Sunteți pe pagina 1din 14

S-Mass® Meter

S-Mass Series Coriolis


Mass Flow and Density Meters and
SMT Series Transmitters
Specifications
Issue/Rev. 0.5 (3/04) Bulletin SS0M019

FMC Measurement Solution’s S-Mass Series Corio-


lis Mass Flow and Density Meters use the Coriolis
effect to directly measure the mass flow of virtually any
liquid. Mass measurement is independent of the fluid
parameters and the ambient conditions in which the
measurement is made. Therefore, the measurement is
essentially unaffected by changes in temperature, pres-
sure, density, viscosity, and flow profile. The meter is
also a highly-accurate vibrating tube densitometer for
measuring process fluid density.
The microprocessor-based electronics measures the
natural frequency of vibration of the sensor flow tubes.
This information is used along with the calibration con-
stants to determine the on-line fluid density.
The SMT050 pulse transmitter is a full-featured product
including multiple frequency/pulse outputs for the vari-
S-Mass Models S25, S50, S100 and S200
ous process variables measured by the sensor.
The SMT100 programmable transmitter is a full-fea-
tured product including multiple outputs for the various Specifications
process variables measured by the sensor including
frequency and analog outputs. Mass Flow (S25, S50, S100, and S200)
In addition, these transmitters incorporate HART com- Accuracy: ±[0.1% + (Zero Stability/Flow Rate x 100)].
munications for ease of configuration and calibration in Repeatability: ±[0.044% + ½ (Zero Stability/Flow Rate)
the field. x 100] % of rate
Zero Stability:
Features S25 - 0.0014 lb/min (0.00064 kg/min).
S50 - 0.011 lb/min (0.005 kg/min).
Direct Mass Measurement - Allows continuous, online, S100 - 0.044 lb/min (0.019 kg/min).
non-inferential measurement of mass—accurate regard-
less of changes in process conditions. S200 - 0.175 lb/min (0.079 kg/min).
Dual S-Tubes - Patented geometry produces a high Maximum Flow Rates:
signal-to-noise ratio and maximum response to Coriolis S25: 0 to 27.5 lb/min (0 to 3 gal/min)
forces for high immunity to external vibration. S50: 0 to 220 lb/min (0 to 26 gal/min)
High Performance - A rated accuracy of ±0.10% and S100: 0 to 882 lb/min (0 to 106 gal/min)
a repeatability of ±0.044% at maximum flow rate. S200: 0 to 3,527 lb/min (0 to 423 gal/min)
Easy Installation - Mounts in line with process piping.
No straight pipe runs required upstream or downstream. Density (S25):
Can be installed in vertical or horizontal piping. Smith Range: 0.2510 to 2.500 g/cc.
Meter S-Mass Models S50, S100, and S200 Accuracy: ±0.001 g/cc from 0.5 to 2.0 g/cc.
Wetted 316L Stainless Steel handles most petroleum ±0.002 g/cc over full range.
applications.
Repeatability: ±0.0005 g/cc.
Density (S50, S100, S200):
Features (S25)
Range: 0.2510 to 2.5000 g/cc.
Secondary Containment - The standard enclosure of Accuracy: ±0.0005 g/cc from 0.6 to 1.16 g/cc.
the S25 sensor is a secondary containment vessel ±0.001 g/cc over full range.
good to 300 psig (2,068 kPa). Repeatability: ±0.00025 g/cc.
Elastomers - Selections of Buna-N, Viton, and Zalak.

