Sunteți pe pagina 1din 8

Analytica Chimica Acta 1042 (2018) 52e59

Contents lists available at ScienceDirect

Analytica Chimica Acta


journal homepage: www.elsevier.com/locate/aca

Development and validation of an open screening method for doping


substances in urine by gas chromatography quadrupole time-of-flight
mass spectrometry
Michael Polet*, Wim Van Gansbeke, Peter Van Eenoo
Ghent University, Department of Clinical Chemistry, Microbiology and Immunology, Doping Control Laboratory, Technologiepark 30 B, B-9052, Zwijnaarde,
Belgium

h i g h l i g h t s g r a p h i c a l a b s t r a c t

 An open and complete GC-MS high


resolution screening method.
 Target analyte list no longer required
due to open screening.
 Expanded list of compounds in the
gas chromatography screening
method.
 The high resolution screening
method is compliant with WADA
regulations.
 Equal or lower LLODs for 80.2% of the
compounds in comparison with CI
GC-QQQ-MS.

a r t i c l e i n f o a b s t r a c t

Article history: In anti-doping, a high number of classes of substances are prohibited and laboratories need to detect
Received 17 July 2018 these at low urinary concentrations. Traditionally, testing is done using complimentary liquid
Received in revised form chromatography mass spectrometry and gas chromatography mass spectrometry. High resolution
22 August 2018
mass spectrometric acquisition has some important advantages over triple quadrupole instruments
Accepted 26 August 2018
(e.g., open screening due to full scan high resolution data acquisition with retrospectivity,
Available online 28 August 2018
compatibility with libraries and a straightforward and effortless addition and validation of new
compounds in the future). Doping samples can be stored for 10 years and retrospective data analysis
Keywords:
Time-of-flight
can be used to re-evaluate previously acquired data (e.g., searching for prohibited (designer) sub-
Gas chromatography stances that were unknown at the initial moment of analysis). During the past decade, these ad-
High resolution vantages have led to the wide-scale transfer of liquid chromatography triple quadrupole mass
Doping spectrometry screening to liquid chromatography high resolution mass spectrometry screening for
Open screening doping control purposes. Up to now, for gas chromatography a similar switch to high resolution
Drugs of abuse screening has not yet occurred, because so far no method has been developed that combines suf-
ficient sensitivity with wide-scale drug detection.
In this work, the current gas chromatography triple quadrupole mass spectrometry screening
method for human doping control purposes was successfully converted into an equivalent and
complete gas chromatography high resolution acquisition screening method. This new screening
method on a gas chromatography quadrupole time-of-flight mass spectrometer has been developed
and validated. The method is compliant with the World Anti-Doping Agency requirements and allows
the detection of 294 target compounds (and 14 internal standards), including diuretics, stimulants,

* Corresponding author.
E-mail address: michael.polet@UGent.be (M. Polet).

https://doi.org/10.1016/j.aca.2018.08.050
0003-2670/© 2018 Elsevier B.V. All rights reserved.
M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59 53

narcotics, beta-2-agonists, beta-blockers, hormone modulators, anabolic agents and the quantification
of 14 endogenous steroids in a single fast run (14.1 min).
© 2018 Elsevier B.V. All rights reserved.

