HPV-HR .
Chip-based Real Time Duplex PCR Test for Human Papillomavirus
High Risk Types 16, 31 and 18, 45 3. PRINCIPLE OF THE TEST
Truenat™ HPV-HR works on the principle of Real Time Polymerase Chain Reaction.
1. INTENDED USE The DNA from the patient sample is first extracted using Trueprep™ MAG Sample
Truenat™ HPV-HR (REF 601220005/601220020) is a chip-based Real Time Prep Device and Trueprep™ MAG Blood Sample Prep kit or using Trueprep™
Polymerase Chain Reaction (PCR) test for the semiquantitative detection of high AUTO Universal Cartridge Based Sample Prep Device and Trueprep™ AUTO
risk Human Papillomavirus (HPV) types 16, 18, 31 and 45 in female cervical Universal Cartridge Based Sample Prep Kit. The Truenat™ HPV-HR chip is placed on
specimens collected by a clinician. It aids in the differential diagnosis of symptomatic the chip tray of the Truelab™ Real Time micro PCR Analyzer. Six (6) µL of the purified
or asymptomatic infection with high risk HPV types 16, 31 and 18, 45. Truenat™ DNA is then dispensed using the provided micropipette and tip into the microtube
HPV-HR runs on the Truelab™ Real Time Quantitative micro PCR Analyzers. containing freeze dried PCR reagents and allowed to stand for 30-60 seconds to get a
clear solution. No mixing by tapping, shaking or by reverse pipetting should be
2. INTRODUCTION done. Six (6) µL of this clear solution is then pipetted out using the same pipette and
Human papillomavirus (HPV) is a non-enveloped, double stranded DNA virus from tip and dispensed into the reaction well of the Truenat™ HPV-HR chip and the test is
the papillomavirus family. It infects squamous epithelia including the skin and started. A positive amplification causes the labeled fluorescent probes in the
mucosae of the upper respiratory and anogenital tracts. There are approximately 170 Truenat™ HPV-HR chip-based Real Time PCR test to release the fluorophores in an
types of HPV, of which about 40 are sexually transmitted and infect the anus and exponential manner which is then captured by the built-in opto-electronic sensor and
genitals. Although most infections are asymptomatic and self-limiting, genital displayed as amplification curve on the analyzer screen, on a real time basis during
infection by HPV is associated with genital warts and anogenital cancers in both men the test run. The Cycle threshold (Ct) is defined as the number of amplification cycles
and women. HPV viruses are classified as either ‘high-risk’ or ‘low-risk’ types required for the fluorescent signal to cross the threshold (i.e. exceed the background
depending on their association with the development of cancer. Low-risk HPV types, signal). Ct levels are inversely proportional to the amount of target nucleic acid in the
such as HPV 6 and 11, can cause common genital warts or benign hyperproliferative sample. (i.e. the lower the Ct level the greater is the amount of target nucleic acid in
lesions with very limited tendency to malignant progression. Persistent infection by the sample). In the case of negative samples, amplification does not occur and a
high-risk HPV types such as HPVs 16, 18, 31 and 45 is associated with the horizontal amplification curve is displayed on the screen during the test run. At the
occurrence of pre-malignant and malignant lesions that are detectable in more than end of the test run, a HPV-HR “DETECTED” or “NOT DETECTED” result is displayed
90% of cervical cancers. Cervical cancer is one of the most common types of cancer and in positive cases, semi quantitative result is also displayed on the screen. Based
in women worldwide. High-risk HPV types are also associated with many penile, on the detection of the Internal Positive Control (IPC), the validity of the test run is also
vaginal, vulvar, anal, head and neck carcinomas and contribute to over 40% of oral displayed. The IPC is a full process control that undergoes all the processes the
cancers. Although persistent infection with oncogenic HPV types is the most specimen undergoes - from extraction to amplification thereby validating the test run