The Most Trusted Name In Measurement


Temperature (S25): Approvals for SMT050 Transmitter
Ranges: Fluid: -31°F to 320°F (-35°C to 150°C). Integral Transmitter (NEMA 4X)
Ambient: -31°F to 149°F (-35°C to 65°C) with “piggy- North America
back” Micro-Pak transmitter module. Class I, Div 2, Groups C, D
-31°F to 302°F (-35°C to 150°C) with internally con- European Community
nected Teflon wire. Zone 2
Accuracy: ±0.4°C (±0.7°F). Ex nA II T4
Temperature (S50, S100, S200): Remote Transmitter
North America
Range: Fluid Temperature:
Class I, Division 2, Groups C, D
-60°F to 275°F (-51°C to 135°C) Standard.
European Community
-60°F to 400°F (-51°C to 200°C) Optional for S200 sensor. Zone 2
Accuracy: ±0.45°F (±0.25°C). Ex nA II T4
Pressure Rating: Remote Transmitter
Flow Tube North America
S25: 1,200 psi (82 bar) Class I, Division 1, Groups C, D
Zone 0
S50: 1,800 psi (124 bar)
Ex d (ia) IIB T4
S100: 1,200 psi (82 bar)
S200: 800 psi (55 bar) Approvals for SMT100 Transmitter (UL/CUL only)
Integral Transmitter (NEMA 4X)
Material of Construction:
North America
Flow Tube
Class I, Div 2, Groups C, D
S25: Hastelloy C
Remote Transmitter
S50 – S200: 316 L Stainless Steel North America
End Connections Class I, Division 2, Groups C, D
S25: 316 L Stainless Steel (standard)/Hastelloy C Remote Transmitter
(option) North America
S50 – S200: 316 L Stainless Steel Class I, Division 1, Groups C, D
Sensor Housing * Approvals for S-Mass sensor S25 pending.
300 Series Stainless Steel, Seal-Welded, and Back-
Filled with an Inert Gas SMT050 Transmitter
Environmental Rating: IP65, DIN 40050, NEMA 4X (Micro-Pak Only - No Barrier Board)
Electronics Housing
Acrylic, Enamel-Coated Aluminum / NEMA 4X/IP66 Outputs:
Explosion Proof / NEMA 7/4 Frequency/Pulse Outputs
Three frequency/pulse outputs are programmable as
follows:
Approvals
Output 1: Mass or Volume low rate.
UL and CUL Output 2: Quadrature of Output 1, Density, or Inactive.
ATEX Output 3: Temperature and Status, or Status only.
Approvals for S-Mass Sensors* Open drain, 30 Vdc maximum, 100 mA maximum, 2.5
KHz maximum.
Sensor with Integral Transmitter (NEMA 4X)
North America Mass and volume outputs are unscaled and propor-
tional to mass or volume flow rate.
Class I, Div 2, Groups C, D
Flow sensor K-factor may be multiplied 1, 2, 4, 8, or 16
European Community to optimize mass or volume output.
Zone 2
Density output is scaled to 100 Hz per 1.00 g/cc (0.1 Hz
Ex nA II T4 per 1.0 kg/m3)
Sensor with Remote Transmitter Minimum frequency approximately 30 Hz (0.3000 g/cc)
North America
Maximum frequency approximately 250 Hz (2.5000 g/cc)
Class I, II, III, Division 1, Groups C, D, E, F, G
Temperature output is scaled to 1.00 Hz/°K
European Community
Minimum frequency approximately 30 Hz
Zone 0 and Zone 1 (-450°F, -243°C, 30°K)
EEx ia IIB T6

Page 2 • SS0M019 Issue/Rev. 0.5 (3/04)