1. Introduction important advantages in comparison with GC-QQQ-MS (e.g., open


screening due to full scan high resolution data acquisition with ret-
The World Anti-Doping Agency (WADA) defines which sub- rospectivity, compatibility with libraries and a straightforward and
stances are regarded as prohibited in sports and these compounds effortless addition and validation of new compounds in the future).
cover a wide range of chemically and pharmacologically different During the past decade, these advantages have led to the wide-scale
classes [1]. The lowest concentrations that WADA accredited labo- transfer of LC-QQQ-MS screening methods to a LC high resolution MS
ratories are required to detect on a day-to-day basis is set by WADA screening [3,19e23]. Up to now, for GC-MS a similar switch to high
as the minimum required performance limit (MRPL) [2]. Nowadays, resolution screening has not yet occurred.
efficient doping control laboratories use one gas chromatography This paper describes the first successful conversion of the cur-
mass spectrometry (GC-MS) method and one liquid chromatography rent GC-QQQ-MS screening method for human doping control
MS (LC-MS) method for their initial testing procedure [3e5]. These purposes into an equivalent and complete GC high resolution
two screening methods are complementary and together they cover acquisition screening method on a new generation GC-QTOF-MS
all low to medium molecular weight drugs of abuse that need to be instrument. The developed and validated method includes 280
detected according to the WADA regulations [1]. For example, some qualitative compounds and 14 quantitative compounds for analysis
compounds are difficult or impossible to detect by using LC-MS (e.g., of human doping control samples. Due to the full scan high reso-
low ionization efficiency for oxymesterone) and these need to be lution acquisition, a large amount of compounds can be added to
included in the GC-MS method [6]. Other compounds are easy on LC- the method, making it a very versatile approach that can easily be
MS, but challenging on GC-MS (e.g., tetrahydrogestrinone) [7]. For transferred and applied in other analytical fields such as environ-
GC-MS screening purposes, prohibited substances can be divided ment, food safety, toxicology and horse doping [24,25]. This open
into two categories: exogenous substances that need to be detected screening method does not rely on a targeted analyte list, allowing
qualitatively (i.e., as sensitive as possible) and endogenous steroids the identification of new compounds more quickly, followed by
that need to be quantified as accurate as feasible within a very wide retrospective data analysis without having to reanalyze samples. In
concentration range [2,8]. Both categories require unique and con- addition, the performance of the method is not restricted by the
flicting method characteristics and a performant screening method amount of compounds that is monitored, as is the case for GC-QQQ-
demands an optimal compromise (e.g., high sensitivity versus high MS screening.
capacity, high chromatographic resolution versus short analysis
time). For example, a higher injection volume may lower the limit of
2. Experimental section
detection (LOD), but will simultaneously deteriorate the quantifica-
tion of endogenous steroid isomers (e.g., etiochanolone and
2.1. Reagents and reference standards
androsterone, two essential endogenous steroid epimers with high
urinary concentrations) due to coelution. As such, the GC-MS
Ethanethiol (reagent grade) was bought from Acros (Geel,
screening method should include as many prohibited substances
Belgium). N-methyl-N-(trimethylsilyl) trifluoroacetamide (MSTFA)
as possible, have a large linear range for the quantitative substances,
was purchased from Chem. Fabrik Karl Bucher (Waldstedt, Ger-
sufficient chromatographic resolution (for accurate quantification of
many); ammonium iodide (NH4I) and dodecane (anhydrous) were
isomers) and a short analysis time. It has to be sensitive enough
obtained from Sigma Aldrich (Bornem, Belgium). Acetonitrile (LC-
(especially for the exogenous compounds) and as cost-effective as
MS grade) and methyl tert-butyl ether (HPLC grade) were from
feasible. GC triple quadrupole MS (GC-QQQ-MS) is capable of
Biosolve (Valkenswaard, The Netherlands). b-Glucuronidase from
meeting these demands and nowadays GC-QQQ-MS instruments
E. coli K12 was obtained from Roche Diagnostics GmbH (Mannheim,
have become the standard GC-MS screening tool, in parallel with
Germany). Sodium hydrogen carbonate (NaHCO3) was from Fisher
many other analytical fields, due to their high sensitivity and excel-
Scientific (Loughborough, UK). Potassium carbonate (K2CO3),
lent linear range [9e16].
disodium hydrogen phosphate (Na2HPO4), sodium dihydrogen
Up to now, high resolution technology in combination with GC
phosphate (NaH2PO4), sodium acetate (NaOAc) and sodium sulfate
has not yet been used as a complete GC-MS screening method for
(Na2SO4) were obtained from Merck (Darmstadt, Germany). XAD-2
doping control analyses. To the best of our knowledge, there are only
was from Serva (Heidelberg, Germany).
two papers that describe the use of GC high resolution as screening
Phosphate buffer (pH 7) was prepared by dissolving 7.1 g of
tool in the doping control field [17,18]. The first paper screens with a
Na2HPO4$2H2O and 1.4 g of NaH2PO4$H2O in 100 mL of deionized
Waters GC orthogonal acceleration time-of-flight MS and the second
water. Carbonate buffer (pH 9.5) was prepared by dissolving 135 g
paper presents a screening method using conventional EI on a Agi-
of K2CO3 and 111 g of NaHCO3 in 900 mL of double-distilled water.
lent 7200 quadrupole time-of-flight MS (QTOF-MS). Both screenings
The origin of all the reference standards is described in detail in our
are primarily for exogenous steroids and a substantial part of the
previous work [3,26].
compounds that need to be covered for a full screening method are
The internal standard (IS) consisted of d3-Testosterone glucu-
not included. Up to now, the available GC high resolution in-
ronide (10 mg mL1), d4-Epitestosterone glucuronide (5 mg
struments were not sensitive enough for a number of compounds
mL1), d4-Androsterone glucuronide (50 mg mL1), d5-
and/or had inadequate linearity to comply with all the required
Etiocholanolone (50 mg mL1), d3-Dihydrotestosterone glucuro-
WADA criteria, meaning that the current GC-QQQ-MS screening
nide (2 mg mL1), d6-Dehydroepiandrosterone (10 mg mL1), d3-5a-
methods could not be replaced by an equivalent method on GC high
Androstane-3a,17b-diol (10 mg mL1), d5-5b-Androstane-3a,17b-
resolution MS. However, high resolution technology has some
diol (10 mg mL1), 17a-Methyltestosteron (5 mg mL1), 19-
54 M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59

Norandrosteron-d4 glucuronide (2 mg mL1), 19- be found in supplementary material S2.


Noretiocholanolone-d4 (2 mg mL1), Nalorphine (2 mg mL1),
Sibutramine (2 mg mL1) and Salbutamol-d3 (10 mg mL1). 2.4. Method validation

