC; Added HP
OfficeJet 6100 Printer information; Updated COM port settings
10 June 1, 2012 Added installation information to enable use of Urostym system with Urocap IV standalone
(see page 150)
9.0 March 7, 2012 Added compatibility of UDS Goby software with HASP key driver installation; Added
reminder to plug the HASP key into all USB ports of PC once Goby software installed; added
HASP key pop-up message that appears with a new install; also updated Device Manager
screenshots; Added compatibility of UDS Goby software with current version of i-LIST Office
reporter software
8.0 January 13, 2012 Updated minimum requirements for installation; Updated service pack information and
Windows OS updates setup; Added step for increasing virtual memory (for desktop PCs
only); Added note at end of each type of UDS120 installation for UDS-94 hardware with a
V1.05 board set; Added software patch installation instructions for each type of
installation; Added note before installation of SeeMore™ ultrasound probe driver
7.0 December 6, 2011 Updated minimum requirements for installation; Added ATI FirePro Video Driver
installation instructions; Added HASP Key Driver installation instructions; Added installation
instructions for NuWav ultrasound probe with UDS120 software(optional feature); Updated
UDS120 installation instructions and removed patch installation instructions; Added TETRA
Manual Date Note
Version
installation note; Added Urocap IV standalone installation; Added EVOX installation
instructions; Updated installation instructions for iLIST Reporter V1.2.2.9; Added SeeMore™
ultrasound probe driver installation instructions; Added VISER installation instructions;
Added Turn off automatic updates step in System Finalization section; Added COM Port
Standardization
6.0 September 13, 2011 Added printer driver installation instructions for the Canon PIXMA™ iP100 series wireless
printer
5.0 August 16, 2011 Added software patch installation instructions for UDS120 V12.0.3 software
4.0 April 1, 2011 Added AquaVu software installation instructions
3.0 February 11, 2011 Updated Minimum Specifications; Added configuration information for laptop and PC
models recommended by LABORIE; Added step to Bluetooth configuration; Updated
Urostym connection diagrams
2.0 December 10, 2010 Updated Minimum Specifications; Added placement of Bluetooth Devices function;
Updated for UDS120 V12 Updated UDS120 installation instructions for version 12; Added troubleshooting
release information
1.0 October 1, 2010 Creation of first version of system configuration guide.
ABOUT THIS INSTALLATION GUIDE
This guide, divided in four parts, describes how to set up and install Laborie software and/or
accessories on your computer with the Windows® 7 OS.
INSTALLATION PROCESS
System Setup
Part 1
Starting on page 1
Starting on page 45
Laborie Software
Part 3 Installation
Starting on page 89
Part 4 System
Finalization
Follow the sections in this guide in the order as outlined for proper installation. If you
encounter any problems during installation, please contact LABORIE Technical Support at
1-800-333-1039. For troubleshooting information see page 157.
IMPORTANT NOTE! Installation and configuration must first
be performed under Administrator mode. The creation of
additional administrator and/or user groups will require a
repeat of the settings outlined in this guide.
LABORIE System Installation Guide for Windows 7 OS
Page 1
LABORIE System Installation Guide for Windows 7 OS
Page 2
LABORIE System Installation Guide for Windows 7 OS
Page 3
LABORIE System Installation Guide for Windows 7 OS
4. Select “Internet Information Services”, under “Internet Information Services” click on “IIS 6
Management Console” and then click on “IIS Metabase and IIS 6 configuration compatibility” as
shown in the next screenshot.
For 32-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727”
For 64-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727” for 64 bit OS,
and press Enter key.
Page 4
LABORIE System Installation Guide for Windows 7 OS
b. Type the command “aspnet_regiis -i” and press Enter key, and wait.
6. Enable 32-Bit Applications from IIS Manager for 64 bit Operating System.
a. Click Start icon and enter “iis” in ‘Search programs and files’, and select Internet Information
Services (IIS) Manger from the list as shown next screenshot.
Page 5
LABORIE System Installation Guide for Windows 7 OS
b. The Internet Information Services (IIS) Manger window will be opened as shown below.
c. Open Advanced Settings window by select Application Pools -> DefaultAppPool, and select
‘Advanced Settings…’ on the right pane. Select True on the Enable 32-Bit Applications option.
Page 6
LABORIE System Installation Guide for Windows 7 OS
In the resulting System Properties window click the Hardware tab and then click the Device
Installation Settings button.
Page 7
LABORIE System Installation Guide for Windows 7 OS
Click the Save Changes button and continue with the next step.
NOTE: Plug in the network cable to connect to the internet as it is now required for the
following steps.
Page 8
LABORIE System Installation Guide for Windows 7 OS
Page 9
LABORIE System Installation Guide for Windows 7 OS
Click the Change advanced power settings heading and make adjustments to each item under
the Advanced settings tab.
Copy the settings for each item as shown in the following images:
Page 10
LABORIE System Installation Guide for Windows 7 OS
Page 11
LABORIE System Installation Guide for Windows 7 OS
Page 12
LABORIE System Installation Guide for Windows 7 OS
Click Apply, then click OK, and then click the Save changes button. Close all open windows.
Continue with the installation of any other software (see page 32).
Continue by installing any other third-party software such as Adobe Reader, PDF creators.
etc… then go to Part 2.
Page 13
LABORIE System Installation Guide for Windows 7 OS
Page 14
LABORIE System Installation Guide for Windows 7 OS
4. Scroll up the list and then select and expand “Internet Information Services”. Under “Web
Management Tools” add a check mark to the “IIS 6 Management Console” option. Then select and
expand ‘IIS 6 Management Compatibility’ and add a check mark to the “IIS Metabase and IIS 6
configuration compatibility” option as shown in the next screenshot.
Page 15
LABORIE System Installation Guide for Windows 7 OS
For 64-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727” for 64 bit OS,
and press Enter key.
b. Type the command “aspnet_regiis -i” and press Enter key, and wait.
6. Enable 32-Bit Applications from IIS Manager for 64 bit Operating System.
a. Click Start icon and enter “iis” in ‘Search programs and files’, and select Internet Information
Services (IIS) Manger from the list as shown next screenshot.
Page 16
LABORIE System Installation Guide for Windows 7 OS
b. The Internet Information Services (IIS) Manger window will be opened as shown below.
c. Open Advanced Settings window by select Application Pools -> DefaultAppPool, and select
‘Advanced Settings…’ on the right pane. Select True on the Enable 32-Bit Applications option.
Page 17
LABORIE System Installation Guide for Windows 7 OS
In the resulting System Properties window click the Hardware tab and then click the Device
Installation Settings button.
Page 18
LABORIE System Installation Guide for Windows 7 OS
Click the Save Changes button and continue with the next step.
Go to Start > Control Panel > Windows Update and click on Change settings.