Maximum frequency approximately 540 Hz Communications:
(512°F, 267°C, 540°K) Intrinsically safe, 3-wire, full duplex, EIA-232 interface.
Status output overrides the temperature output if it is Communications replicated over Safe Area I/O lines.
enabled. Connect directly to Micro-Pak serial interface.
Status alarms indicate one of the following conditions: Zero Output:
1. Zeroing is in progress Intrinsically safe output replicates a Safe Area active
high input. High impedance – active high output.
2. Sensor tube vibration not sufficient (drive is out of
range) Connect directly to Micro-Pak Zero In signal.
3. Temperature probe failure (internal flow tube rtd) Power Out:
Galvanic isolated intrinsically safe power output.
4. Unbalance of the flow tubes
15.75 VDC maximum (open circuit voltage).
Additional Functions 200 mA maximum (short circuit current).
Zero Input Source resistance is 62 ohms.
Zero Input is active high. Zero in signal forces the
Connect directly to Micro-Pak input power terminals.
Micro-Pak to re-calibrate to a zero flow condition.
Power Requirements
Communications Output 12-30 Vdc (minimum and maximum) fuse protected 16
Digital Interface – an EIA-232 compatible port allows
Watt power supply required.
multidrop digital communications between a supervi-
sory device and several Micro-Paks.
Baud rate: 2,400. SMT100 Transmitter (CFM Board)
Power Requirements
12-30 Vdc (minimum and maximum), fuse protected 16 Outputs:
Watt power supply required. Analog
Note: Use Galvanic Barrier for intrinsically safe applications. Two analog outputs independently configured for mass
flow, volume flow, density, or temperature. Selectable
as 4-20 or 0-20mA. Internally-powered loops isolated to
SMT050 Transmitter ±650 VAC, 1000 ohm load. Over-range capability to
(Galvanic Barrier Board) 22mA.
Frequency/Pulse
Outputs: Two scaled frequency/pulse outputs can be configured
Safe Area I/O for mass or volume flow rate, independent of analog
outputs.
Frequency/Pulse Outputs:
Three frequency/pulse outputs replicate the three haz- Internally-powered or open collector, 21 VDC in active
ardous area inputs. Open drain, 30 VDC max, 100 mA state with no load, 30 VDC compliant in passive state.
max, 2.5 kHz max. Mimic Outputs 1, 2, and 3 from the Open collector maximum sink at 100 mA, with mini-
Micro-Pak mum 270 ohm pull-up resistor connected to maximum
supply voltage of 30 volts. Isolated to ±650 VAC. Scal-
Signal Pull-Ups: able to 10,000 Hz.
Two uncommitted internal 5 kohm resistors tied to input
power. Use as pulse signal pull-ups or status LED Frequency output greater than 500 Hz has a 50% duty
power. Connect by wiring together screw terminals. cycle with 25 microsecond minimum pulse width. Over-
range capability to 15,000 Hz. Frequency/pulse output
Communicatons: 1 operates in continuous quadrature mode, 90 degrees
3-wire, full duplex, multi-drop capable, EIA-232 inter- out of phase with frequency/pulse output 2.
face. Serial communications replication. See Bulletin
MN0M008 for S-Mass protocol details. Discrete Output
Two outputs, both represent fault, zero in progress, or
Zero Input: flow direction. Open collector maximum current sink at
Active high Zero In signal replicated at Zero Out termi- 100 mA, maximum supply of 30 volts. Optically iso-
nal. Dry contact type connection to input power supply. lated to ±650 VAC.
Power In: Digital
12 to 24 VDC ± 10%, 10 W power supply required. HART® Bell 202 serial standard superimposed on the
Hazardous Area I/O secondary mA output. Signal is available for interface
Discrete Inputs: to HART® communicator or host system at 1200 baud.
Three discrete intrinsically safe inputs. Dry contact type Requires 250 to 1,000 ohms resistance.
signals are replicated to the Safe Area I/O discrete
outputs. Connect directly to Micro-Pak outputs.