2.2. Sample preparation 2.4.1. Qualitative


The method validation of the qualitative compounds was per-
For the analysis of urine samples, the following procedure was formed on 10 randomly chosen blank urine samples covering a
applied: to 500 mL of urine, 25 mL of IS, 1 mL phosphate buffer (0.1 M; normal range of pH, specific gravity and turbidity for anti-doping
pH ¼ 7) and 25 mL of b-glucuronidase from E. coli were added. The samples. Validation was performed by spiking these urine sam-
samples were incubated for 1 h at 56  C. Afterwards, aqueous ples with reference mixtures at MRPL and ½ MRPL to check WADA
NaHCO3/K2CO3 solution (pH ¼ 9.5, 1 mL) and methyl tert-butyl ether MRPL compliance. These experiments were performed on different
(5 mL) were added. Extraction was performed by rolling the test days and with multiple operators to evaluate the robustness of the
tubes during 20 min. The organic layer was transferred to tubes and method under routine operating conditions as well. These experi-
evaporated to dryness under nitrogen. Then, the residue was ments provided data concerning the repeatability and reproduc-
reconstituted in 150 mL acetone, transferred to vials and evaporated ibility. A compound was regarded as validated if the substance
to dryness under nitrogen. To the dried residues, 40 mL of derivati- could be detected in all 10 urine samples at the 1/2 MRPL level (S/
zation mixture was added and the samples were incubated at 80  C N > 3). Selectivity and specificity were tested by analysis of 10 blank
for 30 min. The derivatization mixture consisted of 2 mL of MSTFA/ urine samples. The lowest LOD (LLOD) is defined as the lowest level
ethanethiol/NH4I (500:4:2, v/v/w) and 1 mL of dodecane. that can be detected in an ideal situation (i.e., urine sample con-
taining very few interferences). LLOD is determined by analysis of
2.3. Method and instrumentation one urine sample with a low specific gravity spiked with reference
mixtures at 2, 1, ½, 1/5, 1/10, 1/25 and 1/100 of the MRPL.
An Agilent 7250 GC-QTOF-MS (Agilent Technologies, Palo Alto,
USA) with back flush system was used and 0.8 mL was injected. The 2.4.2. Quantitative
GC was equipped with a 15 m HP ultra 1 column (200 mm I.D., Six point calibration curves (Table 2) were made by spiking
0.11 mm df) (Agilent Technologies, Palo Alto, USA). The column was steroid stripped urine. Steroid stripped urine was prepared by
divided into two parts: 5 m from the inlet to the back flush unit and pouring negative urine from an infant on a preconditioned XAD-2
10 m from the back flush unit to the detector. The back flush system column. Removal of the naturally present androgenic steroids
was installed to extend the lifetime of the column by removing was necessary for correct quantification. Each validation batch
pollutants during the final stage of the GC run that accumulated consisted of 3 calibration curves and 9 additional spiked samples at
during the run on the first column (5 m). These pollutants are flush the lowest (n ¼ 3), midrange (n ¼ 3) and highest (n ¼ 3) calibration
out towards the inlet (and not towards the MS ion source). This also levels. In total, three validation batches were analysed, each by a
minimizes maintenance cost as the 5 m column simultaneously different analyst and on a different day. Unweighted least squares
functions as a precolumn that can easily be replaced (instead of a regression was used to set up the calibration curves. Accuracy,
complete 15 m column). The oven temperature was programmed expressed as bias, was defined as the relative difference between
from 110  C (0.1 min) to 125  C at 75  C min1, then to 165  C at the theoretical value and the measured concentration. Repeat-
25  C min1, then to 185  C at 40  C min1, then to 210  C at 3.5  C ability and reproducibility were calculated as the relative standard
min1, then to 240  C at 15  C min1 and finally to 325  C (0.9 min) deviation (RSD) and expressed as percentages. Acceptable toler-
at 75  C min1. Afterwards, the back flush was performed for ances (%) for precision were calculated from 2/3
0.5 min at 325  C during which the first part of the column (5 m) RSDmax ¼ 2(10.5logC) [27]. Tolerances for the accuracy, expressed as
was back flushed for 25 void volumes. The total GC run time was bias, were set at maximum of 15% [28].
14.1 min. Helium was used as carrier gas at a constant flow rate of To assess external bias, matrix matched urine samples from the
1 mL min1. Analyses were performed with low energy electron WADA external quality assessment scheme (EQAS) were also ana-
impact (LE-EI) at 18 eV. The instrument operated in full scan mode lysed and z-scores were calculated.
from m/z 50 to 750. During 7 small time intervals the instrument
measured both in full scan mode and MS/MS mode by continuous 3. Results and discussion
switching. Table 1 lists the 7 time intervals and the corresponding
precursor ions that were selected in the Q of the QTOF for MS/MS. 3.1. Conversion of CI GC-QQQ-MS to LE-EI GC-QTOF-MS screening
Nitrogen was used as collision gas at a flow rate of 1 mL min1.
Chromatographic and mass spectrometric details of all analytes can The 7250 GC-QTOF-MS has an optimized source configurations,

Table 1
Time intervals during which the GC-QTOF-MS operates in both full scan (m/z 50e750) and MS/MS mode by
continuous switching.

time interval (min) MS/MS mode: precursor ion (m/z) and collision energy (eV)

10.6e10. 95 446 (25)


339 (25)
442 (25)
11.05e11.40 446 (30)
11.40e11.70 460 (45)
11.75e12.01 506 (45)
428 (30)
12.01e12.25 520 (45)
12.55e12.73 640 (35)
12.73e12.95 490 (30)
M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59 55

Table 2
Target substances for quantitative analysis.