common risk factor in its etiology of cervical cancer and its precursor lesions, only a from sample to result. Absence of or shift of IPC Ct beyond a pre-set range in case of
very small proportion of infections advance to these disease states. Upwards of 70% negative samples invalidates the test run. While IPC will co-amplify in most positive
of healthy young adults, will clear HPV infections within 12 to 24 months. In the subset cases also, in some specimen having a high target load, the IPC may not amplify,
of adults with persistent infection, progression to clinical infection may take years, however the test run is still considered valid. The results can be printed via Bluetooth
providing opportunities for detection and treatment of pre-cancerous lesions. using the Truelab™ micro PCR printer or transferred to the lab computer/or any
Cervical cancer has in the past been shown to be highly preventable when cytological remote computer via Wifi or 3G/GPRS network. Upto 20,000 results in Truelab™
and HPV screening programs are deployed to assist the detection and treatment of Uno Dx/Truelab™ Duo/Truelab™ Quattrocan be stored on the analyzer for future
pre-cancerous lesions. Screening can also detect cancer at an early stage and recall and reference.
treatment has a high potential for cure.
Cervical screening using a Papanicolaou (Pap) test or liquid- 4. TARGET SELECTION
based cytology to detect abnormal cells that may develop into The target sequences for this assay are regions within the E6 and E7 genes of high
cancer has greatly reduced the incidence and fatalities of risk HPV types 16,18,31 and 45.
cervical cancer. However, these tests require interpretation by
highly trained cytopathologists and have a high rate of false 5. CONTENTS OF THE Truenat™ HPV-HR KIT
negatives. HPV is exceptionally difficult to culture in vitro and a A. Individually sealed pouches, each containing
demonstrable antibody response is not found in all patients 1. Truenat™ HPV-HR micro PCR chip.
infected with HPV. Nucleic acid (DNA) testing is a highly sensitive and specific 2. Microtube with freeze dried PCR reagents.
method for determining the presence of infection with high-risk HPV types in cervical 3. DNase & RNase free pipette tip.
specimen. However, nucleic acid based molecular tests have so far been restricted to 4. Desiccant pouch.
centralized reference laboratories as they require skilled manpower and elaborate B. Package Insert
infrastructure. Also the turnaround time for results could take a few days to weeks REF 601220005 601220020
leading to high losses to follow-up.
5T 20T
The Truelab™ Real Time micro PCR System enables decentralization and near
patient diagnosis of high-risk HPV types 16, 18, 31 and 45 by making real time PCR
technology rapid, simple, robust and user friendly and offering “sample to result” 6. CONTENTS OF THE Trueprep™ MAG Swab Specimen Pre-treatment Pack
capability even at resource limited settings. This is achieved through a combination of (only for TrueprepTM MAG users)
lightweight, portable, mains / battery operated Truelab™ Real Time Quantitative A. Sample Pre-treatment tubes (contains lysis cum transport medium).
micro PCR Analyzer and TrueprepTM MAG/AUTO Sample Prep Device and room REF 60202SW20
temperature stable Truenat™ micro PCR chips and TrueprepTM MAG/AUTO
20T
Sample Prep kits so that even the peripheral laboratories with minimal infrastructure
and minimally trained technician can easily perform these tests routinely in their
7 CONTENTS OF THE TrueprepTM AUTO Universal Sample Pre-treatment Pack
facilities and report PCR results in less than an hour. Moreover, with these devices
(only for TrueprepTM AUTO users)
PCR testing can also be initiated in the field level, on site.
Truenat™ HPV-HR is a disposable, room temperature stable, chip-based Real Time A.Lysis buffer.
PCR test with dried MgCl2 in reaction well and freeze dried PCR reagents in B.Disposable transfer pipette(graduated).
microtube for performing Real Time PCR test for detection and diagnosis of HPV
types 16, 18, 31 and 45 and runs on the Truelab™ Real Time Quantitative micro PCR