Issue/Rev. 0.5 (3/04) SS0M019 • Page 3


Additional Functions
Discrete Input indication during a fault condition. Downscale values
Two inputs, one for remote zero, the other for weights are 0 mA, 2 mA, and 0 Hz. Upscale values are 22 mA
and measures security. Input voltage range 5 to 30 and 15,000 Hz.
VDC, 30 mA maximum. Pickup voltage 5 VDC mini- Loop Test
mum. Drop out voltage 1 VDC maximum. Signal must Each output can be commanded to produce an output
be present for 10 microseconds minimum. signal for loop test. Each analog output can be com-
Damping manded to output between 0 and 22 mA. The frequency
Damping is provided independently for mass flow, vol- outputs can be commanded to output between 0 and
ume flow, density, and temperature. Discrete damping 15,000 Hz.
time selections smooth output signals. Damping time Total and Inventory
specifies time to reach 90% of actual signal. Independent, resettable totalizers and non-resettable
Low Flow Cutoff inventory totals are provided for both mass and volume.
Mass flow and volume flow can be independently as- Power
signed low flow cutoff values. Flow rates below the low 12-30 VDC (minimum and maximum), fused protected
flow cutoff value will default outputs to the level repre- 16 Watt power supply required.
senting zero flow.
Fault Conditions
Analog 1, analog 2, and the frequency outputs are each
independently configurable for upscale or downscale

Page 4 • SS0M019 Issue/Rev. 0.5 (3/04)


Apparent Pressure Drop

S25 S50
100.0
100.0
80.0
80.0
60.0
60.0
cp
40.0 00
40.0 10
cp
0
50
cp
20.0 0
20.0 25

cP
Pressure Drop, psi

Pressure Drop, psi


10
10.0 cP 10.0
0

cp
8.0 10

50
8.0
cp

6.0
50

6.0
cp
4.0 25 4.0
P
5c

2.0 2.0
cP
10

1.0
1.0
0.8
0.8
P
1c

0.6
0.6

cP
1
0.4
0.4

0.2
0.2

0.1
0.1
0.2 0.4 0.6 0.8 1.0 2.0 4.0 6.0 8.0 10 20 30
0.2 4.0 6.0 8.0 10 20 40 60 80 100 200
Flow Rate, lb/min Flow Rate, lb/min

S100 S200
100.0 100.0
80.0 80.0

60.0 60.0

40.0 40.0

cp
00
10
20.0 20.0
cp
0
50
cp
Pressure Drop, psi

Pressure Drop, psi

cp 00
10.0 0 10.0 10
20
8.0 8.0 cp
0
cP 50
0
6.0 10 6.0
cp
cp

0
25
50

4.0 4.0

cP
0
10
2.0 2.0
cP

cP
25

25

1.0 1.0
0.8 0.8
0.6
P

0.6
1c

P
1c

0.4 0.4

0.2 0.2

0.1 0.1
10 20 40 60 80 100 200 400 600 800 60 80 100 200 400 600 800 1000 2000 3000

Flow Rate, lb/min Flow Rate, lb/min

Issue/Rev. 0.5 (3/04) SS0M019 • Page 5


S25 Sensor
Dimensions and Weight
Inches (mm)

7.5
(190)
A
6.0
(152)
A

1.5
1.5 6.3 (38)
(38) 1.5
(160)
(38)

3.0 NPT Fitting Detail


2 X 0.5 inch Conduit Openings 4.5
(76) (114)
5.6
(141) 10.1
(257) A

3.6
(91)

3.0 Sanitary Fitting Detail


(76)

4.5 4 X 0.4 inch (10) Holes


(114)

7.9
(200)

Fitting Dimensions Pressure Rating Weight


Code Inches (mm) psig (bar) lb (kg)
S25 Flange Sizes A
1/2" ANSI, 150# Weld Neck Raised Face A1E 13.8" (349) 230 (15.9) 28 (12.7)
1/2" ANSI, 300# Weld Neck Raised Face A3E 14.1" (359) 600 (41.4) 28 (12.7)
3/8" NPT Female F3D 11.5" (292) 600 (41.4) 25 (11.3)
1/2" NPT Male M3E 11.5" (292) 600 (41.4) 25 (11.3)
3/4" Tri-Clamp Style Sanitary Fitting T1F 13.8" (349) 230 (15.9) 27 (12.2)
DIN 15mm, PN40 Weld Neck Raised Face DAE 14.1 (359) 586 (40.4) 30 (13.6)
JIS 15mm, 20K Weld Neck Raised Face JDE 13.8" (349) 484 (33.4) 30 (13.6)
Other (Custom) 999 – – –

Note: Dimensions — inches to the nearest tenth (millimetres to the nearest whole mm), each independently dimensioned from
respective engineering drawings.