Substance Internal standard Calibration levels in Coefficient of RSD and corresponding 2/3RSDmax between Bias at LOQ (%)
urine (ng mL1) determination (r2) brackets at LOQ (%) (n ¼ 18) (n ¼ 18)

Testosterone d3-Testosterone 1-3-10-30-100-400 0.9995 5.8 (30.2) 12.6


glucuronide
Epitestosterone d4-Epitestosterone 1-3-10-30-100-400 0.9997 8.0 (30.2) 3.4
glucuronide
Androsterone d4-Androsterone 24-72-240-720-2400- 0.9993 5.1 (18.7) 9.3
glucuronide 9600
Etiocholanolone d5-Etiocholanolone 24-72-240-720-2400- 0.9987 6.4 (18.7) 3.9
9600
Dihydrotestosterone d3-Dihydrotestosterone 0.5-1.5-5-15-50-200 0.9986 6.8 (33.5) 14.1
glucuronide
Dehydroepiandrosterone d6- 2-6-20-60-200-800 0.9993 5.2 (27.2) 13.9
Dehydroepiandrosterone
4-androstene-3,17-dione d3-Dihydrotestosterone 0.5-1.5-5-15-50-200 0.9988 4.7 (33.5) 11.9
glucuronide
5a-androstane-3a,17b- d3-5a-Androstane- 2-6-20-60-200-800 0.9997 7.0 (27.2) 5.3
diol 3a,17b-diol
5b-androstane-3a,17b- d5-5b-Androstane- 2-6-20-60-200-800 0.9995 4.5 (27.2) 13.6
diol 3a,17b-diol
5a-androstane-3,17- d5-5b-Androstane- 0.5-1.5-5-15-50-200 0.9987 16.0 (33.5) 12.8
dione 3a,17b-diol
5b-androstane-3,17- d5-5b-Androstane- 0.5-1.5-5-15-50-200 0.9986 9.3 (33.5) 5.7
dione 3a,17b-diol
6aOH-androstenedion d3-Dihydrotestosterone 0.25-0.75-2.5-7.5-25- 0.9988 16.1 (37.2) 13.9
glucuronide 100
4OH-androstenedion d3-Dihydrotestosterone 0.25-0.75-2.5-7.5-25- 0.9986 15.5 (37.2) 14.7
glucuronide 100
5b-pregnanediol d5-5b-Androstane- 2-6-20-60-200-800 0.9971 9.5 (27.2) 5.6
3a,17b-diol

lenses, magnets and tuning for LE-EI (12e20 eV) measurements. MS approach for their GC screening [26]. Since 2014, our laboratory
LE-EI allows a softer ionization than conventional EI at 70 eV. Use of uses CI in combination with GC-QQQ-MS because this leads to
LE-EI therefore leads to a higher relative abundance of the high m/z higher sensitivities for more than 90% of the compounds in com-
ions (i.e., ions with high diagnostic value), analogous to chemical parison with EI GC-QQQ-MS [4]. However, CI does also require more
ionization (CI). For a compound such as 6b-OH-4- maintenance and instrument intervention. To convert our current
Chlordehydromethyltestosteron for example, LE-EI at 18 eV leads CI GC-QQQ-MS doping control screening method into an equivalent
to a lower LOD (factor of 2) in comparison with conventional EI at and complete GC high resolution acquisition screening method on
70 eV (Fig. 1). As a consequence, LE-EI at 18 eV was used for the LE- LE-EI GC-QTOF-MS, a small number of chromatographic changes
EI GC-QTOF-MS screening method. were necessary. First, the 12 m HP-1MS (250 mm I.D. x 0.25 mm df)
Most doping control laboratories use a conventional EI GC-QQQ- column was replaced with a 15 m HP ultra 1 (200 mm I.D. x 0.11 mm

Fig. 1. GC-QTOF-MS analysis using LE-EI (18 eV) and conventional EI (70 eV) of a urine sample spiked at different levels of 6b-OH-4-Chlordehydromethyltestosteron.
56 M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59

Fig. 2. Mass accuracy measurements (ppm) of epimetenediol during 48 h: first 100 injections without calibration runs, last 100 injections with calibration runs every 5 injections.
Average mass deviation and 2 x standard deviation in full and dotted line respectively.