Page 19
LABORIE System Installation Guide for Windows 7 OS
Select Never check for updates under Important updates, and select Give me recommended
updates the same way I receive important updates under Recommended updates.
NOTE: Plug in the network cable to connect to the internet as it is now required for the following
steps.
Page 20
LABORIE System Installation Guide for Windows 7 OS
Page 21
LABORIE System Installation Guide for Windows 7 OS
As Bluetooth is typed into the search field, the available settings with automatically refresh. Click
the Change Bluetooth settings heading under Devices and Printers.
In the resulting Bluetooth Settings window, make sure the Show the Bluetooth icon in the
notification area option is selected. This will place the Bluetooth icon shortcut in the task tray.
Page 22
LABORIE System Installation Guide for Windows 7 OS
Configure the Network, Universal Serial Bus Controller Power, and Bluetooth
Settings
Click Start > Control Panel. From the Control Panel window click Device Manager.
Network Settings:
Scroll down the list and double-click the ‘Network Adapters’ entry. Double-click the first hardware item
listed under Network Adapters [example: Intel® 82577LM Gigabit Network Connection] to open the
Properties window. If the Properties window contains Power Management tab then select it and
confirm that no check marks are present beside the settings ‘Allow this device to bring the computer out
of standby’, ‘Allow the computer to turn off this device to save power, ‘Allow this device to wake the
computer’, or ‘Only allow a magic packet to wake the computer’. Close the Properties window.
Click OK.
Repeat for every hardware item listed under the Network Adapters heading.
Page 23
LABORIE System Installation Guide for Windows 7 OS
Page 24
LABORIE System Installation Guide for Windows 7 OS
Page 25
LABORIE System Installation Guide for Windows 7 OS
Bluetooth Settings:
Under the Bluetooth Radios, double-click each item to open its Properties window.
If the Properties window contains Power Management tab then select it and clear the check
mark beside the ‘Allow the computer to turn off this device to save power’ or the ‘Allow this
device to save power’ options.
Page 26
LABORIE System Installation Guide for Windows 7 OS
Click the Change plan settings option on the right to access the Edit Plan Settings window.
Copy the settings as illustrated in the following image.
Page 27
LABORIE System Installation Guide for Windows 7 OS
Double-click on Sleep and confirm that the settings for Sleep are set to Sleep after: Never;
Allow hybrid sleep: Off; Hibernate after: Never; Allow wake timers: Disable.
Page 28
LABORIE System Installation Guide for Windows 7 OS
Double click on Wireless Adapter Settings, and for both On battery & Plugged in, select
‘Maximum Performance’
Double click on Display, and select the options as chosen in the following screens:
Page 29
LABORIE System Installation Guide for Windows 7 OS
Unpin the media player and My TOSHIBA from the taskbar by right-clicking on each icon and
selecting the Unpin option.
Page 30
LABORIE System Installation Guide for Windows 7 OS
Choose to ‘Close gadget’ for Yellow Pages, Norton, or any other gadgets that appear on the
desktop:
Physically, and carefully, move the laptop along the counter or surface it is placed on. In the
resulting pop-up TOSHIBA HDD Protection window, place a check mark next to the ‘Do not
display this message in the future.’ option and click OK.
Change the icon size by right-clicking on the desktop and select View > Small icons
Page 31
LABORIE System Installation Guide for Windows 7 OS
Before starting installation make sure the computer is connected to the internet.
1. If the Microsoft Office Home and Business 2013 application is on a Disc, then once the disc is
inserted the following message must appear.
2. Click Run setup.EXE.
4. Once the product key has been entered and verified to be correct, a tick mark will appear as
shown in the following screenshot. Click Continue.
Page 32
LABORIE System Installation Guide for Windows 7 OS
6. Under the following window, enter the account name and password. Once the password is
entered, click the Sign in button.
Note:
If the following message window appears at the bottom of the screen, choose ‘No’ to disable
the AutoComplete.
If the following message window appears at the bottom of the screen, choose ‘Ask me later’.
If the following message window appears at the bottom of the screen, choose ‘Enable’
7. Under the following window enter the Country and the preferable Language. Click Continue
after making the required selections.
Page 33
LABORIE System Installation Guide for Windows 7 OS
Page 34
LABORIE System Installation Guide for Windows 7 OS
10. Under the following window, enter a valid and active e-mail address and click Sign In.
11. Under the following window, enter the password used to access the e-mail address just entered.
Page 35
LABORIE System Installation Guide for Windows 7 OS
12. Under the following window, select the option Microsoft Office Home and Business 2013 (2).
Click Continue.
Page 36
LABORIE System Installation Guide for Windows 7 OS
15. Click the ‘X’ in the top-right corner to close this window or wait for the video to end in about 1
minute.
Page 37
LABORIE System Installation Guide for Windows 7 OS
17. Select No Background from the drop down box and click Next.
Page 38
LABORIE System Installation Guide for Windows 7 OS
20. In the following window, under the ‘Account’ tab, click ‘Sign out’. If the following window does
not appear, click Start >All Programs > Microsoft Office 2013 > Word 2013 to start the Word
application. Then select ‘Blank Document’. Then click FILE and go into the ‘Account’ tab as
shown below and click ‘Sign out’.
Page 39
LABORIE System Installation Guide for Windows 7 OS
22. Click Start > All Programs > Microsoft Office 2013 > Word 2013 to start the Word application.
When the Word application opens, confirm ‘Sign in to get the most out of office’ appears on the
top right corner of the Word application window, and not the credentials of the e-mail address
entered earlier in the installation process.
Page 40
LABORIE System Installation Guide for Windows 7 OS
1. Click Start > Control Panel and click the Programs and Features heading.
2. Scroll down the list of installed programs to locate the currently installed version of Nitro PDF
(earlier than V10).
3. Click on the line with the name of the Nitro version installed and then click Uninstall on the top
toolbar.
Page 41
LABORIE System Installation Guide for Windows 7 OS
Page 42
LABORIE System Installation Guide for Windows 7 OS
2. Click Run in the resulting Security Warning window and wait for the installation to occur.
Page 43
LABORIE System Installation Guide for Windows 7 OS
5. Double-click the
NitroPDF
Professional
shortcut icon on
the desktop to start
the software.
6. Click the Activate button and type in the 18-digit license number to register the full program.
Page 44
LABORIE System Installation Guide for Windows 7 OS
Install the HP Officejet Mobile 100 Printer (optional) (see page 53)
Install the HP ENVY 4500e All-in-One Printer Driver (optional) (see page 65)
Install the Grid Connect ATC 810 (PER410) USB to Serial Adapter Software (see page 71)
Configure the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURiO](see page 83)
Install HASP Driver for UDS120, UDS-Goby, NuWav Advanced, Urostym, Urocap IV
Standalone, and/or AquaVu (see page 88)
Page 45
LABORIE System Installation Guide for Windows 7 OS
NOTE: Do not power-on and connect the printer to the computer until requested to do so
during the installation process.