REF 60205AB05 60205AB20
Analyzer. It requires only six (6) µL of purified DNA to be added to the reaction well for 5T 20T
the analysis. The intelligent chip also carries test and batch related information
8. CONTENTS OF THE Trueprep™ AUTO Transport Medium for Swab Specimen
including standard values for quantitation. The Truenat™ HPV-HR chip also stores
Pack (only for TrueprepTM AUTO users)
information of used test to prevent any accidental re-use of the test.
NOTE : Truelab™ Uno Dx/Truelab™ Duo/Truelab™ Quattro/ Trueprep™ AUTO/ A. Transport Medium for Swab specimen tubes (contains transport medium).
Truepet™ / Truenat™ are all registered trademarks of Molbio Diagnostics (P) REF 60206TS05 60206TS20
Limited. 5T 20T
The Truelab™ Real Time micro PCR Analyzer is protected by the following
patents and patents pending: IN 2313/CHE/2007, WO 2009/047804 and
while performing sample extraction.
9. STORAGE AND STABILITY
Truenat™ HPV-HR chip is stable for two years from the date of manufacture if stored 14. PROCEDURAL PRECAUTIONS
between 2-30oC. It is also stable for upto six (6) months at temperatures up to 40o C and 1. Check all packages before using the kit. Damage to the packaging does not
one (1) month at temperatures up to 45o C. Avoid exposure to light or elevated prevent the contents of the kit from being used. However, if the outer packaging
temperatures (above recommended levels). Do not freeze. is damaged the user must confirm that individual components of the kit are intact
TrueprepTM MAG Swab specimen Pre-treatment Pack, Trueprep™ AUTO Transport before using them.
Medium for Swab Specimen Pack and TrueprepTM AUTO Universal Sample Pre- 2. Do not perform the test in the presence of reactive vapours (e.g. from sodium
Treatment Pack is stable for two years from the date of manufacture if stored between 2- hypochlorite, acids, alkalis or aldehydes) or dust.
40°C. It is also stable for one (1) month at temperatures upto 45°C. 3. While retrieving the Truenat™ HPV-HR chip, microtube and the DNase and
RNase free pipette tip from the pouch, ensure that neither bare hands nor
10. MATERIALS REQUIRED BUT NOT PROVIDED WITH THE KIT gloves that have been used for previous tests run are used.
Truelab™ Real Time micro PCR Workstation (REF 60301001/62301001/
633010001/64301001) consisting of 15. PROCEDURAL LIMITATIONS
1. Trueprep™ MAG / AUTO Sample Prep Device (REF 603040001/603041001). 1. Optimal performance of this test requires appropriate specimen collection,
2. Truelab™ Uno/Truelab™ Uno Dx/Truelab™ Duo/Truelab™ Quattro Real Time handling, storage and transport to the test site.
micro PCR Analyzer (REF 603020001/603021001/603022001/603023001). 2. Though very rare, mutations within the highly conserved regions of the target
3. Truelab™ micro PCR Printer (REF 603050001). genome where the Truenat™ assay primers and/or probe bind may result in the
4. Truepet™ SP fixed volume precision micropipette - 6 µl (REF 604060006). under-quantitation of or a failure to detect the presence of the concerned
5. Truelab™ Microtube Stand (REF 603070001). pathogen.
Also required additionally are: TrueprepTM MAG Blood Sample Prep Kit (REF 3. The instruments and assay procedures are designed to minimize the risk of
602010005/REF 602010050) / TrueprepTM AUTO Universal Cartridge Based Sample contamination by PCR amplification products. However, it is essential to follow
Prep Kit (REF60203AR05/REF 60203AR25),Truenat™ Universal Control Kit (REF good laboratory practices and ensure careful adherence to the procedures
601100008),DNase and RNase-free pipette tips with filter barrier which may also be specified in this package insert for avoiding nucleic acid contamination from
procured from Molbio, Powder free disposable gloves, nylon flock swabs and waste previous amplifications, positive controls or specimens.
disposal container with lid. 4. A specimen for which the Truenat™ assay reports “Not Detected” cannot be
concluded to be negative for the concerned pathogen. As with any diagnostic
