Shanghai Changji SYD-3536 Flash Point Tester (Closed Cup)
4. Stirring device:
(1) Stirring motor: 45TYZ constant speed motor;
(2) Driving mode: by a soft shaft;
(3) Stirring blade: 8 mm×40 (mm);
5. Standard oil cup:
(1) Inner diameter: 50.8 mm;
(2) Depth: 56 mm;
(3) Scale depth: 34.2 mm;
(4) Oil sample cubage: about 70 ml;
6. Flame application device:
(1) Flame source: coal gas or other combustible gas for civil use;
(2) Aperture of the device: 0.8 mm;
7. Thermometer: Inner calibrated mercury thermometer, as per GB/T514;
3
Shanghai Changji SYD-3536 Flash Point Tester (Closed Cup)
Fig.1
1. Gas inlet: The instrument connects to outer gas source through this inlet. You can adjust gas
rate at gas inlet.
4
Shanghai Changji SYD-3536 Flash Point Tester (Closed Cup)
Fig.2
5
Shanghai Changji SYD-3536 Flash Point Tester (Closed Cup)
R L
R2 E3
V
R1 ~
220V W2 W4 FS
E4
D1
ST
C1 C2
k1
M
Rs
D2
图2:电气原理图
Fig.3
6
Shanghai Changji SYD-3536 Flash Point Tester (Closed Cup)
without air draft, so that you can observe flash conveniently. Please get ready for all test apparatus
and materials.
4. Check the working state of the instrument to make sure it is in accord with working
environment and condition stipulated in the Operation Manual.
5. Check the outer shell of the instrument to make sure it is fine grounded.
6. Pour the oil sample into the oil cup and start the stirring motor. The stirrer is driven by soft
shaft to stir sample.
7. Please adjust the heating rate as per the stipulation of GB/T261 (Rotate the Heat Regulation
knob to control the heating power. The heating voltage is shown on the voltmeter).
8. Please connect the outer gas source to the instrument, ignite the flame application device,
and adjust the flame size before the oil temperature reaches to the expected flash point of oil sample
as per the relative standards.
9. Rotate the spring knob to move shutter. Then the hole on the oil cup cover will open and the
flame application device will move downwards automatically to ignite the oil sample once.
10. Continue making test as per the stipulation of GB/T261 until you get test results.
V. Attention
1. The power supply wire should have fine grounding end for safety.
2. Turn off the outer gas source after test.
3. Please turn off the Power switch and Stir switch after tests and keep it in a dry and ventilated
place.
4. Place the spare oil cup in the oil cup seat when it is not used.
5. The instrument is suitable to make determination for oil sample, the expected closed cup
flash point of which is lower than 280 ℃.
6. If there is any trouble appears, please ask a professional to check and repair it.
Warning: If there is any trouble appears, please cut off power supply immediately. Then ask a
professional to check and repair it. You can use the instrument again only after the problem is solved
to avoid any accidents.