df) column. Both columns have an identic stationary phase. The 3.2. Validation
extension in length is compensated by the reduction of the internal
diameter and film thickness, leading to comparable behaviour on 3.2.1. Qualitative
both the CI GC-QQQ-MS screening method and the new LE-EI GC- An overview of all the compounds that are included in the LE-EI
QTOF-MS screening method. Secondly, on the LE-EI GC-QTOF-MS GC-QTOF-MS method can be found in supplementary material S2.
0.8 mL out of 40 mL of sample is injected whereas with the CI GC- All retention times were compliant with WADA's identification
QQQ-MS method, an injection volume of 2 mL out of 60 mL of sam- criteria [2]. The LLOD defines the lowest level that can be detected
ple was used. in an ideal situation. As most prohibited substances have a zero-
One of the most desirable attributes of a high resolution mass tolerance, anti-doping laboratories need to be able to detect sub-
analyser is its mass accuracy. To improve the mass accuracy of the stances at the lowest possible concentration. This concentration is
instrument, it is advisable to execute an external calibration run at not only dependent on the method and the target substance, but
regular intervals. This automatic external calibration run is pro- also on the sample matrix. In a field where results are highly
vided and incorporated in the operation software of the instru- challenged in court, the LLOD of a method needs to reflect the true
ment and does not jeopardize the high throughput characteristics LLOD as much as possible, as otherwise lawyers will claim that it is
of the approach as it only requires 1 min and is executed during GC impossible to detect a substance below the validated LOD. In
cool down time. Accuracy measurements during approximately practice, this can occur as matrix is an important determining
48 h are presented in Fig. 2 and supplementary material S1 for factor. Nevertheless, it is important that the LLOD reflects as much
17a-methyltestosterone (m/z 446.3031 and 301.1982) and epi- as possible a best case scenario. The LLODs of the CI GC-QQQ-MS
metenediol (m/z 143.0887, 216.1873, 325.2686 and 448.3193). and the LE-EI GC-QTOF-MS screening method are listed in sup-
Before the first injection, a calibration run was performed on the plementary material S2. For 14 out of 68 stimulants, the LE-EI GC-
instrument. During the subsequent 99 injections (±24 h) no cali- QTOF-MS has higher LLODs than the CI GC-QQQ-MS and for 7 out of
bration runs were executed. Starting from injection 101 until in- 20 narcotics the LE-EI GC-QTOF-MS has lower LLODs than the CI
jection 200, a calibration run was performed every 5 injections. GC-QQQ-MS. However, for both these classes, the LLOD is not of
The maximum observed deviations between the measured mass utmost importance, as WADA has recommended not to report
and the theoretical mass during the first 100 injections were 4.9, concentrations below ½ MRPL. However, for steroids it is important
3.3, 2.8, 2.3, 4.2 and 4.2 ppm for m/z 446.3031, 301.1982, 143.0887, to have LLODs which are as low as possible, as lower LODs will
216.1873, 325.2686 and 448.3193 respectively. When calibration mean a longer detection window after administration of a sub-
runs are performed at regular intervals (i.e., injection 101 until stance due to decreasing concentrations in function of time. The LE-
200), the maximum observed deviations drop to 4.0, 2.7, 1.4, 1.4, EI GC-QTOF-MS has equal or lower LLODs for 150 out of 187 (80.2%)
3.6 and 3.1 for m/z 446.3031, 301.1982, 143.0887, 216.1873, of the tested compounds than the CI GC-QQQ-MS screening
325.2686 and 448.3193 respectively. The standard deviation method, while previously it has been shown that CI GC-QQQ-MS
(STDEV) of the mass deviations remain roughly the same (Fig. 2 has lower LODs than EI GC-QQQ-MS [4].
and supplementary material S1). Without regular calibration The high resolution open screening allows retrospectivity and
runs, the STDEVs range from 0.46 until 0.62 ppm. With regular compatibility with libraries. In addition, a large amount of com-
calibration runs, the STDEVs range from 0.50 until 0.65 ppm. In pounds can be added to the method, without effecting other
general, the mass analyser has a tendency to deviate towards a simultaneously detected substances. In contrast, the number of
higher mass than the exact mass and the deviation becomes larger compounds that can be added in a GC-QQQ-MS method is
for ions with a higher m/z (Fig. 2 and supplementary material S1). restricted because it is a targeted approach with a limited number
When very high amounts of compound reach the mass analyser of transitions that can be included. In QQQ-MS, the number of
(i.e., samples with very high concentrations), the mass analyser monitored transitions will therefore affect the sensitivity of the
can get saturated leading to a mass shift towards the higher method. Here, a total of 280 compounds was validated for quali-
masses (data not shown). This phenomenon has to be taken into tative detection, most of which in full scan mode. Only 8 com-
account. If the mass window is set too narrow during the pro- pounds required the use of the MS/MS mode to comply with
cessing of the samples, there is a risk of not detecting substances WADA's stringent criteria. All other compounds had sufficient
that are present in very high concentration due to this mass shift. sensitivity and selectivity by using the full scan high resolution
Therefore, a conservative mass window of 20 ppm was used for mode. The list of compounds that is included in the LE-EI GC-QTOF-
the processing of the compounds. This mass window takes into MS screening method is considerably larger than the list of sub-
account the required selectivity for the respective compound and stances that is included in current CI GC-QQQ-MS screening
the expected possible maximum concentration of that compound methods [4,26,29]. For example, diuretics, a class of doping sub-
in urine. stances that is not covered in the CI GC-QQQ-MS screening method,
M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59 57