11. SPECIMEN PREPARATION FOR EXTRACTION WITH TrueprepTM MAG test, results from the Truenat™ assay should be interpreted in the context of
Swab specimen must be collected as per standard procedures using a standard nylon other clinical and laboratory findings.
flocked swab. Insert the swab with specimen into the Sample pre-treatment tube
provided and mix well by repeatedly twirling the swab in the buffer solution. After mixing, 16. CLEANING AND DECONTAMINATION
squeeze out the excess liquid from the swab by pressing it a few times against the inside 1. Spills of potentially infectious material should be cleaned up immediately with
wall of the tube. Dispose off the swab as per the section on “Disposal and Destruction” absorbent paper tissue and the contaminated area should be decontaminated
(Section 20).Tightly close the cap of the Sample pre-treatment tube. with disinfectants such as 0.5% freshly prepared sodium hypochlorite [10 times
dilution of 5% sodium hypochlorite (household bleach)] before continuing work.
Sample Storage and Transportation:
Sample Pre-treatment decontaminates the specimen and makes it ready for storage/ 2. Sodium hypochlorite should not be used on an acid-containing spill unless the
transportation/ extraction. The specimen in this form is stable for up to 3 days at 40ºC and spill-area is wiped dry first. Materials used to clean spills, including gloves,
1 week at 30ºC. should be disposed off as potentially bio-hazardous waste e.g. in a biohazard
Nucleic acid extraction: Use entire content from the Sample pre-treatment tube waste container.
containing specimen for further procedure with the Trueprep™ MAG Sample Prep
17. TEST PROCEDURE
Device and Trueprep™ MAG Blood Sample Prep Kit. Transfer the entire content into the
extraction tube. Dispose off the Sample pre-treatment tube as per the section on (Please also refer the Truelab™ Real Time Quantitative micro PCR Analyzer
“Disposal and Destruction” (Section 20). Start the extraction from step 4 of “the sample user manual).
extraction and purification procedure” in section 13 of the package insert of Trueprep TM 1. Switch on the Truelab™ Analyzer .
TM
MAG Blood Sample Prep kit. (Refer to the User Manual of Trueprep MAG Sample Prep 2. Select user and enter password.
device and the package insert of Trueprep™ MAG Blood Sample Prep Kit ) for details. 4. For TruelabTM Uno Dx, select the test profile for “HPV” to be run from the Profiles
Screen on the Analyzer screen. For TruelabTM Duo/Quattro, select the Bay
TM
12. SPECIMEN PREPARATION FOR EXTRACTION WITH Trueprep AUTO (Idle1/2) for Duo and (Idle1/2/3/4) for Quattro from the Status Screen to view
Swab specimen must be collected as per standard procedures using a standard nylon the Profiles Screen. Select the test profile for “HPV” to be run from the Profiles
flocked swab. Insert the swab with specimen into the Transport Medium for Swab Screen on the Analyzer screen.
Specimen Tube provided and mix well by repeatedly twirling the swab in the buffer 5. Enter the patient details as prompted in the Truelab™ Analyzer screen.
solution. After mixing, squeeze out the excess liquid from the swab by pressing it a few 6. Press Start Reaction.
times against the inside wall of the tube. Dispose off the swab as per the section on 7. For TruelabTM Uno Dx, Press the eject button to open the chip tray. For
“Disposal and Destruction” (Section 20). Tightly close the cap of the Transport Medium TruelabTM Duo/Quattro, the chip tray opens automatically on tapping the “Start
for Swab Specimen Tube. Reaction” button.
Sample Storage and Transportation: 8. Open a pouch of Truenat™ HPV-HR and retrieve the chip-based Real Time
Transport Medium for Swab Specimen decontaminates the specimen and makes it ready PCR test and the microtube.
for storage/ transportation/extraction. The specimen in this form is stable for up to 3 days 9. Label the chip and the tube with the patient ID using a marker pen at the space
at 40ºC and 1 week at 30ºC. provided on the back side of the chip and the space on the microtube label.
