The gke PCDs are type 2 indicator systems The PMS Compact-PCD®, colour: orange, simu-
according to EN ISO 11140-1, consisting of a lates the hollow load test according to EN 867-5.
„specific test load“ (Process Challenge Device = This type test is included in the standard for large
PCD) and „indicator“ (indicator strip). They have sterilizers (EN 285) as well as in the standard for
been validated from a laboratory accredited ac- small class B sterilizers (EN 13060).
cording to the standard EN ISO 17025 (test re-
ports on request). The standard EN 867-5 describes a test cycle
with 9 injections of steam to 950 mbar. In this test
gke offers a variety of PCDs with different lev- cycle the hollow load test will be successfully
els of difficulty. Using a test with the highest level penetrated, if a vacuum of 450 mbar will be
which will still pass in a validated process the reached in between each steam pulse.
sterilizer is always tested at its limits. Failures that
are not detected with standard methods will now
be visible and the maximum possible safety to 3. HDH-PMS Compact-PCD®, colour: red
test sterilization processes can be guaranteed. If
a process has been validated with a defined load The HDH-PMS Compact-PCD®, colour: red,
and a suitable test system, this method secures (High Demand Hollow) is used for routine moni-
the sterilization at the most difficult locations. This toring of complex instruments or long tubes,
is not possible by only recording the physical data whereas the hollow load test according to
with documentation, e.g. with a printer. EN 867-5 is not anymore sufficient.
Insufficient temperature, no air All gke chemical indicators are protected from
removal and no steam penetration bleeding by a polymer binder and surface
Fig. 4: Colour change of a standard indicator strip. If using an
coating can be disposed with normal garbage.
indicator strip for prion programs the initial colour is pink
Selection between four different versions of
PMS Compact-PCD® depending on the air
This result ensures air removal and steam pene- removal program used and the load to be ster-
tration into the whole load under the condition ilized.
that the PCD is representing the load configura-
tion. The screw cap consists of a highly thermal
resistant material and stainless steel sandwich
construction that protects hands from high
Benefits temperatures. The chemical indicator may be
easily removed and evaluated on completion
Indicator strips for standard and prion steriliza- of each cycle.
tion cycles are available.
The Compact-PCD® can be used for a consid-
Test of a sterilizer, if the specification require- erable number of cycles. All important parts
ments are according to the standard, lower or are made of stainless steel or thermal re-
higher. sistant polymers.
Each start-up kit contains of a Compact-PCD and 100 integrating indicator strips for standard steam sterili-
zation processes. All test devices are also available separately. The indicator strips are available as refill
packs for standard and prion steam sterilization processes and also contain a sealing kit for the screw cap.
1. Start-up kit and process challenge devices / Process monitoring systems (PMS)