Page 46
LABORIE System Installation Guide for Windows 7 OS
Page 47
LABORIE System Installation Guide for Windows 7 OS
4. Deselect all options except for the HP Officejet 6100 Help option and then click Next.
5. Select the No, do not enable the Web Services option and then click OK.
Page 48
LABORIE System Installation Guide for Windows 7 OS
6. Deselect all options in the Internet Connection Usage Settings window, if not already done. Then
click OK.
7. Select the I have reviewed and accept the installation agreements and settings option
and then click Next.
Page 49
LABORIE System Installation Guide for Windows 7 OS
9. Click Finish.
10. Continue with the instructions in the Printer Driver Setup section on page 51.
Page 50
LABORIE System Installation Guide for Windows 7 OS
2. Once the printer has initialized, plug the other end of the USB cable to any available USB
port on the computer. The computer will automatically start to install the necessary
drivers. Please do not disturb the computer during this process.
3. Once the driver installation is complete, go into the Device Manager – Universal Serial
Bus controllers, and confirm the presence of ‘Officejet 6100 (REST)’.
4. Click Start > Control Panel > Devices and Printers to open the Printers window.
Page 51
LABORIE System Installation Guide for Windows 7 OS
5. Check if a tick mark is next to the ‘HP Officejet6100’ printer. If not, then right-click
over this printer and select ‘Set as Default Printer’.
7. Insert the printer USB cable into each USB port on the laptop or PC to initialize the
driver.
8. Print a test page and attach this test page to the Computer QC forms. Plug the cable into
any port to print the page.
Page 52
LABORIE System Installation Guide for Windows 7 OS
1. In the LABORIE installation disc, open the HP Officejet 100 installation program and
click Install USB-Connected Device.
Page 53
LABORIE System Installation Guide for Windows 7 OS
3. Click Next.
4. Select the checkbox “I have reviewed and accept the installation agreements and
settings. Then click Next and wait for installation.
Page 54
LABORIE System Installation Guide for Windows 7 OS
5. Connect the USB cable between the computer and the printer.
6. Click Finish.
7. Click Cancel.
Page 55
LABORIE System Installation Guide for Windows 7 OS
9. Click Exit to finalize setup. If necessary print a test page to ensure printer is properly installed.
Page 56
LABORIE System Installation Guide for Windows 7 OS
2. Turn on Bluetooth radio on the printer. Then select the “The Bluetooth light is on”
checkbox and click Next.
Page 57
LABORIE System Installation Guide for Windows 7 OS
4. Select the checkbox “I have reviewed and accept the installation agreements and
settings. Then click Next and wait for installation to occur.
NOTE: If the printer software was previously installed the printer software, some of the files may need
to be deleted and computer restarted before proceeding. Follow the onscreen instructions if the restart
screen appears. Then continue with the steps below.
Page 58
LABORIE System Installation Guide for Windows 7 OS
5. Make sure Bluetooth radios in the computer and printer are turned on. Click Next.
6. Click Start > Devices and Printers and click Add a printer.
Page 59
LABORIE System Installation Guide for Windows 7 OS
8. Locate and click on the printer name that starts with OJL411 followed by the S/N of
the printer (e.g. if the S/N of the printer is MY26M821TB, the printer name would
be OJL411MY26M821TB). Then click Next.
9. Click Next.
Page 60
LABORIE System Installation Guide for Windows 7 OS
10. Select “Do not share this printer” and click Next.
11. Make sure “Set as default printer” is selected. Then click “Print a test page”
Page 61
LABORIE System Installation Guide for Windows 7 OS
14. Select “I have created a Bluetooth connection between the printer and the computer”. Then click
Next.
Page 62
LABORIE System Installation Guide for Windows 7 OS
Page 63
LABORIE System Installation Guide for Windows 7 OS
Page 64
LABORIE System Installation Guide for Windows 7 OS
Page 65
LABORIE System Installation Guide for Windows 7 OS
4. Select the No, do not enable the Web Services option and click OK.
Page 66
LABORIE System Installation Guide for Windows 7 OS
6. Remove all check marks from the settings and click OK.
7. Select the I have reviewed and accept the installation agreement and settings option and click
Next.
Page 67
LABORIE System Installation Guide for Windows 7 OS
8. Select the Connect later - Do not connect to the printer now. I will connect it later. option and
click Next.
Page 68
LABORIE System Installation Guide for Windows 7 OS
3. Go into the Device Manager of the computer and check that the following entries are found
under each section:
Imaging devices - HP ENVY 4500 series (USB)
Universal Serial Bus controllers - HP ENVY 4500 series (REST), USB Composite Device,
USB Printing Support.
4. Click Start > Control Panel > Devices and Printers.
5. Confirm that HP ENVY 4500 series appears with a check mark next to it as shown in the
following screenshot:
6. If a check mark does not appear as shown above, right-click over the HP ENVY 4500 series icon
and choose to Set as default printer.
7. Double-click the HP ENVY 4500 series icon.
8. Double-click the HP Printer Assistant option to open the HP Printer Assistant window.
Page 69
LABORIE System Installation Guide for Windows 7 OS
9. Double-click the Maintain Your Printer option to open the HP ENVY 4500 series Toolbox
window.
10. Under the Device Services tab, click the icon next to the Print Quality Diagnostics Page option.
Page 70
LABORIE System Installation Guide for Windows 7 OS
Install the Grid Connect ATC 810 (PER410) USB to Serial Adapter Software
IMPORTANT NOTES! READ BEFORE INSTALLATION
When installing a second/additional unit of the GridConnect USB Serial Convertor, do not plug
the second unit into the same USB port of the computer used when installing the first unit.
Instead, choose a different USB port to plug and install the second unit.
Never directly unplug the GridConnect USB Serial Convertor from the USB port of the computer.
Always go first into the notification area and right-click the Safely Remove Hardware and Eject
Media option, choose Eject USB High Speed |RS-232 Converter, and then unplug the unit.
1. Open the Device Manager and under Other devices right-click on USB High Speed RS-232 Converter.
Page 71
LABORIE System Installation Guide for Windows 7 OS
Page 72
LABORIE System Installation Guide for Windows 7 OS
4. Select the Let me pick from a list of device drivers on my computer option.
Page 73
LABORIE System Installation Guide for Windows 7 OS
Page 74
LABORIE System Installation Guide for Windows 7 OS
8. Ensure the correct address is displayed for driver location and click Open.
Page 75
LABORIE System Installation Guide for Windows 7 OS
11. Open the Device Manager once again and right-click USB Serial Port under Other devices.
Page 76
LABORIE System Installation Guide for Windows 7 OS
Page 77
LABORIE System Installation Guide for Windows 7 OS
Page 78
LABORIE System Installation Guide for Windows 7 OS
Page 79
LABORIE System Installation Guide for Windows 7 OS
18. Ensure the correct address is displayed for driver location and click Open.
Page 80
LABORIE System Installation Guide for Windows 7 OS
Page 81
LABORIE System Installation Guide for Windows 7 OS
22. Plug the Grid Connect USB Serial Convertor into every port on the computer and check the
notifications/Device Manager for any COM Port assignment failures or exclamation marks.