Nucleic acid extraction: Transfer the entire content from the Transport Medium for 10. Place the Truenat™ HPV-HR chip on the chip tray without touching the white
Swab Specimen Tube into the Lysis Buffer Tube for further procedure with the reaction well. The reaction well should be facing up and away from the Analyzer.
Trueprep™ AUTO Universal Cartridge Based Sample Prep Device and Trueprep™ Gently press the chip to ensure that it has seated in the chip tray properly.
AUTO Universal Cartridge Based Sample Prep Kit (Refer to the User Manual of 11. Place the microtube containing freeze dried PCR reagents in the microtube
TrueprepTM AUTO Universal Cartridge Based Sample Prep device and the package stand provided along with the TruelabTM Real Time micro PCR workstation after
TM
insert of Trueprep AUTO Universal Cartridge Based Sample Prep kit for details). ensure that white pellet of dried PCR reagents remain at the bottom of the
Dispose off the Transport Medium for Swab Specimen Tube as per the section on microtube. Remove the microtube cap and dispose it off as per the section on
“Disposal and Destruction” (Section 20). “Disposal and Destruction” (Section 20). Using the filter barrier tip provided in
the pouch, pipette out six (6) µL of the purified DNA from the Elute Collected
13. SAFETY PRECAUTIONS Tube into the microtube. Allow it to stand for 30-60 seconds to get a clear
1. For in vitro diagnostic use only. solution. Do not mix it by tapping, shaking or by reverse pipetting. Using the
o
2. Bring all reagents and specimen to room temperature (20 - 30 C) before use. same filter barrier tip, pipette out six (6) µL of this clear solution and dispense
3. Do not use kit beyond expiry date. into the centre of the white reaction well of the Truenat™ HPV-HR chip. Take
4. Carefully read the User Manuals and package inserts of all the components of the care not to scratch the internal well surface and not to spill elute on the outside of
Truelab™ Real Time Quantitative micro PCR System before use. the well. Dispose off the microtip as per the section on “Disposal and
5. All materials of human origin should be handled as though potentially infectious. Destruction” (Section 20).
6. Do not pipette any material by mouth. 12. For TruelabTM Uno Dx, slide the chip tray containing the Truenat™ HPV-HR chip-
7. Do not eat, drink, smoke, apply cosmetics or handle contact lenses in the area where based Real Time PCR test loaded with the sample into the TruelabTM Analyzer.
testing is done.
Press Done on the “Please Load Sample” Alert message.
8. Use protective clothing and wear disposable gloves when handling samples and
For TruelabTM Duo/Quattro, select “YES” at the Please load Sample prompt. Chip (ATCC®65446™) DNA was made and nucleic acids were extracted on TrueprepTM
tray will close automatically and the reaction will start. AUTO sample Prep device followed by TruelabTM Uno Dx Real Time micro PCR
13. Read the result from the screen. Analyzer.
14. After the reaction is completed, for TruelabTM Uno Dx, push the Eject button to eject The assay is found to be linear over 5 orders of magnitude (from 1.00E + 06 IU/ml to
the chip tray. For TruelabTM Duo/Quattro, tap the “Open/Close Tray” button to eject 5.00E + 02 IU.ml) for the NIBSC 1st WHO International Standard for HPV 16 DNA.
the chip tray. The assay is found to be linear over 5 orders of magnitude (from 1.00E + 06 IU/ml to
5.00E + 02 IU.ml) for the NIBSC 1st WHO International Standard for HPV 18 DNA.
15. Take out the Truenat™ HPV-HR chip-based Real Time PCR test at end of the test
The assay is found to be linear over 5 orders of magnitude (from 5.79E + 06 IU/ml to
and dispose it off as per the section on “Disposal and Destruction” (Section 20).
5.79E + 02 IU.ml) for the NIBSC 1st WHO International Standard for HPV Type 31,
16. Turn on Truelab™ micro PCR printer and select print on the screen for printing out
pHPV31 (ATCC®65446™) DNA.
hard copy of the results. Test results are automatically stored and can be retrieved
any time later. (Refer to the Truelab™ Analyzer manual). HPV 16 linearity curve HPV 18 linearity curve
17. Switch off the Truelab™ Analyzer.