Page 6 • SS0M019 Issue/Rev. 0.5 (3/04)


S50 Sensor
Dimensions and Weight
Inches (mm)

7.2 7.6
(193)
(182)

2 X 0.5"
Conduit Openings

B
1.5 14.5 A
(38) (368)
NPT Female Fitting Detail

B
4.4
(112)
A
2.9
(74)
10.1
B
(257)
A
6.0
(153) Sanitary Fitting Detail

Fitting Dimensions Pressure Rating Weight


Code Inches (mm) psig (bar) lb (kg)
S50 Flange Sizes A B
1/2" ANSI, 150# Weld Neck Raised Face A1E 13.0" (330) 2.9" (74) 230 (15.9) 17 (7.7)
1/2" ANSI, 300# Weld Neck Raised Face A3E 13.4" (340) 3.1" (79) 600 (41.4) 17 (7.7)
1/2" ANSI, 600# Weld Neck Raised Face A6E 13.9" (352) 3.4" (85) 1,200 (82.7) 19 (8.6)
1/2" ANSI, 900# Weld Neck Raised Face A9E 14.5" (368) 3.7" (93) 1,800 (124.1) 23 (10.4)
1/2" NPT Female F9E 13.0" (330) 2.9" (74) 1,800 (124.1) 14 (6.4)
1/2" NPT Male M3E 13.0" (330) 2.9" (74) 1,800 (124.1) 14 (6.4)
3/4" ANSI, 150# Weld Neck Raised Face A1F 13.0" (330) 2.9" (74) 230 (15.9) 17 (7.7)
3/4" ANSI, 300# Weld Neck Raised Face A3F 13.5" (343) 3.2" (80) 600 (41.4) 17 (7.7)
3/4" ANSI, 600# Weld Neck Raised Face AGF 14.0" (356) 3.4" (87) 1,200 (82.7) 19 (8.6)
3/4" x 1/2" ANSI, 900# Weld Neck Raised Face A9F 14.7" (375) 3.8" (96) 1,800 (124.1) 25 (11.3)
1" ANSI, 150# Weld Neck Raised Face A1G 13.2" (337) 3.0" (77) 230 (15.9) 19 (8.6)
1" ANSI, 300# Weld Neck Raised Face A3G 13.5" (343) 3.2" (80) 600 (41.4) 21 (9.5)
1" ANSI, 600# Weld Neck Raised Face A6G 14.0" (356) 3.4" (87) 1,200 (82.7) 23 (10.4)
1" ANSI 900# Weld Neck Raised Face A9G 15.0" (381) 3.9" (100) 1,800 (124.1) 31 (14.1)
3/4" Tri-Clamp Style Sanitary Fitting T1F 13.0" (330) 2.9" (74) 230 (15.9) 14 (6.4)
DIN 15mm, PN40 Weld Neck Raised Face DAE 13.4" (340) 3.1" (79) 586 (40.4) 19 (8.6)
DIN 15mm, PN100 Weld Neck Raised Face DBE 13.9" (352) 3.3" (85) 1,465 (100.4) 19 (8.6)
JIS 15mm, 20K Weld Neck Raised Face JDE 13.0" (330) 2.9" (74) 484 (33.4) 19 (8.6)
JIS 15mm, 40K Weld Neck Raised Face JFE 13.4" (340) 3.1" (79) 967 (66.7) 21 (9.4)
JIS 15mm, 63K Weld Neck Raised Face JGE 13.9" (352) 3.3" (85) 1,522 (104.5) 23 (10.4)
Other (Custom) 999 – – – –
Note: Dimensions — inches to the nearest tenth (millimetres to the nearest whole mm), each independently dimensioned from
respective engineering drawings.