is included in the LE-EI GC-QTOF-MS screening. In the CI GC-QQQ- in 6 replicates. This procedure was repeated in triplicate (three
MS method, these compounds could not be included due to the different analysts at three different points in time). All STDEVs were
limited number of transitions that can be monitored without acceptable as they were lower than 2/3 of the maximum residual
effecting the sensitivity of other compounds, illustrating the benefit standard deviation (RSDmax ¼ 2(10.5logC)) as defined by Horwitz
of an open screening method. The LE-EI GC-QTOF method is also [27]. The bias at each of these points was below 15% [28]. The limit
capable of screening for HIF stabilizer and SARMS (e.g., Molidustat, of quantification (LOQ) is defined as the lowest calibration point of
andarine, ostarine, LGD-4033). These are relatively new classes of the corresponding calibration curve. All calibration curves had a
doping substances that can be difficult to detect on LC-MS. How- coefficient of determination r2 > 0.98.
ever, they are important doping substances. For example, SARMS Within WADAs EQAS urine samples are frequently sent to
have anabolic properties, in parallel with anabolic steroids, but can accredited laboratories. Fitness for purpose of the method was
be bought legally and cheap, making them easily accessible for assessed by reanalysis of 17 EQAS samples, each in triplicate, and
athletes. the mean of measured values were compared with the reported
Fig. 3 shows the LE-EI GC-QTOF-MS chromatograms of some consensus values. For 110 out of 119 values, the z-scores were lower
compounds that can be challenging if one uses a conventional EI than 1, indicating satisfactory accuracy of the method (supple-
GC-QQQ-MS screening. The LE-EI GC-QTOF-MS screening does not mentary material S3). Only 2 out of 119 measured values had a z-
have any issues to reach the WADA MRPLs for these compounds. score higher than 2 (i.e., sample 1 Epitestosterone z-score ¼ 2.15,
Fig. 4 visualizes the LE-EI GC-QTOF-MS analyses of a random urine sample 4 Testosterone z-score ¼ 2.43) which is acceptable as sta-
sample spiked with long term metabolites (LTM) at different levels. tistically you would only expect 95% of the values to be within a z-
These LTM are metabolites of anabolic steroids that are excreted in score of 2.
(very) low concentrations in urine, but that have very long detec- A comparison between the CI GC-QQQ-MS screening method
tion times. Therefore, they are of great importance for doping and the LE-EI GC-QTOF-MS screening method is shown in Fig. 5 and
control analyses as the extended detection times resulted in a supplementary material S4. On CI GC-QQQ-MS, the endogenous
4e80-fold increase of adverse analytical findings for anabolic ste- steroid concentrations of 100 urine samples analysed over 5
roids [30]. batches was determined. Subsequently, the same samples were
additionally analysed on a second CI GC-QQQ-MS instrument and
3.2.2. Quantitative on the LE-EI GC-QTOF-MS to verify that the bias and the spread on
For most compounds a zero tolerance policy applies in doping the bias is comparable between the CI GC-QQQ-MS analyses and
control. According to WADA regulations, quantitative data is the analyses on the LE-EI GC-QTOF-MS screening method. The re-
required for a defined set of endogenous steroids [8]. For this sults indicate an excellent correlation (r2) and comparability (slope
purpose, the method was validated quantitatively for 14 endoge- between 0.926 and 1.093) between both techniques.
nous steroids, including the two minor metabolites 4OH-androst-
4-ene-3,17-dione (formestane) and 6aOH-androst-4-ene-3,17-dion 3.3. Application to excretion/routine samples
(Table 2) [31,32]. Six point calibration curves were made by spiking
steroid stripped urine. The concentration ranges differ for the The suitability of the developed method was proven by ana-
different compounds according to their naturally occurring con- lysing 105 urine samples previously declared positive for doping
centrations. The low-, mid- and high-level standards were analysed substances and authentic samples from controlled administration

Fig. 3. LE-EI GC-QTOF-MS analysis of a urine sample spiked at the MRPL.


58 M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59

Fig. 4. LE-EI GC-QTOF-MS analysis of a negative urine sample and a urine sample spiked with LTMs at different levels (MRPL ¼ 5 ng mL1). Each column corresponds to the above
mentioned concentration. The circles indicate the lowest detectable concentration for the corresponding LTM.

Fig. 5. Quantification of endogenous steroids in 100 urines samples determined with the CI GC-QQQ-MS screening method and the LE-EI GC-QTOF-MS screening method: cor-
relation between CI GC-QQQ-MS and LE-EI GC-QTOF-MS and visualization of the CI GC-QQQ-MS vs CI GC-QQQ-MS bias and the CI GC-QQQ-MS vs LE-EI GC-QTOF-MS bias.

studies. All positive cases were detected and no false negative The method allows the replacement and transition of the cur-
samples were identified. In addition, 241 blank urine samples were rent GC-QQQ-MS screening method for human urine doping con-
analysed and no false positive samples were identified. trol purposes into an equivalent and complete GC high resolution
acquisition screening method. The high resolution open screening
4. Conclusions allows retrospectivity and compatibility with libraries and, in
addition, a large amount of compounds can be added to the
A comprehensive high resolution open screening method on LE- method. On top of that, the full scan high resolution acquisition
EI GC-QTOF-MS has been developed and validated for the analysis makes it a versatile method and approach that can easily be
of 294 compounds (and 14 internal standards) in urine. All detec- transferred and applied in other related analytical fields. An open
tion limits were compliant with the required WADA MRPLs. The LE- screening method does not rely on a targeted analyte list, allowing
EI GC-QTOF-MS has equal or lower LLODs for 80.2% of the tested the identification of new compounds more quickly, followed by
compounds compared to the CI GC-QQQ-MS screening method. retrospective data analysis without having to reanalyze samples.
M. Polet et al. / Analytica Chimica Acta 1042 (2018) 52e59 59

