1

Flow Calibration with Adjustment


30352612-3674059
¸
3003802542 FCP-6.H
Purchase order number Calibration rig

US-3005439331-10 / Endress+Hauser Flowtec 132.2774 lb/min ( ≙ 100%)


Order Nº/Manufacturer Calibrated full scale

8FE25-AABSAASAB1NN Service interface


Order code Calibrated output

LPGmass 1" 2.4200


Transmitter/Sensor Calibration factor

KA01B816000 -16
Serial Nº Zero point

- 71.3 °F
Tag Nº Water temperature

Flow Flow Duration m target m meas. ∆ o.r.* Measured error % o.r.


[%] [lb/min] [s] [lb] [lb] [%] Tolerance limit: ±0.2% o.r.* ± Zero stability

24.9 32.871 30.1 16.466 16.467 0.01 1.5

24.9 32.942 30.1 16.505 16.506 0.01


1
100.6 133.021 30.1 66.837 66.837 0.00
100.9 133.463 30.1 67.051 67.056 0.01 0.5

- - - - - - 0
- - - - - -
- - - - - - -0.5

- - - - - - -1
- - - - - -
- - - - - - -1.5

*o.r.: of reading 0 10 20 30 40 50 60 70 80 90 100 [%]


Flow

For detailed data concerning output specifications of the unit under test, see Technical Information (TI), chapter Performance characteristics.
The calibration is traceable to the N.I.S.T. through standards certified at preset intervals.
Endress+Hauser Flowtec operates ISO/IEC 17025 accredited calibration facilities in Reinach (CH), Cernay (FR), Greenwood (USA),
Aurangabad (IN) and Suzhou (CN).

10-09-2015 W. Sprague
Date of calibration Operator

Endress+Hauser Flowtec, Division USA Certified acc. to


2330 Endress Place ISO 9001, Reg.-Nº 030502.2
Greenwood, IN 46143 ISO 14001, Reg.-Nº EMS561046