NOTES:
When installing a second/additional unit of the GridConnect USB Serial Convertor,
do not plug the second unit into the same USB port of the computer used when
installing the first unit. Instead, choose a different USB port to plug and install the
second unit.
Never directly unplug the GridConnect USB Serial Convertor from the USB port of
the computer. Always go first into the notification area and right-click the Safely
Remove Hardware and Eject Media option, choose Eject USB High Speed |RS-232
Converter, and then unplug the unit.
23. Shutdown then power ON the computer. Unplug and then plug the Grid Connect adapter back
in to the same UDS port to which it was previously connected. Open the Device Manager and
confirm that the COM port assignment has not changed.
Page 82
LABORIE System Installation Guide for Windows 7 OS
! IMPORTANT!
COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO] is a Plug-and-Play device;
therefore, after the COM098 is plugged into the USB Port of the computer and the Device
Manager lists TDK Bluetooth USB Adapter and Microsoft Bluetooth Enumerator under Bluetooth
Radio, there will be no need to follow this procedure. The following procedure must be
followed to its completion only if the Device Manager, under Bluetooth Radio, does not list TDK
Bluetooth USB Adapter and Microsoft Bluetooth Enumerator.
1. Plug the power cord of the UDS system into an electrical outlet.
2. Insert the Laborie Released Software disc BuildXX into the CD-ROM drive of the
computer.
3. Copy and paste the folder called Windows 7 to the computer desktop. The folder can be
found inside the CD in the following location:
D:\[UDS Hardware Model]\Software\Bluetooth Drivers\COM098
[Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO]\
OR
D:\COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO]\
4. Insert the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO] Bluetooth
dongle into any USB port of the computer.
Result: The Found New Hardware Wizard window will automatically appear.
Page 83
LABORIE System Installation Guide for Windows 7 OS
8. Choose Bluetooth Radios from the Common hardware types list and click Next.
Page 84
LABORIE System Installation Guide for Windows 7 OS
11. Select the bth.ini file in the Windows 7 folder located on the desktop and click Open.
12. Confirm that the path location appears correctly and then click OK.
Result: The Found New Hardware Wizard window appears with the COM098 driver.
13. Click Next.
14. Select Yes if the following message screen appears.
15. Wait for 2-3 minutes while the wizard searches for the COM098 driver.
Result: The screen with the ‘TDK Bluetooth USB Adapter’ will appear confirming that the driver
has been installed onto the computer.
16. Click Finish. The computer will automatically initialize the driver installation of the Microsoft
Bluetooth Enumerator.
If you see the Found New Hardware Wizard window again, then follow steps below.
a. Choose ‘Install the software automatically [Recommended]’ and click Next.
Page 85
LABORIE System Installation Guide for Windows 7 OS
b. Wait 2-3 minutes while the wizard searches for the Microsoft Bluetooth Enumerator.
c. The following screen with the ‘Microsoft Bluetooth Enumerator’ will appear confirming that the
driver has been installed onto the computer.
d. Click Finish.
e. Plug the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO] into each USB port of
the computer, doing so, you may be asked by the computer to repeat ALL the steps explained
above (i.e. steps 1 to 16).
f. Continue with the following steps.
17. Delete the Windows 7 folder that was copied on the desktop in step 3.
18. Click Start > Control Panel and type the word Bluetooth into the search field at the top-right corner.
As Bluetooth is typed into the search field, the available settings will automatically refresh. Click
the Change Bluetooth settings heading under Devices and Printers.
Page 86
LABORIE System Installation Guide for Windows 7 OS
In the resulting Bluetooth Settings window, make sure the Show the Bluetooth icon in the
notification area option is selected. This will place the Bluetooth icon shortcut in the task tray.
Click OK.
19. Click the Start button , click Control Panel, click Hardware and Sound, and then click
Bluetooth Devices.
Page 87
LABORIE System Installation Guide for Windows 7 OS
Install HASP Driver for UDS120, UDS-Goby, NuWav Advanced, Urostym, Urocap
IV Standalone, and/or AquaVu Software
NOTE: This step does not need to be performed for the DELL E6530 laptop. However, you
must plug the Aladdin Hasp Key into each USB port on the laptop or PC before proceeding
with the other installations. At times it may take a minute for the HASP key to activate when
plugged in to the USB port. Wait for the hasp key light to illuminate before continuing to plug
into another USB port.
Do not connect probes to the computer until instructed to do so.
1. Insert the LABORIE software disc
into the CD or DVD drive and
double-click the
HASPUserSetup.exe application.
3. Click Next in the following screen and wait for installation to occur.
4. Click Finish to complete installation.
Page 88
LABORIE System Installation Guide for Windows 7 OS
Install NuWav Advanced Ultrasound Software with UDS120 Software (see page 90)
Install UDS120 Software (for all Laborie UDS systems except Goby) (see page 92)
Install UDS120 Software for a Single UDS Device Connection (see page 92)
Install UDS120 Software for a Multiple UDS Device Connection (see page 93)Install the GOBY
Software (see page 98)
Page 89
LABORIE System Installation Guide for Windows 7 OS
3. Click Next.
Result: The End User License Agreement will appear.
4. Click I Accept.
Result: The Connection Type window appears.
5. Select the hardware and type of connection and click Next. For Goby KT select UDS94 for
Delphis, Delphis IP, Delphis KT and the Bluetooth (MS Stack) options.
IMPORTANT! If you have selected a wireless option a screen will appear to remind you
to install Bluetooth software. Follow the appropriate Bluetooth Manuals for installing
the Bluetooth support on the computer or laptop.
Page 90
LABORIE System Installation Guide for Windows 7 OS
6. Select NuWav – Advanced plus any other Options and click Next.
9. Click Next.
Result: The program is ready for installation.
Page 91
LABORIE System Installation Guide for Windows 7 OS
Install UDS120 Software (for all Laborie UDS systems except Goby)
1. Double‐click the UDS 120V12.x.x-ServerX Vx.xx.exe program in the Laborie software disc.
Result: The Welcome to UDS 120V12.x.x-ServerX Vx.xx Setup welcome screen will appear.
2. Click Next.
Result: The End User License Agreement will appear.
3. Click I Accept the license agreement and click Next.
Result: The Connection Type window appears.