The performance of the method is not restricted by the amount P. Adsule, Multiresidue determination of 375 organic contaminants including
pesticides, polychlorinated biphenyls and polyaromatic hydrocarbons in fruits
of compounds that is monitored, as is the case for GC-QQQ-MS
and vegetables by gas chromatography-triple quadrupole mass spectrometry
screening. As such, the list of compounds that are enclosed in the with introduction of semi-quantification, J. Chromatogr. A 1270 (2012)
LE-EI GC-QTOF-MS screening method has been expanded consid- 283e295, https://doi.org/10.1016/j.chroma.2012.10.066.
erably in comparison with the CI GC-QQQ-MS screening method. It [15] M. Polet, W. Van Gansbeke, P. Van Eenoo, K. Deventer, Efficient approach for
the detection and identification of new androgenic metabolites by applying
also includes HIF, SARMS and diuretics. These are doping classes SRM GC-CI-MS/MS: a methandienone case study, J. Mass Spectrom. 51 (2016)
that are not covered in the current CI GC-QQQ-MS screening 524e534, https://doi.org/10.1002/jms.3781.
method. [16] M. Polet, L. De Wilde, P. Van Renterghem, W. Van Gansbeke, P. Van Eenoo,
Potential of saliva steroid profiling for the detection of endogenous steroid
abuse: reference thresholds for oral fluid steroid concentrations and ratios,
Funding Anal. Chim. Acta 999 (2018) 1e12, https://doi.org/10.1016/j.aca.2017.11.015.
[17] C.G. Georgakopoulos, A. Vonaparti, M. Stamou, P. Kiousi, E. Lyris, Y.S. Angelis,
G. Tsoupras, B. Wuest, M.W.F. Nielen, I. Panderi, M. Koupparis, Preventive
This work was supported by the Partnership for Clean Compe- doping control analysis: liquid and gas chromatography time-of-flight mass
tition (grant number 2018R1000323G). spectrometry for detection of designer steroids, Rapid Commun. Mass Spec-
trom. 21 (2007) 2439e2446, https://doi.org/10.1002/rcm.3103.
[18] W. Abushareeda, E. Lyris, S. Kraiem, A. Al Wahaibi, S. Alyazidi, N. Dbes,
Appendix A. Supplementary data A. Lommen, M. Nielen, P.L. Horvatovich, M. Alsayrafi, C. Georgakopoulos, Gas
chromatographic quadrupole time-of-flight full scan high resolution mass
Supplementary data related to this article can be found at spectrometric screening of human urine in antidoping analysis, J. Chromatogr.
B 1063 (2017) 74e83, https://doi.org/10.1016/j.jchromb.2017.08.019.
https://doi.org/10.1016/j.aca.2018.08.050.
[19] K. Deventer, O.J. Pozo, A.G. Verstraete, P. Van Eenoo, Dilute-and-shoot-liquid
chromatography-mass spectrometry for urine analysis in doping control and
References analytical toxicology, TrAC Trends Anal. Chem. (Reference Ed.) 55 (2014)
1e13, https://doi.org/10.1016/j.trac.2013.10.012.
[1] WADA, The 2018 Prohibited List International Standard, World Anti-Doping [20] A. Musenga, D.A. Cowan, Use of ultra-high pressure liquid chromatography
Agency, 2018. coupled to high resolution mass spectrometry for fast screening in high
[2] WADA, Technical Document Minimum Required Performance Limits for throughput doping control, J. Chromatogr. A 1288 (2013) 82e95, https://
Detection of Prohibited Substances: TD2018MRPL, World Anti-Doping doi.org/10.1016/j.chroma.2013.03.006.
Agency, 2018. [21] S. Odoardi, V. Valentini, N. De Giovanni, V.L. Pascali, S. Strano-Rossi, High-
n, K. Deventer, K. Roels, P. Van Eenoo, Development and validation of
[3] A.J. Giro throughput screening for drugs of abuse and pharmaceutical drugs in hair by
an open screening method for diuretics, stimulants and selected compounds liquid-chromatography-high resolution mass spectrometry (LC-HRMS),
in human urine by UHPLC-HRMS for doping control, Anal. Chim. Acta 721 Microchem. J. 133 (2017) 302e310, https://doi.org/10.1016/
(2012) 137e146, https://doi.org/10.1016/j.aca.2012.02.002. j.microc.2017.03.050.
[4] W. Van Gansbeke, M. Polet, F. Hooghe, C. Devos, P. Van Eenoo, Improved [22] W. Abushareeda, A. Vonaparti, K. Al Saad, M. Almansoori, M. Meloug, A. Saleh,
sensitivity by use of gas chromatographydpositive chemical ionization triple R. Aguilera, Y. Angelis, P.L. Horvatovich, A. Lommen, M. Alsayrafi,
quadrupole mass spectrometry for the analysis of drug related substances, C. Georgakopoulos, High resolution full scan liquid chromatography mass
J. Chromatogr. B 1001 (2015) 221e240, https://doi.org/10.1016/ spectrometry comprehensive screening in sports antidoping urine analysis,
j.jchromb.2015.07.052. J. Pharmaceut. Biomed. Anal. 151 (2018) 10e24, https://doi.org/10.1016/
[5] M. Mazzarino, X. de la Torre, F. Botre , A screening method for the simulta- j.jpba.2017.12.025.
neous detection of glucocorticoids, diuretics, stimulants, anti-oestrogens, [23] I. Ojanper€a, M. Kolmonen, A. Pelander, Current use of high-resolution mass
beta-adrenergic drugs and anabolic steroids in human urine by LC-ESI-MS/ spectrometry in drug screening relevant to clinical and forensic toxicology