Issue/Rev. 0.5 (3/04) SS0M019 • Page 7


S100 Sensor
Dimensions and Weight
Inches (mm)

8.4 8.7
(221)
(214)

2 X 0.5"
Conduit Openings

1.5 21.0 B
A
(38) (533)
NPT Female Fitting Detail

B
4.9
(125)

4.0
(102)
10.1 B
A
(257)
Sanitary Fitting Detail
6.8
(166)

Fitting Dimensions Pressure Rating Weight


Code Inches (mm) psig (bar) lb (kg)
S100 Flange Sizes A B
1" ANSI, 150# Weld Neck Raised Face A1G 15.0" (381) 3.3" (83) 230 (15.9) 32 (14.5)
1" ANSI, 300# Weld Neck Raised Face A3G 15.5" (394) 3.5" (90) 600 (41.4) 34 (15.4)
1" ANSI, 600# Weld Neck Raised Face A6G 16.0" (406) 3.8" (96) 1,200 (82.7) 34 (15.4)
1" NPT Female Connections F6G 15.0" (381) 3.3" (83) 1,200 (82.7) 26 (11.8)
1" Tri-Clamp Style Sanitary Fitting T1G 15.0" (381) 3.3" (83) 230 (15.9) 26 (11.8)
1-1/2" ANSI, 150# Weld Neck Raised Face A1H 15.2" (387) 3.4" (87) 230 (15.9) 34 (15.4)
1-1/2" ANSI, 300# Weld Neck Raised Face A3H 15.7" (400) 3.7" (93) 600 (41.4) 40 (18.1)
1-1/2" ANSI, 600# Weld Neck Raised Face A6H 16.4" (416) 4.0" (101) 1,200 (82.7) 40 (18.1)
2" ANSI, 150# Weld Neck Raised Face A1J 15.4" (390) 3.5" (88) 230 (15.9) 38 (17.2)
2" ANSI, 300# Weld Neck Raised Face A3J 15.7" (400) 3.7" (93) 600 (41.4) 40 (18.1)
2" ANSI, 600# Weld Neck Raised Face A6J 16.6" (422) 4.1" (104) 1,200 (82.7) 42 (19.1)
DIN 25mm, PN40 Weld Neck Raised Face DAG 15.5" (394) 3.5" (90) 586 (40.4) 32 (14.5)
DIN 25mm, PN100 Weld Neck Raised Face DBG 16.0" (406) 3.8" (96) 1,200 (82.7) 38 (17.2)
JIS 25mm, 20K Weld Neck Raised Face JDG 15.0" (380) 3.3" (83) 484 (33.4) 32 (14.5)
JIS 25mm, 40K Weld Neck Raised Face JFG 13.4" (340) 3.1" (79) 967 (66.7) 34 (15.4)
Other (Custom) 999 – – – –

Note: Dimensions — inches to the nearest tenth (millimetres to the nearest whole mm), each independently dimensioned from
respective engineering drawings.

Page 8 • SS0M019 Issue/Rev. 0.5 (3/04)


S200 Sensor
Dimensions and Weight
Inches (mm)

15.6 10.6
(395) (269)

2 X 0.5"
Conduit Openings

1.5 32.4
(38) (822)

B
5.9
A
B (150)

Sanitary Fitting Detail

A
5.9
(150)
10.1
(257)

7.5
(191)