4. Select the hardware and type of connection and click Next. For Goby KT select UDS94 for
Delphis, Delphis IP, Delphis KT and the Bluetooth (MS Stack) options.
5. Click Ok in the resulting warning screen.
6. Select any Options and click Next. This is where you will select options like video,
Ultrasound (for NuWav), infusion transducer, and/or infrared remote. If any of these are
part of your system you must select them here to ensure proper software configuration.
7. Click Ok in the resulting warning screen.
8. Select the Language and click Next.
9. Confirm the Folders and click Next.
10. Click Next.
Result: The program will be installed on the computer.
11. Click Finish to complete installation.
12. (optional) If the UDS120 software will be used with UDS-94 hardware having a v1.05 board
set, then open the uds120.ini file (located in C:\Windows\uds120.ini\) and change
the EMGDeltaAD value under the EMG section from 837 to 755.
If no other programs are required at this point then go to Part 4 System Finalization to
complete set up.
IMPORTANT NOTE for UDS120 Wired Connection Installation with the Adapter:
Once the UDS installation process for Wired Connection is complete, change the uds120.ini
settings under the [UDS-120] section from CommPortLoopCount=300000 to
CommPortLoopCount=600000
Page 92
LABORIE System Installation Guide for Windows 7 OS
NOTE 1. The ONLY Multiple Device connections supported with this software release are the
following:
1 UDS device (Triton/Aquarius, Delphis/KT) with wireless urocap
1 UDS device (Triton/Aquarius, Delphis/KT) with wireless urocap and a Mini-D device
1 UDS device (Triton/Aquarius, Delphis/KT) with 2 Mini-D devices
Goby-KT with Wireless Uroflow
Tetra NIRS (wired) with Triton
Tetra NIRS (wired) with AquariusTT
Tetra NIRS with Urocap / Wireless Uroflow
Urocap with Mini-D
LIMITATIONS: The pair of Multiple Devices NOT supported in this present release is the
following:
Tetra NIRS with Mini-D
Note 2. Functions ONLY with Bluetooth (MS Stack) and NOT Bluetooth (Widcomm). Therefore,
Bluetooth (Widcomm) should not be using this present release of UDS120 for Multiple Devices.
Note 3. A maximum of three wireless connections are recommended for connection through Bluetooth.
Note 4: To install a multiple device connection with Urostym V6, please contact the Service
department.
IMPORTANT NOTE for UDS120 Wired Connection Installation with the Adapter:
Once the UDS installation process for Wired Connection is complete, change the uds120.ini
settings under the [UDS-120] section from CommPortLoopCount=300000 to
CommPortLoopCount=600000
Page 93
LABORIE System Installation Guide for Windows 7 OS
IMPORTANT! If you have more than two UDS devices to connect and you are not sure of
which combination to select in this screen then follow the list below. Selections are listed in
order of importance.
1. Triton
2. Delphis
3. Tetra
4. Urocap
5. Mini-D
For example:
If you have a Triton device, a Tetra device, and a Urocap device, then you would
select the Tetra NIRS with Triton option and configure the third device (the Urocap)
in the Connection Manager screen in the UDS120 software.
If you have a Delphis device, a Mini-D device, and a Urocap device then you would
select the Delphis with Urocap option and configure the third device (the Mini-D) in
the Connection Manager screen in the UDS120 software.
6. Click Next.
7. Select any Options and click Next.
8. Select the Language and click Next.
9. Click Next in the Destination Folder screen.
10. Click Next to start the installation. Installation may take a few minutes to complete.
11. Click Finish to complete the software installation.
12. (optional) If the UDS120 software will be used with UDS-94 hardware having a v1.05
board set, then open the uds120.ini file (located in C:\Windows\uds120.ini\)
and change the EMGDeltaAD value under the EMG section from 837 to 755.
13. Continue with the configuration of the Multiple Device Configuration as described in
the following section.
Page 94
LABORIE System Installation Guide for Windows 7 OS
1. Start the UDS120 software by double-clicking the UDS 120 shortcut on the desktop.
2. Click Connection Manager to open the Connection Manager window.
3. You will see two UDS devices already listed in the Device column. These were
chosen during the installation process. If you have more than two UDS devices
then select any additional devices in this column.
IMPORTANT!
Do not click the
Play button
until ALL
devices have
first been set up
in the Device,
Connection
Type, and Wired
COM
Step 7 Step 3 Step 4 Step 5
Port/Wireless
Settings
columns.
7. Once all connections are set, close the UDS120 software and then restart the
UDS120 software.
8. Once ALL the devices have been added click the button next to each device to
start the UDS device:
Page 95
LABORIE System Installation Guide for Windows 7 OS
11. When the Calibration button for another UDS device is clicked a message box
appears asking for the selection of the Flowmetry device to use for calibrating
the Pump transducer. For example: If Urocap1 is to be chosen as the Flowmetry
device to use with the pump, then calibrate the Urocap1 device before starting
the calibration of the Pump transducer inside the Delphis1 device.
12. After selecting the Flowmetry device to use with the pump, a Transducer
Calibration window appears. Continue with the calibration of the other UDS
device and then continue with the following steps.
13. Click the Channel Mappings tab. This will display the channels that are needed to
configure the UDS graph for testing with the UDS devices. The channels from the
devices selected during installation will already be listed.
* NOTE: If any changes are made to the Channel Mappings tab after initial configuration, the UDS120
software must be restarted.
* NOTE: If you have more than two UDS devices to connect, you will need to map the channels in this
list for any additional device(s). Scroll down the list to the first empty line and begin by selecting the
Device Name, then select the Device HW Channel and then select the UDS Screen Channel Name.
Refer to the following charts for additional channel setup:
Additional Mini-D1 Device
UDS Screen Channel Name Device Name Device HW Channel
Ch9 Mini-D1 Ch1
Ch10 Mini-D1 Ch2
Ch11 Mini-D1 Ch3
EMGR Mini-D1 EMGR
Additional Mini-D2 Device
Page 96
LABORIE System Installation Guide for Windows 7 OS
15. In the UDS120 main window click Config > Set up/Modify.
16. Create a configuration file that contains all the channels listed in the UDS Screen
Channel Name column from the Channel Mapping tab. For example, following the
channel listing as seen in the screen from step 13 the configuration file will look
like this:
17. Type a name in the Test Name box. This will be the basic configuration file that
you will use for testing with all the connected devices.
18. Click OK.
If no other programs are required at this point then go to Part 4 System Finalization to complete
set up.
Page 97
LABORIE System Installation Guide for Windows 7 OS
Page 98
LABORIE System Installation Guide for Windows 7 OS
Page 99
LABORIE System Installation Guide for Windows 7 OS
10. (optional) Insert the Bluetooth dongle (COM093) into a USB port on the computer and
wait for the configuration of the Bluetooth.
