Documente Academic
Documente Profesional
Documente Cultură
Sources of Error
Offset
Full Scale Span
Pressure Non-Linearity
Accuracy Total
Pressure Hysteresis Error
BFSL
Pressure Non-Repeatability Band
Thermal Effect on Offset
Thermal Effect on Span
Thermal Hysteresis
Characteristic Parameter
Supply voltage -0.3 Vdc to 11.0 Vdc
CAUTION CAUTION
IMPROPER USE PRODUCT DAMAGE
Do not use these products to sense liquid flow. Do not dissamble these products.
Failure to comply with these instructions may result in Failure to comply with these instructions may result in
product damage. product damage.
Shock 30 g, 6 ms
Substrate PCB
Adhesives epoxy
Pneumatic connection manifold mount, 22 mm OD tapered male fitting, G 3/8 female threaded fitting
For example, HAFUHM0020L4AXT defines a Honeywell ZephyrTM Airflow Sensor, unidirectional flow, long port, manifold mount, 20 SLPM, I2C output with
custom 0x49 address, 10% to 90% transfer function, 3 Vdc to 10 Vdc supply voltage.
HAF U HM 0020 L 4 A X T
Product Series Flow Direction Port Style Flow Range Unit Output Format Transfer Reserved for Supply Voltage
Function Future Use
HAF Series—
Digital I2C address:
High Accuracy U Unidirectional 0010 10 L SLPM 2 0x29
A 10% to 90% of input X XXXXX T 3 Vdc to 10 Vdc
Manifold
Airflow Sensor HM mount Digital I2C address:
0015 15 3 0x39
0300 300
1
Apart from the general configuration required, other customer-specific requirements are also possible. Please contact Honeywell.
Figure 4. Nominal Digital Output: 10, 15, 20, 50, 100, 200, 300 SLPM
18018
16380
14742
11466
9828
8190
6552
4914
3276
1638
0
0 100
Full Scale Flow (%)
4.00% 8.00%
TEB = 7.00%
3.00% 7.00% TEB = 4.67% at 300 SLPM
at 200 SLPM
2.00% 6.00%
Error (%FS)
5.00%
1.00% TEB = 2.65%
at 200 SLPM
4.00%
0
3.00% TEB = 0.5%
-1.00% at 43 SLPM
2.00%
-2.00%
Error (%FS)
1.00%
-3.00%
0
-4.00%
0 10 20 30 40 50 60 70 80 90 100 -1.00%
%FS
Ideal -2.00%
-3.00%
Accuracy:
0%FS to 14.3%FS = 0.5% -4.00%
14.3%FS to 100%FS = 3.5%
-5.00%
Total Error Band:
0%FS to 12.5%FS = 0.5% -6.00%
12.5%FS to 100%FS = 4.0%
-7.00%
-8.00%
0 10 20 30 40 50 60 70 80 90 100
%FS
Ideal
Accuracy:
0%FS to 14.3%FS (0 SLPM to 43 SLPM) = 0.5%
14.3%FS to 100%FS (43 SLPM to 300 SLPM) = 3.5%
Figure 6. Flow vs Pressure Drop: 10, 15, 20, 50, 100, 200, 300 SLPM
Flow Typical Pressure Drop
(SLPM) mbar inH2O kPa
0 0.000 0.000 0.000
1 0.103 0.042 0.010
2 0.206 0.082 0.021
4 0.396 0.159 0.040
8 0.803 0.322 0.080
17.237
14 1.549 0.621 0.155
13.790 15 1.686 0.676 0.169
Digital Interface
For additional details on the use of Zephyr with digital output see the Technical Note “I2C Communications with Honeywell
Digital Airflow Sensors”.
The sensor uses the I2C standard for digital communication with a slave address specified in the Nomenclature and Order Guide
in Figure 2. Following sensor power-up, each of the first two read sequences shown in Figure 7 will respond with 2 bytes of the
unique 4-byte Serial Number. The first read after power-up will respond with the two most significant bytes of the Serial Number,
while the second read will respond with the two least significant bytes of the Serial Number. For reliable performance, allow
sensor to be powered for the sensor startup time before performing the first read, then allow a 10 ms command response time
before performing the second read.
After the power-up read sequence described above, the sensor will respond to each I2C read request with a 16-bit (2 byte)
digital flow reading. Read requests taken faster than the Response Time (1 ms) are not guaranteed to return fresh data. The first
two bits of each flow reading will be ‘00’, while non-flow responses (such as error and status codes) will begin with ‘11’. There
are several user commands available as shown in Table 8. Following an I2C write sequence of a user command, the sensor will
respond to the next I2C read request with a 16-bit response. Possible responses to user commands can be seen in Table 9.
Table 8. User Command Descriptions
Command Byte Command Response Time
Command Name Command Description
(Hexadecimal) (Max.)
Next two read requests will each return two bytes of
0x01 GetSerialNumber 10 ms
the sensor’s unique 4-byte Serial Number.
0x02 PowerOnReset Force Power-On reset of sensor microcontroller. 20 ms
Calculates EEPROM Checksum and compares to
production Checksum value. If the values match,
0x03 Checksum 1s
the next read request will respond with 0xCCA5.
Otherwise, the next read will respond with 0xCC90.