Fitting Dimensions Pressure Rating Weight


Code Inches (mm) psig (bar) lb (kg)
S200 Flange Sizes A B
2" ANSI, 150# Weld Neck Raised Face A1J 22.7" (578) 3.6" (91) 230 (15.9) 100 (45.4)
2" ANSI, 300# Weld Neck Raised Face A3J 23.2" (591) 3.8" (98) 600 (41.4) 106 (48.1)
2" ANSI, 600# Weld Neck Raised Face A6J 24.0" (610) 4.2" (107) 1,200 (82.7) 112 (50.8)
2" Tri-Clamp Style Sanitary Fitting T1J 22.7" (577) 3.6" (91) 230 (15.9) 90 (40.8)
3" ANSI, 150# Weld Neck Raised Face A1L 23.2" (591) 3.8" (98) 230 (15.9) 114 (51.7)
3" ANSI, 300# Weld Neck Raised Face A3L 23.7" (603) 4.1" (104) 600 (41.4) 122 (55.3)
3" ANSI, 600# Weld Neck Raised Face A6L 24.5" (622) 4.5" (114) 1,200 (82.7) 130 (58.9)
DIN 50mm, PN40 Weld Neck Raised Face DAJ 23.2" (590) 3.8" (97) 586 (40.4) 104 (47.2)
DIN 50mm, PN100 Weld Neck Raised Face DBJ 23.6" (600) 4.0" (102) 1,450 (100) 114 (51.7)
DIN 80mm, PN40 Weld Neck Raised Face DAL 23.8" (604) 4.1" (104) 586 (40.4) 120 (54.4)
JIS 50mm, 10K Weld Neck Raised Face JCJ 22.8 (578) 3.6" (91) 199 (13.7) 98 (44.5)
JIS 50mm, 20K Weld Neck Raised Face JDJ 22.8 (578) 3.6" (91) 484 (33.4) 100 (45.4)
JIS 50mm, 30K Weld Neck Raised Face JEJ 23.2" (590) 3.8" (97) 725 (50.0) 104 (47.2)
Other (Custom) 999 – – – –
DBJ Flange – 23.6" (600) 4.0" (102) 1,450 (100) 114 (51.7)

Note: Dimensions — inches to the nearest tenth (millimetres to the nearest whole mm), each independently dimensioned from
respective engineering drawings.

Issue/Rev. 0.5 (3/04) SS0M019 • Page 9


Sensor Modeling Code
Model Fitting Elastomer Trans. Cable Safety Custom
Code Code Type1 Config.2 Code Approvals4 Options Description
KMSCC S25, Hastelloy C Tubes
KMSCH S25, Hastelloy C Tubes,
Hastelloy C Fittings
KMSES S50, Stainless Steel
KMSGS S100, Stainless Steel
KMSJS S200, Stainless Steel
*** Refer to Sensor Fitting Codes (Pgs. 6-9)
0 None (S50, S100, S200)
B Buna-N (Standard, S25)
V Viton (S25)
Z Zalak (S25)
W Integral Mount3
0 Remote Mount
0 None (For Integral Mount)
A 10' Teflon
B 20' Teflon
C 50' Teflon
D 100' Teflon
E 10' PVC
F 20' PVC
G 50" PVC
H 100' PVC
0 General Purpose (No Approvals)
2 ATEX
3 UL/CUL
0 Standard
H S200 Enhanced Drive Power Option
(UL/CUL) Division 2)
HT S200 Enhanced Drive & Hi-Temp Option
(General Purpose Only)
X Custom

Example Sensor Model Number


KMSGS A1G 0 W 0 0 0 = KMSGSA1G0W000

S100 1" Sensor, Sainless Steel, 150# Raised Face Flange, Transmitter Integral Mount, General Purpose
(No Approvals), Standard Oprions.

1 Elastomer selection required for S25 sensor only.


2 Cable length selection, required for remotely mounted transmitter only.
3 When integrally mounted, the transmitter approval must match the sensor approval.
4 Approvals for the S-25 pending.