18. RESULTS & INTERPRETATIONS


Three amplification curves are displayed on the Truelab™ Real Time Quantitative micro
PCR Analyzer screen to indicate the progress of the test. Either or both the target and the
internal positive control (IPC)* curves will take a steep, exponential path when the
fluorescence crosses the threshold value in case of positive samples. The Ct will depend
HPV 31 linearity curve
on the number of viral genomes in the sample. The target curve will remain horizontal
throughout the test duration and the IPC curve will take an exponential path in case of
negative samples. In case the IPC curve remains horizontal in a negative sample, the test
is considered as Invalid. At the end of the test run, the results screen will display
“DETECTED” for Positive result or “NOT DETECTED” for Negative result. The result
screen would also display the viral load as “HIGH”, MEDIUM” or “LOW” for positive
specimen. The result screen also displays the validity of the test run as “VALID” or
“INVALID”. Invalid samples have to be repeated with fresh specimen from the sample Limit of detection (Analytical Sensitivity):
preparation stage. *While IPC will co-amplify in most positive cases also, in some The LOD was determined by testing dilutions of NIBSC 1st WHO International
specimen having a high target load, the IPC may not amplify, however the test run is still Standard for HPV 16 DNA, NIBSC 1st WHO International Standard for HPV 16 DNA
considered valid. and HPV Type 31, pHPV31 (ATCC®65446™) DNA . Probit analysis of the data was
used to determine the concentration of the respective RNA with 95% probability of
19. QUALITY CONTROL PROCEDURES detection.
To ensure that the Truelab™ Real Time Quantitative micro PCR Analyzer is working The LOD was found to be 242.86 IU/ml of NIBSC 1st WHO International Standard for
accurately, run positive and negative controls from time to time. The Universal Control HPV 16 DNA.
kit containing Positive Control and Negative Control must be ordered separately. It is The LOD was found to be 277.44IU/ml of NIBSC 1st WHO International Standard for
advisable to run controls under the following circumstances. ● Whenever a new HPV 18 DNA.
shipment of test kits is received. ● When opening a new test kit lot. ● If the temperature of The LOD was found to be 400.94 copies/ml of pHPV31 (ATCC®65446™) DNA for
the storage area falls outside of 2-30oC. ● By each new user prior to performing testing on HPV 31 DNA.
clinical specimen.
Probit Analysis for HPV 16
20. DISPOSAL AND DESTRUCTION HPV 16 Total No of Positivity HPV-16 Probit Plot
IU/mL runs Positives Percentage
1. Submerge the used content such as Truenat™ HPV-HR chip, microtube, microtube
5.00E+01 10 0 0
cap, pipette tips, nylon flocked swab, Sample pre-treatment tube, Transport Medium
for Swab Specimen Tube etc. in freshly prepared 0.5% sodium hypochlorite solution 1.00E+02 10 5 50
Probability

for 30 minutes before disposal as per the standard medical waste disposal 2.00E+02 10 8 80
guidelines. 5.00E+02 10 10 100
2. Disinfect the solutions and/or solid waste containing biological samples before 1.00E+03 10 10 100
discarding them according to local regulations.
1.00E+04 3 3 100
3. Samples and reagents of human and animal origin, as well as contaminated
1.00E+05 3 3 100
materials, disposables, neutralized acids and other waste materials must be
discarded according to local regulations after decontamination by immersion in a 1.00E+06 3 3 100 LOD = 242.86 CI [185.72 - 431.24]
freshly prepared 0.5% of sodium hypochlorite for 30 minutes (1 volume of 5% sodium
hypochlorite for 10 volumes of contaminated fluid or water). Probit Analysis for HPV 18
4. Do not autoclave materials or solutions containing sodium hypochlorite. HPV 18 Total No of Positivity
HPV-18 Probit Plot
5. Chemicals should be handled in accordance with Good Laboratory Practice and IU/mL runs Positives Percentage

disposed off according to the local regulations. 5.00E+01 10 0 0

1.00E+02 10 5 50
Probability

21. SPECIFIC PERFORMANCE CHARACTERISTICS 2.00E+02 10 7 70


Analytical Exclusitivity (Primer specificity): The following viruses and
5.00E+02 10 10 100
microorganisms were evaluated in silico from the NCBI database using the NCBI
1.00E+03 10 10 100
nucleotide blast and primer blast tools to determine potential cross-reactivity in the
Truenat™ HPV-HR assay. Results obtained showed no cross reactivity of the assay with 1.00E+04 3 3 100

the listed organisms. 1.00E+05 3 3 100

1.00E+06 3 3 100
LOD = 277.44 CI [205.95 - 560.33]