11. Insert the HASP key into one of the USB ports on the computer and wait until Windows
configures the HASP key driver. Plug the HASP key into every USB port of the computer.
12. Continue with the installation of the i-LIST Office Reporter Software (see page 131). Once
i-LIST Office Reporter is installed return to this page and continue with the connection of
the GOBY components (as outlined in the following section).
3. Switch the GOBY Hub to ON with the power switch at the bottom of the Hub.
Page 100
LABORIE System Installation Guide for Windows 7 OS
4. Plug the Urocap IV into the charger and plug the charger into an outlet.
If using a GO Cart, place the Urocap
into the charger at the bottom of the
GO Cart and then plug the power
cord of the cart into an outlet. Make
sure the cart is on a level surface and
the wheels are locked.
5. Double-click the GOBY Device Manager
icon on the computer desktop.
7. Start by connecting the Urocap IV Uroflowmeter. Make sure it is near the computer.
Page 101
LABORIE System Installation Guide for Windows 7 OS
15. Repeat steps 7 to 14 to connect the GOBY Hub and the UDS Roam to the system.
When all components are connected and functioning, the GOBY Device Manager window will
look similar to this:
Page 102
LABORIE System Installation Guide for Windows 7 OS
Page 103
LABORIE System Installation Guide for Windows 7 OS
Order of Installation
When installing Urocap IV software, the order for proper installation is as follows:
1. Install the Sentinel HASP key driver (if not already installed). See page 88.
2. Install the Urocap IV software. i-LIST Office Reporter software is part of the Urocap IV
standalone install so there is no need for separate install for i-LIST Office Reporter software.
Page 104
LABORIE System Installation Guide for Windows 7 OS
4. Click Next.
Page 105
LABORIE System Installation Guide for Windows 7 OS
Page 106
LABORIE System Installation Guide for Windows 7 OS
10. (optional) Insert the Bluetooth dongle into a USB port on the computer and wait for the
configuration of the Bluetooth.
11. Insert the HASP key into a USB port on the computer.
Page 107
LABORIE System Installation Guide for Windows 7 OS
Page 108
LABORIE System Installation Guide for Windows 7 OS
When the Urocap IV is connected and functioning, the Goby Device Manager window will look
similar to this:
Setup complete. Continue with the final setup process in Part 4: System Finalization.
Page 109
LABORIE System Installation Guide for Windows 7 OS
Important notes: If the UDS120 software application is to be installed on the same computer
with UROSTYM software, then the UDS software must be installed first; do not register the
modules again through the UROSTYM Utility program, but instead keep using the modules
registered through the UDS120 software.
The HASP key is required for this version of UROSTYM. Do not plug the key in before installation
of the software is completed. If it is already plugged into the computer, remove it and re-insert
it when asked to do so during installation.
1. If the Line Isolation Transformer (LIT) has a power switch, make sure it is in the OFF
position.
2. Connect one end of the power cord (LIT080) to the LIT and connect the other side to the
UROSTYM.
3. Plug the AC adaptor of the laptop computer into an outlet. If using power adaptor
POW320 with a laptop, plug it into the LIT. If using a desktop computer, plug the
computer and monitor power cords into the LIT. NOTE: LIT is recommended to meet
Hospital Grade power protection requirements in the United States.
4. Plug the printer power cord into the LIT.
5. Connect the medical grade power cable between the (LIT) and an electrical outlet. If the
LIT has a power switch, turn on the LIT with the power switch.
Page 110
LABORIE System Installation Guide for Windows 7 OS
Page 111
LABORIE System Installation Guide for Windows 7 OS
Page 112
LABORIE System Installation Guide for Windows 7 OS
The USB-to-Serial Adapter driver must be installed before continuing. If it is not already
installed go to page 71 for installation instructions.
No more than one adapter should be configured and connected to the computer.
Configuration of a second adapter unit to work on the same computer will result in
communication failure and may require re-installing the operating system on the
computer. If for any reason a new adapter is ordered, the aforementioned failures will
result.
If using Urostym with other Laborie UDS equipment, then the wired connection to the
Urostym system must be configured before the Bluetooth connection to UDS or Urocap
systems is configured.
Never change the COM Port number of the ‘Unknown Device’ (i.e. the PER380 or
PER381) located under the Device Manager – Ports (COM & LPT).
4. Click the I accept the license agreement option and click Next.
Page 113
LABORIE System Installation Guide for Windows 7 OS
5. Click Next in the following screen and wait for installation to occur.
During the installation process, files are copied onto the PC and a UROSTYM shortcut
Page 114
LABORIE System Installation Guide for Windows 7 OS
14. Plug the HASP key into a USB port and wait until Windows configures the HASP key
driver.
15. Start the UROSTYM software by double-clicking the UROSTYM shortcut on the desktop.
16. Click OK in the resulting pop-up messages.
17. Set up the UROSTYM components in the resulting CheckDB window section by section.
Page 115
LABORIE System Installation Guide for Windows 7 OS
Page 116
LABORIE System Installation Guide for Windows 7 OS
Then, click the Register Components button and select the select the ServerXMSB.ocx for a
PC with Bluetooth (MS Stack) component to be registered:
Result: The RegSvr32 message will appear. Click OK to close the pop-up message
18. Click the Exit button to close the Application. In the resulting message click Yes to start
the UROSTYM application or click No to exit.
The software installation is complete. Go to the configuring the software section on page
121.
Page 117
LABORIE System Installation Guide for Windows 7 OS
Install the UROSTYM V6 Software for a UROSTYM Unit (PRO180) with a UDS Unit
1. Install the UDS120 software if not already installed.
2. Install the UROSTYM V6 Software Application (see page 114, start from step 7 and go to
step 18 (but do not exit the checkDB utility window).
3. Click the Configure the UDS120.ini file button (in the 6.Path to UDS120.ini file section in
the bottom-left of the screen).
Result: The UDS120.ini File Configurator window appears.
4. Set the path of the UDS120.ini file by clicking the path selection button in the
upper-left of the screen and then selecting the UDS120.INI file located inside the
C:\Windows\ folder.
Page 118
LABORIE System Installation Guide for Windows 7 OS
Install the UROSTYM V6 Software Application for use with a UDS Unit Only
1. Install the UDS120 software if not already installed.
2. Install the UROSTYM V6 Software Application (see page 114, start from step 7 and go to
step 18 (but do not exit the checkDB utility window).
3. Click the Configure the UDS120.ini file button (in the 6.Path to UDS120.ini file section in
the bottom-left of the screen).
Result: The UDS120.ini File Configurator window appears.
Page 119
LABORIE System Installation Guide for Windows 7 OS
If the UDS120.ini file is configured for a Multiple Bluetooth connection, then the
following message will be displayed:
Click OK.
The software installation is complete. Go to the configuring software section on page 121.