Page 10 • SS0M019 Issue/Rev. 0.5 (3/04)


Transmitter Modeling Code
Model Mounting Safety Custom
Code Config. Approvals Options Description
SMT050 S25-S200 Pulse Transmitter
SMT100 S25-S200 Programmable Transmitter

X No Housing
A Integral Mount NEMA 4X
B Remote Mount NEMA 4X
C Remote Mount Explosion Proof
0 General Purpose (No Approvals)
1 ATEX
2 UL/CUL
0 Standard
H High Drive (S200)
B Barrier
X Custom

Example Transmitter Model Number

SMT100 A 0 0 = KMSGSA1G0W000

S25-S200 Programmable Transmitter, Integral Mount in NEMA 4X Enclosure,


General Purpose, Standard Options.

Issue/Rev. 0.5 (3/04) SS0M019 • Page 11


Dimensions (NEMA 4X/IP66 Enclosure)
Inches (mm)

8.7"
(220)
8.0" (204) (Mounting)

3.2" 4.7" Net Weight: 4 lb (1.8 kg).


(82) (120)

3.1" Materials of Construction:


(80) Aluminum with gray acrylic
enamel. Neoprene gasket.
1.4"
(37) 0.8"
(21)
1.5" 0.85" (22) #10
(39) 2.8"
(71) Dia. Holes Mounting
For 1/2" Fittings Screws

Note: Dimensions — Inches to the nearest tenth (millimetres to the nearest whole mm), each independently dimensioned from respective
engineering drawings.

Page 12 • SS0M019 Issue/Rev. 0.5 (3/04)


Dimensions (NEMA 7/4 Explosion-Proof Dual Unit Enclosure)
Inches (mm)

15"
(381)
14"
(356)

4.6"
(116) 3.3"
(83)

UL & CSA Approved enclosure


Use .25" Mounting Bolts for use in hazardous locations
(4 Places) Class I, Groups C & D Class II,
Groups E, F & G Class III
11.5"
.50" - 14 NPT (292) 5.8"
(2) Paces
(146)

4.5"
(114)
2.5"
(64)

Note: Dimensions — Inches to the nearest tenth (millimetres to the nearest whole mm), each independently dimensioned from respective
engineering drawings.

Issue/Rev. 0.5 (3/04) SS0M019 • Page 13


Related Publications
Galvanic Barrier Module Specifications ...................................................................................................... SS0M014
FloMate Specifications ................................................................................................................................ SS09033
CoriolisMass Flow and Density Meter with SMT050 Electronics
Installation/Operation/Calibration Manual ............................................................................................... MN0M016

Revisions included in SS0M019 Issue/Rev. 0.5 (3/04):


Approvals changed from FM/CSA and CENELEC to UL/CUL and ATEX. .

The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the
specifications are currently in effect. Otherwise, the manufacturer assumes no responsibility for the use of specifications which may have been changed and are
no longer in effect.

Headquarters:
1803 Gears Road, Houston, TX 77067 USA, Phone: 281/260-2190, Fax: 281/260-2191
Gas Measurement Products: Liquid Measurement Products:
Houston, TX USA Phone 281/260-2190 Erie, PA USA Phone 814/898-5000 Beijing, China Phone (86) 10/6500-2251
Thetford, England Phone (44) 1842-82-2900 Los Angeles, CA USA Phone 661/702-8660 Singapore Phone (65) 6861-3011
Kongsberg, Norway Phone (47) 32/286-700 Slough, England Phone (44) 1753-57-1515 Chennai, India Phone (91) 44/450-4400
Buenos Aires, Argentina Phone 54 (11) 4312-4736 Ellerbek, Germany Phone (49) 4101-3040
Integrated Measurement Systems: Barcelona, Spain Phone (34) 93/201-0989
Corpus Christi, TX USA Phone 361/289-3400 Moscow, Russia Phone (7) 495/564-8705
Kongsberg, Norway Phone (47) 32/286-700 Melbourne, Australia Phone (61) 3/9807-2818
San Juan, Puerto Rico Phone 787/274-3760
United Arab Emirates, Dubai Phone 971 +4/331-3646

Visit our website at www.fmctechnologies.com