1
Table 9. Sensor Response Descriptions
Sensor Response
Response Name Response Description
(Hexadecimal)
0xCCA5 POSACK non-response command was executed successfully
0xCC99 BadCommand command byte was not recognized
0xCC9A BadParam command sent with incorrect parameter bytes
0xCC9B Failure command failed during execution
0xCC90 BadChecksum checksum did not match stored value
0xCCBB Busy sensor is busy calculating the checksum value
The maximum sink current on SCL or SDA is 2 mA. Therefore, if the pull-up resistors are biased by VDD, and if VDD reaches
the maximum supply voltage of 6 V, then the pull-up resistors for SCL and SDA must be greater than 3.0 kOhm to limit the sink
current to 2 mA. The typical value for SCL and SDA pull-up resistors is 4.7 kOhm (this value depends on the bus capacitance and
the bus speed).
8 Sensing and Internet of Things
Honeywell ZephyrTM Digital Airflow Sensors
HAF Series—High Accuracy: 10, 15, 20, 50, 100, 200, 300 SLPM
22,0
54,4 [0.87] 44,0
[2.14]
14,5 33,0 [1.73]
[0.57] [1.30] 54,4
44,0 [2.14]
[1.73]
B 8
[0.3]
4,0 63,2
[2.49] 2X 13,0 DIA.
[0.16] [0.51]
51,5 87
[2.03] [3.4]
D
54
34,5 [2.1]
[1.36]
A Pin 1.
33
[1.3]
42,4
4X 4,50
[0.177] B 4X 10-32 pan head screws 1,13 N m [20 in-lb] torque.
[1.67]
C 2X Gland for O-Ring AS568-113, 13,94 mm ID x 2,62 mm
[0.549 in ID x 0.103 in W. Two O-rings, AS568A-113 Durometer
22,0 A65 to A80 Silicon or Viton, are required to seal sensor to
[0.87] C
manifold. O-rings are not included.
4,0 63,2
[0.16]
87
[2.49] D Flow channel.
[3.4]
18,3
[0.72]
Mounting Footprint
21,0 2X 5,70
[0.83] [0.22]
14 39.5
[0.6] [1.56]
14,8
[0.58] 14,8
[0.58]
2X 5,20
A [0.205] 14,5 A
[0.57]
54,4
[2.14] 44,0
14,5 35,1 [1.73]
[0.57] 44,0 [1.38]
[1.73] 54,4
[2.14]
2X 18,3
[0.72]
B 2X 21,0 68
[0.83] [2.7]
C
54
[2.1]
18,0 51,5
[0.71] [2.03] A Pin 1.
16 39,5
[0.63] [1.56] Mounting Footprint
2X 18,3 16 39,5
2X 1,5 [0.72] [0.63] [1.56]
[0.06]
2X 5,70 2X 1.5 14,8
[0.22] [0.06] [0.58]
14,8 A
[0.58]
A
44,0
14,5 35,1 [1.73]
2X 5,20 [0.57] [1.38]
[0.205] 44,0 54,4
[1.73] [2.14]
54,4
14,5 [2.14]
[0.57]
2X 18,3
71 [0.72]
[2.8]
B
C
A Pin 1.
51,5
[2.03]
B 2X 10-32 pan head screws 1,13 N m [20 in-lb] torque.
18,0
34,5
[0.71] 54
[2.1]
C ISO 1179 G3/8, recommended torque is 1,7 ±0,6 N m
[1.36] [15 ±5 in lb]. Electrical connector part number is JST
(Japan Solderless Terminals) B6B-PH-K-S. Mating
30,0 connector part numbers are JST PHR-6 (socket) and
[1.18] 35 2X 4,50
[1.4] [0.177] JST SPH-002T-P0.5L (crimp terminal).
71,4
[2.81] 42,4
[1.67]
Warranty/Remedy
Honeywell warrants goods of its manufacture as being free
of defective materials and faulty workmanship during the
applicable warranty period. Honeywell’s standard product
warranty applies unless agreed to otherwise by Honeywell in
writing; please refer to your order acknowledgment or consult
your local sales office for specific warranty details. If warranted
goods are returned to Honeywell during the period of coverage,
Honeywell will repair or replace, at its option, without charge
those items that Honeywell, in its sole discretion, finds defective.
The foregoing is buyer’s sole remedy and is in lieu of all
other warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
For more information
Honeywell Sensing and Internet of While Honeywell may provide application assistance personally,
Things services its customers through a through our literature and the Honeywell web site, it is buyer’s
worldwide network of sales offices and sole responsibility to determine the suitability of the product in
distributors. For application assistance, the application.
current specifications, pricing or the
nearest Authorized Distributor, visit Specifications may change without notice. The information we
sensing.honeywell.com or call: supply is believed to be accurate and reliable as of this writing.
Asia Pacific +65 6355-2828 However, Honeywell assumes no responsibility for its use.
Europe +44 (0) 1698 481481
USA/Canada +1-800-537-6945
Authorized Distributor
Honeywell:
HAFUHH0200L4AXT HAFUHT0020L4AXT HAFUHM0200L4AXT HAFUHH0020L4AXT HAFUHM0020L4AXT
HAFUHT0200L4AXT HAFUHM0200L5AXT HAFUHT0100L4AXT HAFUHH0300L4AXT HAFUHT0010L4AXT
HAFUHM0300L4AXT HAFUHT0050L4AXT HAFUHM0050L4AXT HAFUHM0010L4AXT HAFUHM0100L4AXT
HAFUHH0015L4AXT HAFUHM0015L4AXT HAFUHT0300L4AXT HAFUHT0015L4AXT HAFUHH0050L4AXT
HAFUHH0100L4AXT HAFUHH0010L4AXT HAFUHT0200L2AXT HAFUHM0020L5AXT HAFUHH00200L4AXT
HAFUHT0050L2AXT HAFUHT0050L5AXT HAFUHT0050L3AXT HAFUHT0050L6AXT HAFUHT0100L6AXT
HAFUHH0020L3AXT HAFUHT0015L2AXT