Sunteți pe pagina 1din 3

Characteristics Notes MIURA 200

General

Instrument Type Type Floor standing & Bench top


P.C. Separate Yes
Open/Close system open/closed with smart card
ise module On-board yes/optional
ise parameters Na+, K+, Cl-, Li+
Throughput/hour Without ISE 160
With ISE 250
Number of sampling Arms 1
Reagents on line up to 49
Methods off line more than 200

Sample Treatment
Sample support in the analyser Positions 50 + 50 dilution cups

Paediatric sample support in the Positions 6


analyser
Calibrators Positions 18
Controls Positions 3 (Levy-Jennings Graph / 3-levels)

Bar-code reader On-board yes/optional


Sample primary tube types Without adapter Ø12÷16mm × 75÷100mm
Cups with adapter 3ml sample cups
Dead Volume 12/13mm primary tube 100µl

16mm primary tube 200µl


3ml sample cup 50µl
Sample Volume (µl) Min. max. 1 / 100 µl
Dilution of patient sample Pre-dilution automatic on request
Post-dilution / rerun automatic on request
Automatic standard curve dilution yes, up to 8-points

Diluter Micro-metering pump stainless steel plunger


Characteristics Notes MIURA 200
Liquid level sensor Capacitive yes
Shock sensor yes

Reagent Treatment
Reagent support in the analyser Positions 48 + 1 diluent
Bar-code reader On-board yes/optional
Reagent bottles Without adapter 50ml
Without adapter 20ml
Dead Volume 50ml bottle 1,5ml
20ml 0,5ml
Reagent Volume (µl) Min. max. 180 / 400 µl
Diluter Micro-metering pump
Liquid level sensor Capacitive yes
Shock sensor yes
Reagents to be used per test 3
Reagent cooling 14°C below ambient temp
Multiple reagent position (same lot) yes

Analytical methods end-point, bi-chromatic end-point, fixed-time, kinetic, differential sample blank, differential 2-reagents

Cuvette Treatment
Reaction/Reading Volume (µl) Typical 200 µl
Cuvettes on board Cuvette plate 80 (re-washable)
Cuvettes type optical plastic (Bionex®)
Incubation heater 37°C ± 0.1°C controlled
Cuvette Washing Station yes, over 8 steps

Measurement System
Photometer Type: Light Source Lamp Halogen lamp (2,000hours)

Type of Measurement mono or bichromatic


Wavelenghts Interferential filters 340÷700nm, 8 on board + 1 free
Photometric linearity 0÷3 O.D.

Installation and Use of Analyser


Water Consumption Over 1 hour (typ) 3 liters
Liquid tanks systemic sol. tank , cleaner sol.tank , waste tank
Tank level sensor yes

Control System
Operating System Windows XP® Pro
Characteristics Notes MIURA 200
L.I.S. connection yes, optional
Patient archive yes, up to 15,000 patients
Q.C. archive yes, up to 5,000 tests