Page 120
LABORIE System Installation Guide for Windows 7 OS
3. Depending on the equipment used for the system make the selections as
follows:
Page 121
LABORIE System Installation Guide for Windows 7 OS
Result: The UDS indicator on the main window status bar is green, the
Samples value on the Adjustment form goes up, and the curves in the Flow
and Volume graph areas show a curve deflection.
o Click OK.
5. Click OK.
6. Close the UROSTYM application.
2. Click the Configure the UDS120.ini file button (in the 6.Path to UDS120.ini file section in the
bottom-left of the screen).
Page 122
LABORIE System Installation Guide for Windows 7 OS
Result: The UDS120.ini File Configurator window appears and the warning message may
display. Click OK to continue.
4. In the resulting window, set the path to the UDS120.ini file to C:\Program Files\Laborie
Medical Technologies\Urostym\UDS120.INI.UC-III.
5. In the Connection/Hardware Type section select the Hardware type and connection type.
ii. For a wired connection click the Change COM Port button and select the COM port
that you wish to use:
COM1 – for Triton/Delphis family
Page 123
LABORIE System Installation Guide for Windows 7 OS
Page 124
LABORIE System Installation Guide for Windows 7 OS
8. Click the Transducers Calibration button and calibrate all necessary transducers according
to UDS Manual or follow the on-screen instructions.
UROSTYM software for multiple Bluetooth connection or for Urocap III use is complete.
2. Back up the Urostym.mdb file found on the old computer with the patient data.
Note: By default the Urostym.mdb file is located in the folder:
For version 6.0.3 or older – “C:\Program Files\Laborie\Urostym\"
For version 6.1.0 - "C:\Program Files\Laborie Medical Technologies\Urostym\"
3. Insert a USB memory key into the old computer and copy the Urostym.mdb file to the key.
Remove the memory key.
4. Insert the memory key into the new computer and paste the Urostym.mdb file to the new
computer to the C:\Program Files\Laborie Medical Technologies\Urostym\ folder.
You will need to overwrite the existing Urostym.mdb file.
6. Click the Update DB to version 6.3.0 button in the CheckDB window located in section 1
Database 6.3.0.
Page 125
LABORIE System Installation Guide for Windows 7 OS
If both the UDS120 software application and the Evox software application are to be installed on
one computer, first install the UDS120 software application and then continue by installing the
Evox software application named EvoxV1.1.x_ServerX3.0.x_Setup.exe [located inside the folder
named :\\Laborie Software Disc\Evox\Software\].
If you currently have, or previously had, any EP (Evoked Potential) or Evox software applications
installed on your computer, please back up all configuration (.CFP, .CFT), settings (.INI), and UDS
data (.EDA) files on the computer before EVOX Software installation.
Each file required for installation is located in the LABORIE Software Disc:
o UDS120 v12.X.X-ServerX3.0.XX.exe: \\Laborie Software Disc\[UDS Device
Name]\Software\V12.x.x_ServerX3.0.x\UDS120-V12.x.x_serverX3.0.x.exe.
o EvoxV1.X.X_ServerX3.0.XX_Setup.exe:\\Laborie Software Disc\Evox\Software\
EvoxV1.1.x_ServerX3.0.x_Setup.exe.
When UDS120 V12.1.xx (ServerX3.0.xx) and Evox V1.1.xx (ServerX3.0.xx) are installed on the
same computer, do the following before launching UDS120 software.
i. Plug-in the Aladdin HASP key programmed with the required features.
ii. Install UDS120 software
iii. Install Urostym software (optional)
iv. Install Evox software
v. Install iLIST Office Reporter software
vi. Launch Evox software and then exit the Evox software (to initiate the ServerX
registry)
2. Click Next.
Page 126
LABORIE System Installation Guide for Windows 7 OS
6. Click Next.
7. Click Next.
Page 127
LABORIE System Installation Guide for Windows 7 OS
NOTES:
i. After EVOX Installation, the following directories are created on the PC:
C:\Data\Evox (the path for EVOX data files)
C:\CFG\Evox (the path for EVOX configuration files)
C:\UDS120\ (the path for EVOX program files)
ii. When UDS120 and EVOX Software are installed on the same computer, the EVOX software and the
UDS software CANNOT RUN AT THE SAME TIME. One needs to be closed prior to opening the other
software.
Page 128
LABORIE System Installation Guide for Windows 7 OS
Page 129
LABORIE System Installation Guide for Windows 7 OS
Connected Disconnected
NOTE: You may be required to restart the software and hardware.
If you power off the equipment, please wait at least one minute before powering it on again.
Page 130
LABORIE System Installation Guide for Windows 7 OS
2. Depending on the system components present on the system, the installation process
will occur in one of the following ways:
i. If .Net Framework 3.5 SP1 and SQL Server 2008 Express Edition are
already installed then the i-LIST Office application will automatically
begin. Go to step 10 on page 133 .
ii. If .Net Framework 3.5 SP1 and SQL Server 2008 Express Edition are not
installed go to step 3 on page 131.
3. To begin .Net Framework 3.5 SP1 installation click Accept in the following three setup
windows as they appear on screen.
Page 131
LABORIE System Installation Guide for Windows 7 OS
4. Wait for the setup to continue. Ignore the “Not Responding” message and let the
installation run.
6. Click Yes.
Page 132
LABORIE System Installation Guide for Windows 7 OS
7. After restart, the i-LIST Prerequisites Setup will continue, and the SQL Server
2008 Express Edition installation will automatically start.
10. The i-List Office Reporter 1.3.0.9 Setup Wizard will begin.
Page 133
LABORIE System Installation Guide for Windows 7 OS
14. Wait for installation to occur. The program will show a series of screens that
track the installation processes.
Page 134
LABORIE System Installation Guide for Windows 7 OS
Page 135
LABORIE System Installation Guide for Windows 7 OS
1. Start the UDS120 or UDS Goby software and open the Uroflow configuration file by
either clicking the Uroflow button on the UDS control panel or clicking Config > Open
and then double-clicking the uroflow.cfg file.
2. Click Config > Set up/Modify.
3. Ensure that the Test Type box has the correct test type (for example, Uroflow is set in
the following illustration) selected.
4. Click the ILIST Events Linking tab and then click the Full Test Graph Settings button.
5. Ensure all the settings on the i-LIST Event Graph are set as in the image below:
6. Repeat steps 1 to 5 for all remaining configuration files (Micturition, UPP, etc…). For each
test, enter the correct test type and select the correct channels in the Specify the Channels
section in the i-LIST Event Graph box.
Page 136
LABORIE System Installation Guide for Windows 7 OS
Install Patch
NOTE: Make sure all other applications are closed before installing the patch.
1. In the LABORIE software disc, double-click the RunMe.exe file in the Patch-
ArmReport2And3-for-v1-3-0-18 folder.