MS, Anal. Bioanal. Chem. 392 (2008) 681e698, https://doi.org/10.1007/ and doping control, Anal. Bioanal. Chem. 403 (2012) 1203e1220, https://
s00216-008-2292-5. doi.org/10.1007/s00216-012-5726-z.
[6] M. Polet, W. Van Gansbeke, L. Geldof, K. Deventer, P. Van Eenoo, Identification [24] H.H. Maurer, M.R. Meyer, High-resolution mass spectrometry in toxicology:
and characterization of novel long-term metabolites of oxymesterone and current status and future perspectives, Arch. Toxicol. 90 (2016) 2161e2172,
mesterolone in human urine by application of selected reaction monitoring https://doi.org/10.1007/s00204-016-1764-1.
[25] F. Hernandez, T. Portole
s, E. Pitarch, F.J. Lo
 pez, Gas chromatography coupled to
GC-CI-MS/MS, Drug Test. Anal. (2017), https://doi.org/10.1002/dta.2183.
[7] D.H. Catlin, M.H. Sekera, B.D. Ahrens, B. Starcevic, Y.-C. Chang, C.K. Hatton, high-resolution time-of-flight mass spectrometry to analyze trace-level
Tetrahydrogestrinone: discovery, synthesis, and detection in urine, Rapid organic compounds in the environment, food safety and toxicology, TrAC
Commun. Mass Spectrom. 18 (2004), https://doi.org/10.1002/rcm.1495, Trends Anal. Chem. (Reference Ed.) 30 (2011) 388e400, https://doi.org/
1245e1049. 10.1016/j.trac.2010.11.007.
[8] WADA, Technical Document Endogenous Anabolic Androgenic Steroids: [26] P. Van Eenoo, W. Van Gansbeke, N. De Brabanter, K. Deventer, F.T. Delbeke,
TD2016EAAS, World Anti-Doping Agency, 2016. A fast, comprehensive screening method for doping agents in urine by gas
[9] Z. He, Y. Wang, L. Wang, Y. Peng, W. Wang, X. Liu, Determination of 255 chromatography-triple quadrupole mass spectrometry, J. Chromatogr. A 1218
pesticides in edible vegetable oils using QuEChERS method and gas chroma- (2011) 3306e3316, https://doi.org/10.1016/j.chroma.2010.09.082.
tography tandem mass spectrometry, Anal. Bioanal. Chem. 409 (2017) [27] W. Verwaal, M. van Bavel, A. Boot, J. Bravenboer, F. de Goei, C. Maas, A. van der
1017e1030, https://doi.org/10.1007/s00216-016-0016-9. Putten, Valideren van (fysisch-) chemische en (fysisch-) mechanische meth-
[10] K. Kalachova, T. Cajka, C. Sandy, J. Hajslova, J. Pulkrabova, High throughput oden op het niveau van de wettelijke eis, Ware(n) Chem. 26 (1996) 106.
sample preparation in combination with gas chromatography coupled to [28] F. Bressolle, M. Bromet-Petit, M. Audran, Validation of liquid chromatographic
triple quadrupole tandem mass spectrometry (GC-MS/MS): a smart procedure and gas chromatographic methods. Applications to pharmacokinetics,
for (ultra)trace analysis of brominated flame retardants in fish, Talanta 105 J. Chromatogr. B Biomed. Appl. 686 (1996) 3e10. http://www.ncbi.nlm.nih.
(2013) 109e116, https://doi.org/10.1016/j.talanta.2012.11.073. gov/pubmed/8953186. (Accessed 25 September 2014).
[11] M. Monteleone, A. Naccarato, G. Sindona, A. Tagarelli, A reliable and simple [29] N. De Brabanter, W. Van Gansbeke, L. Geldof, P. Van Eenoo, An improved gas
method for the assay of neuroendocrine tumor markers in human urine by chromatography screening method for doping substances using triple quad-
solid-phase microextraction-gas chromatography-triple quadrupole mass rupole mass spectrometry, with an emphasis on quality assurance, Biomed.
spectrometry, Anal. Chim. Acta 759 (2013) 66e73, https://doi.org/10.1016/ Chromatogr. 26 (2012) 1416e1435, https://doi.org/10.1002/bmc.2714.
[30] H. Geyer, W. Scha €nzer, M. Thevis, Anabolic agents : recent strategies for their
j.aca.2012.11.017.
[12] A. Thomas, C. Widmer, G. Hopfgartner, C. Staub, Fast gas chromatography and detection and protection from inadvertent doping, Br. J. Sports Med. 48 (2014)
negative-ion chemical ionization tandem mass spectrometry for forensic 820e826, https://doi.org/10.1136/bjsports-2014-093526.
analysis of cannabinoids in whole blood, J. Pharmaceut. Biomed. Anal. 45 [31] M. Polet, P. Van Renterghem, W. Van Gansbeke, P. Van Eenoo, Studies on the
(2007) 495e503, https://doi.org/10.1016/j.jpba.2007.08.019. minor metabolite 6a-hydroxy-androstenedione for doping control purposes
[13] L. Truta, A.L. Castro, S. Tarelho, P. Costa, M.G.F. Sales, H.M. Teixeira, Antide- and its contribution to the steroid profile, Drug Test. Anal. 6 (2014) 978e984,
pressants detection and quantification in whole blood samples by GC-MS/MS, https://doi.org/10.1002/dta.1618.
for forensic purposes, J. Pharmaceut. Biomed. Anal. 128 (2016) 496e503, [32] M. Polet, P. Van Renterghem, W. Van Gansbeke, P. Van Eenoo, Profiling of
https://doi.org/10.1016/j.jpba.2016.06.027. urinary formestane and confirmation by isotope ratio mass spectrometry,
[14] K. Banerjee, S. Utture, S. Dasgupta, C. Kandaswamy, S. Pradhan, S. Kulkarni, Steroids 78 (2013), https://doi.org/10.1016/j.steroids.2013.07.011.

S-ar putea să vă placă și