Probit Analysis for HPV 31


HPV 31 Total No of Positivity HPV-31 Probit Plot
Copies/mL runs Positives Percentage

7.20E+01 10 0 0

1.44E+02 10 4 40

2.89E+02 10 7 70

5.79E+02 10 10 100

5.79E+03 10 10 100

5.79E+04 3 3 100
Copies/ml
5.79E+05 3 3 100
Linearity: LOD = 400.94 CI [304.97 - 741.65]
5.79E+06 3 3 100
Serial dilutions of the NIBSC 1st WHO International Standard for HPV 16 DNA, NIBSC 1st
WHO International Standard for HPV 16 DNA and HPV Type 31, pHPV31
Reproducbility:
The reproducibility of TruenatTM HPV-HR assay was determined between three different
users and between three different devices. Three different titres of samples (High,
Medium and Low) were used for this study. The variation in the standard deviation
between the users and devices were calculated. The standard deviation values obtained
for both three user study and three device variation study was within the accepted range
of =<1.5 Ct.

Interference:
The purpose of this study is to determine the effect of potentially interfering substances
on the Truenat™ HPV-HR assay. For this study medium load samples were used. To the
samples different concentrations of blood ranging from 5%, 10% and 30% were spiked
and then the samples were subjected to extraction on TrueprepTM AUTO Sample Prep
Device. DNA was eluted and PCR was performed on TruelabTM Uno Dx real time micro
PCR analyzer using Truenat™ HPV-HR chips. The presence of blood till 10% did not
interfere with the performance of Truenat™HPV-HR assay. The standard deviation
values obtained were within the accepted range of =<1.5 Ct for Truenat™ HPV-HR
assay.

Accuracy of TruenatTM HPV-HR assay:


Accuracy was determined by performing DNA extractions and Truenat™ HPV-HR PCR
for varying titres of samples over 5 consecutive days. The standard deviation values
obtained were within the accepted range of =<1.5 Ct.

Precision of TruenatTM HPV-HR assay:


Precision was tested by performing Truenat™ HPV-HR assay of High, Medium and Low
titre DNA for five consecutive days. Every day PCR for each titre DNA was run in
duplicates. The standard deviation values obtained were within the accepted range of
=<1.5 Ct.

22. REFERENCES
1. http://www.who.int/mediacentre/factsheets/fs380/en/
2. Sherman, Mark E., et al. (1994).Toward objective quality assurance in cervical
cytopathology: correlation of cytopathologic diagnoses with detection of high-risk human
papillomavirus types. American Journal of Clinical Pathology, 102.2 : 182-187.
3. Remmink, Ans J., et al. (1995).The presence of persistent high-risk hpv genotypes in
dysplastic cervical lesions is associated with progressive disease: Natural history up to
36 months. International Journal of Cancer, 61.3: 306-311.
4. Walboomers, J. M. M., et al. (1999).Human Papillomavirus is a necessary cause of
invasive cervical cancer worldwide. Journal of Pathology, 189:12-19.
5. Molijn, Anco, et al. (2005). Molecular diagnosis of human papillomavirus (HPV) infections.
Journal of Clinical Virology, 32:43-51.

SYMBOL KEYS
Consult In vitro Diagnostic Medical Temperature
For
Catalogue This Side
instructions
for use
IVD Device. Not for medicinal
use. Limitation REF Number 2 single
use only Up
Manufacturer

Date of Date of Batch Number Contains sufcient Authorised Representative in


Expiry LOT / Lot Number
Caution EC REP the European Community
Manufacture for <n> tests

Molbio Diagnostics Pvt. Ltd.

H. No. 13, Sagar Society, Dona Paula, Plot No. L-46, Phase II D,
Panaji, North Goa, Goa - 403004, INDIA Verna Industrial Estate, Verna,
www.molbiodiagnostics.com Goa - 403 722, INDIA
TNHCV/0718/V-01

Email: sales@molbiodiagnostics.com
EC REP
Qarad b.v.b.a. Cipalstraat 3, B-2440 Geel, Belgium

S-ar putea să vă placă și