2. Click Run.
Page 137
LABORIE System Installation Guide for Windows 7 OS
Verification of Installation
1. Right-click on the Computer icon on the computer desktop and select Manage.
2. From the Computer Management window check if the snapshotqueue and udsadmin
entries appear in the Message Queuing group as illustrated in the following image:
Page 138
LABORIE System Installation Guide for Windows 7 OS
i-LIST Office Reporter software installation complete. Restart the computer and continue with the
installation of any remaining applications.
Page 139
LABORIE System Installation Guide for Windows 7 OS
Do not connect the probe to the computer until instructed to do so during installation.
NOTE: Before start of installation, make sure that the HASP Key driver is installed. If not already
installed, see page 88 for installation instructions.
1. Double-click the setup.exe file in the AquaVu product software folder on the LABORIE
software disc.
Page 140
LABORIE System Installation Guide for Windows 7 OS
1. Once the Set up Security Permissions for User Mode Support process has
been performed on a computer with AquaVu installed, contact your IT department to request
and to enable user mode support.
5. Once the installation process is complete, the AquaVu software will open. At this point, close
the application, plug in the ultrasound probe into any USB port on the computer, and restart the
AquaVu software application.
AquaVu installation process is now complete and AquaVu enabled to be used under
user mode.
Page 141
LABORIE System Installation Guide for Windows 7 OS
Page 142
LABORIE System Installation Guide for Windows 7 OS
5. Keep the destination folder as assigned by the installation program and click Next.
Page 143
LABORIE System Installation Guide for Windows 7 OS
6. Keep the start menu folder as assigned by the installation program and click Next.
Page 144
LABORIE System Installation Guide for Windows 7 OS
9. Click Install this driver software anyway in the security message window.
Page 145
LABORIE System Installation Guide for Windows 7 OS
11. Click Finish in the Completing the SeeMore USB Application Setup Wizard screen.
Plug the probe into each USB port of the computer and restart the computer after each plug-in.
NOTE: If the following window
appears when probe is plugged in
or software starts then select the
first option and click Next.
Page 146
LABORIE System Installation Guide for Windows 7 OS
2. Click Next.
Result: The User Information screen
appears.
3. Add the user’s Full Name, the name of the Organization and click Next.
Result: The Destination Folder window appears.
Page 147
LABORIE System Installation Guide for Windows 7 OS
Page 148
LABORIE System Installation Guide for Windows 7 OS
2. In the main menu bar click Patient and then click Preferences.
3. Select the Sensors Sets heading on the left side of the window.
add pressure
channel number
here
4. Apply pressure onto the pressure transducer and specify which channel is responding.
For example, if Channel 5 is responding to pressure placement then add the number 5
into the Pressure Channel text box in the Sensors Set 1 section.
5. Click OK.
6. Click OK in the resulting pop-up message.
The configuration is saved on the computer hard drive and will be used as the current setting
for all tests.
Page 149
LABORIE System Installation Guide for Windows 7 OS
The Urostym must use Urostym software V6.3.0.10 with iList V1.2.2.9. Use of other versions of
Urostym software will not be compatible.
Once the above conditions are satisfied, then you can use ONE hasp key programmed with
Urocap-IV and LAB596 Urostym software.
When Urostym and Urocap IV are on the same system with the same HASP key, the
EMG/Uroflow feature in Urostym will not work with the Urocap IV.
1. Install the HASP Key driver. See page 88 for installation instructions.
2. Install Urostym software. See page 110. NOTE that only Urostym V6.0.3.0.10 is
supported for use with this reporting feature.
3. Install Urocap IV Standalone with iLIST V1.2.2.9 software. See page 104.
b. Click Run.
Page 150
LABORIE System Installation Guide for Windows 7 OS
d. Click OK in the
confirmation message.
Installation complete.
Page 151
LABORIE System Installation Guide for Windows 7 OS
Upgrade Software
1. Backup the following files: (this step is needed only to ensure no data is accidentally lost)
C:\windows\uds120.ini file
C:\windows\saturnDirectShow.ini file (if available)
C:\CFG folder
3. If you are updating software on a "live" system and wish to preserve the current
software settings, then go to step 4; otherwise perform the following:
a. Please ensure the following files are removed from the C:\Windows folder
(Please note that some of the items might be already removed when uninstalling
the software in which case it is OK)
uds120.ini
b. Please ensure the following folder is removed:
C:\CFG
4. If you wish to remove the i-LIST Database, perform the following steps:
Click Start and type "cmd" in the Run field.
Type the following command (iLIST database will be removed and existing
iLIST data will not be recoverable after this step): OSQL -E -S .\SqlExpress
-Q "drop database [iList-Snapshot]
Ensure no error messages are displayed. The cmd window should appear
as shown:
Page 152
LABORIE System Installation Guide for Windows 7 OS
Files and Folders backup performed by Service: Note* After installation, the files and
folders, if applicable, should be restored if the Installation has overwritten the existing files
and folders mentioned below. See Upgrading of Existing Installation notes above.
Page 153
LABORIE System Installation Guide for Windows 7 OS
Page 154
LABORIE System Installation Guide for Windows 7 OS
Page 155
LABORIE System Installation Guide for Windows 7 OS
4. You will be asked for a name. Give it a useful name that you will be able to easily
identify later.
5. Click the Create button. The system will start to create the restore point.
When finished, you will see a message saying “The restore point was created successfully”.
System installation and configuration complete. The system is now ready for use.
Page 156
LABORIE System Installation Guide for Windows 7 OS
Appendix
Troubleshoot Installation
Page 157
LABORIE System Installation Guide for Windows 7 OS
1. Open the Zip file (V6_License_Manager_x86.zip) from the Laborie Disc Build1XX\Driver and Utilities\
2. Extract LicenseManager.exe from the Zip file (V6_License_Manager_x86.zip).
IMPORTANT: Make sure you complete steps 3 to 5 otherwise you will receive an error when running
the alternative License Manager.
3. Right click LicenseManager.exe and select Copy from the context menu.
4. Open up My Computer on your PC and browse to the default location of your Nitro Pro installation:
Nitro Pro 8
* 32-bit: C:\Program Files\Nitro\Pro 8
* 64-bit: C:\Program Files (x86)\Nitro\Pro 8
Nitro Pro 6 or 7
* 32-bit: C:\Program Files\Nitro PDF\Professional
* 64-bit: C:\Program Files (x86)\Nitro PDF\Professional)
5. Once the above directory is open, right click and select Paste to move License Manager.exe into that
location.
6. Double click on the License Manager.exe file.
7. Enter your serial number into the License Number field (screenshot below).
8. Select Activate at the bottom of the screen (screenshot below)
Page 158
LABORIE System Installation Guide for Windows 7 OS
You should then be able to double click the Nitro Pro icon on your desktop.
Page 159